Capitulo 3 EPA

Download as pdf or txt
Download as pdf or txt
You are on page 1of 29

CHAPTER THREE

TABLE OF CONTENTS

Section Page

3.0 INTRODUCTION 1
3.1 DEFINITIONS 1
3.2 SAFETY 4
3.3 SAMPLING CONSIDERATIONS 4
3.4 SPECIAL CONSIDERATIONS FOR DETERMINING INORGANIC ANALYTES AT 6
ULTRA-TRACE CONCENTRATION LEVELS
3.5 REAGENT PURITY 11
3.6 SAMPLE DIGESTION METHODS 11
3.7 METHODS FOR DETERMINATION OF INORGANIC ANALYTES 13
3.8 REFERENCES FOR PREVIOUS SECTIONS AND THE TABLES AND FIGURES 16

TABLE
3-1 MATERIALS FOR USE IN SAMPLE COLLECTION FOR INORGANIC ANALYTE 18
DETERMINATIONS
3-2 RECOMMENDED SAMPLE HOLDING TIMES, PRESERVATION, COLLECTION 19
QUANTITIES, AND DIGESTION VOLUMES FOR SELECTED INORGANIC ANALYTE
DETERMINATIONS IN AQUEOUS AND SOLID SAMPLES
3-3 EXAMPLES OF THE ANALYTICAL BLANK INFLUENCE ON ULTRA-TRACE 21
ANALYSIS OF ELEMENTS IN GLASS
3-4 EXAMPLES OF LEAD CONCENTRATIONS IN AIR 21
3-5 CLEANLINESS LEVELS IN INTERNATIONAL STANDARD ISO 14644-1 22
3-6 PARTICULATE CONCENTRATIONS IN LABORATORY AIR 23
3-7 NON-CONTAMINATING MATERIALS AND FOR USE AS DIGESTION VESSELS AND 23
SAMPLE CONTAINERS IN ULTRA-TRACE ANALYSIS

FIGURE
3-1 COMPARISON OF CLEAN VERSUS CONVENTIONAL SAMPLING AND ANALYSIS 24
TECHNIQUES USED IN THE ANALYSIS OF SOUTH TEXAS ESTUARY WATERS
3-2 COMPARISON OF PARTICLE COUNT ANALYSIS OF A CLEAN ROOM AND A 25
STANDARD LABORATORY AT DUQUESNE UNIVERSITY IN PITTSBURGH, PA
3-3 PARTICLE SIZE COMPARISON CHART FOR COMMON PARTICULATES 26

Appendix A SUMMARY OF UPDATES/CHANGES IN CHAPTER 3 27

SW-846 Update VI THREE - i Revision 6


December 2018
CHAPTER THREE

INORGANIC ANALYTES

Prior to employing the methods in this chapter, analysts are advised to consult the
disclaimer statement at the front of this manual and the information in Chapter Two for guidance
on the allowed flexibility in the choice of apparatus, reagents, and supplies. In addition, unless
specified in a regulation, the use of SW-846 methods is not mandatory in response to Federal
testing requirements. The information contained in each procedure is provided by EPA as
guidance to be used by the analyst and the regulated community in making judgments
necessary to meet the data quality objectives or needs for the intended use of the data.

For a summary of changes in Chapter Three, please see Appendix A at the end of this
document.

3.0 INTRODUCTION

This chapter provides guidance for the analysis of inorganic analytes in a variety of
matrices. The analytical methods are written as specific steps in the overall analysis scheme --
sample handling and preservation, sample digestion or preparation, and sample analysis for
specific inorganic components. From these methods, the analyst should assemble a total
analytical protocol which is appropriate for the sample to be analyzed and for the information
required. This introduction discusses the options available in general terms, provides
background information on the analytical techniques, and highlights some of the considerations
to be made when selecting a total analysis protocol.

3.1 DEFINITIONS

The following terms are relevant for the determination of inorganic analytes:

Calibration blank: The calibration blank is a sample of analyte-free media


prepared with the same amounts of acids or other reagents as were the standards
and samples that can be used along with prepared standards to calibrate the
instrument. A calibration blank may also be used to verify absence of instrument
contamination (e.g., initial calibration blank and continuing calibration blank).

Calibration curve: The functional relationship between instrument response and


target analyte concentration determined for a series of calibration standards. The
calibration curve is obtained by plotting the instrument response versus
concentration and performing a regression analysis of the data.

Calibration standards: A series of solutions containing the target analyte at known


and various concentrations used to calibrate the instrument response with respect
to analyte concentration (i.e., preparation of the calibration curve).

Continuing calibration verification (CCV): A solution containing a known


concentration of analyte typically derived from the same source as the calibration
standards. The CCV is used to assure calibration accuracy during each analysis
run. It should be run for each analyte as described in the particular analytical
method. At a minimum, it should be analyzed at the beginning of the run and after

SW-846 Update VI THREE - 1 Revision 6


December 2018
the last analytical sample. The CCV concentration should be at or near the mid-
range levels of the calibration curve.

Dissolved metals: The concentration of metals determined in an aqueous sample


after the sample is filtered through a 0.45-µm filter (see Method 3005).

Initial calibration verification (ICV) standard: A certified or independently-prepared


solution from a source other than used for the calibration standards and used to
verify the accuracy of the initial calibration.

Instrument detection limit (IDL): Typically used in metals analysis to evaluate the
instrument noise level and response changes over time for analytes of
interest. IDLs in µg/L can be estimated as the mean of the blank results plus three
times the standard deviation of 10 replicate analyses of the reagent blank solution.
(Use zero for the mean if the mean is negative). Each measurement should be
performed as though it were a separate calibration standard (i.e., each
measurement must be followed by a rinse and/or any other procedure normally
performed between the analysis of separate samples). IDLs should be determined
at least once using new equipment, after major instrument maintenance such as
changing the detector, and/or at a frequency designated by the project. An
instrument log book should be kept with the dates and information pertaining to
each IDL performed.

Laboratory control sample (LCS): A volume of reagent water spiked with known
concentrations of analytes and carried through the sample preparation and
determinative procedure. It is used to monitor laboratory performance on analyte
loss/recovery in a clean matrix. The LCS should be prepared from the same
source as the calibration standards to remove potential error contribution from
standards of different sources. An independently prepared LCS may also be
obtained as or prepared from a certified reference solution or prepared from a
certified reagent solid or from an alternate lot reagent solid relative to the
calibration standards source if, for each analytical batch, at least one LCS is
prepared from the same source as the calibration standards. In this way, if the
recoveries of both the LCS and the matrix spike are outside the acceptance limits,
the analyst will be able to determine whether the problem is due to calibration error
or matrix interference.

Linear range: In both inductively coupled plasma optical emission spectrometry


(ICP-OES) and inductively coupled plasma mass spectrometry (ICP-MS) analysis,
the concentration range over which instrument response is linear. The linear range
establishes the highest concentration that may be reported without diluting the
sample. Following calibration, the laboratory may choose to analyze a standard at
a higher concentration than the high standard in the calibration. The standard
must recover within 10% of the true value, and if successful, establishes the linear
range. The linear range standards must be analyzed in the same instrument run
as the calibration they are associated with (i.e., on a daily basis) but may be
analyzed anywhere within that run. If a linear range standard is not analyzed for
any specific element, the highest standard in the calibration becomes the linear
range.

NOTE: Other standards exist that have alternative methods for determining the
linear range (e.g., ISO 17025). Therefore, the method used to define and verify the
linear range should meet the requirements of the project.

SW-846 Update VI THREE - 2 Revision 6


December 2018
Lower limit of quantitation (LLOQ): The lowest point of quantitation, or in most
cases, the lowest point in the calibration curve, which is ideally less than or equal
to the desired regulatory action levels based on the stated project requirements.
Laboratory-specific recovery limits should be established when the laboratory has
determined sufficient data points exist. Individual methods specify recovery limits
for use until the laboratory has sufficient data to determine acceptance limits
statistically.

