Manual Capnografo EMMA
Manual Capnografo EMMA
Revision history
Edition Date Description
2 (33)
Contents
1 INTENDED USE ................................................................................................................................ 10
2 SAFETY INFORMATION .................................................................................................................. 11
2.1 WARNINGS ................................................................................................................................... 11
2.2 CAUTIONS .................................................................................................................................... 14
2.3 SYMBOL DESCRIPTION .................................................................................................................. 16
3 DEVICE DESCRIPTION.................................................................................................................... 17
3.1 EMMA EMERGENCY CAPNOMETER OVERVIEW ............................................................................... 17
3.2 PRINCIPLE OF OPERATION ............................................................................................................. 20
3.2.1 EMMA Airway Adapter ........................................................................................................ 23
4 PREPARATION FOR USE................................................................................................................ 25
4.1 SETTING UP .................................................................................................................................. 25
4.2 STARTING UP ................................................................................................................................ 27
4.3 SWITCHING OFF ............................................................................................................................ 29
4.4 CONNECTING THE EMMA EMERGENCY CAPNOMETER TO A BREATHING CIRCUIT .............................. 30
5 USER INTERFACE ........................................................................................................................... 32
5.1 OVERVIEW .................................................................................................................................... 32
5.2 CONTROLS ................................................................................................................................... 34
5.3 MONITORING ................................................................................................................................ 34
5.3.1 RESPIRATORY RATE Display ........................................................................................... 35
5.3.2 ETCO2 Display.................................................................................................................... 35
5.3.3 CO2 BAR GRAPH............................................................................................................... 35
5.4 INDICATORS AND ALARMS.............................................................................................................. 37
5.4.1 Default Limits....................................................................................................................... 37
5.4.2 Battery Status Indicator ....................................................................................................... 38
5.4.3 Alarm Status Indicator (EMMA Monitor only)...................................................................... 39
5.4.4 Alarm Silence (EMMA Monitor only) ................................................................................... 40
5.4.5 Display Indications .............................................................................................................. 41
5.4.6 Adjusting the ETCO2 Alarm Limits (EMMA Monitor only)................................................... 43
5.4.7 Test of Indicators and Alarms ............................................................................................. 46
6 EMMA EMERGENCY CAPNOMETER AND ACCESSORIES......................................................... 47
7 MAINTENANCE AND SERVICE ...................................................................................................... 49
7.1 BATTERY REPLACEMENT ............................................................................................................... 49
7.2 CLEANING .................................................................................................................................... 49
7.3 EMMA AIRWAY ADAPTER ............................................................................................................. 50
7.4 ZERO CALIBRATION ....................................................................................................................... 50
7.5 RECYCLING .................................................................................................................................. 51
7.6 GAS SPAN CHECK ........................................................................................................................ 51
7.7 SERVICE AND PRODUCT RETURN REQUIREMENTS .......................................................................... 52
8 TECHNICAL SPECIFICATIONS....................................................................................................... 53
8.1 GENERAL SPECIFICATIONS ............................................................................................................ 53
8.2 ELECTROMAGNETIC COMPATIBILITY (EMC).................................................................................... 59
8.3 COMPLIANCE ................................................................................................................................ 70
8.4 CLASSIFICATIONS ......................................................................................................................... 70
3 (33)
1 Intended use
The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide
concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in
the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency
transport settings for adult and pediatric patients.
The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and
respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite,
intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult
and pediatric patients.
4 (33)
2 Safety Information
2.1 Warnings
Adhere to the following warnings for safe operation of EMMA Emergency Capnometer.
WARNING! EMMA Emergency Capnometer should only be used for the purpose and in the
manner described in this manual.
WARNING! For clinical applications EMMA Emergency Capnometer is intended for use only by
authorized and trained health care professionals.
WARNING! EMMA Emergency Capnometer must not be used with flammable anesthetic
agents.
WARNING! EMMA Airway Adapters shall not be reused. Used airway adapters shall be
disposed of in accordance with local regulations for medical waste.
WARNING! The user must be aware that the EMMA Airway Adapter will add 6 ml dead space
to the breathing circuit.
WARNING! Remove batteries if EMMA Emergency Capnometer is not likely to be used for a
period of time greater than 90 days.
WARNING! If EMMA Emergency Capnometer is used with a respirator or with harmful gases
such as N2O, always perform a pre-use tightness check of the patient circuit.
