Operation and Maintenance Manual FOR Riken Compact Infrared Gas Monitor Model
Operation and Maintenance Manual FOR Riken Compact Infrared Gas Monitor Model
This is an infrared single gas monitor designed to measure one of the specified gas for
carbon monoxide, carbon dioxide or methan in air.
The gas monitor is a safety instrument, not an analyzer nor measuring instrument to
make quantitative and qualitative analysis or measurement.
Please understand following points and use this monitor correctly.
1. This gas monitor may respond to other gas and vapors than the target gas.
Be care to alarm action by interference.
It may fluctuate against the environmental(temperature, humidity, etc.) change
around the monitor.
Perform zero adjustment with regular interval(approx once a month).
2. Avoid to use this monitor at the place where the temperature is drastically(approx 10
~15℃) changed.
3. Alarm level shall be set within the performance of this monitor.
If set below our standard alarm level(20% of full scale), it may be caused to false
alarm.
4. This is not a control equipment, but a safety instrument.
Use the alarm contact from this monitor only for an external alarm light or buzzer.
And the analog output signal shall be used only for an external indicator or recorder.
We do not assume indemnification for any accident or damage caused by other
control purpose than the above.
5. Gas sampling from places where there are high humidity, dust or acid mist requires
sample pre-conditioner.
Please consult our nearest agent or Riken Keiki for such applications.
6. For maintenance, replacement parts listed in this manual and adjustments are
required.
We also recommend to perform regular maintenance including gas calibration in every
6 months as this is a safety instrument.
CONTENTS
1.PRODUCT OUTLINE
1-1 Preface ………………………………………………………………… 3
1-2 Application for use ……………………………………………………… 3
1-3 Identification of each caution mark …………………………………… 3
3.PRODUCT FUNCTIONS
3-1 Name of each part and functions ………………………………………… 6
3-1-1 Configuration ……………………………………………………… 6
3-1-2 Front panel ……………………………………………………… 8
3-2 Block diagram …………………………………………………………… 9
3-2-1 Electric diagram…………………………………………………… 9
3-2-2 Piping diagram …………………………………………………… 9
4.HOW TO USE
4-1 Before operation ……………………………………………………… 10
4-2 Installation place ……………………………………………………… 10
4-3 Mounting the dust filter ……………………………………………… 12
4-4 Caution in the system engineering……………………………………… 12
4-5 Grounding ……………………………………………………………… 15
4-6 Caution in wiring ………………………………………………………… 15
4-7 Applicable cable ………………………………………………………… 16
4-7-1 Power cable ……………………………………………………… 16
4-7-2 Output signal ……………………………………………………… 16
4-7-3 Cable for alarm contact…………………………………………… 16
4-7-4 Specifications of terminal plate…………………………………… 16
4-8 Arrangement of terminal plate…………………………………………… 16
4-9 Caution in piping works…………………………………………………… 17
5.OPERATION METHOD
5-1 Preparation to start-up …………………………………………………… 18
5-2 Basic performance flow …………………………………………………… 18
5-3 Start-up ……………………………………………………………………… 19
5-3-1 Power ON ……………………………………………………………… 19
5-3-2 Adjusting the sample flow rate……………………………………… 20
5-4 Explanation of performance………………………………………………… 20
5-4-1 Display action ………………………………………………………… 20
5-4-2 External output action………………………………………………… 22
5-5 Adjustments………………………………………………………………… 23
5-5-1 Zero adjustment……………………………………………………… 23
5-5-2 Span adjustment……………………………………………………… 26
5-6 Alarm test (Confirming a transmission status)……………………………… 29
5-7 How to finish operation …………………………………………………… 29
1
6.KIND OF ALARM AND ITS ACTION
6-1 Kind of alarm ……………………………………………………………… 30
6-2 Alarm action ……………………………………………………………… 30
6-3 Gas alarm (The unit with alarm setting)………………………………… 31
6-3-1 Responding to gas alarm……………………………………………… 31
6-3-2 Gas alarm possibility except the case of gas detection …………… 31
7.MAINTENANCE CHECK
7-1 Frequency of maintenance and check items …………………………… 32
7-1-1 Daily check …………………………………………………………… 32
7-1-2 Regular maintenance check ………………………………………… 32
7-1-3 Maintenance contract for regular check ………………………… 32
7-2 Replacing the parts ………………………………………………………… 33
7-3 Treatment when stopping operation or move its installation place … 34
7-3-1 Stopping the normal operation ……………………………………… 34
7-3-2 Changing the installation place ……………………………………… 34
7-4 Storage or treatment when not in use for a long period and treatment
for re-use………………………………………………………………… 34
7-5 Recommendable spare parts for regular maintenance ………………… 35
9. DEFINITION OF TERMS……………………………………………………………… 39
2
1.PRODUCT OUTLINE
1-1.Preface
First of all, we wish to express our heartfelt thanks for your purchase of our RI-557
Compact Infrared Gas Monitor.
