Magnetic Particle Testing

Download as pdf or txt
Download as pdf or txt
You are on page 1of 34

Magnetic Particle Testing

Magnetic particle testing is one of the most widely utilized NDT methods
since it is fast and relatively easy to apply and part surface preparation is
not as critical as it is for some other methods. This method uses magnetic
fields and small magnetic particles (i.e.iron filings) to detect flaws in
components. The only requirement from an inspectability standpoint is
that the component being inspected must be made of a ferromagnetic
material (a materials that can be magnetized) such as iron, nickel, cobalt,
or some of their alloys.
The method is used to inspect a variety of product forms including
castings, forgings, and weldments. Many different industries use
magnetic particle inspection such as structural steel, automotive,
petrochemical, power generation, and aerospace industries. Underwater
inspection is another area where magnetic particle inspection may be
used to test items such as offshore structures and underwater pipelines.

Basic Principles
In theory, magnetic particle testing has a relatively simple concept. It can be
considered as a combination of two nondestructive testing methods: magnetic flux
leakage testing and visual testing. For the case of a bar magnet, the magnetic field is in
and around the magnet. Any place that a magnetic line
of force exits or enters the magnet is called a “pole”
(magnetic lines of force exit the magnet from north pole
and enter from the south pole).
When a bar magnet is broken in the center of its length, two complete bar magnets
with magnetic poles on each end of each piece will result. If the magnet is just cracked
but not broken completely in two, a north and south pole will form at each edge of the
crack. The magnetic field exits the north pole and reenters at the south pole. The
magnetic field spreads out when it encounters the
small air gap created by the crack because the air
cannot support as much magnetic field per unit
volume as the magnet can. When the field spreads
out, it appears to leak out of the material and, thus
is called a flux leakage field.
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 1 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
If iron particles are sprinkled on a cracked magnet, the particles will be attracted to
and cluster not only at the poles at the ends of the magnet, but also at the poles at the
edges of the crack. This cluster of particles is much easier to see than the actual crack
and this is the basis for magnetic particle inspection.
The first step in a magnetic particle testing is to magnetize the component that is to be
inspected. If any defects on or near the
surface are present, the defects will
create a leakage field. After the
component has been magnetized, iron
particles, either in a dry or wet
suspended form, are applied to the
surface of the magnetized part. The
particles will be attracted and cluster
at the flux leakage fields, thus forming
a visible indication that the inspector can detect.

Advantages and Disadvantages


The primary advantages and disadvantages when compared to other NDT methods
are:
Advantages
 High sensitivity (small discontinuities can be detected).
 Indications are produced directly on the surface of the part and constitute a visual
representation of the flaw.
 Minimal surface preparation (no need for paint removal)
 Portable (small portable equipment & materials available in spray cans)
 Low cost (materials and associated equipment are relatively inexpensive)
Disadvantages
 Only surface and near surface defects can be detected.
 Only applicable to ferromagnetic materials.
 Relatively small area can be inspected at a time.
 Only materials with a relatively nonporous surface can be inspected.
 The inspector must have direct access to the surface being inspected.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 2 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Magnetism
The concept of magnetism centers around the magnetic field and what is known as a
dipole. The term "magnetic field" simply describes a volume of space where there is a
change in energy within that volume. The location where a magnetic field exits or
enters a material is called a magnetic pole. Magnetic poles have never been detected
in isolation but always occur in pairs, hence the name dipole. Therefore, a dipole is an
object that has a magnetic pole on one end and a second, equal but opposite,
magnetic pole on the other. A bar magnet is a dipole with a north pole at one end and
south pole at the other.
The source of magnetism lies in the basic building block of all
matter, the atom. Atoms are composed of protons, neutrons and
electrons. The protons and neutrons are located in the atom's
nucleus and the electrons are in constant motion around the
nucleus. Electrons carry a negative electrical charge and produce
a magnetic field as they move through space. A magnetic field is
produced whenever an electrical charge is in motion. The
strength of this field is called the magnetic moment.
When an electric current flows through a conductor, the movement of electrons
through the conductor causes a magnetic field to form around the conductor. The
magnetic field can be detected using a compass. Since all matter is comprised of
atoms, all materials are affected in some way by a magnetic field; however, materials
do not react the same way to the magnetic field.

Reaction of Materials to Magnetic Field


When a material is placed within a magnetic field, the magnetic forces of the material's
electrons will be affected. This effect is known as Faraday's Law of Magnetic Induction.
However, materials can react quite differently to the presence of an external magnetic
field. The magnetic moments associated with atoms have three origins: the electron
motion, the change in motion caused by an external magnetic field, and the spin of the
electrons.
In most atoms, electrons occur in pairs where these pairs spin in
opposite directions. The opposite spin directions of electron pairs
cause their magnetic fields to cancel each other. Therefore, no net
magnetic field exists. Alternately, materials with some unpaired
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 3 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
electrons will have a net magnetic field and will react more to an external field.
According to their interaction with a magnetic field, materials can be classified as:
Diamagnetic materials which have a weak, negative susceptibility to magnetic
fields. Diamagnetic materials are slightly repelled by a magnetic field and the
material does not retain the magnetic properties when the external field is
removed. In diamagnetic materials all the electrons are paired so there is no
permanent net magnetic moment per atom. Most elements in the periodic table,
including copper, silver, and gold, are diamagnetic.
Paramagnetic materials which have a small, positive susceptibility to magnetic
fields. These materials are slightly attracted by a magnetic field and the material
does not retain the magnetic properties when the external field is removed.
Paramagnetic materials have some unpaired electrons. Examples of paramagnetic
materials include magnesium, molybdenum, and lithium.
Ferromagnetic materials have a large, positive susceptibility to an external
magnetic field. They exhibit a strong attraction to magnetic fields and are able to
retain their magnetic properties after the external field has been removed.
Ferromagnetic materials have some unpaired electrons so their atoms have a net
magnetic moment. They get their strong magnetic properties due to the presence
of magnetic domains. In these domains, large numbers of atom's moments are
aligned parallel so that the magnetic force within the domain is strong (this happens
during the solidification of the material where the atom moments are aligned within
each crystal ”i.e., grain” causing a strong magnetic force in one direction). When a
ferromagnetic material is in the
unmagnetized state, the domains are nearly
randomly organized (since the crystals are
in arbitrary directions) and the net magnetic
field for the part as a whole is zero. When a
magnetizing force is applied, the domains
become aligned to produce a strong
magnetic field within the part. Iron, nickel,
and cobalt are examples of ferromagnetic
materials. Components made of these
materials are commonly inspected using
the magnetic particle method.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 4 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Magnetic Field Characteristics
Magnetic Field In and Around a Bar Magnet
The magnetic field surrounding a bar magnet can be seen in the magnetograph below.
A magnetograph can be created by placing a piece
of paper over a magnet and sprinkling the paper
with iron filings. The particles align themselves with
the lines of magnetic force produced by the
magnet. It can be seen in the magnetograph that
there are poles all along the length of the magnet
but that the poles are concentrated at the ends of
the magnet (the north and south poles).

