Idexx Vetlab Ua Analyzer: Operator'S Guide
Idexx Vetlab Ua Analyzer: Operator'S Guide
Operator’s Guide
English Version
Proprietary Rights Notice
Information in this document is subject to change without notice. Companies, names and data used
in examples are fictitious unless otherwise noted. No part of this document may be reproduced or
transmitted in any form or by any means, electronic, mechanical or otherwise, for any purpose, without
the express written permission of IDEXX Laboratories. IDEXX may have patents or pending patent
applications, trademarks, copyrights or other intellectual or industrial property rights covering this
document or subject matter in this document. The furnishing of this document does not give a license
to these property rights except as expressly provided in any written license agreement from IDEXX
Laboratories.
*IDEXX VetLab and IDEXX UA are trademarks or registered trademarks of IDEXX Laboratories, Inc. or its
affiliates in the United States and/or other countries. All other product and company names and logos are
trademarks of their respective holders.
Preface................................................................................................................................................7
Safety Precautions.................................................................................................................................. 7
International Symbol Descriptions.......................................................................................................... 8
D Maintenance . .............................................................................................................................25
Overview............................................................................................................................................... 25
Cleaning the Test Strip Tray ................................................................................................................. 25
Calibrating the IDEXX VetLab UA Analyzer ......................................................................................... 26
Cleaning the Case................................................................................................................................ 27
Replacing the Printer Paper.................................................................................................................. 27
Upgrading the Software....................................................................................................................... 27
E Troubleshooting .........................................................................................................................29
Analyzer Notifications........................................................................................................................... 29
F Appendix.....................................................................................................................................35
Technical Specifications....................................................................................................................... 35
Safety Information................................................................................................................................. 36
IDEXX Technical Support Contact Information..................................................................................... 37
Preface
Safety Precautions
The analyzer does not contain any user-serviceable components. DO NOT disassemble.
DO NOT stack other equipment or containers on top of the analyzer.
Keep analyzer away from sources of heat or flames.
DO NOT place or operate the analyzer near x-ray equipment, photocopiers, or other devices that
generate static or magnetic fields.
PROTECT your equipment from damp conditions or wet weather.
Take care not to spill water or other fluids on the unit.
DO NOT use any of the following liquids, abrasives, or aerosol sprays on or near the analyzer, as
they may damage the outer case and may influence results:
• Organic solvents
• Ammonia-based cleaners
• Ink markers
• Sprays/disinfectants containing volatile liquids (except isopropyl alcohol)
• Insecticides
• Polish
• Room freshener
Use only the power adapter supplied (which is marked as supplying 7.5 V DC, 3000 mA). The
cords from other power adapters may fit in the power socket of the UA analyzer, but may have
the incorrect voltage or current for this analyzer. Using a different power adapter than the one
supplied may impact results or render the analyzer unusable.
Disconnect the power cable:
• If the cable becomes frayed or otherwise damaged.
• If anything is spilled onto the equipment.
• If your equipment is exposed to excessive moisture.
• If your equipment is dropped or the case has been damaged.
• If you suspect that your analyzer needs service or repair.
• Whenever you clean the case.
If the equipment is used in a manner other than specified, the protection provided by the
equipment may be impaired.
2
joints No reutilizarw
Achtung, Begleitdokumente Non riutilizzare
beachten 再利用しないでください。
Attenzione, consultare la
documentazione allegata
Precaución, consultar la
documentación adjunta
注意、添付文書をご参照ください。
Other Symbols
Symbol Description Symbol Description
USB symbol Ethernet/network symbol
Introduction
The IDEXX VetLab* UA* Analyzer (UA analyzer) is a reflectance photometer designed to read and
evaluate IDEXX UA Strips for canine, feline and equine urine only. The UA analyzer reads the test
strips, saves the results to memory and prints them on its own internal printer.
The UA analyzer eliminates factors known to affect the visual evaluation of urine test strips, such
as variable lighting conditions at the workplace, people’s individual skills at matching colors and
corresponding concentration ranges, different reaction times for test pads on the strips, clerical
errors and color artifact imparted by some urine samples.
The IDEXX VetLab UA Analyzer, IDEXX UA* Strips, and IDEXX VetLab* UA* Calibration Strips are
designed for veterinary use only.
Cover:
Press here to open the cover, load
Display menu
printer paper, or load a software
upgrade code plug
Function keys
On/Off switch
Start button:
Serial interface: Power socket:
Starts the reading process
for connection connects the analyzer
NOTE: Press the Start button
to the IDEXX VetLab to the AC power
during any submenu process to
Test strip tray Station adapter
return the analyzer to ready-to-
IDEXX UA Strip 5-pin DIN socket
measure mode.
(not used)
How it Works
A prepared IDEXX UA Strip is placed on the test strip tray, which moves it into the analyzer and
under the reading head. The analyzer reads the gray reference pad first, and then reads each of
the test pads on the strip. Results are available in approximately 80 seconds.
IMPORTANT: Only IDEXX UA Strips can be used with the IDEXX VetLab UA Analyzer.
The analyzer contains light-emitting diodes (LEDs) that emit light at various wavelengths. Reading
is done electro-optically, as follows:
Phototransistor A/D converter
(analog to digital) Microprocessor
LED
Test pad
Overview
The IDEXX VetLab UA Analyzer’s display menu and function keys provide an easy-to-use interface
(Figure 7). The first line of the display menu provides system status and user information. The
second line of the display menu provides options for the three function keys.
