VXvue Operation Manual.V86b27 - EN
VXvue Operation Manual.V86b27 - EN
VXvue Operation Manual.V86b27 - EN
Contents
1. Information ....................................................................................................................................... 9
2. Instruction ....................................................................................................................................... 16
4.1 Introduction............................................................................................................................................ 34
Rev.86b27 Page 2 of 255 RA14-135-012
VXvue Operation Manual
1. Information
This section gives basic information of this manual and the product.
Disclaimer
Document Guide
Document Convention
Version Information
Revision History
1.1 Disclaimer
If the user is not fully aquainted with this manual, the system can be malfunctioned or unsuspected
problem can be happened due to carelessness. To prevent any medical accidents, the user should fully
understand the instructions of this manual before operating the program.
When the user operates this software, take the following precautions. Otherwise, problems may occur or
the product may not function correctly.
1 Roentgenography, image processing, reading of image, and data storage must be performed in
accordance with the law of each country where the program is being used. Also, the user is responsible
for protecting the privacy of image data.
2 In no event will Vieworks be liable for direct or indirect consequential damage arising out of the use of
this program. Vieworks will not be liable for loss of image data due to any reason.
3 Before using this software, be sure to read this manual thoroughly. Also, read the manuals of relevant
instruments in the system. Keep this manual where it is easily accessible.
4 Other systems and product names in this manual are the trademarks from their manufacturer.
The user is responsible for ensuring whether the detector calibration is performed right
after the software installation. Do not try to use the software if the detector calibration
has not been performed.
Contents
This document is the operation manual for VXvue users. You can install VXvue and acquire / manage
images from the detector. It is also available to improve the image quality by using various image
management and adjustment tools.
Target
This manual is intended for the engineers and end users who install and use the VIVIX-S detectors and
VXvue.
Contact Deparment
For any comments or inquiries regarding this document, contact the customer support team in Vieworks.
Item Contents
Department Customer Support Team at Vieworks
E-mail [email protected]
Trademarks
The name “Vieworks” and Vieworks logo are registered trademarks of Vieworks.
1.3.1 Symbols
This symbol is used for indicating a potentially hazardous situation while operating this
product. Users should be well acquainted with this symbol and the related contents.
Information
This symbol is used for indicating the product-related references and complementary
information.
1.3.2 Notations
Bold Types
Words in bold indicate options seen directly within the GUI (graphical user interface) of the VXvue
program.
1.0.0.85b7 1.0.4.4 2.0.0.6 1.3 01.03 01.01 01.18 3.33 1.32 - - - 3.28 01.15 3.28
1.0.0.85b7 1.0.4.4 2.0.0.6 1.3 01.03 01.01 01.18 3.33 1.32 - - - 3.28 01.15 3.28
1.0.0.86b6 1.0.4.11 2.0.0.7 1.4 01.06 01.04 01.21 3.38 1.38 - - - 3.35 01.15 3.35
The versions above are meant to be used for document revision only. They do not
guarantee compatibility across all referenced models.
2. Instruction
Overview
Recommended Specifications
2.1 Overview
VXvue is the image viewer program for VIVIX-S only that acquiring, adjusting and managing images
generated from the VIVIX-S detector. Besides VXvue Viewer, VXvue is organized with several specialized
modules. This system provides high quality medical images to users through various advanced functions
and quick image processing between the modules.
HS8 and PureImpact™ are the image processing algorithm developed by Vieworks.
This section provides information about the recommended specifications for installing and using VXvue.
If the specification cannot meet the recommended one, it may affect the functions and
processing speed of VXvue.
2.2.1 Workstation
3. Browse Around
This section gives instruction on how to start and close VXvue and its basic components.
Preparation
Starting Up
Basic Composition
Registration Mode
Exposure Mode
Database Mode
Viewer Mode
Print Mode
Setting Mode
Closing
3.1 Preparation
Before running VXvue, you should be ready to use the system by prechecking its operating condition and
environments.
Refer to VIVIX-S Detector Operation Manual for preparation before installing VXvue and
for checking status after installing VXvue.
Without performing calibration, it is not possible to start VXvue or get images. Refer to
VIVIX-S Setup Operation Manual for the detailed information about calibration and the
device connection.
Checking License
You need to register the license when you first execute VXvue and XIPL. Make sure to be fully aware of
the related data in advance to issue the online license.
3.2 Starting Up
1 Execute VXvue.
2 Wait until VXvue loads completely.
A new license window can be displayed if you change specifications of the PC where
VXvue is installed.
Contact the person in charge of Vieworks for the way of license issue and input.
3.2.3 Login
1 Task area
All tasks can be processed in the task area. This area has different compositions by each mode, and can be
switched when you select the desired mode tab.
The Registration mode provides functions for registering and managing the patient information.
1 Panel area
The Registration mode is divided into three panels (Scheduled, Unscheduled and Reserved) according to
the registration way of patient information. You can choose the desired panel and register patient
information.
You can make an exposure from the Exposure mode. The Exposure mode supports a general exposure
and macro exposure modes.
General exposure mode – Set the preparation steps to make an exposure.
Macro exposure mode – Make an exposure and adjust the image by magnifying the image view area.
7 Thumbnail area
Check the thumbnail image acquired from the thumbnail area. If you click the thumbnail image, it will
display on the image view area.
4 Thumbnail area
Check the thumbnail image list acquired from the thumbnail area. When you select the thumbnail image, it
is displayed on the image view area.
Name Description
Step Used for indicating and changing steps of the image.
Manupulation Tools for manipulating the image.
Layout Tools for changing layout of the image display.
Annotation Tools for inserting annotations on the image.
Tools Tools for processing the image.
You can check and manage the closed study list / images from the Database mode.
4 Thumbnail area
Check the thumbnail image of the selected study from the thumbnail list.
You can check and adjust the selected images from the Viewer mode. The organization of Viewer mode is
very similar to the macro exposure mode in the Exposure mode.
3 Thumbnail area
Check thumbnail images of the selected study.
You can set the print and adjust an image to be printed from the Print mode.
You can configure various values from the Setting mode. This mode provides 10 setting groups.
1 Panel area
Panel Description
System Sets the system information.
Registration Sets the information releated to the patient registration.
Display Sets the information releated to the Viewer mode.
Tool Sets the information of tools.
Study Sets items for managing studies
Procedure Sets the exposure procedure information.
Integration Sets items for integrating a detector with equipment.
DICOM Sets the DICOM information.
Backup/Reject Checks and restores the backup and deleted information of images.
Account Default Sets an account-related default to use when adding another account
2 Setting area
When you select a desired panel, the screen is switched to the related setting window.
Refer to VXvue Service Manual for more information about the setting areas.
3.10 Closing
Select Logoff button to open the login window to log into VXvue with another account.
Select Shut down button to close the system.
4. User eXperience
This chapter provides information about the improved UX (User eXperience) of VXvue.
Introduction
Skin (Theme)
Layout of Sapphire Blue Skin
Touch-Oriented
Auto Rotation of UI
4.1 Introduction
VXvue provides various UX (User eXperiece) and GUI (Graphic User Interface) for user convenience. With
these functions, the user can build an optimal environment suitable for one’s purpose by using a touch
screen, virtual keyboard and tablet PC away from the conventional keyboard, mouse, monitor and the
general PC environment.
UX Description
VXvue supports the skin (theme) function. You can choose a skin suitable for
Skin (Theme)
the use environment to use the brand new VXvue.
VXvue has various layouts. User can use VXvue with the layout suitable for
Layout
the use environment.
VXvue supports the touch function. If the monitor in use supports a touch
Touch-orientation
screen function, you can select a suitable skin (theme) from VXvue to use.
VXvue supports virtual keyboard. You can input data by touching the virtual
Virtual Keyboard
keyboard displayed on the screen.
VXvue has a layout for a vertical monitor under the HD resolution
Auto Rotation of UI
(1920x1080). The layout changes when the monitor is rotated.
The skin, touch screen and virtual keyboard can be used by installing VXvue to a
Windows based tablet PC which supports these functions. Be sure to ask the person in
charge of Vieworks for the models and specifications of tablet PC suitable for VXvue.
Skin Description
Onyx Classic Default skin
Sapphire Blue Suitable for touch operation.
Provides bigger button and font size than those of Sapphire Blue to fit for a tablet
Sapphire Blue-QHD
PC with small monitor.
Sapphair Blue is a designed skin for touch operation. The size of GUI such as icons is changed for easy
touch as well as designed a user-friendly way.
Sapphire Blue skin is a very simple flat design with a blue background and white GUI. As the icons and
thumbnails are turned to orange color by selecting those items, it is easy to understand their current
status. In addition, the size of selected icons gets bigger for making touch operation easy.
Icon Thumbnail
All functions of Sapphair Blue skin are same as the Onyx Classic skin, except the design.
You can change the skin from Setting System System Info. menu.
Screen Image
Before
expanding
the search
screen
After
expanding
the search
screen
The displaying layout of Exposure mode is changed automatically depending on the screen resolution.
Default Layout
Both the step and view areas are displayed.
Wide Layout
If the step and image view areas cannot be displayed together at the currently set resolution, they will be
displayed separately.
You need to arrange the Exposure mode effectively as this mode is consisted of various
areas for making the exposure. In VXvue, the layout of each area is changed
automatically depending on the screen resolution. You can also change the layout with
using shortcuts like F5, F6, F7, and F8.
Refer to <10.5.2 Shortcuts> for the detailed information about shortcuts.
4.4 Touch-Oriented
If you use devices like a monitor and tablet PC support a touch screen, you can use the touch function in
VXvue.
Although the touch function is supported to all skin themes, the Sapphire Blue skin is
designed to be more suitable for using the function.
VXvue supports a virtual keyboard function. This keyboard is displayed when you choose a text box for
entering data. Touch the virtual keyboard to inputting data directly.
The virtual keyboard function uses On Screen Keyboard (OSK) provided by Windows.
Install Windows Service Pack 1 or the above. If not, the process speed of virtual
keyboard can be lowered.
Set Use virtual keyboard to Yes from Setting System Theme menu to use the
virtual keyboard.
Pin Zoom
VXvue supports the Pin Zoom function, allows the user to zoom in or out the image conveniently.
2 Touch image display region on the screen for 0.5 second, than the screen changes to Pan mode after the
status of mode conversion is indicated.
4 Lift yout finger from the screen to return to the previous mode.
Select the desired icons to display from Setting Tool Image Tool menu.
The color of selected icons becomes blue.
Check the selected icons from Tools panel in the Exposure mode and the Viewer mode.
Click on Detail button to find the hidden icons.
Set the user-defined icon group by collecting frequently-used icons from Setting Tool Quick
Access menu.
The color of selected icons becomes blue.
You can set the displayed order of selected icons.
