DO Analyser Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

Quick Start Guide

00825-0100-3493, Rev AA
May 2019

Rosemount™ 499ATrDO

Trace Dissolved Oxygen Sensor


Quick Start Guide May 2019

Safety information

CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.

CAUTION
Equipment damage
Do not exceed pressure and temperature specifications
Pressure: 65 psig (549 kPa abs) max.
Temperature: 32 to 122 °F (0 to 50 °C)

WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of
end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true
for all systems used within the facility.

Contents
First steps..................................................................................................................................... 3
Install........................................................................................................................................... 6
Wire............................................................................................................................................. 8
Calibrate.................................................................................................................................... 15
Maintenance.............................................................................................................................. 17
Accessories................................................................................................................................ 20

2 Emerson.com/Rosemount
May 2019 Quick Start Guide

1 First steps
1.1 Unpack and inspect
Procedure
1. Inspect the shipping container. If it is damaged, contact the shipper
immediately for instructions.
2. If there is no apparent damage, unpack the container. Be sure all
items shown on the packing list are present. If items are missing,
notify Emerson immediately.

1.2 Product description


Figure 1-1: Rosemount™ 499ATrDOSensor Parts

A. Membrane retainer
B. Membrane assembly
C. O-ring
D. Cathode
E. Electrolyte fill plug (wrap with pipe tape)
F. Pressure equalizing port
G. Sensor cable (integral cable shown)

1.3 Specifications
Table 1-1: Sensor Specifications

Physical characteristics Specifications

Range 0.1 ppb to 20 ppm

Pressure 0 to 65 psig (101 to 549 kPa abs)

Temperature (operating) 0 to 50 °C (32 to 122 °F)

Process connection 1 in. MNPT

Quick Start Guide 3


Quick Start Guide May 2019

Table 1-1: Sensor Specifications (continued)


Physical characteristics Specifications

Wetted parts Noryl®, Viton, EPDM, Teflon® , and silicone

Cathode Gold (not normally wetted)

Accuracy Accuracy depends on the accuracy of the chemical test used to


calibrate the sensor.

Linearity ±5% of reading or ±3 ppb (whichever is greater) at 25 °C (77 °F)

Repeatability ±2% of reading at constant temperature

Response time < 20 sec to 90% of final reading at 25 °C (77 °F) (0 to 200 ppb
oxygen)

Membrane permeability Defined between 0 and 50 °C (32 and 122 °F)


connection

Electrolyte volume 25 mL (approx.)

Electrolyte life 4 to 6 months (approx.)

Sensor life 2 years (approx.)

Accuracy at 25 °C (77 °F) <20 ppb: ±1 ppb


following air calibration >20 ppb: ±5% of reading

Cable length (standard 25 ft (7.6 m)


integral cable)

Cable length (maximum) 300 ft (91 m)

Drift <4% over 60 days

Sample flow 1.6-6.3 gph (100-400 mL/min) recommended. Response changes


less than 2% when flow is maintained at recommended range. At
0.8 gph (50 mL/min), sensor response is about 90% of value at 100
mL/min.

Comedown time to 1 ppb New sensor: < 5 hours


Following membrane change: < 1 hour
Following air calibration: < 1 hour

Shelf life 3 months. If the shelf life has been exceeded, the sensor can still be
used. The comedown time will be longer.

Weight/shipping weight 0.5 kg / 1.5 kg (1 lb / 3 lb)

4 Emerson.com/Rosemount
May 2019 Quick Start Guide

Table 1-2: Other Specifications

Type PN Wetted Process Maximum Maximum


materials connection temperature pressure

Low flow 24091-00 and Polycarbonat Compression 70 °C (158 °F) 90 psig (722
cell (1) 24091-01 e/polyester, fitting for 1/4 kPa abs)
316 stainless in. O.D.
steel, and tubing or 1/4
silicone in. FNPT

(1) Temperature and pressure specifications for the low flow cell exceed the temperature and
pressure specifications for the sensor.

