Establishing Acceptance Criteria For Analytical Methods
Establishing Acceptance Criteria For Analytical Methods
Establishing Acceptance Criteria For Analytical Methods
Methods
Knowing how method performance impacts out-of-specification rates may improve quality risk management and
product knowledge.
Oct 01, 2016
By Thomas A. Little, PhD [1]
BioPharm International
Volume 29, Issue 10
To control the consistency and quality of pharmaceutical products, analytical methods must be developed to measure
critical quality attributes (CQAs) of drug substance/drug product. Analytical method accuracy/bias and precision are
always in the path of drug evaluation and associated acceptance/failure in release testing. The following are three
equations that show how the analytical method is always influencing the quantitation of drug substance/product
(Equations 1–3):
[Eq. 1]
[Eq. 2]
Reportable Result = Test sample true value + Method Bias + Method Repeatability
[Eq. 3]
Knowing what is the allowable contribution of the method error in drug performance becomes crucial when building
product knowledge, process understanding, and the associated long-term product lifecycle control. Mathematically,
the variation of any drug product or drug substance is the additive variation of the method and test sample being
quantitated.
Generally, to control the quality of a product and to manage drug safety and efficacy, there are two key elements:
cinical trials evaluting the pharmacokinetics (PK) response to drug product and dose and specification limits (1) of
drug product and drug substance once clinical trials have demonstated the drug to be safe and effective. This logic is
essentially laid out in two guidance documents: International Council for Harmonization (ICH) Q6B Specifications and
ICH Q9 Quality Risk Management
(2).
Clearly defined method acceptance criteria that evaluate the goodness and fitness of an analytical method for its
indended purpose are mandatory to correctly validate an analytical method and know its contribution when
quantitating product performance or releasing a batch. Methods with excessive error will directly impact product
acceptance out-of-specification (OOS) rates and provide misleading information regarding product quality.
This strategy has its advantages and its drawbacks. The advantage is the lab can develop and evaluate the
goodness of a method independent of the product and the associated acceptance criteria it is intended to measure.
This is particularly of interest during early development when product specification limits (Q6B) are not yet available.
The penalty for solely depending on CV or % recovery is a method may be developed and qualified without knowing if
it is fit-for-purpose or fit-for-use, and knowing its associated influence on product acceptance and release testing.
Further, the traditional approach will often falsely indicate a method is performing poorly at low concentrations, when
in fact it is performing excellently. Conversely, at high concentrations, the method will often appear to be performing
well—as the % CV and % recovery appear to be acceptable—when it is actually unacceptable relative to the product
specification limits it will be used to evaluate.
What therefore should be the basis for measurement goodness, if not comparing method performance to the mean or
the theoretical concentration? The answer is simple: don’t evaluate a method relative to the mean, evaluate it relative
to the product specification tolerance or design margin it must conform to. This concept has been well established for
many years in chemical, automotive, and semiconductor industries and is recommended in the United States
Pharmacopeia (USP) <1033> and <1225> (4, 5). Effectively the question is: how much of the specification tolerance
is consumed by the analytical method? Finally, how does the method contribute to OOS events when releasing
product to the clinic or market?
Method error should be evaluated relative to the tolerance for two-sided limits, margin for one-sided limits, and the
mean or theoretical concentration if there are no specification limits (Equations 4–6).
ICH Q2: Discusses what to quantitate, what to report, study design, and sample size. No mention of
acceptance criteria is made in the standard, although it is implied there will be acceptance criteria generated
(3).
FDA, Analytical Procedures and Methods Validation for Drugs and Biologics (6): “An analytical procedure is
developed to test a defined characteristic of the drug substance or drug product against established
acceptance criteria for that characteristic. Early in the development of a new analytical procedure, the choice
of analytical instrumentation and methodology should be selected based on the intended purpose and scope
of the analytical method. Parameters that may be evaluated during method development are specificity,
linearity, limits of detection (LOD), and limits of quantitation (LOQ), range, accuracy, and precision.”
