Event Codes PF W10
Event Codes PF W10
Event
Group Description
code
E001-E099 0 Rundown failures
E100-E199 1 Event related errors
E200-E299 2 User input events
E300-E399 3 Statistical events
E400-E499 4 Communication events
E500-E599 5 Hardware events (tool)
E600-E699 6 Hardware events
E700-E799 7 Hardware events
E800-E899 8 Software events
E900-E999 9 Events MMI
code
Tool ready
PF ready
E003 Torque measurement The programming is incorrect as Check the programming for the
possibly invalid no movement is possible in the Pset.
second stage. Ensure that there is some movement
in the second stage.
Acknowledge
code
Tool ready
PF ready
E102 Rundown prohibited An attempt to make a tightening is Depending on C210 Lock on reject x
due to Lock on made and tightenings are disabled by configuration, either unlock the tool by
Reject the parameter C210 Lock on reject the direction ring or by digital input
because the result was not OK. Unlock tool or Master unlock tool.
Additional parameter information: If applicable, reconfigure C210 Lock
First integer parameter: 0=Locked on reject.
during tightening, 1=locked during
loosening
Second integer parameter 0=Service
denied, 1=service aborted.
E103 Tool locked by One or several of the digital input Unlock the tool with one of the x
digital input functions lock the tool: Tool disable, following digital input functions:
Tool tightening disable, Tool loosening Tool disable, Tool tightening disable,
disable. Tool loosening disable
Additional parameter information: Note: Unlock tool or Master unlock
First integer parameter: 0=Locked tool will not act as unlock trigger.
during tightening, 1=locked during
loosening
Second integer parameter 0=Service
denied, 1=service aborted.
E107 Rundown prohibited A Job using line control is selected with Unlock the tool with one of the x
due to Line control, parameter Use J330 line control. The following digital input functions: Line
batch not enabled Job does not start until a line control control start signal, Job abort or
start signal is received from digital Master unlock tool.
input function Line control start signal.
Acknowledge
code
Tool ready
PF ready
E117 Tool locked – Power Focus cannot communicate with Acknowledge the event. x x x
cannot access RBU the RBU and must be rebooted to Turn off PF and check that the RBU is
protect data. correctly connected.
Reboot PF.
Additional parameter information:
First integer parameter: 0=Locked
during tightening, 1=locked during
loosening
Second integer parameter 0=Service
denied, 1=service aborted.
E120 Motor tuning failure The connection between the tool and Acknowledge the event. Check that x x x
PF is incorrect or faulty, or the remote the remote start connector on PF is
start connector is missing or incorrectly present and correctly wired.
wired. Check that the tool cable and
connectors are undamaged and
correctly connected.
Service the tool.
E121 Open end tuning This event indicates that the Open end Acknowledge the event. x x x
failure tuning command has not been Check that the remote start connector
successfully performed. on PF is present and correctly wired.
Verify that parameters T421-T422
matches the geometry of the attached
OE head.
Check that the mechanical stop is
undamaged and working properly.
E122 Open end Open end positioning failed due to too Remove the tool from the joint.
positioning failure much torque during positioning phase. Restart the positioning.
E125 Alarm on timeout / This event is issued when parameter Redo the tightening.
Alarm on torque lost P246 Alarm on end time shutoff is If the event reappears, set P141 End
activated and the time defined in P141 time to a higher value or set P246
End time is exceeded before final Alarm on end time shutoff to No.
torque is reached.
The reason can be that the tool has
slipped off the joint, for example.
Acknowledge
code
Tool ready
PF ready
E126 Multistage aborted This event indicates that the current Check the additional information to
running Multistage has not been see which events occurred.
performed entirely (the drive has been Follow the procedure for those events.
shut off or the tool trigger was released
before the end).
If the second parameter in the sub
information is zero (no event is
indicated), activate P244 Alarm on lost
trigger and P246 Alarm on end time
shutoff in the stage in which the
multistage was aborted.
n Stage n nnn No of event that Multistage aborted 1 Rundown angle max exceeded
1
occurred
2 Rehit 2 Rundown angle min not reached
Torque measurement 4 Final torque max exceeded
4
possible invalid
8 Current limit reached 8 Final angle max exceeded
Timeout on end time 16 Self tap max exceeded
16
shutoff
32 Not used 32 Self tap min not reached
64 Other 64 PVT monitoring max exceeded
Remove fastener limit 128 PVT monitoring min not
128
reached
256 Drive disabled 256 PVT Compensate overflow
Transducer lost 512 Current monitoring max
512
exceeded
Transducer short circuit 1024 Post view torque min not
1024
reached
2048 Transducer corrupt 2048 Post view torque max exceeded
4096 Sync timeout 4096 Post view torque angle too small
8192 Dynamic CM Min Tq 8192 Trigger lost
16384 Dynamic CM Max Tq 16384 Torque less than target
32768 Angle Max mon 32768 Tool hot
65536 Yield nut off
131072 Yield too few samples
Acknowledge
code
Tool ready
PF ready
E127 PVT shut off This event indicates that the drive was In the Trace monitor, check the result
shut off in the selftap or prevail phase of and change the settings accordingly.
the rundown, or the compensate conditions
were not fulfilled.