Check the selected icons from Process panel in the Exposure mode and the Viewer mode.
Check the selected icons from user-defined tools area in the Exposure mode.
The Step area in the Exposure mode is provided as two types – Image type and Button type.
Click F8 shortcut in the Exposure mode or change settings in the Setting mode to switch each type in
real time.
The user-defined text buttons are supported which are the pre-
defined ones used frequently.
The user interface of VXvue is rotated automatically when the screen resolution of a device such as a
tablet is changed by rotating it.
If the vertical resolution is bigger than the horizontal one when you rotate a device, its screen is changed
as a vertical layout automatically.
VXvue provides a vertical layout when vertical resolution of the screen is bigger than horizontal resolution.
Vertical Layout is applied normally when horizontal resolution of the screen is more than
1080.
This section provides information about registering the patient prescribing information.
You can bring the patient information from the Scheduled panel by using DICOM Worklist.
1 Input or choose search conditions and click on the Search button.
2 Check the searched Scheduled list and choose the desired item.
3 Process the next step by selecting one of the tool buttons in the Scheduled panel.
DICOM Worklist server should be configured to use Worklist. Refer to MWL memu in
VXvue Service Manual for setting the Worklist server.
Station Name
Choose search condition of Station Name by checking the box.
The items of search condition can be added or deleted from
Setting Registration Scheduled.
Modality
You can select search condition of Modality by checking the box.
Select one of the three options - DX, CR, DR.
Station AE Title
You can select search condition of Station AE Title by checking
the box.
The items of search condition can be added or deleted from
Setting Registration Scheduled.
Go to Setting DICOM General to set the default value of Station Name and search
condition of Station AE Title.
Scheduled Date
Search
Click on the Search button to start search after setting the search conditions.
All study data is searched if you search the data under the initialized state.
The Patient and scheduling information that are searched for in the following three conditions appear in
the checklist.
If the search condition is hidden in Sapphire Blue skin, the Scheduled Start Date and End
Date are not initialized.
Click the searched item to check its scheduled procedure items from the following list.
Headers of the procedure list cannot be modified.
The scheduled item will not be indicated if it does not include a procedure.
Shows or hides the list of procedures that display the Scheduled Procedure entry for the selected item at
the bottom of the screen.
4 Show SPS List: All patient information and schedule information appears.
5 Hide SPS List: Only the patient information with images and the schedule information are displayed.
If the searched item is mapped with a procedure code of VXvue, the steps of mapped procedure for
shooting will be showed as follows.
Refer to <10.4 Managing the List Header and Field> for more information about headers
and fields of the searched list.
Check Auto Refresh to automatically renew the patient information transmitted from DICOM
Worklist Server.
Set the interval for renewing patient information from Setting mode DICOM tab MWL.
Press the ESC key to stop searching Scheduled (while the progress circle is rotated.).
The searched result before stopping search will display on the screen.
Add to Reserved
Save the patient information to Reserved in advance, and use the data when you cannot
use the Scheduled item.
1 Select the Scheduled item to be saved to Reserved. You can select multiple items using
Ctrl or Shift key.
2 Click on the Add to Local button and save the selected items to the local database.
3 Check the saved items from the Reserved tab.
Start Study
1 Selet the scheduled items to make an exposure.You can select multiple items using Ctrl or
Shift key. When you select multiple items, the 1st selected item will start being exposed.
2 Click on the Start to Study button to start taking images.
The exposure will be started with the patient and prescribing information of the selected
Scheduled item.
The screen is switched to the Exposure mode automatically.
Refer to <10.4 Managing the List Header and Field> for more information about headers
and fields of the searched list.
Refer to <VXvue Portable Mode> guide for the information about wireless exposure.
Refer to <10.2 Emergency> for more information about the emergency exposure.
Multi select
Click on the Multi Select button to select multiple Scheduled items.
You can input the patient information from the Unscheduled panel.
The patient information can be stored to the local database.
You can make an exposure after inputting the patient information.
You can start an emergency exposure without inputting the patient information.
You can generate accession number automatically by selecting Use auto access number-
Yes from Setting Registration General.
If you input the pre-existing patient ID, the following information will be input and
displayed automatically.
Patient ID, Patient Name, Birth Date, Sex, Patient Comments
If different patient names have a same ID, a window for patient name selection is
displayed. You can select a patient name from the window.
The input items of Patient Name are divided to five (5) itmes as follows.
Prefix, First Name, Middle Name, Last Name, Suffix
Last Name is the compulsory one represents Patient Name. You can also add the desired
name item to use.
Refer to Registration group in VXvue Service Manual for the details about setting the
patient information.
Put the mouse cursor on a month, day, year / hour, minute, second items of Scheduled
Date Time and press the arrow keys (↑↓) to change each item.
Age is calculated and input automatically when you enter Birth Date first.
The unit options of Height / Weight are Metric (kg / cm) and English (lb / ft).
If you take images after selecting Procedure, the matched Step is registered
automatically.
You can select the exposure priority from Procedure Priority.
STAT / HIGH / ROUTINE / MEDIUM / LOW
Refer to Procedure Manager in VXvue Service Manual for the detailed information about
the study procedure.
Refer to <10.2 Emergency> for more information about the emergency exposure.
From the Reserved panel, you can use the patient information reserved by clicking on the Add to reserved
button in Scheduled or Unscheduled.
Since the patient information as the text (.txt) file is parsed according to the rules for each
row and column, refer to the template files (PatientListSample.txt) by clicking on
button.
The patient information (Patient ID/ Paitent Name/..) is divided into tabs.
The unit of patient is divided into carriage characters (Enter).
You can move and edit the patient information in Excel. When the edit is completed, be
sure to import the file after changing its format to ‘.txt’.
Specify the path of text file that the patient information is recorded.
Check Use First Line to use the first line of text file as patient information.
Uncheck Use First Line to use the first line of text file as the information for
column mapping.
The function of Import Patient Order is supported for the human version of VXvue only.
Edit Patient Modifies the patient prescription information in the Reserved item.
Import Adds patient information read from a file to the Reserved item.
Patient Age
Select the age unit options and insert numbers.
Select one of the options among D (Day), M (Month), Y (Year).
Patient Sex
Select the gender of a patient.
Select one among M (Male), F (Female), O (Other).
Scheduled Date
In Sapphire Blue skin, the scheduled start date and end date are not initialized if the
search condition is hidden.
The Patient and scheduling information that are searched for in the following three conditions appear in
the checklist.
Click the searched item to check its scheduled procedure items from the following list.
Headers of the procedure list cannot be modified.
The scheduled item will not be indicated if it does not include a procedure.
Shows or hides the list of procedures that display the Scheduled Procedure entry for the selected item at
the bottom of the screen.
1 Show SPS List: All patient information and schedule information appears.
2 Hide SPS List: Only the patient information with images and the schedule information are displayed
If the searched item is mapped with a procedure code of VXvue, the steps of mapped procedure for
shooting will be showed as follows.
Select items
Select items in Reserved. You can select multiple items by using Ctrl or Shift key.
Start Study
Start acquiring an image by clicking on the Start to Study button.
The screen is switched to the Exposure mode automatically.
The exposure is started with the patient and prescribing information of the selected item.
When you select multiple items by pressing Ctrl or Shift key, the detector starts taking
images from the item that you selected first.
Refer to Procedure Manager in VXvue Service Manual for more information about the
procedure code.
Refer to <10.2 Emergency> for more information about the emergency exposure.
Multi Select
Click on the Multi Select button to choose multiple items in the Reserved.
You can send the selected items to the Exposure mode.
Export
Export the patient prescription information as a text file (.csv) by clicking the Export button.
All searched prescription information is exported.
This section gives basic information for the way to acquire images.
Item Instruction
1 Register Patient Info. Register the patient information and start a study.
Register the patient information first from the Registration mode for making exposure.
You can register the patient information at the Registration mode by three ways as follows.
Panel Description
Scheduled Import the patient information by using DICOM Worklist.
Unscheduled Input the patient information directly.
Reserved Use the patient information reserved in advance.
Refer to <5 How to Register Patient Prescribing Information> for the way of registering
patient information.
The detector icons are showed differently depending on each detector model. Refer to
<10.6.4 Detector Information Display Icon> for the information of detector icons.
You can use the Receptor icon instead of the detector icon.
The number at the right bottom of the icon indicates a turn of the detector when
you use multiple detectors.
The Receptor has four types of icon – Table, Stand, Table Top, No Bucky
Receptor Description
Table The detector is installed in the Bucky for table.
Stand The detector is installed in the Bucky for stand.
Table Top The detector is mounted on Bucky for table.
No Bucky The detector is not installed in Bucky.
The detector icon is indicated differently depending on each detector model. Refer to
<10.6.4 Detector Information Display Icon> for the detailed information about detector
icons.
Refer to Detector Menu in VXvue Service Manual for the way to change a detector icon
to a receptor icon.
6.3.2 CR Mode
In the CR mode, the following message is displayed on the image view of Exposure mode.
The CR mode icon is not indicated unless you preset the CR mode to use.
Refer to Detector Menu in VXvue Service Manual to preset whether or not to use the CR
mode.
When the detector is switched to the Sleep mode, the detector icon is changed to the
Sleep mode icon.
You cannot make an exposure if the Sleep mode is not disabled.
The Sleep mode can be used if the following conditions are all met.
When using the wireless detector.
When the wireless detector is connected wirelessly.
When the Sleep mode is activated from VXSetup.
Refer to VIVIX Setup Operation Manual for the detailed information about setting the
Sleep mode.
You cannot acquire images even the x-ray is exposed under the Sleep mode. Be sure to
check if the Sleep mode is activated before making an exposure.
A caution mark is shown on the detector icon when the operation environment of the
detector is out of the optimal range.
If you see this mark, take proper measures by checking the following situations.
1 The temperature of detector is too high or too low. (less than 5°C, more than 50°C)
2 The strength of wireless communication signal is too weak when the detetcor is operated wirelessly. (Link
quality is less than 40.)
3 In case of the wireless detector, the battery remains less than 30%.
You can check the detector status information by clicking the right mouse button on the detector icon.
You can import the images stored in the detector from VXvue.
Refer to <6.9 Importing Images from the Detector> for the detailed information.
Detector AP mode
The contents about battery information, Wifi signal strength, the number of images,
image import, and changing the AP Mode are applied to the wireless detector only.
If a detector and a PC are not connected as follows while detector AP is on, the Turn Off
button of the AP mode is not indicated in the Detector Info dialog.
Connecting the detector to SCU with a tether interface cable.
Connecting SCU to PC with a LAN cable directly.
Refer to VIVIX Setup Operation Manual for the detailed information about setting the
AP Mode.
Check the patient size to be exposed and click on the arrow button to select size
options as follows.
