Gs01c50h01-01en - 012 Yta610
Gs01c50h01-01en - 012 Yta610
Gs01c50h01-01en - 012 Yta610
General YTA610
Specifications Temperature Transmitter
GS 01C50H01-01EN
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 3
Safety Requirement Standards ■ PHYSICAL SPECIFICATIONS
EN61010-1, C22.2 No.61010-1
• Installation category: I Enclosure
(Anticipated transient overvoltage 330 V) Material & coating
• Pollution degree: 2 • Low copper cast aluminum alloy
• Indoor/Outdoor use [for aluminum housing]
EN61010-2-030, C22.2 No.61010-2-030 Urethane-cured polyester powder coating
• Measurement category: O (Other) Mint-green paint (Munsell 5.6BG 3.3/2.9 or its
(Measurement Input voltage: 150mVdc max) equivalent)
EU RoHS Directive [for option code /P or /X2]
Applicable standard: EN 50581 Epoxy and polyurethane resin solvent coating
• ASTM CF-8M Stainless steel
■ NORMAL OPERATING CONDITION Degrees of protection
(Optional features or approval codes may affect limits.) IP66/IP67, TYPE 4X
Ambient Temperature Limits Name plate and tag
–40 to 85°C (–40 to 185°F) 316 SST
–30 to 80°C (–22 to 176°F) (with indicator model) Mounting
Ambient Humidity Limits Optional mounting brackets can be used either for
0 to 100% RH at 40°C (104°F) two-inch pipe or flat panel mounting.
Supply Voltage Requirements Terminal screws
M4 screws
HART type
10.5 to 42 V DC for general use and flameproof type Integral Indicator (with indicator model)
10.5 to 32 V DC for lightning protector (option code /A) 5-digit numerical display, 6-digit unit display and bar
10.5 to 30 V DC for intrinsically safe and graph.
non-incendive type Local Parameter Setting (with indicator model)
Minimum voltage limited at 16.6 V DC for digital Parameter configuration by the push button offers
communications HART easy and quick setup for parameters.
With 24 V DC supply, up to a 550Ω load can be Accessible parameters are different with each output
used. See graph below. cord.
Weight
Aluminum housing:
1.3 kg (2.9 lb) without integral indicator and mounting
600 Integral indicator: 0.2 kg (0.4 lb)
Bracket for horizontal pipe: 0.3 kg
E-10.5
R= Bracket for vertical pipe: 1.0 kg
External 0.0244 Digital
load Communication
Stainless housing:
resistance range 3.1 kg (6.8 lb) without integral indicator and mounting
HART Integral indicator: 0.3 kg (0.7 lb)
250 Connections
Refer to “MODEL AND SUFFIX CODE.”
R (Ω)
Fieldbus type
9 to 32V DC for general use, flameproof type, and
non-incendive type
9 to 30 V DC for intrinsically safe type
9 to 17.5 V DC for FISCO field device
Communication Requirements
Supply Voltage: 9 to 32 V DC
Current Draw:
Steady state: 15 mA (max)
Software download state: 24 mA (max)
Load Requirements (HART type)
0 to 1290Ω for operation
250 to 600Ω for digital communication
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 4
Table 1. Sensor type, measurement range, and accuracy
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Jan. 15, 2018-00
<<Contents>> <<Index>> 5
Table 2. Temperature coefficient
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Sep. 30, 2016-00
<<Contents>> <<Index>> 6
MODEL AND SUFFIX CODES
Model Suffix Codes Description
YTA610 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Temperature Transmitter
Output signal -J. . . . . . . . . . . . . . . . . . . . . . . . . . . 4 to 20 mA DC with digital communication HART protocol
-F. . . . . . . . . . . . . . . . . . . . . . . . . . . Digital communication (FOUNDATION Fieldbus protocol)
— A. . . . . . . . . . . . . . . . . . . . . . . Always A
Sensor input 1 . . . . . . . . . . . . . . . . . . . . Single
2 . . . . . . . . . . . . . . . . . . . . Double
Housing code A. . . . . . . . . . . . . . . . Aluminum
C. . . . . . . . . . . . . . . . Stainless
Electrical connection 0 . . . . . . . . . . . . . G1/2 female
2 . . . . . . . . . . . . . 1/2 NPT female
4 . . . . . . . . . . . . . M20 female
Integral indicator D. . . . . . . . . . Digital indicator with Local Operating Switch
N. . . . . . . . . . None
Mounting bracket B. . . . . . SUS304 stainless steel 2-inch horizontal pipe mounting bracket *1
D. . . . . . SUS304 stainless steel 2-inch vertical pipe mounting bracket
J . . . . . . SUS316 stainless steel 2-inch horizontal pipe mounting bracket *1
K. . . . . . SUS316 stainless steel 2-inch vertical pipe mounting bracket
N. . . . . . None
Option codes / Optional specification
*1: For flat-panel mounting, please prepare bolts and nuts.