Quick Start Guide 5


Quick Start Guide May 2019

2 Install
The gray PVC cap contains a solution of sodium sulfite. Remove the cap
before installing the sensor.

WARNING
POISONOUS SUBSTANCE
The cap contains sodium sulfite solution. Avoid contact with skin or eyes. Do
not swallow!

Figure 2-1: Sensor Orientation

6 Emerson.com/Rosemount
May 2019 Quick Start Guide

Figure 2-2: Low Flow Cell (PN 24091-00)

Quick Start Guide 7


Quick Start Guide May 2019

3 Wire
NOTICE
For additional wiring information on this product, including sensor
combinations not shown here, please refer to the Liquid Transmitter Wiring
Diagrams.

Figure 3-1: Rosemount™ 499ATrDO-54 Sensor Wiring to Rosemount


1056 and 56 Transmitters

A. White
B. Green
C. Red
D. Clear
E. Black
F. Resistance temperature device return
G. Resistance temperature device sense
H. Resistance temperature device in
I. Resistance temperature device shield
J. Out
K. Out
L. Anode shield
M. Anode
N. Cathode shield
O. Cathode
P. Clear (not used)

8 Emerson.com/Rosemount
May 2019 Quick Start Guide

Figure 3-2: Rosemount 499ATrDO-54-60 and Rosemount 499ATrDO-54-


VP Sensor Wiring to Rosemount 1056 and 56 Transmitters

A. White
B. White/red
C. Red
D. Clear
E. Gray
F. Orange
G. Resistance temperature device return
H. Resistance temperature device sense
I. Resistance temperature device in
J. Resistance temperature device shield
K. Out
L. Out
M. Anode shield
N. Anode
O. Cathode shield
P. Cathode

Quick Start Guide 9


Quick Start Guide May 2019

Figure 3-3: Rosemount 499ATrDO-54 Sensor Wiring to Rosemount 5081


Transmitter

10 Emerson.com/Rosemount
May 2019 Quick Start Guide

Figure 3-4: Rosemount 499ATrDO-60 and 499ATrDO-VP Sensor Wiring


to Rosemount 5081 Transmitter

Quick Start Guide 11


Quick Start Guide May 2019

Figure 3-5: Rosemount 499ATrDO-54 Sensor Wiring to Rosemount 1066


Transmitter

Note
Connect clear shield wires to sol gnd terminal on TB 2. Use wire nut and
pigtail if necessary.

12 Emerson.com/Rosemount
May 2019 Quick Start Guide

Figure 3-6: Rosemount 499ATrDO-60 and 499ATrDO-54-VP Sensor


Wiring to Rosemount 1066 Transmitter

Note
Connect clear shield wires to solution ground terminal on TB 2. Use wire nut
and pigtail if necessary.

Figure 3-7: Rosemount 499ATrDO Sensor Pin-out Diagram

Quick Start Guide 13


Quick Start Guide May 2019

When making a connection through a junction box (PN 23550-00), wire


point-to-point.

NOTICE
Use a wire nut and pigtail (included) when connecting several wires to the
same terminal.

14 Emerson.com/Rosemount
May 2019 Quick Start Guide

4 Calibrate
4.1 Zero point calibration
Even in the absence of oxygen, the Rosemount 499ATrDO sensor generates
a small signal called the zero current. Normally, the zero current is less than
5 nA, which introduces nor more than a 0.5 ppb error in measurement. Zero
the sensor when it is first placed in service and every time the fill solution is
changed.
To zero the sensor:

Procedure
1. Pour a cup of deionized or bottled water.
2. Add a teaspoon of sodium sulfite to the water.
3. Place the sensor in the water.
4. Wait until the sensor current has reached a stable low value (at least
two hours).
5. Measure the current.
a) If it is less than 5 nA, do not zero the sensor.
b) If it is between 5 and 10 nA, allow the sensor to run overnight.
Zero the sensor if the reading is still between 5 and 10 nA and
is stable.
c) If it is greater than 10 nA, call the factory.