As the patient size is linked with the procedure step information, you can set the
exposure condition for each patient size in advance. Refer to Step Tab in VXvue Service
Manual for the information about setting steps related to a patient size.
The patient size is one of the main factors affects the exposure condition and image
quality. Process the exposure by selecting a patient size matched with the exposure
environment since the x-ray dose collected by the detector can be different according to
the patient size.
You can select a category and step suitable for the exposed bodypart of a patient after checking it.
1 Page Up / Down
Turn the pages of selection dialog of category, step, procedure or projection.
2 Category selection area
Choose the category.
3 Step or procedure selection area
Choose the step configured for each bodypart in the selected category.
4 Exposure projection selection area
Choose the exposure projection of each step.
Refer to Category Tab in VXvue Service Manual for the detailed information about
registering and setting the category.
Step is a basic unit of the exposure condition combinated with the exposed bodypart,
orientation and patient size information.
Procedure is a combinated unit with two or more different steps.
Before taking images, you should preset the step or procedure for each exposure
condition. Use the basic step and procedure in VXvue or set the step in person.
Refer to Step Tab in VXvue Service Manual for the contents about setting each procedure
and step.
When the exposure is ready, the patient and bodypart information are displayed on the image view in the
order of Patient Name, Patient ID, and Bodypart. Also, the icon is indicated in the thumbnail to
notice that the step is selected.
Receiving image
1 Start making an exposure by shooting X-ray to the object.
2 Wait until the detector receives the image.
While the image is being received, the standby icon is displayed and the reception
amount of the image is displayed.
The receiving time can be different depending on the detector connection status or the
PC use environment.
For certain models, the Cancel button will be displayed if image transmission fails or the
reception is delayed. Click this button to stop the image transmission.
Processing image
1 The preview image displays on the screen after the image is
received.
2 The image processing is started by reading the exposure step
information immediately after the preview image displayed.
3 Wait until the image is processed completely.
The standby icon shows while the image is being processed, and the preview image
displays.
When the preview image is displayed during image processing, the preview message
appears at the bottom of the image along with the processing status message.
The setting of font, size, and color of ‘Preview’ letter is the same as that of label.
(Path: Setting Display Information Overlay)
The processing speed can be different depending on the algorithm setting or the PC use
environment.
Completing exposure
Check the final image applied image processing.
Refer to <7 Image Processing> for the information about using the image processing and
adjustment tools.
Refer to <7.30 Closing Study> for the information on how to close the study.
Refer to <7.31 Suspending Study> for the information on how to hold the study.
If the image is captured with the wireless detector without connecting VXvue (Stored), or if the image
transmission fails during shooting with the wireless detector (Backup), the image is saved inside the
detector.
If image processing fails after shooting (Backup - Local), the image is saved on PC.
VXvue generates studies and imports images automatically by using the study information stored in the
detector.
If there is no study information stored in the detector, or if the image is saved in the PC, it will be
displayed as No ID, No name together with the shooting time as below.
1 Click the Import button in Detector Info to open the Import Image From Detector window as above.
2 Choose one or more studies from the list you want to import to VXvue, and click the Auto Import
button.
In case you choose a single study, select images to import by checking boxes and clicking on the Auto
Import button at the lower right.
If you select multiple studies, import all images by clicking the Auto Import button at the lower right.
When images are imported completely, they disappear from the list one by one.
When communication is broken while importing images, a message displays and the
Import Image From Detector window is closed.
In this case, check the connection status between the detector and viewer. After that,
open the Detector Info window again to import images.
We recommend you connect the detector with a tether interface cable when importing
images.
If the image is captured with the wireless detector without connecting VXvue (Stored), or if the image
transmission fails during shooting with the wireless detector (Backup), the image is saved inside the
detector.
You can match the study information when importing images stored in the detector.
If there is no study information stored in the detector, the image will be displayed as No ID, No name
together with the shooting time as below
1 Click on the Import button in Detector Info to open the Import Image From Detector window as
above.
2 Choose one or more studies from the list you want to import to VXvue, and click on the Matching
Import button.
In case you choose a single study, select images to import by checking boxes and clicking on the
Matching Import button in the lower right. Then the selection window of target study shows for
selecting the desired study.
In case you choose multiple studies, import all images by clicking on the Matching Import button in
the lower left. Match each study in the Matching Study window to import images.
The maximum number of study you can match at the same time is 10.
Click one of the four panels display on the Select Study window.
2 Scheduled panel
The patient information imported from DICOM Worklist is indicated on the study list to be matched.
Unlike the Registration mode, this panel does not provide tool buttons.
3 Unschedueld panel
You should input the patient information to the study to be matched.
Unlike the Registration mode, this panel does not provide date and time of the study as well as the
study code.
4 Reserved panel
The patient information reserved by using the Scheduled or Unscheduled menu is indicated on the
study list to be matched.
Unlike the Registration mode, this panel does not provide tool buttons.
Refer to <5 How to Register Patient Prescribing Information> for more information.
From the study list showed on the selected panel window, the study information is matched with the
items selected from the study list. Click on the Import button to import images after the study matching
is compeleted.
5 Select and input the desired study and click Import / OK buttons in order.
When you match a single study, the Import button shows and the image is imported directly.
When you match multiple studies, the OK button shows. If you click the OK button, an ‘O’ mark is
indicated on the matched item of the Matching Study window. Check if the Import button is operated
normally after all studies are matched.
6 In the Study tab of the shooting screen, check if the image is imported successfully after matching the
study.
Refer to <6.3.5 Checking the Detector Information> for the information of Auto Import by
using detector information.
How to Set
You can determine whether or not to use the FTM mode from VIVIX Setup and VXvue.
For more information, refer to <6.9.1 Auto Import> and <7.29 Getting Full-Sized Image>.
FTM mode is supported from F/W V1.0.3.31 for FXRD-1012N detector and F/W V1.0.0.4
for VXTD-2532E detector.
7. Image Processing
This section gives instruction on how to use the image processing tools.
Image View
Thumbail Images
Panel and Buttons
Image Processing Tools
Selecting Image, Adjusting Image Size, Moving Image
Magnifying the Region of Interest
Adjusting Brightness and Contrast of Image
Cropping, Inverting, Rotating Image
Using the Image Layout
Using Annotations and Labels
Stitching Images
Processing the Image
Copying the Image
Changing Target E.I.
Correcting Grid Lines
Changing Order of Images
Using Tools for Print
Using Full View
Suspending Study
You can check the image information by indicating it in the overlaid area. Each Overlay can be assigned on
the top left / top right and at the bottom left / bottom right of the image as you want.
Refer to Information Overlay Menu in VXvue Service Manual for setting Image Overlay.
7.1.2 Scale
The length displayed on the image is configured on the basis of the length measured
from a detector surface. This means that a common difference can be occurred between
the actual length and the measured one, because the actual exposure part has uniform
height from the detector surface. For this reason, the length measured from Viewer
should be used as a reference data only. Vieworks has no responsibility for the result
caused by the measured length information.
Thumbnail information
You can check step information for shooting from the thumbnail.
You can check the series and image number of the thumbnail.
You can check status icons of the thumbnail.
For details on acquiring Full Image, refer to <6.9 Importing Images from the Detector>
and <7.29 Getting Full-Sized Image>.
FTM mode is supported from F/W V1.0.3.31 for FXRD-1012N detector and F/W V1.0.0.4
for VXTD-2532E detector.
Restoring Image
The deleted image is indicated as a red color, and is available
to be restored.
4 Double click the thumbnail of a deleted image.
5 Click on the Yes button on the following message that asking
whether you will restore the image or not.
6 Check the restored image.
The Reject Thumbnail function is supported to the Exposure and Viewer modes only.
You can remove an image in the study unit from the Database mode (Reject Study).
Grouping Thumbnails
Click on the Group Thumbnail button to group the images in the study for each body
part.
Click on the title button of group to show or hide a group.
The order change of thumbnail image is available in the same series only.
The order of images cannot be changed of multiple thumbnails are selected.
If the order of thumbnail image is changed, its instance number is also changed.
Some backed-up images are limited to apply this function.
Selecting Thumbnail
Click on the thumbail you want to choose.
Move arrow keys (←,→, ↑,↓) of the keyboard to choose a thumbnail.
← → Keys
↑↓Keys
Retaking Images
Choose the image to retake and click on the Retake Thumbnail button. A pop-
up window displays for selecting a reason to remove the image. Choose one of
the reasons to remove image, and then a same step for retaking image is added.
The added step for retaking image will be removed when you restore the rejected image.
Before you retake images, uncheck the ‘Use reject reason’ menu from Setting > Study >
Rejected Reason menu to stop displaying ‘Use Reject reason’.
You can change the tools used in the thumbnail list from the Setting mode. Refer to
Thumbnail Setting in VXvue Service Manual for the detailed information.
The Tools panel provides tools for adjusting and controlling images. You can click on the desired tools to
add them as user-defined image processing ones.
Checking Steps
From the displaying region of Step, you can check
the step name of the study or image which is
currently selected.
The selected image means that it is processed by
the current step information.
Changing Steps
1 From the displaying region of Step, click
on the Change Step button to open Step
List window.
2 Choose Bodaypart, Projection and
Patient Size that you want to change.
3 Click OK button to change the step
information of the image. After that, The
image is processed again by the changed
step information.
Refer to <7.5 Using Basic Image Processing Tools> for the detailed information of using
the default tools for image processing.
You can check the user-defined image processing tools in the Exposure mode and Viewer mode.
You can register or delete the image processing tools directly.
With the tools, you can easily process the image.
Multi Layout Displays images in the layout configured in the Setting tab.
Distance Measures distance by drawing a line between two points on the image.
Draws a line between the two points of the image to measure twisted
V.Disc.
part (up-and-down).
Rect ROI Assigns the region of interest of image to measure its pixel information.
Print Sends the image to the Print mode for DICOM printing.
Close Study Saves and close the study in checking among the opened studies.
You can view the original image (RAW) without applying detector
Raw
rotation information and image processing.
Processes the image variously and detailedly as well as manage the
Image Process
image processing parameters.
Grid Suppression
Corrects vertical grid lines.
(Vertical)
Grid Suppression
Corrects horizontal grid lines.
(Horizontal)
VXvue provides the basic tools for image processing. You can secure the satisfactory image quality by
using the tools.
The basic image processing tools predefines three types of parameter for each Step. (Normal, Soft,
Hard) You can improve the image quality by selecting Step and image processing items freely.
The basic image processing tools provide the basic elements of image processing: Contrast, Sharpness,
and Brightness. For images with PureImpact™ image processing, additional Sharpening Size, Tone Type
and Noise Reduction adjustments are available. You can check the image quality while using the
adjustment function directly.