Magnetic Fields in and around Horseshoe and Ring Magnets


Magnets come in a variety of shapes and one of the more common
is the horseshoe (U) magnet. The horseshoe magnet has north and
south poles just like a bar magnet but the magnet is curved so the
poles lie in the same plane. The magnetic lines of force flow from
pole to pole just like in the bar magnet. However, since the poles are
located closer together and a more direct path exists for the lines of
flux to travel between the poles, the magnetic field is concentrated
between the poles.

General Properties of Magnetic Lines of Force


Magnetic lines of force have a number of important properties, which include:
 They seek the path of least resistance between opposite
magnetic poles (in a single bar magnet shown, they attempt
to form closed loops from pole to pole).
 They never cross one another.
 They all have the same strength.
 Their density decreases with increasing distance from the
poles.
 Their density decreases (they spread out) when they move
from an area of higher permeability to an area of lower
permeability.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 5 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
 They are considered to have direction as if flowing,
though no actual movement occurs.
 They flow from the south pole to the north pole
within a material and north pole to south pole in air.

Electromagnetic Fields
Magnets are not the only source of magnetic fields. The flow of electric current
through a conductor generates a magnetic field. When electric current flows in a long
straight wire, a circular magnetic field is generated around the wire and the intensity of
this magnetic field is directly proportional to the amount of
current carried by the wire. The strength of the field is strongest
next to the wire and diminishes with distance. In most
conductors, the magnetic field exists only as long as the current
is flowing. However, in ferromagnetic materials the electric
current will cause some or all of the magnetic domains to align
and a residual magnetic field will remain.
Also, the direction of the magnetic field is dependent on the direction of the electrical
current in the wire. The direction of the magnetic field around a conductor can be
determined using a simple rule called the “right-hand clasp rule”. If a person grasps a
conductor in one's right hand with the thumb pointing in the direction of the current,
the fingers will circle the conductor in the direction of the magnetic field.
Note: remember that current flows from the positive terminal to the negative
terminal (electrons flow in the opposite direction).

Magnetic Field Produced by a Coil


When a current carrying wire is formed into
several loops to form a coil, the magnetic field
circling each loop combines with the fields from
the other loops to produce a concentrated field
through the center of the coil (the field flows
along the longitudinal axis and circles back
around the outside of the coil).

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 6 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
When the coil loops are tightly wound, a uniform magnetic field is developed
throughout the length of the coil. The strength of the magnetic field increases not only
with increasing current but also with each loop that is added to the coil. A long,
straight coil of wire is called a solenoid and it can be used to generate a nearly uniform
magnetic field similar to that of a bar magnet. The concentrated magnetic field inside a
coil is very useful in magnetizing ferromagnetic materials for inspection using the
magnetic particle testing method.

Quantifying Magnetic Properties


The various characteristics of magnetism can be measured and expressed
quantitatively. Different systems of units can be used for quantifying magnetic
properties. SI units will be used in this material. The advantage of using SI units is that
they are traceable back to an agreed set of four base units; meter, kilogram, second,
and Ampere.
 The unit for magnetic field strength H is ampere/meter
(A/m). A magnetic field strength of 1 A/m is produced
at the center of a single circular conductor with a 1
meter diameter carrying a steady current of 1 ampere.

 The number of magnetic lines of force cutting through a plane of a given area at
a right angle is known as the magnetic flux density, B. The flux density or
magnetic induction has the Tesla as its unit. One Tesla is equal to 1
Newton/(A/m). From these units, it can be seen that the flux density is a
measure of the force applied to a particle by the magnetic field.

 The total number of lines of magnetic force in a material is called magnetic flux,
ɸ. The strength of the flux is determined by the number of magnetic domains
that are aligned within a material. The total flux is simply the flux density applied
over an area. Flux carries the unit of a weber, which is simply a Tesla-meter2.

 The magnetization M is a measure of the extent to which an object is


magnetized. It is a measure of the magnetic dipole moment per unit volume of
the object. Magnetization carries the same units as a magnetic field A/m.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 7 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Quantity SI Units SI Units CGS Units
(Sommerfeld) (Kennelly) (Gaussian)
Field H A/m A/m oersteds
(Magnetization
Force)
Flux Density B Tesla Tesla gauss
(Magnetic
Induction)
Flux ɸ Weber Weber maxwell
Magnetization M A/m - erg/Oe-cm3

The Hysteresis Loop and Magnetic Properties


A great deal of information can be learned about the magnetic properties of a material
by studying its hysteresis loop. A hysteresis loop shows the relationship between the
induced magnetic flux density (B) and the magnetizing force (H). It is often referred to
as the B-H loop. An example hysteresis loop is shown below.

