#1G5562-90000C Bravo UG S EN PDF
#1G5562-90000C Bravo UG S EN PDF
#1G5562-90000C Bravo UG S EN PDF
User Guide
Original Instructions
Notices
© Agilent Technologies, Inc. 2018 Warranty (June1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
No part of this manual may be reproduced The material contained in this docu- technical data.
in any form or by any means (including
ment is provided “as is,” and is sub-
electronic storage and retrieval or Safety Noticies
ject to being changed, without notice,
translation into a foreign language) without
prior agreement and written consent from in future editions. Further, to the max- A WARNING notice denotes a
Agilent Technologies, Inc. as governed by imum extent permitted by applicable hazard. It calls attention to an
United States and international copyright law, Agilent disclaims all warranties, operating procedure, practice, or the
laws. either express or implied, with regard like that, if not correctly performed or
to this manual and any information adhered to, could result in personal
User Guide Part Number contained herein, including but not injury or death. Do not proceed
G5562-90000 limited to the implied warranties of beyond a WARNING notice until the
merchantability and fitness for a par- indicated conditions are fully
Edition ticular purpose. Agilent shall not be understood and met.
liable for errors or for incidental or
Revision C, April 2018 A CAUTION notice denotes a hazard. It
consequential damages in connection
calls attention to an operating procedure,
Contact Information with the furnishing, use, or perfor-
practice, or the like that, if not correctly
mance of this document or of any
Technical Support and Customer Service: performed or adhered to, could result in
information contained herein. Should
Go to https://www.agilent.com and click damage to the product or loss of important
Agilent and the user have a separate data. Do not proceed beyond a CAUTION
CONTACT US for your country
representative. written agreement with warranty notice until the indicated conditions are
terms covering the material in this fully understood and met.
Documentation feedback: document that conflict with these
[email protected] terms, the warranty terms in the sep-
Web: arate agreement shall control.
http://www.agilent.com/en-us/products/
automation-solutions Technology Licenses
The hardware and/or software described in
Acknowledgements this document are furnished under a
Adobe® and Acrobat® are trademarks of license and may be used or copied only in
Adobe Systems Incorporated. accordance with the terms of such license.
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
About this guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Accessing product user information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
About the Bravo Platform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Hardware overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connection panel description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Liquid-handling heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Software overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
vi
Bravo Platform User Guide
Bravo Platform
User Guide
Preface
This guide describes how to use the Bravo Automated Liquid
Handling Platform, also known as the Bravo Platform.
This preface contains the following topics:
• “About this guide” on page viii
• “Accessing product user information” on page xi
vii
About this guide
Software version
This guide documents Bravo Diagnostics version 19.1, which is included in the
VWorks Automation Control software v13.1 installer.
The G5562A, G5563A Bravo Platform requires the following software versions
at a minimum:
• Microsoft Windows 10 64- bit operating system
• VWorks Automation Control software v13.1
The earlier model G5523A/G Bravo Platform is validated on the Microsoft
Windows 7 and Windows 10 64- bit operating systems.
Related guides
Use this guide in conjunction with the following:
• G5562A, G5563A Bravo Platform Safety and Installation Guide. Describes
potential safety hazards and how to avoid them, how to install the device,
and how to install the Light Curtain and shields.
For earlier model devices, see the Bravo Automated Liquid Handling
Platform Safety and Installation Guide.
• Automation Solutions Products General Safety Guide. Provides general
safety information and describes potential safety hazards that you might
encounter when using Automation Solutions products. A copy of this safety
guide is included with your shipment.
• VWorks Automation Control Setup Guide. Explains how to define labware
and labware classes, liquid classes, and pipetting techniques, and how to
track and manage labware in storage.
• VWorks Automation Control User Guide. Explains how to create protocols,
and set task parameters for each device in the system.
If you are using the Bravo 96AM Head, see “AssayMAP Bravo Platform” on
page 199.
If the Bravo Platform is a device in a third- party system, see the relevant
third- party system guides.
Related information
1 In the main window of the VWorks software, click the help button .
The pointer changes to . Notice that the different icons or areas are
highlighted as you move the pointer over them.
2 Click an icon or area of interest. The relevant topic or document opens.
Item Feature
1 Navigation area. Consists of four tabs:
• Contents. Lists all the books and the table of contents of the
books.
• Index. Displays the index entries of all of the books.
• Search. Allows you to search the Knowledge Base (all products)
using keywords. You can narrow the search by product.
• Favorites. Contains bookmarks you have created.
Related information
1
Introduction
This chapter gives you an overview of the Bravo Platform and
contains the following topics:
• “Safety notes” on page 2
• “About the Bravo Platform” on page 3
• “Hardware overview” on page 5
• “Connection panel description” on page 9
• “Liquid- handling heads” on page 11
• “Software overview” on page 14
1
1 Introduction
Safety notes
Safety notes
WARNING Using controls, making adjustments, or performing procedures other than
those specified in the user documentation can expose you to moving-parts hazards and
hazardous voltage. Before using the Bravo Platform, make sure you are aware of the
potential hazards and understand how to avoid being exposed to them.
Ensure you have read the G5562A, G5563A Bravo Platform Safety and
Installation Guide and are trained in the safe operation of the device.
Related information
Product description
The Bravo Platform is a versatile liquid handler with a nine plate- location deck
that is suitable for handling 96- well, 384- well, and 1536- well microplates
(plates). The platform uses interchangeable liquid- handling heads.
Figure Bravo Platform front view with optional gripper holding a microplate
Related information
Hardware overview
1 1
2
3
7
6
8
3 Head mount The fixture that provides the mechanical connections and communication
inferface for the liquid- handing head. The head mount attaches to an arm
that extends from the Bravo rear wall of the tie bar. The head mount travels
along the arm between the front and back of the deck (y- axis). The arm
moves the head mount from side to side across the deck (x- axis). For details,
see Axes of motion.
You can move the head mount manually while the Bravo Platform is turned
off.
4 Tie bar The vertical bar at the front of the device that adds structural support to the
Bravo head mount. The tie bar moves at high speed from side to side (x- axis)
across the front of the Bravo deck whenever the head moves to a deck
location.
6 Emergency- The pendant that contains the red emergency- stop button. To stop in an
stop emergency, press the red button. All Bravo motion stops.
pendant
7 Gripper An optional gripper that extends from the head mount to below the pipette
head tips. The gripper picks and places labware on the deck.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
Axes of motion
The Bravo Platform has components that move in the x- , y- , and z- axes, as the
following figure shows.
00232
Bravo
Pinch Hazard
Y
X
00232
If the Bravo Platform is fitted with a gripper, the gripper moves with the
Bravo head. In addition, the gripper has the following axes of motion:
Bravo Platform User Guide 7
1 Introduction
Hardware overview
• G- axis. The opening and closing distance of the gripper fingers, which
enable the gripper to grip and release labware.
• Zg- axis. The vertical distance the gripper arm moves, which enables the
gripper to extend beyond the pipette head.
G Zg
The motion of the displacement of the pipettor inside a liquid- handling head is
referred to as the w- axis (not shown).
Related information
1 2 3 4
00220
Bravo
Back
220c
3 Fuse holder Houses the AC inlet fuse and a spare fuse for the
Bravo Platform.
00212c
Bravo
Acc_IO
Related information
Liquid-handling heads
Disposable-tip heads
Disposable- tip heads allow you to change pipette tips during a run to prevent
cross- contamination. The following table lists the disposable- tip heads available
for the Bravo Platform.
The 96- and 384- barrel heads can dispense fluid into all the wells in a
microplate simultaneously or into a single column, single row, or single well in
the microplate.
The 96LT head use large transfer (LT) tips to dispense up to 250 µL per well.
The 96ST and 384ST heads use small transfer (ST) tips to dispense up to
70 µL per well.
Pipette barrel
Related information
Removing one liquid- handling head “Changing the Bravo head” on page 65
and mounting a different head
Software overview
VWorks software
You use the VWorks software to:
• Set up the system. You create device files for your specific set of devices.
You can also create user accounts with different levels of privileges to
enforce access policies.
• Define labware. Labware definitions describe the type and the geometry of
the labware that the devices will be handling during the protocol runs.
• Specify pipette speed and accuracy.
• Create protocols. Protocols determine the sequence of tasks that you want
to automate in a run.
• Run, pause, monitor, and stop protocols. You can start, pause, monitor, and
stop a protocol run from the host computer.
For detailed descriptions, see the VWorks Automation Control Setup Guide
and the VWorks Automation Control User Guide.
Bravo Diagnostics
Bravo Diagnostics is software that you open from within the VWorks software.
You use Bravo Diagnostics to do the following:
• Create Bravo profiles. Each profile specifies the following:
– Ethernet connection for the LAN communication
– Type of liquid- handling head
– Teachpoint settings. The teachpoints ensure the positional accuracy of
the pipette head and gripper, if included, at each deck location
– Deck configuration
– Whether the Bravo robot and an external robot can operate on the
Bravo deck concurrently
• Establish communication with the Bravo device using a Bravo profile.
• Verify that the device functions properly and troubleshoot problems. The
controls enable you to:
– Run tasks individually. For example, you can perform a Tips On or
Tips Off task independently of a protocol.
– Home the head and send the head to home position.
– Control the motions of an individual axis or disable the motors of
selected axes for manual moves.
Related information
2
Setting up the Bravo Platform
This chapter contains the following topics:
• “Workflow for setting up the Bravo Platform” on page 18
• “Starting up and shutting down” on page 19
• “Creating or adding a Bravo device” on page 21
• “Opening Bravo Diagnostics” on page 26
• “Creating and managing profiles” on page 28
• “Initializing the Bravo Platform” on page 33
• “Setting teachpoints” on page 36
• “Verifying the gripper setup” on page 43
• “Configuring external robot access” on page 48
17
2 Setting up the Bravo Platform
Workflow for setting up the Bravo Platform
Workflow
The general workflow for setting up the Bravo Platform is as follows:
9 Save any changes that you made “Creating and managing profiles”
in steps 3–8 in the Bravo on page 28
profile.
CAUTION Always turn off the Bravo Platform before installing or uninstalling a liquid-
handling head. Failure to do so can damage the head electronics.
CAUTION Always turn off the Bravo Platform before installing or uninstalling a liquid-
handling head. Failure to do so can damage the head electronics.
Related information
6 On the Device File tab, expand the Bravo device icon to show the list of deck
locations, and then click the location 1 icon. The corresponding location
properties appear.
Set the desired values for the following properties. Use the default values
for the remaining properties.
Property Description
IMPORTANT If an external robot will access the Bravo deck, ensure that you
carefully plan which locations can be accessed and where the Bravo head can
be during the external robot’s access. See “Configuring external robot access”
on page 48.
Related topics
Procedure
To open Diagnostics:
1 In the VWorks software window, ensure the correct device file is open.
To open a device file, choose File > Open, and then select the appropriate
device file (*.dev) in the Open dialog box.
2 In the Devices area of the opened device file tab, highlight the device icon,
and then click Device diagnostics.
Alternatively, you can double- click the device icon.
Related information
Profiles defined
IMPORTANT Each Bravo Platform that you install requires a unique profile,
which enables communication settings specific to the given instrument.
CAUTION Using a profile that is associated with teachpoints from a different Bravo
Platform can result in a hardware collision and cause equipment damage. Ensure that you
verify the profile and associated teachpoints for each specific Bravo Platform.
CAUTION Each profile can be used by multiple protocols. Deleting, renaming, or changing
the parameters for a profile based on one protocol can invalidate other protocols that use the
profile.
For information about device files, see “Creating or adding a Bravo device” on
page 21. For a detailed description of the relationships between the device file,
profile, and teachpoints, see the VWorks Automation Control User Guide.
Creating a profile
A Bravo profile consists of a combination of parameters that you set in the
Profiles tab, Configuration tab, Jog/Teach tab, Gripper tab, and the External
Robots tab of Bravo Diagnostics.
3 In the Create Profile dialog box, type a name for the new profile, and click
OK.
Use a profile name that identifies the specific configuration. For example,
include the liquid- handling head type and serial number in the profile
name. If this computer controls multiple Bravo Platforms, also identify the
device, for example, the device serial number.
4 Under Connection, select This Bravo is connected via ethernet, and then click Find
available devices.
Note: If you have an earlier model, you may select This Bravo is connected
via serial, and then select which COM port on the computer you are using.
5 In the Discovered BioNet Devices dialog box that appears:
a Click the Select the Ethernet adapter to use from the list below, and select the
correct adapter for the device connection.
b In the list of devices that appear, select the Bravo device. If multiple
Bravo devices are on the network, use the MAC Address to identify the
Bravo device for this profile. To successfully communicate with the
Bravo device, the Status column must display New or Matched.
c Click OK.
a b c
CAUTION Selecting the wrong Teaching tip type for a disposable-tip head, pin
tool, or Bravo 96AM Head can result in a head crash.
IMPORTANT For the Bravo 96AM Head, the Teaching tip type can be
a cartridge or an 250- µL tip.
c Ensure that the Check head type on initialize check box is selected.
CAUTION If this check box is cleared, the Bravo Platform will run with any head
installed. Therefore, if the profile specifies a 96-channel head type, but a 384-channel
head is installed, the head will crash.
Note: If the Check head type on initialize check box is cleared, the
w- axis controls in the Jog/Teach tab will be disabled even when the
device is initialized.
WARNING When you initialize the Bravo Platform, the head can move. Keep clear
of the Bravo head while it is in motion. Do not touch any of the moving parts or attempt
to move labware while the Bravo Platform is in operation. The device could pinch,
pierce, or bruise you.
Related information
WARNING When you initialize the Bravo Platform, the head and tie bar can move. To
prevent potential injury, keep clear of the device while it is in motion.
CAUTION To prevent potential equipment damage, ensure that the deck is clear of any
obstacles before using the Bravo Platform.
CAUTION Using an incorrect profile can damage the Bravo Platform. Ensure that the
profile is correct for the head type and deck configuration.
Ensure that you have added the Bravo device to a VWorks device file and that
you have a valid Bravo profile.
5 Verify that the selected Head type matches the installed head.
For example, if the 96LT head is installed, the Bravo Platform requires a
profile for the 96LT head type.
6 Click the Configuration tab, and verify that the Location Configuration and
Accessory Configuration matches the physical layout on the Bravo deck.
7 In the Profiles tab, click Initialize this profile to start initializing the selected
profile.
8 If the A microplate-in-gripper error message appears, and the gripper is not
holding labware, click Ignore and continue to continue the homing process.
During the homing process, the gripper opens and closes.
If the gripper is holding labware, the gripper will drop the labware during
this process. If necessary, click Abort to cancel the initialization. See the
instructions for manually opening the gripper to remove the labware before
you attempt to initialize the device.
9 If a Verify that it is safe to home the W-axis (the aspirate/dispense axis) message
appears, do one of the following:
• If the tips do not contain fluid. Click Retry to continue homing the
w- axis.
• If the tips contain fluid. Click Ignore. When the W-axis is not homed
message appears, use the tools in the Jog/Teach tab in the Bravo
Diagnostics dialog box to home the w- axis over a waste location.
Homing the w- axis will empty the tips.
Note: The Miscellaneous profile settings include the Prompt user to home
W-axis on first initialization option and the Ignore plate sensor during pick
and place option. For an explanation of the options, see “Miscellaneous area”
on page 292.
Related information
Setting teachpoints
CAUTION Using a profile that is associated with teachpoints from a different Bravo
Platform can result in a hardware collision and cause equipment damage. Ensure that you
verify the profile and associated teachpoints for each specific Bravo Platform.
Teachpoint defined
A teachpoint is a set of axial coordinates that define a location to which the
liquid- handling head moves. Each of the nine locations on the Bravo deck has
a default teachpoint that can be edited in Bravo Diagnostics. The default
teachpoints are set so that well A1 is at the back left corner of each deck
location.
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should perform the procedures in this topic.
WARNING The red Stop motors button in Bravo Diagnostics does not perform an
immediate stop. The Bravo head can continue to move in the same direction at the same
speed after you click the button. To perform an emergency stop, press the red button on the
pendant.
If you are teaching a pin tool, see “Workflow for non- tip- box locations” on
page 37.
If you are teaching the AssayMAP Bravo 96AM Head, see “Teaching part 1:
Setting teachpoints in the AssayMAP 1 profile” on page 244.
CAUTION To prevent the liquid-handling head from crashing into an accessory, be sure to
start the accessory teaching process from the safe z-axis point above the accessory location.
This is especially important if you use the Set all teachpoints based upon selected
teachpoint feature.
2 Set teachpoints for the remaining locations by doing “Setting other teachpoints based on
one of the following: one teachpoint” on page 40
• Automatically set the remaining teachpoints
based on the first teachpoint.
• For more precise teachpoints (384- or 1536- well “Setting the first teachpoint” on
plates), repeat step 1 for each deck location. page 38
– Nested tip box. Install and configure the Nested “Installing and setting up a
Rack Insert. Nested Rack Insert” on
page 151
2 Set the teachpoint for the location. “Setting the first teachpoint”
on page 38
3 ST tip box locations. Verify the labware definition for the “Verifying teachpoints for
tip box. disposable- tip heads” on
page 41
4 Verify the teachpoint. If tip- box operations are not precise “Verifying teachpoints for
enough, repeat this step. disposable- tip heads” on
page 41
CAUTION Make sure the disposable tip is an Agilent tip and is the same type that you
specified as the Teaching tip type in the profile.
6 Use the Jog Axes controls to move the liquid- handling head to the correct
teachpoint as follows:
Aspirate
+W
Dispense
-W
CAUTION Jogging the head down too far will damage the tip. Be careful not to jog
the head down too far.
CAUTION When jogging down in the z-axis, a warning might appear stating that you
are about to exceed the set teachpoint, which could cause a crash. Before you click OK,
verify that enough space remains to continue jogging down.
a Jog the head down (z- axis) until the A1 tip is approximately 2 mm
above the platepad.
Reduce the increment value when the tip gets close to the deck.
b Jog the head in 0.05 mm increments in the x and y directions, until
the tip is positioned directly above the crosshair mark on the platepad.
c To achieve paper- thin z- axis clearance, slide a sheet of paper between
the tip and the platepad. Set the z- axis increment to 0.05 mm
increments. Jog the head down until the paper is barely pinched, and
then jog up by 0.05 mm.
7 When the tip is in the correct position (directly over the crosshair mark
with paper- thin clearance), click Teach.
Verify the information in the message box before you click OK.
8 To save the teachpoint, click the Profiles tab and click Update this profile.
The teachpoint for the other locations will be set based on this one.