E128 Trigger lost When the function trigger lost is activated For information only.
in the Pset, this event indicates that the
trigger of the tool was released before final
target.
E129 Torque lower than When the function torque lower than For information only.
target target is activated in the Pset, this event
indicates that the torque result is below Note: Check also E128 to see if E129
final target. The torque status is NOK even is issued because the trigger was
if the torque result was larger than final released before final target.
torque min.
E130 Post view torque Post view torque conditions were not In the Trace monitor, check the result x
shut off fulfilled. and change the settings accordingly.
E131 Tool Disconnected This event code is generated when the tool Acknowledge the event. x x x
is disconnected (logically or physically) If the tool has been logically
from PF. The event code is also generated disconnected, disconnect physically
when an attempt to start a disconnected also.
(logically or physically) tool is done.
Additional parameter information: Note: The tool can be logically
First integer parameter: 0=locked during disconnected if the user requests
tightening, 1=locked during loosening, disconnection or if the open end
2=locked during positioning, 3=locked settings are modified.
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
Acknowledge
code
Tool ready
PF ready
E133 Tool locked by The tool is prohibited via ToolsTalk PF. Unlock the tool by x
Tools Talk - sending an “Enable tool” message via
Additional parameter information: ToolsTalk PF protocol or
First integer parameter: 0=locked during - using digital input Master unlock
tightening, 1=locked during loosening, tool.
2=locked during positioning, 3=locked
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
E136 Tool locked by This event code is displayed if the tool is Unlock the tool by x
batch OK locked by the function P152 Lock at batch - using the digital inputs Unlock tool
OK. or Master unlock tool or
Additional parameter information: - sending an “Enable tool” message on
First integer parameter: 0=locked during fieldbus, Open protocol or ToolsTalk
tightening, 1=locked during loosening, PF protocol.
2=locked during positioning, 3=locked
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
E137 Tool locked by The tool is locked by fieldbus. Unlock the tool by x
fieldbus Additional parameter information: - sending an “Enable tool” message via
First integer parameter: 0=locked during fieldbus or
tightening, 1=locked during loosening, - using the digital input Master unlock
2=locked during positioning, 3=locked tool.
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
E138 Wrong tool start It is only possible to start the tool from Check the setting for T200 Tool start
input source selected tool start select source. select source.
E139 Tool locked by The tool is locked by open protocol. Unlock the tool by x
Open protocol Additional parameter information: - sending an “Enable tool” message via
First integer parameter: 0=locked during Open protocol or
tightening, 1=locked during loosening, - using the digital input Master unlock
2=locked during positioning, 3=locked tool.
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
Acknowledge
code
Tool ready
PF ready
E140 Insert user ID card The tool is locked; the user must insert his Insert a valid ID card. x
to release tool ID card in the card reader to release the
tool.
Additional parameter information:
First integer parameter: 0=locked during
tightening, 1=locked during loosening,
2=locked during positioning, 3=locked
during motor tuning, 4=locked during
open end tuning.
Second integer parameter 0=Service
denied, 1=service aborted.
E141 Tool locked during Multiple identifiers; tool is locked during Check the work order settings in the x
work order work order. Tool will be unlocked when Options dialog in the Identifier branch.
all expected identifiers have been received Scan the identifiers in the correct work
in correct order. order without pushing the trigger.
Unlock the tool by
Restoring of identifier configuration is
done
Reset all identifiers command is
performed via digital input, open
protocol or identifier monitor
Reset latest identifier command is
performed after the first type in the
work order is accepted.
E145 Tool locked by The tool is locked by digital input Timer Unlock the tool with digital input x
Timer enable tool enable tool when the Job was not started Timer enable tool.
within the time set in J320 Max time to If applicable, set J320 Max time to start
start Job. Job to a higher value.
E146 No tool cable Incorrect or damaged cable. Ensure that the tool cable is correctly x
detected connected and without damage.
Replace the cable.
E147 Tool locked by A Pset with click wrench tightening Use the correct combination of tool x
click wrench Pset strategy is selected for a tool. and Pset.
Acknowledge
code
Tool ready
PF ready
E149 Tool locked The function C220 Disable loosening at Unlock the tool by selecting forward x
disable loosening OK has locked the tool. on the tool, or by digital input Master
at OK unlock tool.
Additional parameter information: Continue tightening.
First integer parameter: 1=locked during
loosening
Second integer parameter 0=Service
denied, 1=service aborted.
E150 Job client does not When running a Cell Job, this event is Check which controller in the Cell that
respond displayed by the Job reference when one is not responding.
of the Job members does not respond. Check the cables and connections
between the Cell member and the Cell
Additional parameter information: reference.
The first integer parameter indicates the Check that the IP addresses in the Cell
channel ID for the Job client that does not group are correct.
respond. Check that IP addresses within the Cell
is not used by someone else.