The loop is generated by measuring the magnetic flux of a ferromagnetic material


while the magnetizing force is changed. A ferromagnetic material that has never been
previously magnetized or has been thoroughly demagnetized will follow the dashed
line as H is increased. As the line demonstrates, the greater the amount of current
applied (H+), the stronger the magnetic field in the component (B+). At point "a"
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 8 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
almost all of the magnetic domains are aligned and an additional increase in the
magnetizing force will produce very little increase in magnetic flux. The material has
reached the point of magnetic saturation. When H is reduced to zero, the curve will
move from point "a" to point "b". At this point, it can be seen that some magnetic flux
remains in the material even though the magnetizing force is zero. This is referred to as
the point of retentivity on the graph and indicates the level of residual magnetism in
the material (Some of the magnetic domains remain aligned but some have lost their
alignment). As the magnetizing force is reversed, the curve moves to point "c", where
the flux has been reduced to zero. This is called the point of coercivity on the curve
(the reversed magnetizing force has flipped enough of the domains so that the net flux
within the material is zero). The force required to remove the residual magnetism from
the material is called the coercive force or coercivity of the material.
As the magnetizing force is increased in the negative direction, the material will again
become magnetically saturated but in the opposite direction, point "d". Reducing H to
zero brings the curve to point "e". It will have a level of residual magnetism equal to
that achieved in the other direction. Increasing H back in the positive direction will
return B to zero. Notice that the curve did not return to the origin of the graph because
some force is required to remove the residual magnetism. The curve will take a
different path from point "f" back to the saturation point where it with complete the
loop.
From the hysteresis loop, a number of primary magnetic properties of a material can
be determined:
1. Retentivity - A measure of the residual flux density corresponding to the
saturation induction of a magnetic material. In other words, it is a material's
ability to retain a certain amount of residual magnetic field when the magnetizing
force is removed after achieving saturation (The value of B at point b on the
hysteresis curve).
2. Residual Magnetism or Residual Flux - The magnetic flux density that remains in a
material when the magnetizing force is zero. Note that residual magnetism and
retentivity are the same when the material has been magnetized to the saturation
point. However, the level of residual magnetism may be lower than the retentivity
value when the magnetizing force did not reach the saturation level.
3. Coercive Force - The amount of reverse magnetic field which must be applied to a
magnetic material to make the magnetic flux return to zero (The value of H at
point c on the hysteresis curve).
4. Permeability, µ - A property of a material that describes the ease with which a
magnetic flux is established in the material.
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 9 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
5. Reluctance - Is the opposition that a ferromagnetic material shows to the
establishment of a magnetic field. Reluctance is analogous to the resistance in an
electrical circuit.

Permeability
As previously mentioned, permeability (µ) is a material property that describes the
ease with which a magnetic flux is established in a component. It is the ratio of the flux
density (B) created within a material to the magnetizing field (H) and it is represented
by the following equation:
µ = B/H
This equation describes the slope of the curve at
any point on the hysteresis loop. The permeability
value given in letrature for materials is usually the
maximum permeability or the maximum relative
permeability. The maximum permeability is the
point where the slope of the B/H curve for the
unmagnetized material is the greatest. This point is
often taken as the point where a straight line from
the origin is tangent to the B/H curve.
The shape of the hysteresis loop tells a great deal about the material being
magnetized. The hysteresis curves of two different materials are shown in the graph.
 Relative to other materials, a material with a wider
hysteresis loop has:
- Lower Permeability
- Higher Retentivity
- Higher Coercivity
- Higher Reluctance
- Higher Residual Magnetism
 Relative to other materials, a material with a narrower
hysteresis loop has:
- Higher Permeability
- Lower Retentivity
- Lower Coercivity
- Lower Reluctance
- Lower Residual Magnetism
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 10 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
In magnetic particle testing, the level of residual magnetism is important. Residual
magnetic fields are affected by the permeability, which can be related to the carbon
content and alloying of the material. A component with high carbon content will have
low permeability and will retain more magnetic flux than a material with low carbon
content.

Magnetic Field Orientation and Flaw Detectability


To properly inspect a component for cracks or other defects, it is important to
understand that the orientation of the crack relative to the magnetic lines of force
determinies if the crack can or cannot be detected. There are two general types of
magnetic fields that can be established within a component.
 A longitudinal magnetic field has magnetic lines of force
that run parallel to the long axis of the part. Longitudinal
magnetization of a component can be accomplished using
the longitudinal field set up by a coil or solenoid. It can also
be accomplished using permanent magnets or
electromagnets.
 A circular magnetic field has magnetic lines of force that
run circumferentially around the perimeter of a part. A
circular magnetic field is induced in an article by either
passing current through the component or by passing
current through a conductor surrounded by the
component.
The type of magnetic field established is determined by the method used to magnetize
the specimen. Being able to magnetize the part in two directions is important because
the best detection of defects occurs when the lines of magnetic force are established
at right angles to the longest dimension of the defect. This
orientation creates the largest disruption of the magnetic
field within the part and the greatest flux leakage at the
surface of the part. If the magnetic field is parallel to the
defect, the field will see little disruption and no flux
leakage field will be produced.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 11 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
An orientation of 45 to 90 degrees between the magnetic field and the defect is
necessary to form an indication. Since defects may occur in various and unknown
directions, each part is normally magnetized in two directions at right angles to each
other. If the component shown is considered, it is known that passing current through
the part from end to end will establish a circular magnetic field that will be 90 degrees
to the direction of the current. Therefore, defects that have a significant dimension in
the direction of the current
(longitudinal defects)
should be detectable, while
transverse-type defects will
not be detectable with
circular magnetization.

Magnetization of Ferromagnetic Materials


There are a variety of methods that can be used to establish a magnetic field in a
component for evaluation using magnetic particle inspection. It is common to classify
the magnetizing methods as either direct or indirect.
Magnetization Using Direct Induction (Direct Magnetization)
With direct magnetization, current is passed directly through the component. The flow
of current causes a circular magnetic field to form in and around the conductor. When
using the direct magnetization method, care must be taken to ensure that good
electrical contact is established and maintained between the test equipment and the
test component to avoid damage of the the component (due to arcing or overheating
at high resistance ponts).
There are several ways that direct magnetization is commonly accomplished.
- One way involves clamping the component between
two electrical contacts in a special piece of equipment.
Current is passed through the component and a circular
magnetic field is established in and around the
component. When the magnetizing current is stopped,
a residual magnetic field will remain within the
component. The strength of the induced magnetic field
is proportional to the amount of current passed
through the component.
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 12 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
- A second technique involves using clamps or prods, which are
attached or placed in contact with the component. Electrical
current flows through the component from contact to contact.
The current sets up a circular magnetic field around the path of
the current.
Magnetization Using Indirect Induction (Indirect Magnetization)
Indirect magnetization is accomplished by using a strong external magnetic field to
establish a magnetic field within the component. As with direct magnetization, there
are several ways that indirect magnetization can be accomplished.
- The use of permanent magnets is a low cost method of
establishing a magnetic field. However, their use is limited due
to lack of control of the field strength and the difficulty of
placing and removing strong permanent magnets from the
component.

- Electromagnets in the form of an adjustable


horseshoe magnet (called a yoke) eliminate
the problems associated with permanent
magnets and are used extensively in industry.
Electromagnets only exhibit a magnetic flux
when electric current is flowing around the
soft iron core. When the magnet is placed on
the component, a magnetic field is established
between the north and south poles of the
magnet.

