The Right Care in The Right Moment.: Advance

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

advance

The right care


in the right moment.

0434
Technical Data advance
and Specifications
Operative Modes Other Features And Controls
Adults And Pediatrics 12” color touch screen
VCV Volume Control (Assisted/Controlled) Sighs (invcv), Trends (up to 72 hours)
PCV Pressure Control (Assisted/Controlled) PF, PV, FV LOOPS. They can be saved as reference loops.
PSV Pressure Support
Suction %O2: for suction sequence with variable FiO2
CPAP Continuous Positive Airway Pressure
Synchronized Nebulizer
SIMV (VCV) + PSV
SIMV (PCV) + PSV Manual Inspiration, Inspiratory/Expiratory Pause (manual)
MMV + PSV Mandatory Minute Ventilation Standby function
PSV + Tidal Volume Assured Inspiratory relief valve (antisuffocation), Pneumatic safety valve: 120 cmH2O (±5)
APRV Airway Pressure Release Ventilation Complementary Functions
PRVC Pressure Regulated Volume Control
NIV Non-Invasive Ventilation Altitude compensation for volume correction
Neonates And Infants Body temperature volume correction (BTPS)
VCV Volume Control (Assisted/Controlled) Volume correction according to patient circuit compliance
PCV Pressure Control (Assisted/Controlled) Leak compensation available in all operative's modes
PSV Pressure Support Endotracheal or tracheotomy tube compensation: compensation of 10%-100%
CPAP Continuous Positive Airway Pressure for Ø 4-12 mm
SIMV (VCV) + PSV Tidal Volume Setting based on Ideal Body Weight (IBW)
SIMV (PCV) + PSV Intra-hospital transport: facilitates the mobilization when the ventilator can only be
TCPL Time Cycled Pressure Limited supplied with oxygen bottles
SIMV (TCPL) + PSV Capnography Curves of CO2 / Time and Volumetric Capnography (CO2 / VT)
CPAP with Continuous Flow (with leak compensation for NIV)
Measurements of ETCO2 (partial pressure of CO2 at the end of expiration), and
APRV Airway Pressure Release Ventilation
PRVC Pressure Regulated Volume Control their derivatives variables (Alveolar Ventilation, Dead Space, CO2 Elimination
(VCO2), VD/VT Ratio, CO2 expired volume (VTCO2), etc. The capnograph (sensor)
Parameter Selection
is an optional accessory.
(according to operative mode and patient category)
Tidal Volume: 5-2500 mL Respiratory Mechanics
Programmable Minute Volume (MMV + PSV): 1.0-50 L/min Selection by onscreen menu:
Resulting Minute Volume: 0.01-130 L/min AutoPEEP, Trapped volume measurement
Inspiratory Time: 0.1-3 seconds (30 seconds in APRV) Dynamic and static compliance, Inspiratory and Expiratory Resistance
I:E Ratio: 5:1 - 1:599
Slow Vital Capacity (Non-forced), Occluded inspiratory effort during 100 ms (P0.1)
Respiratory Rate:
P/V Inflections Points, Maximum inspiratory pressure (Pi max)
ADL: 1-100 bpm
PED/NEO-INF: 1-150 bpm Trapped Volume Measurement, Physiological Dead Space
FiO2: 21-100% Rapid Shallow breathing index (F/VT Index), Imposed work of breathing (WOBi),
Inspiratory sensitivity: Expiratory time constant (TCexp)
Flow Triggered: 0.2-15 L/min Connectivity
Pressure Triggered: 0.5-20 cmH2O below PEEP
Expiratory sensitivity for PSV: 5%-80% of the initial peak flow, in steps of 5% RS-232C with DB-9 connector
PEEP/CPAP: 0-50 cmH2O Electrical Requirements
Controlled Pressure (PCV): 2-100 cmH2O
Support Pressure (PSV): 0-100 cmH2O Main Power: 100-240 V / 50-60 Hz. Automatic voltage switching
Inspiratory Pause (programmable in VCV): 0-2 seconds Internal Battery: 10,8 V / 4,8 Ah Automatic recharge.
Inspiratory Flow Waveform (in VCV): Rectangular and Descending Ramp Estimated duration: 2 hours when fully charged. Charge level indicator onscreen
Inspiratory Flow (resultant): 0.2-180 L/min Pneumatic Requirements
Continuous Flow (NEO-INF): 2-40 L/min
Limited Pressure in TCPL (NEO-INF): 3-70 cmH2O Gas supply:
Maximum pressure limited (safety limits): up to 120 cmH2O - Oxygen: Pressure 3.5-7 bar (approx 50-100 psi) Connector: DISS 9/16"-18
Alarms - Air: Pressure 3.5-7 bar (approx 50-100 psi) Connector: DISS 3/4"-16
Automatic gas switching when one of them is absent in order to allow patient
Light and audible signals according to priority and messages on the screen. The
ventilation with the remaining gas
system keeps a record of the occurred events with name, date, and time. This
record is printable and cannot be deleted. The system allows the deactivation of: Order Information
Tidal Volume and Minute Volume alarms in NIV GraphNet advance main unit
High and Low Inspiratory Pressure Cart/Trolley
Low Pressure of O2 and Air, or one of them
Exhalation valve with flowsensor
Main Power Loss, Low Battery
Reusable adult patient circuit
High Continuous Pressure
Test lung
Technical Failure, Fan Failure
Disconnection, Leak (non-compensable) Proximal flow sensor (neonatal)
High and Low Minute Volume, High and Low Tidal Volume Optional:
High and Low O2 percentage, Oxygen not adequate Air compressor
Apnea
Capnostat 5® CO2 sensor
High Respiratory Rate, PEEP Loss
High and Low ETCO2 (optional with capnography) Humidifier