2 In the Advanced operations list, select Set all teachpoints based upon selected
teachpoint.
3 Click Perform advanced operation.
The VWorks software calculates the teachpoints for the other eight
locations, keeping the z- axis coordinate the same and changing the
x and y coordinates according to their spacing.
4 Readjust the z- axis height for any positions that are taller than a standard
platepad, such as the Orbital Shaking Station.
5 To save the teachpoints, click the Profiles tab and click Update this profile.
6 In the Jog/Teach tab, click Move. The head moves to the teachpoint you set
in “Setting the first teachpoint” on page 38. You might notice that the A1
tip is not quite in the correct position.
7 Use the Jog Axes controls to move the head A1 tip to the correct position
(directly over the crosshairs with paper- thin clearance).
8 When you are finished, click Teach. Verify the information in the message
box before you click OK.
9 Repeat step 6 to step 8 for each deck location.
10 To save the changes, click the Profiles tab and click Update this profile.
To verify a teachpoint:
1 In the Jog/Teach tab, set the Location field to the location you want to
check.
2 Click Approach.
3 Use the following table to decide your next step:
Related information
Changing the liquid- handling head “Changing the Bravo head” on page 65
IMPORTANT Before you adjust the gripper y- axis offset, ensure that the
pipettor teachpoints (x- , y- , and z- axes) have been verified for the given
profile and no pipette tips or cartridges are installed on the head.
IMPORTANT Ensure that you use a standard platepad for the following
procedure.
4 Click Move to move the gripper to the previously saved teachpoint for the
selected location.
The Bravo head moves above the deck location, and the gripper assembly
extends down (Zg- axis) so that the gripper fingers are just above the top
surface of the platepad.
-Zg
+Zg
-Y
+Y
5 Visually inspect the spacing between the black rubber gripper pads and the
sides of the platepad (y- axis). Make sure that all the gripper pads have
clearance so that the gripper can move down without colliding with the
platepad.
CAUTION Make sure you use small enough jog increments to prevent any gripper
collisions, which could damage the gripper.
a Click the Jog/Teach tab, and ensure that the deck location you selected
in step 3 is still selected.
b Under Jog/Axes, set the Back -Y/Forward +Y increment to a small enough
size, such as 0.2 mm.
7 After ensuring that the gripper pads will not collide with the platepad, set
the Down +Z increment to 5.0 mm, and then click Down +Z.
CAUTION Make sure you do not jog the gripper down more than 5.0 mm at this stage.
Jogging down farther will result in a gripper collision with the setscrew on the platepad,
which could damage the gripper.
At this stage, the gripper fingers should be next to the sides of the
platepad.
8 Fine tune the gripper alignment as follows:
a In the Gripper tab, set the Close +G increment to a small value, such as
0.2 mm, and then click Close +G so that you can clearly see the
alignment of the gripper pads with the platepad.
b In the Jog/Teach tab, set the y- axis increment to 0.2 mm, and then use
the Back -Y and Forward +Y buttons to position the gripper so that the
fingers are at equal distances around the platepad.
Ensure the pair of gripper pads on each side are as close as possible
to the sides of the platepad without actually touching the platepad.
Note: If a gripper finger is not perfectly straight, one of its gripper
pads may be closer than the other to the side of the platepad, as the
following figure shows. In this case, adjust the y- axis using the
gripper pad closest to the platepad and the diagonally opposite gripper
pad for reference.
7 Repeat steps 4 and 6 for the remaining deck locations where the gripper
will be transferring labware.
Related information
Workflow
Perform the following steps to configure an external robot’s access on the
Bravo deck. This workflow assumes that you have already verified the Bravo
Platform teachpoints.
CAUTION Ensure that you plan the external robot access locations carefully to prevent
potential hardware collisions. A collision between an external robot and the Bravo Platform
can damage the Bravo Platform, the external robot, or both robots.
Careful planning of the deck locations that the Bravo robot and any external
robots can access will optimize your throughput and prevent hardware
collisions.
Physical factors to consider
Make sure you consider the following factors when you plan the deck locations
that an external robot may access or when selecting a safe location for the
Bravo head:
Figure Bravo head mount and tie bar motion (front view) and (side
view)
Y
X
External robot’s Some robots can reach any Bravo deck location, while
physical reach other robots can access a subset of deck locations.
For example, the BenchCel robot approaches the Bravo
Platform from the left or right side only and can access
only the left- most and right- most deck locations.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
Front
• Portrait
External robot’s The approach path of the external robot can affect
approach path which Bravo deck locations must be blocked for the
Bravo head during the external robot’s access.
For example, the following figure shows two variations
of an external robot’s approach path to deck location 7.
The blue locations are where the Bravo head could
potentially continue operations concurrently while the
external robot is accessing the deck:
• Example A. The external robot is accessing
location 7. However, locations 1 and 4 are also
blocked to prevent the tie bar from being in the
path of the external robot when it accesses
location 7. The Bravo head can access the remaining
deck locations (blue) without risking a collision with
the external robot during its approach to location 7.
• Example B. In addition to locations 1, 4, and 7, the
Bravo head is blocked from accessing location 8. The
Bravo head cannot access location 8 without risking
a potential collision with the external robot during
its approach to location 7.
Figure Deck locations for Bravo robot (blue) and BenchCel robot (gray) operations
2 3
A
00240
Bravo
Plate Orientation
4 5
7 8 9
00240
Example scenario 2: The Bravo robot, BenchBot Robot, and Direct Drive Robot
operating concurrently
In this example, three robots can move concurrently at the Bravo deck:
• BenchBot Robot with a landscape grip. Approaches location 1 from the
rear opening in the Bravo backplate. To be conservative, location 2 is also
reserved for the BenchBot Robot and blocked for access by the Bravo
robot.
However, if you are certain that the external robot’s approach path would
not cross within the area that the Bravo head requires to access location 2,
location 2 could be unblocked to optimize usage.
• Direct Drive Robot with a portrait grip. Approaches location 3 at the
right side of the Bravo deck. To be conservative, location 6 is also reserved
for this robot and blocked for access by the Bravo robot.
• Bravo robot. Continues operations at deck locations 4, 5, 7, 8, and 9.
The following figure shows the deck location designations for each of the three
robots.
Figure Deck locations for the Bravo robot (blue), BenchBot robot (gray), and Direct Drive Robot
(gray) operations
A
1
1 00240
2 3 A
Bravo
1
Plate Orientation
4 5 6
7 8 9
00240
• Allows one robot to access the Bravo deck at any given time. For example,
if the Bravo robot is performing a liquid- handling task, the external robot
will wait for the Bravo robot to finish before approaching the deck. If the
external robot is placing or removing labware from a deck location, the
Bravo robot will move to its safe location wait for the external robot to
finish.
• Enables the Bravo robot to move to a specified deck location (safe location)
whenever the external robot approaches the deck to perform a task. For
example, if you specified deck location 1 as the safe location, the Bravo
robot will always move to deck location 1 whenever the external robot
approaches.
CAUTION You can specify any deck location as the Bravo safe location. However, to
prevent potential hardware collisions, you should determine the relative position of the
external robot and its approach before specifying the Bravo safe location.
For guidelines on how to select the safe location, see “Planning the robot
access locations” on page 49.
3 In the Move Bravo to this safe location box, select the deck location number
where the Bravo head should move for the duration while the external
robot is accessing the Bravo deck.
Note: Location 5 is the default safe location.
CAUTION To prevent potential hardware collisions, ensure that you plan the external robot
access locations carefully. For guidelines, see “Planning the robot access locations” on
page 49.
Related information
3
Preparing for a protocol run
Before you start a protocol run, you should check the Bravo
Platform teachpoints and the protocol to ensure optimum
operation.
This chapter contains the following topics:
• “Workflow for preparing a protocol run” on page 58
• “Emergency stops” on page 58
• “Planning for the protocol run” on page 59
• “About performing dry runs” on page 60
57
3 Preparing for a protocol run
Workflow for preparing a protocol run
2 Plan for the protocol run. “Planning for the protocol run” on
page 59
Emergency stops
Procedures
To stop the head motion in an emergency:
1 Press the red button on the emergency- stop pendant. The Bravo motion
stops.
2 In the VWorks message box, click one of the action buttons to re- enable
the motors in the head.
Related information
• The accessories and labware required for the protocol are where they
should be positioned.
• The reservoirs that must be filled.
• The waste bins or bottles that should be emptied.
Related information
Related information
4
Maintenance and troubleshooting
This chapter tells you how to keep your Bravo Platform in good
working order through cleaning, inspection, and maintenance. It
also explains what to do when you encounter a problem.
This chapter contains the following topics:
• “Routine maintenance and cleaning” on page 62
• “Cleaning up after a run” on page 64
• “Changing the Bravo head” on page 65
• “Testing the safety equipment” on page 74
• “Replacing the fuse” on page 75
• “Retracting tip box stripper pins” on page 78
• “Manually moving the head or gripper” on page 80
• “Recovering from a head collision” on page 82
• “Troubleshooting hardware problems” on page 84
• “Troubleshooting hardware- related error messages” on page 88
• “Reporting problems” on page 90
61
4 Maintenance and troubleshooting
Routine maintenance and cleaning
WARNING To prevent potential injury, always shut down the Bravo Platform before
performing any maintenance procedure.
CAUTION Use only the recommended cleaning materials. Using other cleaning solutions
and materials can cause damage to the device. Do not use abrasive, corrosive cleaning
agents. Do not use metal brushes.
CAUTION Alcohol and alcohol-based solutions can damage the gripper pads. Use a soft
cloth dampened with water to clean the gripper pads.
Use a clean soft cloth to clean up spills on any part of the device immediately
after a protocol run.
Clean the Bravo Platform weekly or as needed. Use standard laboratory wipes
and a mild detergent or isopropyl alcohol to clean the exterior painted white
surfaces and the metal surfaces of dust, grime, chemical deposits, and other
debris.
Also see “Cleaning up after a run” on page 64.
Related information
Related information
WARNING AssayMAP Bravo Platform only. To prevent potential injury from exposure to
chemical spills, ensure that you empty the syringes before uninstalling the Bravo 96AM
Head. For instructions, see “Uninstalling the Bravo 96AM Head” on page 212.
WARNING Put on gloves to protect your hands from potential exposure to any residual
contaminants.
CAUTION Always turn off the Bravo Platform before installing or uninstalling a liquid-
handling head. Failure to do so can damage the head electronics.
IMPORTANT If a head is being used for the first time, make sure you have a
profile specifically for that head. If necessary, create a new profile before
starting the following procedure. For details, see “Creating and managing
profiles” on page 28.
If disposable tips are loaded on the currently installed head, use Diagnostics to
dispense any liquid and remove the tips before removing the head. For
instructions, see “Performing a task using Diagnostics” on page 104.
Make sure that the head mount is in its home position above deck location 5.
If necessary, click the Jog/Teach tab in Diagnostics, and then click Home XYZ
to reposition the head mount.
IMPORTANT If you are changing the currently installed head for the same
type of head and want to edit the profile, you may use the Change head
wizard. Otherwise, use the following procedure.
Figure Bravo Platform status lights (front) and power switch (right side)
2 Make sure that the head mount is at its home position above deck
location 5, as shown in the preceding figure.
If necessary, manually move the head to the center of the deck.
3 To unlock the mounted head, refer to the following figure:
a Pull out and twist the (1) two head- retainer pins one- quarter turn
(90°) so that they remain retracted.
Note: The straight edge of the retainer pins should be horizontal.
b Turn the (2) head lock counterclockwise until it clicks into position.
Figure Installed Bravo 96AM Head: (1) retainer pin and (2) head lock
4 Reaching from the left side of the Bravo tie bar, grasp the head firmly as
the following figure shows:
a With your left hand firmly grip the left side of the head.
b Place your right hand under the flat area just to the left of the barrel
or syringe array to support the weight of the head.
5 While supporting the head with your hands, slide the head to the left and
out of the Bravo head mount.
CAUTION Support the head securely without touching the barrels, probes, or pins.
Dropping the head or bumping the barrels, probes, or pins will damage the head.
6 Carefully place the head into the storage stand to protect the barrels,
probes, as follows:
• Disposable- tip heads. Rest the bottom of the stand on a clean, dry
surface. Slide the head into the stand, with the barrels facing down, as
the following figure shows.
96lt
• AssayMAP Bravo 96AM Head. Ensure that the top of the head is
resting on a clean, stable surface so that the probes are facing up.
Carefully place the stand (1) onto the head, guiding the side cutouts
onto the two side tabs (2) on the head. Use care to avoid touching the
probes.
• Pin Tools. Rest the top of the head on a clean, dry surface with the
pins facing up. Slide the stand onto the head
Store the head top- side up and resting in the head stand.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
2 Carefully remove the liquid- handling head and protective stand from the
packaging. While the head remains seated in the stand, pull out and twist
the two head- retainer pins one- quarter turn so that they remain retracted.
CAUTION Do not rest the bottom of a liquid-handling head on any surface. Doing so
can damage the barrels, pins, or probes.
96LT
• Bravo 96AM Head. Ensure that the top of the head is resting on a
clean, stable surface so that the probes are facing up.
WARNING The probes of the AssayMAP Bravo 96AM Head are sharp and can
scratch you if they brush across your hand. A probe scratch can expose you to any
contaminants remaining on the probes. Wear gloves and use extreme caution to
avoid brushing against the probes.
Using both hands, carefully lift the stand off of the head while guiding
the stand’s side cutouts (1) off the head side tabs (2). Use care to avoid
touching the probes.
• Pin Tools. Ensure that the top of the head is resting on a clean, stable
surface so that the pins are facing up. Slide the stand off of the head.
4 With your left hand, firmly grip the left side of the head. Place your right
hand under the flat area just to the left of the array to support the weight
of the head.
5 While supporting the head with your hands, slide the head onto the Bravo
head mount. Press the head firmly into place to ensure the head is plugged
into the connector receptacle on the head mount. You should hear the click
when the retaining pins snap into place.
Note: If you do not hear the pins snap into place, check that the straight
edges of the retainer pins are in the vertical position, as the following
figure shows. Attempt to rotate the pins to ensure that they are in the
locked position. The pins should not rotate freely.
Figure Installed Bravo 96AM Head: (1) retainer pin and (2) head lock
6 To lock the head, rotate the head lock clockwise until it reaches its hard
stop. This ensures that the head is fully seated and does not shift position
during operation.
CAUTION Dropping the head or bumping the barrels or probes will damage the head.
If the Bravo head is not properly secured in place, it could drop unexpectedly. Ensure that
the head is securely locked into the head mount.
IMPORTANT If this is the first time to use this Bravo head with this
device, you must create a profile for the new Bravo head.
2 On the side of the Bravo Platform, press the power switch to the on (I)
position.
3 To initialize the Bravo Platform, click Initialize selected devices in the device
file.
4 Open Diagnostics, and in the Jog/Teach tab, make sure that a value is
displayed for each axis.
This confirms communication between the Bravo Platform and VWorks
software.
5 If you opened an existing device file in step 1, go to the Configuration tab in
Diagnostics and verify that the location configuration graphic matches what
you actually have on the Bravo deck.
Related information
Procedure
3 While the Bravo head is in motion, press the red button on the pendant,
and ensure that the Bravo head stops moving.
4 To reactivate the pendant, turn the red button clockwise. The spring-
loaded button pops up.
5 Repeat step 2.
6 While the Bravo is in motion, interrupt the Light Curtain and ensure that
the Bravo head stops moving.
Related information
CAUTION A blown fuse can indicate more serious problems. If the new fuse blows after
replacement, contact Agilent Automation Solutions Technical Support.
CAUTION Using an incorrect fuse can damage the Bravo Platform. Ensure that you have
the correct fuse for the device.
For fuse specifications, see the corresponding safety and installation guide for
your Bravo model:
• G5562A, G5563A Bravo Platform Safety and Installation Guide
Procedure
To replace the fuse in the power switch:
1 Shut down the Bravo Platform, and unplug the power cable from the rear
panel connector.
2 At the rear panel power switch enclosure, use a small flat- head
screwdriver (2.5 mm) to pry open the tab on the fuse enclosure and open
the enclosure cover.
00220
Bravo
Back
00220c
3 Insert the screwdriver head in the enclosure notch to dislodge the fuse
cartridge. Slide the fuse cartridge all the way out of the enclosure.
4 Replace the fuse in the cartridge.
5 Slide the fuse cartridge back into the fuse enclosure.
6 Press the enclosure cover securely into the closed position.
7 Plug in the power cable at the rear panel connector, and then start up the
Bravo Platform.
Related information
Front Back
Pipette barrel
The pins are spring- loaded and should retract until the pin head is nearly
flush with the surface of the pipette head. The tip of each pin, which is hidden
within the pipette head housing, has a semi- circle- shaped locking mechanism
that aligns with a dowel inside the pipette head to lock the pin in the
retracted position.
CAUTION If you apply too much force when pressing down or rotating the pin to retract it,
the locking mechanism at the tip of the pin can bend or break. A damaged pin cannot lock in
the retracted position. Use only gentle pressure to push down and rotate the pin to lock it into
place.
2 Insert the end of a 2- mm hex wrench into the head of one of the pins, and
very gently push the pin down into the pipette head until you feel the pin
stop.
Verify that the pin is properly seated, as the preceding figure shows. If the
pin has an arrow, the square end of the arrow should disappear or be
barely visible.
IMPORTANT If the pin hits a stop before it retracts completely, the pin
locking mechanism is above instead of below the locking dowel. While
removing any downward pressure on the pin head, gently rotate the pin
in quarter- turn increments until you feel the pin drop into the seated
position.
Related information
CAUTION Moving the Bravo head or gripper in any of its axes without first disabling the
servo motors can damage the device.
3 After you complete the manual moves, re- enable the motors in one of the
following ways:
• If the emergency- stop is engaged, release the red button on the
pendant by turning it clockwise.
• If the motors are disabled, click Enable all motors in the Jog/Teach tab of
Bravo Diagnostics.
IMPORTANT The Zg- brake engages when the Zg- axis is disabled, preventing
you from moving the gripper assembly manually in the Zg- axis.
3 While ensuring that one hand is free to support the labware held by the
gripper, use a 2.5- mm flat- blade screwdriver to turn the G- axis lead screw
counterclockwise to open the gripper and release the labware.
Related information
Procedure
To inspect the Bravo Platform:
1 Inspect the impacted parts for visible signs of damage.
Look carefully for any subtle signs of damage.
2 If contact was made to the pipette barrel or tip:
a Compare the impacted region to the surrounding barrels or tips.