The basic image processing tool provides Post LUT information in HS8 image processed image. If the
image is processed with PureImpact™, the Post LUT selection screen is not provided. You can view the
image quality results by changing the Post LUT directly.
VXvue provides three parameters for each step. (Normal, Soft, Hard)
You can select and apply the desired parameter to the image for securing the optimal image quality.
Parameter Description
Normal Basic image processing parameter
Soft Processes the image more softly than Normal.
Hard Processes the image more sharply than Normal.
Besides Normal, Soft, Hard, all parameters can be configured from XIPL.TOOL.
Refer to XIPL User Manual for the detailed information about parameter settings.
The Contrast, Sharpness, Brightness, Sharpening Size, Tone Type, Noise Reduction and
Post LUT values you change in the Process tab are not stored in the parameters.
Please use XIPL.TOOL to save the changed value to parameters.
Refer to XIPL User Manual for the detailed information about the meaning of each
parameter as well as setting and storing it.
You can choose the image / annotations partailly (Select) or choose all images of a
study. (Select All).
When you click on this button, the active function will be deactivated and turned
back to the basic state.
Only the following functions can be applied to the multi-selected images simultaneously.
Invert function (Invert, Mirror, Flip)
Rotation function (Clockwise, Counter clockwise)
Size control function (Fit, 1:1)
Add label function (Add R label, Add L label)
Specialized functions (Image Process, Send, Print, Stitch, Convert Raw, Reject Image)
The other functions are applied to the single image only.
If you choose the other functions, the multi-selected images are cancelled and only one
image will be selected automatically.
How to use
1 Click on the Zoom button on the Tools panel.
2 Move the mouse cursor to the image screen then the cursor will be changed to magic glass shape.
3 Press and drag the left mouse button on the image to control zooming.
Drag the mouse upwards to enlarge image.
Drag the mouse downwards to reduce image.
Close
1 Release the mouse button to finish zooming.
How to use
1 Click on the Fit button on the Tools panel.
2 Check if the image is fit to the main image window.
How to use
1 Click on the Real Size button on the Tools panel.
2 Check if the image is fit to the actual size of the object displayed on the monitor.
You can measure the object size on the monitor with a ruler.
Refer to Monitor Correction Menu in VXvue Service Manual for setting pixel information
of the monitor.
The pixel spacing correction should be performed to display exact size of the image.
For more information, refer to <7.26 Correcting Pixel Spacing of Image> or <11.8.2
Detector Menu>.
Pixel Spacing Correction: As the physical pixel pitch is fixed, measure a magnification for correcting pixel
spacing as follows;
1 Measure the length of an object.
2 Mesure the length of an object projected on the image under the specific SID and OID conditions.
Terms
Real Size: Calculate the object’s real length of the desired part through the magnification calculated by
pixel spacing correction, and display the image in real size.
Image Pixel Spacing Correction: Applies the result of pixel spacing correction for single exposed
images.
Detector Pixel Spacing Correction: Applies the result of pixel spacing correction for all images acquired
from each detector.
Magnification: Image size of the object / Object size = SID/(SID-OID)
Type Description
You can move an image to the desired position with the Pan button.
How to use
1 Click on the Pan button on the Tools panel.
2 Move to the main image window and check if the cursor is changed to a hand shape.
3 Drag the image to the desired position by holding down the left mouse button on the image.
Close
1 Release the mouse button to finish.
You can magnify a selected region of interest with the Magic Glass button.
Annotation on the image is also magnified.
How to use
1 Click on the Magic Glass button on the Tools panel.
2 Move to the main image window and check if the cursor shape is changed as a magnifying glass.
3 Press and hold the left mouse button and drag the region of interest on the image.
You can check the region of interest enlarged twice than the surrounding area.
Close
1 Release the mouse button to finish.
Close
1 Release the mouse button to finish.
Decrease W2
You can set the increase or decrease direction / speed of W/L as you want. W/L Refer to
Display Group in VXvue Service Manual for setting W/L speed and direction.
Histogram
The histogram is indicated labels which are distributed on the image by mapping them with same value.
Select menu
1 Display the pop-up menu by clicking the right mouse button in
the shutter area.
2 Choose either Crop Type or Crop Rect Size menu.
Crop menu
Rectangle
Ellipse
Polygon
Polygon crop
1 Select Polygon and click the starting point that you want to crop.
2 Make the polygon shape by clicking the desired points.
3 Double-click the last point to finish.
The starting and last points are connected.
Click and drag the green dot line with a mouse button to adjust the shutter size.
Touch and drag a gray circle outside of the shutter area to adjust shutter size.
Click or touch an orange circle in the shutter area and drag it to move to the desired
location.
Close
After adjusting the shutter size, click on the Select button to finish crop.
The cropped image will be fit to the main image window automatically.
Refer to Integration Group in VXvue Service Manual for setting a shutter size of the
popup menu.
You can invert the image as black / white, top / bottom and left/right.
How to use
Click on the Invert button on the Tools panel and check if the black / white of an image is inverted.
Click on the Flip button on the Tools panel and check if the top / bottom of an image is inverted.
Click on the Mirror button on the Tools panel and check if the left / right of an image is inverted.
You can rotate the image as clockwise or counter-clockwise way with CW and
CCW buttons.
How to use
Click on the CW button in the Tools panel and check if the image is turned clockwise by 90˚.
Click on the CCW button in the Tools panel and check if the image is turned counter-clockwise by 90˚.
Use the free rotation button to rotate the image freely or by the reference line.
The shutter area and annotation are rotated with the image.
Rotate by Line
1 Click on the Line button on Tools panel.
2 Draw a reference line by clicking and dragging the desired point. After that, adjust direction and degree
to rotate the image.
3 Release the mouse button and check if the image is rotated by the appointed direction and degree.
Rotate freely
4 Click on the Free button on Tools panel.
5 Click and drag the desired point in clockwise or in counter-clockwise to adjust degree to rotate the
image.
The roation degree is based on a clockwise direction.
6 When the degree is indicated as you want, release the mouse button and check if the image is rotated.
Click on the Reset button to return the rotating image to its original one.
Refer to Layout Menu in VXvue Service Manual for the way of editing multi-layout.
You can annotate the image by using the text annotation function.
Input text in the Free Text area and click on the Add button to add
the entered annotations.
Click on the Add button to activate the entered annotation on a
mouse pointer, and click the pointer to a desired position to
annotate the image.
How to add labels by using annotation buttons of R / L label in the Tools panel
1 Click on the R label button.
2 Check if the R label is added to the left top of image.
3 Click on the L label button.
4 Check if the L label is added to the right top of image.
Refer to Information Overlay Menu in VXvue Service Manual for the information about
setting the font, color, and default size of a label.
If you move the label outside of the image area, it is not indicated when sending,
printing, or exporting the image. Therefore, it is necessary to insert the label within the
image area.
The left / right image information should be stored to the DICOM image. One of the four
values - U (Unpaired), L (Left), R (Right), B (Both) is stored to the Image Laterality tag
(0020, 0062). Please use the DICOM Viewer to check the tag information.
Refer to Display Group in VXvue Service Manual for more information about registering
Image Overlay.
A-P, R-L and H-F are sets of orientation information which is symmetrical to each other.
If you choose one orientation from the set, the rests are selected to the opposite site
automatically.
The patient orientation information should be stored to the DICOM image. One of the
five values - A (Anterior), P (Posterior), R (Right), L (Left), H (Head), F (Foot) is stored to
the Patient Orientation tag (0020, 0020). Please use the DICOM Viewer to check the tag
information.
Refer to <9.1 Stitching Images> for the detailed information about stitching.
Click on the Image Process button to call out the image processing tools - XIPL.TOOL.
After processing images from XIPL.TOOL, you can update Step information of the selected
image.
This item provides more various functions than those of the Process tab.
How to use
1 Choose the image and click on the Image Process button to call out XIPL.TOOL.
We recommend you to understand this manual and XIPL User Manual fully before trying
to process images. If you process images without understanding the aboves correctly, it
may affect the image quality.
Log on to VXvue with a service account or an account with administrative rights to use
XIPL.TOOL through the Image Process menu.
Copy a selected image to the last position of series where the selected image is included.
Check if the copied image is added to the last position of thumbnail.
When the image is copied, the created time (content time) of image is changed to the
copied time.
The image can be copied differently according to the setting of image copy options.
Refer to Tool Group in VXvue Service Manual for the information about the options.
You can change the Target E.I value configured in the step of selected image for each
patient size.
Click on the Tools > Target E.I. button to display the Target E.I.
3
dialogue.
You can check the Target E.I. value for each patient size.
If you click on the Save button, Target E.I. value of the
checking items will be changed only.
Refer to <9.5 Exposure Index> for the detailed information about the exposure index.
When you select an image, you can check the state of Grid Suppression button applied to the image.
If you change Grid Suppression button to another, the related function is applied to the image.
If HS1, HS2, HS4 algorithms are applied to the image already, the Grid Suppression icon
becomes deactivated. In this case, process the image with HS8 or PureImpact™ algorithm
and then use the Grid Suppresion function.
Type
Grid lines are corrected from XIPL and the result is returned to VXvue. Therefore, you can
directly set the processing standards of Grid Suppression from XIPL to improve image
quality. Refer to XIPL User Manual for the detailed information about Grid Suppression
and its setting method.
Some specific patterns or noise on the image can be recognized and processed as grid
lines.
The sharpness of image can be lowered because of the grid line correction.
This is the preceding function of the Real size function to display the real size of an object.
This function corrects pixel spacing generated by magnification to display the object of each
image as a real size.
1 Select an image and click on the Image Pixel Spacing button in the Tools menu.
2 When the cursor is changed to a ruler shape, click and drag a mouse button at the start point of the
object. Release the button at its end point.
Image information (Exmaple)
Object: Vernier calipers fixed as 100mm
SID: 130cm
OID: 60cm
3 When the Correction for Real Size Measurement (Image) dialog appears, enter the real length of the
object to Length of the New menu. The result of converted pixel spacing value is also indicated.
Name Description
Current Based on the detector currently connected.
Pixel Spacing The pixel spacing preset to the current detector.
A converted segment length in magnification by pixel spacing set to the current
Length
detector.
New Based on the current image.
The pixel spacing applied to the current image.
Pixel Spacing The virtual distance between pixels according to the converted magnification
by the number of pixels correspond to the segment of actual measurement.
Length Input the length of actual measurement.
4 You can measure the segment again by lengthening or shortening it after moving its position.
5 The configured pixel spacing is not applied to the next shooting.
In case of real shooting, it is easy to measure the actual length by putting the object like
a flat ruler on the same SID and OID with the subject.
The length of object for measurement is recommended 100mm or longer.
The correction result of pixel spacing is applied to the corresponding image only.
