Capitulo 3 EPA
Capitulo 3 EPA
Capitulo 3 EPA
TABLE OF CONTENTS
Section Page
3.0 INTRODUCTION 1
3.1 DEFINITIONS 1
3.2 SAFETY 4
3.3 SAMPLING CONSIDERATIONS 4
3.4 SPECIAL CONSIDERATIONS FOR DETERMINING INORGANIC ANALYTES AT 6
ULTRA-TRACE CONCENTRATION LEVELS
3.5 REAGENT PURITY 11
3.6 SAMPLE DIGESTION METHODS 11
3.7 METHODS FOR DETERMINATION OF INORGANIC ANALYTES 13
3.8 REFERENCES FOR PREVIOUS SECTIONS AND THE TABLES AND FIGURES 16
TABLE
3-1 MATERIALS FOR USE IN SAMPLE COLLECTION FOR INORGANIC ANALYTE 18
DETERMINATIONS
3-2 RECOMMENDED SAMPLE HOLDING TIMES, PRESERVATION, COLLECTION 19
QUANTITIES, AND DIGESTION VOLUMES FOR SELECTED INORGANIC ANALYTE
DETERMINATIONS IN AQUEOUS AND SOLID SAMPLES
3-3 EXAMPLES OF THE ANALYTICAL BLANK INFLUENCE ON ULTRA-TRACE 21
ANALYSIS OF ELEMENTS IN GLASS
3-4 EXAMPLES OF LEAD CONCENTRATIONS IN AIR 21
3-5 CLEANLINESS LEVELS IN INTERNATIONAL STANDARD ISO 14644-1 22
3-6 PARTICULATE CONCENTRATIONS IN LABORATORY AIR 23
3-7 NON-CONTAMINATING MATERIALS AND FOR USE AS DIGESTION VESSELS AND 23
SAMPLE CONTAINERS IN ULTRA-TRACE ANALYSIS
FIGURE
3-1 COMPARISON OF CLEAN VERSUS CONVENTIONAL SAMPLING AND ANALYSIS 24
TECHNIQUES USED IN THE ANALYSIS OF SOUTH TEXAS ESTUARY WATERS
3-2 COMPARISON OF PARTICLE COUNT ANALYSIS OF A CLEAN ROOM AND A 25
STANDARD LABORATORY AT DUQUESNE UNIVERSITY IN PITTSBURGH, PA
3-3 PARTICLE SIZE COMPARISON CHART FOR COMMON PARTICULATES 26
INORGANIC ANALYTES
Prior to employing the methods in this chapter, analysts are advised to consult the
disclaimer statement at the front of this manual and the information in Chapter Two for guidance
on the allowed flexibility in the choice of apparatus, reagents, and supplies. In addition, unless
specified in a regulation, the use of SW-846 methods is not mandatory in response to Federal
testing requirements. The information contained in each procedure is provided by EPA as
guidance to be used by the analyst and the regulated community in making judgments
necessary to meet the data quality objectives or needs for the intended use of the data.
For a summary of changes in Chapter Three, please see Appendix A at the end of this
document.
3.0 INTRODUCTION
This chapter provides guidance for the analysis of inorganic analytes in a variety of
matrices. The analytical methods are written as specific steps in the overall analysis scheme --
sample handling and preservation, sample digestion or preparation, and sample analysis for
specific inorganic components. From these methods, the analyst should assemble a total
analytical protocol which is appropriate for the sample to be analyzed and for the information
required. This introduction discusses the options available in general terms, provides
background information on the analytical techniques, and highlights some of the considerations
to be made when selecting a total analysis protocol.
3.1 DEFINITIONS
The following terms are relevant for the determination of inorganic analytes:
Instrument detection limit (IDL): Typically used in metals analysis to evaluate the
instrument noise level and response changes over time for analytes of
interest. IDLs in µg/L can be estimated as the mean of the blank results plus three
times the standard deviation of 10 replicate analyses of the reagent blank solution.
