3260B User Manual Iss G PDF
3260B User Manual Iss G PDF
3260B
User Manual
Issue G
Part No 9H3260B
USA UK Asia
Wayne Kerr Electronics Inc. Wayne Kerr Electronics Microtest
165L New Boston Street Vinnetrow Business Park 14F-6, No.79, Hsin Tai Wu Road, Sec. 1,
Woburn MA 01801-1744 Vinnetrow Road Hsi-chih, Taipei 221, Taiwan, R.O.C.
Tel: 781 938 8390 Chichester Tel: +886-2-2698-4104
Fax: 781 933 9523 West Sussex PO20 1QH Fax: +886-2-2698-0716
email: [email protected] Tel: +44 (0)1243 792200 Email: [email protected]
www.waynekerrtest.com Fax: +44 (0)1243 792201 www.waynekerrtest.com
email: [email protected]
email: [email protected]
www.waynekerrtest.com
Wayne Kerr Electronics Limited 2003
The copyright in this work is vested in Wayne Kerr Electronics Limited and this document is
issued for the purpose only for which it is supplied. No licence is implied for the use of any
patented feature. It must not be reproduced in whole or in part, or used for tendering or
manufacturing purposes except under an agreement or with the consent in writing of and then
only on the condition that this notice is included in any such reproduction. Information
furnished is believed to be accurate but no liability in respect of any use of it is accepted by
Wayne Kerr Electronics Limited.
CONTENTS
1. SAFETY............................................................................................................................... 1–1
1.1 General............................................................................................................................ 1–1
1.2 AC Power Supply ........................................................................................................... 1–2
1.3 Adjustment, Maintenance and Repair............................................................................. 1–2
1.4 Static Electricity.............................................................................................................. 1–3
2. INTRODUCTION............................................................................................................... 2–1
3. INSTALLATION................................................................................................................ 3–1
3.1 AC Line Connections...................................................................................................... 3–1
3.2 Location .......................................................................................................................... 3–1
3.2.1 Rack Mounting......................................................................................................... 3–1
3.3 Wayne Kerr Electronics Measurement Leads ................................................................ 3–2
3.3.1 Other Test Leads...................................................................................................... 3–3
4. OPERATION ...................................................................................................................... 4–1
4.1 The Rear Panel................................................................................................................ 4–1
4.1.1 Voltage Selector....................................................................................................... 4–1
4.1.2 IEC Socket and Fuse Holder ................................................................................... 4–1
4.1.3 Rear Panel Control Connections............................................................................. 4–2
4.1.4 Safety Interlock ........................................................................................................ 4–2
4.1.5 External Trigger ...................................................................................................... 4–3
4.1.6 Parallel Printer Connector...................................................................................... 4–3
4.1.7 GPIB Connector ...................................................................................................... 4–5
4.1.8 Auxiliary In .............................................................................................................. 4–5
4.1.9 Auxiliary Control Out .............................................................................................. 4–5
4.1.10 Auxiliary AC Out ................................................................................................... 4–6
4.1.11 Bin Handler Interface (Option) ............................................................................. 4–6
4.2 The Front Panel............................................................................................................... 4–8
4.2.1 Switching the Instrument ON................................................................................... 4–9
4.2.2 Switching the Instrument OFF................................................................................. 4–9
4.2.3 The Soft Keys ........................................................................................................... 4–9
4.2.4 The Navigation Keys.............................................................................................. 4–10
4.2.5 The Control Keys ................................................................................................... 4–11
4.2.6 The Data Entry Keypad ......................................................................................... 4–12
4.3 Trimming ...................................................................................................................... 4–16
4.3.1 Performing an O/C Trim (Pri) or S/C Trim (Pri).................................................. 4–17
4.3.2 Performing an O/C Trim (Pri–Sec) ....................................................................... 4–18
4.3.3 Performing an S/C Trim (Sec) ............................................................................... 4–19
4.4 HF Lead Compensation ................................................................................................ 4–19
4.4.1 Performing HF Lead Compensation...................................................................... 4–19
4.5 Self Calibration............................................................................................................. 4–21
4.5.1 Performing Self Calibration .................................................................................. 4–22
4.6 Handler Calibration ...................................................................................................... 4–23
4.6.1 O/C Trim and S/C Trim ......................................................................................... 4–23
4.6.2 HF Lead Compensation ......................................................................................... 4–25
4.7 Measuring a Component in IMPEDANCE MODE ...................................................... 4–26
4.7.1 Example.................................................................................................................. 4–27
4.7.2 IMPEDANCE MODE Parameters......................................................................... 4–29
5. ADVANCED OPERATION............................................................................................... 5–1
5.1 Front Panel Connections ................................................................................................. 5–1
5.1.1 Connection Protocol ................................................................................................ 5–1
5.1.2 Two-, Three- and Four-Terminal Connections........................................................ 5–2
5.2 In-Circuit Measurements................................................................................................. 5–3
5.3 Measurement of Very Small Capacitors ......................................................................... 5–4
5.4 Measurement of Very Small Inductors ........................................................................... 5–5
5.5 Measurement of Iron-Cored and Ferrite Inductors ......................................................... 5–5
5.6 Measurement of Transformers ........................................................................................ 5–5
5.7 TRANSFORMER MODE .............................................................................................. 5–6
5.7.1 Example.................................................................................................................... 5–7
5.7.2 TRANSFORMER MODE Parameters...................................................................... 5–8
5.8 INSULATION MODE (Optional) ................................................................................ 5–10
5.8.1 Connections............................................................................................................ 5–11
5.8.2 Trimming................................................................................................................ 5–13
5.8.3 Operation ............................................................................................................... 5–14
5.8.4 Example.................................................................................................................. 5–15
5.8.5 INSULATION MODE Parameters......................................................................... 5–17
5.9 BINNING MODE (Optional)........................................................................................ 5–18
5.9.1 BINNING MODE – Set .......................................................................................... 5–18
5.9.2 BINNING MODE – Sort......................................................................................... 5–22
5.9.3 BINNING MODE – Count ..................................................................................... 5–23
5.10 SEQUENCE EDIT MODE ......................................................................................... 5–24
5.10.1 Sequence Mode Program Files ............................................................................ 5–25
5.10.2 Sequence Mode Test Program ............................................................................. 5–28
5.11 SEQUENCE RUN MODE.......................................................................................... 5–30
5.11.1 Selecting a Program............................................................................................. 5–30
5.11.2 Program Development ......................................................................................... 5–30
5.11.3 Running a Program.............................................................................................. 5–31
5.12 HANDLER MODE ..................................................................................................... 5–32
5.12.1 Use of HANDLER MODE.................................................................................... 5–32
5.13 TELECOMS MODE................................................................................................... 5–32
5.13.1 Example................................................................................................................ 5–33
5.13.2 TELECOMS MODE Parameters ......................................................................... 5–34
5.14 MULTI FREQ MODE ................................................................................................ 5–36
5.14.1 MULTI FREQ – Set.............................................................................................. 5–36
5.14.2 MULTI FREQ – Run ............................................................................................ 5–40
5.15 GRAPH MODE .......................................................................................................... 5–41
5.15.1 GRAPH MODE – Set Parameters........................................................................ 5–42
5.15.2 Parameters Available when the Graph is Displayed........................................... 5–44
5.15.3 Example................................................................................................................ 5–45
5.16 RESONANCE MODE ................................................................................................ 5–47
5.17 DEMAG MODE ......................................................................................................... 5–49
5.17.1 DEMAG MODE Parameters ............................................................................... 5–49
5.18 The SETTINGS Page.................................................................................................. 5–49
5.18.1 The SETTINGS Page Parameters........................................................................ 5–50
5.19 The CAL STATUS Page ............................................................................................ 5–52
6. GENERAL PURPOSE INTERFACE BUS (GPIB) ........................................................ 6–1
6.1 GPIB Control .................................................................................................................. 6–1
6.1.1 Introduction ............................................................................................................. 6–1
6.1.2 Interface Specification............................................................................................. 6–1
6.1.3 Changing GPIB Address.......................................................................................... 6–2
6.1.4 Message Syntax........................................................................................................ 6–2
6.1.5 Data Output ............................................................................................................. 6–5
6.1.6 Status Reporting....................................................................................................... 6–6
6.1.7 Common Commands .............................................................................................. 6–12
6.1.8 Standard Operation Status Commands ................................................................. 6–13
6.2 Device-Specific Commands.......................................................................................... 6–14
6.2.1 Command Summary ............................................................................................... 6–14
6.3 Example Programs ........................................................................................................ 6–91
6.3.1 Example 1 .............................................................................................................. 6–93
6.3.2 Example 2 .............................................................................................................. 6–94
6.3.3 Example 3 .............................................................................................................. 6–95
6.3.4 Example 4 .............................................................................................................. 6–97
6.3.5 Example 5 .............................................................................................................. 6–98
7. SPECIFICATION............................................................................................................... 7–1
7.1 Measurement Functions.................................................................................................. 7–1
7.1.1 IMPEDANCE MODE .............................................................................................. 7–1
7.1.2 HANDLER MODE ................................................................................................... 7–1
7.1.3 TRANSFORMER MODE ......................................................................................... 7–1
7.1.4 RESONANCE MODE .............................................................................................. 7–1
7.1.5 DEMAGNETISATION MODE................................................................................. 7–1
7.1.6 INSULATION MODE (Option)................................................................................ 7–1
7.1.7 BINNING MODE ..................................................................................................... 7–2
7.1.8 SEQUENCE MODE................................................................................................. 7–2
7.1.9 LF TELECOM (Option)........................................................................................... 7–2
7.2 Additional Measurement Facilities................................................................................. 7–2
7.2.1 2/4 Terminal............................................................................................................. 7–2
7.2.2 Transformer Ratio Correction................................................................................. 7–3
7.2.3 Low Magnetization Measurement............................................................................ 7–3
7.2.4 Measurement Speeds ............................................................................................... 7–3
7.2.5 Repeat Measurement ............................................................................................... 7–3
7.2.6 Frequency Steps....................................................................................................... 7–3
7.3 Test Conditions............................................................................................................... 7–3
7.3.1 Measurement Range ................................................................................................ 7–3
7.3.2 Frequency (AC Measurements) ............................................................................... 7–3
7.3.3 Drive Level............................................................................................................... 7–3
7.3.4 DC Bias Current (Impedance modes only).............................................................. 7–4
7.3.5 Insulation (Option) .................................................................................................. 7–4
7.3.6 LF Telecom (Option)................................................................................................ 7–4
7.4 Basic Accuracy ............................................................................................................... 7–4
7.4.1 Rdc ........................................................................................................................... 7–4
7.4.2 L, R, Z, C .................................................................................................................. 7–4
7.4.3 Dissipation Factor (D)............................................................................................. 7–4
7.4.4 Quality Factor (Q) ................................................................................................... 7–4
7.4.5 Insulation (Option)................................................................................................... 7–5
7.4.6 Insertion Loss (LF Telecom Option)........................................................................ 7–5
7.4.7 Return Loss (LF Telecom Option) ........................................................................... 7–5
7.4.8 Accuracy Chart ........................................................................................................ 7–6
7.5 General Data ................................................................................................................... 7–7
7.5.1 Power Supply ........................................................................................................... 7–7
7.5.2 Display ..................................................................................................................... 7–7
7.5.3 Measurement Connections....................................................................................... 7–7
7.5.4 Remote Control (Option) ......................................................................................... 7–7