This instruction manual is just a guide book to operate our gas monitor Model RI-557.
Your kind reading of this manual is requested not only for first user but for already
experienced staff.
The Riken Model RI-557 is a compact, single gas infrared monitor designed to measure one
of the specified gas and range for 1) carbon monoxide, 2) carbon dioxide or 3) methane in
air. The sample gas is drawn into this lightweight and rugged unit by an internal pump and is
then analyzed through a Non-Dispersive Infrared Ray Absorption detector. A recorder
output is standard to enable convenient and easy recording of your readings. Model RI-557
is powered on AC mains provide with LCD digital display to meet the extensive utility such
as in automotive, environmetal control, laboratories, university laboratory, research
institutes and many others.
3
2.IMPORTANT INSTRUCTION
FOR THE SAFETY
2-1.Danger items
DANGER
WARNING
○ Handling this monitor
Perform the operation and maintenance of this monitor by enough knowledgeable and
skilled person.
○ Power source
Before turning the power on, check that the specified voltage is within the supplied voltage.
Then, turn the power on.
○ Necessity of Grounding
Do not cut the protective ground both inside and outside of instrument. Or, do not
disconnect the connection of the ground terminals.
○ Fuse
To protect the instrument from fire, use the specifically rated fuse. When replacing the
fuse, turn off the power inside of the panel, and the power source as well. Do not use
fuse that is not specified, or do not make a short-circuit inside of the fuse holder.
○ Operation in Gas
Do not operate the instrument in a atmosphere of combustible gases, explosive gases
or vapor. It is very dangerous to operate the instrument under such circumstances.
○ External connections
Confirm that the outer equipment is connected to this monitor correctly.
4
2-3.Caution items
CAUTION
○ Do not use this monitor to control other equipment by use of output from this monitor.
This is not a control equipment.
Do not connect outer equipment for control purpose.
5
3. PRODUCT FUNCTIONS
3-1-1.Configuration
【 Front 】 【 Side 】
Air vent
【 Back 】 【 Top 】
④ Gas inlet
③ Dust filter
⑤ Fuse holder
6
① LCD display : Displaying gas concentration and various messages.
② Flowmonitor : To confirm the suction flow.
③ Dust filter : To remove the dust containing in the sampled gas.
④ Gas inlet : Sampled gas inlet to this monitor.
⑤ Fuse holder : 2A fuses are mounted.
⑥ Gas outlet : Sampled gas outlet from this monitor.
⑦ Power switch : ON/OFF switch for power supply
⑧ Terminal plate(3P): Terminal strip for power source.
⑨ Terminal plate(8P): Terminal strip for outputs/relay contacts.
⑩ Calibration gas inlet: Inlet for canned calibration gas(zero/span adjustments).
CAUTION
・Put the cap onto calibration gas inlet in case you do not
use calibration gas.