Dented barrels can affect tip sealing.
b Replace damaged tips or barrels if disposable.
3 If contact was made with a:
a Standard platepad, make sure the alignment tabs did not become
damaged and still allow plates to be easily picked or placed.
b Alignment Station, make sure the rollers and springs are not damaged
and still allow plates to be easily picked and placed.
c Other deck locations, make sure the alignment tabs are not damaged
and still allow plates to be easily picked and placed.
Make sure the specific functionality of the deck location is not
compromised. If the operation is hindered, contact Agilent Automation
Solutions Technical Support. Be ready to describe how the collision
occurred.
Related information
The device is not connected to Ensure that you are using the
the power source, or the power Agilent- supplied country- specific
cord is damaged. power cord and plug. Inspect the
power cord and connections, and
ensure that the device is
connected to the power source.
The Bravo Platform cannot be The communications cable is Check the Ethernet cable
initialized. not properly connected, or the connections.
wrong cable is being used. If necessary, power cycle the
device and then try initializing
again.
The gripper pads (black rubber) Gently clean the gripper pads
on the interior side of each using a soft cloth dampened with
gripper finger are dirty, worn, water. Inspect the gripper pads
or partially out of alignment. for wear and ensure they are
aligned.
If necessary, contact Agilent
Automation Solutions Technical
Support to replace the gripper
pads.
The robot attempts to grip too If this problem occurs at all Verify that the labware definition
high or low on the labware. Bravo locations, the Bravo is correct.
gripper range in the labware See the VWorks Automation
definition requires adjustment. Control Setup Guide.
The labware does not sit level An obstruction is preventing Verify that the setscrews in the
within the platepad. proper seating of the labware. platepad tabs are recessed and
not protruding into the plate nest.
A protruding setscrew will prevent
proper seating of the labware.
The gripper cannot pick or The labware does not match the Verify that the correct labware is
place labware at a location. specified labware definition, or on the deck, in good condition,
the labware might be damaged and corresponds to the selected
or deformed. labware definition.
The head or gripper The Bravo Platform might Contact Agilent Automation
movement is rough or require service. Solutions Technical Support.
strained, or rubbing or
scraping noises are audible
during moves.
The labware definition for the Verify that the correct tip box is
tip box is incorrect. in the correct location and that
the labware definition is correct.
The tip box raises off the The tip box stripper pins on the Release the tip box stripper pins
platepad during Tips On liquid- handling head are on the liquid- handling head. See
operations. retracted. “Retracting tip box stripper pins”
on page 78.
Tips Off operations are not Stacked tips on a barrel. Static To resolve Tips Off problems
successful. electricity is causing the tip to caused by static electricity, select
cling to the barrel during Tips the Enable tips-off tip-touch option in
Off operations, and then a the profile.
second tip is pressed on during
Tips On operations.
Tips are askew in the tip box The tip box on the deck does Verify that the correct tip box is
or not returned to their not match the labware in the correct location.
designated well locations after definition specified in the
Tips Off operations. protocol.
The pipetting accuracy is The bias table in the liquid See the Calibrating the pipettor
inconsistent or poor. class requires adjustment for topic in the VWorks Automation
your pipetting device. Control Setup Guide.
The liquid in the wash trays The tubing is kinked or the Inspect the tubing. If necessary,
or reservoirs is overflowing Weigh Station might require recalibrate the Weigh Station.
causing flooding. recalibration.
Related information
Troubleshooting table
The There appears to be a The Ignore plate sensor If the gripper is not gripping labware, click
plate present error message during pick and place Ignore and Continue to continue the homing
appears during option is not selected in process.
initialization. the profile.
Error message related to Communications are • Turn off the Bravo Platform off, and
communication problems interrupted. then turn it back on.
or failure. • Initialize the Bravo Platform.
Related information
Reporting problems
Related information
5
Using diagnostics software
This chapter explains how to use the Diagnostics software to
control the Bravo Platform. Only administrators and experienced
personnel should use the procedures in this chapter to diagnose
errors with the Bravo Platform.
This chapter contains the following topics:
• “Homing the liquid- handling head” on page 94
• “Jogging the liquid- handling head” on page 95
• “Using the Move and Approach commands” on page 99
• “Using the gripper controls” on page 100
• “Performing a task using Diagnostics” on page 104
For a quick reference to the controls and indicators, see
“Diagnostics quick reference” on page 275.
For instructions on how to set teachpoints, see “Setting
teachpoints” on page 36.
93
5 Using diagnostics software
Homing the liquid-handling head
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should use the Jog/Teach tab controls.
WARNING The red Stop motors button in Bravo Diagnostics does not perform an
immediate stop. The Bravo head can continue to move in the same direction at the same
speed after you click the button. To perform an emergency stop, press the red button on the
emergency-stop pendant.
About homing
The liquid- handling head has a defined home position for each axis of motion.
Homing sends the head to the home position for the axes. Home the liquid-
handling head to reset the axes. For example, if you notice the Bravo Platform
is not moving to locations or teachpoints accurately, home the head.
Homing the liquid- handling head is typically done automatically on first
initialization after startup.
There are four homing axes.
WARNING Keep away from the Bravo Platform when the liquid-handling head is
moving or about to move, especially in the z-axis direction. The liquid-handling head might
not stop immediately in a collision and a pipette tip could pierce your hand.
CAUTION Before homing an individual axis, home the z-axis first so that the head avoids
potential obstacles on the deck. A collision between the head and obstacles on the deck can
result in equipment damage.
Related information
Moving the liquid- handling head “Manually moving the head or gripper”
manually on page 80
Changing the liquid- handling head “Changing the Bravo head” on page 65
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should use the Jog/Teach tab controls.
WARNING To perform an emergency stop, press the red button on the emergency-stop
pendant.
CAUTION To prevent damage to the gripper and other hardware components, do not dock
the gripper if you want to jog the liquid-handling head along the y-axis and z-axis.
Axes of motion
You can use the Jog/Teach tab controls to jog the Bravo head in the following
four directions.
Axis Description
X The horizontal distance (left- to- right) in millimeters, along the rail
from the x- axis home position.
00232
Bravo
Pinch Hazard
Y
X
00232
WARNING Keep away from the Bravo Platform when the liquid-handling head is
moving or about to move, especially in the z-axis direction. The head’s z-axis motor is
particularly powerful. It might not stop immediately in a collision and a pipette tip could
pierce your hand.
CAUTION Before you jog the liquid-handling head, remove objects from its path. Move the
head slowly and keep the jog increment small until you are certain that there is no
obstruction.
Related information
Moving the liquid- handling head “Manually moving the head or gripper”
manually on page 80
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should use the Jog/Teach tab controls.
WARNING The red Stop motors button in Bravo Diagnostics does not perform an
immediate stop. The Bravo head can continue to move in the same direction at the same
speed after you click the button. To perform an emergency stop, press the red button on the
emergency-stop pendant.
Commands defined
• Move. Instructs the liquid- handling head to move to the teachpoint for the
selected location.
• Approach. Instructs the liquid- handling head to move to the teachpoint for
the selected location, with a vertical offset equal to the value specified in
the Approach height above teachpoint box.
Note: The Approach height parameter in the profile is the height above
the labware when pipetting.
WARNING Before you command the Bravo head to move, make sure that everyone is
clear of the Bravo Platform.
CAUTION Before using the Move or Approach command, make sure the path is clear of
any obstacles that you have not defined in the profile. The move-to-position commands
present a higher damage risk to the liquid-handling head, because the head can move in large
increments.
Procedure
To use the Move command:
1 Open Diagnostics, and then click the Jog/Teach tab.
2 Select a deck location from the Location box.
Note: The teachpoint coordinates for the selected location are displayed
under the Location list.
3 Click Move.
The liquid- handling head moves to the teachpoint for the selected location.
Related information
Using the Jog/Teach tab controls to • “Homing the liquid- handling head”
move the head on page 94
• “Jogging the liquid- handling head”
on page 95
G Zg
Related information
Using the Jog/Teach tab controls to • “Homing the liquid- handling head”
move the liquid- handling head on page 94
• “Jogging the liquid- handling head”
on page 95
Moving the liquid- handling head “Manually moving the head or gripper”
manually on page 80
WARNING The red Stop motors button in Bravo Diagnostics does not perform an
immediate stop. The Bravo head can continue to move in the same direction at the same
speed after you click the button. To perform an emergency stop, press the red button on the
emergency-stop pendant.
Procedure
Performing a task requires that you:
• Choose a location
• Select a task and task parameters
• Select labware
To perform a task:
1 Select a location. You can do one of the following:
• Click the location in the graphical display.
• Select from the Location list.
2 Select the labware from the Labware at selected location list.
3 If the task involves two locations, repeat step 1 and step 2.
For example, if you select the Assemble Vacuum task for a Vacuum
Filtration Station, you could have a collection plate at Location 1 and a
filter plate at Location 2.
4 Series III disposable- tip heads and Bravo 96AM Head only. To use one
row or column of channels instead of all channels in the head:
a Click Set head mode.
b In the Head Mode Selector dialog box, select the barrels, and then click
OK.
Note: If you select a single row on a disposable- tip head, make sure you
retract the head stripper pins.
5 If you are using the head in serial mode or the head has fewer tips than
the plate has wells, select the quadrant or quadrants in the Well Selection
and Head Mode area.
Click a representative well in the plate graphic to select the corresponding
quadrant of wells. The selection message appears below the plate graphic.
6 In the Command Parameters area:
a Select the process from the Command to execute list.
b Set the parameter values for the command.
Related information
A
Accessories and platepads
This chapter describes the accessories platepads that can be used
with the Bravo Platform and provides the installation and removal
procedures.
This chapter contains the following topics:
• “Bravo accessories overview” on page 108
• “Installing and setting up an Orbital Shaking Station” on
page 110
• “Using the teach plate to set the teachpoint for an accessory”
on page 117
• “Installing and setting up a Vacuum Filtration Station” on
page 118
• “Setting up a Trash or Filter Plate Holder accessory” on
page 138
• “Configuring a platepad for delidding” on page 143
• “Installing and configuring a platepad or Alignment Station” on
page 144
• “Installing and configuring a Magnetic Bead Accessory” on
page 147
• “Installing and setting up a Nested Rack Insert” on page 151
• “Using the Manual Fill Reservoir” on page 154
• “Setting up the Thermal Station (cooling pad)” on page 155
• “Installing a Heating Shaking Station” on page 158
• “Setting up thermal and shaking stations (Inheco controller)”
on page 161
• “Installing and configuring the Peltier Thermal Station” on
page 167
• “(Bravo SRT only) Setting up an LT tip box location” on
page 173
For details on how to set up accessories on the AssayMAP Bravo
Platform, see “About installing the AssayMAP Bravo Platform” on
page 209.
107
A
Bravo accessories overview
Description
You can add accessories to the Bravo Platform to enhance existing functions
and facilitate operation. The accessories and platepads include:
SRT platepad for Enables tips on and tips off “(Bravo SRT only)
250- µL tip boxes tasks using the 250- µL tip box Setting up an LT tip
on the shorter SRT Bravo box location” on
Platform. page 173
Related information
Description
The Orbital Shaking Station is a microplate shaker that can be installed at any
deck location on top of a short adapter pad instead of a platepad. The Orbital
Shaking Station has the option of using its own control module, or an Inheco
Single TEC controller (STC) or Multi TEC controller (MTC). For a full
description of the Orbital Shaking Station, see the manufacturer’s
documentation.
Figure Orbital Shaking Station with its own control module and an adapter pad
Setup workflow
WARNING Turn off the Bravo Platform before you install or remove the accessory.
Ensure that you remove the platepad from the deck location where you plan to
install the Orbital Shaking Station.
3 Place the Orbital Shaking Station atop the adapter pad so that the cable is
towards the back of the Bravo deck.
Make sure the station is completely centered on top of the adapter and sits
level.
4 Insert the two M3 x 18- mm socket- head cap screws into the holes at the
front and back end of the adapter pad and use a 2.5- mm hex driver to
tighten the screws.
Tighten screws evenly to ensure the feet on the bottom of the Orbital
Shaking Station are pressed evenly into the adapter pad and the pad
remains level.
Note: The feet of the Orbital Shaking Station are compressible, and these
screws do not reach a hard stop as a result.
CAUTION The Orbital Shaking Station can shift slightly when tightening. Ensure that
the sides of the Orbital Shaking Station are parallel to the sides of the adapter plate. If the
Orbital Shaking Station is rotated, it could cause the Bravo gripper to collide with the
accessory.
2 1
2 Connect the serial cable with the red stripe to the leftmost port on the
control module (figure, item 2). Connect the other end of the cable to the
serial port, if available, on the controlling computer.
Next steps:
1 Install a platepad or accessory at this deck location.
2 In Diagnostics, select the appropriate profile, and then click the Configuration
tab and reconfigure this location.
3 Adjust the teachpoint of the location.
2 Type the RPM, select the Stir direction, and then click Start.
3 At the Orbital Shaking Station on the deck, verify that the accessory
operates correctly.
4 In the Bravo Accessory Diagnostics dialog box, click Stop, and then click OK.
Related information
IMPORTANT When editing the teachpoint for the accessory, you must account
for the teach plate height before setting the teachpoint.
Related information
Figure Vacuum Filtration Station components: A collar options, B base and optional inserts
A
1 2
3 Inserts The white plastic spacers that can be placed in the base
manually before starting the protocol, if required, to raise
the height of a collection plate that sits underneath the
collar. Two sizes are available. You may use multiple
inserts in combination for a combined height of 15 mm,
maximum.
For more details about the Vacuum Filtration Station collar and base, see the
Millipore user documentation.
• Configuration B. The filter plate is not part of the station assembly. The
robot will move the filter plate to the station during the protocol run after
the assembly process is finished.
• Configuration C. The filter plate is not part of the station assembly. The
robot will move the filter plate to the station during the protocol run after
the assembly process is finished.
A B C
Filter plate
Collar
Filter plate
Collar
Filter plate
Collar
Collection plate Collection plate
Figure Removal of (1) plastic plug from vacuum port and (2) rubber feet from base
Figure Installation options on the Bravo deck: location 1, 2, or 3 on the back row
3 Connect the Vacuum Filtration Station as required for the type of vacuum
source:
• House vacuum with pinch- valve module.
• Agilent VARIO Pump.
2 Connect the pinch- valve tubing “Connecting the pinch valve tubing”
and cables. on page 125
4 Connect the coupling insert on the end of the vacuum pressure gauge
tubing (4) to the quick- disconnect fitting on the long arm of the four- way
connector.
5 Optional. Disconnect the manual on/off toggle switch (5) from the end of
the tubing. This toggle switch is not required for the setup.
Connecting the pinch valve tubing
Note: The pinch- valve tubing can collapse with regular use over time, resulting
in inadequate vacuum flow for the Vacuum Filtration Station. Replace the
tubing periodically to keep the Vacuum Filtration Station functioning as
intended.
The following figure shows the connections from the Vacuum Filtration Station
to a vacuum source via a pinch- valve module.
3
5
4
6 7
• Vent port Routes the vent tubing on the pinch- valve module.
• Valve port Routes the tubing from the vacuum source to the
Vacuum Filtration Station through the waste
bottle.
3 Waste bottle and Collects the waste solution from the filtration
tubing process.
IMPORTANT If you are installing the control valve- and- gauge assembly
with the optional on/off toggle switch, ensure that the toggle switch is
in the on position (handle in line with valve body), as the following
figure shows.
Note: The remaining tubing from the three- way connector is pre-
assembled: One tube connects to the vent port, and the second tube
connects through the valve port.
3 Connect the open end of the pinch- valve tubing from the valve port (2) to
a port on the waste bottle (3).
4 Connect tubing between the second port of the waste bottle (3) and an in-
line filter (4).
5 Connect the tubing between the in- line filter and the vacuum source (5).
6 Connect the power cable (6) from the pinch- valve module to an AC outlet.
7 Connect the communications cable (7) from the pinch- valve module to a
com port on the controlling computer.
Setting the vacuum pressure
To set the regulated vacuum pressure for the pinch valve:
1 Place a solid- bottom collection plate or the teach plate on top of the
installed collar on the Vacuum Filtration Station.
2 At the vacuum control valve, turn the knob fully counterclockwise (CCW)
so that the valve is completely open, preventing any vacuum pressure in
the tubing.
If the vacuum pressure exceeds 8.46 inHg, the pinch valve tubing may
collapse. If this occurs, turn off the vacuum source to the pinch valve and
turn the control- valve knob counterclockwise to eliminate the vacuum
pressure. You may need to massage the tubing slightly to open it up again.
Now you are ready to configure the Vacuum Filtration Station in the software.
See “Configuring the Vacuum Filtration Station for the VWorks software” on
page 132.
The following figure shows the Agilent ME4C NT VARIO Vacuum Pump.
CVC 300
0
For detailed instructions about the VARIO pump, see the manufacturer’s
Instructions for use, provided with the pump.
The following figure shows the connections for a Vacuum Filtration Station that
uses the ME4C NT VARIO Vacuum Pump.
Figure Vacuum Filtration Station connections for ME4C NT VARIO Vacuum Pump
5 7
2
4
9
3
6
Item Description
1 Vacuum Contains a vacuum port on that base that
Filtration connects to the tubing.
Station
2 Three- way Branches the tubing from the station’s base to the
connector pump and the waste bottle.
To connect the Vacuum Filtration Station to the ME4C NT VARIO Vacuum Pump:
1 Make sure the Vacuum Filtration Station base is installed on the platepad.
2 Using the three- way tubing connector:
a Connect the tubing with the quick- disconnect fitting to the port on the
station base (1), and connect the other end to the three- way connector
(2).
b Connect the input tubing from the waste bottle (3) to the three- way
connector.
c Connect tubing from the three- way connector to the port on the back
of the controller (5).
3 Connect the output tubing from the waste bottle (3) to the filter (4), and
connect the output tubing from the filter to the inlet on the front of the
pump (6).
4 Carefully, connect the pump’s power supply cable (7) to the 4- pin
connector on the back of the controller (5).
CAUTION Do not try to force the cable connector into the port on the controller.
Doing so can damage the connector.
5 Connect the communications cable (8) from the pump controller to a com
port on the controlling computer.
6 Connect the power cord (9) from the pump to an AC outlet.
Bravo Platform User Guide 129
A
Installing and setting up a Vacuum Filtration Station
CVC 3000
2 1
2 If the controller power is off, press the power button to turn on the
display.
Note: The controller display will turn on automatically if the controller
power button was already set to on.