7.5.5 Binning Interface (Option)....................................................................................... 7–7
7.5.6 Printer Output .......................................................................................................... 7–7
7.5.7 Environmental Conditions ....................................................................................... 7–7
7.5.8 Safety........................................................................................................................ 7–8
7.5.9 EMC ......................................................................................................................... 7–8
7.5.10 Mechanical............................................................................................................. 7–8
7.5.11 Accessories Supplied.............................................................................................. 7–8
7.5.12 Options and Accessories ........................................................................................ 7–8
8. THEORY REFERENCE.................................................................................................... 8–1
8.1 Abbreviations .................................................................................................................. 8–1
8.2 Formulae ......................................................................................................................... 8–1
8.3 Series/Parallel Conversions ............................................................................................ 8–2
8.4 Polar Derivations............................................................................................................. 8–2
9. MAINTENANCE, SUPPORT AND SERVICES ............................................................. 9–1
9.1 Guarantee ........................................................................................................................ 9–1
9.2 Maintenance .................................................................................................................... 9–1
9.2.1 Cleaning ................................................................................................................... 9–1
9.2.2 Safety Checks ........................................................................................................... 9–1
9.3 Support and Service ........................................................................................................ 9–2
ILLUSTRATIONS
Figure 2-1 PMA3260B Precision Magnetics Analyzer ............................................................ 2–1
Figure 3-1 Procedure for Attachment of Rack Mounting Brackets.......................................... 3–2
Figure 3-2 4 Terminal Measurement ........................................................................................ 3–3
Figure 3-3 2 Terminal Measurement ........................................................................................ 3–3
Figure 4-1 The 3260B Rear Panel ............................................................................................ 4–1
Figure 4-2 Typical Bias Interlock Fixture ................................................................................ 4–2
Figure 4-3 Standard Bin Handler Timing ................................................................................. 4–7
Figure 4-4 The 3260B Front Panel ........................................................................................... 4–8
Figure 4-5 The Soft Keys........................................................................................................ 4–10
Figure 4-6 The Navigation Keys............................................................................................. 4–10
Figure 4-7 The Control Keys .................................................................................................. 4–11
Figure 4-8 Bias Safety Hazard Notice .................................................................................... 4–11
Figure 4-9 The3260B Main Menu (pages 1 and 2) ................................................................ 4–11
Figure 4-10 Single Shot Mode................................................................................................ 4–12
Figure 4-11 Repetitive Mode.................................................................................................. 4–12
Figure 4-12 The Data Entry Keypad....................................................................................... 4–12
Figure 4-13 Example of an Error Message from an Invalid Keypad Entry............................ 4–13
Figure 4-14 'Nearest Available' Error Message ..................................................................... 4–13
Figure 4-15 Connections for O/C trimming of Kelvin clips................................................... 4–16
Figure 4-16 Connections for S/C trimming of Kelvin clips ................................................... 4–16
Figure 4-17 Calibrate Mode.................................................................................................... 4–17
Figure 4-18 Settings Lost........................................................................................................ 4–17
Figure 4-19 O/C Trim (Pri) Options ....................................................................................... 4–18
Figure 4-20 S/C Trim(Pri) Options......................................................................................... 4–18
Figure 4-21 High Voltage ON warning .................................................................................. 4–19
Figure 4-22 HF Lead Compensation....................................................................................... 4–20
Figure 4-23 HF Lead Compensation Data Lost...................................................................... 4–21
Figure 4-24 Self-Calibration Reminder .................................................................................. 4–21
Figure 4-25 Calibration Data Lost .......................................................................................... 4–22
Figure 4-26 Handler Cal Mode............................................................................................... 4–24
Figure 4-27 Front Panel BNC Connections for Handler Mode .............................................. 4–24
Figure 4-28 Handler Factory Cal Lost Message..................................................................... 4–24
Figure 4-29 HANDLER CAL MODE O/C Trim Options...................................................... 4–25
Figure 4-30 HANDLER CAL MODE S/C Trim Options ...................................................... 4–25
Figure 4-31 HF Lead Compensation....................................................................................... 4–26
Figure 4-32 Impedance Mode ................................................................................................. 4–27
Figure 4-33 Example Display when Measuring the Inductance and Quality Factor of a
100µH Inductor ................................................................................................. 4–28
Figure 4-34 Non-Soft Key IMPEDANCE MODE Parameters............................................... 4–31
Figure 4-35 Nearest Drive Level Warning ............................................................................. 4–33
Figure 4-36 Cannot Set Level Warning.................................................................................. 4–33
Figure 5-1 Connection Protocol for IMPEDANCE MODE ..................................................... 5–2
Figure 5-2 Connection Protocol for TRANSFORME/TELECOM MODE ............................. 5–2
Figure 5-3 Connection Protocol for HANDLER MODE ......................................................... 5–2
Figure 5-4 In-Circuit Measurements......................................................................................... 5–3
Figure 5-5 Effect of shunt loading on current terminal ............................................................ 5–4
Figure 5-6 Example of Measuring Turns Ratio in TRANSFORMER MODE......................... 5–8
Figure 5-7 INSULATION MODE: Current Too Large .......................................................... 5–11
Figure 5-8 Unit Not Available ................................................................................................ 5–11
Figure 5-9 Active Connection for 2-TERM INSULATION MODE (Pri–GND) .................. 5–12
Figure 5-10 Active Connection for 4-TERM INSULATION MODE (Pri–GND) ................ 5–12
Figure 5-11 Active Connection for 2-TERM INSULATION MODE (Sec–GND) ............... 5–12
Figure 5-12 Active Connection for 4-TERM INSULATION MODE (Sec–GND) ............... 5–13
Figure 5-13 Active Connections for 2-TERM INSULATION MODE (Pri–Sec) ................ 5–13
Figure 5-14 Active Connections for 4-TERM INSULATION MODE (Pri–Sec) ................. 5–13
Figure 5-15 High Voltage ON warning................................................................................... 5–14
Figure 5-16 INSULATION MODE: display readouts from above example .......................... 5–17
Figure 5-17 BINNING MODE – Set Main Screen ................................................................. 5–19
Figure 5-18 BINNING MODE Nested Percentage Limits .................................................... 5–19
Figure 5-19 BINNING MODE Stacked Absolute Limits ...................................................... 5–19
Figure 5-20 BINNING MODE – Set: Reset Warning ............................................................ 5–20
Figure 5-21 BINNING MODE – Set, Measurement Options ................................................. 5–21
Figure 5-22 Unit Not Available Message ............................................................................... 5–21
Figure 5-23 BINNING MODE – Sort ..................................................................................... 5–22
Figure 5-24 Delete Last Result Message ................................................................................ 5–22
Figure 5-25 BINNING MODE – Count.................................................................................. 5–23
Figure 5-26 Delete Bin Counts Warning ................................................................................ 5–23
Figure 5-27 SEQUENCE EDIT MODE: Program Template ................................................. 5–25
Figure 5-28 SEQUENCE EDIT MODE:Sequence File Screen.............................................. 5–25
Figure 5-29 SEQUENCE EDIT MODE:New Program .......................................................... 5–26
Figure 5-30 SEQUENCE EDIT MODE:Program Label ........................................................ 5–26
Figure 5-31 SEQUENCE EDIT MODE:Additional Test Step ............................................... 5–28
Figure 5-32 SEQUENCE EDIT MODE:Change Test Parameter ........................................... 5–29
Figure 5-33 SEQUENCE EDIT MODE:Moving a Test Step................................................. 5–30
Figure 5-34 SEQUENCE RUN MODE: Connection Prompt................................................. 5–31
Figure 5-35 SEQUENCE RUN MODE: Failure Message...................................................... 5–32
Figure 5-36 TELECOMS MODE: Simple Termination......................................................... 5–33
Figure 5-37 TELECOMS MODE: Direct Damped Termination............................................ 5–33
Figure 5-38 TELECOMS MODE: Damped Termination with DC Blocking Capacitor........ 5–33
Figure 5-39 Example of Measuring Insertion Loss and Return Loss in TELECOMS
MODE ............................................................................................................... 5–34
Figure 5-40 MULTI FREQ – Set Display With No Parameters Set up.................................. 5–36
Figure 5-41 MULTI FREQ – Set Display Example ............................................................... 5–39
Figure 5-42 Delete Frequency Message.................................................................................. 5–39
Figure 5-43 Initial MULTI FREQ – Run Display (from example in section 5.14.1.1) .......... 5–40
Figure 5-44 MULTI FREQ – Run........................................................................................... 5–41
Figure 5-45 GRAPH MODE – Set.......................................................................................... 5–42
Figure 5-46 Graph Plotted From Example.............................................................................. 5–47
Figure 5-47 Series Equivalent Circuit..................................................................................... 5–47
Figure 5-48 Parallel Equivalent Circuit .................................................................................. 5–47
Figure 5-49 RESONANCE MODE ........................................................................................ 5–48
Figure 5-50 RESONANCE MODE Results............................................................................ 5–48
Figure 5-51 Resonance Not Found Message .......................................................................... 5–48
Figure 5-52 DEMAG MODE.................................................................................................. 5–49
Figure 5-53 The SETTINGS Page.......................................................................................... 5–50
Figure 5-54 The CAL STATUS Page..................................................................................... 5–53
Figure 6-1 GPIB Message Structure ......................................................................................... 6–2
Figure 6-2 GPIB Command Structure ...................................................................................... 6–3
Figure 6-3 GPIB Data Output ................................................................................................... 6–5
Figure 6-4 GPIB RMU Structure.............................................................................................. 6–6
Figure 6-5 Status Byte Register ................................................................................................ 6–7
Figure 6-6 Standard Event Status Register ............................................................................... 6–8
Figure 6-7 Event Status Byte Register...................................................................................... 6–9
Figure 6-8 Standard Operation Status Group ......................................................................... 6–10
Safety 1–1
1. SAFETY
1.1 General
This equipment has been designed to meet the requirements of EN61010-1 ‘Safety
requirements for electrical equipment for measurement, control & laboratory use’ and has left
the factory in a safe condition.
The following definitions in EN61010-1 are applicable:
The RESPONSIBLE BODY must ensure that this equipment is only used in the manner
specified. If it is not used in such a manner, the protection provided by the equipment may be
impaired.
This product is not intended for use in atmospheres which are explosive, corrosive or adversely
polluted (e.g. containing conductive or excessive dust). It is not intended for use in safety
critical or medical applications.
The equipment can cause hazards if not used in accordance with these instructions. Read them
carefully and follow them in all respects.
Do not use the equipment if it is damaged. In such circumstances the equipment must be
made inoperative and secured against any unintentional operation.
The user must also ensure that the protective ground lead would be the last to break should the
cable be subject to excessive strain.
If the plug is fused, a 3-amp fuse should be fitted.
If the power cable electrical connection to the AC power plug is through screw terminals then,
to ensure reliable connections, any solder tinning of the cable wires must be removed before
fitting the plug.
Before switching on the equipment, ensure that it is set to the voltage of the local AC power
supply.
WARNING!
Any interruption of the protective ground conductor inside or outside the equipment or
disconnection of the protective ground terminal is likely to make the equipment
dangerous. Intentional interruption is prohibited.
WARNING!
The equipment must be disconnected from all voltage sources before it is opened for any
adjustment, replacement, maintenance, or repair.
When the equipment is connected to the local AC power supply, internal terminals may be live
and the opening of the covers or removal of parts (except those to which access can be gained
by hand) is likely to expose live parts.
Safety 1–3
Capacitors inside the equipment may still be charged even if the equipment has been
disconnected from all voltage sources.
Any adjustment, maintenance, or repair of the opened equipment under voltage must be carried
out by a skilled person who is aware of the hazards involved.
Service personnel should be trained against unexpected hazards.
Ensure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and short-circuiting of fuse holders is prohibited.
2. INTRODUCTION
3. INSTALLATION
The supply voltage setting can be checked by looking through the transparent window on the
rear panel next to the power inlet socket. This can be changed by first disconnecting the unit
from the electrical supply, removing the window and adjusting the switch to read the required
voltage. Replace the window and ensure that the fuse rating is correct:
230V 1A–T
115V 2A–T
3.2 Location
The 3260B is intended for use either on the bench or in a rack. The power modules are
convection cooled and care must be taken not to restrict any of the air paths.
measurements can be performed without moving the measuring leads: stable lead positioning is
important when measuring low value inductors.
RD OR
4. OPERATION
WARNING!
This equipment is intended for use by suitably trained and competent persons.
This product can cause hazards if it is not used in accordance with these instructions.
Read them carefully and follow them in all respects. Double check connections to the unit
before use.
Bias Safety
Interlock Bin Handler
Trigger in (option) Voltage Selector
Parallel Printer
GPIB Standard GPIB For remote operation. Sections 4.1.7 and 4.2.5
Duplicates action of
TRIGGER IN BNC Section 4.1.5
front panel trigger key.
OPTIONAL - to interface
HANDLER 25-way D-type (male) Sections 4.1.11 and 5.9
with bin sorting equipment.
Only when the fixture door is closed, and the microswitch therefore made, can DC bias be
activated. At this time, the relay is energized, activating an AC supply for a solenoid which can
be used to lock the door while bias remains available.
The solenoid drive relay should have a 5V DC coil of resistance not less than 200Ω. Diode coil
protection is provided within the bias circuitry.
The door lock is activated when Bias On is selected on the instrument. If the door switch or
interlock lead is broken, DC bias is inhibited.
Operation 4–3
Note
If the safety interlock is not required, it is necessary to insert the 3.5mm jack plug, with the ring
and sleeve connections linked, into the socket on the back panel. Failure to do this will result in
bias being inhibited, and the message, ✷ Bias Interlock ✷ being displayed.
Note:
The printer must be enabled before results can be output to it: enter Code 30 from the MAIN
MENU as described in section 4.2.6.1. If printer output is enabled with no printer connected or
with the printer power switched off, a message will be displayed and printer output will be
disabled. Printer output will also be disabled when the instrument is switched off or goes to
remote control. To manually disable the printer output enter Code 31 from the MAIN MENU.
Status
Whenever a mode is selected or changed, the status of the selected mode will be printed, for example:
Test Conditions
A summary of the test conditions will be output whenever the type of test is changed, e.g. from
Rdc Meas to AC Meas in IMPEDANCE MODE.