Method blank: A volume of reagent water equal to that used for aqueous samples,
or, otherwise, a clean, empty container, equivalent to that used for actual solid
samples, processed through each sample preparation and determinative
procedure. Analysis of a method blank is used to assess contamination from the
laboratory environment, sample processing equipment, and/or reagents.

Reagent blank – Unprocessed (not run through the extraction procedure)


extraction fluid should be analyzed for each new batch of extraction fluid. The
reagent blank is considered within control limits if its result is less than the lower
limit of quantitation (LLOQ). The corrective action for a blank hit above LLOQ
should include preparing a new batch of extraction fluid and reprocessing any
samples that were prepared with the failing reagent fluid

Method of standard addition (MSA): An alternative calibration procedure employed


when the instrument response of the analyte of interest is different in a particular
matrix than when it is in reagent water. The procedure is generally reserved for
analyzing complex matrices. The standard addition technique involves the addition
of known amounts of the target analyte to each of a series of replicate sample
aliquots. The final concentrations of the sample replicates should span the
calibration range of the method. The analytical responses versus the standard
addition concentration for each of the replicates is plotted. After performing a
linear regression, the curve is extrapolated to the x-axis. The analyte
concentration in the original unspiked sample is equal to the inverse of the x-
intercept. See Method 7000, for more information.

Sample holding time: The storage time allowed between sample collection and
sample analysis when the designated preservation and storage techniques are
employed. Different times may be specified for holding field samples prior to
extraction, digestion, or other such preparation procedures versus holding
prepared samples (e.g., an extract or a digestate) prior to analysis.

Sensitivity: The ability of an analytical technique or instrument to discriminate


between small differences in analyte concentration (Reference 1). For metals
analysis, the following methods are commonly employed to determine sensitivity.

(a) Atomic absorption (AA): The concentration of metal, in mg/L, that


produces a transmission of 1%.

(b) Graphite furnace AA (GFAA): The mass of analyte required to give a


response of 0.044 absorbance-seconds.

(c) Inductively coupled plasma (ICP-OES or ICP-MS): Sensitivity is


measured in terms of the instrument detection limits as defined elsewhere
in the definitions.

SW-846 Update VI THREE - 3 Revision 6


December 2018
Spectral Interference Check solution (SIC): A solution containing both interfering
and analyte elements of known concentration that can be used in ICP-OES and
ICP-MS analysis to verify background and inter-element correction factors.

Suspended metals: The concentration of metals determined in the portion of an


aqueous sample that is retained by a 0.45-µm filter (Method 3005).

Total acid soluble/recoverable metals: The concentration of metals determined in


an unfiltered sample following digestion using hot mineral acid by Methods 3005,
3010, 3015, 3020, 3050, or 3051.

NOTE: Methods 3010 and 3020 state they are used to determine “total metals”.
However, because these methods do not use hydrofluoric acid, they are unable to
break down silicates and have been defined here for the “total acid
soluble/recoverable metals” fraction.

Total metals: The concentration of metals determined in a sample following


digestion by Method 3052.

Bioaccessibility: An in vitro measure of the physiological solubility of the metal that


may be available for absorption into the body (see Method 1340).

3.2 SAFETY

The methods in this chapter do not address all safety issues associated with their use.
The laboratory is responsible for maintaining a safe work environment and a current awareness
file of OSHA regulations regarding the safe handling of the chemicals specified in this method.
A reference file of material safety data sheets (MSDSs) should be available to all personnel
involved in these analyses.

The toxicity or carcinogenicity of each reagent used in these methods has not been
precisely defined. However, each chemical compound should be treated as a potential health
hazard. From this viewpoint, exposure to these chemicals should be reduced to the lowest
possible level by whatever means available. The following additional references to laboratory
safety are available:

1. "Carcinogens - Working with Carcinogens," Department of Health, Education, and


Welfare, Public Health Service, Centers for Disease Control, National Institute for
Occupational Safety and Health, Publication No. 77-206, August 1977.

2. "Handbook of Chemical Health and Safety," American Chemical Society, Oxford


University Press, New York, 2001.

3. "NIOSH Pocket Guide to Chemical Hazards," Department of Health and Human


Services, Centers for Disease Control, National Institute for Occupational Safety and
Health, Publication No. 2005-149, September 2005.

4. "Occupational Safety and Health Standards," 29 CFR Part 1910, Occupational Safety
and Health Administration, Department of Labor.

SW-846 Update VI THREE - 4 Revision 6


December 2018
5. "Safety in Academic Chemistry Laboratories," 7th Edition, Volumes 1 and 2, American
Chemical Society, Committee on Chemical Safety, Washington, D.C., 2003.

3.3 SAMPLING CONSIDERATIONS

3.3.1 Sample Collection

The fundamental goal of all field sampling activities is to collect samples that are
representative of the water, soil or waste from which they were collected. Thus,
representative sampling may be considered to be the sampling analog to analytical
accuracy. Of equal importance is sampling precision for ensuring consistency both within
a single sampling event and between sampling events conducted over time. Sampling
imprecision can be a significant source of measurement error. High quality field practices
are, therefore, necessary for generating representative samples on a consistent basis.
Sampling quality assurance includes the development of a quality assurance plan, data
quality objectives and the generation of field quality control samples including equipment
rinsates, trip blanks and field duplicates. Regardless of the specific program needs, the
documentation of all relevant field and sample information is the final essential component
of a sampling event for providing evidence that proper procedures and quality assurance
were performed during sample collection. Use of inadequate field procedures and
documentation can jeopardize an entire sampling program despite adequate planning,
analytical facilities, and personnel.

While advances in analytical sensitivity are continuing to be made that allow for
quantification of environmental contaminants at ultra-trace levels (i.e., < 0.1 ppb), clean
sampling techniques are consequently being devised and practiced in order to minimize or
eliminate sources of contamination during the collection of samples intended for ultra-trace
contaminant testing. Such clean sampling and analysis techniques are not generally
needed or required under the RCRA program and are beyond the scope of this chapter.
However, as an introduction to this topic, Sec. 3.4 provides a more detailed discussion on
the special category and requirements of clean analysis for determining constituents at
ultra-trace levels.

3.3.2 Sample Containers

Sample container materials can introduce either positive or negative errors in


measurement, particularly at low or ultra-trace levels, by contributing contaminants
through leaching or surface desorption, or by depleting concentrations through adsorption.
Additionally, the sample containers should be compatible with the reagents used for
sample preservation. Thus, the collection and containment of the sample prior to analysis
requires particular attention. Sample contamination introduced through field collection
activities including sample containment and shipment can be assessed from the analysis
of equipment rinsates. Guidelines on the selection of appropriate sample container
materials for the collection of inorganic analytical samples are provided in Table 3-1.

3.3.3 Cleaning of Sample Containers

Sample containers should be scrupulously clean so as not to introduce


contaminants that could interfere with quantification of the target analyte(s). This is of
particular importance when determining trace or ultra-trace analyte concentration levels.
The following cleaning sequence has been determined to be adequate to minimize

SW-846 Update VI THREE - 5 Revision 6


December 2018
contamination in the sample bottle, whether borosilicate glass, linear polyethylene,
polypropylene, or PTFE:

• Phosphate-free detergent
• Tap water
• 1:1 HNO3
• Tap water
• 1:1 HCl
• Tap water
• Reagent water

NOTE: Chromic acid should not be used to clean glassware, especially if chromium is
to be included in the analytical scheme. Commercial, non-chromate products may be
used in place of chromic acid, if adequate cleaning is documented by an analytical quality
control program. Chromic acid should not be used with plastic bottles.