7
3-1-2.Front panel
② WARNING
light ① POWER light
③ ALARM light
④ MAINTENANCE
light
⑨ CAL.switch
⑧ SETswitch
⑦ TEST/▼Switch
⑥ ZERO/▲Switch
⑤ FLOW/ESC switch
8
3-2.Block diagram
3-2-1.Electric diagram
Gas concentration display
Detector
C P U
Voltage output
External output
Current
output
POWER light
WARNING light
ALARM light
MAINTENANCE light
WARNING contact
Noise filter
Fuse Power switch
Switching
power unit AC power source
3-2-2.Piping diagram
Cal. gas inlet
CAL. GAS IN
Stop valve
GAS outlet
Gas inlet
GAS OUT
Protective Flow Protective
Pump Detector
dust filter monitor filter
Flame Dust Flame
arrestor filter arrestor
*NOTE
・The flame arrestor is a optional accessory.
・Contact our nearest agent or Riken Keiki for flame arrestor.
9
4.HOW TO USE
4-1.Before operation
Be sure to keep caution items of use not only for first user but for already experienced staff.
If not keep these caution items, the unit may be defective and correct gas detection may
not be performed.
4-2.Installation place
CAUTION
・ Do not move this monitor during operation.
Reading may be fluctuated by vibration.
WARNING
, Do not install at places where direct sun drought or where the temperature is
drastically changed.
Keep away it at places where the temperature is drastically changed such as direct sun
drought, around the ventilation blower outlet or direct wind blow.
It may not be followed to the drastically temperature change as the monitor inside is
condensed. Also keep away from the place where the temperature is below 0℃ or above
40℃.
Keep the system from the equipment which may generate the noise (Unit &
cable).
Keep the system from the equipment which may give a high frequency in the
surroundings and install it.
・Do not put the system jointly each. ・Do not wire the cable in parallel nor take
access.
10
Do not install it where the target gas is deposited around.
・ It is required to make regular maintenance. Do not install it where it will be danger for
maintenance such as near the high electric power cable, et.
11
4-3.Mounting the dust filter
The performance of this unit will be interfered by dusts. Make sure to attach the dust filter
to the detection gas inlet.
Mount
WARNING
・ Be sure to mount the dust filter. If it is used without dust filter, water or dust will enter into
the monitor and correct measurement cannot be performed. And it may be the cause of
trouble.
The dust filter is equipped to the monitor at the time of shipment.
・ When replacing the filter, turn OFF the power to stop the pump suction before replacement.
CAUTION
・Unstable power and noise may cause error of performance and alarm.
・For the system to use this unit, it is required to make design based on this manual
descriptions.
Power
Within 100VAC±10% or within 220VAC±10%
voltage
Power failure Approx 10msec.
tolerance (For power failure above 10msec, it re-starts)
time To warrant the continuous operation, install the standby battery outside.
Do not share a power source with the power involving high power load and
Others high frequency noise.
Use line filter and separate it from noise source according to the need.
12
(2)Designing to consider radiation
When install the closed self-standing control panel, mount the fans in the upper
and lower part.
(3)Lightning measures
There is the problem “Lightning”. When make outdoor wiring of cable at factory
or plants etc or when make a parallel wiring in the same duct with the cable in
from outdoor even at the indoor wiring. If the lightning is a huge generation
Lightning
source, the cable is a reception antenna and there is the case that cable
surge
connecting instrument is broken. It is impossible to prevent the generation of
lightning. If the cable should put in metal tube or laid in the underground, it is
impossible to prevent the inductive lightning surge generating from the thunder.
There is not the complete countermeasure for it but the following method can be
considered.
Make the suitable treatment accordingly.
a)The transmission signal route is arranged for connection by the optical fiber
cable etc.
b)Countermeasure by the lighting arrester (Cable safety retainer). There is the
Lightning
way to install the lightning arrester just before the field apparatus and the
measures
central control station. The position of the lightning arrester installation is at
each point of cable laid out from the outdoor to the indoor. The lightning
arrester builds in the circuit to remove the surge voltage to be the source for
the damage of field apparatus (Protection resistor, zero diode etc.) and is
designed to protect the apparatus. But as the signal may be attenuated due
to the lightning arrester, check the action and it is required to use.
Surge noise shall be generated from the thunder lightning or except it.
Grounding
To protect detector from these cause, be sure to make grounding.