OPTIONAL SPECIFICATION
Item Description Code
Lightning protector *4 Allowable current: Max. 6000A(8×20µs), repeating 1000A(8×20µs), 100 times A
Painting*1*6 Color and coating change Color: Munsell code N1.5 Black
P1
Amplifier cover only*3 Coating: High anti-corrosion coating
Color: Munsell code 7.5BG4/1.5, Jade green
P2
Coating: High anti-corrosion coating
Color: Metallic silver
P7
Coating: High anti-corrosion coating
Color and coating change Color: Munsell code 7.5 R4/14, Red
PR
Amplifier and terminal Covers*3 Coating: High anti-corrosion coating
Coating change High anti-corrosion coating X2
Output signal low-side in Output signal Low-side: –5%, 3.2 mA DC or less.
C1
transmitter failure*2 Sensor burnout is also set to 'Low': –2.5%, 3.6 mA DC.
NAMUR NE43 compliant*2 Output signal limits: Failure alarm down-scale: output status at CPU failure and
3.8 mA to 20.5 mA hardware error is –5%, 3.2 mA or less. C2
Sensor burnout is also set to Low: –2.5%, 3.6 mA DC.
Failure alarm up-scale: output status at CPU failure and hardware
error is 110%, 21.6 mA or more. C3
In this case Sensor burnout is High: 110%, 21.6 mA DC.
Data configuration*2 Description into “Descriptor” parameter of HART protocol (max. 16 characters) CA
Wired tag plate SUS316 stainless steel tag plate wired onto transmitter N4
Sensor matching RTD sensor matching function CM1
Attached flameproof Electrical connection G1/2 female
2pc. V52
packing adapter*5*7 Applicable cable: O.D.8.0 to 12 mm
EAC approval and Russian EAC approval and Russian pattern approval marking
VR
pattern approval marking*8
EAC approval marking EAC approval marking without Russian pattern approval marking
without Russian pattern VE
approval marking*8
Note: The indication of the nameplate shows an initial shipment state.
*1: Not applicable for Stainless housing.
*2: Not applicable for Fieldbus type.
*3: Except for Amplifier and terminal cover, color and coating are general specification.
*4: Lightning protector (surge absorber) can be remove from, or added to the equipment.
*5: Combination with other Explosion protected other than TIIS flameproof is not possible.
*6: The combination of X2 and P is not possible.
*7: Applicable for Electrical Connection code 4. (The thread of connection between YTA and CABLE GLAND is M20,
and the thread of connection between CABLE GLAND and CABLE is G1/2.)
*8: Combination with other Explosion protected other than EAC Explosion-proof type is not applicable.