3.3.4 Sample Handling and Preservation

Recommended materials to use for sample collection are listed in Table 3-1.
Sample holding times, recommended collection volumes or masses and recommended
digestion volumes, and preservatives are listed in Table 3-2. The sample collection and
digestion amounts depend on the combination of digestion or extraction and determinative
procedures that will be employed for a given sample as well as the sensitivity and tolerable
sampling errors that are required for a specific project (see References 20 through 22).
Likewise, the use of alternative preservatives to those indicated in Table 3-2 may be
necessary depending on the objectives of the project. In all cases, the sample quantity
that is collected should be representative of the bulk material whenever feasible.

3.3.5 Sample Preparation

For all non-speciated digestion methods, great reduction in analytical variability can
be achieved through the use of appropriate sample preparation procedures. Generally, a
reduction in subsampling variance can be accomplished by increasing the subsample
aliquot size, reducing the sample particle size, and/or thoroughly mixing the resulting fines
(see References 20 through 23 for additional information on sampling and sub-sampling).
Under most circumstances, it is recommended that the sample be analyzed without drying.
If it is necessary to report the analytical data on a dry-weight basis, then a separate aliquot
may be analyzed for moisture content and the wet-weight data corrected accordingly.

If the sample cannot be well-mixed and homogenized in the form in which it was
received by the laboratory, then air- or oven-drying at 30 °C or less, crushing, sieving,
grinding, and mixing should be performed as needed or feasible to homogenize the
sample until the subsampling variance is less than the data quality objectives of the
analysis. While proper sample preparation generally produces great reduction in
analytical variability, it should be noted that in certain unusual circumstances there could
be loss of volatile metals (e.g., mercury, organometallics) or irreversible chemical changes
( e.g., precipitation of insoluble species, change in valence state) caused by inappropriate
sample preparation procedures.

Variability due to sample heterogeneity is assessed by analyzing individually


prepared sample replicates such as field duplicates (FD), laboratory matrix duplicates
(MD), and matrix spike (MS)/matrix spike duplicate (MSD). Variability inherent in the
SW-846 Update VI THREE - 6 Revision 6
December 2018
analytical determinative procedure can be assessed by matrix spiking of individually
digested samples (i.e., post-digestion matrix spike) or by analyzing laboratory splits of the
digestate.

3.4 SPECIAL CONSIDERATIONS FOR DETERMINING INORGANIC ANALYTES AT ULTRA-


TRACE CONCENTRATION LEVELS

3.4.1 Clean Sampling Techniques

For the determination of ultra-trace analyte concentrations in environmental


samples, it is essential that samples be collected and subsequently managed using
techniques specifically designed to minimize sample contamination from field collection
activities and to ensure target analyte stability. Such techniques represent a special
category of sampling procedures designed specifically for ultra-trace analyses and are
commonly referred to as clean or ultra-clean sampling procedures. Clean sampling
methods are generally not intended for the determination of discharges from industrial
facilities. Rather, they are primarily applicable for the determination of ambient element
concentrations at levels of 0.1 ppb or less. At these concentrations, the opportunity for
sample contamination during sample collection or analysis in the laboratory is significant
and should be managed accordingly. Figure 3-1 provides a demonstration of the impact of
clean sampling and analysis techniques on data obtained for estuarine waters. Clean
sampling typically involves the following key steps:

• Special container pre-cleaning and pre-packaging requirements


• Specific sampling equipment and container materials selection
• Specific cleaning protocols for sampling equipment
• Equipment and container blank determinations prior to field use
• "Clean hands/dirty hands" sample collection techniques based on a 2-person
sampling crew
- Dirty hands sampler manages sampling equipment only
- Clean hands sampler manages the sample container
• Special sample packaging prior to shipment
• Use of a laboratory trained and properly equipped to perform clean analysis of
the analytes of interest

Given the laboratory resources required to perform clean analysis techniques, it is


paramount that samples be collected using ultra-clean techniques and conditions in the
field. Otherwise, subsequent analytical efforts become futile. The information provided in
this section is intended only as an introduction to the topic of clean sampling. Specific
guidelines for clean sampling may be found in Reference 2 and other sources.

3.4.2 Clean Analysis and the Analytical Blank

The significant role of the analytical blank in chemical analysis of trace metals
cannot be overemphasized. Sensitive instrumentation such as ICP-MS, ICP-OES, and
GFAA requires that sample preparation be at least as sophisticated as the instruments
used for analysis. The analytical blank is normally a primary source of error in ultra-trace
element analysis. Ultra-trace analysis is as dependent on control of the analytical blank as
it is on the bias and precision of the instrument making the measurement. Inability to
control contamination is frequently the limiting factor in trace (parts per million (ppm) to
parts per billion (ppb)) and ultra-trace (ppb to parts per trillion (ppt)) analysis. Analytical
blank contributions occur from the following four major sources (References 3 through 7):

SW-846 Update VI THREE - 7 Revision 6


December 2018
• The atmosphere in which the sample preparation and analysis are conducted
• The purity and the quantities of the reagents used in sample preparation,
including all reagents added directly to the sample
• The materials and equipment used in digestion or extraction vessels that come
in contact with the sample during the sample preparation and analysis (i.e., the
cleanliness and maintenance as well as their material properties).
• The analyst’s technique and skill in preparing the samples and performing the
analyses

The four primary areas that affect the analytical blank can be demonstrated using
standard reference materials in analysis. Table 3-3 illustrates and isolates the main
influencing parameters on the blank: environment, reagents, materials, and analyst skills.
The skill of the analyst was kept constant as the same analyst changed the environment,
reagents, and combinations of these parameters in the analysis (see Reference 6). The
trace elements in glass (TEG) standard reference material from the National Institute of
Standards and Technology (NIST) was used to keep sample homogeneity constant and to
permit removal of the sampling error by using sample sizes in which appropriate
homogeneity had previously been demonstrated. It is important to note that the
relationship of the precision and measurement remained relatively constant. The
significance of the first two major sources of contamination, environment and reagents, is
evaluated. The contamination in the laboratory air and in the acid used for the reagent
blank altered the accuracy of the example above by over two orders of magnitude for both
lead and silver. The larger influence of the two sources in this example is the laboratory
environment in which the samples were prepared.

3.4.2.1 Sample Preparation and Analysis Atmosphere

The atmosphere in which the sample is prepared is a major source of


contamination for most target analytes when analyzing at ultra-trace levels. With
the exception of some rare constituents, contamination from airborne sources
represents the most significant of the four main contamination sources. For
example, the mercury in a dental amalgam filling may cause blank
contamination. To illustrate this point, Table 3-4 presents concentrations of lead
found in samples of ambient air.

This contamination can also be seen in the comparison of 58,000


particles per liter of air measured in a normal laboratory in Pittsburgh, PA, and
inside a clean chamber in an adjacent laboratory five meters away. Figure 3-2
demonstrates the dramatic difference between the two environments. Cost-
effective methods of creating clean chambers for sample preparation are
documented along with this data in Reference 4.

Any laboratory air that comes into contact with the sample may deposit
airborne contamination into the sample. The sample is especially vulnerable
when it is being digested with acid. When samples are being digested with
acid(s), contamination of the sample by particulates in the laboratory air can
cause results that are biased high.

To prevent air from contaminating a sample for ultra-trace analysis, the


sample should be processed in a clean environment. This is much easier to
accomplish than it may appear at first. These precautions are becoming state-of-
the-art in many analytical and environmental laboratories. The prevention of
airborne contamination is most frequently dealt with by employing a laminar flow
SW-846 Update VI THREE - 8 Revision 6
December 2018
clean bench or a clean laboratory facility. Instructions are referenced for the
construction of both from component parts; both are relatively inexpensive and
uncomplicated, once the concepts are understood (Reference 4).

There are many sources of airborne contamination. Several of the


sources have been described and their particle size ranges are provided in
Figure 3-3. These sources primarily provide particulates in discrete size ranges.
Depending on whether the laboratory is located in an industrial, urban, or rural
area, or near the sea, the distribution of these source particles will be different,
as will their composition. The vertical dashed line in Figure 3 indicates the
particle size cutoff, usually 0.5 µm, for the high efficiency particulate air (HEPA)
filter used to prevent particulate contamination. Particles above this size cannot
pass through a HEPA filter that is in good working order. These filters are in
common use today (References 4 and 8).

