A-Scan DGH6000
A-Scan DGH6000
The DGH 6000 Scanmate-A combines the accuracy ofDGH's proven A-Scan alignment algorithm with the process-
ing power, data storage and connectivity advantages of a personal computer. Patient Data can be stored on a local
computer, or in a centralized network location where it can be accessed by multiple users. Patient records are fully
searchable and can be exported in a format that is compatible with EMRlEHR systems. The IOL Calculator is easy
to use and includes modem formulas such as SRK®/T, Hoffer® Q , Holladay 1 and Haigis. Lens constants can be
optimized automatically based on Post-Operative results. The included Post Refractive Calculator assists the doctor
in estimating the true corneal power for patients that have undergone refractive surgery. The DGH 6000 provides
portability, accuracy and intuitive data management, all for a competitive price.
A-Scan Features
Measurement Range Measurement Accuracy
Average (mm)
• AXL 15.0 mm to 40.0 mm • Repeatability: ± O.03mmSTDEV (Immersion)
AXL 22.24 A~~ ~;;
• ACD 2.0 mm to 6.0 mm • Resolution: O.Olmm
Std. Deviation (mm)
• LT 2.0 mm to 7.51mn
ACD 0.00
AXL 0.01 LT 0.00 Transducer
Measurement Modes • Length 55.7 mm x Diameter 6.6 mm
Operalor~
• Immersion I Contact
Lens ~ct • Frequency 10.0 MHz (Nominal)
Vitreous @Normat
Eye Types • Focused Acoustic Beam (23.0 mm Nominal)
l'"E~:"II~)ll • Aphakic
Other
• Normal
• Cataract • Audible Feedback for Probe Alignment
• Dense Cataract • Corneal Compression Detection
• Pseudophakic (Silicone) • Fully Configurable Velocities
• Pseudophakic (PMMA) • Save Video or Individual A-Scan Waveforms
• Pseudophakic (Acrylic) • Automatic & Manual Measurement Modes
• Normal I Silicone Oil • Customizable - Each Doctor Can Configure
Vitreous Preferred A-Scan Settings
~DGHScanm.te
~
File Edit Prefer.nc"" Report, Help
25.00 +1.01
+{l.66
• Clinically Derived
Type Posterior Type Posterior
(Shammas)
Other
FonnUlalHoiladaYI ...1 Formula IHOfferQ .1 • Refraction Derived
IE] Personalized E1Pers()nalized
• Customizable - Each Doctor Can Configure
pACO 5.261
• Contact Lens Three (3) Preferred IOLs and Formulas, for
~TargeIP()Wer ~TargetPower Fast and Accurate Calculations
gmmatroplc 26.79 Emmetropic
• Integrates Seamlessly with DGH 8000 B-Scan
6000-LAB-BRCH1-R2