AC Meas:
L Q Parallel
1.02Vac 1.0000kHz
DC Bias 0.000 A OFF
Range Auto
Speed Fast
ALC off
4–4 Operation
If, on selecting a mode, the required test type is already selected, press the highlighted key once
to obtain this test condition summary.
Measurement Results
Measurement results will be output to the printer in IMPEDANCE, TRANSFORMER,
INSULATION and BINNING modes whenever a single-shot test is performed. The value of
the frequency, test level or DC bias will be printed if highlighted in the test set up. For example,
printing L and Q variation with frequency.
To obtain a print out of test results without the additional parameter, move the test set-up
highlight to Range or Speed.
Binning
A PRINT soft key label will be displayed in BINNING MODE – COUNT if a printer is
connected and printing is enabled (Code 30 from the MAIN MENU).
In BINNING MODE – SET, Code 34 will print a list of the current bin set-up (the printer
must first be enabled by entering Code 30 from the MAIN MENU).
In BINNING MODE – SORT and single-shot mode, measurement results and bin numbers
will be printed if printing is enabled.
13 Select
4.1.8 Auxiliary In
For future expansion.
The two output signal lines BUSY and BDA (Bin Data Available) will at any time assume
one of four different states:
4.1.11.3 Busy
Both BUSY and BDA are low in this state. The BUSY line goes low to acknowledge the
trigger and also to indicate that the component between its terminals is in the process of being
measured and should not be removed until the BUSY line goes high again, when the
instrument enters the next state.
If the process has been completed without interruption, the instrument will re-enter the ‘Ready
for Trigger’ state waiting to measure the next component. The bin handler hardware should
respond to the falling edges of the BDA line and the relevant BIN line, which will occur only
when a component has been successfully measured and sorted.
When the set up details are showing on the screen (in some modes, there is a soft key which
toggles between Hide Setup and Show Setup: this soft key can be seen in Figure 4-5), the left
and right navigation keys, and , allow each parameter to be selected in turn. When a
parameter is selected, the up and down navigation keys, and , step the numeric value for
AC level, frequency and DC Bias: the steps vary according to the value but are always
multiples of 1, 2 or 5. Finer frequency steps can be achieved by using the data entry keypad, see
section 4.2.6. For other parameters, the and navigation keys change the settings, e.g.
Auto Range/[fixed range], Slow/Med/Fast/Max, ALC off/on.
Operation 4–11
Pressing Local restores control to the front panel when the instrument is under GPIB control.
Bias toggles the DC bias current ON and OFF. Before bias can be switched ON, the Bias
parameter must be selected with the left and right navigation keys, and , and a Bias current
value entered, either with the up and down navigation keys, and , or using the data entry
keypad: see section 4.2.6. During AC measurements the 3260B can supply a DC bias current
which is variable between 1mA and 1A (DC Bias NORM) and when used with an external
3265B DC Bias Unit, up to 125A DC bias current is available (DC Bias BOOST). Whenever
DC bias is on, the message shown below in Figure 4-8 will be displayed at the top of the screen.
Pressing the Menu key displays the MAIN MENU, from where each mode of operation can be
selected with the soft keys. To select the second page of menu options, press the Menu control
key again or press the MORE . . soft key.
The 2/4 Term key selects 2- or 4-terminal measurement. When 2-terminal measurement is
selected the associated indicator lights and the display shows 2-TERM… MODE at the top of
the screen. Switching between 2- and 4-terminal mode will require the analyzer to be
retrimmed, see section 4.3. 4-terminal measurements are recommended when measuring low
4–12 Operation
impedance devices. Quoted accuracies assume 4-terminal connection whenever the DUT
impedance is below 1kΩ.
Sngl/Rep toggles between Single shot mode and Repetitive mode. When Sngl/Rep is pressed
the display briefly indicates the mode selected as shown in Figure 4-10 and Figure 4-11 below.
Single shot mode is also indicated by the lack of a continuously flashing asterisk (*) in the top
left corner of the screen. Conversely, the presence of a continuously flashing asterisk indicates
that the instrument is in repetitive mode. The asterisk flashes once every time the instrument
makes a measurement.
When in single shot mode, the Trigger key initiates a single measurement. If it is pressed and
held, the analyzer will fall into repetitive measurement mode until the key is released.
The data-entry keypad is a multi-function key set permitting manual entry of data values,
measurement units and control codes.
The Units key must be used prior to keying a unit or multiplier. Where more than one unit is
available on a key, e.g. D/Q, pressing the key will display the first unit, pressing the key again
will display the second unit. Terminate the units mode with Enter to accept the key sequence.
Pressing Clear will delete the whole key sequence; pressing will delete the last key press.
An invalid keypad entry may cause the entry line to be cleared and an error message, such as
the one shown in Figure 4-13, to be displayed, in which case the existing settings will be
preserved; or the nearest available value may be set, accompanied by the error message shown
in Figure 4-14.
Operation 4–13
Figure 4-13 Example of an Error Message from an Invalid Figure 4-14 'Nearest Available' Error Message
Keypad Entry
The +/- key may be used before or after a value to change its sign. If the key is pressed more than once,
the value will toggle between + and -. For numbers which are positive only, this key is disabled.
MAIN MENU
Code Description
0.1 Display the character set. Press any key to display more characters, when all characters
have been shown, the main menu will be displayed and normal operation can be
resumed.
9.1 Load default values of non-volatile RAM variables, measurement conditions and trim
values.
SETTINGS MENU
Code Description
34 Print the settings page. Printer must be enabled first—enter code 30 in MAIN MENU.
Code Description
Code Description
34 Print the full list of the current program. Code 30 must be entered first from the MAIN
MENU page.
Code Description
Code Description
18 Single-shot mode
19 Repetitive mode
Code Description
Code Description
19 Repetitive mode
Operation 4–15
Code Description
34 Print the present bin set-up. Code 30 must be entered first from the MAIN MENU page.
Code Description
19 Repetitive mode
[.] [0] [2] [7] [3] [9] [Units] [H] check data entry line is correct, then press [Enter]
or
[2] [7] [.] [3] [9] [Units] [m] [H] [Enter]
If a mistake is made in a key sequence, before pressing Enter, press to delete the last key
press or Clear to delete the whole key sequence.
4.3 Trimming
The purpose of trimming is to eliminate the effects of stray capacitance or series impedance in
the connecting leads or fixture.
The trim values are held in non-volatile stores and for most measurements no retrimming is
necessary. The exceptions are:
• when the lead set or fixture is changed;
• when the highest possible accuracy is required for measurements of very high or very low
impedances;
• when maximum accuracy is required when switching between modes, in which case
maximum accuracy will be obtained by trimming from the mode which is to be used for
component measurement;
• when the instrument is switched between 4-terminal and 2-terminal operation;
• when the instrument is switched to, or from, HANDLER MODE, see section 4.6—
Handler Calibration.
Depending on the trim option selected, the analyzer trims by making measurements at a number
of frequencies, including the measurement frequency in use when the trim was initiated, and
storing the corrections for each. If the measurement frequency is changed the analyzer
automatically applies a new correction value by interpolation of the stored values. Corrections
for the Rdc functions are also stored.
For O/C Trim the Kelvin clips or fixture jaws should be separated by a distance equivalent to
the DUT pin separation.
For S/C Trim the connector jaws should be clipped to a piece of wire or a component lead as
close together as possible. Do not connect the clips directly together: this does not provide the
necessary 4-terminal short circuit and will lead to trim errors.
Figure 4-15 Connections for O/C trimming of Kelvin clips Figure 4-16 Connections for S/C trimming of Kelvin clips
For HF Lead Compensation the transfer standard capacitor supplied with the analyzer should
be placed in the Kelvin clips or fixture jaws. Both O/C and S/C Trims must be performed
before performing HF lead compensation.
Operation 4–17
Note:
If the instrument is switched OFF during the trim, the messages shown in Figure 4-18 and
Figure 4-23 will be displayed when the instrument is next switched ON. The component
measurement modes will be reset to the default settings and or
will be displayed at the top of the screen. These messages will only be
cleared by performing the appropriate trims. The instrument can be used with the default
settings but it is recommended that O/C trim and/or S/C trim is run for full measurement
accuracy.
Figure 4-18 will also be displayed when power is removed during other critical routines,
such as calibration and data entry.
4–18 Operation
Figure 4-19 O/C Trim (Pri) Options Figure 4-20 S/C Trim(Pri) Options
All freq trims at a number of frequencies, including the frequency set when the trim was
initiated. For most measurements made using standard test leads and fixtures this is the normal
trim option to use. The other trim options are normally only used in exceptional circumstances,
such as when a special test fixture fails O/C or S/C trim at certain frequencies outside of the
component test parameters.
Spot trim trims only at the frequency set in the last selected mode.
<= 10kHz trims at a number of frequencies up to and including 10kHz.
<= 100kHz trims at a number of frequencies up to and including 100kHz.
Abort cancels the trim and displays the CALIBRATE MODE main screen.
Note:
If, after trimming with an option other than All freq, a measurement frequency is selected
which is outside of the trim parameters, or will be
displayed at the top of the screen and no trim corrections will be applied for the frequency
selected. The analyzer can be used without trim correction but full measurement accuracy will
not be available until the analyzer is retrimmed using an option which covers the new
measurement frequency.
measurement leads, connect them to the Primary and Secondary BNC connectors as
appropriate.
4) Open-circuit the Kelvin clips (see Figure 4-15) or fixture jaws.
5) Select the trim option required and wait until the analyzer has finished trimming. The trim
options are described in section 4.3.1.1.
WARNING!
If the optional INSULATION MODE is fitted, part of the O/C Trim (Pri–Sec)
performs an INSULATION MODE trim. The test connections are at a high voltage
during this trim. While high voltage is applied, a message, shown in Figure 4-21
below, will be displayed on the screen. To minimize the risk of electric shock, AVOID
TOUCHING THE TEST CONNECTIONS during the trimming operation.
return the analyzer to that mode). The analyzer will enter CALIBRATE MODE, shown in
Figure 4-17.
2) Select the HF Lead Compensation soft key.
The following message will be displayed:
3) Assuming that O/C and S/C trims have already been performed, connect the transfer
standard capacitor supplied with the instrument to the Kelvin clips or fixture jaws.
4) Press the START soft key. When the CALIBRATE MODE main screen is redisplayed,
HF lead compensation has finished.
Notes:
1) If the HF lead compensation routine fails for any reason, e.g. a test lead connection error or
a power failure during the routine, Figure 4-18 (above) may be displayed when the
instrument is next switched ON, and Figure 4-23 will be displayed every time the
instrument is switched ON. When component measuring modes are selected,
will be displayed at the top of the screen when the measurement
frequency is set to 200kHz or above. These messages will only be cleared by successfully
performing the HF lead compensation routine preceded, if necessary, by O/C and S/C trims.
The instrument can be used with the default settings but it is recommended that the trims
are run for full measurement accuracy.
Operation 4–21
2) If the instrument is switched to HANDLER MODE and then back to any other mode,
will be displayed at the top of the screen when the measurement
frequency is set to 200kHz or above. This message will only be cleared, and full
measurement accuracy restored, by successfully performing the HF lead compensation
routine preceded, if necessary, by O/C and S/C trims.
Note:
If the self calibration routine fails for any reason, e.g. a test lead connection error or a power
failure during the routine, or if the self calibration data becomes corrupted, Figure 4-18 (above)
may be displayed when the instrument is next switched ON, followed by Figure 4-25, which
will be displayed every time the instrument is switched ON. When component measuring
modes are selected, will be displayed at the top of the screen. These
messages will only be cleared by performing self calibration. The instrument can be used with
the default settings but it is recommended that self calibration is run for full measurement
accuracy.
Notes
1) If the instrument is switched to HANDLER MODE and then back to any other mode,
will be displayed at the top of the screen when the measurement
frequency is set to 200kHz or above. This message will only be cleared, and full
measurement accuracy restored, by successfully performing the HF lead compensation
routine preceded, if necessary, by O/C and S/C trims.
2) If the instrument is switched to HANDLER MODE and then back to any other mode, and
is then switched OFF, Figure 4-18 may be displayed when the instrument is next switched
ON, and Figure 4-23 will be displayed every time the instrument is switched ON. When
component measuring modes are selected, will be displayed at the top
of the screen when the measurement frequency is set to 200kHz or above. These messages
will only be cleared by successfully performing the HF lead compensation routine
preceded, if necessary, by O/C and S/C trims. The instrument can be used with the default
settings but it is recommended that the trims are run for full measurement accuracy.
3) Self Calibration is also an option from HANDLER CAL MODE. This is run without
leads or a fixture connected to the instrument front panel and is therefore valid for all
modes and all types of tests. The self calibration routine is also available from the normal
CALIBRATION MODE menu: see section 4.5.
4) Select the trim option required and wait until the analyzer has finished trimming. The
HANDLER MODE trim options are described in section 4.6.1.1.