5 (33)
WARNING! To prevent “rain-out” and moisture from draining into the EMMA Airway Adapter,
place the adapter in the vertical position as shown in the picture below.
WARNING! Do not use EMMA Emergency Capnometer with nebulized medications as this may
affect the light transmission of the EMMA Airway Adapter windows.
WARNING! Audible alarm volume of any monitor may not be heard in some loud environments,
such as when sirens are in use and the care provider is more distant from the alarm source.
Alarm volume should be tested with the extremes of your noise environment to confirm ability or
limitations to hear an alarm in all circumstances of the environment.
WARNING! The user must be aware that light transmission can be affected by water rain-
out/condensation in the airway adapter if a heated humidifier is used. Dry or change airway
adapter if necessary.
2.2 Cautions
CAUTION! If the EMMA Emergency Capnometer is used in a manner other than that for which it
was intended, unpredictable behavior could result.
CAUTION! The EMMA Airway Adapters are non-sterile devices. Do not autoclave the devices
as this will damage them.
CAUTION! Do not operate EMMA Emergency Capnometer at ambient temperatures less than
23°F (-5°C) or greater than 122 °F (50 °C).
CAUTION! Federal law restricts this device to sale by or on the order of a physician.
CAUTION! The use of EMMA Emergency Capnometer is restricted to one patient at a time.
6 (33)
2.3 Symbol Description
Catalog number.
7 (33)
3 Device Description
3.1 EMMA Emergency Capnometer overview
The EMMA Emergency Capnometer is a quantitative carbon dioxide mainstream monitor comprised of a
sensor body that snaps in place on top of a disposable EMMA Airway Adapter. The EMMA Emergency
Capnometer is available in two different models: EMMA Analyzer with measurement and display
functionality and EMMA Monitor with measurement, display and alarm functionality.
Battery Status
Indicator
Power On
button
ETCO2 Display
Respiratory Rate
Display EMMA Sensor Body
Note: A figure of the EMMA Analyzer is shown in Section 5.1, Figure 9a.
8 (33)
3.2 Principle of operation
The measurement of CO2 in the breathing gas mixture is based on the fact that different gas
components absorb infrared light at specific wavelengths. A beam of invisible infrared light is directed
through the respiratory gas flow in the EMMA Airway Adapter. As the beam passes through the airway
adapter, some of the light is absorbed by the gas mixture. The amount of absorbed light is measured by
a miniaturized two channel spectrometer positioned to receive the infrared light beam.
The spectrometer incorporates a filter wheel fitted with two different optical "color" filters. The
wavelength ranges of these filters are chosen such that one filters out colors where carbon dioxide has
very strong absorption and the other filters out colors where carbon dioxide has no absorption.
The spectrometer also incorporates an infrared detector that converts the light beam to an electrical
signal. The electrical signal is converted to a digital value that is fed to a microprocessor. The ratio of the
light measured through the two filters is then used by the microprocessor to calculate the carbon dioxide
concentration in the breathing gas mixture.
Spectrometer with
optical filter wheel and
infrared detector
Infrared light source
9 (33)
3.2.1 EMMA Airway Adapter
EMMA Emergency Capnometer snaps in place on top of the EMMA Airway Adapter. The airway adapter
may, for example, be inserted between the endotracheal tube and the resuscitation bag or between the
resuscitation bag and the patient mask. Respiratory gas measurements are, as described in the previous
section, obtained by continuously measuring the infrared light absorption through the airway adapter.
The EMMA Airway Adapter is therefore fitted with optical XTP™ windows that are transparent to light in
the wavelength ranges of interest.
XTP™ window
10 (33)
4 Preparation for Use
4.1 Setting up
Unpack and inspect the EMMA Emergency Capnometer for external damage.
1. Press the battery cover release button into the EMMA Emergency Capnometer body until the
battery cover pops off.
2. Open the battery compartment and insert two (2) AAA batteries. Make sure the batteries are
fitted according to the indicated polarity. After battery installation, snap battery cover back into
place.
11 (33)
4.2 Starting up
Attach the EMMA Airway Adapter to the EMMA Emergency Capnometer. The method to attach the
EMMA Airway Adapter to the EMMA Emergency Capnometer Sensor Body is intuitive upon visual
inspection. The EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body are designed
so that they may only be connected in the proper orientation. Physical barriers prevent attachment in an
improper orientation. The EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body will
remain attached only if placed in the proper orientation and snapped into place, using light finger-
pressure force. If the EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body are not
properly fitted together they will not remain attached. It is also possible to first position the Emma Airway
Adapter in the breathing circuit and then fit the EMMA Emergency Capnometer to the Adapter. Check
that it is properly fitted and press the power on button. A start-up sequence will be initiated to check the
proper functioning of all display segments.