CAUTION
・The each relay contact for WARNING, ALARM and TROUBLE is set at OFF(not
functioning) as a standard version.
・Normally-closed contact (Break contact) at de-energized condition may change to
open contact in a moment due to physical shock.
・Whenever alarm signals from gas detectors are used with normally-closed
contact, please put delayed-circuit (for about one second) to receiver side of
normally-closed contact to avoid such phenomenon.
13
When control the external load, the bad influence may be given to the system according to
the load characteristics. In such case, the following countermeasure shall be taken to
stabilize the action and protect the contents.
・Relayed by the low voltage relay and operate by connecting CR circuit (Spark Killer: SK)
(Diode etc for DC) suited for relay coil directly to relay.
* NOTE
・Add CR circuit to the load side of relay on the request.
・It may be better to add CR circuit to the contact side depending on the load.
◎ V V
Relayed relay Power supply
Power supply
(Low voltage relay)
・Contact part of relay is melted adhesively and the contacts can not work.
・High voltage is put inside of this monitor and then, electrical parts may be damaged.
・As it is big noise, the trouble action may be taken by the reckless drive of CPU.
CAUTION
・The inductive load shall not be connected in principle (Do not connect fluorescent
lamp, motor, etc.).
・When use the inductive load, make the contact amplification outside. But as the
outside relay coil belongs to the inductive load, use the relay driven by the low
voltage (within AC100V) and it is protected by an appropriate surge killer.
* As the inductive load, there are following samples.
・Revolving light ・External relay ・Buzzer ・Siren ・Fan ・Fluorescent lamp
・Motor etc.
14
4-5.Grounding
WARNING
Use the power cord(3P) or earth terminal at the rear side of this monitor
when make grounding.
Be sure to make grounding for safety purpose and to keep stable operation of
this monitor. Do not connect each wire with gas pipe. Make grounding
corresponding to D-class grounding (Grounding resistance : below 100 Ω)
4-6.Caution in wiring
CAUTION
・Do not lay down the power cable and signal cable together with motive cable.
15
4-7.Applicable cable
4-7-1.Power cable
Use the power cable supplied with the unit.
4-7-2.Output signal(4~20mA)
Use the CVVS shielded cable or equivalent.
DANGER
・Set the terminal cover after wiring to avoid electrical shock.
16
4-9.Caution in piping works
WARNING
・This unit is designed to sample the ambient gas in an atmospheric pressure.
If the excessive pressure is added to gas inlet port and outlet port, it is dangerous
that inner tubing will be disconnected and gas will be leaked.
Use this unit not to adding excessive pressure.
・Connect the tube for exhaust gas to the gas outlet (GAS OUT) and exhaust the sampled
gas to be the safety place.
CAUTION
・Do not use this monitor in a high corrosive atmosphere.
17
5.OPERATION METHOD
5-1.Preparation to start-up
Before connecting the power cord, take care of the following. If not, there will be the
danger of electric shock and the unit will be damaged.
① Make grounding
② Check that the wiring with the outer unit is made correctly.
DANGER
・Be sure to put the terminal cover after wiring to the terminal plate
not to get an electric shock.
Start-up preparation
Power on
FLOW/ESC switch
(Press approx 3 sec.)
Flow rate adjusting mode
ZERO/▲ switch
(Press approx 3 sec)
Zero adjusting mode
Detection TEST/▼ switch
mode (Press approx 3 sec)
Alarm test mode
CAL.switch
(Press approx 3 sec)
Span adjusting mode
18
5-3.Start-up
5-3-1.Power ON
① Make sure to place the unit with properly before power on.
② Confirm location of the power switch.
③ The mark (|) means “ON” and (○) means “OFF”.
ON
○ OFF
④ Push the power switch(|)”ON” to power on. Then, POWER light turns on and internal
pump starts working.
⑤ During initial clear (2 minutes), the following actions are carried out.
・System check
・Operation for output : Approx. 2.5mA output
・Operation for alarming.: Gas alarm(optional), trouble action(light, contact) shut-off
⑥ Check flow rate with flowmonitor. The small ball in the flowmonitor should be set between
the RED lines.