- Another way of indirectly inducting a magnetic field in a material is


by using the magnetic field of a current carrying conductor. A
circular magnetic field can be established in cylindrical components
by using a central conductor. Typically, one or more cylindrical
components are hung from a solid copper bar running through the
inside diameter. Current is passed through the copper bar and the
resulting circular magnetic field establishes a magnetic field within
the test components.

- The use of coils and solenoids is a third method of indirect magnetization. When the
length of a component is several times larger than its diameter, a longitudinal
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 13 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
magnetic field can be established in the
component. The component is placed
longitudinally in the concentrated
magnetic field that fills the center of a coil
or solenoid. This magnetization technique
is often referred to as a "coil shot".

Types of Magnetizing Current


As mentioned previously, electric current is often used to establish the magnetic field
in components during magnetic particle inspection. Alternating current (AC) and direct
current (DC) are the two basic types of current commonly used. The type of current
used can have an effect on the inspection results, so the types of currents commonly
used are briefly discussed here.
Direct Current
Direct current (DC) flows continuously in one direction at a constant voltage. A battery
is the most common source of direct current. The current is said to flow from the
positive to the negative terminal, though electrons flow in the opposite direction. DC is
very desirable when inspecting for subsurface defects because DC generates a
magnetic field that penetrates deeper into the material. In ferromagnetic materials,
the magnetic field produced by DC generally penetrates the entire cross-section of the
component.
Alternating Current
Alternating current (AC) reverses its direction at a rate of 50 or 60 cycles per second.
Since AC is readily available in most facilities, it is convenient to make use of it for
magnetic particle inspection. However, when AC is used to induce a magnetic field in
ferromagnetic materials, the magnetic field will be limited to a thin layer at the surface
of the component. This phenomenon is known as the "skin effect" and it occurs
because the changing magnetic field generates eddy currents in the test object. The
eddy currents produce a magnetic field that opposes the primary field, thus reducing
the net magnetic flux below the surface. Therefore, it is recommended that AC be used
only when the inspection is limited to surface defects.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 14 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Rectified Alternating Current
Clearly, the skin effect limits the use of AC since many inspection applications call for
the detection of subsurface defects. Luckily, AC can be converted to current that is very
much like DC through the process of rectification. With the use of rectifiers, the
reversing AC can be converted to a one directional current. The three commonly used
types of rectified current are described below.

Half Wave Rectified Alternating Current (HWAC)


When single phase alternating current is passed through a rectifier, current is allowed
to flow in only one direction. The reverse half of each cycle is blocked out so that a one
directional, pulsating current is produced. The current rises from zero to a maximum
and then returns to zero. No current flows during the time when the reverse cycle is
blocked out. The HWAC repeats at same rate as the unrectified current (50 or 60 Hz).
Since half of the current is blocked out, the amperage is half of the unaltered AC. This
type of current is often referred to as half wave DC or pulsating DC. The pulsation of
the HWAC helps in forming magnetic particle indications by vibrating the particles and
giving them added mobility where that is especially important when using dry
particles. HWAC is most often used to power electromagnetic yokes.
Full Wave Rectified Alternating Current (FWAC) (Single Phase)
Full wave rectification inverts the negative current to positive current rather than
blocking it out. This produces a pulsating DC with no interval between the pulses.
Filtering is usually performed to soften the sharp polarity switching in the rectified

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 15 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
current. While particle mobility is not as good as half-wave AC due to the reduction in
pulsation, the depth of the subsurface magnetic field is improved.
Three Phase Full Wave Rectified Alternating Current
Three phase current is often used to power industrial equipment because it has more
favorable power transmission and line loading characteristics. This type of electrical
current is also highly desirable for magnetic particle testing because when it is rectified
and filtered, the resulting current very closely resembles direct current. Stationary
magnetic particle equipment wired with three phase AC will usually have the ability to
magnetize with AC or DC (three phase full wave rectified), providing the inspector with
the advantages of each current form.

Magnetic Fields Distribution and Intensity

Longitudinal Fields
When a long component is magnetized using a solenoid having a shorter length, only
the material within the solenoid and
about the same length on each side of
the solenoid will be strongly magnetized.
This occurs because the magnetizing
force diminishes with increasing distance
from the solenoid. Therefore, a long
component must be magnetized and
inspected at several locations along its
length for complete inspection coverage.

Circular Fields
When a circular magnetic field forms in and around a conductor due to the passage of
electric current through it, the following can be said about the distribution and
intensity of the magnetic field:
- The field strength varies from zero at the center of the component to a maximum
at the surface.
- The field strength at the surface of the conductor decreases as the radius of the
conductor increases (when the current strength is held constant).
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 16 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
- The field strength inside the conductor is dependent on the current strength,
magnetic permeability of the material, and, if ferromagnetic, the location on the
B-H curve.
- The field strength outside the conductor is directly proportional to the current
strength and it decreases with distance from the conductor.
The images below show the magnetic field strength graphed versus distance from the
center of the conductor when current passes through a solid circular conductor.
 In a nonmagnetic conductor carrying DC, the internal field strength rises from zero
at the center to a maximum value at the surface of the conductor.

 In a magnetic conductor carrying DC, the field strength within the conductor is
much greater than it is in the nonmagnetic conductor. This is due to the
permeability of the magnetic material. The external field is exactly the same for the
two materials provided the current level and conductor radius are the same.

 When the magnetic conductor is carrying AC, the internal magnetic field will be
concentrated in a thin layer near the surface of the conductor (skin effect). The
external field decreases with increasing distance from the surface same as with DC.

The magnetic field distribution in The magnetic field distribution in The magnetic field distribution in
and around a solid conductor of a and around a solid conductor of a and around a solid conductor of a
nonmagnetic material carrying magnetic material carrying direct magnetic material carrying
direct current. current. alternating current.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 17 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
In a hollow circular conductor there is no magnetic field in the void area. The magnetic
field is zero at the inner surface and rises until it reaches a maximum at the outer
surface.
 Same as with a solid conductor, when DC current is passed through a magnetic
conductor, the field strength within the conductor is much greater than in
nonmagnetic conductor due to the permeability of the magnetic material. The
external field strength decreases with distance from the surface of the conductor.
The external field is exactly the same for the two materials provided the current
level and conductor radius are the same.

 When AC current is passed through a hollow circular magnetic conductor, the skin
effect concentrates the magnetic field at the outside diameter of the component.

The magnetic field distribution in The magnetic field distribution in The magnetic field distribution in
and around a hollow conductor of a and around a hollow conductor of and around a hollow conductor of
nonmagnetic material carrying a magnetic material carrying direct a magnetic material carrying
direct current. current. alternating current.