3 Access the controller configuration display as follows:
a Press the selection knob to display the settings menu. Turn the knob to
move the selector to Configuration, and then press the knob to make the
selection.
b In the Configuration display, turn the knob to move the selector to
RS-232, and then press the knob to make the selection.
c In the RS-232 configuration display, ensure that the parameters are set
as follows:
RS-232 parameters Values
Baud 9600
Parity 8- N- 1
Handshake None
The PC icon will appear after you initialize the Vacuum Filtration
Station in your automation software, indicating that the pump is under
remote control.
CVC 3000
WARNING Before using the VARIO pump, refer to the safety information in the
manufacturer’s user documentation.
Note: While the PC icon appears on the controller display, the Vacuum
Filtration Station is under the control of the computer, and you can no longer
control the pump locally using the controller display.
If you want to reset the controller, press the power button to turn it off, set
the power switch on the back of the pump to off. Wait a minute, and then turn
on the pump and turn on the controller again.
For additional information on the VARIO pump, see the manufacturer’s user
documentation.
WARNING To avoid potential injury and damage to the device, turn off the device
before installing or removing an accessory.
3 If applicable, disconnect the pinch- valve module from the vacuum source.
4 Disconnect the tube from the Vacuum Filtration Station.
5 Remove the set screw that is holding the station base to the platepad.
6 Remove the Vacuum Filtration Station from the platepad.
Pump Description
Standard Select this option if the station is connected to the
pinch- valve module.
Pump Description
ME4CVario Select this option if the station is connected to the
ME4C NT VARIO Vacuum Pump.
To configure the ME4C NT VARIO Vacuum Pump and test the station:
1 Verify that the pre- assembled Vacuum Filtration Station is in place on the
target deck location and connected to the VARIO Vacuum Pump.
2 In Diagnostics, click the Configuration tab.
3 In the Accessory Configuration area, highlight Vacuum Filtration Station, and then
click Diagnose accessory.
Figure Accessory Diagnostics dialog box for the ME4C NT VARIO Vacuum Pump
Related information
Figure Bravo Platform on deck risers and top view of deck location cutouts
4 6
3 Position the Filter Plate Holder in the cutout, aligning the two dowel pins
in the holes.
4 Install the four M3 screws that secure the Filter Plate Holder to the deck.
The image of the accessory appears at the specified deck location in the
Configuration tab, as the following figure shows.
Figure Example of VWorks protocol with static labware configuration for tip trash at deck location 6
Related information
Related information
Descriptions
Standard platepad
The standard platepad can accommodate any microplate- sized labware. The
platepad has a crosshairs (1) in one corner, as the following figure shows,
which is used when setting pipettor teachpoints.
Alignment Station
An Alignment Station is a platepad that has three springed rollers on two of
the sides. The springed rollers are designed to push a microplate into the
opposing corner, securing the microplate position and ensuring precise
pipetting. Like the other platepads, it contains a crosshairs (1) for setting
pipettor teachpoints.
The Alignment Station can accommodate 1536- well microplates, small- transfer
(ST) tip boxes, and any other microplate requiring precision location.
When installing the platepad or Alignment Station, ensure that you orient the
crosshairs towards the northwest corner of the deck, as the following figure
shows.
WARNING Turn off the Bravo Platform before you install or remove the accessory.
3 Verify the teachpoint for this location. For instructions, see “Setting
teachpoints” on page 36.
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to set up Magnetic Bead Accessory and teach the Bravo Platform should perform the
following procedures.
2 Perform the following steps to securely mount the magnet to its backing
plate without the springs to eliminate any warping:
a Remove the four corner screws from the magnet (item 1 in the following
figure.
b Remove the springs from between the black plastic magnet mounting
(figure, item 2) and metal backing plate (item 3).
c Replace the screws with shorter ones, or use small washers (size M3
works well) to fill the gap and allow the screws to be tightened so that
they press the plastic magnet mounting firmly onto backing with the
stickerless side facing down.
3 Set the Magnetic Bead Accessory aside temporarily.
In subsequent procedures, you will set the teachpoint at the platepad
crosshairs, secure the Magnetic Bead Accessory in place atop the platepad,
and then adjust the teachpoint for the accessory.
Click Teach.
4 Click Move to safe height at current X/Y position.
6 Using the 1.5 mm wrench, tighten the set screw in the tab, as shown in
preceding figure (2), to secure the base to the platepad.
7 In the Jog/Teach tab image of the Bravo deck, click the accessory Location.
8 In the Approach height above teachpoint list, select 2 mm, and then click
Approach.
The Bravo head moves to the deck location and moves down so that the
pipette tip is approximately 2 mm above the Magnetic Bead Accessory.
9 Verify that the pipette tips are centered above the magnet rings in the
x- and y- axes.
If necessary, adjust the x- and y- axes at this point, and then jog the head
down by 2 mm to the teachpoint z- axis and click Teach.
10 In the Profiles tab, click Update this profile to save the changes.
IMPORTANT Verify the this accessory teachpoint for your microplate type
using a full set of tips. For guidance, contact Agilent Automation Solutions
Technical Support.
Related information
6 Set the teachpoint for the Nested Rack Insert using the same procedure
that you would for a platepad.
The Nested Rack Insert is approximately 5 mm taller than a standard
platepad.
7 In the Profiles tab, click Update this profile.
8 To verify the setup, perform a tips- on task.
Related information
Description
The Manual Fill Reservoir is an open tray that you can install on a platepad to
supply reagents to 96- and 384- channel pipette heads. You must manually refill
and empty the reservoir.
Automation Solutions reservoirs are approved for use with most reagents and
solvents. If you have questions on the use of a particular chemical or solvent
in a Automation Solutions reservoir, contact Automation Solutions Technical
Support prior to use.
Figure Two types of Manual Fill Reservoirs: 96-well and 384-well options
96-well 384-well
WARNING Before you place labware on the deck, ensure that the pipette head is not
moving.
Related information
Setup workflow
2 Edit the teachpoint using the teach “Using the teach plate to set
plate. Be sure to consider the 10- mm the teachpoint for an
thickness of the teach plate when you accessory” on page 117
edit the teachpoint.
3 Connect the power cord from cooling cube Liquid chilling unit
to an AC outlet. manufacturer’s user
documentation
5 Make sure the unit is on, and set the Liquid chilling unit
temperature. manufacturer’s user
Verify that the cooling pad achieves the set documentation
temperature.
Related information
Description
The Heating Shaking Station provides fluid mixing and heating for microplates.
The Heating Shaking Station uses a short adapter pad to attach to the Bravo
deck. The station uses the Inheco Single TEC Control (STC) or Multi TEC
Control (MTC) to connect to the computer and power source.
WARNING To prevent potential injury and damage to the device, turn off the Bravo
Platform before you install or remove an accessory.
The Heating Shaking Station may be installed at any deck location using the
short adapter pad instead of a platepad.
• Heating Shaking Station, including adapter pad, M6 x 12 flathead screw (to
attach the adapter pad), and two M3 x 18 socket- head cap- screws
• 2.5- mm and M5 hex wrenches
• Inheco STC or MTC Controller with the corresponding PCA installed for
the Heating Shaking Station, power and communication cables, and Inheco
user documentation
Use the following workflow.
After installing the accessory, you configure it in the Bravo profile, establish
communication, and set the teachpoint for the location.
Related information
The STC can control a single Inheco device, and the MTC can control multiple
Inheco devices. For a detailed description of the Inheco STC or MTC, see the
Inheco STC or MTC user documentation.
WARNING Burn hazard. During the heating process, the thermal stations can have hot
surfaces that can cause a burn injury. Allow the thermal stations to cool before attempting
to touch them.
The following warning symbol appears on the thermal stations, warning you of
the potential burn hazard.
4 In the Bravo Accessory Diagnostics dialog box, verify that the correct device
appears in the display at the top.
Note: The software obtains the device name from the STC or MTC, and
then the Inheco name of the device appears in the display of the dialog
box, for example, CPAC Ultraflat or Teleshake 95.
5 Shaking accessories only. Set the parameters in the Shake area:
a In the Shake speed box, type the shaking speed.
b In the Shake duration box, type the time duration (seconds).
c In the Shake type box, select the type of stirring motion.
d Click Start to test the shake parameter settings. At the device, verify
that the accessory operates correctly.
Click Stop to turn off the shaking feature.
6 Thermal accessories only. In the Heat area, set the parameters to test the
temperature function:
a In the Target temp box, type the temperature (°C).
b To specify a time limit within which to reach the target temperature,
select the Wait for temperature to be reached check box, and then set the
following:
• In the Temperature tolerance box, type the ±°C.
• In the Time limit box, type a time limit (minutes).
c Click Start to test the Heat parameter settings. At the device, verify that
the accessory operates correctly.
Click Turn Off to stop the heating or cooling.
Bravo Platform User Guide 163
A
Setting up thermal and shaking stations (Inheco controller)
IMPORTANT If you have a Peltier Thermal Station with a custom plate nest
and adapter, Contact Automation Solutions Technical Support for guidance on
how to teach your combination of microplate, plate nest, and adapter.
Figure Skirted microplate sitting in Peltier Thermal Station with generic plate nest and adapter
(cut-away side view)
1
2
As the figure shows, the height difference between the adapter top (1) and the
plate nest top (2) enables a skirted microplate to nest in the station. The well
bottoms must sit directly on the adapter top to enable an efficient temperature
transfer. However, the teachpoint must account for this height difference to
ensure that the Bravo robot grips the microplate at the correct position and
performs pipetting tasks at the correct height.
CAUTION If your microplate does not sit in the plate nest as the figure shows, contact
Automation Solutions Technical Support for guidance on whether your microplate requires a
custom plate nest and adapter.
To set the teachpoint for a generic Peltier Thermal Station or Heating Shaking Station:
CAUTION Use the following procedure only on a Bravo Platform with a disposable-tip
head.
1 Place the microplate securely in the plate nest of the thermal station.
Verify that the microplate sits in the plate nest so that the well bottoms
are resting on the adapter and the skirt rests in the plate nest bottom
surface. Remove the microplate.
2 Using calipers, measure the height difference between the adapter top (1)
and the plate nest top (2) on all four sides of the plate nest interior. Make
a note of the smallest number. Later in the procedure, you will add this
value to the z- axis value for the teachpoint.
3 Place the teach plate in the plate nest.
Note: Alternatively, you can use a 384- well microplate instead of the
teach plate to set the teachpoint xy coordinates. In this case, you set the
xy coordinates based on the center of the A1 well quadrant, as the
following figure shows.
1
A
4 In Bravo Diagnostics, click the Profiles tab, and initialize the desired profile.
5 In Bravo Diagnostics, click the Processes tab, select a tip box, and perform a
Tips On command. For details, see “Performing a task using Diagnostics” on
page 104.
CAUTION Using the wrong tip for teaching can result in a head crash. Make sure the
disposable tip is an Automation Solutions tip and is the same as the Teaching tip type
specified in the profile.
6 In Bravo Diagnostics, click the Jog/Teach tab, and set the initial teachpoint as
follows:
a Select the deck location of the station from the Location list.
b Use the controls to jog the head so that the A1 pipette tip is directly
over the crosshairs with paper- thin clearance. For details, see “Setting
the first teachpoint” on page 38.
c Click Teach to set the teachpoint.
d In the Profiles tab, click Update this profile.
7 Adjust the teachpoint z- axis value for the teach plate height as follows:
a Jog the head up in the z- axis so that you have enough room to remove
the teach plate. Remove the teach plate.
b In the Jog/Teach tab, use the controls to jog the head down in the
z- axis until there is paper- thin clearance between the end of the tip
and the top surface of the adapter.
Note: When the z- axis error message appears, you may click Ignore.
c Click Teach to set the teachpoint.
d In the Profiles tab, click Update this profile.
8 In Bravo Diagnostics, click the Processes tab, select the empty tip box, and
perform a Tips Off command. For details, see “Performing a task using
Diagnostics” on page 104.
9 Adjust the teachpoint z- axis value for the height difference between the
adapter and the plate nest as follows:
CAUTION Before you perform this step, ensure that the tips are off. Otherwise, a
crash can occur between the tips and the station.
a In the Jog/Teach tab, select the deck location of the station, and click
Move.
b Jog the head down in the z- axis by the measured value from step 2.
c Click Teach to set the teachpoint.
d In the Profiles tab, click Update this profile.
CAUTION After adjusting the teachpoint to account for the height difference
between the adapter and the plate nest, do not use tips to verify the teachpoint.
Otherwise, a crash can occur between the tips and the station.
Related information
Description
The Peltier Thermal Station uses Peltier technology to provide temperature
control uniformly across a microplate. The station can be fitted with a variety
of plate nests and adapters to provide efficient heating and cooling control for
different types of microplates.
This accessory requires the Bravo risers to raise the Bravo deck surface and
accommodate the height of the accessory below the deck. To install Bravo
risers, contact Agilent Automation Solutions Technical Support.
The following figure shows the Peltier Thermal Station fitted with the custom
plate nest.
inheco
IMPORTANT Ensure that you have the correct plate nest and adapter for the
requirements of your assay. Your microplate must nest in the station correctly
to ensure efficient heat transfer. For more information, contact Agilent
Automation Solutions Technical Support.
CAUTION Ensure that the connector end of the cable on the Peltier Thermal Station does
not knock against the metallic table top or other areas as it could get permanently damaged.
Handle the Peltier Thermal Station cable with care during all stages of installation.
1 2 3
4 5 6
7 8 9
inheco
3 Uninstall the plastic interface from the top of the Peltier Thermal Station
by removing the four screws located at the four corners.
4 Refer to the following figure, and then place the Peltier Thermal Station
assembly into the cutout hole as follows:
Figure Placing the Peltier Thermal Station into the deck hole
a Carefully slide the cable into the cutout hole first before you place the
Peltier Thermal Station assembly into the cutout.
b Ensure that the safety label faces the front of the Bravo deck.
c Press and hold from the bottom to squeeze in the metal as you insert
the Peltier Thermal Station assembly into the hole.
d Position the bracket around the outside edges of the top area of the
assembly.
e If there is slight resistance when fitting the assembly into the cutout or
when positioning the bracket, you may use a rubber mallet to tap the
metallic base or bracket into place.
CAUTION Using too much force to knock the base into place can damage the
Peltier Thermal Station.
f Ensure that the Peltier Thermal Station sits flush in the cutout and
that the bracket is seated on the Bravo deck.
5 Install the four M3 bolts that secure the bracket to the deck.
6 Install the custom plate nest on the plastic top of the Peltier Thermal
Station as follows:
a Position the plate nest and ensure that it is fully seated.
b Install the M3 screw in the center of the plate nest to secure it to the
Peltier Thermal Station.
Note: The accessory cable is color- coded to match the card installed
in the STC or MTC.
b Connect the STC or MTC Controller power and communication cables.
See the Inheco user documentation for details on how to connect the
STC or MTC.
c Press the STC or MTC Controller power switch to the on (|) position.
After installing the accessory, you configure it in the Bravo profile, establish
communication, and set the teachpoint for the location.
Related information
Figure SRT platepads: (A) ST tip loading station, and (B) SRT platepad for 250 µL LT tip boxes
A B
CAUTION To prevent potential hardware crashes when using 250-µL tips, make sure you
understand the following limitations of the z-axis clearance on the Bravo SRT.
• Be aware that the Bravo SRT has insufficient clearance to move labware if
the 250- µL tips are installed on the liquid- handling head. The Bravo SRT
can move labware only if 250- µL tips are not on the liquid- handling head.
• Limit tips on and tips off tasks to deck locations where an SRT 250- µL tip
box platepad is physically installed and configured in the Bravo Diagnostics
software. This special platepad provides additional z- axis clearance, and is
designed for a tip box that holds 96 250- µL tips.
• Use labware classes for each deck location so that the 250- µL tip box
operations are not attempted at deck locations with insufficient z- axis
clearance.
• Use VWorks Task Groups to ensure that tips on and tips off tasks are
performed before plate movement tasks in the protocol.
Ensure that you have the following:
• M5 hex wrench for removing a standard SRT platepad
• SRT platepad for 250- µL tip box
• Star- head screw and wrench for installing the SRT 250- µL tip box platepad
Workflow
4 In the Location is configured as list, select (SRT only) platepad for 250 uL tips.
5 Click the Jog/Teach tab, and teach the deck location using the crosshairs
on the platepad. See “Setting teachpoints” on page 36.
If you previously taught this deck location using the Set all teachpoints based
upon selected teachpoint feature, you can:
a Select the platepad Location, and click Move to go to the preset
teachpoint.
b Using small increments, jog down in the z- axis until you have paper-
thin z- axis clearance between the pipette tip and the crosshairs.
c When the tip is in the correct position (directly over the crosshairs
with paper- thin clearance), click Teach.
Note: Although the 250- µL tip box actually sits on the Bravo SRT deck, the
software makes an adjustment in the z- axis to compensate for the height
difference.
6 In the Profiles tab, click Update this profile.
Related information
B
Autofilling accessories
This chapter describes the autofilling accessories that can be used
with the Bravo Platform and provides the installation and removal
procedures. This chapter contains the following topics:
• “Autofilling station overview and setup workflow” on page 178
• “Configuring an autofilling station for the Bravo Platform” on
page 179
• “Setting up a Weigh Station for the Bravo Platform” on
page 184
For details on how to set up accessories on the AssayMAP Bravo
Platform, see “About installing the AssayMAP Bravo Platform” on
page 209.
177
B
Autofilling station overview and setup workflow
Autofilling accessories
Accessory Description
Pump Module Pumps fluids to and from one of the Agilent autofilling
type reservoirs.
For more details on the autofilling accessories, see the Pump Module User
Guide.
Related information
Parameter Settings
Fill module Specify the Pump Module that this deck location
uses. For example, select 1 if only one Pump
Module is connected to the device.
One Pump Module can function as both the Fill
module and Empty module.
Parameter Settings
Pump for filling Select which of the two pumps on the Pump
Module will fill the container, where,
• Pump 1 controls the upper pump.
• Pump 2 controls the lower pump.
Direction for Select the pumping direction, where,
filling • Forward moves the pumped fluid in the
forward direction of the pump.
• Reverse moves the fluid in the opposite
direction.
Empty module Specify the Pump Module that this deck location
uses. For example, select 1 if only one Pump
Module is connected to this device.
One Pump Module can function as both the Fill
module and Empty module.
Pump for Select which of the two pumps on the given
emptying Pump Module will empty the container on the
deck location, where,
• Pump 1 controls the upper pump.
• Pump 2 controls the lower pump.