⑦ It is not in the RED lines, adjust flow rate according to 【5-3-2 Adjusting the sample flow
rate】
RI-557
PREPARATION
SYSTEM CHECK
Display alternately
Waiting
19
5-3-2.Adjusting the sample flow rate
RED lines
*NOTE
・Model RI-557 has automatic flow adjustment function. With internal flow sensor, it can
check the low flow and then enhances suction ratio of internal pump. However, it alarms
“FAIL FLOW” on LCD when it cannot get enough flow even after automatic flow
adjustment.
・In this case, as the flow ratio was set maximum value by automatic flow adjustment, you
cannot enter the Flow adjusting mode. Then, check the flow pass and filter if there are
clogging and turn OFF the unit. You can enter the flow adjusting mode with the unit
turn ON again. If you could not recover, contact our nearest agent or Riken Keiki.
5-4.Explanation of performance
5-4-1.Display action
20
*NOTE
・In order to explain the indication of LED lights, we use following notation and abbreviation.
Working conditions of LCD and LED will be different from the status of the unit.
Minus indication
! CAUTION
・Zero adjustment is necessary to operate this unit properly when LCD display shows
“-0.000vol%” according to 【5-5-1 Zero adjustment】.
Alarm
When the unit detects the target gas which ● ◎ ◎ ○
concentration is higher than WARNING or PW AL1 AL2 MNT CO2: 3.500 vol%
ALARM, these lights will be blinking. (Example)
*NOTE
・WARNING (1st alarm) and ALARM (2nd alarm) are optional functions.
Failure
LCD display shows following messages if
some trouble would be happened. ◎ ○ ○ ○
FAIL SENSOR
<LCD display>・・・・<Causes> PW AL1 AL2 MNT
FAIL SYSTEM ・・・・ System failure (Example)
21
5-4-2.External output action
4-20mA version
(1) Signal transmission method : Electric current transmission(Max loop load 300Ω)
(2) Signal level
① Detection mode : 4-22mA(Depends on gas concentration)…(0-1V for 0-1V version)
② Initial clear : 2.5mA(Fixed)………………………………(0V for 0-1V version)
③ Zero/Span adjustment mode : 2.5mA(Fixed)………………………………(0V for 0-1V version)
④ Alarm test : 4-22mA(Depends on gas concentration)…(0-1V for 0-1V version)
⑤ Failure alarm : 0.5mA(Fixed)………………………………(0V for 0-1V version)
⑥ Power OFF : 0mA…………………………………………(0V for 0-1V version)
Relations between gas concentration and signal output are shown below:
Signal output
22mA
20mA
4mA
2.5mA
Gas concentration
0
Full scale
Detection mode
Initial clear /Zero, Span adjustment
mode
! CAUTION
・4~20mA signal output has been adjusted before delivery. Do not attempt to adjust
this output.
Contact our authorized service agent if re-adjustment after installation is required.
22
5-5.Adjustments
5-5-1.Zero adjustment
CAUTION
・ Zero point will not be stable just after power on(During warm-up operation).
Do not adjust zero during warm-up operation, you cannot adjust it correctly.
Zero adjustment shall be done after warm-up operation (More than one hour).
・ When performing zero adjustment with zero gas, carry out it after reading gets stable.
・ Zero adjustment shall be done in gas free places(fresh atmosphere)
・ Normal adjustment cannot be performed if the detector is condensed.
・ Use the designated calibration gas.
・ Confirm the remainder when canned calibration gas is used (Use the canned gas
containing above 0.1MPa)
*NOTE
23
④ LCD display shows current gas
concentration with blinking. Apply zero gas CO2: 0.010 vol%
and wait until the reading gets stable. (Example)
*NOTE
In case of using canned gas C
y
→ Apply zero gas as indicated the figure.
l
i
n
d
Calibration gas inlet
e
r
Inlet
24
* NOTE
When you use this unit first time or you have not used this unit for rather long term,
zero point might be shifted. In such case, The unit automatically goes to fundamental
adjustment mode from zero adjustment mode. See below instruction.