E151 Job in OFF mode It is not possible to select a new Job. The In the Job monitor, clear the Job off
Power Focus is in Job off mode. check box.
Select new Job.
E152 PF locked in Job This event appears when, in a forced Cell Activate the controller. x
mode Job, an attempt is made to tighten with a If possible, disable J302 Lock at Job
controller which is not currently active or done.
when a controller has performed all
tightenings in the Job.
Acknowledge
code
Tool ready
PF ready
E156 Job members lost When running a Cell Job, this is displayed Check cables and connections to the
by the Job reference when it has lost lost Job client.
contact with one of the Job members. In the Controller menu of the lost Job
client, open Network and check that
Additional parameter information: the Job reference IP-address is correct,
The first integer parameter indicates the and that the Job member is part of the
channel ID for the lost Job client. same cell as the Job reference.
In the Job menu of the Job reference,
check that the all PFs are present in the
Job list.
E157 Job reference lost When running a Cell Job, this is displayed Check cables and connections to the
by the Job clients when they have lost Job reference.
contact with their Job reference. In the Controller menu, check the
network settings for the Job reference
and the Job clients. All Job clients and
the Job reference must belong to the
same cell.
E158 Invalid Job ID When the selected Job does not exist. In the Job reference, open the PF Map
and check the ID of all defined Jobs.
E159 No Pset in selected When the selected Job does not contain In the Job reference, open PF Map >
Job any Pset. Job, and double-click the selected Job.
Add the correct Psets and store the
settings.
E160 Job select source Attempt to select a Job with the wrong In the Job reference PF Map, open
not valid input source. Controller>Configuration.
In the Select source view, check the
settings for C201 Job.
E161 Line control alert The line control has been activated, and For information only.
1 the first control alert limit has been
reached.
E162 Line control alert The line control has been activated, and For information only.
2 the second control alert limit has been
reached.
Acknowledge
code
Tool ready
PF ready
E166 Job aborted This event will be displayed when a For information only.
running Job is aborted.
Additional parameter information.
First integer parameter:
2 Digital input
3 Ethernet/serial (PC
5 Field bus
6 Keyboard
7 Reserved
8 Reserved
9 Manual mode (the current active
job is aborted and a new job
becomes active when selected from
digital input)
10 Timer 1 (max time to start job)
11 Timer 2 (max time to complete job)
12 Deleted (a pending job is deleted
from PF database)
13 Job off mode
14 Line control
Alarm 2
15 Job override
E167 Max coherent Not When the number of NOT OK tightenings For information only. x
OK tightenings in a row is reached, then the tool will be
reached locked and can be unlocked only via a
digital input (reset NOK counter).
E175 Configuration will The Power Focus will erase the Accept or reject the memory re-
be erased at next configuration in the Power Focus memory configuration.
reboot and the RBU since the memory setup has
been changed.
E177 Functionality not The user is trying to use functionality that Replace the RBU or use another
in RBU is not enabled in the RBU. functionality.
E178 Memory allocation The user is trying to allocate more Check settings for Controller >
error memory than exists in the Power Focus. Memory.
Decrease the number of results, Psets
or Jobs until the memory usage is
below 100%.
Acknowledge
code
Tool ready
PF ready
E179 Totally Not possible to configure this functionality Save the PF configuration to file.
configurable without function “totally configurable Open Controller > Memory.
memory needed memory”. C600 Memory type must be set to
T T
Acknowledge
code
Tool ready
PF ready
E206 Pset number An attempt was made to perform a Check parameter C200 Pset and select
T T x
invalid tightening with the wrong Pset number. a Pset from the correct source.
Additional parameter information:
First integer parameter: 0=Locked during
tightening, 1=locked during loosening
Second integer parameter 0=Service
denied, 1=service aborted.
E207 Wrong Pset Attempt to select Pset from a source not Check parameter C200 Pset and select
T T
Acknowledge
code
Tool ready
PF ready
E209 Pset configuration The configuration of the Pset has been For information only.
updated changed.
E210 Identifier source An Identifier source is disabled. All Switch back to Automatic Mode to
disabled Identifiers sent by a disabled source are enable the Identifier source.
discarded.
E211 Wrong identifier Wrong input source for an identifier Open Identifier>Type setup and
input source string. check that the input source is correct.
Scan the identifier string.
E212 Pset/Multistage Event displayed every time a Pset or For information only.
stored Multistage is stored or restored.
Additional parameter information:
First integer parameter: Pset or Multistage
number
Second integer parameter 0=Pset,
1=Multistage
E213 Pset number not An attempt to start a Job containing a Pset Choose correct Job or reprogram the
present not present in PF has been performed. Job.
E220 Fastener broken Displayed if fastener (screw or nut) breaks Ensure that more than one comparison
during yield control. is obtained before yield is reached.
In ToolsTalk, open the trace monitor
and analyze the tightening.
Change the applicable parameter.