The definition of clean air is derived from International Standard ISO


14644-1, which defines cleanliness levels. Table 3-5 lists these conditions.
"Laminar flow" is directed coherent air movement that does not contain any
turbulence.

A dramatic reduction in airborne contaminants can be achieved by using


HEPA-filtered air in laminar-flow clean hoods or entire clean laboratories. Table
3-6 demonstrates the dramatic differences in airborne contaminant
concentrations in an ordinary laboratory, a clean laboratory, and a clean hood
inside a clean laboratory.

3.4.2.2 Reagent Purity for Ultra-trace Analysis

The purity of the reagents used for acid digestion, leaching, and
extraction is extremely important to the overall level of the blank. Reagents have
very different purities, depending on their processing grade and purpose.
Frequently, the analyst should purchase special reagents, or purify lesser-grade
reagents prior to use, in order to minimize the analytical blank.

In addition to the purity of the reagents, the quantity of reagent that is


added to the sample may have a significant impact. When reagents are added,
they bring with them elemental and molecular components that exist as
contaminants. Increasing the reagent used in excess of the stoichiometric
reaction, leads to greater potential for blank contamination. Reagents of high
purity should either be purchased or produced in the laboratory.

In the preparation of high purity reagents, there is only one significant and
practical choice for the method of purification, i.e., sub-boiling distillation
(References 9 through 11). Different from normal distillation, sub-boiling
distillation uses an infrared radiation source to heat the reagent to a temperature
just below the boiling point. This use prevents the formation of bubbles that rise
and burst at the surface of the liquid. Thus, the aerosolized solution particles are
left in solution and prevented from physically transporting contaminants
throughout the distillation apparatus. Sub-boiling distillation is a slower but very
reliable method of purifying all of the common mineral acids and many organic
reagents used in analytical methods. It relies exclusively on the vapor pressure
of the reagent, and contaminant, and can therefore be specifically optimized for
purification of the mineral acids if the object is to remove metal ions. Of all acids,

SW-846 Update VI THREE - 9 Revision 6


December 2018
nitric acid, for a variety of reasons, can be purified to excellent quality. Sources
for sub-boiling apparatus equipment and methods for constructing one are
provided in the references. Purchasing sub-boiling acids from commercial
sources is also an option. Construction or purchase of sub-boiling reagent
purification equipment may be cost effective for some laboratories depending on
the quantity of reagents required for sample throughput.

3.4.2.3 Materials for Sample Preparation, Storage, and Analysis

For ultra-trace analyses, only certain materials are preferred for use in the
construction of sample vessels and instrument components that come into
contact with the sample. Over the past two decades, materials identified as
being non-contaminating have become the top choices for bottles, beakers,
reaction vessels, storage containers, nebulizers, and instrument components for
trace and ultra-trace analysis. The materials are the same as those currently
being used in many digestion vessels, bomb liners, and microwave vessels. The
materials are characterized by being thermally durable, chemically resistant or
inert, non-contaminating, and possessing appropriate compression and tensile
strength. Table 3-7 lists, in order of preference, several types of, non-
contaminating materials that are chemically inert to most acid reactions. These
materials have been evaluated and tested extensively for their potential to
contaminate (References 4, 6, 7, 12, and 13).

With the exception of polyethylene, the materials listed in Table 3-7 are
those most commonly used for sample preparation vessels, both atmospheric
pressure vessels and closed vessel liners that come into contact with the sample.
These materials are the most stable to acid reactions (with the exception of
quartz and glass if hydrofluoric acid is used). Fluoropolymers are the most
common and were adapted from other chemical uses for application in pressure
systems. The fluoropolymers have the highest range of use temperatures for
most plastics, ranging from 270-300 °C. They are also chemically inert to the
majority of mineral acids and combinations thereof. Sulfuric acid has a boiling
point of approximately 330 °C and can damage all fluoropolymers by melting
them. Quartz and glass can safely contain sulfuric acid at these high
temperatures, but borosilicate glass is not appropriate for ultra-trace elemental
analysis (References 7 and 13). Glass actually forms a gel layer that hydrates
and leaches, transferring contaminants from the glass to the sample solution.
While these quantities may be considered minute, they would be detected in
blanks and samples undergoing ultra-trace analyses.

Polyethylene is suitable for storage of diluted samples after digestion, but


it does not have a thermal-use temperature appropriate for digestion. It is also
not sufficiently inert to be useful as a digestion vessel or vessel liner, similar to
polycarbonate and polypropylene. The low cost of polyethylene and its relative
inertness to cool, weakly acidic solutions make it an excellent storage container
for trace element solutions (Reference 4).

3.4.2.4 Analytical Technique and Synergistic Equipment

The fourth significant source of analytical blank contamination is the skill


of the analyst and the appropriateness of the technique being performed.
Analytical blank control has been explained as the combination of atmosphere,
reagent, material, and protocol being performed. Also, the skill and awareness of

SW-846 Update VI THREE - 10 Revision 6


December 2018
the analyst as well as the way in which the combinations of the aforementioned
clean chemistry techniques are applied will have a significant effect on the final
contamination error and analytical blank control. Sample preparation
instrumentation may also assist in these protocols. For example, microwave
sample preparation assists each of these parameters in synergistic ways, thus
lowering the analytical blank, improving blank precision, and enhancing overall
quality control and transferability of methods. Some instrumentation and
fundamental processes involved in specific sample preparation procedures
assists the analyst by incorporating useful clean chemistry concepts into
instrumentation and method structure. Such instrumentation is pertinent since
microwave methods now exist that provide sample preparation for leaching or
total analysis of many target analytes simultaneously. As an example, the skill of
the analyst with regard to clean chemistry is assisted by the method structure
and microwave equipment as indicated below:

• If a closed or controlled atmospheric microwave vessel is prepared in a clean


hood and sealed before leaving the clean environment, the sample will not be
affected by atmospheric contamination during the reaction, since it has not
been removed from a clean environment.

• The vessel materials described previously might not normally be used by


many laboratories, and therefore the advantages of the fluoropolymers would
not be realized if they were not required in most microwave reaction vessels
as they commonly are.

• The time that the sample spends in digestion, leaching, or extraction may be
reduced from hours to minutes, thus reducing the potential leaching of
contaminants from the container walls.

• Because most microwave systems are sealed systems, evaporation of the


reagent before it reacts productively is prevented and smaller quantities of
reagents are used, thus preventing excess and unnecessary accumulation of
contaminants in the blank.

By reducing the exposure variables, the blank variability is consequently


reduced in size and is more consistent. An example of these components
working together has been provided in the literature, where analysis under
different conditions has verified these conclusions (References 4, 14 and 15).
The example illustrates the isolation of the blank optimization areas:
environment, reagents, materials, and analyst skills. The skill of the analyst is
kept more constant as the instrument dictates cleaner, chemically-appropriate
procedures.

3.5 REAGENT PURITY

The purity of the reagents used for sample preservation, acid digestion, leaching,
extraction and analysis is extremely important for preventing or minimizing sample
contamination. Reagents have very different purities, depending on their processing grade and
purpose. Reagent grade, ACS grade or better are recommended for use with most SW-846
methods. Sample contamination introduced through sample preservation, handling, preparation
and analysis is assessed from the analysis of method blanks.

SW-846 Update VI THREE - 11 Revision 6


December 2018
3.6 SAMPLE DIGESTION METHODS

Many of the methods listed below employ HCl in the digestion process. Chlorine is an
interferant in ICP-MS analysis and its use in sample digestion is discouraged except when
absolutely necessary or when the instrument manufacturer has indicated that the use of HCl will
not adversely affect the equipment and accurate quantitation of the desired target analytes.
However, please note that recoveries of certain metals may be biased low when using nitric acid
only. Interference from chlorine may be minimized by use of collision or reaction cell
instrumentation.