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 7
OPTIONAL SPECIFICATION (For Explosion Protected type)
Item Description Code
ATEX 4-20mA:
[Intrinsically safe ia approval]
Applicable Standard: EN 60079-0:2012+A11:2013, EN 60079-11:2012
Certificate No. FM16ATEX0019X
II 1 G Ex ia IIC T5…T4 Ga
Ambient Temperature: –40 to 70°C for T4, –40 to 50°C for T5
Enclosure: IP66/IP67
Electrical parameters:
Supply/Output circuit: Terminals: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Intrinsically safe ic]
Applicable Standard: EN 60079-0:2012+A11:2013, EN 60079-11:2012
Certificate: Not Applicable as per Annex VIII to ATEX 2014/34/EU
II 3 G Ex ic IIC T5...T4 Gc
KU2
Ambient Temperature: –30 to 70°C for T4, –30 to 50°C for T5
Enclosure: IP66/IP67
Overvoltage category: I
Electrical parameters:
Supply/Output circuit: Terminals: +, –
Ui=32V, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Flameproof and Dust Ignition Proof Approval]
Applicable Standard: EN 60079-0:2012+A11:2013, EN 60079-1:2014, EN 60079-31:2014
Certificate: KEMA 07ATEX0130
II 2 G Ex db IIC T6/T5 Gb, II 2 D Ex tb IIIC T70°C/T90°C Db
Ambient Temperature for Gas Atmospheres: –40 to 75°C for T6, –40 to 80°C for T5
Ambient Temperature for Dust Atmospheres: –30 to 65°C for T70°C, –30 to 80°C for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
Fieldbus:
[Intrinsically safe ia approval]
Applicable Standard: EN 60079-0:2012+A11:2013, EN 60079-11:2012
Certificate No. FM16ATEX0019X
II 1 G Ex ia IIC T4 Ga
Ambient Temperature: –55 to 60°C
Enclosure: IP66/IP67
Electrical parameters:
Supply/Output circuit: Terminals: +, –
FISCO field device or
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Intrinsically safe ic] KU25
Applicable Standard: EN 60079-0:2012+A11:2013, EN 60079-11:2012
Certificate: Not Applicable as per Annex VIII to ATEX 2014/34/EU
II 3 G Ex ic IIC T4 Gc
Ambient Temperature: –30 to 70°C
Enclosure: IP66/IP67
Electrical parameters:
Supply/Output circuit: Terminals: +, –
FISCO field device or
Ui=32V, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Flameproof and Dust Ignition Proof Approval]
Same as KU2
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 8
Item Description Code
IECEx 4-20mA:
[Intrinsically safe ia, ic pproval]
Applicable Standard: IEC 60079-0:2011, IEC 60079-11:2011
Certificate No. IECEx FMG 16.0014X
Ex ia IIC T5…T4 Ga
Ex ic IIC T5...T4 Gc
Ambient Temperature: –40 to 70°C for T4, –40 to 50°C for T5(Ex ia)
Ambient Temperature: –30 to 70°C for T4, –30 to 50°C for T5(Ex ic)
Enclosure: IP66/IP67
Over voltage category: I
Electrical parameters (Ex ia)
Supply/Output circuit: Terminal: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Electrical parameters (Ex ic) SU2
Supply/Output circuit: Terminal: +, –
Ui=30V, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Flameproof and Dust Ignition Proof Approval]
Applicable standard: IEC 60079-0:2011, IEC 60079-1:2014, IEC 60079-31:2013
Certificate: IECEx KEM 07.0044
Ex db IIC T6/T5 Gb, Ex tb IIIC T70°C/T90°C Db
Ambient Temperature for Gas Atmospheres: –40 to 75°C (–40 to 167°F) for T6,
–40 to 80°C (–40 to 176°F) for T5
Ambient Temperature for Dust Atmospheres: –30 to 65°C (–22 to 149°F) for T70°C,
–30 to 80°C (–22 to 176°F) for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
Fieldbus:
[Intrinsically safe ia approval]
Applicable Standard: IEC 60079-0: 2011, IEC 60079-11: 2011
Certificate No. IECEx FMG 16.0014X
Ex ia IIC T4 Ga
Ex ic IIC T4 Gc
Ambient Temperature: –55 to 60°C (Ex ia)
Ambient Temperature: –30 to 70°C (Ex ic)
Enclosure: IP66/IP67
Overvoltage category: I
Electrical parameters (Ex ia)
Supply/Output circuit: Terminals: +, –
FISCO field device or SU25
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Electrical parameters (Ex ic)