When you take an image with one detector, refer to Detector Menu in VXvue Service
Manual if it is necessary to correct pixel spacing to apply the value of magnification
correction according to the fixed SID and OID.
Use the tools for printing images from the Print mode only.
Refer to <9.2 Printing> for the detailed information about the way to use tools for print
only.
How to Use
Double click the image or click on the Full View button.
This function can be used under Zoom, Pan, Fit, and W/L menu. When you change W/L, it is applied to
the image.
Use this function to get a full-sized image of the selected preview or FTM image.
How to Use
You can select multiple FTM or preview images to import a full Image, and automatically import images
without selecting a study.
If you cannot import a full-sized preview image automatically, you can select a full size image through
the window below.
When VXvue is used for vehicle inspection, the function to automatically import images
without selecting a study is not available.
Refer to <8.15 Sending Image> for the details about automatic settings for transmitting
images and closing studies.
Refer to <7.31 Suspending Study> for the details about suspending studies.
How to use
1 Click on the Suspend Study button from Tools panel.
If steps exist for shooting, click button in the Study tab to select whether you close or suspend the
study.
This function only applies to the study being shot.
2 From the study management list, check if the study you select is suspended normally.
The suspended study can be opened or closed again from the list.
Not only the suspended study but the steps which are set to be shot can be checked from the list.
If you shot the suspended study additionally, the steps set to be shot are also added to the list.
Refer to <8.2.3 Checking the Result of Searched Study> for checking the list of
suspended studies.
Refer to <8.5 Insert Image> for instruction on how to keep shooting or for additional
shooting.
Refer to <8.20 Closing Study> for the information about closing the suspended study.
You cannot use this function if there are more than one image exist.
You cannot apply this function to the study for correcting pixel spacing.
In case of stitch step, the steps set to be shot are not stored when you suspend a study.
You can use the guide button to display the guide lines, and check the symmetry of the
image and the position of the area of interest.
Guide Lines
When cropping images (Crop mode), the guide line mode is canceled.
8. Study Management
This section describes the way to use the study management tools.
You can search the stored study list by using the study search tools.
It is available to search studies quickly and exactly by inputting various search conditions.
You can filter the search result as needed.
You can find the study search tools from the Database mode.
You can use the study data in various ways by using the study management tools.
Perform a new study by using patient information in the study data.
Modify or manage the information and images in the study data.
Check the image by connecting VXvue with QXLink if QXLink is installed normally.
Store the study data to the storage medium or print it.
Send the study data to DICOM or print is as a DICOM format.
Available to use the Statistics function.
Available to compare two seletected studies.
You can find the study management tools from the Database mode.
You can search the study data using the study search tools in the Database mode.
Eight search conditions are provided for searching studies.
AND arithmetic operation is applied to the inserted search condition.
Search
Search studies by clicking on the Search button after setting the search condition.
All study data is searched under the initialized search condition.
Refer to <10.4 Managing the List Header and Field> for the detailed information about
the search list header and field management.
Insert Image Opens the existing study and make an additional exposure.
QXLink Report Searches report of the relevant study while running QXLink 3.
Import Study Imports study and image from the storage medium.
Multi Send Sends the image by selecting the server set to Protocol List.
Print Adds images to the Print mode for making DICOM print.
QXLink 3 is the image viewer program developed by Vieworks. Refer to QXLink 3 User
Manual for the detailed information about QXLink 3.
You can start a new study by using the patient information in the selected study.
How to use
1 Choose the study to be used as patient information.
2 Click on the New Study button and check if the screen is converted to the Exposure mode.
You can make an exposure if the new study is registered successfully.
If you select more than two studies, they will be registered as new studies.
The time of registering new study is applied as Scheduled Date Time.
It is not needed to register the patient information additionally since this function uses
patient information from the selected study.
You can open the selected study again and make an additional shooting.
How to use
1 Choose the study to make an additional shooting.
2 Click on the Insert button and check if the screen is converted to the Exposure mode.
The existing study is registered again, and you can make an additional shooting.
If you select two studies or more, the existing studies are registered again as many as you select.
In case of a suspended study, it is resgistered with the images of existing studies if there are steps to
be shot.
You cannot make the additional shooting if the relevant study is in use from the other
mode.
You cannot make the additional shooting with the backed-up study.
1 Scheduled panel
The prescribing information is modified as the one brought on by using DICOM Worklist.
Unlike the Registration mode, this panel does not offer any tool buttons.
2 Unscheduled panel
You can input and modify the prescribing information directly.
You cannot modify Scheduled Date Time.
It is necessary to enter a value for the required field.
3 Reserved panel
The prescribing information is changed to the reserved one by using Schduled or Unscheduled.
Unlike the Registration mode, this panel does not offer any tool buttons.
Refer to <5 How to Register Patient Prescribing Information> for the instruction on how
to register the patient prescribing information.
If you input the registered ID, the other information in that ID is registered automatically.
If several patient information exists since Patient ID is overlapped in the database, you
can select patient information from the following pop-up window.
Refer to Registration Group in VXvue Service Manual about how to edit items from the
Edit Patient window.
You can move all or partial images of the selected study to another study.
This function can be used for combining several images to a single image, or for dividing a
study into several ones.
The study can be moved only when you use VXvue for Vehicle.
Refer to <9.4.9 Additional Functions> for the way of moving the study.
1 StudyList panel
In the Database mode, items in the study management list displayed on the window are showed only
in the target study list, except the items include the images you selected.
In the Exposure or Viewer mode, studies displayed on the window are showed only in the target study
list, except the items include the images you selected.
If this ‘Move Image’ function is used for examination, all items of the study management list in
Location displayed on the window are showed in the target study list, except the items include the
images you selected.
Refer to <9.4 Using VXvue for Vehicle> for the information about how to use this
function for examination.
2 Scheduled panel
The prescribing information brought on by using DICOM Worklist displays on the target study list.
Unlike the Registration mode, this panel does not offer any tool buttons.
3 Unscheduled panel
Input the patient prescribing information as a target study item.
Unlike the Registration mode, this panel does not offer any tool buttons.
4 Reserved panel
The patient prescribing information reserved by using Schduled or Unscheduled displays on the target
study list.
Unlike the Registration mode, this panel does not offer any tool buttons.
Refer to <5 How to Register Patient Prescribing Information> for the instruction on how
to register the patient prescribing information.
Move images to the target study item among the study lists showed on the selected panel window.
Select and input the target study item and click on the Move button.
Check if the images in all study lists are moved correctly.
You can also check the image movement from the thumbnail.
No image but the study information can be left only if all images are moved.
The images cannot be moved if the relevant study is in use from the other mode.
The backed-up image cannot be moved.
This function cannot be applied to the study for correcting pixel spacing.
This function cannot be applied to the rejected image.
This function cannot be applied to the preview and FTM images.
This function cannot be applied to the stitch step.
You can open the study from the study list and review its image.
When you open the selected study, the screen is converted to the Viewer mode
automatically.
How to use
1 Choose the study from the study list.
2 Click on the Open Study button and check if the screen is switched to the Viewer mode automatically
and the image is displayed.
You can select and open two or more studies at the same time.
In case of the suspended study, it is opened as the additional exposure mode. If there is a step to be
shot, the suspended study will be registered with the images of existing studies.
You can delete the study selected from the study list.
You cannot delete the opened study.
How to use
Choose the study from the study list and click on the Reject Study button.
You can select and reject more than two studies at the same time.
Input reason
1 Click on the Reject Study button to open the dialog for selecting
a Reject reason
2 Choose one of the reasons to reject the study.
Refer to Rejected Reason Menu in VXvue Service Manual for the information about
changing the Pre-defined reasons.
Close
Click one of the reject reasons to reject a study.
If you search studies again, the rejected one will be indicated in gray or will not be searched depeding
on the setting.
When you reject a study, the Restore button becomes activated at the same time. Click on
the Restore button to restore the rejected study.
If you choose more than two studies, the Restore button becomes deactivated.
Use this function to call QXLink 3 and use functions of QXLink 3 Viewer.
You can search images through QXLink Viewer only when the images have been sent to
QXLink Server.
Use this function to search a report at the same time when QXLink Viewer searches
images.
Set QXLink Server to check if the images have been sent to QXLink Server.
Refer to Storage Menu in VXvue Service Manual for the information of setting DICOM
Storage.
The QXLink 3 icon is not showed if QXLink 3 is not installed.
The available version of QXLink 3 which can be called from VXvue should be 3.2 or
higher.
Introducing QXLink 3
QXLink 3 is the PACS program developed by Vieworks. You can send the acquired images from VXvue to
save and manage them as well as use the various functions of QXLink 3.
Refer to QXLink 3 User Manual for the detailed information about installing and using
QXLink 3.
Use this function to export study items selected from the study list to local or to the external
storage medium.
It supports various storage medium like HDD, USB, CD, DVD, etc.
It also supports various image format conversions.
You can execute other actions while importing images. (e.g. shooting or sending images)
How to use
1 Select studies to be exported from the
study list.
You can choose serveral studies by
using Ctrl or Shift key.
2 Click on the Export button.
The Export Manager window
displays.
3 Set various options such as image
format, storage path, etc.
Click on the Options button to set
additional options.
Click on the StudyList button to
check the list of studies to be
exported.
4 Click on the Start button to export
images.
You can suspend exporting images
while converting the files.
5 You cannot change the configured
options if the exporting is stopped or
failed.
Close
1 Click on the Close button or to finish exporting the image.
If you click while exporting images, the progress window is hidden but the images continue to be
exported.
2 The Export Manager window is closed after exporting images successfully. Check the exported files from
the storage media.
While the images are being exported, the Export Manager is not closed even you close
VXvue.
A pop-up window will appear to ask if you continue to export reject / preview / FTM
images if they are included in the study list to be exported.
It is not available to export the reject / preview / FTM images.
The TEXT file is not exported when the file is exported to DICOM and DICOMDIR.
If inapplicable characters (e.g. /\:*?"<>|) are included in the patient ID, they cannot be
included in a file name.
When exporting a file, the generation rule of file name is as follows.
Common use (for hospital) : Selected path\
- Study Date]_[Study Time]_[Patient ID]_[Patient Name]_[Study Key]\[Acquisition
Date]_[Acquisition Time]_[Patient ID]_[SeriesKey]_[InstanceKey].File extension
Common use (for examination) : Selected path\
- [Location Name]_[Location Scheduled Date]_[Location Key]\[Acquisition
Date]_[Acquisition Time]_[Patient ID]_[StudyKey]_[SeriesKey]_[InstanceKey].File extension
Exception 1 – IMG : Selected path\Study_[StudyKey]/S_[SeriesKey]_I_[InstanceKey].img
Exception 2 – DICOMDIR : Selected path\I_No (‘No’ is the integer in 8 digits, increased
sequentially.)