When the EMMA Emergency Capnometer is ready the endtidal CO2 (ETCO2) display indicates 0, the
Respiratory Display indicates “- -“, the first segment of the bar graph and the green battery status
indicator are lit.
If the ETCO2 display shows a non-zero value, ensure that there has not been an accumulation of CO2
between the EMMA Emergency Capnometer and the EMMA Airway Adapter by removing and
reattaching the EMMA Airway Adapter. If the ETCO2 still displays a non-zero value after this procedure,
perform a Zero Calibration as described in chapter 7.4 prior to using the EMMA Emergency Capnometer
with a patient.
12 (33)
4.4 Connecting the EMMA Emergency Capnometer to a breathing
circuit
The EMMA Emergency Capnometer can be connected to a patient circuit in different ways. The
following pictures illustrate the most common methods of connection.
Figure 7. EMMA Emergency Capnometer connected between an endotracheal tube and a resuscitation
bag
13 (33)
5 User Interface
5.1 Overview
Alarm Indicator
(EMMA Monitor only)
14 (33)
5.2 Controls
EMMA Emergency Capnometer has one Power On push button. On the EMMA Monitor, the Power On
push button may also be used for adjusting the lower alarm limit for ETCO2 and the Alarm Silence
Button may be used for adjusting the upper alarm limit for ETCO2.
EMMA EMMA
Control Description
Analyzer Monitor
Power On Turn on the device. √ √
Alarm Silence Silence active alarms for 2 minutes. √
Allow the user to change the upper high and
Upper ETCO2 Limit Adjustment low ETCO2 limit. √
5.3 Monitoring
The EMMA Emergency Capnometer has three displays: the ETCO2 display, the Respiratory Rate
display and the CO2 Bar Graph.
15 (33)
5.4 Indicators and Alarms
The EMMA Analyzer is equipped with an indicator system according to the table below. The EMMA
Monitor is equipped with an Alarm Status Indicator and an audible alarm that may be silenced for a
period of 2 minutes. The EMMA Monitor Status Indicator and audible alarm operate according to the
table below.
Low High
Respiratory Rate (RR) 3 bpm -
ETCO2 (EMMA Monitor) OFF 50 mmHg (7.0 kPa)
ETCO2 values are displayed after one breath and have a continually updated breath average.
RR is displayed after two breaths and the averaged value is updated every breath.
16 (33)
5.4.3 Alarm Status Indicator (EMMA Monitor only)
The Alarm Status Indicator is normally not lit. When an alarm is present it is lit with a steady yellow light.
When the alarms No Breath or ETCO2 High/Low has been present for 40 seconds the Alarm Status
Indicator will start to blink.
Caution Blinking Yellow 3 tone beep and then silent for 24 seconds
If an alarm is active, pressing the Alarm Silence button will silence the
alarm for a period of 2 minutes.
The Alarm Silence status is indicated by the upper segment of the bar
graph being lit.
If a No Breath alarm is generated, the alarm can be acknowledged and
turned off by pressing the alarm silence button. The No Breath alarm
should only be turned off when a patient is not connected to the EMMA
Emergency Capnometer.
17 (33)
5.4.5 Display Indications
ETCO2 Display Meanings Suggested action(s)
--
Check Adapter Check that the EMMA Airway
Adapter is not occluded.
(Blinking) Check that the EMMA Airway
No Adapter
Adapter is correctly attached.
--
No Breath Check patient
(Blinking)
Check Adapter Check that the EMMA Airway
Adapter is not occluded.
18 (33)
5.4.6 Adjusting the ETCO2 Alarm Limits (EMMA Monitor only)
5.4.6.1 Adjusting the High ETCO2 Alarm Limit
Press and hold the Alarm Silence button until the RR display shows “Hi” and the ETCO2 display shows
the current high ETCO2 alarm limit. Release the button. To adjust the alarm limit: press the Alarm
Silence button (▲) to increase, or the Power On button (▼) decrease the value. It is possible to switch
off the high ETCO2 alarm by adjusting the limit above 99 mmHg (9.9 kPa). EMMA Monitor will indicate
this setting by showing "- -" on the ETCO2 display during the adjustment routine.