(Use zero for the mean if the mean is negative). Each measurement should be
performed as though it were a separate calibration standard (i.e., each
measurement must be followed by a rinse and/or any other procedure normally
performed between the analysis of separate samples). IDLs should be determined
at least once using new equipment, after major instrument maintenance such as
changing the detector, and/or at a frequency designated by the project. An
instrument log book should be kept with the dates and information pertaining to
each IDL performed.
Laboratory control sample (LCS): A volume of reagent water spiked with known
concentrations of analytes and carried through the sample preparation and
determinative procedure. It is used to monitor laboratory performance on analyte
loss/recovery in a clean matrix. The LCS should be prepared from the same
source as the calibration standards to remove potential error contribution from
standards of different sources. An independently prepared LCS may also be
obtained as or prepared from a certified reference solution or prepared from a
certified reagent solid or from an alternate lot reagent solid relative to the
calibration standards source if, for each analytical batch, at least one LCS is
prepared from the same source as the calibration standards. In this way, if the
recoveries of both the LCS and the matrix spike are outside the acceptance limits,
the analyst will be able to determine whether the problem is due to calibration error
or matrix interference.
NOTE: Other standards exist that have alternative methods for determining the
linear range (e.g., ISO 17025). Therefore, the method used to define and verify the
linear range should meet the requirements of the project.
Method blank: A volume of reagent water equal to that used for aqueous samples,
or, otherwise, a clean, empty container, equivalent to that used for actual solid
samples, processed through each sample preparation and determinative
procedure. Analysis of a method blank is used to assess contamination from the
laboratory environment, sample processing equipment, and/or reagents.
Sample holding time: The storage time allowed between sample collection and
sample analysis when the designated preservation and storage techniques are
employed. Different times may be specified for holding field samples prior to
extraction, digestion, or other such preparation procedures versus holding
prepared samples (e.g., an extract or a digestate) prior to analysis.
NOTE: Methods 3010 and 3020 state they are used to determine “total metals”.
However, because these methods do not use hydrofluoric acid, they are unable to
break down silicates and have been defined here for the “total acid
soluble/recoverable metals” fraction.
3.2 SAFETY
The methods in this chapter do not address all safety issues associated with their use.
The laboratory is responsible for maintaining a safe work environment and a current awareness
file of OSHA regulations regarding the safe handling of the chemicals specified in this method.
A reference file of material safety data sheets (MSDSs) should be available to all personnel
involved in these analyses.
The toxicity or carcinogenicity of each reagent used in these methods has not been
precisely defined. However, each chemical compound should be treated as a potential health
hazard. From this viewpoint, exposure to these chemicals should be reduced to the lowest
possible level by whatever means available. The following additional references to laboratory
safety are available:
4. "Occupational Safety and Health Standards," 29 CFR Part 1910, Occupational Safety
and Health Administration, Department of Labor.
The fundamental goal of all field sampling activities is to collect samples that are
representative of the water, soil or waste from which they were collected. Thus,
representative sampling may be considered to be the sampling analog to analytical
accuracy. Of equal importance is sampling precision for ensuring consistency both within
a single sampling event and between sampling events conducted over time. Sampling
imprecision can be a significant source of measurement error. High quality field practices
are, therefore, necessary for generating representative samples on a consistent basis.
Sampling quality assurance includes the development of a quality assurance plan, data
quality objectives and the generation of field quality control samples including equipment
rinsates, trip blanks and field duplicates. Regardless of the specific program needs, the
documentation of all relevant field and sample information is the final essential component
of a sampling event for providing evidence that proper procedures and quality assurance
were performed during sample collection. Use of inadequate field procedures and
documentation can jeopardize an entire sampling program despite adequate planning,
analytical facilities, and personnel.
While advances in analytical sensitivity are continuing to be made that allow for
quantification of environmental contaminants at ultra-trace levels (i.e., < 0.1 ppb), clean
sampling techniques are consequently being devised and practiced in order to minimize or
eliminate sources of contamination during the collection of samples intended for ultra-trace
contaminant testing. Such clean sampling and analysis techniques are not generally
needed or required under the RCRA program and are beyond the scope of this chapter.