E221 Too few samples Too few samples for mean torque Increase parameter P262 Window angle
T T
for yield control calculation in yield algorithm. or reduce speed during yield control.
E233 Final Target This event code is displayed when an Change the value for “Final Target” to x
range error attempt is made to tighten with a final a value within tool limits.
target out of range (larger than tool max Store and try again.
torque or 9999).
Acknowledge
code
Tool ready
PF ready
E234 Start final angle This event code is displayed when an Change the value for P120 Start Final
T x
range error attempt is made to tighten with a P120
T Angle to a value within tool limits.
T
Source
Acknowledge
code
Tool ready
PF ready
E250 Max time for first This message is displayed and the Job is If needed, open PF Map > Job >
tightening run out terminated if the first tightening is not Timer and change the parameter J320T
(Job) performed within the specified time. Maxtime to start Job, or set to 0 to shut
T
off timer.
Store Job and try again.
E251 Max time to This message is displayed and the Job is If needed, open PF Map > Job >
complete Job run terminated if the Job is not completed Timer and change the parameter J321T
out within the specified time in Job Maxtime to complete Job, or set to 0 to
T
E335 Not enough data Not enough data is available to calculate the statistic Do more tightenings to get
control limits. enough data.
Calculation used: nbrofsubgroups
* subgroupsize * (1 + subgroup
frequency)
Check and if necessary change
P503 No of subgroups.
T T
E336 Mem alloc fail It is not possible to allocate enough memory for the Open Controller > Memory and
statistic subscription. set C600 Type to Totally
T T
configurable.
Check settings.
E340 Xucl tq The last subgroup mean torque value is larger than the The statistic events give an
upper control limit. indication of irregularities in the
tightening process.
If needed:
E342 Rucl tq The last subgroup range torque value is larger than the See E340 Xucl tq
T T
E343 Rlcl tq The last subgroup range torque value is lower than the See E340 Xucl tq
T T
E345 Cpk tq The torque Cpk is lower than 1.33. See E340 Xucl tq
T T
E346 7inc x tq Trend deviation alarm, the subgroup torque mean value See E340 Xucl tq
T T
E347 7dec x tq Trend deviation alarm, the subgroup torque mean value See E340 Xucl tq
T T
E348 7inc r tq Trend deviation alarm, the subgroup torque range value See E340 Xucl tq
T T
E350 7above x tq Trend deviation alarm, the subgroup torque mean value See E340 Xucl tq
T T
E355 2sigma r tq Trend deviation alarm, the last subgroup torque range See E340 Xucl tq
T T
E360 Xucl ang The last subgroup mean angle value is larger than the See E340 Xucl tq
T T
E361 Xlcl ang The last subgroup mean angle value is lower than the See E340 Xucl tq
T T
E362 Rucl ang The last subgroup range angle value is larger than the See E340 Xucl tq
T T
E363 Rlcl ang The last subgroup range angle value is lower than the See E340 Xucl tq
T T
E365 Cpk ang The angle Cpk is lower than 1.33. See E340 Xucl tq
T T
E366 7inc x ang Trend deviation alarm, the subgroup angle mean value See E340 Xucl tq
T T
E367 7dec x ang Trend deviation alarm, the subgroup angle mean value See E340 Xucl tq
T T
E368 7inc r ang Trend deviation alarm, the subgroup angle range value See E340 Xucl tq
T T
E369 7dec r ang Trend deviation alarm, the subgroup angle range value See E340 Xucl tq
T T
E375 2sigma r ang Trend deviation alarm, the last subgroup angle range See E340 Xucl tq
T T
E380 Xucl cm The last subgroup mean CM value is larger than the See E340 Xucl tq
T T
E381 Xlcl cm The last subgroup mean CM value is lower than the lower See E340 Xucl tq
T T
control limit.
E382 Rucl cm The last subgroup range CM value is larger than the See E340 Xucl tq
T T
E383 Rlcl cm The last subgroup range CM value is lower than the See E340 Xucl tq
T T
E385 Cpk cm The CM Cpk is lower than 1.33. See E340 Xucl tq
T T
E386 7inc x cm Trend deviation alarm, the subgroup CM mean value has See E340 Xucl tq
T T
E387 7dec x cm Trend deviation alarm, the subgroup CM mean value has See E340 Xucl tq
T T
E388 7inc r cm Trend deviation alarm, the subgroup CM range value has See E340 Xucl tq
T T
E389 7dec r cm Trend deviation alarm, the subgroup torque range value See E340 Xucl tq
T T
E395 2sigma r cm Trend deviation alarm, the last subgroup CM range See E340 Xucl tq
T T
Acknowledge
code
Tool ready
PF ready
E401 Duplicate device Two I/O devices with the same ID are Acknowledge the event. x x x
IDs on I/O bus present on the I/O bus. This event occurs Make sure that all of the devices on the
even if the devices are not configured in bus have a unique ID.
PF.
E403 I/O device not I/O device is not properly connected or the Check IO cables. x x x
responding ID of the device is not the same as the one Check the configuration of the
configured. accessories on the IO bus and compare
Additional parameter information: it with the devices on the IO bus.