Note
If the instrument is switched OFF during O/C trim or S/C trim, the message shown in Figure 4-18
will be displayed when the instrument is next switched ON, followed by Figure 4-23, which will be
displayed every time the instrument is switched ON. The component measurement modes will be
reset to the default settings and or will be displayed at the
top of the screen. Figure 4-28 will be displayed whenever the instrument is switched ON in
HANDLER MODE after power is interrupted during O/C Trim or S/C Trim. These messages will
only be cleared by performing the appropriate trims. The instrument can be used with the default
settings but it is recommended that O/C trim and/or S/C trim is run for full measurement accuracy.
Figure 4-29 HANDLER CAL MODE O/C Trim Options Figure 4-30 HANDLER CAL MODE S/C Trim Options
All freq trims at a number of frequencies, including the frequency set when the trim was
initiated. For most measurements this is the normal trim option to use. The other trim options
are normally only used in exceptional circumstances, such as when a fixture or scanner fails
O/C or S/C trim at certain frequencies outside of the component test parameters.
Spot trim trims only at the frequency set in HANDLER MODE.
<= 10kHz trims at a number of frequencies up to and including 10kHz.
<= 100kHz trims at a number of frequencies up to and including 100kHz.
Abort cancels the trim and displays the HANDLER CAL MODE main screen.
Note:
If, after trimming with an option other than All freq, a measurement frequency is selected
which is outside of the trim parameters, or will be
displayed at the top of the screen and no trim corrections will be applied for the frequency
selected. The analyzer can be used without trim correction but full measurement accuracy will
not be available until the analyzer is retrimmed using an option which covers the new
measurement frequency.
b) Select CALIBRATE from HANDLER MODE (in which case pressing the bottommost
soft key, which will be labelled HANDLER, will return the analyzer to HANDLER
MODE). The analyzer will enter HANDLER CAL MODE (Figure 4-26).
2) Select the HF Lead Compensation soft key.
The following message will be displayed:
3) Assuming that O/C and S/C trims have already been performed, connect the transfer standard
capacitor supplied with the instrument to the fixture or scanner component contacts.
4) Press the START soft key. When the CALIBRATE MODE main screen is redisplayed,
HF lead compensation has finished.
See Notes 1) and 2) of section 4.6.
3) Use the soft keys, shown in Figure 4-4 and Figure 4-5, to set the required measurement
parameters: these are described in section 4.7.2 below. Do not exceed the limitations of the
component to be measured.
4) Connect the component to be measured to the test leads or fixture.
5) If the analyzer is in Repetitive mode, the measured values will be displayed and updated
according to the Speed setting selected. A continuously flashing asterisk (*) in the top left-
hand-corner of the screen indicates that the analyzer is in repetitive mode.
6) If the analyzer is in Single shot mode, the front panel Trigger key must be pressed to
initiate a measurement; alternatively, a suitable trigger pulse may be applied to the
TRIGGER IN socket on the rear panel, see section 4.1.5. If the Trigger key is pressed and
held, the analyzer will make repetitive measurements at the Speed setting selected until the
key is released. When in single shot mode, the asterisk (*) in the top left-hand-corner of the
screen only flashes when a measurement is triggered.
4.7.1 Example
This example will take the user through the process of measuring the inductance (L) and quality
factor (Q) of a 100µH inductor. The settings used are examples only and the user may
substitute other settings, subject to the limitations of the component to be measured.
The analyzer should be powered up with the test leads or fixture connected to the front panel
BNC connectors. If the test leads or fixture have been changed since the analyzer was last used,
they should be trimmed as described in section 4.3. If measurements will be made at 200kHz
and above, repeat the HF lead compensation as described in section 4.4.1.
1) Press the front panel Menu control key. The MAIN MENU will be displayed.
2) Press the IMPEDANCE soft key. IMPEDANCE MODE will be displayed.
3) Ensure that the analyzer is in Repetitive mode (if there is no continuously flashing asterisk (*) in
the top left-hand-corner of the screen press the front panel Sngl/Rep control key—the analyzer will
briefly indicate which mode it is entering (shown in Figure 4-10 and Figure 4-11)).
4–28 Operation
4) Use the soft keys to select the following parameters. Pressing the soft keys will either
toggle between two options or, where more than two options are available, scroll through
the options from left to right, one option at a time.
AC Meas
L
Q
Series
5) Using the navigation keys, highlight and set each of the following parameters in turn. Use
the and navigation keys to highlight a parameter and the and navigation keys to
alter the highlighted parameter setting. Settings may be altered one step at a time, or
continuously by holding the navigation key down.
100mVac
10.000kHz
DC Bias OFF set with the front panel Bias control key (shown in Figure 4-4 and Figure 4-7)
NORM can be set to BOOST only when a 3265B External Bias Unit is connected
Range Auto
Speed Med
ALC on
6) Connect the component to be measured to the test leads or fixture. The screen will display
the measured values of L and Q. The display should be similar to Figure 4-33 below.
Figure 4-33 Example Display when Measuring the Inductance and Quality Factor of a 100µH Inductor
Operation 4–29
Parallel/Series Parallel or Series equivalent circuit. This soft key is not available
when Z is selected: see the narrative on L C Z (above).
Show Scale/Hide Scale Toggles between Show Scale and Hide Scale. The selection either
shows a diagram of the equivalent circuit, i.e. Parallel or Series,
or shows a bar graph representation of either of the measurement
terms (selectable by setting the nominal and limits, see Abs %
below). The bar graph scale can either be used as a quick visual
verification that the component is within the limits set, or can be
used for adjustment of variable components. When the
measurement falls within the centre band the analyzer reports
PASS; when the measurement falls above or below the centre band
the analyzer reports HI or LO.
Notes:
1) The centre portion of the scale length is proportional to the
measured value, but scale compression is used above and
below the centre band.
2) If the binning option is fitted, an external output is available to
indicate PASS or FAIL, see sections 4.1.11 and 4.1.11.5 for
details.
4–30 Operation
Abs % Only available when the bar graph scale is displayed. Toggles
between Abs and %. When Abs is selected, absolute Hi and Lo
limits (i.e. units of the measured parameter) are displayed. When
% is selected, a nominal value together with Hi and Lo percentage
limits are displayed.
The limits and nominal value (if applicable) must be set using the
and navigation keys to highlight each parameter and the data
entry keypad to set each value (the use of the data entry keypad is
described in section 4.2.6). When in % mode, the bar graph scale
Hi and Lo limits can easily be set equidistant about the nominal by
setting either of the limits then highlighting the other limit and
pressing the keypad Enter key twice. This mimics the setting of
the other limit but with the opposite sign.
Save Nom Only available when the bar graph scale is displayed. If a standard
component exists, it can be connected to the test leads or fixture
and measured by the analyzer. Pressing Save Nom (when either
abs or % is selected) enters the most recent analyzer measurement
of the component as the nominal test value for comparing all
subsequent components with against % limits.
Notes:
1) To change this function from the first to the second measured
parameter (or vice versa), first enter a dummy value with units
via the keypad; e.g. to change from L to R, enter
[1] [units] [Ω] [Enter] then press the Save Nom key.
2) Do not use the Save Nom function if the measured value is
negative (e.g. an inductor measured above its self resonant
frequency).
Show Setup/Hide Setup Once the measurement parameters have been set, Hide Setup can
be selected to clear them from the screen. The parameter settings
are still valid and will be used for component measurements. The
bar graph scale and limits will still be displayed. Hide Setup is
used primarily to unclutter the display, making it more easily
readable. Selecting Show Setup will redisplay the parameter
settings.
The following IMPEDANCE MODE parameters are those displayed in the bottom left-hand-
corner of the screen, shown in Figure 4-34. They are only visible when Hide Setup is NOT
SELECTED.
Drive Level Only available when AC Meas is selected. Set by highlighting the
parameter with the and navigation keys, then altering the
setting in pre-determined steps with the and navigation
keys, or by using the data entry keypad. The range is:
Rdc Meas mode Drive Level not displayed—Fixed at 100mV
(short circuit current 10mA)
AC Meas mode Variable between: 1mV–10V
or 50µA–200mA
(appropriate for low
impedance components)
See also ALC, below
Measurement Frequency Set by highlighting the parameter with the and navigation
keys, then altering the setting in pre-determined steps with the
and navigation keys, or by finer increments using the data entry
keypad. The range is:
20Hz to 3MHz
Fine or coarse frequency steps are available. Coarse steps vary in
increments of between 20% and 33%; fine steps vary in
increments of 1% or less. Set fine or coarse steps from the
SETTINGS page—see section 5.18, or use code 10 (fine steps) or
code 11 (coarse steps)—see section 4.2.6
4–32 Operation
DC Bias The DC Bias is turned ON and OFF with the Bias control key.
Before bias can be turned ON, the level must be set by
highlighting the parameter with the and navigation keys, then
altering the setting in pre-determined steps with the and
navigation keys, or by finer increments using the data entry
keypad. 1mA to 1A is available internally (NORM mode). With
one or more 3265B External Bias Units connected (max 5), the
unit may be toggled between NORM (internal bias) and BOOST
(external bias). Up to 25A per 3265B is available in BOOST
mode. For more information see section 4.2.5.
Range Toggles between auto range and manual range selection, set by
highlighting the parameter with the and navigation keys and
altering the setting with the and navigation keys. Auto
range automatically selects the most accurate range for the
measurement. Circumstances where manual ranging may be more
appropriate include:
• measuring non-linear components (auto range may hunt);
• to avoid the short auto range delay, for example when using
max speed with an auto handler.
The manual range is set using the data entry keypad. Ranges 1 to 7
are valid. When a manual range is selected, the equivalent
measurement range is shown on the display: although range
boundaries are impedance values they are converted to appropriate
L or C values.
At higher frequencies or reduced levels, availability of the highest
or lowest ranges is restricted. If a previously selected range is
changed due to a change in drive conditions, the selection will be
remembered by the analyzer and reapplied when drive conditions
allow it.
Speed Four measurement speeds are available: Slow, Med, Fast and
Max. Selecting slower measurement speeds increases the display
resolution and decreases measurement noise by averaging. The
measurement speed is set by highlighting the parameter with the
and navigation keys and altering the setting with the and
navigation keys.
The following measurement periods apply for Rdc Meas and AC
Meas ≥100Hz:
Max speed makes measurements at ≈40ms intervals and is
intended for automatic sorting under GPIB control.
Fast speed makes measurements at ≈100ms intervals and is
intended for non-critical measurements.
Med speed makes measurements at ≈300ms intervals and gives
Operation 4–33
drive source. Therefore, the drive level at the fixture jaws will rise
when a component is removed and will return to the correct level
when a new component of the same impedance is inserted
( may be reported at the top of the display with no
component in the fixture jaws). The range is also maintained at
that used for measuring the representative component. When using
ALC hold therefore, the analyzer does not spend time computing
the correct ALC compensation and range, thus making
measurements quicker.
Advanced Operation 5–1
5. ADVANCED OPERATION
This section will provide the user with a guide to:
• front panel connections;
• in-circuit measurements;
• measurement of very small capacitors;
• measurement of very small inductors;
• measurement of iron-cored and ferrite inductors;
• measurement of transformers;
• using the various modes available from the MAIN MENU, with the exception of
CALIBRATE MODE which is covered in sections 4.3 to 4.6, and IMPEDANCE MODE
which is covered in section 4.7.
leads marked not used may be left connected to the analyzer but are not used for the insulation test
Connect one lead set to transformer primary Connect 2nd lead set to transformer secondary
If the DUT has a large area of metal not connected to either of its terminals (e.g. a screen or
core), this should be separately connected to ground using the green clip lead; but if there is a
relatively large unscreened conducting surface which is connected to one of its terminals (e.g.
an air-spaced tuning capacitor), this should be connected to the ORANGE signal source lead to
minimize noise pick-up.
Measurement connections to transformers are simplified by providing separate primary and
secondary connections. These are automatically reconfigured by the instrument to suit the
selected test mode.
The front panel BNC LEDs indicates which BNC connections are active in each case
The presence of Zd introduces a small measurement error, dependant on the frequency and
impedance range in use. Figure 5-5 shows the minimum shunt impedance (i.e. R, ωL or ωC) for
an additional error (magnitude or phase) not exceeding 1%. Note that when measuring high
impedances it may be beneficial to use a drive level greater than 1V or to manually select a
lower measurement range (see section 4.7.2—IMPEDANCE MODE Parameters).
5–4 Advanced Operation
The main effect of adding Zs alone is to reduce the available drive signal. When measuring
high impedances, this effect is dominated by the fixed 50Ω output impedance of the signal
source. For example, a shunt resistance of 50Ω may be expected to halve the available output
level. When measuring components with an impedance below 50Ω, the degree of reduction will
be less. Note that when ALC is turned ON (see section 4.7.2) the displayed level will always
correspond to the actual level at the measurement terminals.