Direction for Select the pumping direction: Forward or Reverse.
emptying
Use Weigh If you are using a Weigh Station at this location,
Station? select Yes.
Weigh Station If connecting the Weigh Station serial cable to the
module number Pump Module SHELF A input port, select 1.
If more than one Weigh Station is connected, the
module number corresponds to the order in
which they are connected to the controlling
computer.
Note: Agilent Technologies recommends using one
Pump Module for each pairing of reservoir and
Weigh Station.
e Click Next, and then click Finish on the Summary page.
5 When the accessory configuration message appears and advises you to set
the teachpoint to a safe value, do one of the following:
• If you are using a Weigh Station at this location, click Yes to move the
teachpoint to a safe value.
On the Profiles tab, click Initialize this profile , and then reset the
teachpoint for the installed Weigh Station.
• If you are placing the reservoir on a standard deck location, click No to
preserve the existing teachpoint. On the Profiles tab, click Update this
profile to save the settings.
To test autofilling:
1 In Diagnostics, initialize the profile, and then click the Configuration tab.
2 In the Accessory Configuration area, highlight the Autofill Station icon, and then
click Diagnose accessory.
Note: The Autofill Station is represented by a Pump Module graphic in
the software.
Control Description
Pump on time (s) Specifies the duration that the selected pump on
the Pump Module runs when you click Run
pumps.
Fill module Specifies the Pump Module that this deck location
uses. One Pump Module can function as both the
Fill module and Empty module. For example,
select 1 if only one Pump Module is connected.
Fill pump Specifies which of the two pumps on the Pump
Module to use for filling the container, where
1 controls the upper pump.
2 controls the lower pump.
Fill direction Forward. Moves the pumped fluid in the forward
direction of the pump.
Reverse. Moves the fluid in the opposite direction.
Pump fill speed Specifies how quickly the pump moves the fluid,
(0–100%) where 100% is the fastest setting.
Empty module Specifies the Pump Module that this deck location
uses. One Pump Module can function as both the
Fill module and Empty module. For example,
select 1 if only one Pump Module is connected.
Empty pump Specifies which of the two pumps on the Pump
Module to use for emptying the container, where
1 controls the upper pump.
2 controls the lower pump.
Empty direction Forward. Moves the pumped fluid in the forward
direction of the pump.
Reverse. Moves the fluid in the opposite direction.
Pump empty speed Specifies how quickly the pump moves the fluid,
(0–100%) where 100% is the fastest setting.
Run pumps Starts the pumps on the selected Pump Module in
real time.
Stop pumps Stops the pumps in real time.
4 Click Run pumps and verify that the pump is functioning properly.
5 Click Stop pumps, and then click OK.
Related information
00268
Bravo
weigh pad
3 In the Accessory Configuration area, highlight Autofill Station, and then click
Diagnose accessory.
Note: The Autofill Station is represented by a Pump Module graphic in
the software.
5 At the Bravo Platform, place the reservoir on the Weigh Station. Make sure
the reservoir is empty.
6 In the Accessory Diagnostics dialog box, click Set Tare to configure the empty
setting.
8 Verify that the Level reading decreases as the liquid level decreases while
you run the designated empty pump to remove any liquid from the
reservoir.
9 Click OK.
10 On the Profiles tab, click Update this profile.
Related information
C
Setting up the Barcode Reader
This section contains the following topics:
• “About setting up the Barcode Reader” on page 188
• “Installing or removing the Barcode Reader” on page 190
• “Creating a profile for the Barcode Reader” on page 191
• “Specifying the Barcode Reader location” on page 193
• “Testing and optimizing barcode scanning” on page 195
187
C
About setting up the Barcode Reader
Figure Barcode Reader oriented to scan barcodes on the east side of microplates
Mirror
00274
Overall Bar Code Reader with mirror
Setup workflow
The following table presents the workflow for installing and setting up the
Barcode Reader. The procedures can vary depending on the brand of barcode
reader sensor head (for example, Keyence or Microscan).
Note: The symbology must be stored in the barcode reader in order for the
barcode reader to decode the barcode symbol.
Related information
WARNING Turn off the Bravo Platform before you install or remove the accessory.
Related information
Creating a profile for the Barcode “Creating a profile for the Barcode
Reader Reader” on page 191
Procedure
To add the Barcode Reader device and create a profile:
1 In the VWorks window, create a new or open an existing Bravo device file.
2 In the Available Devices area, double- click the Microscan Barcode Reader.
The Microscan Barcode Reader icon appears in the Devices area.
3 In the Devices area, select the Microscan Barcode Reader icon, and click Device
diagnostics. The Barcode Reader Diagnostics dialog box opens.
4 In the Profiles tab under Profile Management, click Create a new profile. The
Create Profile dialog box opens.
5 Type a name for the new profile and click OK. The new profile name
appears in the Profile name list.
6 In the Profile Settings area, select the Accessories Hub Serial port that is
connected to the Barcode Reader.
7 Microscan only. In the Default barcode side list, select East or West to
indicate the side of the labware that has the barcode label.
Note: The Bravo Barcode Reader can scan barcode labels on the labware
east and west sides only.
8 Click Update this profile to save the newly created profile.
9 Click Initialize this profile to establish communication with the Barcode
Reader.
10 In the VWorks software window, choose File > Save and save the device file.
Related information
Procedure
To specify the Microscan Barcode Reader location in the Bravo device file:
1 Open the Bravo device file that contains the Barcode Reader device.
2 In the Devices area, expand the Bravo device. The list of deck locations
appear.
3 Select the deck location at which you installed the Barcode Reader. In the
example in step 2 the Barcode Reader is installed at deck location 7, so
deck location 7 is selected.
4 In the location properties table, select the Barcode Reader for one of the
following:
• BCR on west side
5 When you are finished, click Initialize selected devices to re- establish
communication with the Bravo and use the updated device file.
Related information
Creating a profile for the Barcode “Creating a profile for the Barcode
Reader Reader” on page 191
WARNING Class II laser hazard. Looking directly at the laser beam can result in
serious eye injury. Do not look directly at the laser beam.
WARNING Laser emission from the reader does not automatically turn off if the sensor
head is disassembled. Do not disassemble the barcode reader sensor head.
5 Click Scan. The barcode reader laser turns on briefly to scan the barcode
label. One of the following messages appears next to Result:
6 If the No barcode found message appears, click Scan again, and check that
the barcode reader laser beam is projected onto the vertical center of the
barcode label.
If the laser beam is not at the vertical center of the label, adjust the scan
angle of the Barcode Reader mirror.
7 To optimize the scanner settings, click Calibrate.
Note: During the calibration process, the scanner attempts various
settings to determine the optimum decode rate for the given conditions.
00275
Bar code reader with mirror detail
2 Tilt the Barcode Reader mirror slightly. If the angle is set up properly, you
can see the laser beam at the vertical center of the barcode label each time
you click Scan barcode.
Tilt
00275
Bar code reader with mirror detail
3 Tighten the screws on both sides of the barcode reader sensor head.
4 Repeat the procedures to test the barcode scanning and adjust the scan
angle until the reader is able to scan the barcode successfully (the barcode
text appears next to Scan result).
Troubleshooting
If the reader is unable to scan the barcode after many adjustments, try one or
more of the following:
• Make sure the barcode label is applied properly onto the microplate.
• Make sure the microplate is sitting level on the Barcode Reader platepad.
• Make sure the barcode meets the requirements. See “Barcode label
specifications” on page 189.
• Repeat the adjustment process using a new spare microplate.
Related information
Creating a profile for the Barcode “Creating a profile for the Barcode
Reader Reader” on page 191
D
AssayMAP Bravo Platform
This section describes the AssayMAP Bravo Platform and includes
the following topics:
• “AssayMAP Bravo Platform overview” on page 200
• “Bravo 96AM Head overview” on page 202
• “AssayMAP cartridge overview” on page 206
• “About installing the AssayMAP Bravo Platform” on page 209
• “Installing and using the Bravo 96AM Head” on page 211
• “Installing the 96AM Wash Station” on page 219
• “Turning on the AssayMAP Bravo Platform” on page 224
• “Setting up the AssayMAP Bravo Platform” on page 225
• “Verifying the profile settings for the AssayMAP Bravo
Platform” on page 227
• “Verifying the AssayMAP Autofill Station configuration” on
page 234
• “Verifying the AssayMAP Peltier Thermal Station configuration”
on page 236
• “Verifying the AssayMAP Orbital Shaking Station configuration”
on page 238
• “Testing the AssayMAP accessories” on page 240
• “Teaching part 1: Setting teachpoints in the AssayMAP 1
profile” on page 244
• “Teaching part 2: Adjusting AssayMAP teachpoints for
accessories” on page 249
• “Teaching part 3: Updating the other profiles based on
AssayMAP 1” on page 255
• “Adjusting the gripper offset for AssayMAP profiles” on
page 259
• “Manually updating teachpoints in each AssayMAP profile” on
page 260
• “Adjusting the teachpoint for the Bravo Plate Riser” on
page 261
• “Adjusting the Peltier Thermal Station for the red PCR plate
insert” on page 264
• “Verifying the setup in the Protein Sample Prep Workbench” on
page 266
• “Opening the Workbench and accessing the utilities” on
page 266
• “Requirements for partial- head liquid handling” on page 269
199
D
AssayMAP Bravo Platform overview
The following figure and table show the components of the AssayMAP Bravo
Platform, which is required for running the AssayMAP protocols.
6 Risers, 146 mm
The Bravo Platform must include the gripper assembly, which is required for
removing the mounted AssayMAP Bravo cartridges and disposable tips from
the Bravo 96AM Head. For a detailed description of the Bravo Platform and
gripper assembly, see “Hardware overview” on page 5.
CAUTION Using a model of the Tip Wash Station other than the 96AM Wash Station can
result in a potential crash with the Bravo 96AM Head. Ensure that you use only the 96AM
Wash Station.
The Pump Module supplies liquid to the 96AM Wash Station for washing
operations, and it removes the waste liquid during washing operations. The
inline filter is required to help prevent any tubing particulates from potentially
clogging the chimneys of the 96AM Wash Station.
Software overview
The Protein Sample Prep Workbench is an interface that provides a collection
of AssayMAP workflows, protocols, and forms for running protein sample prep
applications on the AssayMAP Bravo Platform. The Protein Sample Prep
Workbench includes the following:
• Workflow Library
• Application Library
• Utility Library
• Literature Library
The VWorks software is required for the Protein Sample Prep Workbench. You
set up the AssayMAP Bravo Platform using the VWorks software.
Related information
Cartridge & Tip Seating Station “About seating the cartridges on the
probes” on page 208
Figure Bravo 96AM Head, syringe, and detail of probe tip (cross-section)
• Mechanism for removing mounted cartridges and tips. The head has a
stripper plate, which is a spring- loaded mechanism that is actuated by the
Bravo gripper assembly. The stripper plate enables the Bravo 96AM Head
to remove the cartridges while still holding liquid in the syringes.
Note: The w- axis is the motion of the displacement of the pipettor
inside the head. The w- axis is not engaged during cartridge mounting or
removal from the head so that fluid can be held in the syringes and
probes. For the Tips On and Tips Off tasks, the Bravo w- axis goes to
the zero position to empty any fluid contained in the syringes.
Figure Stripper plate on Bravo 96AM Head (front-side view and side view)
Liquid-handling modes
The Bravo 96AM Head features include the ability to:
• Transfer liquids directly with the bare probes, through AssayMAP Bravo
cartridges mounted on the probes, and using mounted 250- µL large-
transfer (LT) tips
• Mount or remove the cartridges while holding liquid in the syringes
• Provide precise control of flow rates and volume required for moving fluid
through the packed resin in the cartridges
• Access 96- well and 384- well microplates
• Perform 96- channel liquid handling or partial- head liquid handling using a
subset of the 96 channels
Protocols for the Bravo 96AM Head can employ a combination of the following
liquid- handling modes:
• Bare probes
• AssayMAP Bravo cartridges mounted on the probes
• 250- µL LT tips mounted on the probes
Capabilities of each liquid-handling mode
The following figure shows a side- by- side comparison of a bare probe, a probe
with a mounted cartridge, and a probe with a mounted 250- µL LT tip. The
following table provides the details for each.
Bravo Platform User Guide 203
D
Bravo 96AM Head overview
Figure Comparison of probe, probe with cartridge, and probe with 250-µL LT tip
1 2 3
CAUTION Extended use of the Bravo 96AM Head with dry syringes for disposable-tip
liquid handling can shorten the life of the syringe seals.
Probe
Syringe
plunger
Cartridge
Inlet seal
5 µL bed
Related information
Description
The AssayMAP cartridges are microchromatography cartridges for small- scale
sample preparation. Each cartridge features a 5- µL bed that is packed with
chromatographic resin to support a variety of applications. Liquid can be
aspirated or dispensed through the packed resin in the cartridges. The
AssayMAP Bravo Platform provides the precision flow- rate control required for
aspirating or dispensing the fluid through the packed resin in the cartridges.
Cartridge anatomy
The following figure shows the parts of the cartridge.
Figure Cross-sectional side view of cartridge (left) and cartridge mounted on probe (right)
2
3
4
2 Inlet seal The tapered seal at the top of the packed resin bed.
When the probe is inserted into the cartridge, the tip of
the probe engages the inlet seal. This configuration
enables the Bravo 96AM Head probe to make or break a
fluid- tight connection between the packed- resin bed and
the liquid inside the syringe.
4 Bed Frits at the top and bottom of the resin bed provide a
supports, membrane that retains the resin beads within the bed
upper and while allowing a bidirectional flow of fluid to pass
lower through the cartridge.
Figure AssayMAP cartridge rack lid (1), rack with cartridges (2), and receiver plate (3)
The lid rests directly on top of the cartridges and provides protection during
shipping and storage. The lid can also prevent cartridges from being pushed
out when the cartridge rack is inserted into a receiver plate.
In the VWorks software, the 96AM Cartridge Rack and Receiver Plate is
defined as labware. This labware definition is installed when you install the
provided AssayMAP protocols and forms.
Related information
Workflow
3 Install the Bravo risers to elevate the Agilent Automation Solutions Technical Support
height of the deck.
Ensure the y- axis shipping stop is
removed from the Bravo head mount.
4 Connect the Bravo Platform power cord, G5562A, G5563A Bravo Platform Safety and
communications cable, and pendant cable. Installation Guide
5 Install the Bravo 96AM Head. “Installing and using the Bravo 96AM Head” on
page 211
IMPORTANT You will use the AssayMAP- specific instructions in step 10 to configure the
accessories in the software.
• 96AM Wash Station at deck “Installing the 96AM Wash Station” on page 219
location 1.
9 Set up the software and configure the “Setting up the AssayMAP Bravo Platform” on
device files. page 225
10 Set the AssayMAP head teachpoints in “Teaching part 1: Setting teachpoints in the
every one of the AssayMAP profiles. AssayMAP 1 profile” on page 244
Verify the gripper offset in every profile. “Teaching part 2: Adjusting AssayMAP
teachpoints for accessories” on page 249
“Teaching part 3: Updating the other profiles
based on AssayMAP 1” on page 255
“Adjusting the gripper offset for AssayMAP
profiles” on page 259
11 Install the Bravo shields and Standard G5562A, G5563A Bravo Platform Safety and
Light Curtain. Installation Guide
12 Verify the installation and setup. “Verifying the setup in the Protein Sample Prep
Workbench” on page 266
WARNING The probes of the Bravo 96AM Head are sharp and can scratch you if they
brush across your hand. A probe scratch can expose you to any contaminants remaining on
the probes. Be careful to avoid touching the probes.
CAUTION Extended use of the Bravo 96AM Head with dry syringes for disposable-tip
liquid handling can shorten the life of the syringe seals.
CAUTION To prevent potential damage, ensure the Bravo 96AM Head probes, tips, or
cartridges do not touch the bottom of the 96AM Wash Station. To adjust the distance from the
bottom, you can use the Bravo liquid-handling task parameters in the VWorks software.
CAUTION Before leaving the Bravo 96AM Head idle overnight, wash the syringes
thoroughly with deionized water to prevent salt buildup that could damage the syringe seals.
IMPORTANT To aspirate more than 140 µL of liquid, use the bare probes
instead of disposable tips. The syringes can hold up to 250 µL. If you aspirate
more than 140 µL using 250- µL tips, the excess aspirated liquid will enter the
syringes. Ensure that you wash the syringes as appropriate for your protocol to
prevent contamination.
For information about using a subset of the 96 channels, see “Requirements for
partial- head liquid handling” on page 269.
CAUTION If the Bravo 96AM Head tip box stripper pins are extended, the pins will crash
into the wash station during syringe washing. To prevent potential equipment damage, ensure
that the pins are always retracted during use.
Figure Bravo 96AM Head (side view) with tip box stripper pins retracted and extended
CAUTION Placing the probes of the Bravo 96AM Head (bottom of head) on any surface
can damage the probes. To protect the probes when the head is not mounted on the Bravo
Platform, place the top of the head on a clean, stable surface so that the probes are facing
upward. For storage, use the storage stand.
3 On the side of the Bravo Platform, press the power switch to the off (o)
position.
4 Gently, move the Bravo head manually to position it over deck location 6,
which will provide easier access for removing the head.
Figure Installed Bravo 96AM Head: (1) retainer pin and (2) head lock
6 Reaching from the left side of the Bravo tie bar, grasp the head firmly as
follows:
a With your left hand firmly grip the left side of the head.
b Place your right hand under the flat area just to the left of the syringe
array to support the weight of the head.
7 While supporting the head with your hands, slide the head to the left and
out of the Bravo head mount.
CAUTION Support the head securely without touching the probes. Dropping the head
or bumping the probes will damage the syringes.
8 Carefully place the head upside down on a clean, stable surface so that the
probes are facing up.
9 Carefully lift the stand onto the head, guiding the stand’s side cutouts (1)
onto the head side tabs (2). Use care to avoid touching the probes.
CAUTION Placing the probes of the Bravo 96AM Head (bottom of head) on any surface
can damage the probes. To protect the probes when the head is not mounted on the Bravo
Platform, place the top of the head on a clean, stable surface so that the probes are facing
upward. For storage, use the storage stand.
2 With your left hand, firmly grip the left side of the head. Place your right
hand under the flat area just to the left of the syringe array to support the
weight of the head.
3 While supporting the head with your hands, slide the head onto the Bravo
head mount. Press the head firmly into place to ensure the head is plugged
into the connector receptacle on the head mount. You should hear the click
when the retaining pins snap into place.