First integer parameter: Number of device
not responding.
E407 IO device has An I/O device has selected ID 0, which is On the device, select an ID between 1 x x x
invalid ID invalid. and 15 (in ToolsTalk) or 1-F (on
selected Additional parameter information: selector)
First integer parameter: Number of device
not responding.
Acknowledge
code
Tool ready
PF ready
E417 Too many There are too many connections at the Wait until the other sessions are done
communication same time (ToolsTalk PF, Tools Net, and close their connection.
sessions at one Operator Panel etc.) The most common
time problem is that an external program
allocates all the communication.
E420 ST Accessory not A device connected to the ST-bus is not Acknowledge the event. x x x
supported. recognized by PF. Upgrade PF software.
E421 Safety trigger not The tool software version does not support Acknowledge the event. x x x
supported by tool safety trigger function. Upgrade the tool software.
software
E433 No RBU present No RBU detected at start-up or RBU Acknowledge the event. x x x
found missing at runtime. Check that there is an RBU and that it
is connected properly.
E434 RBU Timeout RBU did not respond to command from Acknowledge the event. x x x
Power Focus. Check that the RBU is connected
properly.
Reboot PF.
If the event reappears, replace the
RBU
E436 RBU file system RBU file system warning because of Reboot PF.
warning download interruption. If the event reappears, replace the
RBU
E437 RBU file system RBU file system fault. An attempt will be Acknowledge the event. x x x
corrupt made to repair it by means of erasing the Reboot PF.
RBU. Replace the RBU.
E438 RBU file A file with incorrect name or size was Acknowledge the event. x x x
mismatch returned to a read request. Reboot PF.
Replace the RBU.
E440 RBU no files RBU file system fault. Acknowledge the event. x x x
Reboot PF.
Replace the RBU.
E441 RBU access too RBU has answered to the poll but Power Acknowledge the event. x x x
busy Focus is too busy to receive the reply. Acknowledge the event.
Upgrade PF software.
E442 RBU no such file RBU file system fault. Acknowledge the event. x x x
Reboot PF.
Replace the RBU.
Acknowledge
code
Tool ready
PF ready
E444 RBU packet RBU responds with an answer not Acknowledge the event. x x x
rejected matching the last request. This is probably Reboot PF.
due to duplicates sent when the Power Replace the RBU.
Focus does not acknowledge packages.
E445 RBU corrupt RBU software fault. Acknowledge the event. x x x
Reboot PF.
Replace the RBU.
E446 RBU read error RBU read operation failed. Acknowledge the event. x x x
Try read again
Reboot PF
Replace the RBU
E447 RBU write error RBU write operation failed. Acknowledge the event. x x x
Try write again.
Reboot PF.
Replace the RBU.
E448 RBU delete error RBU delete operation failed. Acknowledge the event. x x x
Try delete again.
Reboot PF.
Replace the RBU.
E449 RBU flush error RBU erase operation failed. Acknowledge the event. x x x
Try erase again.
Reboot PF.
Replace the RBU.
E450 RBU list error RBU update operation failed. Acknowledge the event. x x x
Try update again.
Reboot PF.
Replace the RBU.
E451 RBU update Database synchronization between Power Acknowledge the event. x x x
failed Focus and RBU failed. Try update again.
Reboot PF.
Replace the RBU.
E460 Fieldbus type Error detected in initialization of fieldbus. Contact the Atlas Copco service
mailbox message representative.
fault
E461 Fieldbus type gen Error detected in initialization of fieldbus. Contact the Atlas Copco service
com fault representative.
E462 Fieldbus mailbox Error detected in initialization of fieldbus. Contact the Atlas Copco service
message fault representative.
Acknowledge
code
Tool ready
PF ready
E463 Fieldbus gen com Error detected in initialization of fieldbus. Contact the Atlas Copco service
fault representative.
E464 Fieldbus The fieldbus module is broken and has to Acknowledge the event. x x x
hardware fault be replaced. Contact the Atlas Copco service
representative.
E465 Fieldbus dip The software tries to configure the value of Switch off PF.
switch error node address or baud rate, but the address See chapter Fieldbus for information
switch on the fieldbus module is not in the about how to set the switch position.
right position to enable software setting. Correct the switch and turn on PF.
E466 Fieldbus offline The fieldbus went from online to offline. Check if PF is connected to the
network.
Check if the PLC is down.
E467 Fieldbus Error detected in initialization of fieldbus. Contact the Atlas Copco service
configuration representative.
fault
E468 Fieldbus The fieldbus module installed in Power Check the configured filedbus type.
hardware Focus is not the same fieldbus type as Check which fieldbus type is installed
mismatch configured with ToolsTalk PF. Change in PF.
fieldbus module or configuration to get a Reconfigure to match the installed
match. fieldbus type.
E469 Fieldbus init Error detected in initialization of fieldbus. Contact the Atlas Copco service
error representative.