4.2 Full scale


The Rosemount 499ATrDO sensor is best calibrated by exposing the sensor
to water-saturated air.

Procedure
1. Pour a small amount of water into a cup.
2. Suspend the sensor, keeping the membrane dry, about 1/4 in. (6
mm) above the surface of the water.
3. Once readings are stable, which should take no longer than 20
minutes, follow the analyzer prompts to complete the calibration.
The analyzer automatically calculates the equilibrium solubility of
atmospheric oxygen in water under the prevailing temperature and
barometric pressure.
4. After calibration, go to the Diagnostics menu and check the
sensitivity.
The sensitivity should be between 3,600 and 6,100 nA/ppm.

Quick Start Guide 15


Quick Start Guide May 2019

For more information, refer to the transmitter manual.


Prolonged exposure to air may affect the linearity of the sensor. If the
cumulative exposure to air is less than about five hours per year,
sensor linearity should be within specification. If cumulative exposure
exceeds five hours per year, restore linear response of the sensor by
draining the electrolyte solution and replacing it with fresh.

16 Emerson.com/Rosemount
May 2019 Quick Start Guide

5 Maintenance
Periodic maintenance and cleaning are required for best performance of the
sensor. Generally, the membrane and fill solution should be replaced every
four to six months. The optimum maintenance frequency is best determined
by experience. Periodically check the zero current and sensitivity. If the zero
current is less than about 5 nA, and the current in the air is between 30 and
45 nA with a variability of less than 2%, the sensor does not need
maintenance.

WARNING
PRESSURIZED SPRAY INJURY
Before removing the sensor, be absolutely certain that the process pressure
is reduced to 0 psig and the process temperature is lowered to a safe level!

5.1 Cleaning the membrane


Keep the membrane clean and free from solid corrosion products. Clean the
membrane with water sprayed from a wash bottle or gently wipe the sensor
with a soft, clean tissue.

5.2 Replacing the electrolyte solution and membrane


WARNING
HARMFUL SUBSTANCE
Fill solution may cause irritation. May be harmful if swallowed. Read and
follow manual.

Procedure
1. Unscrew the membrane retainer.
2. Remove the membrane assembly and O-ring.
See Figure 1-1.
3. Hold the sensor over a container with the cathode pointing down.
4. Remove the fill plug.
5. Allow the electrolyte solution to drain out.
6. Inspect the cathode.
a) If it is tarnished, clean it by gently rubbing in the direction of
the existing scratches (do not use a circular motion) with
400-600 grit silicon carbide finishing paper.

7. Remove the old pipe tape from the plug.

Quick Start Guide 17


Quick Start Guide May 2019

8. Wrap the plug with one or two turns of pipe tape..


9. Prepare a new membrane.
a) Hold the membrane assembly with the cup formed by the
membrane and membrane holder pointing up.
b) Place a drop of isopropyl alcohol in the cup.
c) Slowly add about 20 drops of electrolyte solution to the cup.
This step is important, because alcohol wets the inside
surface of the membrane and ensures that no air bubbles will
be trapped when the membrane assembly is placed over the
cathode.

d) Fill the cup with electrolyte solution.


e) Leave the membrane assembly filled with electrolyte solution
and set it aside.

10. Hold the sensor at about a 45° angle with the cathode end pointing
up.
11. Add electrolyte solution through the fill hole until the liquid
overflows.
12. Tap the sensor near the threads to release trapped air bubbles.
13. Add more electrolyte solution if necessary.
14. Place the fill plug in the electrolyte port and begin screwing it in.
15. After several threads have engaged, rotate the sensor so that the
cathode is pointing up and continue tightening the fill plug.
Do not overtighten.
16. Place a new O-ring in the groove around the cathode post.
17. Cover the holes at the base of the cathode stem with several drops of
electrolyte solution.
18. Insert a small blunt probe, like a toothpick with the end cut off,
through the pressure equalizing port.
See Figure 1-1.