The methods in SW-846 for sample digestion or dissolution include:

Method 3005A: Acid Digestion of Waters for Total Recoverable or Dissolved Metals for
Analysis by FLAA or ICP Spectroscopy

This method may be used for the preparation of ground water and surface water samples
for total recoverable and dissolved metal determinations by FLAA, ICP-OES, or ICP-MS. The
unfiltered or filtered sample is heated with dilute HCl and HNO3 prior to metal determination.

Method 3010A: Acid Digestion of Aqueous Samples and Extracts for Total Metals for Analysis
by FLAA or ICP Spectroscopy

This method may be used for the preparation of waste samples for total recoverable metal
determinations by FLAA, ICP-OES, or ICP-MS. The samples are vigorously digested with nitric
acid followed by dilution with hydrochloric acid. The method is applicable to aqueous samples,
leachates, and mobility-procedure extracts.

Method 3015A: Microwave Assisted Acid Digestion of Aqueous Samples and Extracts

This method may be used for the preparation of aqueous samples, mobility-procedure
extracts, and wastes that contain suspended solids for total recoverable metal determinations
by FLAA, GFAA, ICP-OES, or ICP-MS. Nitric acid and hydrochloric acid are added to the
sample in a PTFE digestion vessel and heated in a microwave unit prior to metals
determination.

Method 3020A: Acid Digestion of Aqueous Samples and Extracts for Total Metals for Analysis
by GFAA Spectroscopy

This method may be used for the preparation of waste samples for total recoverable
metals determinations by GFAA or ICP-MS. The samples are vigorously digested with nitric
acid followed by dilution with nitric acid. The method is applicable to aqueous samples,
leachates, and mobility-procedure extracts.

Method 3031: Acid Digestion of Oils for Metals Analysis by Atomic Absorption or ICP
Spectrometry

This method may be used for the preparation of waste oils, oil sludges, tars, waxes,
paints, paint sludges and other viscous petroleum products for analysis by FLAA, GFAA, and
ICP-OES. The samples are vigorously digested with nitric acid, sulfuric acid, hydrochloric acid,
and potassium permanganate prior to analysis.

Method 3040A: Dissolution Procedure for Oils, Greases, or Waxes

SW-846 Update VI THREE - 12 Revision 6


December 2018
This method may be used for the preparation of oily waste samples for determination of
soluble metals by FLAA and ICP-OES. The samples are dissolved and diluted in organic
solvent prior to analysis. The method is applicable to the organic extract in the oily waste EP
procedure and other samples high in oil, grease, or wax content.

Method 3050B: Acid Digestion of Sediments, Sludges, and Soils

This method may be used for the preparation of waste samples for total recoverable
metals determinations by FLAA and ICP-OES, or GFAA and ICP-MS depending on the options
chosen. The samples are vigorously digested in nitric acid and hydrogen peroxide followed by
dilution with either nitric or hydrochloric acid. The method is applicable to soils, sludges, and
solid waste samples.

Method 3051A: Microwave Assisted Acid Digestion of Sediments, Sludges, Soils, and Oils

This method may be used for the preparation of sludges, sediments, soils and oils for total
recoverable metal determinations by FLAA, GFAA, ICP-OES or ICP-MS. Nitric acid and
hydrochloric acid are added to the representative sample in a fluorocarbon digestion vessel and
heated in a microwave unit prior to metals determination.

Method 3052: Microwave Assisted Acid Digestion of Siliceous and Organically Based Matrices

This method may be used for the preparation of siliceous and organically based matrices
including ash, biological tissue, oil, oil contaminated soil, sediment, sludge, and soil for total
metals analysis by FLAA, CVAA, GFAA, ICP-OES, and ICP-MS. Nitric acid and hydrofluoric
acid are added to a representative sample in a fluorocarbon digestion vessel and heated in a
microwave unit prior to analysis.

Method 3060A: Alkaline Digestion for Hexavalent Chromium

This method may be used for the preparation of soils, sludges, sediments and similar
waste materials for hexavalent chromium determination. The samples are digested and heated
to dissolve the Cr(VI) and stabilize it against reduction to Cr(III).

Method 3200: Mercury Species Fractionation and Quantification by Microwave Assisted


Extraction, Selective Solvent Extraction and/or Solid Phase Extraction

This method uses sequential extraction and separation procedures to differentiate


mercury species that are present in soils and sediments into four distinct fractions: extractable
organic mercury, extractable inorganic mercury, semi-mobile mercury and non-mobile mercury.
Extraction is accomplished with the aid of either microwave irradiation or ultrasound.
Quantification of mercury in the different fractions may be performed using any appropriate
determinative method, (e.g., Method 7473, 1631, or Methods 7470 and 7471). The test method
also contains provisions for separating and quantifying individual extractable mercury species
using HPLC.

Method 1340: In Vitro Bioaccessibility Assay for Lead in Soil.

The purpose of this method is to define the proper analytical procedure for the validated in
vitro bioaccessibility (IVBA) assay for lead in soil, to describe the typical working range and
limits of the assay, quality assurance (QA), and to indicate potential interferences. At this time,

SW-846 Update VI THREE - 13 Revision 6


December 2018
this method has only been validated for lead-contaminated soil under field conditions and not for
other matrices (e.g., water, air, amended soils, dust, food, etc.).

3.7 METHODS FOR DETERMINATION OF INORGANIC ANALYTES

This section of the manual contains analytical techniques for trace inorganic analyte
determinations. Instrumental techniques include:

- Inductively coupled argon plasma optical emission spectrometry (ICP-OES),


- Inductively coupled plasma mass spectrometry (ICP-MS),
- Direct-aspiration or flame atomic absorption spectrophotometry (FLAA),
- Graphite furnace atomic absorption spectrophotometry (GFAA),
- Hydride-generation atomic absorption spectrometry (HGAA),
- Cold-vapor atomic absorption spectrometry (CVAA),
- Cold-vapor atomic fluorescence spectrometry (CVAF),
- Anodic stripping voltammetry (ASV),
- X-ray fluorescence (XRF),
- Ion chromatography (IC)
- Capillary electrophoresis (CE)
- Speciated isotope dilution mass spectrometry (SIDMS) and
- Several procedures for hexavalent chromium analysis.

Each of these (except the individual hexavalent chromium analyses) is discussed briefly below.
Some advantages, disadvantages, and cautions for the analysis of wastes are provided.

ICP-OES allows simultaneous or rapid sequential determination of many elements in a


short time. Aerosol samples are introduced into an extremely hot plasma source which
vaporizes, atomizes, and ionizes the sample components. Upon exiting the plasma, the
electronically excited analytes emit characteristic photons that are detected via emission
spectrometry. A primary disadvantage of ICP-OES is the occurrence of background radiation
from other elements and the plasma gases. Although all ICP-OES instruments utilize high-
resolution optics and background correction to minimize these interferences, analysis of trace
levels of inorganic analytes in the presence of a large excess of a single analyte is difficult.
Examples would be trace levels of inorganic analytes in an alloy or trace metals in a limed (high
calcium) waste. ICP-OES and FLAA have comparable detection limits (within a factor of 4)
except that ICP-OES exhibits greater sensitivity for refractories (Al, Ba, etc.). Further, the wide
dynamic ranges and multi-element capabilities make ICP-OES instruments more productive.
GFAA, in general, will exhibit lower detection limits than either ICP-OES or FLAA.

ICP-MS allows sensitive, simultaneous determination of many elements in a short time


frame using MS detection in place of OES. In general ICP-MS exhibits greater sensitivity than
GFAA, FLAA or ICP-OES for most elements. The main disadvantage of ICP-MS is isobaric
elemental interferences. These are caused by different elements or molecular species forming
ions with the same nominal mass-to-charge ratio. Mathematical correction for interfering ions
can minimize these interferences, but the most effective means of removal of molecular isobaric
interferences is the collision or reaction cell.