However, as an introduction to this topic, Sec. 3.4 provides a more detailed discussion on
the special category and requirements of clean analysis for determining constituents at
ultra-trace levels.
• Phosphate-free detergent
• Tap water
• 1:1 HNO3
• Tap water
• 1:1 HCl
• Tap water
• Reagent water
NOTE: Chromic acid should not be used to clean glassware, especially if chromium is
to be included in the analytical scheme. Commercial, non-chromate products may be
used in place of chromic acid, if adequate cleaning is documented by an analytical quality
control program. Chromic acid should not be used with plastic bottles.
Recommended materials to use for sample collection are listed in Table 3-1.
Sample holding times, recommended collection volumes or masses and recommended
digestion volumes, and preservatives are listed in Table 3-2. The sample collection and
digestion amounts depend on the combination of digestion or extraction and determinative
procedures that will be employed for a given sample as well as the sensitivity and tolerable
sampling errors that are required for a specific project (see References 20 through 22).
Likewise, the use of alternative preservatives to those indicated in Table 3-2 may be
necessary depending on the objectives of the project. In all cases, the sample quantity
that is collected should be representative of the bulk material whenever feasible.
For all non-speciated digestion methods, great reduction in analytical variability can
be achieved through the use of appropriate sample preparation procedures. Generally, a
reduction in subsampling variance can be accomplished by increasing the subsample
aliquot size, reducing the sample particle size, and/or thoroughly mixing the resulting fines
(see References 20 through 23 for additional information on sampling and sub-sampling).
Under most circumstances, it is recommended that the sample be analyzed without drying.
If it is necessary to report the analytical data on a dry-weight basis, then a separate aliquot
may be analyzed for moisture content and the wet-weight data corrected accordingly.
If the sample cannot be well-mixed and homogenized in the form in which it was
received by the laboratory, then air- or oven-drying at 30 °C or less, crushing, sieving,
grinding, and mixing should be performed as needed or feasible to homogenize the
sample until the subsampling variance is less than the data quality objectives of the
analysis. While proper sample preparation generally produces great reduction in
analytical variability, it should be noted that in certain unusual circumstances there could
be loss of volatile metals (e.g., mercury, organometallics) or irreversible chemical changes
( e.g., precipitation of insoluble species, change in valence state) caused by inappropriate
sample preparation procedures.
The significant role of the analytical blank in chemical analysis of trace metals
cannot be overemphasized. Sensitive instrumentation such as ICP-MS, ICP-OES, and
GFAA requires that sample preparation be at least as sophisticated as the instruments
used for analysis. The analytical blank is normally a primary source of error in ultra-trace
element analysis. Ultra-trace analysis is as dependent on control of the analytical blank as
it is on the bias and precision of the instrument making the measurement. Inability to
control contamination is frequently the limiting factor in trace (parts per million (ppm) to
parts per billion (ppb)) and ultra-trace (ppb to parts per trillion (ppt)) analysis. Analytical
blank contributions occur from the following four major sources (References 3 through 7):
The four primary areas that affect the analytical blank can be demonstrated using
standard reference materials in analysis. Table 3-3 illustrates and isolates the main
influencing parameters on the blank: environment, reagents, materials, and analyst skills.
The skill of the analyst was kept constant as the same analyst changed the environment,
reagents, and combinations of these parameters in the analysis (see Reference 6). The
trace elements in glass (TEG) standard reference material from the National Institute of
Standards and Technology (NIST) was used to keep sample homogeneity constant and to
permit removal of the sampling error by using sample sizes in which appropriate
homogeneity had previously been demonstrated. It is important to note that the
relationship of the precision and measurement remained relatively constant. The
significance of the first two major sources of contamination, environment and reagents, is
evaluated. The contamination in the laboratory air and in the acid used for the reagent
blank altered the accuracy of the example above by over two orders of magnitude for both
lead and silver. The larger influence of the two sources in this example is the laboratory
environment in which the samples were prepared.