Exception 3 – DICOM (including a portable viewer) : Selected path\dcm\Common (for
hospital or for examination)
Item Description
File Format Sets multiple image formats (more than 1 format)
BMP Bitmap image format
Compressed format for image loss developed by JPEG (Joint Photographic
JPEG
Experts Group).
TIFF image storage foramt
TIFF
Both TIFF-8 and TIFF-16 cannot be selected at the same time.
Vieworks medical image format. The exported IMG file can be imported from
IMG
VXvue installed to the other systems.
RAW Raw image data without including image-related information.
Standard format of medical image by DICOM.
DICOM
Cannot be selected with DICOM DIR at the same time.
Standard format of medical image directory by DICOM.
DICOMDIR
Cannot be selected with DICOM at the same time.
Export Path Sets the storage path where the exporting images are saved.
Selects one of the lists of optical device drive (ODD) such as a pre-
configured network drive, local drive, USB, CD, DVD, etc.
Selecting a drive If you select USB, CD, or DVD, insert the medium to the selected drive in
advance and check if it is operated normally.
The network drive must be mapped as a drive letter (C, D..).
In case of a local drive or USB, click button to apply the directoy of the
Selecting a path
selected drive.
Burning Option Sets options to be recored when exporting images.
Annotations Includes the annotation information to the image. (If checked)
Information Includes the Overlay information to the image. (If checked)
Sets whether the patient and hospital information is processed to be
Anonymous
anonymous. (If checked)
Raw Applies raw image without image processing. (If checked)
Software Collimation Sets the image area.
None Applies the image to its original size.
Cut Applies the cropped area of image only.
Fill Applies the outside area of the cropped image as a black color.
Displays the cropped image from the original image in the center, and
Fill & Center
applies the area outside the cropped image to black.
Image Modality Sets the image modality information.
Current Uses the modality information without change.
DX/CR/DR Uses the modality information by fixing it among DX/CR/DR.
The DAP value to be transferred to the DICOM image is changed according
DAP Unit Type
to the set unit.
mgy/cm2 -
ugy/m2 -
Rev.86b27 Page 162 of 255 RA14-135-012
VXvue Operation Manual
dgy/cm2 -
Language Applies DICOM Language (16 languages).
Default Not applying DICOM Language Encoding.
Status Checks the processing condition of image exporting and its result.
Item Description
Sets whether to include DICOM Portable Viewer.
Include Portable Viewer
This item activates only by checking DICOM or DICOMDIR.
Sets JPEG2000 compression way of DICOM image.
Compression
This item activates only by checking DICOM or DICOMDIR.
No Compression Not apply the JPEG2000 compression way.
JPEG2000 Lossless Compress the DICOM image as JPEG2000 without a loss.
JPEG2000 Lossy Compress the DICOM image as JPEG2000 with a loss.
You can set the JPEG2000 compression ratio of the DICOM image.
Compression Quality This item activates only when selecting JPEG2000 Lossy Compression.
You can input the compression ratio directly, or control the slide bar.
The image can be damaged seriously when the compression ratio of JPEG2000 Lossy is
higher, and this can affect to the image quality.
Start
1 Click on the Import Study button to open the Import Study window.
2 Click on button to select the location of the study file folder to be imported.
When you select the location, all study lists in the folder are showed automatically.
3 Select the desired study to be imported.
You can select multiple studies by using the Ctrl or Shift key.
Import
4 Click on the Import button.
5 Select All Image or Selected button from the Import Study window.
All Image – All studies are imported regardless of the selected item.
Selected – Only the selected studies are imported.
Cancel – Cancel importing studies.
Close
1 Click on the OK button to finish importing images.
2 Check the importing result by searching the study list again.
Refer to <9.1 Stitching Images> for the detailed information about stitching images.
You can select study or image and send it to the DICOM storage server.
Send: Send images to the server configured in SCP List.
Multi Send: Send images to the server configured in Protocol List.
Refer to Storage Menu in VXvue Service Manual for the information about SCP List and
Protocol List.
Send
1 Select the study or image to be sent.
You can select multiple studies or images with the Ctrl or Shift key.
Check the connection state of the, DICOM Storage server before sending images.
2 Click on the Send button or Multi Send button.
Select Protocol from the Storage Protocol box if you click on the Multi Send button.
3 Click on the All Image, Selected or Cancel button from the Send box.
Refer to Queue Menu in VXvue Service Manual for the detailed information about the
Queue tab.
Refer to General Menu in VXvue Service Manual for the way to set auto sending and
closing functions.
The auto sending function cannot be applied to the study for pixel spacing.
The rejected image is not included to the auto sending and closing functions.
If preview / FTM images and the step which is not shot are included in the study, the
auto closing functions cannot be used.
You can select study or image and send the image to the Print mode.
The DICOM print can be available in the Print mode only.
Refer to <9.2 Printing> for the detailed information about using the Print mode.
8.17 Statistics
Refer to <9.6 Using Statistics Function> for the detailed information about the statistics
funtion.
You can select multiple images in a study by clicking on the Multi-Select Thumbnail
button or press Ctrl and Shift.
Refer to <9.7 Comparing Studies> for the detailed information about comparing studies.
How to use
1 Choose a study which is on hold from the list.
2 Click on the Close Study button to check if the selected study is closed.
The image is sent and stored to external storage media (External Save) depending on the settings at
the time of closing the suspended study.
In the DICOM server, the closed study is classified as the one being taken.
All suspended studies are closed when VXvue closes.
9. Advanced Functions
This section provides instruction for using the advanced functions of VXvue.
Stitching Images
Printing
Using VXvue for Vet
Using VXvue for Vehicle
Exposure Index
Using Statistics Function
Stitching images is a function that connecting two or more images as a unified one.
You can stitch up to 6 images.
The stitched image cannot be connected to other images again.
Only the same size images can be stitched together.
When stitching images which have different brightness and contrast, the conditions are adjusted
automatically.
When adding steps after stitching, the stitched image is added as a new image.
This function provides two type of stitching – Physical Matching (auto stitch) and Pattern Matching
(manual stitch).
This document explains the way to use Pattern Matching.
The way to input OID for measuring actual image size is as follows.
When you create a study for the first time by using the step registered in Stitch
Protocol, the Auto Stitch Option dialog displays.
This option is used for measuring the distance between a centerline of an object and
surface of a detector (or bucky). Enter the distance in Object Distance for calculating
actual measurement and applying it to the images to be stitched.
Set OID value from the Setting mode > System > Intergration tab to calculate the
distance of actual measurement in every time you stitch images.
Besides OID, it is necessary to perform the pixel spacing of a detector to get the correct
image size. Refer to Detector Menu in VXvue Service Manual for the detailed information.
9.1.1 Start
1 Select more than two images to be stitched and then click on the Stitch Images button.
You cannot select only one image for using the stitch function.
2 After clicking on the Stitch Images button, check if the screen for stitching image displays.
The Stitch Images icon can be shifted to the Quick Access tab.
2 The original image is displayed as applying W/L which was set from the Viewer mode.
Sort Down Moves the order of the selected original image downward.
Zoom for stitched Area Enlarges the stitched part between each original image.
Blend Links the boundary between the two orgianal images softly.
Move Up
Move Right
Top
If an overlapping region is made because of the image
movement, show or hide the selected image.
Back
Selects an image.
Select
Return the mouse button back to a default status.
Auto Stitch
1 Assign the order of each original image correctly by using the buttons for changing orders.
Sort Up, Sort Down, Ascending Sort and Descending Sort button
2 Click on the Pattern Matching button and check the result.
3 Adjust the position of each original image by clicking on the following buttons.
Move Up, Move Down, Move Left and Move Right button
4 Match each original image to be overlapped properly by clicking on the Top and Back buttons.
5 Click on the Pattern Matching button to check the result again.
6 Repeat the steps 3~5 if you are not satisfied with the result of stitch.
It is not needed to adjust the stitched image additionally if you are satisfied with the first
result of Auto Stitch.
You can get the satisfying stitched image as the quality of original images is similar and
their quality is fine.
Apply Step
After checking the result of stitched image, click on the Change button to display the Step List screen.
Select a step (Bodypart, Projection) to be applied to the stitched image and click on the OK button.
Close
Click on the Save button and save the stitched image.
Check if the stitched image is added to the thumbnail as a new one.
Overlapping images
Click on the Stitch Top and Back buttons to show or hide the selected image.
Top button – Raise the selected image up to the overlapped area. (The selected image is
showed.)
Back button – Lower the selected image down to the overlapped area. (The selected
image is not showed.)
Top Back
Top Back
9.2 Printing
Click on the Print button and send the study or image to the Print mode.
Print button is located in the Exposure mode, Database mode and Viewer mode.
You can send the DICOM print from the Print mode only.
Precheck
Check if the DICOM print server information is set.
Check the connection state to the DICOM print server (ECHO).
Refer to Setting Common Items of DICOM server in VXvue Service Manual for checking
the setting and connection status of DICOM print.
Refer to Print Menu in VXvue Service Manual for the detailed information about print
settings.
Print
You can print the image set in print view by sending it to DICOM print.
Click the Print or Print &Close button to print image.
Check the process from Setting mode DICOM panel Print Queue.
After printing the image, close all images in the Print mode.
Refer to <7 Image Processing> for the detailed information about other image
adjustment tools and POST LUT.
Refer to Print Menu in VXvue Service Manual for the detailed information about print
settings.
You can print the images set in print view by sending them to the printer.
Click the Print / Print & Close button.
You can use the VXvue program not only for human but also for animals (Vet, Equine).
The followings are additional functions of VXvue for animals only.
Veterinary patient information
Veterinary Category
Veterinary Bodypart / Projection
Species Code
Breed Code
Breed Registration Code
Key word
Term Meaning
A unit of biological classification organized with a relationship of interbreeding and has
Species
common characteristic.
A group within a species that shares definable and identifiable characteristics (visual,
Breed performance, geographical, and / or cultural) which allow it to be distinguished from other
groups within the same species.
When installing VXvue, set the viewer mode as Vet or Equine if you want to use VXvue for animals.
The items of veterinary patient information are more than those of human because of various kinds of
animal.
This chapter explains only about the difference of exposure procedures and preparation
when preparing the veterinary version. As most of the other parts are the same as the
human version, please refer to <6How to Shoot Images> for the detailed information.
Refer to Step Tab in VXvue Service Manual for the detailed information about setting the
patient size for each step.
Choose a category and a bodypart to register step. After that, an auto label is indicated
on the bodypart name located in the step.
Choose the registered step and take a shot, then a label is added on the image
automatically.
Change Category
1 Click on the category text on the step selection dialog of the Exposure mode.
2 Select a category to use when the dialog for selecting a category shows.
The built-in options of VXvue for Vet are Dog, Cat, Exotics, Equine, and Pre-purchase exam.