FLAA direct-aspiration determinations, as opposed to ICP-OES or ICP-MS, are normally


completed as single-element analyses and are relatively free of interelement spectral
interferences. Either a nitrous-oxide/acetylene or air/acetylene flame is used as an energy
source for dissociating the aspirated sample into the free atomic state, making analyte atoms
available for absorption of light and spectrophotometric detection. In the analysis of some

SW-846 Update VI THREE - 14 Revision 6


December 2018
elements, the temperature or type of flame used is critical. If the proper flame and analytical
conditions are not used, chemical and ionization interferences can occur.

GFAA replaces the flame with an electrically-heated graphite furnace. The furnace allows
for gradual heating of the sample aliquot in several stages. Thus, the processes of dissolution,
drying, decomposition of organic and inorganic molecules and salts, and formation of atoms,
which should occur in a flame or plasma in a few milliseconds may be allowed to occur over a
much longer time period and at controlled temperatures in the furnace. This allows an
experienced analyst to remove unwanted matrix components by using temperature
programming and/or matrix modifiers. The major advantage of this technique is that it affords
extremely low detection limits. It is the easiest to perform on relatively clean samples. Because
this technique is so sensitive, interferences can be a real problem; finding the optimum
combination of digestion, heating times and temperatures, and matrix modifiers can be a
challenge for complex matrices.

HGAA utilizes a chemical reduction to reduce and separate arsenic, selenium, or antimony
selectively from a sample digestate. The technique therefore has the advantage of being able
to isolate these elements from complex samples which may cause interferences for other
analytical procedures. Significant interferences have been reported when any of the following is
present: (1) easily reduced metals (Cu, Ag, and Hg); (2) high concentrations of transition metals
(>200 mg/L); (3) oxidizing agents (oxides of nitrogen) remaining following sample digestion.

CVAA uses chemical reduction to reduce mercury to its volatile elemental form. The
procedure is sensitive, though not as sensitive as CVAF, but is subject to interferences from
some volatile organics, chlorine, and sulfur compounds. CVAA may be more appropriate than
CVAF for samples with higher mercury concentrations.

CVAF uses chemical reduction to reduce mercury to its volatile elemental form. The
procedure is extremely sensitive, approximately one to two orders of magnitude more sensitive
than CVAA. CVAF may be more appropriate than CVAA for samples with low mercury
concentrations. Because of the extreme sensitivity of CVAF, contamination during sampling
and analysis is a major concern.

ASV is an electrochemical method that can be used for determining free dissolved metals
such as arsenic and mercury. The analyte of interest is reduced at the electrode surface during
a deposition step and oxidized from the electrode during a stripping step; the current measured
during the stripping step is proportional to the analyte concentration. ASV detects dissolved
divalent mercury species and cannot directly distinguish between organic and inorganic mercury
compounds. The sensitivity of this method for mercury is similar to CVAA. ASV detects
dissolved inorganic arsenic species and can distinguish between As(III) and As(V) by adjusting
the potential applied during the deposition step. The sensitivity of this method for arsenic is
about an order of magnitude more sensitive than ICP-MS, GFAA, or HGAA. This method is
subject to interferences from suspended solids, organics, and other metals. Samples can be
modified (e.g., digested, addition of chelating agent, etc.) to determine the concentration of
various species of the analyte of interest (e.g., total fraction, methylmercury, etc.)

XRF uses sealed radioisotope sources or other suitable X-ray source to irradiate samples
with X-rays. When a sample is irradiated with X-rays, the source X-rays may undergo either
scattering or absorption by sample atoms. This later process is known as the photoelectric
effect. When an atom absorbs the source X-rays, the incident radiation dislodges electrons
from the innermost shells of the atom, creating vacancies. The electron vacancies are filled by
electrons cascading in from outer electron shells. Electrons in outer shells have higher energy
states than inner shell electrons, and the outer shell electrons give off energy as they cascade

SW-846 Update VI THREE - 15 Revision 6


December 2018
down into the inner shell vacancies. This rearrangement of electrons results in emission of X-
rays characteristic of the given atom. The emission of X-rays, in this manner, is termed X-ray
fluorescence.

IC generally refers to the separation of ions through ion exchange chromatography. In this
technique, an aqueous sample is injected into a mobile solution that is carried into a
chromatography column. As the sample travels through the column, the sample analytes are
temporarily retained on the column, the stationary phase, via electrostatic forces. The
separated analytes are identified as they are released from the column based on their retention
time. Detection and quantification in IC are most commonly performed using conductivity
detection. IC is typically used for the determination of anionic analytes in waste samples.

CE refers to the electrophoretic separation of ions dissolved or suspended in an


electrolyte. Samples are introduced into a capillary tube containing an electrolytic buffer. Under
the application of an electric field the cations in the sample migrate toward the negatively
charged electrode (cathode) and the anions migrate toward the positively charged electrode
(anode). This technique may be coupled with a variety of determinative techniques for
quantitative analysis. Inorganic anions can be determined in environmental samples using CE
and indirect UV detection, in which analytes are detected and quantified based on proportional
decreases in the absorbance of the buffer solution. CE is a complementary technique to IC and
typically offers shorter analysis times than IC.

SIDMS is a quantitative method for determining elemental species based on the


measurement of isotope ratio(s) in each species of a nuclide using mass spectrometry after
speciated isotope dilution. Samples are mixed with one or more isotopic spikes which have
different isotopic abundances and are artificially converted to chemical forms corresponding to
the species to be analyzed. The spiked samples are then subjected to the separation of the
species and the measurement of the altered isotope ratios in each species. Both species
concentrations and species conversions can be mathematically derived.

The following methods are included in this section:

Method 6010D: Inductively Coupled Plasma-Optical Emission Spectrometry


Method 6020B: Inductively Coupled Plasma-Mass Spectrometry
Method 6200: Field Portable X-Ray Fluorescence Spectrometry for the Determination
of Elemental Concentrations in Soil and Sediment
Method 6500: Dissolved Inorganic Anions in Aqueous Matrices by Capillary Ion
Electrophoresis
Method 6800: Elemental and Speciated Isotope Dilution Mass Spectrometry
Method 7000B: Flame Atomic Absorption Spectrophotometry
Method 7010: Graphite Furnace Atomic Absorption Spectrophotometry
Method 7061A: Arsenic (Atomic Absorption, Gaseous Hydride)
Method 7062: Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)
Method 7063: Arsenic in Aqueous Samples and Extracts by Anodic Stripping
Voltammetry (ASV)
Method 7195: Chromium, Hexavalent (Coprecipitation)
Method 7196A: Chromium, Hexavalent (Colorimetric)
Method 7197: Chromium, Hexavalent (Chelation/Extraction)
Method 7198: Chromium, Hexavalent (Differential Pulse Polarography)

SW-846 Update VI THREE - 16 Revision 6


December 2018
Method 7199: Determination of Hexavalent Chromium in Drinking Water,
Groundwater and Industrial Wastewater Effluents by Ion
Chromatography
Method 7470A: Mercury in Liquid Waste (Manual Cold-Vapor Technique)
Method 7471B: Mercury in Solid or Semisolid Waste (Manual Cold-Vapor Technique)
Method 7472: Mercury in Aqueous Samples and Extracts by Anodic Stripping
Voltammetry (ASV)
Method 7473: Mercury in Solids and Solutions by Thermal Decomposition,
Amalgamation, and Atomic Absorption Spectrophotometry
Method 7474: Mercury in Sediment and Tissue Samples by Atomic Fluorescence
Spectrometry
Method 7580: White Phosphorus (P4) by Solvent Extraction and Gas
Chromatography
Method 7741A: Selenium (Atomic Absorption, Gaseous Hydride)
Method 7742: Selenium (Atomic Absorption, Borohydride Reduction)

3.8 REFERENCES FOR PREVIOUS SECTIONS AND THE TABLES AND FIGURES

1. Skoog, D. A., "Principles of Instrumental Analysis," 3rd ed, Saunders College Publishing,
Philadelphia, PA, 1985.

2. EPA, U.S., "Sampling Ambient Water for Trace Metals at EPA Water Quality Criteria
Levels," Method 1669, U.S. Environmental Protection Agency, Office of Water,
Washington, D.C., 1996.