When Zd and Zs are connected simultaneously, an additional measurement error occurs due to
the impedance of the guard lead (Zg). This error may become significant if the DUT is larger
than Zd and Zs, and is given by
DUT x Zd
Error % = 100 x
Zs x Zd
At low frequencies Zg is up to 40mΩ for lead types 1EVA40100, 1EVA40180 or 1EV1505. At
frequencies above 10kHz the series inductance, which depends to some extent on lead and
component positioning, may become significant. For lowest inductance, minimize the area of
the loop formed by the Red (current detector) lead, via Zd and the Green lead to neutral. In this
case the inductance should not exceed 0.25µH.
Note that at low frequencies (<2kHz) the effective guard resistance can be reduced by a factor
of 2:1 or more by returning Zs and Zd directly to the outer of the Red BNC connector. However
this technique increases the loop inductance and any benefit is lost at frequencies above 5kHz.
may be selected. Use 4-terminal measurement if the primary impedance is particularly low. If
accurate level control is required, select ALC on.
5.7.1 Example
This example will take the user through the process of making a 2-terminal measurement of the
turns ratio of a transformer. The settings used are examples only and the user may substitute
other settings, subject to the limitations of the component to be measured. The analyzer should
be powered up with the test leads or fixture connected to the front panel BNC connectors
according to the connection protocol for transformer mode (see Figure 5-2).
1) Press the front panel Menu control key. The MAIN MENU will be displayed.
2) Press the TRANSFORMER soft key. TRANSFORMER MODE will be displayed.
3) Press the CALIBRATE soft key and perform the trims listed in section 5.7 paragraph 5).
When the trims are complete, press the TRANSFORMER soft key to return to
TRANSFORMER MODE.
4) Ensure that the analyzer is in Repetitive mode (if there is no continuously flashing asterisk
(*) in the top left-hand-corner of the screen press the front panel Sngl/Rep control key—the
analyzer will briefly indicate which mode it is entering (shown in Figure 4-10 and Figure
4-11).
5) If the 2/4 Term LED is not lit, press the 2/4 Term control key. The 2/4 Term LED should
light and 2-TERM TRANSFORMER MODE should be displayed at the top of the screen.
6) Press the Turns Ratio soft key.
7) Using the navigation keys, highlight and set each of the following parameters in turn. Use
the and navigation keys to highlight a parameter and the data entry keypad or and
navigation keys to alter the highlighted parameter setting. Settings may be altered one
step at a time, or continuously by holding the navigation key down.
100mVac
10.000kHz
Ratio: Np/Ns
Range Auto
Speed Med
ALC off
8) Connect the component to be measured to the test leads or fixture. The screen will display
the measured value of turns ratio.
5–8 Advanced Operation
L + Q (Pri) Measures the inductance (L) and quality factor (Q) of the
transformer primary winding. The Equ cct parameter allows
selection of either Parallel or Series resistance in the transformer
primary winding. When Equ cct is set to Auto, the analyzer makes
a measurement of the primary winding resistance and sets series or
parallel equivalent circuit according to the following criterion:
Primary winding resistance >250Ω = Parallel equivalent circuit
Primary winding resistance <250Ω = Series equivalent circuit
Turns Ratio Measures the turns ratio of a transformer. Turns ratio can be
displayed as Np/Ns or Ns/Np. When maximum resolution is
required, choose the display which is greater than unity. The value
displayed is the ratio of measured voltages, hence non-integer
results are likely. Trimming and high frequency lead compensation
are not required for turns ratio measurements.
A negative reading when measuring turns ratio implies a reverse-
connected winding. Check the winding sense convention.
Secondary turns (Ns) can be displayed. This requires the primary
turns (Np) to be preset via the data entry keyboard. Allowable
values of Np are between 0.001 and 10000.
For accurate measurement of turns ratio, software correction can
Advanced Operation 5–9
Leakage (Pri) The series leakage inductance and resistance of the transistor
primary winding. Best results will be obtained by short-circuiting
the transformer secondary winding(s). A message is displayed to
this effect when the mode is initially selected. Leakage (Pri) and
Leakage (Sec) tests share the same measurement conditions.
Leakage (Sec) The series leakage inductance and resistance of the transistor
secondary winding. Best results will be obtained by short-
circuiting the transformer primary winding. A message is
displayed to this effect when the mode is initially selected.
Leakage (Pri) and Leakage (Sec) tests share the same measurement
conditions.
The following TRANSFORMER MODE parameters are those displayed in the bottom left-
hand-corner of the screen, shown in Figure 5-6. They are only visible when Hide Setup is NOT
SELECTED.
5–10 Advanced Operation
repetitive measurements until the key is released. To minimize the risk of electric shock, high
voltages are applied to the test leads only during testing.
WARNING!
Unless global measurement conditions are selected (see section 5.18—The SETTINGS Page),
the test voltage and other settings may be set differently for each of the three test types.
The insulation test cannot be performed if the test current exceeds 1mA, i.e. if the impedance is
too low. This corresponds to a minimum resistance of 100kΩ at 100V, 200kΩ at 200V or
500kΩ at 500V. In this case a message, shown in Figure 5-7, will be displayed.
Note
If the INSULATION soft key is pressed when the insulation option is not fitted, the analyzer
will report Unit Not Available (Figure 5-8).
5.8.1 Connections
Except when using a handler or 4-wire scanner, the standard connection protocol for
TRANSFORMER/TELECOM MODE may be used for all insulation test types. This is
shown in Figure 5-2. The tables below show the active connection for each test. Leads marked
not used may be left connected to the analyzer but are not used for the insulation test.
For operation with a 4-wire scanner the scanner relays should be configured as for turns ratio
measurement. During Insulation mode operation, set the instrument to 2-terminal operation.
Since a 4-wire scanner is connected to the analyzer with HANDLER MODE lead
configuration (see Figure 5-3), it is not possible to perform a primary to secondary winding test.
5–12 Advanced Operation
Red
LEADS not used not used not used not used not used not used not used
Drive Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the RED lead to the transformer primary, and the GREEN CLIP LEAD to the transformer metal case or shell.
Brown
LEADS not used not used not used not used not used not used not used
Sense Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the BROWN lead to the transformer primary, and the GREEN CLIP LEAD to the transformer metal case or
shell.
Red
LEADS not used not used not used not used not used not used not used
Drive Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the RED lead to the transformer secondary, and the GREEN CLIP LEAD to the transformer metal case or shell.
Brown
LEADS not used not used not used not used not used not used not used
Sense Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the BROWN lead to the transformer secondary, and the GREEN CLIP LEAD to the transformer metal case or
shell.
Red Red
LEADS not used not used not used not used not used not used
Drive Low Drive Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the Primary RED lead to the transformer primary, and the Secondary RED lead to the transformer secondary.
Brown Brown
LEADS not used not used not used not used not used not used
Sense Low Sense Low
Leads marked not used may be left connected to the analyzer but are not used for the insulation test.
Connect the Primary BROWN lead to the transformer primary, and the Secondary BROWN lead to the transformer
secondary.
5.8.2 Trimming
For INSULATION MODE operation, trimming compensates for any residual leakage current
in the test leads or fixture. For maximum accuracy the trimming operation should be performed
with the test leads connected to the instrument but isolated from each other.
WARNING!
5–14 Advanced Operation
The test connections are at a high voltage during this trim. While high voltage is applied,
a message, shown in Figure 5-15 below, will be displayed on the screen. To minimize the
risk of electric shock, AVOID TOUCHING THE TEST CONNECTIONS DURING THE
TRIMMING OPERATION.
Notes:
1) The stored trim values will be applied during 2-terminal insulation tests, even if AC
impedance and Rdc trimming correspond to 4-terminal operation.
2) The Insulation Trim soft key is not available when CALIBRATE is selected from
INSULATION MODE.
3) If Insulation Trim is attempted with Kelvin clip leads attached to the analyzer in
HANDLER MODE configuration, the trim will fail because the RED and BROWN leads
are shorted by the Kelvin clip.
5.8.3 Operation
The general procedure when using INSULATION MODE is as follows.
1) Select the Menu control key, followed by the INSULATION soft key.
Advanced Operation 5–15
2) Connect the measurement leads to the analyzer according to the connection protocol
detailed in section 5.8.1.
3) Select 2- or 4-terminal measurement.
4) Refer to section 5.8.2 and trim the leads/fixture.
5) Select the insulation test required, i.e. Pri–Sec, Pri–GND or Sec–GND, with the
appropriate soft key.
6) If the setup parameters are hidden, press Show Setup.
7) Change the setup parameters, i.e. Level, Range and Speed, as necessary for the test.
8) Connect the measurement leads/fixture to the transformer (see section 5.8.1—
Connections).
9) Select the measurement unit to display with the µA MΩ soft key.
10) Press the Trigger control key to initiate a measurement.
5.8.4 Example
This example will take the user through the process of performing all three insulation test types
on a transformer. The settings used are examples only and the user may substitute other settings
as required. The analyzer should be powered up with two sets of Kelvin clip leads connected to
the Primary and Secondary front panel BNC sockets according to the connection protocol
shown in Figure 5-2 .
1) Press the front panel Menu control key. The MAIN MENU will be displayed.
2) Press the INSULATION soft key. INSULATION MODE will be displayed.
3) Select either 2- or 4-terminal measurement, as required for subsequent tests, with the 2/4
Term control key. For this example, 4-terminal measurement will be selected.
4) Press the CALIBRATE soft key and perform an O/C Trim (Pri–Sec). If necessary refer to
section 4.3.2 for trimming details. When the trim is complete, press the INSULATION soft
key to return to INSULATION MODE.
WARNING!
The test connections are at a high voltage during this trim. While high voltage is
applied, a message, shown in Figure 5-15 above, will be displayed on the screen. To
minimize the risk of electric shock, AVOID TOUCHING THE TEST CONNECTIONS
DURING THE TRIMMING OPERATION.
Level 500V this can also be set with the data entry keypad.
Range Auto
Speed Med
8) Connect the Primary BROWN BNC connector to one end of the transformer primary
winding. Connect the Secondary BROWN BNC connector to one end of the transformer
secondary winding. The LEDs above the BROWN BNC connectors should be lit to show
that they are the active connections. There should be no other connections to the
transformer.
9) Press the Trigger control key to initiate the measurement. Figure 5-15, above, will be
displayed while high voltage is applied to the measurement terminals and the measurement
result will be displayed.
WARNING!
10) Press the Pri–GND soft key and, if necessary, re-establish the measurement conditions
outlined in steps 6) and 7).
11) Connect the Primary BROWN BNC connector to one end of the transformer primary
winding. Connect the GREEN CLIP LEAD to the transformer metal case or screen The
LED above the Primary BROWN BNC connector should be lit to show that it is the active
connection. There should be no other connections to the transformer.
12) Press the Trigger control key to initiate the measurement. Figure 5-15, above, will be
displayed while high voltage is applied to the measurement terminals and the measurement
result will be displayed.
13) Press the Sec–GND soft key and, if necessary, re-establish the measurement conditions
outlined in steps 6) and 7).
14) Connect the Secondary BROWN BNC connector to one end of the transformer secondary
winding. Connect the GREEN CLIP LEAD to the transformer metal case or screen The
LED above the Secondary BROWN BNC connector should be lit to show that it is the
active connection. There should be no other connections to the transformer.
15) Press the Trigger control key to initiate the measurement. Figure 5-15, above, will be
displayed while high voltage is applied to the measurement terminals and the measurement
result will be displayed.
Advanced Operation 5–17
The following INSULATION MODE parameters are those displayed in the bottom left-hand-
corner of the screen, shown in Figure 5-16. They are only visible when Hide Setup is NOT
SELECTED.
Level The DC voltage level used for the insulation test. This parameter is
selected with the and navigation keys and is altered with the
and navigation keys or with the data entry keypad. Levels
of 100V, 200V or 500V are available.
In Figure 5-17 above, the bins are shown ready to receive percentage limits. Absolute limits can
be entered by highlighting Abs with the Abs % soft key. Either nested or stacked limits can be
entered. When entering nested limits, the second limit in a row can be entered by highlighting it
and pressing the keypad Enter key twice. This mimics the setting of the first limit but with the
opposite sign. Since the analyzer will accept one set of percentage limits and another set of
absolute limits it is possible to enter a set of nested percentage limits and a set of stacked
absolute limits, or vice versa. The only limitation to this is that the minor term, when displayed,
is common to both percentage and absolute limits.
When using percentage limits, the nominal component value must also be entered by pressing
the Nominal soft key, then entering the nominal component value with the data entry keypad.
Any limit set to zero is ignored during subsequent sorting. Bins can therefore be set up with no
minor term limit.
The minor term limit is a single maximum or minimum value as appropriate, e.g. Q would have
a minimum limit, indicated by Minor Q > x.
