09-Wato Series Anesthesia Software Upgrade Instruction
09-Wato Series Anesthesia Software Upgrade Instruction
------------------------------------------------------------------------------------------------------------------------------------------------------------------
Contents
1 The Basics.....................................................................................................................................................................1
2 Install PC-based software: “SystemUpdateTool.exe”..............................................................................................1
3 Connect the PC and the anesthesia machine..........................................................................................................2
4 Upgrade the software of the anesthesia machine ...................................................................................................3
5 The version relationship between software: .............................................................................................................5
6 NOTICE: ........................................................................................................................................................................5
1 The Basics
1.1 Tools required: PC、a standard direct network cable or cross network cable.
1.2 The WATO SERIES anesthesia machines contain WATO EX-55\65, EX-20\25\30\35.
1.3 The anesthesia Software include:
01_SystemUpdateTool SAME
04_eflowmeter √ X X
Page 1 of 6
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 2 of 6
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
------------------------------------------------------------------------------------------------------------------------------------------------------------------
192.168.23.250
3.3 Set the IP address of the computer to “192.168.23.5” , subnet mask to “255.255.255.0”as below:
3.4 Switch on the machine, check the status of the local area connection at the lower right corner of
computer ( OR ). If the connection is unusual, please check the cable and ports.
4 Upgrade the software of the anesthesia machine
4.1 Double-click the “Mindray Anesthesia Machine Update Tool” icon on the computer
desktop.
4.2 Choose “Wato Series”, then click the “OK”.
Page 3 of 6
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
------------------------------------------------------------------------------------------------------------------------------------------------------------------
4.4 Click the button to select the update file in your hard disk, and click the
4.5 Click “Ok” to confirm the update file, and then click “Start (Single)” button.
Page 4 of 6
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
------------------------------------------------------------------------------------------------------------------------------------------------------------------
4.6 Switch on the anesthesia machine to start upgrade. Wait for a few minutes till you can see “Update
system successfully!” information on the PC screen.
4.7 Reboot the machine to check the software version of the machine.
4.8 Repeat the steps 4.3-4.6 to upgrade the other software.
5 The version relationship between software:
host system software version 03.00.01 version 04.10.03(same Ver, different file)
Note:
1: For the machines with bios 2.0,you cannot upgrade the bios to 3.1, you can just upgrade the system to
03.00.01. If you want to upgrade the software upper than it, you need to change the CPU board also.
2: For the machines with bios 3.0, you can first upgrade it to 3.1, then the other software to the latest.
3: For the machines with bios 3.1, you can upgrade all to the latest software as table shows.
6 NOTICE:
Before upgrade, please make sure the connection between PC and machine is well as the step 3.4.
Before upgrade, please record the old software version of machine in case of recovery.
Go to “Maintenance””factory Maintenance” ”system information””version information”
Page 5 of 6
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
------------------------------------------------------------------------------------------------------------------------------------------------------------------
Ensure that you click "Start (Single)" button first, then switch on the anesthesia machine to upgrade.
For the machine with BOOT software Ver3.1, you can force it to “upgrade mode” by continue pressing any
key button when switch on the machine.