Technical Information Indumax CLS50D/CLS50
Technical Information Indumax CLS50D/CLS50
Technical Information
Indumax CLS50D/CLS50
Inductive conductivity sensor for standard, Ex and
high-temperature applications
Application
Indumax CLS50/CLS50D is particularly suitable for use in the chemical industry and
process engineering. The six-decade measuring range and the excellent chemical
resistance properties of the materials in contact with the medium (PFA or PEEK)
make it possible to use this sensor in a wide ranges of applications, such as:
• Concentration measurement of acids and bases
• Quality monitoring of chemical products in tanks and pipes
• Phase separation of product/product mixtures
The CLS50D digital sensor is used with Liquiline CM44x or Liquiline M CM42, while
the CLS50 analog sensor is used with the Liquiline M CM42, Liquisys CLM223/253
or Mycom CLM153 transmitters.
Your benefits
• High durability
– High chemical resistance thanks to PFA coating
– PEEK version for temperatures up to 180 °C (356 °F)
• Low risk of soiling
– Dirt-repellent PFA surface
– Large sensor opening
• Easy installation
– Can be installed in pipes ≥ DN 80
– Total cable length up to 55 m (180 ft)
• Large measuring range from 2 μS/cm to 2000 mS/cm
• Integrated, coated Pt 100 temperature sensor, error class A
• Ex approval Ex ia IIC T4/T6
Other advantages of Memosens technology
• Maximum process safety
• Data security thanks to digital data transmission
• Very easy to use as sensor data saved in the sensor
• Predictive maintenance possible as sensor load data logged in the sensor
Indumax CLS50D/CLS50
1 Oscillator
1 2
2 Receiver
3 Secondary coil
4 Current flow in the medium
5 Primary coil
Advantages of inductive conductivity measurement:
• No electrodes and therefore no polarization effects
5 3 • Accurate measurement in media with a high degree of
pollution and a tendency to form buildup
• Complete galvanic isolation of the measurement and the
medium
4
A0024926
1 2
4 3
A0024929
2 Endress+Hauser
Indumax CLS50D/CLS50
CLS50
1 2
5 3
4
A0024930
Endress+Hauser 3
Indumax CLS50D/CLS50
Input
Measured variables • Conductivity
• Temperature
CLS50
Pt 100 (Class A according to IEC 60751)
4 Endress+Hauser
Indumax CLS50D/CLS50
Power supply
Electrical connection The sensor is supplied with a fixed cable. The cable between the sensor and transmitter can be
extended using the CYK11 (CLS50D) or CLK6 (CLS50) special measuring cable (does not apply for
use in a hazardous environment).
GY Shield RD
WH BU
WH U - RD
Memosens
BN U+ BU
GN Com A RD
GN
Pt 100
YE Com B WH
GY - Additional YE
PK + sensor supply BN n.c.
A0017984 A0024937
3 CYK11 for extension for CLS50D 4 CLK6 for extension for CLS50
Max. total cable length: 100 m (330 ft) Max. total cable length: 55 m (180 ft)
CLS50 only:
The residual coupling of the sensor increases when the fixed cable is extended.
Performance characteristics
Conductivity response time t95 ≤ 2 s
Maximum measured error -20 to 100 °C (-4 to 212 °F): ±(5 μS/cm + 0.5 % of reading)
> 100 °C (212 °F): ±(10 μS/cm + 0.5 % of reading)
Endress+Hauser 5
Indumax CLS50D/CLS50
Installation
Orientation When installing, align the sensor in such a way that the medium flows through the flow opening of
the sensor in the direction of medium flow. The sensor head must be completely immersed in the
medium.
1 2
max. 100
(3.94)
a
A0024950
Installation factor In confined installation conditions, the conductivity measurement is affected by the pipe walls. The
installation factor compensates for this effect. The transmitter corrects the cell constant by
multiplying by the installation factor. The value of the installation factor depends on the diameter
and the conductivity of the pipe nozzle as well as the sensor's distance to the wall. The installation
factor f can be disregarded (f = 1.00) if the distance to the wall is sufficient (a > 15 mm (0.59"), from
DN 80). If the distance to the wall is smaller, the installation factor increases for electrically
insulating pipes (f > 1) and decreases for electrically conductive pipes (f < 1). It can be measured
using calibration solutions, or a close approximation determined from the following diagram.
1.40
2
1.20
1.00
1
0.80
0 5 10 15 20 25 a [mm]
6 Endress+Hauser
Indumax CLS50D/CLS50
CLS50
To compensate residual coupling in the cable and between the two sensor coils, zero adjustment in
air ("air set") must be performed before installing the sensor. Follow the instructions provided in the
Operating Instructions of the transmitter used.
Installation with flange The sensor is suitable for installation in T-pieces ≥ DN 80, with the outgoing diameter reduced to ≥
DN 50.
LWARNING
Leakage
Risk of injury if medium escapes
‣ Tighten the sensor nut with a torque of 20 Nm.
‣ To avoid leakages, regularly check the tightness of the nut.