Figure Installing the Bravo 96AM Head in the Bravo head mount
Note: If you do not hear the pins snap into place, check that the straight
edges of the retainer pins are in the vertical position, as the following
figure shows. Attempt to rotate the pins to ensure that they are in the
locked position. The pins should not rotate freely.
Figure Installed Bravo 96AM Head: (1) retainer pin and (2) head lock
4 To lock the head, rotate the head lock clockwise until it reaches its hard
stop. This ensures that the head is fully seated and does not shift position
during operation.
CAUTION Dropping the head or bumping the probes will damage the syringes. If the
Bravo head is not properly secured in place, it could drop unexpectedly. Ensure that the
head is securely locked into the head mount.
Related information
Workflow
CAUTION The AssayMAP Bravo Platform requires precise positioning of the 96AM Wash
Station relative to the pipette tips, cartridges, and probes. Ensure that the teachpoint is
correct and that the 96AM Wash Station is secured to the platepad to prevent potential
collisions.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
2 On the platepad side that is next to the crosshairs, using a hex driver
uninstall the two 2- mm screws and remove the bracket.
Note: The wash station is easier to install if you remove the outer
bracket, and then replace the bracket after the wash station is in
position.
Figure 96AM Wash Station fill line (A) and empty line (B)
Source Waste
B
To ensure proper chimney filling and waste emptying characteristics for the
96AM Wash Station, ensure that you use the proper tubing combination for the
Pump Module. See the following figure and table.
6 7
4 4 4 6 8
4 5 3 3 5 4 6 4
8
6
1 2
2 Waste bottle Collects the waste liquid that is pumped from the reservoir.
3 Tubing, 8- mm The length of tubing in the lower pump head, which is used for emptying
(5/16- in) the wash station. The larger diameter tubing ensures that the fluid is
removed efficiently from the wash station
4 Tubing, 6.4- mm The fill- line tubing from the source bottle to the wash station, and
(1/4- in) empty- line tubing connections on either side of the 5/16- in diameter
tubing.
Note: The 5/16- in diameter tubing is too large for the quick- disconnect
fittings and the wash station ports. So, you use a union connector (5) to
splice the 5/16- in tubing to the 1/4- in tubing.
7 Inline pump Removes the particulates that can clog the chimneys in the wash station.
filter
8 3- way Enables one tube to branch into two tubes at the input and output ports
connector on the wash station.
For detailed Pump Module connection instructions, see the Pump Module User
Guide.
Related information
Procedure
IMPORTANT If the VWorks software has not yet been configured for your
instrument, you must configure the software before operating the AssayMAP
Bravo Platform.
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the AssayMAP Bravo Platform should perform the setup procedures.
Software requirements
For the computer requirements, see the software release notes or the VWorks
Knowledge Base at www.agilent.com/chem/askb.
The AssayMAP Bravo Platform requires compatible versions of the VWorks
software and the Protein Sample Prep Workbench. See the software release
notes for compatibility requirements and software installation instructions.
Setup workflow
You must set up the AssayMAP Bravo Platform in the VWorks software before
you run the protocols in the Protein Sample Prep Workbench. Make sure that
you perform the following steps in the order given.
CAUTION Using a profile that is associated with teachpoints from a different AssayMAP
Bravo Platform or different Bravo 96AM Head can result in a hardware collision and cause
equipment damage. Ensure that you verify the profiles and teachpoints for your specific
instrument and Bravo 96AM Head.
CAUTION Using the wrong device profile or an improperly created profile can result in
equipment damage. If you are uncertain about which device file or profile to use, see your lab
administrator.
Each AssayMAP application requires a specific Bravo device file and profile.
The device profile specifies the head type (Bravo 96AM Head), Ethernet
communication settings for the device, the configuration of the platepads and
accessories on the deck, and the associated teachpoints for the specific
combination of instrument and Bravo 96AM Head.
Each device profile is associated with a device file. The following table lists the
device files and associated profiles installed with the Protein Sample Prep
Workbench v3.0.
GlykoPrep- Plus_AM Bravo.dev GlykoPrep- Plus profile N- Glycan Sample Prep applications
InSolution Digestion- AM Single Plate InSolution In- Solution Digestion: Single Plate
Bravo.dev Digestion
The device profile must specify the correct deck configuration and the
associated teachpoints for the Bravo deck layout.
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the AssayMAP Bravo Platform should perform the setup procedures.
3 In the VWorks window, open the corresponding AssayMAP device file for
the profile that you are verifying.
To open a device file, choose File > Open. In the Open dialog box, ensure the
file type is Device Files (*.dev), select the file, and then click Open.
The device files are installed in the following default location:
C:\VWorks Workspace\Workbench\Device Files
4 Click No when the Would you like to initialize devices now? message appears.
If necessary, select the correct name from the Profile list, and then click the
Save icon.
6 In the <name>.dev tab, click Device diagnostics. The Bravo Diagnostics dialog
opens.
7 In the Profiles tab, verify the Profile name.
8 Verify the following Head Information properties.
Property Setting
Head type 96AM
Teaching tip type 250 µL
Check head type on initialize checked
9 In the Connection area, select This Bravo is connected via ethernet, and then click
Find available device to select the device to associate with the profile.
In the Discovered BioNet Devices dialog box that appears, perform the
following tasks:
a b c
a Click the Select the Ethernet adapter to use from the list below, and select the
correct adapter for the device connection.
b In the list of devices that appear, select the Bravo device. If multiple
Bravo devices are on the network, use the MAC Address to identify the
Bravo device for this profile. To successfully communicate with the
Bravo device, the Status column must display New or Matched.
c Click OK.
10 In the Miscellaneous area, verify the property settings. The following figure
shows the settings for each profile.
11 If you have made any changes, click Update this profile.
Figure Miscellaneous area of the Profiles tab showing variations in property settings for different AssayMAP profiles
GlykoPrep- Plus By default, the GlykoPrep- Plus profile • “Verifying the AssayMAP
profile includes the following configurations: Autofill Station
• Autofill Station (96AM Wash Station) configuration” on page 234
• Heating Shaking Station for the Peltier • “Verifying the AssayMAP
Thermal Station (location 4) Peltier Thermal Station
configuration” on page 236
• Orbital Shaking Station (location 9)
• “Verifying the AssayMAP
If these accessories are not installed on the Orbital Shaking Station
AssayMAP Bravo deck, reconfigure the configuration” on page 238
corresponding deck locations for a
standard platepad.
AssayMAP Peptide By default, the AssayMAP Peptide Sample • “Verifying the AssayMAP
Sample Prep profile Prep profile includes the following Autofill Station
configurations: configuration” on page 234
• Autofill Station (96AM Wash Station) • “Verifying the AssayMAP
• Heating Shaking Station for the Peltier Peltier Thermal Station
Thermal Station (location 4) configuration” on page 236
• Orbital Shaking Station (location 9) • “Verifying the AssayMAP
Orbital Shaking Station
If these accessories are not installed on the configuration” on page 238
AssayMAP Bravo deck, reconfigure the
corresponding deck locations for a
standard platepad.
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the AssayMAP Bravo Platform should perform the setup procedures.
6 Verify the following property settings in the Autofill Station properties table,
and then click Next.
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the AssayMAP Bravo Platform should perform the setup procedures.
3 To verify the settings, right- click deck location 4 in the image. In the
shortcut menu that appears, click Configure location > Accessory. The Bravo
Accessories Wizard opens.
4 In the Location for accessory list, verify the location and then click Next.
5 In the Accessory already configured page, click Yes, reconfigure existing accessory,
and then click Next.
Properties Setting
MTCSTC Id 0
Type STC
Slot Id 1
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the AssayMAP Bravo Platform should perform the setup procedures.
3 To verify the settings, right- click deck location 9 in the image. In the
shortcut menu that appears, click Configure location > Accessory. The Bravo
Accessories Wizard opens.
4 In the Location for accessory list, verify the location and then click Next.
5 In the Accessory already configured page, click Yes, reconfigure existing accessory,
and then click Next.
Property Setting
Serial Port 1*
Module Number 1
*The Serial Port parameter will be automatically defined when a USB- to-
serial adapter is connected to the computer. You might need to lookup the
port assigned to this device in the Windows Device Manager.
Note: Occasionally, a port number is reassigned for a USB- to- serial
connection if you disconnect and reconnect the accessory. If this
happens, you must verify that the correct Serial Port is set in each
profile.
7 Click Finish. The Bravo Accessories Wizard closes.
8 If the accessory configuration message appears, click Yes to move the
teachpoint to a safe height.
9 If you made any changes to the configuration settings, click the Profiles tab,
and then click Update this profile to save the changes.
Make sure that you test the accessory in each profile where it is configured.
For instructions, see “Testing the Orbital Shaking Station” on page 115.
Related information
00306
3 In the Configuration tab, click the Autofill Station icon, and then click Diagnose
accessory. The Bravo Accessory Diagnostics dialog box opens.
3 In the Bravo Accessory Diagnostics dialog box, verify that the correct device
(CPAC Ultraflat) appears in the display at the top.
4 In the Heat area, set the parameters to test the temperature function:
a In the Target temp box, type the temperature (°C).
For example, set the CPAC at deck location 4 to 45 °C.
b Click Start.
c At the Peltier Thermal Station on the deck, verify that the accessory
operates correctly.
You can also monitor the readout in the Current temperature display.
d Click Turn Off to stop the heating or cooling, and click OK to close the
dialog box.
2 Type the RPM, select the Stir direction, and then click Start.
3 At the Orbital Shaking Station on the deck, verify that the accessory
operates correctly.
4 In the Bravo Accessory Diagnostics dialog box, click Stop, and then click OK.
IMPORTANT Ensure that you verify the profile settings and test the
accessories in each of the other profiles before you set the teachpoints.
Related information
Teachpoint requirements
A teachpoint is a set of axial coordinates that define a location to which the
Bravo head moves. Each of the nine deck locations has a teachpoint.
CAUTION Using a profile that is associated with teachpoints from a different instrument
can result in a hardware collision and cause equipment damage. Ensure that you verify the
teachpoints associated with each profile for your specific instrument and Bravo 96AM Head.
IMPORTANT Every device profile in the Protein Sample Prep Workbench must
be adjusted to the correct teachpoints for the specific AssayMAP Bravo
Platform that you are installing.
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should perform the procedures in this topic.
WARNING The red Stop motors button in Bravo Diagnostics does not perform an
immediate stop. The Bravo head can continue to move in the same direction at the same
speed after you click the button. To perform an emergency stop, press the red button on the
emergency-stop pendant.
CAUTION To prevent the head from crashing into an accessory, be sure to start the
accessory teaching process from the safe z-axis point above the accessory location. This is
especially important if you use the Set all teachpoints based upon selected teachpoint
feature.
IMPORTANT Make sure that you remove all labware from the deck except for
the 96AM Wash Station at deck location 1. The wash station remains installed
at deck location 1 in the following procedure while you set the teachpoint for
deck location 2. In a subsequent procedure, you will adjust the wash station
teachpoint at deck location 1.
IMPORTANT Ensure that the pipette tip is new and in good condition.
The following procedures assume that you have basic knowledge of the Bravo
teaching process and have been trained to teach the Bravo Platform.
CAUTION To prevent potential equipment damage, ensure that the deck is clear of any
obstacles before initializing the AssayMAP Bravo Platform.
2 Set the teachpoint for deck a. “Setting the teachpoint for deck
location 2, derive the location 2 and deriving the
remaining teachpoints for the remaining teachpoints” on
AssayMAP 1 profile, and fine page 246
tune the teachpoints. b. “Fine tuning the teachpoints” on
page 247
Setting the teachpoint for deck location 2 and deriving the remaining teachpoints
To set the teachpoints:
1 Manually seat a 250- µL pipette tip on probe A1 (back left corner of the
Bravo 96AM Head). Ensure that the pipette tip locks into place with the
probe.
Ensure a standard platepad is installed at deck location 2.
2 In the Bravo Diagnostics dialog box, click the Jog/Teach tab.
3 In the image of the Bravo deck in the upper left corner, click Location 2.
4 Carefully, use the Jog Axes controls to adjust the position of the pipette tip
above the crosshairs on platepad 2. Use incrementally smaller jog
increments as you jog the pipette tip down closer to the crosshairs.
As the pipette tip moves down along the z- axis towards the crosshairs, the
following error message appears. Visually inspect the clearance remaining
between the tip and the crosshairs. If the tip is not yet at the crosshairs,
adjust the jog increment as required, and click Ignore to continue jogging
the head down.
CAUTION Before you click Ignore in the error message, verify that enough space
remains to continue jogging the head down. Jogging the head down too far will result in
a collision and cause equipment damage.
Note: This error message appears every time you jog in the z- axis to a
point that is lower than the default z- axis teachpoint (60.00 mm).
5 When the pipette tip is 1–2 mm above the crosshairs, do the following:
a Verify that the A1 pipette tip is perfectly centered above the platepad
crosshairs in the x- and y- axes.
b Set the Jog Axes Up/Down increment to 0.2- mm. Carefully, jog the head
down in 0.2- mm increments just until the feeler gauge cannot fit under
the pipette tip.
c Raise the head by 0.05- mm increments until you reach the lowest
possible height that allows the 0.010- inch feeler gauge to slide under
the pipette tip without deflecting it.
6 Click Teach.
7 To derive the teachpoints for the remaining locations from deck location 2:
a In the Advanced operations list, select Set all teachpoints based upon the
selected teachpoint.
b Ensure that deck location 2 is still selected, and then click the Perform
advanced operations.
c Click OK in the confirmation message.
3 Carefully, use the Jog Axes controls to adjust the position of the A1 pipette
tip above the crosshairs. Use incrementally smaller jog increments as you
jog the tip down closer to the crosshairs.
4 When the pipette tip is 1–2 mm above the crosshairs, do the following:
a Verify that the A1 pipette tip is perfectly centered above the crosshairs
in the x- and y- axes.
b Set the Jog Axes Up/Down increment to 0.2- mm. Carefully, jog the head
down in 0.2- mm increments just until the feeler gauge cannot fit under
the pipette tip.
c Raise the head by 0.05- mm increments until you reach the lowest
possible height that allows the 0.010- inch feeler gauge to slide under
the pipette tip without deflecting it.
5 Click Teach.
6 Repeat steps 1 to 5 for deck locations 4, 5, 6, 7, and 8.
• 250- µL pipette tip seated on probe A1 (back left corner) of the Bravo
96AM Head.
IMPORTANT Ensure that the pipette tip is new and in good condition.
3 In the Approach height above teachpoint box, type 15, and then click Approach.
The Bravo head moves to deck location 9 and moves down so that the
pipette tip is approximately 5 mm above the crosshairs on the 10- mm
teaching plate.
4 Carefully, use the Jog Axes Down +Z control to move the pipette tips down
until the pipette tip is 1–2 mm from the crosshairs, and then use the Jog
Axes X/Y controls to center the A1 tip above the crosshairs.
5 When the A1 tip is perfectly centered above the crosshairs, click Teach to
save this value temporarily.
6 Click the Move to safe height at current X/Y position. Remove the teaching plate
from the Orbital Shaking Station.
7 Ensure that deck location 9 is still selected, and then click Move. The head
moves down so that the tip is approximately 10 mm above the top surface
of the plate nest in the Orbital Shaking Station.
8 Use the Down +Z control to lower the pipette head incrementally.
9 When the pipette tip is 1–2 mm above the flat surface of the Orbital
Shaking Station, do the following:
a Set the Down +Z increment to 0.2 mm. Carefully, jog the head down in
0.2- mm increments just until a 0.254- mm (0.010- inch) feeler gauge
cannot fit under the pipette tip.
b Raise the head by 0.05- mm increments until you reach the lowest
possible height that allows the feeler gauge to slide under the pipette
tip without deflecting it.
Note: Alternatively, you may use the Jog Axes controls to move the head
down in the z- axis until the pipette tip is as close as possible to the
top surface of the plate nest, without making contact.
10 Click Teach to set the new teachpoint values for deck location 9.
11 To save the new teachpoint for this location, click the Profiles tab, and then
click Update this profile.
CAUTION Failure to remove the pipette tip from the A1 probe will cause the pipette tip to
collide with the wash station chimney, damaging the Bravo 96AM Head and 96AM Wash
Station. Ensure that you remove the pipette tip before starting the procedure.
For this procedure, you will center the syringe probes above the wash station
chimneys to adjust the x- and y- axes for the teachpoint, but you will use the
existing z- axis teachpoint.
4 In the Approach height above teachpoint box, type 50, and then click Approach.
5 Carefully, use the Jog Axes controls to position the syringe probes so that
they are perfectly centered over the corresponding chimneys in the wash
station.
The tips of the probes should be approximately 1 mm above the tops of the
chimneys.
6 When the probes and chimneys are perfectly aligned, use the Jog Axes
controls to move the head down slowly along the z- axis to the original
Z-axis teachpoint value that you recorded in step 3.
Note: This step will insert the probes into the wash station chimneys,
which is a normal part of the teaching process.
7 When the position of the z- axis (as indicated by the Z- axis readout in the
bottom right of the screen) matches the value that was written down from
above, click Teach to set the new teachpoint for deck location 1.
8 Click the Move to safe height at current X/Y position to remove the probes from
the wash station chimneys, and return the head to the safe Z- height.
9 In the Profiles tab, click the Update this profile to save the new teachpoints.
Adjusting the teachpoint at deck location 2 for the Cartridge & Tip Seating Station
You adjust the teachpoint for deck location 2 to enhance the precision of the
probes as they move into the cartridges or pipette tips in the Cartridge & Tip
Seating Station.
To adjust the teachpoint for the Cartridge & Tip Seating Station:
1 Ensure that no labware is on the deck other than the wash station at deck
location 1.
2 In the Bravo Diagnostics Jog/Teach tab, click Location 2.
3 Write down the value that appears in the Z-axis box for location 2. You will
use this value in a later step.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
6 Carefully, use the Jog Axes controls to jog the head down incrementally to
the Cartridge & Tip Seating Station.
7 When the syringe probes are just above the wells, use the Jog Axes controls
for the x- and y- axes to position the probes so that they are perfectly
centered within the openings of the corresponding wells in the Cartridge &
Tip Seating Station.
8 When the probes are perfectly centered within the well openings:
a Click Teach to save these values temporarily.
b Click Move to safe height at current X/Y position to return the head to a safe
z- axis height.
9 Remove the Cartridge & Tip Seating Station from the deck.
10 Carefully, use the Down +Z control to jog the head down incrementally until
the value in the Z-axis box matches the value that you noted in step 3.
11 Click Teach to set the new teachpoint.
12 In the Profiles tab, click Update this profile.
Related information
IMPORTANT Ensure that the pipette tip is new and in good condition.