3. Adeloju, S. B. and Bond, A. M., "Influence of Laboratory Environment on the Precision and
Accuracy of Trace Element Analysis," Anal. Chem., 1985, 57, 1728-1733.

4. Kingston, S. H. M., et al., "Microwave Enhanced Chemistry: Fundamentals, Sample


Preparation, and Applications," American Chemical Society, Washington, DC, 1997, p.
257-279.

5. Moody, J. R., "NBS Clean Laboratories for Trace Element Analysis," Anal. Chem., 1982,
54, 1358A-1376A.

6. Murphy, T., "National Bureau of Standards Special Publication 422 Accuracy in Trace
Analysis: Sampling, Sample Handling, and Analysis," National Bureau of Standards,
Gaithersburg, MD, 1976, p. 509-539.

7. Tolg, G. and Tschopel, P., "Determination of Trace Elements," VCH, New York, NY, 1994,
p. 1-38.

8. Zief, M. and Mitchell, J. W., "Chemical Analysis," Vol. 47, John Wiley & Sons, New York,
NY, 1976.

9. Kuehner, E. C., et al., "Production and Analysis of Special High-Purity Acids Purified by
Sub-Boiling Distillation," Anal. Chem., 1972, 44, 2050-2056.

10. Moody, J. R., "Purified Reagents for Trace Metal Analysis," Talanta, 1982, 29, 1003-1010.

11. Kuehner, E. C. and Freeman, D. H., "Purification of Inorganic and Organic Materials,"
Marcel Dekker, New York, NY, 1969, p. 297-306.

SW-846 Update VI THREE - 17 Revision 6


December 2018
12. Moody, J. R., "The Sampling, Handling and Storage of Materials for Trace Analysis,"
Philos. Trans. R. Soc. London, 1982, 305, 669-680.

13. Moody, J. R. and Lindstrom, R. M., "Selection and Cleaning of Plastic Containers for
Storage of Trace Element Samples," Anal. Chem., 1977, 49, 2264-2267.

14. Prevatt, F. J., "Clean Chemistry for Trace Analysis," Environmental Testing and Analysis,
1995, 4, 24-27.

15. Skelly, E. M. and DiStefano, F. T., "Clean Room and Microwave Digestion Techniques:
Improvement in Detection Limits for Aluminum Determination by GF-AAS," Appl.
Spectrosc., 1988, 42, 1302-1306.

16. Rabinowitz, M. B. and Wetherill, G. W., "Identifying Sources of Lead Contamination by


Stable Isotope Techniques," Environ. Sci. Technol., 1972, 6, 705-709.

17. Maienthal, E. J., "U.S. National Bureau of Standards Technical Note 545," National
Bureau of Standards, Gaithersburg, MD, 1970, p. 53-54.

18. Boothe, P. N. and Nicolau, B.A. "The Coastal Bend Bays Project," 26th Annual
Conference on Analysis of Pollutants in the Environment, Chicago, IL 2003.

19. Miller, G. T., "Living in the Environment, Wadsworth Inc., Belmont, CA, 1994.

20. Gy, P., "Sampling for Analytical Purposes" Wiley, New York, NY, 1998.

21. Pitard F.F. “Pierre Gy’s Sampling Theory and Sampling Practice,” Second Edition, CRC
Press, Inc., Boca Raton, FL, 1993.

22. EPA, U.S., “Correct Sampling Using the Theories of Pierre Gy,” Office of Research and
Development, Las Vegas, NV, 1999.

23. EPA, U.S., “Guidance for Obtaining Representative Laboratory Analytical Subsamples
from Particulate Laboratory Samples.” Office of Research and Development, 2003.

SW-846 Update VI THREE - 18 Revision 6


December 2018
TABLE 3-1

MATERIALS FOR USE IN SAMPLE COLLECTION FOR


INORGANIC ANALYTE DETERMINATIONS

Analyte Recommended Container Material


Metals PTFE, plastic, glass
Chloride PTFE, plastic, glass
Cyanide PTFE, plastic
Fluoride PTFE, plastic
Nitrate PTFE, plastic, glass
pH PTFE, plastic, glass
Specific Conductance PTFE, plastic, glass
Sulfate PTFE, plastic, glass
Sulfide PTFE, plastic, glass

a
These recommendations are intended as guidance only and not inclusive of all possible analytes and
materials. The selection of sample container should be made based on the nature of the sample, the
intended end use of the data and the project data quality objectives.

SW-846 Update VI THREE - 19 Revision 6


December 2018
TABLE 3-2
RECOMMENDED SAMPLE HOLDING TIMES, PRESERVATION, COLLECTION QUANTITIES,
AND DIGESTION VOLUMES/MASS FOR SELECTED INORGANIC ANALYTE
DETERMINATIONS IN AQUEOUS AND SOLID SAMPLESa,b

Minimum
Collection Digestion Holding
Analyte Matrix Fraction Volume/Mass Preservation1 Volume Time2
Metals Aqueous Total/total 600 mL HNO3 to pH<2 100 mL 6 months
(except Hg recoverable
and Cr6+)
Dissolved 600 mL Filter on site; 100 mL 6 months
HNO3 to pH<2

Suspended 600 mL Filter on site; 100 mL 6 months


Solid Total 200 g None 2g 6 months

Hexavalent Aqueous 400 mL ≤6 °C 100 mL 24 hours


chromium
Solid 100 g ≤6 °C 30 days to
extraction
≤6 °C 2.5 g 7 days from
extraction to
analysis
Mercury Aqueous Total 400 mL HNO3 to pH<2 100 mL 28 days
Dissolved 400 mL Filter; 100 mL 28 days
HNO3 to pH<2

Solid Total 200 g ≤6 °C 0.2 g 28 days

Solid Species 200 g ≤6 °C 0.2 g 5 days

Chloride Aqueous 50 mL ≤6 °C  28 days

Cyanide Aqueous 500 mL ≤6 °C;  14 days


NaOH to
pH>12
Solid 100 g ≤6 °C  14 days

Fluoride Aqueous 300 mL ≤6 °C  28 days

Nitrate Aqueous 1000 mL ≤6 °C  28 days

Hexane Aqueous 1000 mL ≤6 °C  28 days


Extractable HCl or H2SO4
Material to pH <2
(HEM; Oil &
Grease) Solid 100 g ≤6 °C 28 days

HCl or H2SO4 when


to pH <2 practical
SW-846 Update VI THREE - 20 Revision 6
December 2018
TABLE 3-2
RECOMMENDED SAMPLE HOLDING TIMES, PRESERVATION, COLLECTION QUANTITIES,
AND DIGESTION VOLUMES/MASS FOR SELECTED INORGANIC ANALYTE
DETERMINATIONS IN AQUEOUS AND SOLID SAMPLESa,b

Minimum
Collection Digestion Holding
Analyte Matrix Fraction Volume/Mass Preservation1 Volume Time2

pH Aqueous 25 mL NA  Analyze
immediately
Solid 20 g NA  Analyze
immediately
Specific Aqueous 100 mL NA  Analyze
Conductance immediately
Sulfate Aqueous 50 mL ≤6 °C  28 days

Sulfide Aqueous 100 mL 4 drops 2N  7 days


zinc
acetate/100
mL sample;
NaOH to
pH>9;
Minimize
Solid 100 g aeration;  7 days
Store
headspace
free at
≤6 °C

Fill sample
surface with
2N zinc
acetate until
moistened;
Store
headspace
free at ≤6 °C
Organic Aqueous 200 mL ≤6 °C  28 days
Carbon, Total store in dark
(TOC) HCl or H2SO4
to pH <2;
Solid 100 g  28 days
≤6 °C
a These recommendations are intended as guidance only. The selection of sample and digestion volumes/mass and
preservation and holding times should be made based on the nature of the sample, the intended end use of the data
and the data quality objectives.
b Additional sample quantities may need to be collected in order to allow for the preparation and analysis of QC

samples, such as matrix spikes and duplicates.