As can be seen from these three field distribution images, the field strength at the
inside surface of hollow conductor is very low when a circular magnetic field is
established by direct magnetization. Therefore, the direct method of magnetization is
not recommended when inspecting the inside diameter wall of a hollow component
for shallow defects (if the defect has significant depth, it may be detectable using DC
since the field strength increases rapidly as one moves from the inner towards the outer
surface).

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 18 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
 A much better method of magnetizing hollow
components for inspection of the ID and OD surfaces
is with the use of a central conductor. As can be seen
in the field distribution image, when current is passed
through a nonmagnetic central conductor (copper
bar), the magnetic field produced on the inside
diameter surface of a magnetic tube is much greater
and the field is still strong enough for defect detection
on the OD surface.

Demagnetization
After conducting a magnetic particle inspection, it is usually necessary to demagnetize
the component. Remanent magnetic fields can:
- affect machining by causing cuttings to cling to a component.
- interfere with electronic equipment such as a compass.
- create a condition known as "arc blow" in the welding process. Arc blow may
cause the weld arc to wonder or filler metal to be repelled from the weld.
- cause abrasive particles to cling to bearing or faying surfaces and increase wear.
Removal of a field may be accomplished in several ways. The most effective way to
demagnetize a material is by heating the material above its curie temperature (for
instance, the curie temperature for a low carbon steel is 770°C). When steel is heated
above its curie temperature then it is cooled back down, the the orientation of the
magnetic domains of the individual grains will become randomized again and thus the
component will contain no residual magnetic field. The material should also be placed
with its long axis in an east-west orientation to avoid any influence of the Earth's
magnetic field.
However, it is often inconvenient to heat a material above its curie temperature to
demagnetize it, so another method that returns the material to a nearly unmagnetized
state is commonly used.
Subjecting the component to a reversing and decreasing magnetic field will return the
dipoles to a nearly random orientation throughout the material. This can be
accomplished by pulling a component out and away from a coil with AC passing
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 19 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
through it. With AC Yokes, demagnetization of local areas may be accomplished by
placing the yoke contacts on the surface, moving them in
circular patterns around the area, and slowly withdrawing
the yoke while the current is applied. Also, many
stationary magnetic particle inspection units come with a
demagnetization feature that slowly reduces the AC in a
coil in which the component is placed.
A field meter is often used to verify that the residual flux
has been removed from a component. Industry standards
usually require that the magnetic flux be reduced to less
than 3 Gauss (3x10-4 Tesla) after completing a magnetic
particle inspection.

Measuring Magnetic Fields


When performing a magnetic particle inspection, it is very important to be able to
determine the direction and intensity of the magnetic field. The field intensity must be
high enough to cause an indication to form, but not too high to cause nonrelevant
indications to mask relevant indications. Also, after magnetic inspection it is often
needed to measure the level of residual magnetezm.
Since it is impractical to measure the actual field strength within the material, all the
devices measure the magnetic field that is outside of the material. The two devices
commonly used for quantitative measurement of magnetic fields n magnetic particle
inspection are the field indicator and the Hall-effect meter, which is also called a gauss
meter.
Field Indicators
Field indicators are small mechanical devices that utilize a soft iron
vane that is deflected by a magnetic field. The vane is attached to
a needle that rotates and moves the pointer on the scale. Field
indicators can be adjusted and calibrated so that quantitative
information can be obtained. However, the measurement range of
field indicators is usually small due to the mechanics of the device
(the one shown in the image has a range from plus 20 to minus 20
Gauss). This limited range makes them best suited for measuring
the residual magnetic field after demagnetization.
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 20 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Hall-Effect (Gauss/Tesla) Meter
A Hall-effect meter is an electronic device that provides a digital readout of the
magnetic field strength in Gauss or Tesla units. The meter uses a very small conductor
or semiconductor element at the tip of the probe.
Electric current is passed through the conductor. In a
magnetic field, a force is exerted on the moving electrons
which tends to push them to one side of the conductor.
A buildup of charge at the sides of the conductors will
balance this magnetic influence, producing a measurable
voltage between the two sides of the conductor. The
probe is placed in the magnetic field such that the
magnetic lines of force intersect the major dimensions of
the sensing element at a right angle.

Magnetization Equipment for Magnetic Particle Testing


To properly inspect a part for cracks or other defects, it is important to become
familiar with the different types of magnetic fields and the equipment used to
generate them. As discussed previously, one of the primary requirements for detecting
a defect in a ferromagnetic material is that the magnetic field induced in the part must
intercept the defect at a 45 to 90 degree angle. Flaws that are normal (90 degrees) to
the magnetic field will produce the strongest indications because they disrupt more of
the magnet flux. Therefore, for proper inspection of a component, it is important to be
able to establish a magnetic field in at least two directions.
A variety of equipment exists to establish the magnetic field for magnetic particle
testing. One way to classify equipment is based on its portability. Some equipment is
designed to be portable so that inspections can be made in the field and some is
designed to be stationary for ease of inspection in the laboratory or manufacturing
facility.
Portable Equipment
Permanent Magnets
Permanent magnets can be used for magnetic particle inspection as the source of
magnetism (bar magnets or horseshoe magnets). The use of industrial magnets is not
popular because they are very strong (they require significant strength to remove them
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 21 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
from the surface, about 250 N for some magnets) and thus they
are difficult and sometimes dangerous to handle. However,
permanent magnets are sometimes used by divers for inspection
in underwater environments or other areas, such as explosive
environments, where electromagnets cannot be used. Permanent
magnets can also be made small enough to fit into tight areas
where electromagnets might not fit.

Electromagnetic Yokes
An electromagnetic yoke is a very common piece of equipment that is used to establish
a magnetic field. A switch is included in the electrical circuit so that the current and,
therefore, the magnetic field can be turned on
and off. They can be powered with AC from a wall
socket or by DC from a battery pack. This type of
magnet generates a very strong magnetic field in
a local area where the poles of the magnet touch
the part being inspected. Some yokes can lift
weights in excess of 40 pounds.