Any laboratory air that comes into contact with the sample may deposit
airborne contamination into the sample. The sample is especially vulnerable
when it is being digested with acid. When samples are being digested with
acid(s), contamination of the sample by particulates in the laboratory air can
cause results that are biased high.
The purity of the reagents used for acid digestion, leaching, and
extraction is extremely important to the overall level of the blank. Reagents have
very different purities, depending on their processing grade and purpose.
Frequently, the analyst should purchase special reagents, or purify lesser-grade
reagents prior to use, in order to minimize the analytical blank.
In the preparation of high purity reagents, there is only one significant and
practical choice for the method of purification, i.e., sub-boiling distillation
(References 9 through 11). Different from normal distillation, sub-boiling
distillation uses an infrared radiation source to heat the reagent to a temperature
just below the boiling point. This use prevents the formation of bubbles that rise
and burst at the surface of the liquid. Thus, the aerosolized solution particles are
left in solution and prevented from physically transporting contaminants
throughout the distillation apparatus. Sub-boiling distillation is a slower but very
reliable method of purifying all of the common mineral acids and many organic
reagents used in analytical methods. It relies exclusively on the vapor pressure
of the reagent, and contaminant, and can therefore be specifically optimized for
purification of the mineral acids if the object is to remove metal ions. Of all acids,
For ultra-trace analyses, only certain materials are preferred for use in the
construction of sample vessels and instrument components that come into
contact with the sample. Over the past two decades, materials identified as
being non-contaminating have become the top choices for bottles, beakers,
reaction vessels, storage containers, nebulizers, and instrument components for
trace and ultra-trace analysis. The materials are the same as those currently
being used in many digestion vessels, bomb liners, and microwave vessels. The
materials are characterized by being thermally durable, chemically resistant or
inert, non-contaminating, and possessing appropriate compression and tensile
strength. Table 3-7 lists, in order of preference, several types of, non-
contaminating materials that are chemically inert to most acid reactions. These
materials have been evaluated and tested extensively for their potential to
contaminate (References 4, 6, 7, 12, and 13).
With the exception of polyethylene, the materials listed in Table 3-7 are
those most commonly used for sample preparation vessels, both atmospheric
pressure vessels and closed vessel liners that come into contact with the sample.
These materials are the most stable to acid reactions (with the exception of
quartz and glass if hydrofluoric acid is used). Fluoropolymers are the most
common and were adapted from other chemical uses for application in pressure
systems. The fluoropolymers have the highest range of use temperatures for
most plastics, ranging from 270-300 °C. They are also chemically inert to the
majority of mineral acids and combinations thereof. Sulfuric acid has a boiling
point of approximately 330 °C and can damage all fluoropolymers by melting
them. Quartz and glass can safely contain sulfuric acid at these high
temperatures, but borosilicate glass is not appropriate for ultra-trace elemental
analysis (References 7 and 13). Glass actually forms a gel layer that hydrates
and leaches, transferring contaminants from the glass to the sample solution.
While these quantities may be considered minute, they would be detected in
blanks and samples undergoing ultra-trace analyses.
• The time that the sample spends in digestion, leaching, or extraction may be
reduced from hours to minutes, thus reducing the potential leaching of
contaminants from the container walls.
The purity of the reagents used for sample preservation, acid digestion, leaching,
extraction and analysis is extremely important for preventing or minimizing sample
contamination. Reagents have very different purities, depending on their processing grade and
purpose. Reagent grade, ACS grade or better are recommended for use with most SW-846
methods. Sample contamination introduced through sample preservation, handling, preparation
and analysis is assessed from the analysis of method blanks.
Many of the methods listed below employ HCl in the digestion process. Chlorine is an
interferant in ICP-MS analysis and its use in sample digestion is discouraged except when
absolutely necessary or when the instrument manufacturer has indicated that the use of HCl will
not adversely affect the equipment and accurate quantitation of the desired target analytes.
However, please note that recoveries of certain metals may be biased low when using nitric acid
only. Interference from chlorine may be minimized by use of collision or reaction cell
instrumentation.