E470 Fieldbus PCP InterBus PCP length error. Maximum Check the length of the process data
error length is 20 bytes. and the PCP data.
If the data exceeds 20 bytes, then
correct the configuration.
E471 Fieldbus claim Too much data traffic has caused one Reduce the data from the PLC that
area failed, lost packet lost. needs acknowledgement from PF.
one message
E472 Fieldbus release Release area command timed out. Information event.
area timeout If the event occurs frequently, then
reduce the data from the PLC that
needs acknowledgement from PF.
E473 Fieldbus Communication buffer full due to too Check if there is a frequent input
communication many input signals or incorrect fieldbus signal, or if F200 FB Update Interval
T T
Acknowledge
code
Tool ready
PF ready
E475 Fieldbus PsetID Fieldbus selected Pset does not match the In the PLC, check the Pset ID.
mismatch lifted socket. If the Pset and socket do not
Selector lifted correspond, then lift the correct socket
socket or change the Pset.
E480 Channel ID not The channel ID configured is not within Acknowledge the event. x x x
valid the limits permitted. The channel ID must Open PF Map > Controller >
be configured between 1 and 20. Network > Cell and change parameter
C310 Channel ID to be within 1 and 20
T T
in the cell use the same release as the cell Ensure that all cell members have the
reference. same SW version.
E490 Cell ID not valid The Cell ID configured is not within the Acknowledge the event. x x x
limits permitted. The Cell ID must be Open PF Map > Controller >
configured between 1 and 999. Network > Cell and set C313 Cell ID
T T
references in the net use the same release Ensure that all cell members have the
as the net reference. same SW version.
Acknowledge
code
Tool ready
PF ready
E501 Tool overheated The tool is too hot and the drive is Let the tool rest until it is cool enough x x
disabled. to run.
If the event reappears, reconsider the
Additional parameter information: workstation set-up and the tightening
First integer parameter indicates at which strategy.
stage the tool was locked: Service the tool.
0=Tightening, 1=Loosening,
2=Positioning, 3=Motor tuning, 4=open-
end tuning,
Second integer parameter: 0=Service
denied, 1=service aborted.
E502 Tool service Service the tool. See parameter T413 T Acknowledge the event.
interval expired Service interval.
T
Service the tool.
Additional information: Open Tool > Maintenance > Service,
The event appears, for 3 seconds, every click Service counter reset and confirm
100 tightenings after T412 No of
T by clicking OK.
tightenings since service exceeds T413
T
Note: To remove the event and unlock
Service interval, and T411 Service indicator
T T
the tool, set T411 Service indicator to
T T
alarm is checked.
T
E503 Calibration date The calibration date has expired. Use a torque measurement equipment
expired to calculate a calibration value.
Update tool calibration value using
T402 Set calibration value.
T T
alarm date.T
E504 Tool wear Tool wear indicator alarm indicates that the Service the tool.
indicator alarm tool should be serviced. Connect tool to ToolsTalk Service and
reset the wear indicator.
E505 Tool rebuilt Tool redesigned. E.g. angled tool type For information only. x x
to/from OE-type redesigned to an OE-tool type.
E510 Tool type not The tool type is not supported by the RBU. Acknowledge the event. x x x
supported by Check which RBU-level the tool
RBU requires.
Replace the RBU.
Acknowledge
code
Tool ready
PF ready
E511 Tool-Power Mismatch between the tool and PF 3000 Acknowledge the event. x x x
Focus 3000 size controller (for example S4/S7 tool Replace the PF3000 with the correct
mismatch connected to a S9 controller). PF3000 or a PF4000, to match the
tool.
E513 Tool EEPROM Checksum indicates that tool memory is Acknowledge the event. x x x
corrupt - service corrupt. Tool must be serviced. Reprogram the tool and restore the
tool memory, using Tools Talk service.
E514 Tool EEPROM Checksum indicates that the motor tuning Acknowledge the event. x x x
corrupt -Motor or the open-end tuning table is corrupt. Open Tool>Maintenance and do a
tune Additional parameter information: motor tuning or an open-end tuning.
First integer parameter indicates at which
stage the tool is locked: 0= tightening, 1=
loosening, 2=positioning, 3=motor tuning,
4=open-end tuning, 5=tracking
Second integer parameter 0=Service
denied, 1=service aborted.
E517 Tool max torque Tool max torque invalid, service tool. Acknowledge the event. x x x
invalid Set parameter T102 to an appropriate
value.
For S/DS/ETX tools: use TTPF or
Graph.
Other tools: use TTS.
E518 Tool Tool normalization value out of range, Acknowledge the event. x x x
normalization service tool. Use a torque measurement equipment
out of range to calculate a normalization value.
Use the strategy DS-control during
measurements.
Open Tool > Maintenance >
Calibration and update T403
T
Normalisation value.
T
E519 Tool calibration Tool calibration value out of range, service Acknowledge the event. x x x
value out of tool. Recalibrate the tool, using ACTA.
range If the event reappears, replace the tool.
