D 2149 - 97 - Rdixndktotc - PDF
D 2149 - 97 - Rdixndktotc - PDF
D 2149 - 97 - Rdixndktotc - PDF
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
D 2149
long. Precious metal alloys such as type B silver-magnesium- electrodes may be useful. Additional information on the
nickel have better overall properties but require special heat suitability of various electrode systems may be found in Test
treating. Methods D 150.
5.3.2 The insulators may be aluminum oxide, beryllium
oxide, or polytetrafluoroethylene. 7. Sampling
5.3.3 Use electrodes 50 mm in diameter and at least 5 mm 7.1 See ASTM standards for specific materials.
thick, with sharp corners. Maintain electrode parallellism to
within 0.01 mm. 8. Test Specimen
5.3.4 Select a length and cross-section for the lower tube so 8.1 Use a disk test specimen with a diameter of 40.006 0.01
that the temperature of each insulator does not exceed 100°C mm and a thickness of 2 to 3 mm. Finish the surfaces to 1.8 µm
when the oven is at 500°C. Select a length and crosssection for or better and maintain parallel surfaces to within 0.01 mm. The
the upper tube so that the drive nut can be touched with the samples should be free of bubbles and other defects.
operator’s fingers (keep the drive nut less than 60°C) when the
oven is at 500°C. 9. Standard Test Frequencies
5.3.5 Use a micrometer or dial gage with a precision of 9.1 Unless otherwise specified, make measurements at one
0.005 mm to determine electrode separation and to monitor or more of the following frequencies:
specimen expansion. 60 Hz 100 000 Hz
100 Hz 1 MHz
6. Electrodes 400 Hz 10 MHz
1000 Hz
6.1 Prior to measurement, apply conducting film or foil
electrodes to both flat surfaces of the specimen. (The specimen Common test frequencies are 60 Hz, 1000 Hz, and 1 MHz.
thickness should be determined before applying electrodes.)
Silver paint, tin or tin-lead foil, or evaporated metal electrodes 10. Temperature Control
have ranges of usefulness. Evaporated metal electrodes are the 10.1 Take measurements at frequent temperature intervals
most suitable. When the specimen is porous sprayed-on metal (not to exceed 20°C), until the required temperature range has
2
D 2149
been traversed. Reduce the temperature to the lowest required Ds 5 C t/Cs ~Di 2 Dv! (3)
test temperature and leave until equilibrium has been achieved.
Determine equilibrium by clamping a specimen between the where:
Co 5 capacitance of the specimen holder with the specimen
holder electrodes and balancing or peaking the measuring
out,
circuit until no change takes place between balances made 2
Ci 5 capacitance of the electrodes set at the average
min apart. After the required measurements have been made at
measured thickness of the specimen (Note 2),
the lowest test temperature increase the temperature at the rate Cv 5 equivalent geometric vacuum capacitance of the
of 2 6 0.5°C/min to the next test temperature. Follow this specimen,
procedure for achieving the test temperature until the required Ct 5 total capacitance at the unknown terminals of the
temperature range has been traversed. Take measurements at measuring circuit,
approximately the same test temperatures as the temperature is Di 5 dissipation factor of the measuring circuit as indicated
increasing and as the temperature is decreasing. Measurements by the measuring circuit when the specimen is be-
as temperature is being increased and decreased are necessary tween the electrodes, and
to guard against possible hysteresis in electrical properties due Do 5 dissipation factor of the measuring circuit as indicated
to such factors as moisture and chemical change. by the measuring circuit when the circuit has been
rebalanced with the specimen removed from the
11. Conditioning electrodes.
11.1 Prior to applying electrodes condition the specimens at
NOTE 2—If the secondary electrodes are quite thin and the maximum
23 6 1°C and 50 6 2 % relative humidity for a minimum of 40
thickness of the specimen is close to the average thickness, this setting can
h. Carry out room-temperature tests in the Standard Laboratory be considered the same as the micrometer reading when the specimen is
Atmosphere of 23 6 1°C and 50 6 2 % relative humidity. clamped between the electrodes.
