13 Dpconv
13 Dpconv
13 Dpconv
97
When installing a moisture analyzer, sometimes the reading is 1. Since the moisture table is in °C, convert the 53.6°F to
different than what you expected. This difference is often due to 12°C.
a change in the pressure at the measurement point where the
new probe has been installed. For example, a probe is installed 2. Look up the vapor pressure for 12°C dew point — 10.518
at line pressure in a process where the previous method of mmHg.
measurement was at atmospheric pressure. The problem then 3. Convert the original psig value into atmospheres by adding
becomes one of predicting the pressure effect on the probe and 14.7 psia to the 180 psig to get psia. This gives 194.7 psia,
achieving the desired result. which we divide by 14.7 to get 13.2449 atmospheres.
Dew point temperature is always directly related to the partial 4. Since the total pressure change is a factor of 13.2449 to 1,
pressure of water vapor in the gas. In other words, if you know the resulting vapor pressure goes through the same
the dew point, the partial pressure of water for that dew point change. Dividing the 10.518 mmHg by 13.2449
temperature can be found in published tables. S e e G e n e r a l atmospheres gives 0.794 mmHg vapor pressure at the
Eastern’s Humidity Handbook. reduced pressure.
The ability to predict the dew point change from the total 5. Referring to a vapor pressure table, the 0.794 mmHg vapor
pressure change is based on physical laws which govern the pressure converts into -19.76°C dew point. This converts
behavior of gases. The Ideal Gas Law and Dalton's Law are into -3.57°F dew point at atmospheric pressure.
both in operation here. The Ideal Gas Law gives the relationship
of pressure, temperature, and volume in PV = nRT. Dalton's Law
states that total pressure is equal to the sum of all the partial
pressures of the component gases.
This flow cell installation has inlet and outlet valves that can
be used to operate the probe at line or
atmospheric pressure. If the inlet valve is opened completely
and the outlet valve is used to throttle the sample, the probe
will operate at line pressure. If the outlet valve is opened and
the inlet valve is used for throttling, the probe will operate at
atmospheric pressure. This installation provides the flexibility
for reading at a pressure that fits the application.