Calibrate-New-IV-Set-on-MP60 Infuse Pump Medcaptain MP-60
Calibrate-New-IV-Set-on-MP60 Infuse Pump Medcaptain MP-60
Calibrate-New-IV-Set-on-MP60 Infuse Pump Medcaptain MP-60
4). Click [ON] key, then [Enter] key will be activated as shown below.
Step 2. Calibrate (add) new brands IV set
Take calibrating (adding) a B.Braun’s 20d/ml infusion set of model 123 for example.
Preparation:
1) A new IV set to be customized, connecting it to an IV bag with more than 50ml liquid
(pure water or saline solution).
2) A measuring cup graduation in ml (accurate to 1ml)
2). The screen will show as below after above operations. Click [Producer], [Model], [Size]
one by one to enter the IV set information.
3).Click [NEXT]→[NEXT]→[CAL.] after editing [Producer], [Model] and [Size] to enter the
screen shown as below.
Note: Don’t forget to return to the Maintenance screen to switch the Technical Maintenance
off.