Prods
Prods are handheld electrodes that are pressed against the
surface of the component being inspected to make contact for
passing electrical current (AC or DC) through the metal. Prods
are typically made from copper and have an insulated handle to
help protect the operator. One of the prods has a trigger switch
so that the current can be quickly and easily turned on and off.
Sometimes the two prods are connected by an insulator, as
shown in the image, to facilitate one hand operation. This is
referred to as a dual prod and is commonly used for weld
inspections.
However, caution is required when using prods because electrical arcing can occur and
cause damage to the component if proper contact is not maintained between the
prods and the component surface. For this reason, the use of prods is not allowed
when inspecting aerospace and other critical components. To help prevent arcing, the
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 22 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
prod tips should be inspected frequently to ensure that they are not oxidized, covered
with scale or other contaminant, or damaged.
Portable Coils and Conductive Cables
Coils and conductive cables are used to establish a
longitudinal magnetic field within a component. When a
preformed coil is used, the component is placed against
the inside surface on the coil. Coils typically have three
or five turns of a copper cable within the molded frame.
A foot switch is often used to energize the coil.

Also, flexible conductive cables can be wrapped around


a component to form a coil. The number of wraps is
determined by the magnetizing force needed and of
course, the length of the cable. Normally, the wraps are
kept as close together as possible. When using a coil or
cable wrapped into a coil, amperage is usually expressed
in ampere-turns. Ampere-turns is the amperage shown
on the amp meter times the number of turns in the coil.

Portable Power Supplies


Portable power supplies are used to provide the
necessary electricity to the prods, coils or cables.
Power supplies are commercially available in a
variety of sizes. Small power supplies generally
provide up to 1,500A of half-wave DC or AC. They
are small and light enough to be carried and operate
on either 120V or 240V electrical service.
When more power is necessary, mobile power
supplies can be used. These units come with wheels
so that they can be rolled where needed. These units
also operate on 120V or 240V electrical service and
can provide up to 6,000A of AC or half-wave DC.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 23 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Stationery Equipment
Stationary magnetic particle inspection equipment
is designed for use in laboratory or production
environment. The most common stationary system
is the wet horizontal (bench) unit. Wet horizontal
units are designed to allow for batch inspections of
a variety of components. The units have head and
tail stocks (similar to a lathe) with electrical contact
that the part can be clamped between. A circular
magnetic field is produced with direct magnetization.
Most units also have a movable coil that can be moved
into place so the indirect magnetization can be used to
produce a longitudinal magnetic field. Most coils have
five turns and can be obtained in a variety of sizes. The
wet magnetic particle solution is collected and held in
a tank. A pump and hose system is used to apply the
particle solution to the components being inspected.
Some of the systems offer a variety of options in
electrical current used for magnetizing the component
(AC, half wave DC, or full wave DC). In some units, a
demagnetization feature is built in, which uses the coil and decaying AC.

Magnetic Field Indicators


Determining whether a magnetic field is of adequate strength and in the proper
direction is critical when performing magnetic particle testing. There is actually no
easy-to-apply method that permits an exact measurement of field intensity at a given
point within a material. Cutting a small slot or hole into the material and measuring the
leakage field that crosses the air gap with a Hall-effect meter is probably the best way
to get an estimate of the actual field strength within a part. However, since that is not
practical, there are a number of tools and methods that are used to determine the
presence and direction of the field surrounding a component.

Hall-Effect Meter (Gauss Meter)


As discussed earlier, a Gauss meter is commonly used to measure the tangential field
strength on the surface of the part. By placing the probe next to the surface, the meter
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 24 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
measures the intensity of the field in the air adjacent to the component when a
magnetic field is applied. The advantages of this device are: it provides a quantitative
measure of the strength of magnetizing force tangential to the surface of a test piece,
it can be used for measurement of residual magnetic fields, and it can be used
repetitively. The main disadvantage is that such devices must be periodically
calibrated.

Quantitative Quality Indicator (QQI)


The Quantitative Quality Indicator (QQI) or Artificial Flaw Standard
is often the preferred method of assuring proper field direction
and adequate field strength (it is used with the wet method only).
The QQI is a thin strip (0.05 or 0.1 mm thick) of AISI 1005 steel
with a specific pattern, such as concentric circles or a plus sign,
etched on it. The QQI is placed directly on the surface, with the
itched side facing the surface, and it is usually fixed to the surface
using a tape then the component is then magnetized and particles
applied. When the field strength is adequate, the particles will
adhere over the engraved pattern and provide information about
the field direction.

Pie Gage
The pie gage is a disk of highly permeable material divided into four, six, or
eight sections by non-ferromagnetic material (such as copper). The divisions
serve as artificial defects that radiate out in different directions from the
center. The sections are furnace brazed and copper plated. The gage is
placed on the test piece copper side up and the test piece is magnetized.
After particles are applied and the excess removed, the indications provide
the inspector the orientation of the magnetic field. Pie gages are mainly
used on flat surfaces such as weldments or steel castings where dry powder
is used with a yoke or prods. The pie gage is not recommended for precision
parts with complex shapes, for wet-method applications, or for proving
field magnitude. The gage should be demagnetized between readings.

Slotted Strips
Slotted strips are pieces of highly permeable ferromagnetic material with slots of
different widths. These strips can be used with the wet or dry method. They are placed

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 25 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
on the test object as it is inspected. The indications produced on the strips give the
inspector a general idea of the field strength in a particular area.

Magnetic Particles
As mentioned previously, the particles that are used for magnetic particle inspection
are a key ingredient as they form the indications that alert the inspector to the
presence of defects. Particles start out as tiny milled pieces of iron or iron oxide. A
pigment (somewhat like paint) is bonded to their surfaces to give the particles color.
The metal used for the particles has high magnetic permeability and low retentivity.
High magnetic permeability is important because it makes the particles attract easily to
small magnetic leakage fields from discontinuities, such as flaws. Low retentivity is
important because the particles themselves never become strongly magnetized so
they do not stick to each other or the surface of the part. Particles are available in a dry
mix or a wet solution.

Dry Magnetic Particles


Dry magnetic particles can typically be purchased in red, black, gray, yellow and several
other colors so that a high level of contrast between the particles and the part being
inspected can be achieved. The size of the magnetic particles is also very important.
Dry magnetic particle products are produced to include a range of particle sizes. The
fine particles have a diameter of about 50 µm while the course particles have a
diameter of 150 µm (fine particles are more than 20 times lighter than the coarse
particles). This makes fine particles more sensitive to the leakage fields from very small
discontinuities. However, dry testing particles cannot be made exclusively of the fine
particles where coarser particles are needed to bridge large discontinuities and to
reduce the powder's dusty nature. Additionally, small particles easily adhere to surface
contamination, such as remnant dirt or moisture, and get
trapped in surface roughness features. It should also be
recognized that finer particles will be more easily blown away by
the wind; therefore, windy conditions can reduce the sensitivity
of an inspection. Also, reclaiming the dry particles is not
recommended because the small particles are less likely to be
recaptured and the "once used" mix will result in less sensitive
inspections.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 26 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
The particle shape is also important. Long, slender particles tend to align themselves
along the lines of magnetic force. However, if dry powder consists
only of elongated particles, the application process would be less
than desirable since long particles lack the ability to flow freely.
Therefore, a mix of rounded and elongated particles is used since
it results in a dry powder that flows well and maintains good
sensitivity. Most dry particle mixes have particles with L/D ratios
between one and two.