Supply/Output circuit: Terminals: +, –
FISCO field device or
Ui=32V, Ci=2.2nF, Li=0mH
Sensor circuit:Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Flameproof and Dust Ignition Proof Approval]
Same as SU2
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 9
Item Description Code
FM (US) 4-20mA:
[Intrinsically safe approval/non-incendive approval]
Applicable standard: FM Class 3600:2011, FM Class 3610:2015, FM Class 3611:2004,
FM Class 3810:2005, ANSI/ISA-60079-0:2013, ANSI/ISA-60079-11:2014, NEMA 250:2003,
ANSI/IEC 60529:2004
Intrinsically safe for
Class I, II, III Division 1, Groups A, B, C, D, E, F, G, T5…T4
Class I Zone 0 AEx ia IIC T5…T4
Non-incendive for
Class I, II, Division 2, Groups A, B, C, D, F, G, T5…T4
Class III, Division 1 T5…T4
Class I Zone 2 Group IIC T5…T4
Ambient Temperature: –40 to 70°C for T4, –40 to 50°C for T5
Enclosure Type 4X, IP66/IP67
Electrical parameters:
Intrinsically safe for
Supply/Output circuit: Terminals: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
FU1
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Non-incendive for
Supply/Output circuit: Terminals: +, –
Ui=30V, Ci=22nF, Li=0mH
Sensor circuit:Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Explosionproof approval]
Applicable standard: FM Class 3600: 2011, FM Class 3615: 2006, FM Class 3810: 2005,
NEMA250: 2014
Class I, Division 1, Groups A, B, C and D.;
Class II/III, Division 1, Groups E, F and G.
"FACTORY SEALD, CONDUIT SEAL NOT REQUIRED."
Enclosure Ratings: TYPE 4X
Temperature Class: T6
Ambient Temperature: –40 to 60°C (–40 to 140°F)
Electrical Connection: 1/2NPT female*2
Fieldbus:
[Intrinsically safe approval/non-incendive approval]
Applicable standard: FM Class 3600:2011, FM Class 3610:2015, FM Class 3611:2004,
FM Class 3810:2005, ANSI/ISA-60079-0:2013, ANSI/ISA-60079-11:2014, NEMA 250:2003,
ANSI/IEC 60529:2004
Intrinsically safe for
Class I, II, III Division 1, Groups A, B, C, D, E, F, G T4
Class I Zone 0 AEx ia IIC T4
Non-incendive for
Class I, II, Division 2, Groups A, B, C, D, F, G T4
Class III, Division 1 T4 Class I Zone 2 Group IIC T4
Ambient Temperature: –55 to 60°C
Enclosure Type 4X, IP66/IP67
Electrical parameters:
FU15
Intrinsically safe for
Supply/Output circuit: Terminals: +, –
FISCO field device or
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Non-incendive for
Supply/Output circuit: Terminals: +, –
Ui=32V, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Explosionproof approval]
Same as FU1
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 10
Item Description Code
CSA*3 4-20mA:
[Intrinsically safe approval/non-incendive approval]
FM (Canada)*4 Applicable standard: CAN/CSA-C22.2 No. 94.2-07, C22.2 No.213:1987,
CAN /CSA-C22.2 No. 60079-0:11, CAN/CSA-C22.2 No. 60079-11:14,
CAN/CSA-C22.2 No. 60529:05, CAN/CSA-C22.2 No. 61010-1-12,
CAN/CSA-C22.2 No. 61010-2-030-12
Intrinsically safe for
Class I, II, III, Division 1, Groups A, B, C, D, E, F, G, T5…T4
Ex ia IIC T5…T4 Ga
Non-incendive for
Class I, II, Division 2, Groups A, B, C, D, F, G T5…T4
Class III Division 1 T5…T4
Ambient Temperature: –40 to 70°C for T4, –40 to 50°C for T5
Enclosure Type: 4X, IP66/IP67
Electrical parameters:
Intrinsically safe for
Supply/Output circuit: Terminals: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH CU1
Non-incendive for
Supply/Output circuit: Terminals: +, –
Ui=30V, Ci=22nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Explosionproof approval]
Applicable standard: C22.2 No. 0-10, C22.2 No. 0.4-04, C22.2 No. 25-M1966, C22.2 No. 30-M1986,
C22.2 No. 94-M1991, C22.2 No. 142-M1987, C22.2 No. 157-92, C22.2 No. 213-M1987,
C22.2 No.61010-1-12, C22.2 No. 61010-2-030-12
Class I, Groups B, C and D,
Class II, Groups E, F and G,
Class III.