Method 3005A: Acid Digestion of Waters for Total Recoverable or Dissolved Metals for
Analysis by FLAA or ICP Spectroscopy
This method may be used for the preparation of ground water and surface water samples
for total recoverable and dissolved metal determinations by FLAA, ICP-OES, or ICP-MS. The
unfiltered or filtered sample is heated with dilute HCl and HNO3 prior to metal determination.
Method 3010A: Acid Digestion of Aqueous Samples and Extracts for Total Metals for Analysis
by FLAA or ICP Spectroscopy
This method may be used for the preparation of waste samples for total recoverable metal
determinations by FLAA, ICP-OES, or ICP-MS. The samples are vigorously digested with nitric
acid followed by dilution with hydrochloric acid. The method is applicable to aqueous samples,
leachates, and mobility-procedure extracts.
Method 3015A: Microwave Assisted Acid Digestion of Aqueous Samples and Extracts
This method may be used for the preparation of aqueous samples, mobility-procedure
extracts, and wastes that contain suspended solids for total recoverable metal determinations
by FLAA, GFAA, ICP-OES, or ICP-MS. Nitric acid and hydrochloric acid are added to the
sample in a PTFE digestion vessel and heated in a microwave unit prior to metals
determination.
Method 3020A: Acid Digestion of Aqueous Samples and Extracts for Total Metals for Analysis
by GFAA Spectroscopy
This method may be used for the preparation of waste samples for total recoverable
metals determinations by GFAA or ICP-MS. The samples are vigorously digested with nitric
acid followed by dilution with nitric acid. The method is applicable to aqueous samples,
leachates, and mobility-procedure extracts.
Method 3031: Acid Digestion of Oils for Metals Analysis by Atomic Absorption or ICP
Spectrometry
This method may be used for the preparation of waste oils, oil sludges, tars, waxes,
paints, paint sludges and other viscous petroleum products for analysis by FLAA, GFAA, and
ICP-OES. The samples are vigorously digested with nitric acid, sulfuric acid, hydrochloric acid,
and potassium permanganate prior to analysis.
This method may be used for the preparation of waste samples for total recoverable
metals determinations by FLAA and ICP-OES, or GFAA and ICP-MS depending on the options
chosen. The samples are vigorously digested in nitric acid and hydrogen peroxide followed by
dilution with either nitric or hydrochloric acid. The method is applicable to soils, sludges, and
solid waste samples.
Method 3051A: Microwave Assisted Acid Digestion of Sediments, Sludges, Soils, and Oils
This method may be used for the preparation of sludges, sediments, soils and oils for total
recoverable metal determinations by FLAA, GFAA, ICP-OES or ICP-MS. Nitric acid and
hydrochloric acid are added to the representative sample in a fluorocarbon digestion vessel and
heated in a microwave unit prior to metals determination.
Method 3052: Microwave Assisted Acid Digestion of Siliceous and Organically Based Matrices
This method may be used for the preparation of siliceous and organically based matrices
including ash, biological tissue, oil, oil contaminated soil, sediment, sludge, and soil for total
metals analysis by FLAA, CVAA, GFAA, ICP-OES, and ICP-MS. Nitric acid and hydrofluoric
acid are added to a representative sample in a fluorocarbon digestion vessel and heated in a
microwave unit prior to analysis.
This method may be used for the preparation of soils, sludges, sediments and similar
waste materials for hexavalent chromium determination. The samples are digested and heated
to dissolve the Cr(VI) and stabilize it against reduction to Cr(III).
The purpose of this method is to define the proper analytical procedure for the validated in
vitro bioaccessibility (IVBA) assay for lead in soil, to describe the typical working range and
limits of the assay, quality assurance (QA), and to indicate potential interferences. At this time,
This section of the manual contains analytical techniques for trace inorganic analyte
determinations. Instrumental techniques include:
Each of these (except the individual hexavalent chromium analyses) is discussed briefly below.
