Sulfatos PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Sulfate DOC316.53.

01316

SulfaVer 4 Method1 Method 10248


2 to 70, 20 to 700, 200 to 7000 mg/L SO42– Powder Pillows
Scope and application: For oil and gas field waters.
1 Adapted from Standard Methods for the Examination of Water and Wastewater.

Test preparation

Instrument specific information


The table in this section shows all of the instruments that have the program for this test.
Table 1 shows sample cell and orientation requirements for reagent addition tests, such
as powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the corresponding
information for this test.
Table 1 Instrument-specific information for powder pillows
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 5000 The fill line is toward the user.
DR 3900
DR 900 The orientation mark is toward the user. 2401906

Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
Use the Standard Adjust option with each new lot of reagent for the best results.
For best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in
the test procedure to get the reagent blank value. Subtract the reagent blank value from the sample results automatically
with the reagent blank adjust option.
Filter samples that are turbid with filter paper and a funnel.
Do not use the Pour-Thru Cell or sipper module (for applicable instruments) with this test.
The reagents that are used in this test contain barium chloride. Collect the reacted samples for proper disposal.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used and use any recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Use the Safety Data Sheets for disposal
information for unused reagents. Consult the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.

1
Items to collect
Description Quantity
®
SulfaVer 4 Reagent Powder Pillows, 10-mL 1
Sample cells (For information about sample cells, adapters or light shields, refer to Instrument
1
specific information on page 1.)

Refer to Consumables and replacement items on page 5 for reorder information.

Sample collection and storage


• Collect samples in clean glass or plastic bottles.
• To preserve samples for later analysis, keep the samples at or below 6 °C (43 °F) for
up to 28 days.
• Let the sample temperature increase to room temperature before analysis.

Powder pillow procedure

Start

1. Start program 2. Add the sample volume 3. If the sample volume is 4. Swirl to mix.
680 Sulfate. For information that is specified for the test less than 10-mL add
about sample cells, range to a sample cell: deionized water to the 10-
adapters or light shields, mL line.
refer to Instrument specific • 2–70 mg/L: 10 mL For the dilution factor, refer
information on page 1. • 20–700 mg/L: 1.0 mL to Set the dilution factor
Note: Although the program • 200–7,000 mg/L: on page 3.
name may vary between 0.1 mL
instruments, the program
Use a TenSette Pipet or
number does not change.
glass pipet to measure
0.1 mL or 1.0 mL.

Zero

5. Clean the blank. 6. Insert the blank into the 7. Push ZERO. The display 8. Add the contents of one
cell holder. shows 0 mg/L SO42–. SulfaVer 4 Reagent Powder
Pillow to the sample cell.
The sample will get cloudy if
sulfate is present in the
sample.

2 Sulfate, SulfaVer 4 (multi-range: 70, 700, 7000 mg/L)


9. Swirl the sample cell to 10. Start the instrument 11. Clean the prepared 12. Within five minutes after
mix. Undissolved powder timer. A 5-minute reaction sample. the timer expires, insert the
will not affect accuracy. time starts. prepared sample into the
Do not move the sample cell cell holder.
during the reaction period.

Read

13. Push READ. Results 14. Clean the sample cell


show in mg/L SO42–. immediately after each test
with soap, water and a
brush.

Interferences
Interfering substance Interference level
Barium Interferes at all levels. The higher the barium concentration when compared to the sulfate
concentration, the higher the error. Samples with high barium concentrations will generally give a
result that is 20% lower than the actual sulfate concentration.
Calcium More than 20,000 mg/L as CaCO3
Chloride More than 40,000 mg/L as Cl–
Magnesium More than 10,000 mg/L as CaCO3
Silica More than 500 mg/L SiO2

Set the dilution factor


Instruments that have a dilution factor option can include the dilution factor in the result
and show the concentration of the original, undiluted sample. For example, if the sample
is diluted by a factor of 10, the instrument multiplies the result by 10 and shows the
calculated result in the instrument display.

1. Select Options>More>Dilution factor from the instrument menu.


Note: Colorimeters include a dilution factor when the chemical form is set. Go to
Options>Advanced Options>Chemical Form and select LR, MR or HR.
2. Enter the dilution factor:
• 1 mL sample diluted to 10 mL: dilution factor is 10.
• 0.1 mL sample diluted to 10 mL: dilution factor is 100.

Sulfate, SulfaVer 4 (multi-range: 70, 700, 7000 mg/L) 3


3. Push OK to confirm. Push OK again.
4. Push RETURN to go back to the measurement screen.

Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• Sulfate Ampule Standard Solution, 2500 mg/L sulfate
• Ampule breaker
• Pipet, TenSette®, 0.1–1.0 mL and tips
• Mixing cylinders (3), 25 mL

1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 25-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.

Standard solution method


Use the standard solution method to validate the test procedure, reagents and
instrument.
Items to collect:
• Sulfate standard solution, 1000 mg/L
• 100-mL volumetric flask, Class A
• 5-mL volumetric pipet, Class A and pipet filler
• Deionized water

1. Prepare a 50 mg/L sulfate standard solution as follows:


a. Use a pipet to add 5.0 mL of 1000 mg/L sulfate standard solution into the
volumetric flask.
b. Dilute to the mark with deionized water. Mix well. Prepare this solution daily.
2. Use the test procedure to measure the concentration of the prepared standard
solution.
3. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are slight
variations in the reagents or instruments.

4 Sulfate, SulfaVer 4 (multi-range: 70, 700, 7000 mg/L)


Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users may get different
results under different test conditions.
Program Standard Precision (95% Confidence Interval) Sensitivity
Concentration change per 0.010 Abs change
680 40 mg/L SO42– 30–50 mg/L SO42– 0.4 mg/L SO42–

Summary of method
Sulfate ions in the sample react with barium in the SulfaVer 4 and form a precipitate of
barium sulfate. The amount of turbidity formed is proportional to the sulfate concentration.
The measurement wavelength is 450 nm for spectrophotometers or 520 nm for
colorimeters.
Pollution prevention and waste management
Reacted samples contain barium and must be disposed of as a hazardous waste.
Dispose of reacted solutions according to local, state and federal regulations.
Consumables and replacement items
Required reagents

Description Quantity/test Unit Item no.


®
SulfaVer 4 Reagent Powder Pillows, 10-mL 1 100/pkg 2106769

Recommended standards

Description Unit Item no.

Sulfate Standard Solution, 1000-mg/L as SO4 500 mL 2175749


Sulfate Standard Solution, 2500-mg/L, 10-mL Ampules as SO4 16/pkg 1425210

Optional reagents and apparatus

Description Unit Item no.

Cylinder, mixing, 25-mL each 189640


Cylinder, mixing, 50-mL each 189641
®
Ampule Breaker, Voluette ampules each 2196800
Pipet, volumetric 5.00-mL each 1451537
Pipet Filler 1 1465000
®
Pipet, TenSette , 0.1–1.0 mL each 1970001
Pipet tips for TenSette Pipet 1970001 50/pkg 2185696
®
Pipet, TenSette , 1.0 to 10.0 mL each 1970010
Pipet tips for TenSette Pipet 1970010 50/pkg 2199796
Flask, volumetric, Class A, 100-mL each 1457442

Sulfate, SulfaVer 4 (multi-range: 70, 700, 7000 mg/L) 5


FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY
In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – [email protected] FAX: (970) 669-2932

© Hach Company/Hach Lange GmbH, 1989–2013. All rights reserved. 04/2013, Edition 8

You might also like