For Class I, Division2 Groups ABCD Locations “FACTORY SEALED, CONDUIT SEAL NOT
REQUIRED”
Enclosure TYPE 4X
Temperature Class: T6
Ambient Temperature: –40 to 60°C
Electrical Connection: 1/2 NPT female*2
Fieldbus:
[Intrinsically safe approval/non-incendive approval]
Applicable standard: CAN/CSA-C22.2 No. 94.2-07, C22.2 No.213:1987,
CAN /CSA-C22.2 No. 60079-0:11, CAN/CSA-C22.2 No. 60079-11:14,
CAN/CSA-C22.2 No. 60529:05, CAN/CSA-C22.2 No. 61010-1-12,
CAN/CSA-C22.2 No. 61010-2-030-12
Intrinsically safe for
Class I, II, III, Division 1, Groups A, B, C, D, E, F, G T4
Ex ia IIC T4 Ga
Non-incendive for
Class I, II, Division 2, Groups A, B, C, D, F, G T4
Class III Division 1 T4
Ambient Temperature : –55 to 60°C
Enclosure Type: 4X, IP66/IP67
Electrical parameters: CU15
Intrinsically safe for
Supply/Output circuit: Terminals: +, –
FISCO field device or
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Non-incendive for
Supply/Output circuit: Terminals: +, –
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
[Explosionproof approval]
Same as CU1
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
<<Contents>> <<Index>> 11
Item Description Code
NEPSI 4-20mA and Fieldbus:
[Flameproof and Dust Ignition Proof Approval]
Applicable Standard: GB3836.1-2010, GB3836.2-2010, GB12476.1-2013, GB12476.5-2013
Certificate No. GYJ16.1396X
Ex d IIC T6/T5 Gb, Ex tD A21 IP66/IP67 T70°C/T90°C NF2
Ambient Temperature for Gas Atmospheres: –40 to 75°C for T6, –40 to 80°C for T5
Ambient Temperature for Dust Atmospheres: –30 to 65°C for T70°C, –30 to 80°C for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
4-20mA:
[Intrinsically safe approval ]
Applicable Standard: GB 3836.1-2010, GB 3836.4-2010 GB 3836.20-2010
Certificate No. GYJ16.1423X
Ex ia IIC T4/T5 Ga
Ambient Temperature: –40 to 70°C for T4, –40 to 50°C for T5
Enclosure: IP66/IP67 in accordance with only IEC 60529 NS2
Entity Parameters:
Supply/Output circuit: Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit: Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min
[+, -, C, 1, 2, 3, 4, 5] to Earth terminal
[+, -, C] to [1, 2, 3, 4, 5]
Fieldbus:
[Intrinsically safe approval]
Applicable Standard: GB 3836.1-2010, GB 3836.4-2010, GB3836.20-2010
Certificate No. GYJ16.1423X
Ex ia IIC T4 Ga
Ambient Temperature: –55 to 60°C for T4
Enclosure: IP66/IP67 in accordance with only IEC 60529
NS25
FISCO field device
Entity Parameters:
Supply/Output circuit: Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit: Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min
[+, -, 1, 2, 3, 4, 5] to Earth terminal
[+, -] to [1, 2, 3, 4, 5]
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Sep. 13, 2017-00
<<Contents>> <<Index>> 12
Item Description Code
INMETRO [4-20mA & Fieldbus: Flameproof and dust ignition proof approval]
Applicable Standard: ABNT NBR IEC 60079-0:2013 Versão Corrigida 2: 2016,
ABNT NBR IEC60079-1:2016, ABNT NBR IEC 60079-31:2014
Certificate: DEKRA 16.0009
Ex db IIC T6/T5 Gb, Ex tb IIIC T70°C/ 90°C Db UF1
Ambient Temperature for Gas: –40 to +75°C for T6, –40 to +80°C for T5
Ambient Temperature for Dust: –30 to +65°C for T70°C, –30 to +80°C for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
4-20mA:
[Intrinsically safe approval ]
Applicable Standard: ABNT NBR IEC 60079-0:2013 Versão Corrigida 2:2016,
ABNT NBR IEC 60079-11:2013
Certificate: ABNT 17.0001X
Ex ia IIC T5...T4 Ga
Ambient Temperature: –40 to 70 °C for T4, –40 to 50 °C for T5 US1
Enclosure: IP66/IP67
Supply/Output circuit:
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit:
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
Fieldbus:
[Intrinsically safe approval "ia" ]
Applicable Standard: ABNT NBR IEC 60079-0:2013 Versão Corrigida 2:2016,
ABNT NBR IEC 60079-11:2013
Certificate: ABNT 17.0001X
Ex ia IIC T4 Ga
Ambient Temperature: –55 to 60 °C
Enclosure: IP66/IP67
Supply/Output circuit:
FISCO field device and
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit:
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
US15
[Intrinsically safe approval "ic" ]
Applicable Standard: ABNT NBR IEC 60079-0:2013 Versão Corrigida 2:2016,
ABNT NBR IEC 60079-11:2013
Certificate: ABNT 17.0001X
Ex ic IIC T4 Gc
Ambient Temperature: –30 to 70 °C
Enclosure: IP66/IP67
Overvoltage category: I
Supply/Output circuit:
FISCO field device and
Ui=32V, Ci=2.2nF, Li=0mH
Sensor circuit:
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Dec. 18, 2017-00
<<Contents>> <<Index>> 13
Item Description Code
KOSHA [4-20mA & Fieldbus: Flameproof and dust ignition proof approval]
Applicable Standard: Notice of Ministry of Labor No. 2016-54, harmorized with IEC
60079-0:2011, IEC 60079-1:2014, IEC 60079-31:2013
Certificate: 17-AV4BO-0457 (Flameproof)
17-AV4BO-0458 (Dust Ignition Proof)
PF2
Ex d IIC T6/ T5, Ex tD A21 IP66/IP67 T70°C/ T90°C
Ambient Temperature for Gas Atmospheres: –40 to 75°C for T6, –40 to 80°C for T5
Ambient Temperature for Dust Atmospheres: –30 to 65°C for T70°C, –30 to 80°C for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
4-20mA:
[Intrinsically safe approval]
Applicable Standard: Notice of Ministry of Labor No. 2016-54, harmorized with
IEC 60079-0: 2011, IEC 60079-11: 2011
Certificate: 17-AV4BO-0459X
Ex ia IIC T5…T4
Ambient Temperature: –40 to 70 °C for T4, –40 to 50 °C for T5
Enclosure: IP66/IP67
PS2
Supply/Output circuit:
Terminals: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit:
Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
Electrical Connection: 1/2 NPT female and M20 female*1
Fieldbus:
[Intrinsically safe approval]
Applicable Standard: Notice of Ministry of Labor No. 2016-54, harmorized
with IEC 60079-0: 2011, IEC 60079-11: 2011
Certificate: 17-AV4BO-0459X
Ex ia IIC T4
Ambient Temperature: –55 to 60°C
Enclosure: IP66/IP67
Supply/Output circuit: PS25
Terminals: +, –
FISCO field device and
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit:
Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
Electrical Connection: 1/2 NPT female and M20 female*1
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Jan. 15, 2018-00
<<Contents>> <<Index>> 14
Item Description Code
EAC [4-20mA & Fieldbus: Flameproof and dust ignition proof approval]
Applicable Standard: ГОСТ 31610.0-2014
ГОСТ IEC 60079-1-2013
ГОСТ IEC 60079-31-2013
Certificate: TC RU C-JP.ПБ98.B.00040
GF1*5
1Ex db IIC T6…T5 Gb, Ex tb IIIC T70 °C…T90 °C Db
Ambient Temperature for Gas Atmospheres: –40 to 75°C for T6, –40 to 80°C for T5