If no button has been activated for a short period of time, EMMA Monitor will automatically resume
normal operation.
Figure 11. Adjusting the High and Low ETCO2 alarm limits
Low High
ETCO2 displayed in mmHg OFF; 1 – 89 mmHg 11 – 99 mmHg; OFF
ETCO2 displayed in kPa OFF; 1 – 8.9 kPa 1.1 – 9.9 kPa; OFF
If the high ETCO2 limit is decreased close to the low ETCO2 limit, the low limit will be automatically
adjusted in order to maintain a minimum difference of 10 mmHg (1.0 kPa) between the high and low
alarm limit. Similarly, if the low ETCO2 limit is increased close to the high ETCO2 limit, the high limit will
be automatically adjusted to maintain a minimum difference of 10 mmHg (1.0 kPa) between the high and
low alarm limit.
19 (33)
5.4.7 Test of Indicators and Alarms
To test the visible status indicator and the audible alarm (EMMA Monitor only) perform the following
steps.
This will start the Indicator on EMMA Analyzer and the Alarm on EMMA Monitor.
When the test is completed, attach the Airway Adapter to the Sensor Body as described in chapter 4.2.
The EMMA Emergency Capnometer is ready for use.
20 (33)
6 EMMA Emergency Capnometer and Accessories
Below is a list of device models, versions and approved accessories. For an up-to date list of
accessories visit www.phasein.se
21 (33)
7 Maintenance and Service
7.1 Battery Replacement
The EMMA Emergency Capnometer is equipped with a green Battery Status indicator. The Battery
Status indicator starts blinking when the remaining lifetime of the batteries is approximately 30 min.
Insert two new AAA type batteries into the battery compartment.
NOTE: Make sure that the batteries are fitted according to the polarity marking!
When the batteries are properly in place, gently snap the battery cover back into place.
7.2 Cleaning
• Remove the batteries before cleaning.
• EMMA Emergency Capnometer can be cleaned using a cloth moistened with isopropyl alcohol.
• Treat used EMMA Airway Adapters in accordance with hospital protocol for disposable items.
EMMA Monitor: Press and hold down simultaneously the ON and Alarm Silence button until the
ETCO2 display shows “10” and the RR displays “C0”. Keep both buttons depressed while the
ETCO2 Display starts "counting down" i.e. displaying "9" - "8" - "7" etc and until "0" is displayed.
• Once “0” is displayed on the ETCO2 display, zero calibration of the EMMA Emergency
Capnometer is completed.
Gas readings should be verified with a reference instrument at regular intervals. A zero reference
calibration of the IR measurement should be performed whenever an offset in gas readings is
discovered.
22 (33)
7.5 Recycling
The EMMA Emergency Capnometer and the EMMA Airway Adapter shall be disposed in accordance to
local regulations for recyclable products. Dispose the old batteries according to local regulations.
23 (33)
8 Technical Specifications
8.1 General Specifications
General EMMA Emergency Capnometer
Description Compact, battery powered, quantitative capnometer for mainstream
CO2 monitoring of adults and pediatrics.
Measurements Non-dispersive IR absorption
Models EMMA Analyzer (non-alarm)
EMMA Monitor (full-alarm)
Versions CO2 displayed in kPa or mmHg
Warm up In operation and full accuracy within 5 sec.
Calibration No routine calibration required
Certifications CE marked per 93/42/EEC, FDA 510(k) and UL/CSA 60601-1
Dimensions 52 x 39 x 39 mm (2.1 x 1.5 x 1.5 inches)
Weight Approximately 60 g (2.1 oz) with batteries
Shock absorption Withstands repeated 1 m drops
Environment
Operating conditions Temperature: -5 - +50°C (23 to 122°F)
Humidity: 10 - 95% RH (non-condensing)
Atmospheric pressure: 70 - 120 kPa (1) (2)
Storage conditions Temperature: -30 - +70°C (-22 to 158°F)
Humidity: 5 - 100% RH (condensing)
Atmospheric pressure: 50 - 120 kPa
Displays
ETCO2 LED Numeric Display
Respiratory rate LED Numeric Display
CO2 Bar Graph 14 segment LED bar graph
Controls
Power ON key
Alarm Silence 2 min. alarm silence key (EMMA Monitor only)
ETCO2 Up/Down key for setting alarm limits (EMMA Monitor only)
CO2
Range 0 - 99 mmHg (0 - 9.9 kPa) (3)
Accuracy 0-40 mmHg ± 2 mmHg, 41-99 mmHg 6% of reading, during standard
conditions.(4) (5)
Rise time ≤ 60 ms
Total system response time < 0.5 sec.