The AssayMAP Bravo.dev file opens, and the AssayMAP 1 profile initializes.
4 Ensure that the AssayMAP Teachpoint Update protocol is open in the
VWorks window, and then click Start.
5 In the Run Configuration dialog box, click Finish to start the protocol.
6 When the Update Teachpoints dialog box appears, click Continue update
teachpoints in all the other workbench profiles based on the AssayMAP 1
profile.
7 When the WARNING dialog box appears, click Continue to update the
teachpoints in all the Workbench profiles based on the AssayMAP 1 profile
teachpoints.
Related information
IMPORTANT After you have finished setting the teachpoints for all the
profiles in the AssayMAP Bravo Platform, use the procedures in this topic to
adjust the gripper y- axis offset and test the gripper performance in every
profile.
IMPORTANT Before you adjust the gripper y- axis offset, ensure that the
pipettor teachpoints (x- , y- , and z- axes) have been verified and no pipette tips
or cartridges are installed on the head.
Workflow
3 Test the gripper pick- and- place “Checking the gripper pick- and-
operations. place operations” on page 47
6 Test the gripper pick- and- place “Checking the gripper pick- and-
operations. place operations” on page 47
Related information
CAUTION Using a profile that is associated with teachpoints from a different instrument
can result in a hardware collision and cause equipment damage. Ensure that you verify the
teachpoints associated with each profile for your specific instrument and Bravo 96AM Head.
2 In the AssayMAP 1 profile, set the teachpoint for “Setting the teachpoint for deck
deck location 2 and derive the teachpoints for the location 2 and deriving the
other eight deck locations automatically. remaining teachpoints” on
page 246
5 GlykoPrep- Plus profile only. If the Peltier Thermal “Adjusting the Peltier Thermal
Station is installed at deck location 4, adjust the Station for the red PCR plate
teachpoint to accommodate the red PCR plate insert. insert” on page 264
6 Orbital Shaking Station only. If the accessory is “Teaching the Orbital Shaking
installed at deck location 9, fine tune the teachpoint. Station at deck location 9” on
page 249
7 Adjust the teachpoint for the 96AM Wash Station. “Teaching the 96AM Wash Station
at deck location 1” on page 250
CAUTION Placing the Bravo Plate Riser at a deck location that does not have the
corresponding teachpoint in the device profile will result in a collision, causing equipment
damage. Ensure that the device profile includes the corresponding teachpoint for any deck
location where the Bravo Plate Riser is installed.
1 00240
2 3
Bravo
Plate Orientation
4 5 6
7 8 9
00240
IMPORTANT When setting the teachpoint, ensure that a 250- µL pipette tip is
seated on probe A1 (back left corner) of the Bravo 96AM Head.
CAUTION A hardware collision can occur if you initialize a device profile that does not
include the teachpoints for the Bravo Plate Riser while the Bravo Plate Riser remains on the
deck. Ensure that you remove the Bravo Plate Riser from the deck when initializing a different
device profile.
Related information
Adjusting the teachpoint for the “Adjusting the Peltier Thermal Station
red PCR plate insert for the red PCR plate insert” on
page 264
Adjusting the Peltier Thermal Station for the red PCR plate insert
Use the following procedure only if the deck configuration includes the Peltier
Thermal Station and you plan to use the red PCR plate insert.
The AssayMAP GlykoPrep- plus applications require the red PCR plate insert to
be present in the plate nest of the Peltier Thermal Station for the entire
workflow. Use the following teaching procedure to accommodate this
requirement for deck location 4 in the GlykoPrep- plus profile.
IMPORTANT Ensure that the PCR plate insert is not installed in the Peltier
Thermal Station at deck location 4 for the following teaching procedure.
2 Click Move to move the head to the previously defined teachpoint for
location 4.
3 Using the Jog Axes controls, jog the head up (- Z) by exactly 4.4 mm.
For example, if the value for the teachpoint’s Z- axis coordinate from the
fine tuning was 104.00 mm, the new value after adjusting by 4.4 mm will
be 99.60 mm.
4 Click Teach to set the new teachpoint.
5 To save the new teachpoint for this location, click the Profiles tab, and then
click Update this profile.
Related information
Workflow
3 Run the Syringe Wash utility Instructions for the utility in the
for 60 cycles. Protein Sample Prep Workbench
5 Run the Syringe Test utility. Instructions for the utility in the
Protein Sample Prep Workbench
Alternatively, on the Windows desktop, click Start > All Programs > Agilent
Technologies > Protein Sample Prep Workbench> Protein Sample Prep Workbench.
To access a utility:
1 In the Protein Sample Prep Workbench, do one of the following:
• If you have not yet logged in, click Utility Library. The VWorks software
starts, and then the User Authentication dialog box opens.
Type your VWorks user name and password, and then click OK.
Note: The default login user name is administrator.
The default password is administrator.
• If you have already logged in to the software, click the Utility Library
button in any of the other libraries of the Protein Sample Prep
Workbench.
2 In the Utility Library, locate the name of the utility you want to use, and
then click Utility to open the utility.
3 Click Instructions to display the user guide for the corresponding utility.
Related information
CAUTION Attempting to use the bare probes for liquid-handling tasks that employ fewer
than 96 channels can cause a crash, resulting in damage to the Bravo 96AM Head.
CAUTION Attempting to perform a task at the 96AM Wash Station while a portion of the
Bravo 96AM Head is offset above an adjacent location may result in a crash, causing potential
damage to the head. Ensure that you use the entire plate well selection for any task at the
96AM Wash Station.
CAUTION Using a single cartridge or an irregular array of cartridges can cause a potential
head crash. Ensure that the selection of cartridges originates at probe A1 and that the
selection of probes is contiguous. Carefully plan the deck layout and robot moves to prevent
potential collisions between the remaining bare probes and the items on the Bravo deck.
When fewer than 96 cartridges are mounted on the head, careful planning of
the deck layout and the robot moves is required to prevent potential collisions
between the remaining bare probes and the items on the Bravo deck. A
mounted cartridge extends only 9 mm beyond the tip of a bare probe, as the
following figure shows.
Figure Vertical distance between the tips of a bare probe and a mounted cartridge
9 mm
Because the distance between the tip of the bare probe and a mounted
cartridge is relatively small, you must ensure the following:
• The depth at which the cartridges enter the wells is no more than 9 mm.
• When doing partial- plate operations where the head is offset horizontally
from the target deck location (not centered over the platepad), ensure
sufficient vertical clearance for the unused probes:
– Target deck location. If the labware has deep wells and the head goes
down too far, a collision between the unused probes and the perimeter
of the labware can occur.
– Adjacent deck location. If a labware or other object at the adjacent
location is taller than the labware at the target location, a collision
between the unused probes and the tall object or labware can occur.
Note: No horizontal offset of the head is required for partial- plate operations
where the selection of channels in the head aligns with the labware well
selection. For example, probe A1 enters well A1.
The following figures show examples of partial- plate operations where the head
is offset horizontally from the target deck location. In this example, cartridges
are mounted on the probes in columns 1 and 2. To access the wells in columns
11 and 12 of the labware at deck location 7, the head is offset over deck
location 8.
CAUTION If the head is offset from deep-well labware at the target deck location, the
unused probes can collide with the labware perimeter if the head goes down too far.
If the labware at location 7 was a deep- well microplate and you tried to
aspirate from the bottom of the wells in columns 11 and 12, the bare probes in
column 3 of the head would crash into the right side wall of the deep- well
microplate.
CAUTION If the head has fewer than 96 cartridges and is offset from the target deck
location, the unused probes can have undesired contact with surrounding objects or samples.
To prevent equipment damage and contamination, verify that sufficient clearance exists at the
adjacent deck location.
Figure Example of partial-head horizontal offset with taller labware at the adjacent location
Although the VWorks software considers the geometry of the labware at the
target deck location during the task, the software is unaware of the labware
height on the adjacent deck location where the head will be offset.
CAUTION A collision will result if the Bravo 96AM Head is offset from the tip box location
when pressing on tips. When picking up tips from a tip box, ensure that the Bravo 96AM Head
is centered over the tip box.
CAUTION To ensure consistent seating of tips on the probes, use the 96AM Cartridge &
Tip Seating Station to seat the tips on the probes.
The Bravo 96AM Head must be centered over the tip box when picking up tips
to be transfered to the 96AM Cartridge & Tip Seating Station.
Related information
Using and storing the Bravo 96AM “Installing and using the Bravo 96AM
Head Head” on page 211
E
Diagnostics quick reference
This appendix contains the following topics:
• “Diagnostics dialog box” on page 276
• “Configuration tab quick reference” on page 277
• “Accessory Diagnostics dialog box” on page 280
• “External Robots tab quick reference” on page 281
• “IO tab quick reference” on page 282
• “Jog/Teach tab quick reference” on page 284
• “Gripper tab quick reference” on page 288
• “Profiles tab quick reference” on page 291
• “Processes tab quick reference” on page 295
• “Aspirate task parameters” on page 299
• “Dispense task parameters” on page 302
• “Dispense to Waste task parameters” on page 304
• “Mix task parameters” on page 307
• “Pump Reagent parameters” on page 310
• “Shake task parameters” on page 311
• “Vacuum Filtration Station task parameters” on page 313
• “Wash Tips task parameters” on page 316
• “Pin Tool task parameters” on page 320
• “AM Aspirate task parameters” on page 323
• “AM Dispense task parameters” on page 328
• “AM Mix task parameters” on page 333
275
E
Diagnostics dialog box
Buttons
WARNING To perform an emergency stop, press the red button on the emergency-stop
pendant.
Button Description
About Displays the version numbers of Diagnostics and,
if initialized, the Bravo firmware.
Stop motors Disables all the motors that drive the Bravo head
(keyboard SPACEBAR) and engages the brakes to stop the motion.
To re- enable the head movement, click Enable all
motors on the Jog/Teach tab.
Tabbed pages
“IO tab quick reference” Provides controls and indicators for clearing
on page 282 motor faults and checking the head type.
“Jog/Teach tab quick Provides controls for jogging the liquid- handling
reference” on page 284 head and setting teachpoints.
“Gripper tab quick Provides controls for fine- tuning the gripper
reference” on page 288 movements.
Procedures
The Configuration tab enables you to configure and test accessories. For the
procedures, see:
• “Accessories and platepads” on page 107
• “Autofilling accessories” on page 177
Contents
The Configuration tab in Diagnostics contains the following areas:
• “Location Configuration area” on page 277
• “Accessory Configuration area” on page 278
Related information
Related information
Procedures
The External Robot tab provides the controls for configuring how an external
robot can access the Bravo deck. You can configure whether to permit only one
robot to operate on the Bravo deck at any given time or to allow concurrent
operation.
For configuration procedures, see “Configuring external robot access” on
page 48.
Contents
The External Robot tab contains the following controls and indicators:
Move Bravo to this safe Allows you to specify the deck location where
location the Bravo head will move if an external robot is
accessing the Bravo deck.
Does an external robot Specifies whether the selected deck location may
access this location? be accessed by an external robot.
Select the locations that Allows you to specify locations on the Bravo
should be blocked during deck where the Bravo robot is prohibited from
robot access moving during the external robot access.
Related information
Contents
The IO tab in Diagnostics contains the following controls and indicators:
Plate present in gripper Available only for a Bravo Platform that has a
gripper. This indicator lights bright green when
the plate presence sensor detects a plate in the
gripper.
Related information
WARNING To avoid potential injury and damage to the device, only personnel trained in
how to teach the Bravo Platform should use the Jog/Teach tab controls.
WARNING The red Stop motors button does not perform an immediate stop. The Bravo
head can continue to move in the same direction at the same speed. To perform an
emergency stop, press the red button on the emergency-stop pendant.
Procedures
The Jog/Teach tab enables you to:
Procedure See…
Home the liquid- handling head “Homing the liquid- handling head” on
page 94
Move the liquid- handling head to a “Using the Move and Approach
safe distance above a teachpoint, or commands” on page 99
move to a teachpoint
Move the liquid- handling head “Jogging the liquid- handling head” on
incrementally in each of its axes page 95
Contents
The Jog/Teach tab contains the following:
• “Teachpoints area” on page 284
• “Jog and Home Axes areas” on page 286
Teachpoints area
The Teachpoints area contains the following controls and indicators.
X- axis, Y- axis, and Z- axis Provide the current teachpoint coordinates for
displays the selected deck location.
Move to a safe height at Moves the liquid- handling head to the z- axis
current X/Y position safe position that is specified in the profile.
Left –X, Right +X, Enables you to select the incremental distance
Back –Y, Forward +Y (mm) to move the liquid- handling head, and
buttons and then move the head the specified distance
increment (mm) list along the x- or y-axis.
Home XYZ Homes the Bravo head in the side- to- side (x),
forward- backward (y), and vertical (z) axes.
Enable all motors/ Activates the Bravo head motors and turns off
Disable all motors the head motors.
For example, you must disable the motors
before moving the head manually.
Related information
WARNING The red Stop motors button does not perform an immediate stop. The Bravo
head can continue to move in the same direction at the same speed. To perform an
emergency stop, press the red button on the emergency-stop pendant.
Procedures
You can use the Gripper tab to fine- tune the gripper movements using the
following procedures:
• “Verifying the gripper setup” on page 43
• “Using the gripper controls” on page 100
Teach Y offset for gripper Calculates the offset from the selected location
teachpoint to where the gripper is and applies
it as the y- offset.
Plate present in gripper Lights when the plate sensor detects labware in
the gripper.
Dock gripper below head Moves the gripper to the docked position to
ensure that there is not interference when
moving the liquid- handling head.
Related information
Procedures
The Profiles tab enables you to:
Procedure See…
Create a profile, or manage the “Creating and managing profiles” on
available profiles page 28
Change the liquid- handling head “Changing the Bravo head” on page 65
Contents
The Profiles tab contains the following:
• “Profile Management area” on page 291
• “Connection area” on page 292
• “Modified Variables area” on page 295
CAUTION Each profile can be used by multiple protocols. Deleting, renaming, or changing
the parameters for a profile based on one protocol can invalidate other protocols that use the
profile.
Control Description
Profiles name list Specifies the active profile.
Select the profile that you want to use from the
list.
Create a new profile Displays the Create Profile dialog box so that
you can name the new profile.
To add a profile, see “Creating and managing
profiles” on page 28.
Control Description
Rename this profile Displays the Rename Profile dialog box so that
you can rename the profile selected in the
Profile name list.
Connection area
The Connection area contains the following controls.
Control Description
This Bravo is connected Specifies an Ethernet communication connection
via ethernet between the device and the computer.
Find available device Opens the Discovered BioNet Devices dialog box,
which lists the connected Bravo devices that the
software found.
Miscellaneous area
The Miscellaneous area contains the following controls.
Control Description
Approach height Specifies the z- axis distance or height (safety
margin) above a microplate that the tips move to
before moving to another quadrant of the
microplate.
Control Description
Z- axis safe position Sets the safe z- axis distance or height of the
liquid- handling head.
A z- axis safe position of 0 is the highest position
possible for the liquid- handling head. A z- axis
safe position of 10 means the liquid- handling
head is 10 mm below the highest point.
If you select the Always move to safe Z before each
process option, the Bravo Platform moves to the
z- axis safe height when moving between labware
locations. Increasing this value may shorten run
times by minimizing the z- axis movement
between tasks. Making this value too large
causes the liquid- handling head to crash into
labware when moving between plate locations.
Always move to safe Z Moves the liquid- handling head to the specified
before each process z- axis safe position between processes.
If you select this option, the Bravo head moves
to the specified z- axis safe height when moving
between labware locations. Select this option if
you have not specified the labware on the deck.
If you do not select this option, the Bravo
Platform will automatically determine the safe
z- axis point based on the labware specified,
thereby optimizing the processing time.
To prevent damage to the labware or the Bravo
Platform components, do not select this option if
you have not specified the labware.
Ignore plate sensor Bravo Platform with gripper only. Enables the
during pick and place software to ignore the microplate sensor during
a pick- and- place movement.
Select this option if the microplate sensor is
broken but you still want to test the pick- and-
place function.
Enable tips- off tip- touch Performs a tip touch in the tip box after a Tips
Off task to ensure that the tips fall from the
head into the tip box.
Control Description
This is a Bravo SRT Bravo SRT model only. Specifies that this Bravo
device is the SRT model, which is shorter than
the standard model.
If you are using 250- µL tips on the SRT model,
select this check box, configure the tip box
location, and install the special SRT platepad for
250- µL tip boxes to ensure greater z- axis
clearance and prevent potential hardware
crashes. See “(Bravo SRT only) Setting up an LT
tip box location” on page 173.
Control Description
Head type Specifies the liquid- handling head type for the
profile.
Teaching tip type Specifies the type of pipette tip to use for setting
the teachpoints.
If you change the Teaching tip type in a profile,
you must reteach the deck locations using the
new tip type.
Check head type on Verifies that the liquid- handling head specified
initialize in the profile matches the head mounted on the
Bravo Platform when the profile is used by a
protocol.
CAUTION If this check box is cleared, the Bravo
Platform will run with any liquid-handling head installed.
Therefore, if the profile specifies a 96-channel head type,
but a 384-channel head is installed, the head will crash.
Note: If this check box is cleared, the w- axis
controls in the Jog/Teach tab will be disabled
even when the device is initialized.
Related information
WARNING The red Stop motors button does not perform an immediate stop. The Bravo
head can continue to move in the same direction at the same speed. To perform an
emergency stop, press the red button on the emergency-stop pendant.
Procedure
To run a process, see “Performing a task using Diagnostics” on page 104.
Contents
The Processes tab contains the following:
• “Location area” on page 295
• “Miscellaneous area” on page 296
• “Well Selection and Head Mode area” on page 297
• “Command Parameters area” on page 298
Location area
Note: If the selected Command to execute is Disassemble Vacuum, the controls
in this area are unavailable. In this case the command controls the location.
Miscellaneous area
To select a quadrant:
Click a well in that quadrant. All the wells in
that quadrant turn green, indicating the selected
wells.
Alternatively, right- click the graphical display to
access the following shortcut menu commands:
• Select all wells
• Clear all selected wells
• Select all wells in highlighted row
• Clear all selected wells in highlighted row
• Select all wells in highlighted column
• Clear all selected wells in highlighted column
Set head mode Opens the Head Mode Selector dialog box.
Series III disposable- tip heads and Bravo 96AM
Head only. To use one row or column of
channels instead of all channels in the head, use
the controls in the dialog box to select which
channels on the head to use.