Wet Magnetic Particles


Magnetic particles are also supplied in a wet suspension such as water or oil. The wet
magnetic particle testing method is generally more sensitive than the dry because the
suspension provides the particles with more mobility and makes it possible for smaller
particles to be used (the particles are typically 10 µm and smaller) since dust and
adherence to surface contamination is reduced or eliminated. The wet method also
makes it easy to apply the particles uniformly to a relatively large area.
Wet method magnetic particles products differ from dry powder
products in a number of ways. One way is that both visible and
fluorescent particles are available. Most non-fluorescent particles are
ferromagnetic iron oxides, which are either black or brown in color.
Fluorescent particles are coated with pigments that fluoresce when
exposed to ultraviolet light. Particles that fluoresce green-yellow are
most common to take advantage of the peak color sensitivity of the
eye but other fluorescent colors are also available.
The carrier solutions can be water or oil-based. Water-based carriers
form quicker indications, are generally less expensive, present little or
no fire hazard, give off no petrochemical fumes, and are easier to
clean from the part. Water-based solutions are usually formulated
with a corrosion inhibitor to offer some corrosion protection.
However, oil-based carrier solutions offer superior corrosion and
hydrogen embrittlement protection to those materials that are prone
to attack by these mechanisms.
Also, both visible and fluorescent wet suspended particles are available in aerosol
spray cans for increased portability and ease of application.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 27 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Dry Particle Inspection
In this magnetic particle testing technique, dry particles are
dusted onto the surface of the test object as the item is
magnetized. Dry particle inspection is well suited for the
inspections conducted on rough surfaces. When an
electromagnetic yoke is used, the AC current creates a pulsating
magnetic field that provides mobility to the powder.
Dry particle inspection is also used to detect shallow subsurface
cracks. Dry particles with half wave DC is the best approach when
inspecting for lack of root penetration in welds of thin materials.
Steps for performing dry particles inspection:
 Surface preparation - The surface should be relatively clean but this is not as
critical as it is with liquid penetrant inspection. The surface must be free of grease,
oil or other moisture that could keep particles from moving freely. A thin layer of
paint, rust or scale will reduce test sensitivity but can sometimes be left in place
with adequate results. Specifications often allow up to 0.076 mm of a
nonconductive coating (such as paint) or 0.025 mm of a ferromagnetic coating
(such as nickel) to be left on the surface. Any loose dirt, paint, rust or scale must
be removed.

o Some specifications require the surface to be coated with a thin layer of white
paint in order to improve the contrast difference between the background and
the particles (especially when gray color particles are used).

 Applying the magnetizing force - Use permanent magnets, an electromagnetic


yoke, prods, a coil or other means to establish the necessary magnetic flux.

 Applying dry magnetic particles - Dust on a light layer of magnetic particles.

 Blowing off excess powder - With the magnetizing force still applied, remove the
excess powder from the surface with a few gentle puffs of dry air. The force of the
air needs to be strong enough to remove the excess particles but not strong
enough to remove particles held by a magnetic flux leakage field.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 28 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
 Terminating the magnetizing force - If the magnetic flux is being generated with
an electromagnet or an electromagnetic field, the magnetizing force should be
terminated. If permanent magnets are being used, they can be left in place.

 Inspection for indications - Look for areas where the magnetic particles are
clustered.

Wet Suspension Inspection


Wet suspension magnetic particle inspection, more
commonly known as wet magnetic particle inspection,
involves applying the particles while they are
suspended in a liquid carrier. Wet magnetic particle
inspection is most commonly performed using a
stationary, wet, horizontal inspection unit but
suspensions are also available in spray cans for use
with an electromagnetic yoke.
A wet inspection has several advantages over a dry inspection. First, all of the surfaces
of the component can be quickly and easily covered with a relatively uniform layer of
particles. Second, the liquid carrier provides mobility to the particles for an extended
period of time, which allows enough particles to float to small leakage fields to form a
visible indication. Therefore, wet inspection is considered best for detecting very small
discontinuities on smooth surfaces. On rough surfaces, however, the particles (which
are much smaller in wet suspensions) can settle in the surface valleys and lose mobility,
rendering them less effective than dry powders under these conditions.
Steps for performing wet particle inspection:
 Surface preparation - Just as is required with dry particle inspections, the surface
should be relatively clean. The surface must be free of grease, oil and other
moisture that could prevent the suspension from wetting the surface and
preventing the particles from moving freely. A thin layer of paint, rust or scale will
reduce test sensitivity, but can sometimes be left in place with adequate results.
Specifications often allow up to 0.076 mm of a nonconductive coating (such as
paint) or 0.025 mm of a ferromagnetic coating (such as nickel) to be left on the
surface. Any loose dirt, paint, rust or scale must be removed.

o Some specifications require the surface to be coated with a thin layer of white
paint when inspecting using visible particles in order to improve the contrast
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 29 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
difference between the background and the particles (especially when gray
color particles are used).

 Applying suspended magnetic particles - The suspension is gently sprayed or


flowed over the surface of the part. Usually, the stream of suspension is diverted
from the part just before the magnetizing field is applied.

 Applying the magnetizing force - The magnetizing force should be applied


immediately after applying the suspension of magnetic particles. When using a
wet horizontal inspection unit, the current is applied in two or three short bursts
(1/2 second) which helps to improve particle mobility.

 Inspection for indications - Look for areas where the magnetic particles are
clustered. Surface discontinuities will produce a sharp indication. The indications
from subsurface flaws will be less defined and lose definition as depth increases.