1 The exact sample extract, and standard storage temperature should be based on project-specific requirements

and/or manufacturer’s recommendations for standards. Alternative temperatures may be appropriate based on
demonstrated analyte stability within a matrix, provided the data quality objectives for a specific project are still
attainable.

SW-846 Update VI THREE - 21 Revision 6


December 2018
2A longer holding time may be appropriate if it can be demonstrated that the reported analyte concentrations are not
adversely affected by preservation, storage and analyses performed outside the recommended holding times.

TABLE 3-3

EXAMPLES OF THE ANALYTICAL BLANK INFLUENCE ON


ULTRA-TRACE ANALYSIS OF ELEMENTS IN GLASS

Conditions Pb (ng) Ag (ng)

Initial analysis of TEG* standard 330 ± 250 970 ± 500

Analysis using sub-boiled distilled acids 260 ± 200 -

Analysis in a Class 100 hood 20 ± 8 207 ± 200

Analysis using sub-boiled acids in a Class 100 hood 2±1 3±2

* TEG = Trace element in glass

Data are taken from Reference 6.

TABLE 3-4

EXAMPLES OF LEAD CONCENTRATIONS IN AIR

Site Lead Concentration (µg/m3) Source

Downtown St. Louis, MO 18.84 Reference 16

Rural park, Southeastern MO 0.77 Reference 17

NIST Laboratory, MD 0.4 Reference 6

SW-846 Update VI THREE - 22 Revision 6


December 2018
TABLE 3-5

CLEANLINESS LEVELS IN INTERNATIONAL STANDARD ISO 14644-1

Maximum concentration limits (particles/m-3 of air) for particles equal to and


ISO larger than the considered sizes shown below
classification
number 0.1 µm 0.2 µm 0.3 µm 0.5 µm 1 µm 5 µm

ISO Class 1 10 2

ISO Class 2 100 24 10 4

ISO Class 3 1,000 237 102 35 8

ISO Class 4 10,000 2,370 1,020 352 83

ISO Class 5 100,000 23,700 10,200 3,520 832 29

ISO Class 6 1,000,000 237,000 102,000 35,200 8,320 293

ISO Class 7 352,000 83,200 2,930

ISO Class 8 3,520,000 832,000 29,300

ISO Class 9 35,200,000 8,320,000 293,000

NOTE: Uncertainties related to the measurement process require that concentration data with no more
than three significant figures be used in determining the classification level

© ISO. This material is reproduced from ISO 14644-1:1999 with permission of the American
National Standards Institute (ANSI) on behalf of the International Organization for
Standardization (ISO). No part of this material may be copied or reproduced in any form,
electronic retrieval system or otherwise or made available on the Internet, a public network, by
satellite or otherwise without the prior written consent of ANSI. Copies of this standard may be
purchased from ANSI, 25 West 43rd Street, New York, NY 10036, (212) 642-4900,
http://webstore.ansi.org.

SW-846 Update VI THREE - 23 Revision 6


December 2018
TABLE 3-6

PARTICULATE CONCENTRATIONS IN LABORATORY AIR

Concentration (µg/m3)

Location Iron Copper Lead Cadmium


Ordinary
0.2 0.02 0.4 0.002
laboratory
Clean room 0.001 0.002 0.0002 ND

Clean hood 0.0009 0.007 0.0003 0.0002

ND = Not Detected

Data are taken from Reference 17.

TABLE 3-7

NON-CONTAMINATING MATERIALS AND FOR USE AS DIGESTION VESSELS


AND SAMPLE CONTAINERS IN ULTRA-TRACE ANALYSES

Listed from highest to lowest preference for use in sample containment

Fluoropolymers: PFA*, TFM, TFE*, FEP*, Tefzel*

Quartz - Synthetic

Polyethylene (suitable for storage only, not for acid digestion)

Quartz - Natural

Borosilicate Glass

* Various forms of PTFE

Data are taken from Reference 8.

SW-846 Update VI THREE - 24 Revision 6


December 2018
FIGURE 3-1

COMPARISON OF CLEAN VERSUS CONVENTIONAL SAMPLING AND ANALYSIS


TECHNIQUES USED IN THE ANALYSIS OF SOUTH TEXAS ESTUARY WATERS

40

Clean Conventional 26 fold

30
Concentration (ppb)

280 fold

20

10 fold

10
11 fold
36 fold
0
Cd Cu Hg Pb Zn
Element

Taken from Reference 18.

SW-846 Update VI THREE - 25 Revision 6


December 2018
FIGURE 3-2

COMPARISON OF PARTICLE COUNT ANALYSIS OF A CLEAN ROOM AND


A STANDARD LABORATORY AT DUQUESNE UNIVERSITY IN PITTSBURGH, PA

Taken from Reference 4.

SW-846 Update VI THREE - 26 Revision 6


December 2018
FIGURE 3-3

PARTICLE SIZE COMPARISON CHART FOR COMMON PARTICULATES

Taken from Reference 4, 19.

SW-846 Update VI THREE - 27 Revision 6


December 2018
APPENDIX A
SUMMARY OF UPDATES/CHANGES IN CHAPTER THREE

1. The entire document was formatted in Microsoft Word .docx format from the original
.wpd and .pdf files.
2. The revision number was changed to five and the date published to July 2014.
3. Minor editorial and grammatical changes were made throughout the Chapter.
4. Method 3200 was added to Section 3.6.
5. Method numbers were revised to reflect the assigned number for Update V.
6. Definition for instrument detection limit (IDL) was revised to be consistent with methods
6010D and 6020B.
7. The term “accuracy” was replaced by “bias” where appropriate.
8. Definition for linear range was revised to be consistent with methods 6010D and 6020B.
9. All references to atomic emission spectrometry were changed to optical emission
spectrometry throughout the chapter.
10. Replaced the definition of interference check sample (ICS) with the spectral interference
check (SIC) solution to be consistent with methods 6010D and 6020B.
11. Added a note to clarify the methods that can be used for total acid soluble/recoverable
metals.
12. References 20-22 were added as guidance for minimizing sampling bias.
13. The term “decomposition” was replaced by “digestion” where appropriate.
14. Descriptions of cold vapor atomic fluorescence and anodic stripping voltammetry were
added to Section 3.7.
15. Collision/reaction cell technology was added to Sections 3.6 and 3.7 as an effective
method for removing isobaric interferences when analyzing by ICP-MS.
16. The terms "analytical" and "analysis" were changed to "instrument" or "determinative",
where appropriate.
17. The term “signal” was changed to “instrument response”.
18. The method for assessing sensitivity for ICP-OES and ICP-MS in Section 3.1 was
modified.
19. The term “optimum concentration range” was removed from Section 3.1.
20. Antimony was added to the HGAA method described in Section 3.7.
21. The definition for “laboratory control sample” was revised to recommend the use of a
spiking solution from the same source as the calibration standards.
22. A minimum collection mass of 100 g was added to Table 3-2 for solid samples collected
for sulfide analysis.
23. Definition of Bioaccessibility was added to Section 3.1.
24. Definition of Reagent Blank was added to Section 3.1.
25. Method 1340 was added to Section 3.6.
26. Updated Revision# and Date.

SW-846 Update VI THREE - 28 Revision 6


December 2018

You might also like