Ambient Temperature for Dust Atmospheres: –30 to 65°C for T70°C, –30 to 80°C for T90°C
Enclosure: IP66/IP67
Electrical Connection: 1/2 NPT female and M20 female*1
4-20mA:
[Intrinsically safe approval]
Applicable Standard: ГОСТ 31610.0-2014
ГОСТ 31610.11-2014
Certificate: TC RU C-JP.ПБ98.B.00040
0Ex ia IIC T4…T5 Ga X
Ambient Temperature: –40 to 70 °C for T4, –40 to 50 °C for T5
Enclosure: IP66/IP67
GS1*5
Supply/Output circuit:
Terminals: +, –
Ui=30V, Ii=200mA, Pi=1.0W, Ci=22nF, Li=0mH
Sensor circuit:
Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
Electrical Connection: 1/2 NPT female and M20 female*1
Fieldbus:
[Intrinsically safe approval]
Applicable Standard: ГОСТ 31610.0-2014
ГОСТ 31610.11-2014
Certificate: TC RU C-JP.ПБ98.B.00040
0Ex ia IIC T4 Ga X
Ambient Temperature: –55 to 60°C
Enclosure: IP66/IP67
Supply/Output circuit: GS15*5
Terminals: +, –
FISCO field device and
Ui=30V, Ii=300mA, Pi=1.2W, Ci=2.2nF, Li=0mH
Sensor circuit:
Terminals: 1, 2, 3, 4, 5
Uo=6V, Io=90mA, Po=135mW, Co=10μF, Lo=3.9mH
Dielectric strength: 500 V a.c.r.m.s.,1 min (Without /A)
Electrical Connection: 1/2 NPT female and M20 female*1
*1: Applicable for Electrical Connection Code 2 and 4.
*2: Applicable for Electrical Connection Code 2.
*3: For Explosionproof approval.
*4: For Intrinsically safe approval/non-incendive approval.
*5: GF1, /GS1 and /GS15 shall be combined with either /VE or /VR.
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Jan. 15, 2018-00
<<Contents>> <<Index>> 15
DIMENSIONS
Unit: mm (Approx. inch)
2-inch horizontal pipe mounting
111(4.37) 65.4(2.57) 66.1(2.60)
Electrical Connection
Electrical Connection 47.1 18.5
(Input signal) Terminal Cover
(Output signal) (1.85) (0.73)
With Indicator
(Optional)
ø93(3.66)
102(4.02)
164(6.46)
(1.57)
(For Explosionproof type)
40
Tag Plate
Horizontal Pipe
(0.98)
25
Mounting Bracket
(Optional)
ø93(3.66)
Shrouding Bolt
(For Explosionproof type)
209.5(8.25)
191.5(7.54)
Ground Terminal
Tag Plate
Vertical Pipe
Mounting Bracket
(1.81)
(Optional)
46
2-inch pipe
64(2.52)
ø60.5(ø2.38) 101(3.98)
70(2.76)
98(3.86)
F02E.ai
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Jan. 15, 2018-00
<<Contents>> <<Index>> 16
Terminals
Terminal Configuration
Input Wiring
Thermocouple and RTD and resistance Thermocouple and RTD and resistance
DC voltage (2-wire) DC voltage (3-wire)
(A) (A) (A1)
1 1 1 (+) 1
(B) (A) (B1)
2 2 2 (–) 2
(B) (B)
3 3 3 (B2) 3
(B) (B2)
4 4 4 (B2) 4
(A2)
5 5 5 (A2) 5
RTD and resistance RTD and resistance Thermocouple + RTD and resistance
(3-wire) (4-wire) RTD and resistance (2-wire)
(3-wire)
(A1)
1 (B1)
2 (B1)
3 (–)
4
5 (+)
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Jan. 15, 2018-00
<<Contents>> <<Index>> 17
< Ordering Information > 6. Output scale unit
Specify the following when ordering Model, suffix Specify only one unit from the table, ‘Settings when
codes, and optional codes. shipped.’ For sensor input code-2, specify the unit
If not specified, the instrument is shipped with the for two inputs.
settings shown in Table A and B. Specify the following 7. Tag Number
when necessary. Specify Tag number (up to 16 Characters) to be
engraved on the tag plate.