・Keep on applying zero gas and wait until the reading gets
stable. After that, adjust the reading between 2950 mV VPP: 2734 mV
and 3050 mV by pressing ▲ / ▼ switch. Wait for (Example)
approx. 3 minutes until this adjusted value become
stable.
Press SET switch, and both zero adjustment and VPP: 3000 mV
fundamental adjustment are completed. The unit is (Example)
back to normal detector mode.
25
5-5-2.Span adjustment
Span adjustment is done in span adjustment mode. Perform this adjustment periodically or if
the sensitivity would be changed.
CAUTION
・ Span adjustment shall be done with calibration gas after the reading gets stable.
・ Use the designated calibration gas for this adjustment.
・ Confirm the remainder when cannned calibration gas is used (Use the canned gas
containing above 0.1MPa.)
*NOTE
26
④Current gas concentration is shown
with blinking. Apply calibration gas. SPAN: 2.500 vol%
(Example)
* NOTE
When canned gas is used
→ Apply the gas from calibration gas inlet.
Then, supply the gas in 2~4 times every
↑
2~3 seconds as shown below graph. I
n ↓Adjustment
d
i
c
a
t
i
o
n
1 2 3 4 Time
Apply canned gas
27
When SET switch is pressed in atmosphere without supplying the calibration gas.
28
5-6.Alarm test(Confirming a transmission status)
This section is used to confirm the transmission status to the outer equipment by giving
same gas concentration signal output. Alarm will be activated when test level exceeds preset
alarm point if the unit provides an alarm function.
! WARNING
・When make alarm test (transmission test), announce it to respective department
beforehand. Carry it out after making proper treatment (External signal output
and alarm contact.)
① When finish this operation, check the external output and action of outer equipment to
be connected with external alarm contact.
② Turn power switch of this unit to off position (○).
③ Stop the power supply to this unit.
! WARNING
・When finish this operation, do it after making proper treatment such as point skip
with the upper system(Centralized system) etc.
・In case of detecting high concentration gas, high adsorptive gases or alcohol, turn off
the power after sampling fresh air for about 5 minutes and making cleaning inside
the unit.
29
6. KIND OF ALARM AND ITS ACTION
6-1.Kind of alarm
There are two kinds of gas alarm (optional) and trouble alarm.
(1) Gas alarm (WARNING, ALARM …Optional)
When the detection gas reaches to preset alarm level or exceeds it, this starts to active.
* NOTE
・ We will set alarm action(WARNING, ALARM) on request.
To prevent error of performance, it is required to set an alarm delay time
separately.
6-2.Alarm action
(1) Gas alarm (Optional)
When alarm function is provided, WARNING light(yellow) and /or ALARM light(red) turns
on. The standard version is not set an alarm.
WARNING light and ALARM light are turned off in normal operation. When gas
concentration exceeds preset alarm level, this light flashes and starts gas alarm action.
* NOTE
・ Alarm action is reset automatically when gas concentration decreases under
preset alarm level.
・ This unit has no audible alarm.
“Alarm pattern”
G
a AL2 Alarm level (ALARM)
s
c
o AL1 Alarm level (WARNING)
n
c
e
n
0
t
r
a
t
i
Alarm light (AL1)
o
n
Alarm contact(Terminal 1,2)
警報遅延時
Alarm delay time (3sec.)
Alarm light (AL2)
Alarm contact(Terminal 3,4)
Alarm delay time
(3sec.)
30
(2) Trouble alarm
When the unit detects abnormality, POWER light (Green) flashes and relay contact for
trouble alarm works.
* NOTE
・ When the unit return to normal condition from trouble condition, it starts
again from the action (initial clear) after power on, except the system trouble
“FAIL SYSTEM.”. See 【8-1 Trouble alarm】 for details.
・ This unit has no audible alarm.
②Based on gas alarm control concentration, it keeps the safety by keeping away people
from the monitoring area.
③When gas concentration display is continued, close the gas valve and confirm that gas
concentration gets lower enough.