The built-in options of VXvue for Equine are Equine and Pre-purchase exam.
3 Check if the background image is changed after selecting the category.
Refer to <6.5 Selecting Category and Step> for the detailed information about adding
and changing categories.
You can add categories for Vet to VXvue for Equine if necessary.
This chapter explains the difference of settings when the user operates VXvue as a veterinary version.
As the most contents are same as the human version, refer to Settings in VXvue Service
Manual for the detailed information about each item.
Procedure Manager Species Code tab / Breed Code tab / Breed Registration Code tab
The veterinary version provides three (3) additional tabs in the Procedure panel as follows.
Species Code tab – Species code management
Breed Code tab – Breed code management
Breed Registration Code tab – Breed Registration code management
The codes registered to the three tabs are used for registering prescribed information of studied
patients.
Tab Screen
Species Code
Breed Code
How to use
The use way of codes registered in the three (3) tabs is same.
Click on the Add button to add a code to be registered newly.
Click on the Hide button to hide the code which it not used.
Each code tab is supported the Show / Hide function, not deletion.
The item that registered once cannot be deleted.
Check the Show All checkbox to see the hidden items.
The VXvue program can be used for the patients who take an X-ray in a vehicle as well as in the
hospital.
Choose the option VXvue for Vehicle when installing VXvue to use it in a vehicle.
The composition of VXvue for Vehicle is similar to its general version mostly, but the following functions
are specialized for the examination in a vehicle.
Function for registering location and study
Serial radiography
External save function, a substitute for the DICOM sending function.
Barcode for registering and searching studies
Term Description
Vehicle A vehicle with the radiography environment
Location A place or institution for examination
To use VXvue for vehicle inspection, select the Viewer mode as Vehicle during the installation phase.
The way to install VXvue for Vehicle is the same as that of VXvue general version. Refer
to Installation in VXvue Service Manual for the detailed information.
VXvue for Vehicle registers and manages the prescribing information based on the place or institution
where X-ray is taken.
In the Location mode, you can use the functions for registering / managing prescribing information, as well
as managing locations and the acquired images. Before taking X-rays, it is necessary to register location
and the prescribing information to each location.
The Database mode and the Work mode in VXvue is combined as the Location mode in
VXvue for Vehicle.
The functions in the Database and the Work mode are same as those of the Location
mode.
Button Function
Enter the exposure mode with the location selected from the location list.
You can export the selected location by using the multi buttons as follows.
Button Function
Input the location name and click on the Search button to check the
search result from the location list.
Input the search condition and click on the Search button to check the
result from the location list.
Button Function
Indicates the date and time of Scheduled Date which was set when
Date
generating the location.
Location Name Indicates the name of the generated location.
Indicates the number of study that the radiography has completed and
Count
the number of study registered in the location.
Indicates a check mark when radiography for all studies registered in the
Completed
location is completed.
Description Indicates the texts which was entered while generating the location.
Range: The period configured within the search period.
Total: The total number of locations corresponded with the search
condition.
Searches the study under three types of condition and indicates the
result on the study list.
All: Indicates all studies.
Taken: Indicates studies that the radiography has been completed.
Untaken: Indicates studies that the radiography has not been
completed.
Searches the study under three types of condition and indicates the
result on the study list.
All: Indicates all studies.
Rejected: Indicates rejected studies.
No rejected: Indicates non-rejected studies.
You can add studies to location by using the multi buttons as follows.
Button Function
Moves specific images of the selected study to another study in the same location.
Exports the study list information in the related location as a text file (.svc).
You can export the selected studies by using the multi buttons as follows.
Exposed: Number of studies that have been taken images / Whole number
of studies
Filter Result: Number of studies that have been searched with specific
conditions / Wholw number of studies
6 Thumbnail Area
You can check the thumbnail images of studies selected from the study list.
You can make an exposure from the Exposure mode. The Exposure mode supports a general exposure
and macro exposure modes.
Image Function
Searches the study under the two conditions and indicate the result in the area of
study list of patients.
All: Indicates both taken and untaken studies.
Untaken: Indicates the untaken studies only.
9 Thumbnail Area
You can check thumbail images from the thumbnail area. When you select a thumbnail image, it displays
on the image view area.
Macro Exposure mode: You can take and adjust images by expanding the image view area.
Image Function
5 Thumbnail Area
You can check thumbail images from the thumbnail area. When you select a thumbnail image, it displays
on the image view area.
In the Viewer mode, you can search and adjust images of the selected study.
Image Function
6 Thumbnail Area
You can check thumbail images from the thumbnail area. When you select a thumbnail image, it displays
on the image view area.
Panel Description
Tools Available to use the image adjustment tools.
Process Available to use the image processing tools.
Label Available to use the labelling tools on the image.
In VXvue for Vehicle, the prescribing information of pateints is registered based on the shooting location
or institution. Therefore, you should register the location first and the prescribing information next before
taking images under VXvue for Vehicle.
2 Click on the Add button on the left top of the location management area.
Refer to <9.4.2 Location Mode> for the detailed information about the location
management area.
3 Choose the step information to acquire images and input the location information to Location Info
dialog.
The location name can be entered up to 64 letters, excluding some special texts (^ * ? <>
/ \ ,.) which cannot be used as a file name.
The address can be entered up to 64 letters.
The telephone number can be entered up to 16 letters.
The description can be entered up to 1024 letters.
Click on at the right side of Scheduled Date to open the Date Selection window. Choose the
reserved date and click on the OK button to close.
Choose a step from the step list and click on button to remove the step.
Refer to <9.4.2 Location Mode> for the detailed information about the study
management area.
This process of adding study to Import Study is the same as that of the general version
of VXvue. Refer to <5.3.1 Tools in Reserved Panel> for more information.
This process is same as that of the Scheduled panel in VXvue general version, using the
DICOM worklist. (Refer to <5.1 Scheduled Panel (Worklist)>)
The Open, Emergency, Send to Detector, Add to Local buttons in VXvue general version
are not indicated on the VXvue for Vehicle.
You can take images after generating a location and registering studies in the location. Be sure to check if
the location and prescribing information are registered before taking images.
The shooting process in VXvue for Vehicle is same as that of the VXvue general version.
Refer to <6How to Shoot Images> for the detailed information.
This chapter gives an instruction about the functions of VXvue for Vehicle mainly.
Refer to ‘Search Area of Location’ in <9.4.2 Location Mode> for the information on how
to search locations.
3 Click on the Start button on the left top in the management area to enter the Exposure mode. If the
other location is being shot images, close the location first and then enter the Exposure mode.
In the Exposure mode, you cannot take images from more than two locations at the
same time.
You can also enter the Exposure mode by double-clicking the location to take images on
the location list.
Refer to <9.4.3 Exposure Mode> for the information about composition of the Exposure
mode.
The detector icons are showed differently depending on each detector model. Refer to
<10.6.4 Detector Information Display Icon> for the information of detector icons.
Refer to <6.4 Selecting Patient Size> for the detailed information about selecting the
patient size.
7 Check if the patient is ready to be shot from the image view area and the study list in the Exposure
mode.
VXvue for Vehicle supports continuous shooting specialized for the patients examined in
a vehicle.
When shooting for steps in the study is finished, the screen is converted to standby status
for the next patient automatically.
If you want to start shooting a study in the middle of the list, click the related study or
search it from the search field.
10 Check the ready status of the next patient from the following area located at the top of the Exposure
mode screen.
Image Description
Checks the information about the current study.
Patient name, patient ID, age, gender
You can modify the information by clicking on the pencil
icon.
If the two icons are activated, the screen is shifted to
standby status for the next patient automatically after
shooting all steps in the current study.
Click on the turning button of auto exposure to cease
the next shooting of a study automatically.
If the two icons are deactivated, the screen is not shifted
to the next study but the steps are added to the current
study after shooting all steps in the study and trying to
take images again.
If you want to stop shooting images, click on the Finish button at the left top of the
location information area to restart shooting later.
12 Click on the Finish button to enter the Location mode after shooting images for all studies.
When you complete shooting images for all studies in the location, the Auto Generate
item is selected automatically to keep shooting.
13 Check if the shooting is completed from the location list in the Location mode.
The ‘Completed’ part in the Location list area is changed as after taking images of
all studies in the location.
If the images acquired in a vehicle cannot be made DICOM transmission, you can save them to the external
storage media or the external route.
Click on button at the left top in the Location mode and click on Ext.Save icon .
Save all images in the location according to the options preset in the Setting mode.
Refer to External Save Menu in VXvue Service Manual for the detailed information about
the applied options when saving images with the External Save function.
While the images are being taken, the External Save function is applied automatically according to the
preset options.
Set the Use Auto External Save on Exposure Mode option as Yes.
Target Directory: Click on to set the storage path of images.
Default path: Database folder path\ExternalSave
Generation rule of the folder when saving images to external storage: \ ExternalSave \ Location
Name_Location Scheduled Date[YYYYMMDD]_Location Key
Since External Save is performed in background, it can be carried out simultaneously with
other tasks.
The writable and readable media (USB, HDD, pre-configured network drive) can be used
only as the exteral storage.
You cannot use an optical disk such as CD-ROM.
Image Format: Set the image format to save the images to the external device.
Be sure to click on button to save the changed settings.
Refer to External Save Menu in VXvue Service Manual for the information about external
save function.
This chapter explains about the additional functions of VXvue for Vehicle.
2 Choose a step to add to the default step list from Candidate at the left.
3 Click on the right arrow button to add the step to Location Default Step.
4 Click on the Add button at Location Info to check the added step.
2 You can search studies or indicate patient ID / patient name you want to add through the barcode
reader.
3 With the entered barcode, search studies in the location which is ready for being taken images.
4 The studies are searched in turn, and the one matched with the condition is selected automatically.
5 If there is no matched study in the location, a new study is generated and entered exposure ready state
automatically.
The patient ID of the new study is the entered barcode.
1 Choose a study from the study list to add steps while shooting images in the Exposure mode.
3 Choose a step from the Step Selector window and click on the Add button to close the window.
4 Check the added step in the thumbnail area. (ex. Check LAT)
9 Choose a path to store the location and click on the Save button.
Birth Date
Modality
Patient Comment
Study Description
External Save Status
External Save Study Count
The column information of exported location is as follows.
Location Name
Scheduled Date
Total Study Count
Exposed Study Count
4 Select All Studies or Selected button from the window for study movement.
Click on the Cancel button to cancel movement of studies.
5 Choose a location you want to move the studies and click the OK button.
You can search location quickly by using the input window at the top of the screen.
6 From the Location list or Study list, check if the studies are moved correctly.
VXvue can measure the intensity of x-ray dose reaching a detector through the subject and indicate it as
the exposure index by making a correlation calculation with gray level.