Some advantages, disadvantages, and cautions for the analysis of wastes are provided.
GFAA replaces the flame with an electrically-heated graphite furnace. The furnace allows
for gradual heating of the sample aliquot in several stages. Thus, the processes of dissolution,
drying, decomposition of organic and inorganic molecules and salts, and formation of atoms,
which should occur in a flame or plasma in a few milliseconds may be allowed to occur over a
much longer time period and at controlled temperatures in the furnace. This allows an
experienced analyst to remove unwanted matrix components by using temperature
programming and/or matrix modifiers. The major advantage of this technique is that it affords
extremely low detection limits. It is the easiest to perform on relatively clean samples. Because
this technique is so sensitive, interferences can be a real problem; finding the optimum
combination of digestion, heating times and temperatures, and matrix modifiers can be a
challenge for complex matrices.
HGAA utilizes a chemical reduction to reduce and separate arsenic, selenium, or antimony
selectively from a sample digestate. The technique therefore has the advantage of being able
to isolate these elements from complex samples which may cause interferences for other
analytical procedures. Significant interferences have been reported when any of the following is
present: (1) easily reduced metals (Cu, Ag, and Hg); (2) high concentrations of transition metals
(>200 mg/L); (3) oxidizing agents (oxides of nitrogen) remaining following sample digestion.
CVAA uses chemical reduction to reduce mercury to its volatile elemental form. The
procedure is sensitive, though not as sensitive as CVAF, but is subject to interferences from
some volatile organics, chlorine, and sulfur compounds. CVAA may be more appropriate than
CVAF for samples with higher mercury concentrations.
CVAF uses chemical reduction to reduce mercury to its volatile elemental form. The
procedure is extremely sensitive, approximately one to two orders of magnitude more sensitive
than CVAA. CVAF may be more appropriate than CVAA for samples with low mercury
concentrations. Because of the extreme sensitivity of CVAF, contamination during sampling
and analysis is a major concern.
ASV is an electrochemical method that can be used for determining free dissolved metals
such as arsenic and mercury. The analyte of interest is reduced at the electrode surface during
a deposition step and oxidized from the electrode during a stripping step; the current measured
during the stripping step is proportional to the analyte concentration. ASV detects dissolved
divalent mercury species and cannot directly distinguish between organic and inorganic mercury
compounds. The sensitivity of this method for mercury is similar to CVAA. ASV detects
dissolved inorganic arsenic species and can distinguish between As(III) and As(V) by adjusting
the potential applied during the deposition step. The sensitivity of this method for arsenic is
about an order of magnitude more sensitive than ICP-MS, GFAA, or HGAA. This method is
subject to interferences from suspended solids, organics, and other metals. Samples can be
modified (e.g., digested, addition of chelating agent, etc.) to determine the concentration of
various species of the analyte of interest (e.g., total fraction, methylmercury, etc.)
XRF uses sealed radioisotope sources or other suitable X-ray source to irradiate samples
with X-rays. When a sample is irradiated with X-rays, the source X-rays may undergo either
scattering or absorption by sample atoms. This later process is known as the photoelectric
effect. When an atom absorbs the source X-rays, the incident radiation dislodges electrons
from the innermost shells of the atom, creating vacancies. The electron vacancies are filled by
electrons cascading in from outer electron shells. Electrons in outer shells have higher energy
states than inner shell electrons, and the outer shell electrons give off energy as they cascade
IC generally refers to the separation of ions through ion exchange chromatography. In this
technique, an aqueous sample is injected into a mobile solution that is carried into a
chromatography column. As the sample travels through the column, the sample analytes are
temporarily retained on the column, the stationary phase, via electrostatic forces. The
separated analytes are identified as they are released from the column based on their retention
time. Detection and quantification in IC are most commonly performed using conductivity
detection. IC is typically used for the determination of anionic analytes in waste samples.
3.8 REFERENCES FOR PREVIOUS SECTIONS AND THE TABLES AND FIGURES
1. Skoog, D. A., "Principles of Instrumental Analysis," 3rd ed, Saunders College Publishing,
Philadelphia, PA, 1985.