24 (33)
General EMMA Emergency Capnometer
Respiratory Rate (RR)
Respiratory rate 3 - 150 breaths/min
Accuracy ± 1 bpm
Breath detect Adaptive threshold, minimum 1 kPa CO2 change
EMMA Airway Adapters
Adult 6 ml dead space
Indicators and Alarms
EMMA Analyzer Indicators for: No Adapter, Check Adapter, No Breath Detected, Low
Battery
EMMA Monitor Alarms for: No Adapter, Check Adapter, No Breath Detected, Low
Battery, Low ETCO2, High ETCO2
Sound Intensity Level ≥ 80 dBA (EMMA Monitor only)
Power requirements
Batteries Two (2) AAA Cell Alkaline Batteries (2 x 1.5VDC) (IEC Type LR03)
Battery life time 8 hours of normal use
Power status LED indicator
Notes:
(1)
EMMA Emergency Capnometer displays CO2 in partial pressure units (kPa or mmHg) and thus does
not require barometric pressure compensation.
(2)
High altitude measurements are affected by pressure broadening. To compensate for this effect, add
offset to displayed ETCO2 value according to the following table based on altitude and range of ETCO2
being displayed by the EMMA Emergency Capnometer:
ETCO2 Offset
Based on Range Displayed
Altitude 5-40 mmHg / 0.6-5.3 kPa 40 + mmHg / 5.3 + kPa
5 000 ft / 1 500 m + 1 mmHg / 0.1 kPa + 2 mmHg / 0.3 kPa
6 000 ft / 1 800 m + 2 mmHg / 0.3 kPa + 4 mmHg / 0.5 kPa
7 000 ft / 2 100 m + 3 mmHg / 0.4 kPa + 6 mmHg / 0.8 kPa
8 000 ft / 2 450 m + 4 mmHg / 0.5 kPa + 8 mmHg / 1.1 kPa
9 000 ft / 2 750 m + 5 mmHg / 0.7 kPa + 10 mmHg / 1.3 kPa
10 000 ft / 3 050 m + 6 mmHg / 0.8 kPa + 12 mmHg / 1.6 kPa
(3)
Gas reading showing actual partial pressure at current humidity level.
(4)
To include quantitative effect on gas reading from variations in environment conditions and presence
of N2O, anesthetic agents, Ethanol, Isopropyl alcohol, He, Acetone and Methane, then the CO2
accuracy range should be increased to ± 4 mmHg or 10% of reading whichever is the greater.
(5)
CO2 was tested at a RR of 40. As RR rates increases above 60 the accuracy range will also increase.
25 (33)
8.2 Electromagnetic Compatibility (EMC)
The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in
such an environment.
26 (33)
Guidance and PHASEIN’s declaration – electromagnetic immunity –
for EMMA Emergency Capnometer
The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in
such an environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
27 (33)
Guidance and PHASEIN’s declaration – electromagnetic immunity –
for EMMA Emergency Capnometer
The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in
such an environment.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the EMMA Emergency Capnometer is used exceeds the applicable RF compliance
level above, the EMMA Emergency Capnometer should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the EMMA Emergency Capnometer.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 20 V/m.
28 (33)
Recommended separation distances between portable and mobile RF communications
equipment and the EMMA Emergency Capnometer
The EMMA Emergency Capnometer is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the EMMA Emergency
Capnometer can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the EMMA Emergency
Capnometer as recommended below, according to the maximum output power of the communications
equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
29 (33)
8.3 Compliance
MDD 93/42/EEC
ISO 21647:2004
EN/IEC 60601-1:1990, Amendment1 (1991), Amendment2 (1995)
EN/IEC 60601-1-2:2001
ISO 5356-1
EN/IEC 60601-1-8:2003
prEN 1789:2004
8.4 Classifications
According to the type of protection against electric shock
INTERNALLY POWERED EQUIPMENT (Battery power)
According to sterility
No part of the EMMA Emergency Capnometer is sterile.
30 (33)
Notes:
31 (33)
Notes:
32 (33)
PHASEIN AB
Svärdvägen 15
182 33 Danderyd
Sweden
www.phasein.se
33 (33)