VXvue can measure the exposure index automatically through specific algorithm (extracted effective
value of the subject area from the image) and indicate it on the Overlay area of the image view.
E.I. (Exposure Index)
Target E.I (Target Exposure Index)
D.I. (Deviation Index)
If you try to calculate exposure index without a subject, take images by selecting a specific step (EI –
Pattern 25). This step calculates E.I of 25% area of the image from its center.
E.I is the correlation exponent between x-ray dose and gray level displayed on the image.
The E.I range of Vieworks’ detector is between 0 ~ 10000.
The range can be different depending on each detector model.
We use E.I defined from IEC62494-1.
Target exposure index (Target E.I) is the expected optimal exposure index.
You can preset Target E.I for each step and find the difference with the actual E.I.
Target E.I is set differently depending on the detector model, diagnostic purpose, and image processing
parameter.
Deviation index indicates the difference between Target E.I and E.I. Therefore, set both Target E.I and E.I
to measure deviation index.
With D.I, you can judge the exposure to each body part against the target exposure index.
Formula
You can measure the deviation index as follows.
EI
DI = 10 * log10 Ex) D.I. = 10 * log (1361/1300) = 0.199
EI T
Move to Exposure or Viewer mode > Click on the Target E.I button.
Refer to <7.23 Changing Target E.I.> for the detailed information about setting the target
exposure index with Target E.I button.
Refer to the guide ‘Exposure Index Settings’ provided by Vieworks to check the
recommended target E.I for each body part.
The allowed input range is 100 ~ 5000 and the default value is 286 ~ 500 depending on each body
part.
The default value of target E.I for each body part in VXvue is 500 if you create a new
body part. As target E.I is largely affected by the actual exposure environment, you should
set the target E.I accorded with the environment when it needs to be changed.
You can check each index values from overlaid information on the image view.
1 Move to Setting Display Information Overlay tab.
2 Choose Exposure Index, Deviation Index, Target Exposure Index from the Item list and add them to
Top Right or Bottom Right list.
3 Click on the Update button to apply the following items.
Exposure Index (E.I.) = 537
Target Exposure Index (Target E.I.) = 800
Deviation Index (D.I.) = -1.7 (10 * log (537/800))
You can calculate the desired statistics result from the exposure, study, and patient information stored in
VXvue.
Execution file
The statistics program is the stand-alone type one called by VXvue.
The database update for extracting statistic data is needed for using the statistics
program.
If the first installation version of VXvue is older than V1.0.0.82, the database will be
updated automatically when you execute the statistics program for the first time.
If the VXvue version is above V1.0.0.82, the database update is not needed since this
version supports the database structure for extracting statistics data.
You should backup the database file in advance and store it safely before updating
database.
9.6.1 Starting
Execute program
1 Click on the Statistics button in the Database mode to execute the statistics program.
Login
Enter the login account.
The user account information of statistics program complies
with the user account registered in VXvue.
If you log on to VXvue by the administrator account (Service,
Admin), this login procdure is skipped.
The specific patient information can be used with the
performing physician ID only.
Database update
Click on the Update DB button to update database.
This procedure is skipped if the database update is not
needed.
The update speed can be different depending on the PC
specifications and the number of acquired images.
9.6.2 VW.STATISTICS
Item Description
General Statistics data of study, exposure and image
Reject Statistics data of the deleted study and image
Dose Statistics data of X-ray dose and the amount of radiaton.
Personal Statistics data for each patient
9.6.3 Search
Search
1 Click on the Search button.
2 Check statistics data of each tab matched with the search condition.
Export image
1 Click on the Save as Image button.
2 Specify the storage path and save the statistics data as a BMP image file.
The study statistics data and graph are indicated according to the search conditions.
1 Summary
Total number of studies / Total number of exposed images / Average number of images per study
The rejected study statistics data and graph are indicated on according to the search conditions.
1 Summary
The number of removed studies / The average number of removed studies per study
The statistics data and graph about the amount of exposed x-ray dose are indicated according to the
search conditions.
1 Summary
Total amount of x-ray dose / Average dose per image / Average dose per study
The statistics data and graph about the study, exposure, and exposed dose related to the searched
patient are indicated.
1 Summary
Basic patient information (ID / Name / Gender / Birth date / Last study date)
Total number of exposed studies
Totak number of exposed images, number of rejected images, average number of exposed images per a
study
Total amount of exposed dose, average amount of exposed dose per a study
Selects an image.
Select
Returns a mouse button to its default state.
W/L (Window Level) Adjusts the brightness and contrast of the image.
Processes the image with various tools from XIPL, the image
Image Process
processing program connected with VXvue.
Status Bar
Emergency
Switching the User Account
Managing the List Header and Field
Mouse and Shortcuts
Icon
Cursors
You can check the status of VXvue from the status bar.
VXvue logo
Current login account
Signal strength of the wireless detector
Battery remains of the wireless detector
Free space of hard disk
Strorage status of the External Save
Transmission status of DICOM image
Print status of DICOM image
Current date and time
Refer to Status Bar menu in VXvue Service Manual for editing status bar display
information.
(1/2/3/4/5/6/7/8)
The transmission status to the DICOM print server is indicated as a number beside of the icon. (Success
/ All / Fail)
When you click on this icon, the tab is converted to Setting mode DICOM panel Queue tab.
You can check the transmission status of print from the Queue tab in detail.
8 Indicates the transmissoin status to DICOM MWL server after logging to VXvue.
The icon shows connecting status of the DICOM MWL server. (No server list / Normal / Disconnected /
Unknown in connecting status)
If the connecting status is ‘Echo’ or ‘Not transmitting images’, it renews its stauts in every 5 minitues.
Refer to <10.6.1 Status Bar Icon> about the icon status information.
10.2 Emergency
You can use the following functions of VXvue in a medical emergency situation.
Make an emergency exposure without a login account.
You can take images first without entering the information of an emergency patient if there is no time to
figure it out.
In the ER mode, you can only enter into the Registration mode and the Exposure mode.
The following functions are unable to use from the Registration mode and the Exposure
mode.
DICOM sending or DICOM printing from the Exposure mode.
Process or delete the image from the Exposure mode.
You can update the patient information after completing the exposure.
If the radiologist in charge of taking images is changed, use the Log Off function to change the user
account without closing VXvue.
The statistics results of Quick Access Tool, and Button Setting are different depending
on the settings by each user account.
1 Select the Exit mode and click on the Logoff button to close the account being accessed currently.
2 Enter the ID and password of the radiologist in charge and click on Login when a login window displays.
The Offset Refresh for each detector is performed from the login window automatically if the following two
conditions are satisfied at the time of rebooting VXvue or switching its user account.
Over 30 minitutes has passed from the last Offset Refresh.
The temperature difference of detector is over 3°C from the last Offset Refresh.
The Offset Refresh is not perfomed in case the user logs in ER mode.
The Offset Refresh is not perfomed from the demo version, though the related message
is displayed.
Total 61 62 68
Click the left mouse button on the desired field to align it on the header of study list.
You can align the list as an ascending or descending order based on the selected field.
1 Check a checkbox or select a desired item, and click on the Show button. You can add a field from the
header of a study list.
2 Uncheck a checkbox or select a desired item, and click on the Hide button. You can hide a field from the
header of a study list
3 Selct a desired item and click on the Move Up or Move Down button. You can adjust the field order
from the header of a study list.
4 Selct a desired item and input the width of a field directly from Width of selected column (in pixels).
You can set the width of each field in a pixel unit.
5 Click on the OK button and check if the field is changed.
If you remove a field, it becomes invisible from the header, but the data is not removed
actually.
10.5.1 Mouse
Movement Description
Click left button Activates a specific function by selecting image, menu, or a button.
Indicates options of the selected menu.
Click right button
This action is same as a long touch.
Ctrl + Click left button Selects multiple items.
Shift + Click left button Clicks on the first and the last items to select all items between them.
Double click the left button Conducts a special function defined to each menu.
Scroll Changes to select an image by using the mouse wheel.
Press and hold the left
Adjusts the parameter or move the object / change the objsct size.
button+ Drag
Press and hold the right Controls W/L (Window level).
button+ Drag This function can be used while another function is in use.
10.5.2 Shortcuts
Shortcut Description
F2 Makes a virtual exposure from VXvue Demo.
F5 Changes the direction of Projection to the left or right in the Exposure mode.
Adjusts the width of Bodypart and Projection area narrowly or widely in the Exposure
F6
mode.
F7 Changes the Quick Access tools in clockwise direction in the Exposure mode.
Converts UI type of the step selection area in the Exposure mode.
F8
Image Type Button Type
ESC The annotation mode is canceled if the annotation of figure or measurement is being made.
10.6 Icon
Icon Description
Very good
Good
Normal
Bad
Very bad, Disconnected
Icon Description
Connected
Not initialized
Disconnected
Icon Description
Not initialized
75~100%
50~75%
25~50%
25% and less
Very low
Icon Description
Errors in the local HDD
The remaining capacity of local HDD is 25% or less.
The remaining capacity is 50% or less compared to the total capacity of local HDD.
The remaining capacity is 75% or less compared to the total capacity of local HDD.
The remaining capacity is 100% or less compared to the total capacity of local HDD.
Status of DICOM image storage server
Status of DICOM print server
Icon Description
No DICOM image storage server list
All selected DICOM image storage servers are connected normally.
One or more of the selected DICOM image storage servers are disconnected.
The connecting status of one or more of the selected DICOM image storage servers is unknown
Icon Description
No list of DICOM and paper printer servers.
Selected DICOM print server is connected normally.
The selected DICOM print server is disconnected.
The connection status of the selected DICOM print server is unknown.
Icon Description
No list of DICOM MWL servers.
Selected DICOM MWL server is connected normally.
The selected DICOM MWL server is disconnected.
The connection status of the selected DICOM MWL server is unknown.
Icon Description
External Save can be performed because the path specified by External Save normally exists.
External Save is stopped because the path specified by External Save does not exist normally.
The icons indicate status of the study items which are displayed on the database search list.
Icon Description
Backed up
The backed up images are removed from system.
Transmitted to the DICOM print storage server.
Printed to the DICOM print server.
Icon Description
Rejected
Icon Description
Battery is full.
Temperature error
Normal temperature
High temperature
Wifi error
Icon Description
Some images are not sent to the viewer but remained in the internal memory of a detector.
The wireless detector model (FXRD-W, FXRD-N) is connected with a tether interface cable.
10.7 Cursors
When you select image processing or annotation tools, the mouse cursor is changed to the
corresponding shape.
Cursor Description
Crop
Zoom
Magic Glass
Window Level
Pan
Move
Cursor Description
Angle
Arrow
Window Level
Ellipse
Rectangle
Text