2. EPA, U.S., "Sampling Ambient Water for Trace Metals at EPA Water Quality Criteria
Levels," Method 1669, U.S. Environmental Protection Agency, Office of Water,
Washington, D.C., 1996.
3. Adeloju, S. B. and Bond, A. M., "Influence of Laboratory Environment on the Precision and
Accuracy of Trace Element Analysis," Anal. Chem., 1985, 57, 1728-1733.
5. Moody, J. R., "NBS Clean Laboratories for Trace Element Analysis," Anal. Chem., 1982,
54, 1358A-1376A.
6. Murphy, T., "National Bureau of Standards Special Publication 422 Accuracy in Trace
Analysis: Sampling, Sample Handling, and Analysis," National Bureau of Standards,
Gaithersburg, MD, 1976, p. 509-539.
7. Tolg, G. and Tschopel, P., "Determination of Trace Elements," VCH, New York, NY, 1994,
p. 1-38.
8. Zief, M. and Mitchell, J. W., "Chemical Analysis," Vol. 47, John Wiley & Sons, New York,
NY, 1976.
9. Kuehner, E. C., et al., "Production and Analysis of Special High-Purity Acids Purified by
Sub-Boiling Distillation," Anal. Chem., 1972, 44, 2050-2056.
10. Moody, J. R., "Purified Reagents for Trace Metal Analysis," Talanta, 1982, 29, 1003-1010.
11. Kuehner, E. C. and Freeman, D. H., "Purification of Inorganic and Organic Materials,"
Marcel Dekker, New York, NY, 1969, p. 297-306.
13. Moody, J. R. and Lindstrom, R. M., "Selection and Cleaning of Plastic Containers for
Storage of Trace Element Samples," Anal. Chem., 1977, 49, 2264-2267.
14. Prevatt, F. J., "Clean Chemistry for Trace Analysis," Environmental Testing and Analysis,
1995, 4, 24-27.
15. Skelly, E. M. and DiStefano, F. T., "Clean Room and Microwave Digestion Techniques:
Improvement in Detection Limits for Aluminum Determination by GF-AAS," Appl.
Spectrosc., 1988, 42, 1302-1306.
17. Maienthal, E. J., "U.S. National Bureau of Standards Technical Note 545," National
Bureau of Standards, Gaithersburg, MD, 1970, p. 53-54.
18. Boothe, P. N. and Nicolau, B.A. "The Coastal Bend Bays Project," 26th Annual
Conference on Analysis of Pollutants in the Environment, Chicago, IL 2003.
19. Miller, G. T., "Living in the Environment, Wadsworth Inc., Belmont, CA, 1994.
20. Gy, P., "Sampling for Analytical Purposes" Wiley, New York, NY, 1998.
21. Pitard F.F. “Pierre Gy’s Sampling Theory and Sampling Practice,” Second Edition, CRC
Press, Inc., Boca Raton, FL, 1993.
22. EPA, U.S., “Correct Sampling Using the Theories of Pierre Gy,” Office of Research and
Development, Las Vegas, NV, 1999.
23. EPA, U.S., “Guidance for Obtaining Representative Laboratory Analytical Subsamples
from Particulate Laboratory Samples.” Office of Research and Development, 2003.
a
These recommendations are intended as guidance only and not inclusive of all possible analytes and
materials. The selection of sample container should be made based on the nature of the sample, the
intended end use of the data and the project data quality objectives.