E520 Wear indicator Data in tool memory used for wear Open Tool > Maintenance > Wear
data in tool indicator is invalid. Service tool. and set T434 Min wear tightenings to
T T
invalid <100000.
E535 Torque Cables to transducer cut off or shorted. Acknowledge the event. x x x
transducer error Check that the tool cable and
connectors are undamaged and
correctly connected.
If the event reappears, service the tool.
Acknowledge
code
Tool ready
PF ready
E536 No transducer Cables to transducer cut off or shorted. Check that the tool cable and x
(sensor) connectors are undamaged and
correctly connected.
E537 Calibration not Calibration offset outside limits. Service the tool.
OK, offset
outside limits
E538 Calibration not Calibration value outside limits. Service the tool.
OK, CalVal
outside limits
E539 Calibration not Calibration not OK, offset changed > 5% Try to calibrate again.
OK, Offset of max value since last calibration. If the event reappears, service the tool.
changed > 5% of
max value since
last calibration
E540 Calibration not Calibration not OK, calibration value Service the tool.
OK, CalVal changed > 5% of max value since last
changed > 5% of calibration.
max value since
last calibration
E541 Transducer Cables to transducer cut off or shorted. Acknowledge the event. x x x
(sensor) lost Service the tool.
during tightening
E550 Radio contact Radio contact with tool established, tool For information only.
with tool accessible.
established, tool
accessible
E551 Radio contact Radio contact with tool lost, tool Re-establish connection.
with tool lost, inaccessible.
tool inaccessible
E552 Communication An intermittent communication For information only.
error with tool disturbance between tool and controller. If occurring frequently, reposition
Serial Port Adapter.
If re-occurring, contact your Atlas
Copco representative.
E553 Pairing attempt An attempt to pair the Power Focus with a Set the tool in pairing mode according
with wireless wireless tool failed. to the pairing procedure.
tool failed Perform another pairing attempt.
If the event reappears, replace the tool.
E554 Command not A command (batch increment, reset batch Re-establish connection.
performed, etc.) could not be performed and was Perform the action causing the event
wireless tool discarded because the tool was currently again.
inaccessible inaccessible.
Acknowledge
code
Tool ready
PF ready
E555 Condition A condition change in the controller (Pset Re-establish connection.
change not selection, Job selection, Job aborted etc.) The condition change is now
reflected in tool, could not be reflected in the tool. The tool automatically transferred to the tool.
wireless tool was currently inaccessible.
inaccessible
E556 Tool battery low Tool battery voltage is low. Replace the tool battery.
Place the discharged battery in the
battery charger.
E557 Tool battery Tool battery empty. Tool locked. Replace the battery.
empty The tool is unlocked automatically.
E558 Tool software The implemented protocol version in Upgrade the software in tool or PF to
version Power Focus for connected tool differs match each other.
mismatch from the protocol version in tool.
Additional parameter information:
First parameter is Power Focus protocol
version. Second parameter is tool protocol
version.
E559 IRC pairing An IRC paring sequence has been For information only.
completed successfully completed.
E560 Tool RBU error An RBU error was detected by the tool Check that the RBU is correctly
software. mounted, otherwise service tool.
Tool ready
PF ready
E601 System voltage DC voltage too low. Tool locked. Acknowledge the event. x x x
low Power Focus: Verify that the input
voltage is within specified limits and
reboot the system.
STB: Replace battery. If occurring
frequently, check Pset, service tool.
E602 DC-bus HI DC voltage too high Acknowledge the event.
during rundown Verify that the input voltage is within
specified limits.
Acknowledge
code
Tool ready
PF ready
E603 Drive The DC drive is too hot, drive deactivated. Let the system cool for 15 minutes x
overheated Continue operation.
E604 24V output error 24 V output disabled (overload or short Acknowledge the event. x x x
circuit) Check that the cable and connectors
are undamaged and correctly
connected.
Reduce overload by disconnecting one
or more devices.
Connect an external isolated 24V
power source to the 24V DC input of
the devices.
E619 Drive DC drive deactivated during operation. Contact the Atlas Copco service
deactivated representative.
E620 Tool stall Tool motor stall. Check that the tool cable and
connectors are undamaged and
correctly connected.
If possible use a shorter tool cable.
For STB, see E601
E621 High current High current offset at tightening start. Contact the Atlas Copco service
offset representative.
E622 Angle transducer The controller has received invalid angle Acknowledge the event.
error information from the tool. Open Tool>Maintenance>Motor
tuning and perform a motor tuning.
code
Tool ready
PF ready
E700 PF started This event code is only visible in the event For information only.
log and indicates when PF is started.
E701 Backup battery The backup battery level is low. The Replace the backup battery.
low level battery should be changed.
E702 Backup battery The backup battery level is very low or the Replace the backup battery.
empty or battery is missing. A new battery should be
missing inserted as soon as possible.
Acknowledge
code
Tool ready
PF ready
E710 ACTA/QRTT An ACTA/QRTT calibration has been For information only.
calibration performed. See below for additional
information.