Flange, not in contact with medium
1 2
A0024949
7 Fixed flange, not in contact with medium (for order option: "Process connection" = 5, 6, 7)
1 Flange (stainless steel)
2 Nut
3 Sealing disk (GYLON)
4 O-ring
5 Sensor
Endress+Hauser 7
Indumax CLS50D/CLS50
1 2
4 3
A0024953
8 Fixed flange, in contact with medium (for order option: "Process connection" = 3, 4)
1 Flange (stainless steel)
2 Nut
3 O-ring
4 Sensor
1 2
5 3
4
A0024954
9 Lap joint flange, not in contact with medium (for order option: "Process connection" = A, B, C)
1 Lap joint flange (PP-GF)
2 Nut (stainless steel)
3 Flange (PVDF)
4 O-ring
5 Sensor
8 Endress+Hauser
Indumax CLS50D/CLS50
Installation in assembly
1 2 3 4
A0024960
Environment
Ambient temperature range CLS50D
-10 to +60 °C (+10 to +140 °F)
CLS50
-10 to +70 °C (+10 to +160 °F)
Degree of protection IP 68 / NEMA type 6 (sensor in installed state with genuine seal)
Process
Process temperature CLS50D
Sensor CLS50D-*1/2 CLS50D-*3/4/5/6/8 CLS50D-*7 CLS50D-*A/B/C
material Without flange DN50, ANSI 2" JIS PVDF lap joint flange
CLS50
Sensor CLS50-*1/2 CLS50-*3/4/5/6/8 CLS50-*7 CLS50-*A/B/C
material Without flange DN50, ANSI 2" JIS PVDF lap joint flange
Endress+Hauser 9
Indumax CLS50D/CLS50
Process pressure (absolute) Max. 21 bar (305 psi), depending on the sensor version, see pressure-temperature ratings
Temperature-pressure CLS50D
ratings
p (abs.)
PFA Ex
[psi] [bar]
1
305 21
2 3
247 17
4
160 11
5
15 1
T
110 125 [°C]
–20 0 +20 40 60 80 100
–4 +23 +68 230 257 [°F]
A0024981
CLS50
p (abs.)
[psi] [bar]
1
305 21
2 3
247 17
4 6
160 11
5
15 1
T
125
–20 0 +20 40 60 80 100 140 160 180 [°C]
–4 +23 +68 257 356 [°F]
A0024979
10 Endress+Hauser
Indumax CLS50D/CLS50
Mechanical construction
Dimensions
SW/AF 36
35 (1.38)
95 (3.74)
NPT 1"
G3/4
SW / AF 46
151 (5.94) / 153 (6.02)*
Ø 39 (1.54)
Ø 23.3 (0.92)
Ø 24 (0.94)*
Ø4
7.5
(1.8
7)
9)
(0.5 *
1 5.0 0.61)
Ø (
5.5
Ø1 27 (1.06)
A0024984 A0024985
13 Version with G¾ thread, dimensions in mm 14 Version with NPT 1" thread, dimensions in mm
(inch) (inch)
* Dimension for PEEK version
Process connections
Endress+Hauser 11
Indumax CLS50D/CLS50
Flange dimensions
1 2
b
d2 d2
a a
Øk Øk
D D
A0024986
15 Flange dimensions
1 Lap joint flange (PP-GF)
2 Fixed flange (stainless steel)
Dimensions in mm
Lap joint flange PP-GF DN50 PN10 ANSI 2" 150 lbs JIS 10K 50A
d2 4 x 18 8 x 19 4 x 19
b 18 18 18
a 78 78 78
Dimensions in mm
Fixed flange SS 316 L DN50 PN10 ANSI 2" 300 lbs JIS 10K 50A
d2 4 x 18 8 x 19 4 x 19
b 18 22.2 16
a 27 27 27
12 Endress+Hauser
Indumax CLS50D/CLS50
Endress+Hauser 13
Indumax CLS50D/CLS50
Ordering information
Product page www.endress.com/cls50d
www.endress.com/cls50
Product Configurator The navigation area is located on the right of the product page.
1. Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.
2. Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
For CLS50
Measuring cable CLK6
• Extension cable for inductive conductivity sensors, for extension via VBM junction box
• Sold by the meter, order number: 71183688
VBM
• Junction box for cable extension
• 10 terminal strips
• Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
• Material: aluminum
• Degree of protection: IP 65
• Order numbers
– Cable entries Pg 13.5 : 50003987
– Cable entries NPT ½": 51500177
Dipfit CLA140
• For the CLS50/CLS50D inductive sensor
• Immersion assembly with flange connection for very demanding processes
• Product Configurator on the product page: www.products.endress.com/cla140
Technical Information TI00196C
Flexdip CYA112
• Immersion assembly for water and wastewater
• Modular assembly system for sensors in open basins, channels and tanks
• Product Configurator on the product page: www.endress.com/cya112
Technical Information TI00432C
14 Endress+Hauser
Indumax CLS50D/CLS50
Endress+Hauser 15
www.addresses.endress.com