The example shown in Figure 5-18 above would sort transformers on the basis of their %
deviation from primary winding inductance and Q factor. In this case, a transformer with a
5–20 Advanced Operation
primary winding inductance of 100mH ±0.5% and a Q factor greater than 0.5 would be sorted
into bin 2.
Figure 5-19 is an alternative set-up with stacked limits, sorting the transformers by absolute
value. In this case, a transformer with a primary winding inductance of 112mH and a Q factor
greater than 0.5 would be sorted into bin 3.
Nominal This soft key operates only when percentage limits are displayed.
The nominal value is displayed below the percentage limits and
can be seen, set to 100mH, in Figure 5-18. The nominal is set by
pressing the Nominal soft key and entering the value using the
data entry keypad.
Advanced Operation 5–21
Set Test Pressing the Set Test soft key displays the available measurement
options. The required option is highlighted with the and
navigation keys and selected by pressing the Enter test soft key.
Set Bin Moves the cursor, and confines it to, the bin limits area of the
screen. The limits are set by highlighting each field using the
navigation keys, then entering the limits with the data entry
keypad. The and navigation keys move the cursor left and
right one step at a time; the and navigation keys move the
cursor up and down the column. See section 5.9.1 for more
information about setting bin limits.
Set Condition Moves the cursor, and confines it to, the test conditions area of the
screen (bottom-left-corner). The test conditions are set by
highlighting the parameter with the and navigation keys, then
altering the setting in pre-determined steps with the and
navigation keys, or by using the data entry keypad.
CALIBRATE See section 4.3. Note: this is not a binning calibration.
Note: If the BINNING soft key is pressed when the binning option is not fitted, the analyzer
will report Unit Not Available (Figure 5-22).
Selecting the Yes soft key will delete the last result.
Advanced Operation 5–23
1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.
2) Use the and navigation keys to scroll through the list and press the Load soft key
when the desired program is highlighted.
5–26 Advanced Operation
3) To return to the SEQUENCE EDIT MODE press the Edit soft key.
1) Display the SEQUENCE FILE screen. See section 5.10.1.1 Sequence File Screen.
2) Select the New soft key.
3) Type a unique program number between 1 and 65535, followed by the keypad Enter key.
A single step test program will be created assigned the number entered.
1) Open the program to be edited. See section 5.10.1.2 Loading an Existing Program.
2) Use the and navigation keys to select the insertion point of the new test step in the
test program. The new step will be added to the program after the highlighted test step.
3) Press the Add soft key to list the available tests
4) Use the and navigation keys to highlight the type of test to be added.
5) Pressing the Ok soft key adds the test step to the program.
Notes:
• New primary winding test steps are always added after the last primary winding test
in the list.
• New secondary winding test steps are always added after the last secondary winding
test.
• New test steps mimic the settings of the step which was highlighted when the new
step was added.
Advanced Operation 5–29
1) Open the program to be edited. See section 5.10.1.2 Loading an Existing Program.
2) Use the and navigation keys to highlight the setup or limit parameter to be changed.
3) Parameter settings are modified using the and navigation keys or numeric keypad.
Some parameters, e.g. frequency, are changed by entering the new value from the data entry
keypad; other parameters, e.g. speed, are changed by repeatedly pressing the keypad Enter
key to scroll through the available settings.
1) Open the program to be edited. See section 5.10.1.2 Loading an Existing Program.
2) Use the and navigation keys to highlight the test step to be moved.
3) Press the soft key labelled Move and use the Up and Down softkeys to reposition the test
step.
4) Complete the move of the test step by pressing the Ok softkey.
If the transformer passes all test steps, PASS will be reported at the end of the test run. Failures
are reported on a test-by-test basis, with a failure message, such as that shown in Figure 5-35,
reported in each case. When a failure occurs, various options are available.
a) Select the RETRY soft key to rerun the failed test step.
b) Press the Trigger control key to continue the sequence program.
c) Select the NEXT TRANSFORMER soft key to discard any remaining test steps and
allow connection of the next transformer to be tested.
5–32 Advanced Operation
Selecting the Stop Running PROGRAM No. X soft key will allow a new sequence program to
be selected
Two sets of measurement leads are required; see Figure 5-2 for the connection protocol. To
meet the specified accuracy, 4-terminal connection must be used. The general procedure when
using TELECOMS MODE is as follows.
1) Select the Menu control key, followed by the TELECOM soft key.
2) Connect the measurement leads, see Figure 5-2.
3) Select 4-terminal measurement.
4) From TELECOMS MODE select the CALIBRATE soft key.
5) Refer to section 4.3 and perform the following trims.
a) O/C Trim (Pri)
b) S/C Trim (Pri)
c) O/C Trim (Pri–Sec)
d) S/C Trim (Sec)
6) Select the TELECOM soft key to return to TELECOMS MODE.
7) Select the terminating options required..
8) Change the setup parameters as required, i.e. drive level, frequency and terminating
component values, being careful not to exceed the limitations of the transformer.
9) Connect Primary and Secondary leads to the appropriate transformer windings.
10) Press the Start soft key to initiate the test. The measured values of insertion loss (I.L.) and
return loss (R.L.) will be displayed on the screen.
5.13.1 Example
This example will take the user through the process of measuring the insertion loss and return
loss of a line matching transformer. The settings used are examples only and the user may
substitute other settings, subject to the limitations of the component to be measured. The
analyzer should be powered up with the test leads or fixture connected to the front panel BNC
connectors according to the connection protocol for transformer/telecom mode (see Figure 5-2).
1) Press the front panel Menu control key. The MAIN MENU will be displayed.
2) Press the TELECOM soft key. TELECOMS MODE will be displayed.
5–34 Advanced Operation
3) Press the CALIBRATE soft key and perform the trims listed in section 5.13 paragraph 5).
When the trims are complete, press the TELECOM soft key to return to TELECOMS
MODE.
4) Using the Simple/Damped soft key, select Damped termination.
5) Using the Direct/Blocked soft key, select Blocked termination.
6) Using the navigation keys, highlight and set each of the following parameters in turn. Use
the and navigation keys to highlight a parameter and the data entry keypad or and
navigation keys to alter the highlighted parameter setting. Settings may be altered one
step at a time, or continuously by holding the navigation key down.
-10.0dBm (default value)
4.0000kHz
Z0 : 600.0 Ω (default value)
Rt : 600.0 Ω (default value)
Ct : 4.700nF (default value)
Rd : 600.0 Ω (default value)
Cd : 4.700nF (default value)
Cb : 1.000MF (default value)
7) Connect the transformer to the test leads or fixture.
8) Press the Start soft key. The measured values of insertion loss (I.L.) and return loss (R.L.)
will be displayed on the screen.
Figure 5-39 Example of Measuring Insertion Loss and Return Loss in TELECOMS MODE
The following TELECOMS MODE parameters are those displayed in the bottom left-hand-
corner of the screen, shown in Figure 5-39.
Drive Level Set by highlighting the parameter with the and navigation
keys, then altering the setting in pre-determined steps with the
and navigation keys, or by using the data entry keypad. The
range is:
–28.0dBm to 16.0dBm.
Measurement Frequency Set by highlighting the parameter with the and navigation
keys, then altering the setting in pre-determined steps with the
and navigation keys, or by finer increments using the data entry
keypad. The range is:
100Hz to 20kHz
Fine or coarse frequency steps are available. Coarse steps vary in
increments of between 20% and 33%; fine steps vary in
increments of 1% or less. Set fine or coarse steps from the
SETTINGS page—see section 5.18, or use code 10 (fine steps) or
code 11 (coarse steps)—see section 4.2.6.1
Up to eight frequencies can be defined by highlighting the frequency, then entering the
frequency with the data entry keypad. The and navigation keys scroll through each
frequency in turn. Also available, depending upon the setting of the Off Abs % soft key, are
High, Low and Minor term limits and a Nominal parameter. The High, Low, Minor and
Advanced Operation 5–37
Nominal settings are accessed by pressing either of the or navigation keys when one of the
frequency settings is highlighted. The Nominal value is common to all frequencies but the
High, Low and Minor term limits may be different for each frequency set.
Any limit set to zero is ignored when the multi-frequency test is run. Therefore either the major
or minor term test may be omitted by setting the appropriate limits to zero.
5.14.1.1 Example
This example will illustrate the procedure for setting MULTI FREQ parameters using
different limits for each set frequency. The sequence used in this example is not the only way to
set the parameters but is intended to familiarize the user with this mode of operation. For this
illustration, percentage limits will be used.
1) Enter MULTI FREQ – Set mode by pressing the MULTI FREQ soft key from the MAIN
MENU. If MULTI FREQ – Run mode is displayed, press the SET soft key. If no
parameters have previously been set, the display will look like Figure 5-40 above.
2) If the test leads or fixture have been changed since the last time the analyzer was used,
press the CALIBRATE soft key and perform the following trims with reference to sections
4.3 and 4.4.1. When finished, press the MULTI FREQ soft key to return to MULTI
FREQ – Set mode.
O/C Trim (Pri)
S/C Trim (Pri)
HF Lead Compensation
3) Use the Off Abs % soft key to highlight %. This sets the display ready to accept
percentage limits.
4) Press the IMPEDANCE soft key. This will return the instrument to IMPEDANCE
MODE where the appropriate measurement parameters must be set prior to running a
MULTI FREQ test. Enter the parameters required for the test. For this example they are
set to:
AC Meas
L
Q
Parallel
1.00Vac
1.0000kHz—this will be the first set frequency in MULTI FREQ – Set mode
DC Bias 0.000 A OFF
NORM
Range Auto
Speed Med
ALC off
Note: Where a component is to be measured over a wide frequency range, setting Range to
Auto is recommended.
When the measurement parameters have been set, press the RETURN soft key to return the
instrument to MULTI FREQ – Set mode.
5–38 Advanced Operation
5) Highlight the first frequency, shown highlighted in Figure 5-40 (the and navigation
keys scroll through each frequency in turn) and enter the required frequency with the data
entry keypad.
6) Highlight and enter the next frequency. Continue to highlight and enter up to eight
frequencies in this way. This example will enter frequencies of 1kHz, 3kHz, 10kHz, 30kHz,
100kHz, 300kHz and 1MHz.
7) With the first (top) frequency highlighted, press either of the or navigation keys until
the Nominal parameter is highlighted (if using absolute limits there is no nominal
parameter). Enter the Nominal value with the data entry keypad; for this example the
nominal will be set to 100µH.
8) Still using the and navigation keys, highlight the High limit then enter the required
limit with the data entry keypad. For this example all the limits will be set to ±10%, though
they could be set to different values for each frequency. Highlight the Low limit and enter
the required limit. Pressing the Enter key twice will echo the High limit but with the
opposite sign.
9) Highlight the Minor term with the and navigation keys and enter the required value.
Note that the Minor term limit is either an upper or lower limit depending on what the
parameter is (e.g. <D, >Q). For this example the Q term will be set to >2 at 1kHz, i.e.
anything more than or equal to 2 will pass the minor term parameter and anything below 2
will fail.
10) Press the navigation key: the symbol will move down and point to the second
frequency (3kHz in this example). Note that the limits showing at the bottom of the screen
change as each frequency is selected in turn. Using the and navigation keys highlight
and set the High, Low, and Minor limits for the second frequency. Press the navigation
key again and the symbol will point to the third frequency and the limits for the third
frequency can be set. Continue in this way until the limits have been set for each frequency.
The limits set in this example are as follows:
Sort If the frequencies entered were not in sequence, pressing the Sort
soft key will sort them into ascending order. Pressing Sort again
toggles the frequency sequence, i.e. the top frequency becomes the
bottom frequency and vice versa. The limits will stay with the
frequency they relate to.
5–40 Advanced Operation
Figure 5-43 Initial MULTI FREQ – Run Display (from example in section 5.14.1.1)
When the Start soft key or the Trigger key is pressed, the analyzer will measure the
component at the frequencies and measurement parameters previously set and the measurement
Advanced Operation 5–41
values will be displayed. If either Abs or % was selected in MULTI FREQ – Set mode, the
analyzer will report PASS, FAIL, HI or LO according to the table below. Figure 5-44 shows
the results of running the multi-frequency test set up in section 5.14.1.1.
PASS Major and minor terms are within the limits set.
FAIL Major and minor terms are outside of the limits set.
HI (X), e.g. HI D, HI C The parameter indicated is above the upper limit.
LO (X), e.g. LO L, LO Q The parameter indicated is below the lower limit.
When the bin handler option is fitted, the bin handler Pass/Fail output corresponds to the
, , and results. The Pass/Fail output goes low only when a
measurement has passed all set limits, see section 4.1.11.5 for the bin handler interface pin
assignment.
Figure 5-45 shows GRAPH MODE set up to plot a graph of inductance against frequency on a
linear scale between 100kHz and 600kHz with the limits set in absolute terms.