Accessories shown are not part of standard configuration.Technical data are suhject to change without notice. Manufactured by TECME, Argentina
For enquiries contact: [email protected] Website: www.schillerindia.com
Toll-Free: 1800 2098998 Authorised Dealer
Indian Corporate Office: Schiller Healthcare India Pvt. Ltd., Advance House, 2nd Floor, Makwana
Road, Off Andheri Kurla Road, Andheri (East), Mumbai 400 059 Tel: 022-61523333/29209141 Fax:
022-29209142 Email: [email protected], [email protected]
Factory: No. 1 7, Balaji Nagar, Puducherry - 605 010.
Swiss H.Q.: Schiller AG, Altgasse 68, RO. Box 1052, CH-6341 Baar, Switzerland
Regional/Branch Offices: Ahmedabad: 079-30612966/09327031678 Email: [email protected] Bengaluru: 080-26564045/
0 9 3 4 2 2 4 7 2 5 0 / 0 9 3 7 9 0 3 6 9 2 3 E m a i l : s a l e s. s o u t h 2 @ s c h i l l e r i n d i a . c o m B h o p a l : 0 7 5 5 - 4 2 7 4 4 2 1 / 0 9 3 0 3 1 2 4 9 7 3
Email: [email protected] Chennai: 044-28232648, 09383620520 Email: [email protected]
Hyderabad: 040-32952969/09379036923 Email: [email protected] Kathmandu: 00977-9841490096
Email: [email protected] Kochi: 0484-2203264/ 09383620520 Email: [email protected] Kolkata: 033-23593102/
09874766668 Email: [email protected] Lucknow: Tel: 0522-4076659/ 09312432205 E-mail: [email protected]
Mumbai: 022-61523333/09821233314 Email: [email protected] New Delhi: 011-32600582/09312432205
Email: [email protected] Puducherry: 09383620520 Email: [email protected] Pune: 020-24210657/09371020520
Email: [email protected] All registered trademarks acknowledged.

You might also like