PEAK 1: Area and Migration Time Peak 2: Area and Migration Time

Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 7

PEAK 1: Area and Migration Time Peak 2: Area and Migration time

Area Migration Time Area Migration Time


Rep1 466581 3.820 114461 3.600000
Rep2 467498 3.800 114863 3.591667
Rep3 464999 3.790 115842 3.587500
Rep4 468469 3.780 116024 3.587500
Rep5 467325 3.760 115546 3.587500
Rep6 469863 3.740 117509 3.579167

467456 3.78166666666667 AVERAGE 115708 3.588889


1652 0.02857738033247 STANDARD DEVIATION 1064 0.006804
0.0035335 0.00755682159519 RSD 0.009193 0.001896

0.353% 0.756% %RSD 0.919% 0.190%


on time
Voltage

3.0 KV 15.0 KV 30.0 KV 3.0 KV


Measurement 1 2.99 15.1 30.3 3.1
Measurement 2 2.97 15.2 30.3 3.09
Measurement 3 2.98 15.2 30.3 3.09
Measurement 4 2.99 15.2 30.3 3.07
Measurement 5 2.99 15.2 30.3 3.07
Measurement 6 2.99 15.2 30.3 3.08

AVERAGE 2.99 15 30 3
STANDARD DEVIATION 0.008367 0 0 0
RSD 0.002803 0.002689 0.000000 0.003928

%RSD 0.28% 0.27% 0.00% 0.39%


oltage

15.0 KV 30.0 KV
15.2 30.3
15.2 30.3
15.2 30.3
15.2 30.3
15.2 30.3
15.2 30.3

15 30
0 0
0.000000 0.000000

0.00% 0.00%
Capillary and Sample Storage temperature calibration

Check temperature
Capillary Temperature
TD 15 25 50
TM 14.6 51.48

New coefficients
A= 1.05
B= -1.21

Sample Storage Temperature


D
T 7 25 60
TM 3.5 62.2

New coefficients
A= 1.11
B= -4.25 -2.54 -6.79

Note: replace coefficient A, and add on coefficient B to existing value.

Input in EEPROM

Capillary Cooler [30] & [31]


Sample Storage [32] & [33]

EETEMPUPDATE

Then Test temperatures again


Capillary Temperature Stability
0 minutes 24.9
2 minutes 25.0
4 minutes 24.9
6 minutes 25.0
8 minutes 25.1
10 minutes 25.1
Mean= 25.0

Sample Cooler Temperature Stability


0 minutes 24.6
2 minutes 24.6
4 minutes 24.7
6 minutes 24.7
8 minutes 24.7
10 minutes 24.7
Mean= 24.7
Forward Pressure Test
0.50 1.00 7.50 20.0 50.0 80.0
Measurement 1
Measurement 2
Measurement 3
Mean #DIV/0! #DIV/0! #DIV/0! #DIV/0! #DIV/0! #DIV/0!

Reverse Pressure Test


0.50 1.00 7.50 20.0 50.0 80.0
Measurement 1
Measurement 2
Measurement 3
Mean #DIV/0! #DIV/0! #DIV/0! #DIV/0! #DIV/0! #DIV/0!

Vacuum Test
-0.50 -1.00 -5.00
Measurement 1
Measurement 2
Measurement 3
Mean #DIV/0! #DIV/0! #DIV/0!

Low Pressure Calibration


Note: Attatch Pressure Vial and Meter to calibrate

Debug 0_1

clop *will apply 7.0 psi

inlo XX.x *should store to eeprom

High Pressure Calibration


Debug 0_1

chip *will apply 70 psi

inhi XX.x