④Suppose that the leak gas is to remain and provide yourself with protection attire and tool
away from danger, go to the leak site and check the gas residual condition by portable gas
detector.
⑤After checking that there is no danger, the treatment for gas leak shall be taken.
31
7. MAINTENANCE CHECK
This is an important for security and safety. To maintain the security and enhance the
reliability of safety, the regular maintenance check for it shall be absolutely necessary.
7-1-1.Daily check
This is the check items carried out by customer.
① Check of flowmonitor
Check that the ball in flowmonitor of this unit is within two red lines. If not, adjust flow
rate(See 【5-3-2 Adjusting the sample flow rate】).
② Check of PW/TR light
Check that the power light is at on condition. If it is flashing, check the contents of
trouble.
③ Check of LCD reading
Check that LCD reading is zero (0) in fresh atmosphere or when zero gas is supplied.
If it is not “0”, make zero adjustment. (See 【5-5-1 Zero adjustment】)
④ Check and replacement of dust filter
Check the dirt of dust filter by eyes. If it is not within two red lines, adjust it according
to 【5-3-2 Adjusting the sample flow rate】.
Even though it cannot move up, the filth of dust filter can be considered. Replace it with
new one or check the flow path (clog, bend or crack of tubing).
32
7-2.Replacing the parts
CAUTION
・ Contact our nearest agent for replacement of sensor and other parts.
Flat-blade
screwdriver
DANGER
・ Be sure to use the designated fuse.
・ There is the fear of fire and it is very danger if undesignated fuse is used or the fuse
is sort-circuited.
・ If the fuse would be broken again after replacement of fuse, contact our nearest
agent or Riken Keiki.
33
7-3.Treatment when stopping operation or move its
installation place
7-3-1.Stopping the normal operation
See 【5-7 How to finish operation】
CAUTION
・ Do not move the unit while in operation.
Reading will be changed due to the vibration.
・ Be sure to perform gas calibration when changing the installation place.
・ Ask to our nearest agent for re-adjustments including gas calibration.
CAUTION
・ Be sure to perform zero and span adjustments when the unit is used again.
・ The re-adjustment and parts replacement shall be contacted to our nearest agent.
・ When installation place is moved from cold place to warm place, the inside of the unit will
be condensed. Operation under such condition causes malfunction or trouble.
In such case, leave the unit for a while, and then use it after inside temperature of the
unit is equilibrated with ambient temperature.
34
7-5.Recommendable spare parts for regular maintenance
Check Replacement
No. Description Q’ty/unit
interval interval
1 Pump (Excluding diaphragm) 6 months 1~3 years 1 pce
* NOTE
・ The replacement interval will change depending on operating condition and it does
not mean the guarantee period.
・ Replacement of printed circuit board is required due to the deterioration of
capacitor.
・ For printed circuit board, use pre-adjusted one.
35
8.RESPONDING TO THE ABNORMAL CASE
8-1.Trouble alarm
Flashing of PW/TR light
・When PW/TR light is flashing, it is the condition that the following trouble may be
caused.
Research its cause and contact our nearest agent.
・In case of trouble alarm, trouble alarm contact activates and 0.5mA(fix) signal is
output.
When any trouble was found on this unit, contact our nearest agent or Riken Keiki.
If trouble should take place, we try best to search its cause and prevent its cause.
8-3.Before it is thought to be trouble
It may be originated from disconnection and short circuit between units.
・Check the wirings and periphery of related equipment to be connected with this
unit.
There is the case that operation of the unit is not correct including piping work and
installation work etc.
・Read operation manual and re-check operation procedure. Contact our nearest
agent if there is still unclear point.
・This will be treated by the re-check of stand-by battery, electrical line filter,
insulation transformer etc or those addition.
36
8-4.Troubleshooting
* NOTE
Rating of the fuse for this unit is 2A.
37
(5) The unit has a indefinite action
・ Is there outside noise or not ?
→ Turn off the power and turn on the power again for re-start of operation.
When such takes place oftentimes, reduce the outside noise by referring 【4-4
Caution in system engineering】.