E720 Earth failure in An earth (ground) fail detected, probably a Acknowledge the event. x x x
ST tool cable failure. Service the tool.
E721 Internal ST tool Diode (LED) board in ST tool Acknowledge the event. x x x
failure disconnected. Tool service needed. Service the tool.
E722 Illegal ring The direction ring is not in a defined Place the direction ring correctly. x
position position.
Acknowledge
code
Tool ready
PF ready
E805 PF Model This event indicates that the RBU revision Acknowledge the event. x x x
unknown is not supported by the controller (for Check that the correct RBU is
example an RBU bronze in a PF Graph connected to PF.
controller), the RBU is missing, or the If the event reappears, then upgrade the
RBU license file is missing or corrupt. PF software.
E808 Error condition An error condition was detected in Power Focus: Close external programs
detected by software. connected to Power Focus.
software Power Focus: The probable cause is that an STB: Return tool to service.
external application has started too many
subscriptions.
STB: Software error
E822 No Job ACK Tightening initialization phase failed. Acknowledge the event. x x x
from control Hardwar e failure. Contact the Atlas Copco service
representative.
E823 No calibration No calibration was received from the drive Acknowledge the event. x x x
result during the tightening initialization phase Check tool cable.
(timeout 100 ms). Reboot PF.
E824 Cycle abort No tightening result was received from the If necessary, increase the P144 Cycle
T
timeout drive before the cycle abort timeout. abort timer in Pset.
T
E831 Autoset wrong Autoset could not be executed. The Acknowledge the event. x
angle calculated angle was too small. Program the Pset manually or by
Quick programming.
E833 Autoset tight The Autoset could not be executed, four Acknowledge the event. x
NOK result NOK tightenings were performed. Program the Pset manually or by
Quick programming.
E834 Autoset rejected Auto set is only permitted on a CW Pset. Program the Pset manually or by
Pset in CCW Quick programming.
direction If Quick programming is used, then
change the direction from CW to CCW
after programming.
E840 Feature Not This event is displayed when an attempt is Reconfigure to an allowed
Available In made to run Multistage with a RBU configuration or an available function.
Software bronze, or handle a Pset with a PsetID
Revision larger than the max number of Psets.
E841 SW function The software function is not available for Change to the correct tool type.
Not Available this tool type.
for this tool type
Acknowledge
code
Tool ready
PF ready
E851 Connection with The connection with the Tools Net server Check the network configuration and
Tools Net server is lost and the Power Focus is trying to in particular the ToolsNet IP address,
lost reconnect. The Tools Net server might be Subnet mask and Default router.
down or it might be an Ethernet cable Check if the ToolsNet server is down.
problem. Check that the Ethernet cables are OK.
Check that the routers and switches are
correctly connected and without
damage.
E856 Router The router in the configuration setting Check the network configuration and
unreachable could not be reached. Check the network in particular the Subnet mask and the
configuration and in particularly the subnet Default router.
mask and the default router. Ping the router to see if it is alive.
Check the Ethernet cable.
E857 Reboot needed The configuration changes will not take Acknowledge the event. x
before changes effect until after a system reboot. Reboot PF.
take effect If the event reappears, the PF is faulty.
Contact the Atlas Copco service
representative.
E858 IP address The IP address of this PF is already in use Acknowledge the event. x
already in use by another system. Change the IP-address in PF.
E859 IP address Another system attempts to use the same Check the network settings.
collision IP address as this PF.
E862 Ethernet The Ethernet driver of the PF is Check the network.
overload Error temporarily switched off due to an
overloaded network.
E863 IP Port already An attempt was made to bind a TCP or Acknowledge the event. x
in use UDP socket to a port already in use. For Check all network settings.
instance, a user protocol might be using
the same port as ToolsTalk PF.
E864 Ethernet echo An Ethernet packet was discarded, since its Check the network.
detected source MAC address was equal to the
Power Focus’ own MAC address. Notice
that this message only occurs once during
runtime, even if several such packets are
detected. There is most likely a problem in
the network.
Acknowledge
code
Tool ready
PF ready
E870 Sync member The Sync member registration failed In the sync reference, open PF Map > x
registration because of a mismatch in the sync Sync > Options > Sync group list
failure configuration list. dialog box, and check that all IDs are
correct and unique.
In the sync reference, open Controller
> Network settings, and check that the
IP address and ID are correct.
E871 Sync reference The channel ID of the Sync reference is In the sync reference, open PF Map > x x
configuration not first in the Sync list as required. Sync > Options, and check that the
failure sync reference is at the top of the Sync
group list.
E872 Sync Tightening synchronization initialization Check the I/O bus cable between all x
initialization failed. sync members.
failure Check that all Sync members have the
same active Pset/multistage.
Check the external start bridge.
E874 Sync members Only Sync reference is defined in the Sync Acknowledge the event. x
missing group list. In the sync reference, open PF Map >
Sync > Options and add sync
members to the Sync group list.
code
Tool ready
PF ready