Parameter Description
Command to execute Provides the following list of tasks that you can
run:
• “Aspirate task parameters” on page 299
• “Dispense task parameters” on page 302
• “Dispense to Waste task parameters” on
page 304
• “Mix task parameters” on page 307
• “Pump Reagent parameters” on page 310
• “Shake task parameters” on page 311
• Tips On and Tips Off
• “Vacuum Filtration Station task parameters”
on page 313 (Assemble Vacuum, Disassemble
Vacuum, and Move and Filter Plate)
• “Wash Tips task parameters” on page 316
• “Pin Tool task parameters” on page 320
• Bravo 96AM Head only
– AM Cartridges On and AM Cartridges Off
– “AM Aspirate task parameters” on
page 323
– “AM Dispense task parameters” on
page 328
– “AM Mix task parameters” on page 333
Other commands may be available depending on
the installed accessories. For a complete
description of task parameters, see the VWorks
Automation Control User Guide.
Parameter description
The Aspirate task has the following parameters.
Parameter Description
Volume (µL) The volume of liquid to be drawn into each
pipette tip.
Pre- aspirate volume (µL) The volume of air to be drawn before the
pipette tips enter the liquid.
Post- aspirate volume (µL) The volume of air to be drawn after the
liquid is drawn.
Distance from well bottom The distance between the end of the pipette
(0–100 mm) tips and the well bottoms during the Aspirate
task.
If you specify dynamic tip extension, this is
the distance at the end of the Aspirate task.
Parameter Description
Dynamic tip extension The rate at which the pipette head moves
(0–20 mm/µL) during the Aspirate task. The software
calculates the distance over which the tips
will move without crashing.
Use dynamic tip extension to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTE > 1/A
• Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume,
the software automatically moves downward
toward the well bottom with each aspirate
action.
Perform tip touch The option to touch the pipette tip on one or
more sides of the well.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the pipette tips to
(–20 to 50 mm) rise before touching the sides of the wells.
Parameter Description
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter descriptions
The Dispense task has the following parameters.
Parameter Description
Empty tips The option to empty all liquid from the tips
instead of using the dispense volume
specification.
Distance from well bottom The distance between the end of the pipette
(0–100 mm) tips and the well bottoms during the
Dispense task.
If you specify dynamic tip retraction, this is
the starting distance.
Parameter Description
Dynamic tip retraction The rate at which to raise the pipette head
(0–20 mm/µL) during the Dispense task.
Use dynamic tip retraction to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip retraction (DTR) as
follows:
DTR = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTR > 1/A
• Slower than the volume change.
DTR < 1/A
The starting and ending positions can be
calculated as follows:
(Vdispensed * DTR) + Distancewell bottom
Perform tip touch The option to touch the pipette tip on one or
more sides of the well.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the pipette tips to
(–20 to 50 mm) rise before touching the sides of the wells.
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter descriptions
The Dispense to Waste task has the following parameters.
Parameter Description
Empty tips The option to empty all the liquid from the
head into the Tip Wash Station outside the
chimneys.
Parameter Description
Override dispense flow rate The option to override the dispense velocity
from liquid class specified in the liquid class. Selecting this
option enables you to specify a value for the
dispense flow rate without changing the
liquid class.
Distance from well bottom The absolute distance between the end of the
(mm) pipette tips and the bottom of the Tip Wash
Station during the Dispense to Waste task.
If you specify dynamic tip retraction, this is
the starting distance.
Parameter Description
Dynamic tip retraction The rate at which to raise the pipette head
(mm/µL) during the Dispense to Waste task.
Use dynamic tip retraction to prevent spills
as the pipette tips displace the liquid.
To move the pipettes:
• At the same rate as the volume change.
Calculate dynamic tip retraction (DTR) as
follows:
DTR = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTR > 1/A
• Slower than the volume change.
DTR < 1/A
The starting and ending positions can be
calculated as follows:
(Vdispensed * DTR) + Distancewell bottom
Perform tip touch on North/ The option to touch the tip on an outer side
East side of the adjacent northeast chimney or wall in
the Tip Wash Station after performing the
dispense.
Tip touch retract distance The vertical distance for the pipette tips to
(–20 to 50 mm) rise before performing the tip touch.
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter descriptions
The Mix task parameters include the following.
Parameter Description
Volume The volume of liquid to be mixed in each
(0–200 µL) well.
Parameter Description
Dispense at different The option to dispense at a pipette tip height
distance that is different than the aspirate distance.
Select the check box to enter a value for the
dispense distance.
Dynamic tip extension The rate at which the pipette head moves
(0–20 mm/µL) during the Aspirate task. The software
calculates the distance over which the tips
will move without crashing.
Use dynamic tip extension to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTE > 1/A
• Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume,
the software automatically moves downward
toward the well bottom with each aspirate
action.
Perform tip touch The option to touch the pipette tip on one or
more sides of the well.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the pipette tips to
(–20 to 50 mm) rise before touching the sides of the wells.
Parameter Description
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter descriptions
To run the Pump Reagent task, see “Performing a task using Diagnostics” on
page 104.
The Pump Reagent task has the following parameters.
Parameter Description
Reservoir mode The action of the task:
• Fill
• Empty
Use weigh station/shelf The option to use the Weigh Station or Weigh
Shelf.
Weigh station stop action The maximum fluid weight, in percent of the
threshold full weight that was calibrated on the Weigh
Station or Weigh Shelf.
For example, you can set the stop threshold
at 60% so that when the fluid reaches 60% of
the full weight, fluid starts to drain or pump
out of the reservoir.
Related information
Parameter descriptions
The Shake task contains the following parameters.
Parameter Description
Mode The action of the task:
• On. Turns on the Orbital Shaking Station.
• Off. Turns off the Orbital Shaking Station.
• Timed. Turns on the shaking timer. You
must specify the length of time to shake.
Time for operation in Timed The length of time, in seconds, you want to
mode (s) leave the shaking on. At the end of the
period, the shaking will turn off.
Related information
Parameter Description
Assembly order The order, from bottom to top, in which you
want to stack the station components. The
selections are:
• Base- Collection plate- Filter plate- Collar
(configuration A)
• Base- Collection plate- Collar
(configuration B)
• Base- Collar (configuration C)
For details, see “Installing and setting up a
Vacuum Filtration Station” on page 118.
Parameter Description
Insert height (0- 15 mm) The height of the white plastic spacer, if
present, that can be placed in the base
manually before starting the protocol. The
insert is used to raise the height of the
collection plate and reduce the spacing
between the filter plate nozzles and the
collection plate underneath.
Measure the height (mm) from the bottom
edge to the top edge of the insert. If you are
using multiple inserts, measure the combined
height of the stack of inserts.
Parameter Description
Vacuum Filtration Assembly The location of the assembled Vacuum
Filtration Station.
Parameter Description
Mode The action of the task. The options are On,
Off, and Timed.
• On and Off. If you are not timing the
filtering process, add two Move and
Filter Plate tasks in the protocol for
each filtering operation. One task turns
on the vacuum (Mode = On), and the
other task turns off the vacuum (Mode
= Off).
• Timed. If you plan to time the filtering
process, add only one Move and Filter
Plate task. The task turns on the
vacuum, and then turns off the vacuum
automatically at the end of the time
period.
Parameter Description
When filtration timing begin The different options for when to start
ME4C VARIO Vacuum Pump timing the filtration process:
only • When pressure is achieved
• When the vacuum pump starts
Time for operation in Timed The duration, in seconds, that you want to
mode (s) leave the vacuum on. At the end of the
period, the vacuum will turn off.
Hold or tap down filter plate The different options for whether to have
Bravo Platform only the Bravo gripper hold down the filter
plate when the vacuum is turned on to
ensure a secure vacuum seal:
• None
• Tap down. The Bravo gripper will hold
down the filter plate from the top only
for the time period specified in the
Duration for tap down parameter.
• Hold down. The Bravo gripper will hold
down the filter plate from the top for
the duration of the task.
Note: If you select None or Tap down, the
protocol can perform other tasks in
parallel. Concurrent operation is not an
option if you select Hold down.
Duration for tap down (1- 30 s) The length of time, in seconds, that the
Bravo Platform only Bravo gripper will hold down the filter
plate if the Tap down option is selected.
Parameter Description
Target pressure The difference between the pressure of the
ME4C VARIO Vacuum Pump outside atmosphere above the filter and the
only pressure in the Vacuum Filtration Station
manifold, including the enclosure beneath
the filter.
For example, if you set the Target pressure
to 600 mbar and the ambient pressure
displayed on the Vario pump is 1000 mbar,
the vacuum will remain on until the
reading on the Vario pump reaches
400 mbar.
Related information
Parameter descriptions
The Wash Tips task has the following parameters.
Parameter Description
Empty tips The option to empty the entire contents of
the pipette tips, including fluid and air. The
Volume parameter is ignored if this option is
selected.
Pre- aspirate volume (µL) The volume of air to be drawn before the
pipette tips enter the liquid.
Distance from well bottom The distance between the end of the pipette
(mm) tips and the well bottoms during the Wash
Tips task.
Parameter Description
Dynamic tip extension The rate at which the pipette head moves
(mm/(µL) during the Wash Tips task. The software
calculates the distance over which the tips
will move without crashing.
Use dynamic tip extension to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTE > 1/A
• Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(Vdispensed * DTE) + Dwell bottom
(Vaspirated * DTE) + Dwell bottom
Perform tip touch The option to touch the pipette tip on one or
more sides of the well.
Which side to perform tip The wall or walls for tip touch: North, South,
touch East, West, North/South, West/East, West/
East/South/North.
If you also select the Dispense to waste
during wash option, the tip touch is
performed on the northeast side only.
Tip touch retract distance The vertical distance the pipette tips rise
before touching the sides of the wells.
Tip touch horizontal offset The horizontal distance the tips move. The
value is based on the well diameter specified
by the labware definition.
The value of the parameter determines the
direction of movement:
• 0. Tips move a horizontal distance equal
to the well radius.
• Great than 0. Tips attempt to move past
the well radius, which results in a more
forceful tip touch.
• Less than 0. Tips move a distance less
than the radius of the well, resulting in a
lighter tip touch.
Parameter Description
Pump fill speed (%) The speed, in percent of maximum speed, of
liquid flow into the reservoir.
For the MicroWash Reservoir, this value
should be high enough for the washing liquid
to just bubble over the tops of the chimneys.
Related information
Parameter descriptions
The Pin Tool task contains the following parameters.
Parameter Description
Dwell time (s) The time duration that the pins remain at the
specified height (First distance or Second
distance) within the well.
For example, you might start with the
following values:
• Adsorb, Dispense into fluid, or Mix—0.5 s
or longer for more viscous fluids
• Blot—2 s, or longer for more viscous
fluids
Descriptive label A text label that you can add to the task icon
in the protocol. Click the arrow in the
Descriptive label box to choose an option.
The options include:
• Enter a JavaScript variable or script.
• Use a predefined label: Adsorb, Dispense,
Wash, Blot, or Mix
• Type your own label in the box.
Parameter Description
First distance (mm) The first height for the pin tool during the
Pin Tool task. The value is the distance
between the pin tips and the well bottoms.
For example, during an adsorb step, you
might set this value to 0 mm so that the pin
tips touch the bottom of the wells. This
parameter can affect the quantity adsorbed.
Use two distances The option to specify a second height for the
pins during the Pin Tool task.
For example, you could cycle the pin
positions between two heights within the
wells repeatedly to perform mixing or to
wash the pins.
Default: Not selected
Second distance (mm) The distance between the pin tips and the
well bottoms at the second height for the
pins.
Perform tip touch The option to touch the pins on one or more
sides of the well, or to enable the pins to
make lateral stirring moves inside the fluid,
for example during a wash task.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the pins to move
(mm) before moving laterally within the well,
where
• 0 is the vertical distance equal to the well
bottom
• > 0 is the vertical distance the pins rise
above the bottom
• < 0 is the vertical distance the pins
attempt to move past the well bottom
Parameter Description
Tip touch horizontal offset The horizontal distance that the pins move.
(mm) The value is based on the well diameter
specified by the labware definition,
where
• 0 is a distance equal to the well radius
• > 0 is the distance the pins attempt to
move past the well radius, which results
in a more forceful tip touch
• < 0 is a distance less than the radius of
the well, resulting in a lighter tip touch
or no tip touch
Related information
Parameter description
The AM Aspirate task has the following parameters.
Parameter Description
Volume (µL) The volume of liquid to be drawn into each
probe, cartridge, or tip.
Pre- aspirate volume (µL) The volume of air to be drawn before the
probes, cartridges, or tips enter the liquid.
Post- aspirate volume (µL) The volume of air to be drawn after the
liquid is drawn.
Parameter Description
Aspirate flow rate (0–500 The numerical value or the JavaScript
µL/s) variable that will override the aspirate
velocity setting in the liquid class.
A JavaScript variable enables the value to be
assigned later. For example, using a VWorks
form, an operator could easily change the
flow rate for an aspirate step in increments
from as low as 1 µL/min up to 2000 µL/min
or more using the same liquid class.
If the task is included in a VWorks macro, a
JavaScript variable enables you to change the
value for the task at the macro level.
Parameter Description
Distance from well bottom The distance between the tips of the probes,
(- 20–100 mm) cartridges, or disposable tips and the well
bottoms during the AM Aspirate task.
If you specify dynamic tip extension, this is
the distance at the end of the AM Aspirate
task.
Use a positive number for tasks that are
performed with mounted cartridges, mounted
tips, or bare probes at labware other than a
cartridge rack.
If the task is performed in the upper cup of
unmounted cartridges, you can use a negative
number for the parameter value. In this case,
the well bottom is the top of the cartridge
cup, as shown in the following figure.
Parameter Description
Dynamic tip extension The rate at which the Bravo 96AM Head
(0–20 mm/µL) moves during the AM Aspirate task. The
software calculates the distance over which
the tips will move without crashing.
Use dynamic tip extension to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTE > 1/A
• Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume,
the software automatically moves downward
toward the well bottom with each aspirate
action.
Perform tip touch The option to touch the pipette tip on one or
more sides of the well.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the pipette tips to
(–20 to 50 mm) rise before touching the sides of the wells.
Parameter Description
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter description
The AM Dispense task has the following parameters.
Parameter Description
Empty tips The option to empty all liquid from the
probes, cartridges, or tips instead of using
the dispense volume specification.
Parameter Description
Dispense flow rate (0– The numerical value or the JavaScript
500 µL/s) variable that will override the dispense
velocity setting in the liquid class.
A JavaScript variable enables the value to be
assigned later. For example, using a VWorks
form, an operator could easily change the
flow rate for an aspirate step in increments
from as low as 1 µL/min up to 2000 µL/min
or more using the same liquid class.
If the task is included in a VWorks macro, a
JavaScript variable enables you to change the
value for the task at the macro level.
Parameter Description
Distance from well bottom The distance between the tips of the probes,
(- 20–100 mm) cartridges, or disposable tips and the well
bottoms during the AM Dispense task.
If you specify dynamic tip extension, this is
the distance at the end of the AM Dispense
task.
Use a positive number for tasks that are
performed with mounted cartridges, mounted
tips, or bare probes at labware other than a
cartridge rack.
If the task is performed in the upper cup of
unmounted cartridges, you can use a negative
number for the parameter value. In this case,
the well bottom is the top of the cartridge
cup, as shown in the following figure.
Parameter Description
Dynamic tip retraction The rate at which to raise the pipette head
(0–20 mm/µL) during the Dispense task.
Use dynamic tip retraction to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip retraction (DTR) as
follows:
DTR = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTR > 1/A
• Slower than the volume change.
DTR < 1/A
The starting and ending positions can be
calculated as follows:
(Vdispensed * DTR) + Distancewell bottom
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the probes,
(–20 to 50 mm) cartridges, or tips to rise before touching the
sides of the wells.
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
Parameter description
The AM Mix task has the following parameters.
Parameter Description
Volume (µL) The volume of liquid to be mixed in each
well.
Pre- aspirate volume (µL) The volume of air to be drawn before the
pipette tips enter the liquid.
Blowout volume (µL) The volume of air to dispense after the main
volume has been dispensed while the tips are
still in the wells.
Typically, the blowout volume is the same as
the pre- aspirate volume.
Override flow rates from The option to override the aspirate and
liquid class dispense velocity in the selected liquid class.
Selecting this option enables you to specify a
value for the flow rate without changing the
liquid class.
Parameter Description
Aspirate/dispense flow The numerical value or the JavaScript
rate (0–500 µL/s) variable that will override both the aspirate
and dispense velocity in the selected liquid
class.
A JavaScript variable enables the value to be
assigned later. For example, using a VWorks
form, an operator could easily change the
flow rate for a mix step in increments from
as low as 1 µL/min up to 2000 µL/min or
more using the same liquid class.
If the task is included in a VWorks macro, a
JavaScript variable enables you to change the
value for the task at the macro level.
Dynamic tip extension The rate at which the Bravo 96AM Head
(0–20 mm/µL) moves during the aspirate action. The
software calculates the distance over which
the tips will move without crashing.
Use dynamic tip extension to prevent spills
as the pipette tips displace the liquid.
To move the tips:
• At the same rate as the volume change.
Calculate dynamic tip extension (DTE) as
follows:
DTE = (well depth)/(well vol) = 1/A,
where A is the cross- sectional area of a
well with straight walls
• Faster than the volume change.
DTE > 1/A
• Slower than the volume change.
DTE < 1/A
The starting and ending positions can be
calculated as follows:
(Vaspirated * DTE) + Distancewell bottom
Note: Instead of a negative aspirated volume,
the software automatically moves downward
toward the well bottom with each aspirate
action.
Parameter Description
Pipette technique The pipette location offset you want to use
for the Mix task.
The list of pipette techniques are defined in
the Pipette Technique Editor.
Perform tip touch The option to touch the tips on one or more
sides of the well.
Which sides to use for tip The side or sides of the well to use during
touch tip touch: North, South, East, West, North/
South, West/East, West/East/South/North.
Tip touch retract distance The vertical distance for the tips to rise
(–20 to 50 mm) before touching the sides of the wells.
Tip touch horizontal offset The horizontal distance the tips move. The
(–5 to 5 mm) value is based on the well diameter specified
by the labware definition.
For example, if you set a value of:
• 0, the tips move a horizontal distance
equal to the well radius
• > 0, the tips attempt to move past the
well radius, which results in a more
forceful tip touch
• < 0, the tips move a distance less than
the radius of the well, resulting in a
lighter tip touch
Related information
User Guide
G5562-90000
Revision C, April 2018