Quality & Process Control

Particle Concentration and Condition


Particle Concentration
The concentration of particles in the suspension is a very important
parameter and it is checked after the suspension is prepared and
regularly monitored as part of the quality system checks. Standards
require concentration checks to be performed every eight hours or at
every shift change.
The standard process used to perform the check requires agitating the
carrier for a minimum of thirty minutes to ensure even particle
distribution. A sample is then taken in a pear-shaped 100 ml centrifuge
tube having a graduated stem (1.0 ml in 0.05 ml increments for
fluorescent particles, or 1.5 ml in 0.1 ml increments for visible particles).
The sample is then demagnetized so that the particles do not clump
together while settling. The sample must then remain undisturbed for a
period of time (60 minutes for a petroleum-based carrier or 30 minutes
for a water-based carrier). The volume of settled particles is then read.
Acceptable ranges are 0.1 to 0.4 ml for fluorescent particles and 1.2 to
Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 30 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
2.4 ml for visible particles. If the particle concentration is out of the acceptable range,
particles or the carrier must be added to bring the solution back in compliance with the
requirement.

Particle Condition
After the particles have settled, they should be examined for brightness and
agglomeration. Fluorescent particles should be evaluated under ultraviolet light and
visible particles under white light. The brightness of the particles should be evaluated
weekly by comparing the particles in the test solution to those in an unused reference
solution that was saved when the solution was first prepared. Additionally, the
particles should appear loose and not lumped together. If the brightness or the
agglomeration of the particles is noticeably different from the reference solution, the
bath should be replaced.

Suspension Contamination
The suspension solution should also be examined for contamination which may come
from inspected components (oils, greases, sand, or dirt) or from the environment
(dust). This examination is performed on the carrier and particles collected for
concentration testing. Differences in color, layering or banding within the settled
particles would indicate contamination. Some contamination is to be expected but if
the foreign matter exceeds 30 percent of the settled solids, the solution should be
replaced. The liquid carrier portion of the solution should also be inspected for
contamination. Oil in a water bath and water in a solvent bath are the primary
concerns.

Water Break Test


A daily water break check is required to evaluate the surface wetting performance of
water-based carriers. The water break check simply involves flooding a clean surface
similar to those being inspected and observing the surface film. If a continuous film
forms over the entire surface, sufficient wetting agent is present. If the film of
suspension breaks (water break) exposing the surface of the component, insufficient
wetting agent is present and the solution should be adjusted or replaced.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 31 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
Electrical System Checks
Changes in the performance of the electrical system of a magnetic particle inspection
unit can obviously have an effect on the sensitivity of an inspection. Therefore, the
electrical system must be checked when the equipment is new, when a malfunction is
suspected, or every six months. Listed below are the verification tests required by
active standards.
Ammeter Check
It is important that the ammeter provide consistent and correct readings. If the meter
is reading low, over magnetization will occur and possibly result in excessive
background "noise." If ammeter readings are high, flux density could be too low to
produce detectable indications. To verify ammeter accuracy, a calibrated ammeter is
connected in series with the output circuit and values are compared to the
equipment's ammeter values. Readings are taken at three output levels in the working
range. The equipment meter is not to deviate from the calibrated ammeter more than
±10 percent or 50 amperes, whichever is greater. If the meter is found to be outside
this range, the condition must be corrected.
Shot Timer Check
When a timer is used to control the shot duration, the timer must be calibrated.
Standards require the timer be calibrated to within ± 0.1 second. A certified timer
should be used to verify the equipment timer is within the required tolerances.

Magnetization Strength Check


Ensuring that the magnetization equipment provides sufficient magnetic field strength
is essential. Standard require the magnetization strength of electromagnetic yokes to
be checked prior to use each day. The magnetization strength is checked by lifting a
steel block of a standard weight using the yoke at the maximum pole spacing to be
used (10 lb weight for AC yokes or 40 lb weight for DC yokes).

Lighting
Magnetic particle inspection predominately relies on visual inspection to detect any
indications that form. Therefore, lighting is a very important element of the inspection
process. Obviously, the lighting requirements are different for an inspection conducted

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 32 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
using visible particles than they are for an inspection conducted using fluorescent
particles.
Light Requirements When Using Visible Particles
Visible particles inspections can be conducted using natural lighting or artificial lighting.
However, since natural daylight changes from time to time, the use of artificial lighting
is recommended to get better uniformity. Artificial lighting should be white whenever
possible (halogen lamps are most commonly used). The light intensity is required to be
100 foot-candles (1076 lux) at the surface being inspected.

Light Requirements When Using Fluorescent Particles

Ultraviolet Lighting
When performing a magnetic particle inspection using fluorescent particles, the
condition of the ultraviolet light and the ambient white light must be monitored.
Standards and procedures require verification of lens condition and light intensity.
Black lights should never be used with a cracked filter as the output of white light and
harmful black light will be increased. Also, the cleanliness of the filter should also be
checked regularly. The filter should be checked visually and cleaned as necessary
before warming-up the light. Most UV light must be warmed up prior to use and
should be on for at least 15 minutes before beginning an inspection.
For UV lights used in component evaluations, the normally accepted intensity is 1000
µW/cm2 at 38cm distance from the filter face. The required check should be
performed when a new bulb is installed, at startup of the inspection cycle, if a change
in intensity is noticed, or every eight hours of continuous use.

Ambient White Lighting


When performing a fluorescent magnetic particle inspection, it is important to keep
white light to a minimum as it will significantly reduce the inspector’s ability to detect
fluorescent indications. Light levels of less than 2 foot-candles (22 lux) are required by
most procedures. When checking black light intensity a reading of the white light
produced by the black light may be required to verify white light is being removed by
the filter.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 33 of 34

These materials have been adapted from https://www.nde-ed.org and are reused with permission.
White Light for Indication Confirmation
While white light is held to a minimum in fluorescent inspections, procedures may
require that indications be evaluated under white light. The white light requirements
for this evaluation are the same as when performing an inspection with visible
particles. The minimum light intensity at the surface being inspected must be 100
foot-candles (1076 lux).

Light Measurement
Light intensity measurements are made using a radiometer (an instrument that
transfers light energy into an electrical current). Some radiometers have the ability to
measure both black and white light, while others require a separate sensor for each
measurement. Whichever type is used, the sensing area should be clean and free of
any materials that could reduce or obstruct light reaching the sensor. Radiometers are
relatively unstable instruments and readings often change considerable over time.
Therefore, they should be calibrated at least every six months.

Introduction to Non-Destructive Testing Techniques Instructor: Dr. Ala Hijazi

Magnetic Particle Testing Page 34 of 34


Best PDF Encryption Reviews
These materials have been adapted from https://www.nde-ed.org and are reused with permission.

You might also like