12. Procedure A (Using Micrometer Electrode) 14.2 Procedure B—Calculate the capacitance of the speci-
men as follows:
12.1 Refer to Test Methods D 150. Center the specimen
between the electrodes and rotate the drive nut until the friction Cs 5 DC 1 C RT 2 CT 1 Cv (4)
is felt to suddenly decrease. Read this micrometer setting and where:
check it against the setting at which the friction first increases CRT 5 capacitance of the precision capacitor at room
on increasing the electrode spacing. Balance or peak the temperature when the measuring circuit is bal-
measuring circuit. Open the electrodes and remove the speci- anced, and
men. Then restore the balance of the measuring apparatus CT 5 capacitance of the precision capacitor at a tempera-
without changing its capacitance setting by reducing the ture test point when the measuring circuit is bal-
spacing between the electrodes and adjusting the measuring anced.
circuit to balance the loss component. Note the new dissipation 14.3 Calculate the dissipation factor as in Procedure A (Eq
factor and micrometer setting, etc. 3).
12.2 At each test temperature and each required frequency
determine the capacitance and dissipation factor of each 15. Lead Length Correction
specimen. 15.1 In both Procedures A and B keep the length of the leads
to a minimum between the measuring circuit and the specimen
13. Procedure B (Using Precision Capacitor)
holder. The DC from Procedure A will be correct, but the
13.1 Procedure B can be used when the frequency can be dissipation factors as seen by the measuring instrument and the
kept constant or when the measuring circuit, as is the case with change from DC as in Procedure B will be in error if the leads
the bridges, is stable with frequency changes. In this procedure are long. The amount of error will depend on the frequency,
determine the DC at room temperature for each frequency lead length, and the capacitance of the specimen. To correct for
required as in Procedure A. Then center and clamp the the lead error it is necessary to calibrate the measuring circuit
specimen between the electrodes and change the temperature to by calibrating the specimen holder at low frequency and using
the first temperature, taking measurements at each required its capacitance ( CH) to calibrate the measuring circuit at higher
frequency to determine the change in capacitance of the frequencies. A typical curve is shown in Fig. 2. This type of
specimen. curve at the measuring frequency can be used to correct for
13.2 Procedure B requires a variable-precision capacitor lead errors. The change in capacitance in Procedure B can be
with a precision of 0.01 pf in parallel with the specimen holder directly corrected by the curve Cm versus CH. The dissipation
to determine the change in specimen capacitance with tempera- factors from both Procedures A and B can be corrected by the
ture and frequency. following expression if the dissipation factor of the specimen is
less than 0.10:
14. Calculation
Ds 5 ~C t/Cs!~Dt 2 Do!~C H/Cm!2 (5)
14.1 Procedure A—Calculate the capacitance, Cs, and dis-
sipation factor, Ds, of the specimen as follows: where:
Cs 5 C o 2 C t 1 C v (1) CH 5 capacitance of the specimen holder, and
Cm 5 capacitance indicated by the measuring circuit.
Cs 5 DC 1 C v (2)
3
D 2149
If CH/Cm is not less than 0.99, no correction for lead length 16.1.6 Permittivity and dissipation factor at each tempera-
is required. ture and frequency reported.
16. Report 17. Precision and Bias
16.1 Report the following: 17.1 Precision—The precision of this test method has not
16.1.1 Description of the specimen, including name, grade, been determined.
color, and manufacturer, 17.2 Bias—The bias of this test method has not been
16.1.2 Dimensions of test specimen, determined.
16.1.3 Conditioning of test specimen and type of secondary
electrodes, 18. Keywords
16.1.4 Measuring circuit and procedure, 18.1 dielectric constant; dissipation factor; micrometer elec-
16.1.5 Test voltage, and trode; permittivity; precision capacitor; relative permittivity
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website (www.astm.org).