Lin(Z) Log(Z) This soft key is displayed only when Log(unit) is selected (above)
and when Mode is set to Impedance (Z+θ). It allows selection of
Lin or Log vertical scale, allowing log/log graphs to be plotted.
Note that negative or zero scale limits cannot be plotted on a log
scale. Both limits must be set to positive values before pressing
Start, or a warning message will be displayed.
Sweep The Start and Stop settings for the graph, which are set by
highlighting each in turn with the and navigation keys, then
entering the value with the data entry keypad. When Sweep is
highlighted the and navigation keys are used to set the
Sweep parameter to Frequency, Bias, Bias (Boost) or Drive
Level.
Advanced Operation 5–43
Step Size The Step Size is the number of LCD display pixels used for a
single point along the graph. The higher the number selected, the
coarser the graph but the more quickly it is plotted. Set by
highlighting Step Size with the and navigation keys, then
select from 1 2 4 8 with the and navigation keys.
Abs % This soft key toggles between Abs and % (major term only).
When Abs is selected, absolute Hi and Lo limits (i.e. units of the
measured parameter) are displayed. When % is selected, a
Nominal value together with Hi and Lo percentage limits is
displayed.
The limits and nominal value (if applicable) must be set using the
and navigation keys to highlight each parameter and the data
entry keypad to set each value.
Note: The Abs % soft key is not available when the vertical scale
is logarithmic.
Major/Minor This soft key toggles between either of the measurement terms
Term Select selected in IMPEDANCE MODE or TRANSFORMER MODE
depending on whether the Mode parameter is set to Impedance…
or Transformer…. The graph will be plotted according to
whichever of these terms is highlighted.
5–44 Advanced Operation
Start When all parameters have been set up, pressing the Start soft key
will plot the graph.
As the graph is being plotted, a progress marker at the bottom of
the screen shows how complete the graph is and the speed of the
marker gives an indication of how long the graph will take to plot.
If the graph is progressing too slowly, perhaps because too small a
Step Size has been selected, it can be aborted by pressing and
holding the Abort key.
Notes:
1) If the range is manually selected and the component
characteristics take the graph outside the boundaries of the
selected range, that portion of the graph will not be plotted.
2) If incorrect y axis settings are selected, no graph will be seen
even though the progress marker will indicate that a graph is
being plotted: the graph will actually be ‘plotted’ outside of
the display area. The display can be corrected by using the
FIT autoscale function, or by pressing the RETURN soft key
and selecting more appropriate limits.
3) An error message shown at the top of the screen indicates that
the error occurred somewhere on the sweep.
4) The graph is plotted using the full measurement resolution.
View When the graph has been plotted and GRAPH MODE – Set has
been reselected with the RETURN soft key, View can be used to
redisplay the last graph plotted by the analyzer.
VIEW Pressing VIEW hides the available soft keys and displays the
whole graph.
FIT The FIT soft key replots the graph, auto scaling the vertical axis
for the best available resolution. This may be done separately for
both major and minor terms.
Advanced Operation 5–45
TOGGLE When a graph is plotted on the major or minor term, the other term
is automatically stored in the background and the TOGGLE soft
key can be used to toggle between the two graphical displays.
PEAK Aligns the marker with the highest peak in the set frequency range.
DIP Aligns the marker with the lowest trough in the set frequency
range.
NOM>MK Replots the graph with the marker position set as the component
nominal value. To aid positioning of the marker, its position in
both axes is shown at the bottom of the screen. It is moved with
the and navigation keys. When GRAPH MODE – Set is re-
entered with the RETURN soft key, the Nominal and Hi/Lo
limits will reflect the replotted graph.
Note
If the DUT exhibits high Q resonances, the graph function may miss the peak or trough due to
quantized frequency steps. For accurate values at resonance, use the RESONANCE function.
5.15.3 Example
This example will illustrate the procedure for using GRAPH MODE and is intended to
familiarize the user with this mode of operation. For this example a linear scale will be used to
plot inductance over the frequency range of 100kHz to 600kHz.
1) Enter GRAPH MODE – Set by pressing the GRAPH soft key from the MAIN MENU.
2) Highlight the Sweep parameter with the or navigation key, then using the and
navigation keys set it to Frequency.
3) Highlight the Start frequency with the or navigation key and enter the required
frequency with the data entry keypad. For this example 100kHz will be entered.
4) Highlight the Stop frequency with the or navigation key and enter the required
frequency with the data entry keypad. For this example 600kHz will be entered.
5) Highlight Step Size with the or navigation key, then use the and navigation
keys to select 1 2 4 or 8. For this example 2 is selected.
6) Highlight the Mode parameter with the or navigation key, then using the and
navigation keys set it to Impedance….
5–46 Advanced Operation
7) Press the IMPEDANCE soft key. This will display IMPEDANCE MODE where the
appropriate parameters must be set prior to plotting the graph. Enter the parameters
required for the test. For this example they are set to:
AC Meas
L
R
1.00Vac
1.0000kHz—this setting will be overridden when the graph is plotted
DC Bias 0.000 A OFF NORM
Range Auto
Speed Med
ALC off
When the measurement parameters have been set, press the RETURN soft key to return the
instrument to GRAPH MODE– Set.
8) Highlight the L and R Hi and Lo limits with the or navigation key, then set each in
turn with the data entry keypad. For this example they will be set to:
L Hi: 100.00m
Lo: -100.00m
R Hi : 20.000Ω
Lo : 1.000Ω
9) Use the Lin(Hz) Log(Hz) soft key to highlight Lin(Hz).
10) Use the Abs % soft key to highlight Abs.
The screen should now look like Figure 5-45.
11) If necessary, press the CALIBRATE soft key and perform the appropriate trims for the
Mode selected. See section 4.3 for details of the trim procedure.
12) Connect the component to the test leads or fixture.
13) Press the Start soft key. A graph will be plotted showing the component characteristics
using the measurement parameters and frequency range set (Figure 5-46).
Advanced Operation 5–47
14) By stepping the marker along the graph with the and navigation keys, the measurement
result and frequency can be seen at the marker position in the bottom-left corner of the
screen.
15) Pressing the FUNCTION soft key will display the other soft key functions available. These
are described in section 5.15.2—Parameters Available when the Graph is Displayed
Figure 5-47 Series Equivalent Circuit Figure 5-48 Parallel Equivalent Circuit
At the selected resonant frequency, the capacitive and inductive reactances are equal and
opposite, giving a measured impedance that is purely resistive. By making measurements above
and below this frequency, the exact resonant characteristics are found by interpolation. For
resonances with reasonably high Q factors, accurate results are returned for the resonant
frequency (fo), self inductance (Ls) or self capacitance (Cp). The Q value and effective
resistance at resonance (Zo) are also indicated.
5–48 Advanced Operation
For simple components, where the above models are valid, the series or parallel frequency
obtained will also correspond to the minimum or maximum impedance point. The results may
not be valid for more complex components, especially those exhibiting more than one
resonance.
Before initiating a resonance search, enter values for the Start and Stop frequencies
corresponding to the range in which resonance is expected to occur. The search process will
complete more quickly if close limits are entered. The search will be performed at the AC test
level specified in IMPEDANCE MODE, but other test parameters will be selected
automatically and need not be specified.
Start the search by pressing the Find Series or Find Parallel soft key. If a resonance is found,
the results will appear after a few seconds, as shown in Figure 5-50. With the Find Series or
Find Parallel label highlighted, a further search may be initiated by pressing the Trigger key.
Alternatively, press the soft key again.
If the resonance lies outside the frequency range of the analyzer, it is possible to use an
extrapolated search, but the results will be at a reduced accuracy. If no resonance is detected
within the specified frequency limits, Figure 5-51 will be displayed and must be acknowledged
with either the Yes or No soft key. Answering Yes will cause the analyzer to look for a
resonance at frequencies up to 2:1 beyond the limits. For best accuracy, always adjust the limits
and repeat if this is possible. Accuracy of extrapolated results is undefined, as it is not possible
to verify the validity of the circuit model. If no resonance is detected, or if the No soft key was
pressed, the analyzer will report Resonance not found.
Figure 5-50 shows the result of a series resonant frequency search on a 10µF capacitor with the
bandwidth search limits set to 100Hz and 1MHz.
Start Level Set by highlighting the parameter with the and navigation
keys, then altering the setting in pre-determined steps with the
and navigation keys, or by using the data entry keypad. The
range is:
1mV to 10V
At frequencies above 300kHz the maximum start level is
restricted.
There are eight parameters which may be altered from within the settings page: GPIB address,
Freq steps, Connection, Measurement, Test Conditions, Ratio Correction,
Demagnetisation, Low Magnetisation.
Freq Steps This sets the frequency steps used when the measurement
frequency is altered using the navigation keys. Two options are
available: Coarse or Fine. Set by highlighting the Freq steps
parameter with the and navigation keys, then using the or
navigation keys to toggle between the two choices. With Coarse
steps selected, the frequency steps are 33% or less; with Fine steps
selected, the frequency steps are 1% or less.
Even with Coarse frequency steps selected, the data entry keypad
can be used to set the measurement frequency with the maximum
possible resolution and accuracy.
Advanced Operation 5–51
Measurement Toggles the analyzer between Single shot mode and Repetitive
mode operation. Set by highlighting the Measurement parameter
with the and navigation keys, then using the or
navigation keys to toggle between the two choices.
Alternatively, the Sngl/Rep control key can be used to select either
single shot or repetitive mode (see section 4.2.5).
Test Conditions Toggles the analyzer between Global and Non-global test
conditions. With Global set, any parameter, e.g. drive level,
frequency, set in one mode of operation, is automatically reflected
in all other modes. For example if the IMPEDANCE MODE
frequency is set to 300Hz, the TRANSFORMER MODE
frequency will automatically be changed to 300Hz too. With Non-
global set, the parameters in each mode may be set independently
of the parameters in all other modes. Set by highlighting the Test
Conditions parameter with the and navigation keys, then
using the or navigation keys to toggle between the two
choices.
Ratio Correction Sets the ratio correction to Off, Normal transformer or Auto
transformer during transformer ratio measurements. Set by
highlighting the Ratio Correction parameter with the and
navigation keys, then using the or navigation keys to select the
setting required.
When measuring the turns ratio of transformers with a floating
secondary winding, the ratio correction should be set to Normal
transformer or Off. If the primary impedance is low, it is highly
recommended to turn ratio correction Off. When measuring the
turns ratio of auto transformers (transformers with a common
connection between one end of the primary and secondary
windings) the ratio correction should be set to Auto transformer.
5–52 Advanced Operation
6.1.1 Introduction
The GPIB is a parallel port designed to be used for communication between instruments
(listeners) and control devices (talkers) such as PCs fitted with a suitable interface card. The
interface protocol is defined by the IEEE488.1 standard. Some additional generic capabilities of
the listeners and talkers are defined by IEEE488.2. The SCPI standard defines the highest level
of command structure including a number of standard commands for all instruments.
C0 No controller
6–2 General Purpose Interface Bus (GPIB)
The path command prefix allows access to commands in the SCPI command tree. Using this
approach greatly simplifies GPIB programming by allowing related commands to be grouped
together. The next part of the string is the command itself which has the structure shown in
Figure 6-2. Multiple commands can be sent in one message by separating them with a
semicolon (maximum length 256 bytes). The terminator indicates the end of the command
string to the instrument: this can be the sending of the line-feed character (ASCII 0Ah) and/or
the assertion of the EOI handshake line on the GPIB bus.
General Purpose Interface Bus (GPIB) 6–3
Each instrument command begins with a mnemonic that describes the required action, e.g.
FREQ for changing the frequency.
If the command requires a parameter, then the next character should be a white space character
(ASCII 20h), although any character in the range 00h-20h can be used with the exception of
line-feed (ASCII 0Ah).
The parameter itself can take one of three forms depending on the command:
1) Discrete data
This includes words like ON, OFF and ABS.
2) Real Number
A floating point number that can be in engineering format or a number with a multiplier
suffix K (kilo-), M (mega-) or G (giga-).
For example:
FREQ 1000.0
FREQ 1E+3
FREQ 0.1E4
FREQ 1k
are all valid ways of setting a frequency of 1kHz.
3) Integer
A single integer number. Often used to indicate a Boolean state.
For example:
RANGE 1
will select range 1.
If invalid data is supplied then a command error will be generated. If data is supplied but the
instrument is not able to apply the setting, an execution error will be generated. If the
instrument is unable to exactly comply with the command and can only apply the nearest
available, a device specific error is generated. Details of these error codes can be found in
Figure 6-6.
access a specific command in a specific mode the user must supply the ‘path’ to reach that
particular command within the tree.
When the unit is powered up the initial path is ‘root’ which is the top level from which all paths
must start.
Note that common commands (which by convention always start with the ‘*’ character) are not
part of the tree and can be accessed regardless of the current path.