Minimum
Collection Digestion Holding
Analyte Matrix Fraction Volume/Mass Preservation1 Volume Time2
Metals Aqueous Total/total 600 mL HNO3 to pH<2 100 mL 6 months
(except Hg recoverable
and Cr6+)
Dissolved 600 mL Filter on site; 100 mL 6 months
HNO3 to pH<2
Minimum
Collection Digestion Holding
Analyte Matrix Fraction Volume/Mass Preservation1 Volume Time2
pH Aqueous 25 mL NA Analyze
immediately
Solid 20 g NA Analyze
immediately
Specific Aqueous 100 mL NA Analyze
Conductance immediately
Sulfate Aqueous 50 mL ≤6 °C 28 days
Fill sample
surface with
2N zinc
acetate until
moistened;
Store
headspace
free at ≤6 °C
Organic Aqueous 200 mL ≤6 °C 28 days
Carbon, Total store in dark
(TOC) HCl or H2SO4
to pH <2;
Solid 100 g 28 days
≤6 °C
a These recommendations are intended as guidance only. The selection of sample and digestion volumes/mass and
preservation and holding times should be made based on the nature of the sample, the intended end use of the data
and the data quality objectives.
b Additional sample quantities may need to be collected in order to allow for the preparation and analysis of QC
and/or manufacturer’s recommendations for standards. Alternative temperatures may be appropriate based on
demonstrated analyte stability within a matrix, provided the data quality objectives for a specific project are still
attainable.
TABLE 3-3
TABLE 3-4
ISO Class 1 10 2
NOTE: Uncertainties related to the measurement process require that concentration data with no more
than three significant figures be used in determining the classification level
© ISO. This material is reproduced from ISO 14644-1:1999 with permission of the American
National Standards Institute (ANSI) on behalf of the International Organization for
Standardization (ISO). No part of this material may be copied or reproduced in any form,
electronic retrieval system or otherwise or made available on the Internet, a public network, by
satellite or otherwise without the prior written consent of ANSI. Copies of this standard may be
purchased from ANSI, 25 West 43rd Street, New York, NY 10036, (212) 642-4900,
http://webstore.ansi.org.
Concentration (µg/m3)
ND = Not Detected
TABLE 3-7
Quartz - Synthetic
Quartz - Natural
Borosilicate Glass
40
30
Concentration (ppb)
280 fold
20
10 fold
10
11 fold
36 fold
0
Cd Cu Hg Pb Zn
Element
1. The entire document was formatted in Microsoft Word .docx format from the original
.wpd and .pdf files.
2. The revision number was changed to five and the date published to July 2014.
3. Minor editorial and grammatical changes were made throughout the Chapter.
4. Method 3200 was added to Section 3.6.
5. Method numbers were revised to reflect the assigned number for Update V.
6. Definition for instrument detection limit (IDL) was revised to be consistent with methods
6010D and 6020B.
7. The term “accuracy” was replaced by “bias” where appropriate.
8. Definition for linear range was revised to be consistent with methods 6010D and 6020B.
9. All references to atomic emission spectrometry were changed to optical emission
spectrometry throughout the chapter.
10. Replaced the definition of interference check sample (ICS) with the spectral interference
check (SIC) solution to be consistent with methods 6010D and 6020B.
11. Added a note to clarify the methods that can be used for total acid soluble/recoverable
metals.
12. References 20-22 were added as guidance for minimizing sampling bias.
13. The term “decomposition” was replaced by “digestion” where appropriate.
14. Descriptions of cold vapor atomic fluorescence and anodic stripping voltammetry were
added to Section 3.7.
15. Collision/reaction cell technology was added to Sections 3.6 and 3.7 as an effective
method for removing isobaric interferences when analyzing by ICP-MS.
16. The terms "analytical" and "analysis" were changed to "instrument" or "determinative",
where appropriate.
17. The term “signal” was changed to “instrument response”.
18. The method for assessing sensitivity for ICP-OES and ICP-MS in Section 3.1 was
modified.
19. The term “optimum concentration range” was removed from Section 3.1.
20. Antimony was added to the HGAA method described in Section 3.7.
21. The definition for “laboratory control sample” was revised to recommend the use of a
spiking solution from the same source as the calibration standards.
22. A minimum collection mass of 100 g was added to Table 3-2 for solid samples collected
for sulfide analysis.
23. Definition of Bioaccessibility was added to Section 3.1.
24. Definition of Reagent Blank was added to Section 3.1.
25. Method 1340 was added to Section 3.6.
26. Updated Revision# and Date.