8. Software Tag (PD_TAG)
4-20mA type (HART type) Specify software tag (up to 32 Characters) to be
1. Sensor type. written on the amplifier memory.
For RTD and resistance input, specify the number 9. Node Address
of wire as well. (Example: Pt100 3-wire system) 10. Operation Functional Class
For sensor input code-2, specify the type for two Select ‘BASIC’ or ‘LINK MASTER’.
inputs. In this case, PV (Sensor1 value) and SV 11. Sensor back up
(Sensor2 value) are alternately displayed on the Select ‘DISABLE’ or ‘ENABLE’.
integral indicator with sensor back up DISABLE.
2. Calibration range and unit
1) Calibration range can be specified within the < Factory Setting >
measurement range shown in Table 1. Also, set
the upper limit is larger than the lower limit. Table A. 4-20mA type
2) Specify one range from °C, K, °F or °R for
temperature input. It is not necessary to specify Input sensor type *1 “Pt100” or as specified in order
the unit of mV and ohm inputs, for these units Wire connection “3-wire system” or as specified in
automatically will be mV or Ω. (For RTD) *1 order
3. Tag Number (if required) Calibration range “0 to 100” or as specified in order
Specify Tag number (up to 16 Characters) to be Calibration unit “°C” or as specified in order
engraved on the tag plate. The specified letters (TC and RTD)
are written on Tag (16 Characters) in the amplifier Sensor burnout *2 High (110%, 21.6 mA DC)
memory. Output in Transmitter High (110%, 21.6 mA DC or more)
4. Software tag failure *2
Specified characters (up to 32 characters) are Sensor backup ‘DISABLE’ or as specified
set as “Tag” (the first 8 characters) and “Long
Tag No. “Blank” or as specified in order
tag (32 characters) in the amplifier memory.
Use alphanumeric capital letters. When the *1: For sensor input code-2, sensor 2 is set in
“SOFTWARE TAG” is not specified, “TAG NO” is set nonconnection.
as “Tag” (the first 8 characters) and “Long tag” (32 *2: Except when Optional code C1 or C2 is specified.
characters) in the amplifier memory.
5. Other factory configurations Table B. Fieldbus type
Specifying option code CA will allow further
configuration at factory. Input sensor type *1 “Pt100” or as specified in order
Descriptor (up to 16 characters) Wire connection “3-wire system” or as specified in
6. Sensor back up (For RTD) *1 order
Select ‘DISABLE’ or ‘ENABLE’. Calibration range *1 “0 to 100” or as specified in order
Calibration unit “°C” or as specified in order
Fieldbus type (TC and RTD)
1. Sensor type. Output Scale *1 “0 to 100%”
For RTD and resistance input, specify the number Output mode “Direct”
of wire as well. For sensor input code-2, specify the
Node Address ‘0xF3’ unless otherwise specified in
type for two inputs. (Example: Pt100 3-wire system) (in hexadecimal) order
In this case, AI1 (Sensor1) value and AI2 (Sensor2)
Operation functional ‘BASIC’, or as specified in order
value are alternately displayed on the integral
Class
indicator with sensor back up DISABLE.
2. Calibration range (XD_SCALE) Sensor backup *2 ‘DISABLE’ or as specified in order
For sensor input code-2, specify the range for two Tag Number “Blank” or as specified in order
inputs. (Tag plate)
3. Units of calibration range: Software Tag ‘TT1001’ unless otherwise specified
Specify only one unit from the table, ‘Settings when (PD_TAG) in order.
shipped.’ For sensor input code-2, specify the unit *1: For sensor input code-2, sensor 2 is set in
for two inputs. nonconnection.
4. Output mode (L_TYPE) *2: Channel settings of AI function blocks depend on
Select ‘Direct’ or ‘Indirect .’ sensor backup setting.
5. Output scale (OUT_SCALE) Disable: AI1= Sensor1,
AI2= If Sensor2 type is specified then
For sensor input code-2, specify the scale for two
Sensor2 else Terminal temperature
inputs. Enable: AI1= Backup temperature,
AI2= Terminal temperature
All Rights Reserved. Copyright © 2016, Yokogawa Electric Corporation GS 01C50H01-01EN Mar. 2, 2018-00
Subject to change without notice.