USP <1225>: “When validating physical property methods, consider the same performance characteristics
required for any analytical procedure. Evaluate use of the performance characteristics on a case-by-case
basis, with the goal of determining that the procedure is suitable for its intended use. The specific
acceptance criteria for each validation parameter should be consistent with the intended use of the method”
(5).
USP <1033>: “The validation target acceptance criteria should be chosen to minimize the risks inherent in
making decisions from bioassay measurements and to be reasonable in terms of the capability of the art.
When there is an existing product specification, acceptance criteria can be justified on the basis of the risk
that measurements may fall outside of the product specification” (4).
Table I: Method
validation and acceptance criteria.
Identification, demonstrate it is measuring the specific analyte and not some other protein or substance
Bias in the presence of interfering compounds or matrices.
Identification, 100% detection, report detection rate and 95% confidence limits
Reportable Specificity = Measurement - Standard (units) (in the matrix of interest)
Specificity/Tolerance *100, Excellent Results <= 5%, Acceptable Results <=10%
To set the limit of linearity, the following is recommended. Fit a linear regression line when correlating signal versus
theoretical concentration. Save the studentized residuals from the curve. Add a line at +1.96 (95% sure the response
is linear) and -1.96. Fit a quadratic fit to the studentized residuals. As long as the curve remains within +-1.96 of the
studentized residuals, the response of the assay is linear. When the curve exceeds the 1.96 limit, one is 95% sure
the assay is no longer linear. For Figure 1, one is 95% sure this assay is linear up to 30 ug/mL.
Figure 1: Studentized
residuals of a linear fit. (Courtesy of the author)
Recommended acceptance criteria for analytical methods for repeatability are less than or equal to 25% of tolerance.
For a bioassay, they are recommended to be less than or equal to 50% of tolerance.
Figure 2: Influence of
repeatability on capability (out-of-specification [OOS] rate in parts per million [PPM]). (Courtesy of author)
Recommended acceptance criteria for analytical methods for bias are less than or equal to 10% of tolerance. For a
bioassay, they are recommended to also be less than or equal to 10% of tolerance.
If the specification is two-sided and the LOD and LOQ are below 80% of the lower specification limit, then the LOD
and LOQ are considered having no impact on product quality determination and thus acceptable.
Criteria for IP % of tolerance or % margin: less than or equal to 25% Excellent, less than or equal to 30%
Acceptable.
IP should be evaluated at each concentration, variance components for the intra- and inter-assay error should be
reported (4) and IP % CV is report only.
Reporting Robustness
A robustness study has no acceptance criteria; however, the robustness study should indicate the method is accurate
and repeatable at the recommended best set point and across a defined range. It is expected that the robustness
study will be used to determine settings and ranges that will ensure bias less than 10% of tolerance and repeatability
less than 25% of tolerance.
Reporting Stability
A stability study on critical reagents such as standards and/or bulk materials has no acceptance criteria; however, the
study should indicate the expiry of pre-mixes, bulks, or standards.
References
1. ICH, Q6B Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products (ICH,
March 1999).
2. ICH, Q9 Quality Risk Management (ICH, 2006).
3. ICH, Q2(R1) Validation of Analytical Procedures: Text and Methodology (ICH, November 2005).
4. USP, <1033> Biological Assay Validation, USP 38 (USP, 2010).
5. USP, <1225> Validation of Compendial Procedures, USP 38 (USP, 2015).
6. FDA, Analytical Procedures and Methods Validation for Drugs and Biologics, Guidance for Industry (CDER, July
2015).
7. T. Little, Accuracy to Precision (ATP) Profiler. [2]
Article Details
BioPharm International
Vol. 29, No. 10
Pages: 44–48
Citation
When referring to this article, please cite it as T. Little, "Establishing Acceptance Criteria for Analytical
Methods," BioPharm International 29 (10) 2016.
Links:
[1] http://www.biopharminternational.com/thomas-little-phd-0
[2] http://thomasalittleconsulting.com/tools/generalstatistics/index.php