CAUTION
EQUIPMENT DAMAGE
Do not use a sharp probe. It will puncture the bladder and destroy
the sensor.

19. Gently press the probe against the bladder several times to force
liquid through the holes at the base of the cathode stem. Keep

18 Emerson.com/Rosemount
May 2019 Quick Start Guide

pressing the bladder until no air bubbles can be seen leaving the
holes. Be sure the holes remain covered with electrolyte solution.
20. Place a drop of electrolyte solution on the cathode; then place the
membrane assembly over the cathode.
21. Screw the membrane retainer in place.
The sensor may require several hours operating at the polarizing
voltage to equilibrate after the electrolyte solution has been
replenished.

5.3 Storage
Store the sensor with the membrane immersed in a fresh solution of
saturated sodium sulfite. You can use the PVC cap shipped with the sensor.
Leave the power to the transmitter turned on.

Quick Start Guide 19


Quick Start Guide May 2019

6 Accessories
Part number Description

23747-06 Interconnecting cable, VP 6, 2.5 ft (0.8 m)

23747-04 Interconnecting cable, VP 6, 4 ft (1.2 m)

23747-02 Interconnecting cable, VP 6, 10 ft (3.0 m)

23747-07 Interconnecting cable, VP 6, 15 ft (4.6 m)

23747-08 Interconnecting cable, VP 6, 20 ft (6.1 m)

23747-09 Interconnecting cable, VP 6, 25 ft (7.6 m)

23747-10 Interconnecting cable, VP 6, 30 ft (9.1 m)

23747-03 Interconnecting cable, VP 6, 50 ft (15.2 m)

23747-11 Interconnecting cable, VP 6, 100 ft (30.5 m)

24091-00 Low flow cell with 1/4 in. OD tubing compression


fittings

9390004 Rotameter: 0.5 - 5.0 gph

22550-00 Junction box, 12 terminals

9200266 Extension cable, standard, unprepped

9200275 Extension cable for optimum EMI/RFI cable, unprepped

23747-00 Extension cable for optimum EMI/RFI cable, prepped

23501-04 Trace dissolved oxygen membrane kit: includes 1


membrane assembly and 1 O-ring

25302-04 Trace dissolved oxygen membrane kit: includes 3


membrane assemblies and 3 O-rings

9210264 Dissolved oxygen sensor fill solution, 4 oz (125 mL)

33521-02 Membrane retainer

33523-03 Electrolyte fill plug

9390094 O-ring, Viton 2-014

20 Emerson.com/Rosemount
May 2019 Quick Start Guide

Quick Start Guide 21


Quick Start Guide May 2019

22 Emerson.com/Rosemount
May 2019 Quick Start Guide

Quick Start Guide 23


*00825-0100-3493*
Quick Start Guide
00825-0100-3493, Rev. AA
May 2019

GLOBAL HEADQUARTERS NORTH AMERICA


Emerson Automation Solutions Emerson Automation Solutions
6021 Innovation Blvd 8200 Market Blvd
Shakopee, MN 55379, USA Chanhassen, MN 55317
+1 800 999 9307 or +1 952 906 8888 Toll Free +1 800 999 9307
F +1 952 949 7001 F +1 952 949 7001
[email protected] [email protected]

EUROPE MIDDLE EAST AND AFRICA


Emerson Automation Solutions Emerson Automation Solutions
Neuhofstrasse 19a P.O. Box 1046 Emerson FZE
CH-6340 Baar Jebel Ali Free Zone
Switzerland Dubai, United Arab Emirates, P.O. Box
T + 41 (0) 41 768 6111 17033
F + 41 (0) 41 768 6300 T +971 4 811 8100
[email protected] F +971 4 886 5465
[email protected]

©2019 Emerson. All rights reserved.


Linkedin.com/company/Emerson-
Automation-Solutions The Emerson logo is a trademark and service
twitter.com/rosemount_news mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson family of companies.
Facebook.com/Rosemount
All other marks are the property of their
youtube.com/RosemountMeasurement respective owners.

You might also like