Figure 12
Overview
Some of the features on the IDEXX VetLab UA Analyzer have factory-default settings. This chapter
describes how to customize these settings to meet the needs of your practice.
To Clear Memory
The Clear Memory option lets you clear all results from memory.
1. Press Menu.
2. Press Memory.
3. Press Clear and then press Yes to confirm. After the memory has been cleared, the “No
Results Stored” message appears when you press the Memory key.
Printing Settings
The Printer options let you determine what is printed immediately after a test strip is read.
NOTE: If your analyzer is connected to the IDEXX VetLab Station and set to bidirectional mode,
results will be automatically transferred to the patient’s record and included in the patient’s
integrated diagnostic report.
1. Press Menu.
2. Press Setup 1.
3. Press Modes.
4. Press Printer.
Overview
It is recommended that you:
• Clean the test strip tray daily.
• Calibrate the analyzer weekly.
• Keep the exterior parts and surfaces of the analyzer clean.
• Upgrade software promptly.
Calibration Printout
If the calibration results are within the defined range, the “Calibration
OK” message automatically prints with the date and time, as well
as a list of reflectance values for measuring positions 1–11 for the
orange LED (middle column of values) and green LED (right column
of values) (Figure 18).
Figure 18
Figure 20
Analyzer Notifications
Notifications for the IDEXX VetLab UA Analyzer are shown on the display menu, but are not
printed. After three minutes of inactivity, the UA analyzer switches to the ready-to-measure mode.
If the situation is not resolved, the notification appears again when you press the Start button.
Note: When a notification is displayed, press the Start button to clear the notification and then
perform the application action (see below).
The test strip is not properly Repeat the test using a new
positioned on the test strip tray. IDEXX UA strip. Ensure the
test strip is properly aligned in
the middle of the tray before
processing.
For more information, see
“Reading IDEXX UA Strips” on
page 15.
The urine on the test strip has Repeat the test using a new
dried. IDEXX UA strip.
For more information, see
“Reading IDEXX UA Strips” on
page 15.
The pads on the test strip were Repeat the test using a new
not completely moistened with IDEXX UA strip. Ensure that all
urine. of the test pads are completely
moistened with urine.
For more information, see
“Reading IDEXX UA Strips” on
page 15.
E4 Calibration Error Calibration values differ from Repeat the calibration with
those obtained in the last valid a new IDEXX VetLab UA
calibration. Calibration Strip.
For more information, see
“Calibrating the IDEXX VetLab
UA Analyzer” on page 26.
E5 Calibration Invalid Calibration values are out of Turn off the analyzer and remove
tolerance. the test strip tray. Carefully clean
and dry the tray (for instructions,
see “Cleaning the Test Strip Tray”
on page 25).
If the notification persists,
contact IDEXX Technical
Support to obtain a replacement
test strip tray.
For assistance, contact IDEXX Technical Support (see “IDEXX Technical Support Contact
Information” on pages 37–38).
Technical Specifications
Dimensions
Width: approximately 6 in (150 mm)
Depth: approximately 11.5 in (290 mm)
Height: approximately 3.7 in (94 mm)
Weight: approximately 1.7 lb (0.8 kg)
Power supply
External power adapter, Model SA 125A-0735U-S (Sino-American)
Input: 100–240 V AC, 50/60 Hz, 800 mA
Output: 7.5 V DC, 3000 mA
Consumption
Operating: maximum 15 W
Standby: 1.3 W
System description
Type: reflectance photometer
Light source: six LEDs
Wavelengths: 565 nm (green) 3x
610 nm (orange) 3x
Reader head: 1 head with six LEDs
Measuring cycle: approximately 80 seconds
Max. throughput: approximately 45 strips/hour
Printer: thermal printer
Display: liquid crystal display, two lines of 24 characters each
Memory: 100 samples
Date/time: integrated clock
Operating conditions
Operating Storage
Temperature: 59°–90°F (15°–32°C) -4°–158°F (-20°–70°C)
Relative humidity: 20%–80% 20%–85%
Telephone/Voice Mail:
The Netherlands
IDEXX Europe B.V. Spain
Scorpius 60 Building F IDEXX Laboratorios, S.L.
2132 LR Hoofddorp c/ Plom, n° 2–8, 3°
The Netherlands 08038 Barcelona
Spain
Toll-Free Technical Support....... 00800 1234 3399
Telephone................................... (31) 23 558 7000 Toll-Free Technical Support....... 00800 1234 3399
Toll-Free Fax............................... 00800 1234 3333 Telephone................................... (34) 93 414 26 08
Fax............................................... (31) 23 558 7233 Fax............................................... (34) 93 414 74 78
www.idexx.nl www.idexx.es
Germany
IDEXX GmbH Australia
Mörikestraße 28/3 IDEXX Laboratories Pty. Ltd.
D-71636 Ludwigsburg Metro Centre
Germany Unit 20, 30-46 South Street
Rydalmere, New South Wales 2116
Toll-Free Technical Support..... 00800 1234 3399 Australia
Fax............................................. 0800 6645627
www.idexx.de Toll-Free Technical Support....... 1300 44 33 99
Telephone................................... (61) 2 9898 7300
Fax............................................... (61) 2 9898 7302
www.idexx.com.au
Japan
IDEXX Laboratories KK
3-3-7 Kitano, Mitaka-shi, Tokyo
181-8608 Japan
Toll-Free Technical Support....... 0120-71-4921
Telephone................................... (81) 422 71 4921
Fax............................................... (81) 422 71 4922
www.idexx.co.jp