* NOTE
・ The fuse rating for this unit is 2A (Normal fusing).
38
9. DEFINITION OF TERMS
Flowmonitor
This is a monitor to confirm the sampled gas flow.
Output signal
Signals in general being output from this unit.
・ Gas concentration signal : Electric current 4~20mA (or 0~1V as optional)
・ Status signal : Electric current 4~20mA (or 0~1V as optional)
・ Alarm contact : Relay contacts for WARNING, ALARM and TROUBLE.
Dust filter
The goods to remove dust which gives bad influence to the functions by placing it in front of
the unit.
Full scale
The maximum value for detection range.
Adjustment
To adjust instrument reading to the standard gas concentration using calibration gas, etc.
Initial clear
The reading will be unstable for a few minutes after power on.
This means the initial action to prevent a fault alarm during this period.
During this period, a false signal (2.5mA for 4~20mA output or 0V for 0~1V output) is
output and alarm contact shall be cut off.
Zero suppression
This is a function to blur the environment change or the influence by interference gas.
39
10. DETECTION PRINCIPLE
Gas cell
Gas conc. display
Analog output
IR light Optical filter
source
Structure
(IR Detector)
40
11. LIST OF PRODUCT SPECIFICATIONS
11-1.Standard specifications
Model : RI-557
Detection principle : Non-Dispersive Infrared Ray Absorption (NDIR) method
Target gas : Refer to the list of detection gas and range
Detection range : Refer to the list of detection gas and range
Repeatability : Within 2% of full scale (at constant conditions)
Linearity : Within 3% of full scale (at constant conditions)
Response time : Within 30 sec to get 90% of final reading (from the sample gas inlet
of the unit).
Detection method : Automatic sample drawing by internal pump.
Sample flow rate : 0.8~1.2L/min
Sample tubing connection : Sample side φ6 ID tubing (Gas inlet φ7)
Exhaust side φ5 ID tubing (Gas outlet φ6)
Gas alarm (Optional) : 2 levels (WARNING/ALARM), adjustable between 20% of full scale
and full scale, non-latching.
Relay contact : Dry contact, normally-open or normally-close contact for each step,
rated 0.4A at 125VAC/2A at 30VDC(Resistive load).
Display functions :
Gas concentration ……16 digit LCD
Status indication ……… 16 digit LCD and LED lights
① POWER (Green) : Normal…ON, Trouble…Flashing
② WARNING(Yellow) : Normal…OFF, Warning…Flashing
③ ALARM (Red) : Normal…OFF, Alarming…Flashing
④ MAINTENANCE (Green) : Zero adjusting …Flashing
Span adjusting…Flashing
Pump OFF … Lights ON
(During zero/span adjustments)
Self-diagnostic function : System failure, sensor connection failure, zero point failure.
Trouble alarm : Flashing of POWER light, display of trouble contents.
Relay contact : Not provided or Provided(Optional), dry contact, normally-open or
normally-close(Optional)
Rated 0.4A at 125VAC/2A at 30VDC(Resistive load)
Analog output : 4~20mA(non-isolated, max signal load 300Ω) or
0~1V(non-isolated).
Warm-up time : Approx 60 min after power on (Available to detect the gas 2
minutes later after power on)
Power requirement : 100VAC±10%, 50/60Hz, Max 25VA or 220V±10%,
50/60Hz, Max35VA.
Operating temp & humidity: 0~40℃, below 90%RH (Non-condensing)
Dimensions : Approx 220(W) x 200(H) x 320(D) mm (Excluding projection)
Weight : Approx 5.7kg.
41
11-2.Product compositions
Standard compositions
・ The main unit(RI-557) ……………………… 1 unit
・ Standard accessories
Power cable …………………… 1 pce
Operating instruction manual …… 1 vol.
Inspection certificate …………… 1 sheet
Rubber tubing (1m) ……………… 1 pce
・ Optional accessories
Dust filter
Fuse (2A)
Gas sampling bag
Protective dust filter (for the inside of the unit)
Protective filter (for the inside of the unit)
Flame arrestor