So to select the impedance measurement function in deviation mode, the path must describe the
command tree as below:
The ‘:’ character is used as the path separator so the command string will be:
:DEV:FUNC:Z
Note that the string starts with ‘:’. This tells the instrument to start from the ‘root’ path.
Whenever a terminator is reached (line-feed and/or EOI) the path is reset to the root path, so
each new GPIB command string must state the full path in order to work correctly, for example:
To set a measurement frequency of 1kHz at a level of 1.0V, the following string can be used:
:IMP:FREQ 1k <line-feed>
:IMP:LEV 1.0 <line-feed>
However, the following will not work as the second command will be run from the ‘root’ path,
not the measurement path which was required:
:IMP:FREQ 1k <line-feed>
LEV 1.0 <line-feed>
General Purpose Interface Bus (GPIB) 6–5
Summary: The following are the rules for negotiating the command hierarchy
• On power-up or reset, the current path is set to the root.
• Message terminator, line-feed (ASCII 0Ah) or EOI, sets the current path to the root.
• When a colon is the first character of a command, it specifies that the next command
mnemonic is a root level command.
• When a colon is placed between two path mnemonics, the current path is moved down one
level in the command tree if the path name is valid.
• A semicolon separates two commands in the same message without changing the current
path.
• If a command requires more than one parameter, the separate adjacent parameters must be
specified using a comma. Commas do not affect the current path.
• Common commands, such as *RST, *RCL, are not part of the tree. An instrument interprets
them in the same way, regardless of the current path setting.
7 Operation Status Event Register summary bit. This bit is true when
measurement or trimming etc., is in progress
6 RQS – ReQuest for Service. When the bit in the Service Request Enable
mask is set with the corresponding bit in the status register true, this will
trigger a service request to the controller.
MSS – Master Summary Status bit. The version of the request for service bit
which appears in the Status Byte.
5 ESB – Event Summary Bit. When unmasked by the ESE register, this bit will
be set whenever the corresponding bit or bits are set in the Event Status
Register.
General Purpose Interface Bus (GPIB) 6–7
3 A summary bit from Questionable Data. This bit is not used, so is always 0.
2 This is a summary bit of error and instrument status messages. True if any
new status information is available.
1 Always 0.
0 Always 0.
7 Power On (PON) True when the instrument power supply has been turned OFF and
then ON since the last time this register was read.
4 Execution Error (EXE) True when a parameter following a header of a GPIB command
was evaluated by the instrument as being outside of its legal input
range or is otherwise inconsistent with the instrument’s capabilities.
3 Device Dependent Error True when any bit is set in the Encoded Message Register.
(DDE)
2 Query Error (QYE) True when attempting to read data from the output buffer in which
no data was present, or when the data was lost.
0 Operation Complete (OPC) True when the instrument has completed all selected pending
operations before sending the *OPC command
D7 D6 D5 D4 D3 D2 D1 D0
D4 is reserved.
D6 is reserved.
D7 is reserved.
Any of the above messages will set bit 2 of the Service Request Register. If ‘Range Error’ or
‘Connection Error’ occurs, pseudo-measurement results ‘999.9E+15, 999.9E+15’ or
‘999.9E+15’ will be produced dependent on the measurement function.
*CLS Clear Status Clears the Event Status Register and associated status
data structure.
*ESE <NR1> Event Status Enable Sets the Event Status Enable Register to the value of the
data following the command.
*ESE? Event Status Enable Query Returns the current contents of the Standard Event
Status Enable Register as an integer in the range 0 to
255.
*ESR? Event Status Register Query Returns the current contents of the Standard Event
Status Register as an integer in the range 0 to 255. It
also clears ESR.
*SRE <NR1> Service Request Enable Sets the Service Request Enable Register to the value
following the command. The register is set except that bit
6 is ignored.
General Purpose Interface Bus (GPIB) 6–13
*SRE? Service Request Enable Query Returns the current contents of the Service Request
Enable Register as an integer in the range 0 to 63 and
128 to 255.
*STB? Status Byte Query Returns the current contents of the Status Byte with the
Master Summary bits as an integer in the range 0 to 255.
Bit 6 represents Master Summary Status rather than
Request Service.
*IDN? Identification Query Returns the data identifying the instrument. (e.g. the data
output will be: ‘WAYNE KERR,3260B,0,1.0’ where the
first field is the manufacturer, then the model number,
then a zero and the software revision number: here
represented as Issue 1.0).
*TRG Trigger Triggers a direct measurement, but does not return the
results to the controller. This is the same as a GET
(Group Execute Trigger) command.
*OPT Option Identification Query Returns the hardware options installed in the instrument.
*OPC Operation Complete Command Sets the OPC bit of the ESR register.
*OPC? Operation Complete Query Always returns 1 as instrument commands are always
processed sequentially.
:TRAN:RANGE <disc> Select measurement range condition for transformer tests. 6–37
:TRAN:FUNC:NS <real> Select the display of Turns Ratio Ns measurement with Np = x. 6–39
:TRAN:RATio-CORRection <disc> Select the type of transformer for turns ratio correction. 6–40
:BINning:MINOR <real> Set minor term limit if it is applicable on the selected test. 6–46
General Purpose Interface Bus (GPIB) 6–17
:BINning:DEL-ALL Reset all the bin counters in BIN COUNT mode. 6–46
:BINning:LOAD <integer> Load a set of bin limits from non-volatile memory. 6–47
:BINning:RES? Return the counts from all the bin counters. 6–48
:SEQ:LIST? Returns a list of the sequence programs stored in the analyzer. 6–50
:SEQ:CLR Clears all the sequence programs stored in the analyzer. 6–50
:HAN:LEVel? Returns the test level of the currently selected test. 6–52
:HAN:FUNC:NS <real> Select the display of Turns Ratio Ns measurement with Np = x. 6–58
:HAN:RATio-CORRection <disc> Select the type of transformer for turns ratio correction. 6–60
:TELecom:Z0 <real> Set the telecom test line impedance value. 6–62
:TELecom:Rt <real> Set the telecom test termination resistor value. 6–62
:TELecom:Ct <real> Set the telecom test termination capacitor value. 6–63
General Purpose Interface Bus (GPIB) 6–19
:TELecom:Cd <real> Set the telecom test damping network capacitor value. 6–63
:TELecom:Rd <real> Set the telecom test damping network resistor value. 6–63
:TELecom:Cb <real> Set the telecom test blocking capacitor value. 6–64
:MULTI:FREQuency <real> Set the frequency for the currently selected step. 6–66
:MULTI:HIgh-LIMit <real> Set the higher test limit of the currently selected step. 6–66
: MULTI:LOw-LIMit <real> Set the lower test limit of the currently selected step. 6–67
: MULTI:MINor <real> Set the minor test limit of the currently selected step. 6–67
: MULTI:MINor? Returns the minor limit value of the currently selected step. 6–67
:MULTI:SORT <disc> Sorts the current frequency list into the required order. 6–69
:MULTI:RES? <integer> Query the result of the selected frequency step. 6–69
:GRAPH:StarT <real> Set the start frequency for the sweep. 6–70
:GRAPH:StoP <real> Set the stop frequency for the sweep. 6–70
:GRAPH:MarKerF <real> Move the marker to the frequency nearest the supplied value. 6–73
:GRAPH:MAJor-LOw <real> Set the Y-axis start point for the first measurement type. 6–73
Query the current Y-axis start point for the first measurement
:GRAPH:MAJor-LOw? 6–73
type.
:GRAPH:MAJor-HIgh <real> Set the Y-axis stop point for the first measurement type. 6–74
Query the current Y-axis stop point for the first measurement
:GRAPH:MAJor-High? 6–74
type.
:GRAPH:MINor-LOw <real> Set the Y-axis start point for the second measurement type. 6–74
:GRAPH:MINor-HIgh <real> Set the Y-axis stop point for the second measurement type. 6–75
Set the nominal value for use when graphs are being plotted
:GRAPH:NOMinal <real> 6–75
in percentage mode.
:GRAPH:STEP? Query the current step size for the plot. 6–76
:GRAPH:FIT Fit the Y-axis scale to the current measurement data. 6–77
:GRAPH:PEAK Move the marker to the highest point on the current graph. 6–78
:GRAPH:DIP Move the marker to the lowest point on the current graph. 6–78
:RESOnance:StarT <real> Set the start frequency for the search. 6–80
:RESOnance:StoP <real> Set the stop frequency for the search. 6–80
:RESOnance:EQU-CCT <disc> Select the equivalent circuit type for resonance search. 6–81
IMPEDANCE MODE
:IMPedance
Select impedance mode.
Parameters:
None.
Response:
None.
:IMP:TEST
Select test sub-path within impedance mode.
Parameters:
None.
Response:
None.
:IMP:TEST:AC
Select AC measurement.
Parameters:
None.
Response:
None.
:IMP:TEST:RDC
Select Rdc measurement.
Parameters:
None.
Response:
None.
6–24 General Purpose Interface Bus (GPIB)
IMPEDANCE MODE
:IMP:TEST?
Measurement test query.
Parameters:
None.
Response:
0 AC measurement type.
1 Rdc measurement type.
:IMP:TRIGger
Trigger a measurement using the current settings.
Parameters:
None.
Response:
For AC measurements the response will be the first and second measurements separated by
a comma.
Example: 68.860E-9 , 13.0E+6
For Rdc measurements the response will be a single measurement result.
Example: 6.2295E+3
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: :IMP:FREQ 1k
Returns the current test frequency in
:IMP:FREQ 1000 Hz
engineering format.
:IMP:FREQ 1E3
Example: +.10000000E+04
are all equivalent commands and set
for a test frequency of 1kHz.
the test frequency to 1kHz.
Response:
None.
General Purpose Interface Bus (GPIB) 6–25
IMPEDANCE MODE
Parameters: Parameters:
For AC tests supply the required drive None.
level in either Volts or Amps.
Response:
Example: :IMP:LEV 1.2V
Returns the current test level in
:IMP:LEV 1E-2A
engineering format.
will select drive levels of 1.2V and
Example: +.20000000E-01
10mA respectively.
for a test level of 20mV.
Response:
None.
:IMP:DRIVE?
Test level drive type query.
Parameters
None
Response
0 Current drive
255 Voltage drive
6–26 General Purpose Interface Bus (GPIB)
IMPEDANCE MODE
Parameters: Parameters:
ON Turn on bias. None.
OFF Turn off bias.
Response:
INT Select internal bias.
Returns bias status in two integers
EEXT Select external bias. delimited by a comma:
Example: :IMP:BIAS INT First integer: 0 Bias OFF.
:IMP:BIAS ON 1 Bias ON.
will select internal bias and turn it on. Second integer: 0 Internal bias.
1 External bias.
Response:
Example: 1,0
None.
would indicate that internal voltage
bias is turned on.
Parameters: Parameters:
MAX Maximum speed. None.
FAST Fast speed.
Response:
MED Medium speed.
Returns the test speed as an integer
SLOW Slow speed. according to the table:
Example: :IMP:SPEED SLOW 0 Maximum
will select slow speed for 1 Fast
measurements.
2 Medium
Response: 3 Slow
None. Example: 1
indicates that Fast measurements are
selected.
General Purpose Interface Bus (GPIB) 6–27
IMPEDANCE MODE
Response:
None.
IMPEDANCE MODE
:IMP:FUNC:C, L, Z, Q, D, R
Select first or second AC measurement function.
Parameters
Selecting first measurement:
:IMP:FUNC:C Capacitance.
:IMP:FUNC:L Inductance.
:IMP:FUNC:Z Impedance.
Response
None
General Purpose Interface Bus (GPIB) 6–29
IMPEDANCE MODE
:IMP:FUNC:MAJOR?
First AC function query.
Parameters:
None.
Response:
Returns the measurement type according to this table:
0 Inductance.
1 Capacitance
2 Impedance.
Example: 2
indicates that the first measurement is impedance (Z).
:IMP:FUNC:MINOR?
Second AC function query (non-polar measurements).
Parameters:
None.
Response:
Returns the measurement type according to this table:
0 Q-Factor.
1 D-Factor.
2 Resistance.
Example: 1
indicates that the second measurement is dissipation factor (D). Note that if the first
measurement is polar (Z), this query will return the last non-polar setting.
6–30 General Purpose Interface Bus (GPIB)
IMPEDANCE MODE
Parameters: Parameters:
The following parameters are valid: None.
ON Show scale.
Response:
OFF Hide scale.
Returns scale setting according to this
Example: :IMP:SCALE OFF table:
will turn off the scale. 0 Scale hidden.
1 Scale visible.
Response:
Example: 0
None.
indicates that the scale is currently
hidden.
Parameters: Parameters:
The required nominal value. If a unit is None.
supplied it must that of either the first
or second measurement otherwise the Response:
unit mismatch error will be set. If no Returns the nominal in engineering
unit is supplied the current nominal format.
unit will be used.
Example: +.10000000E-01
Examples:
would indicate a nominal of 10mH if
:IMP:NOMINAL 1e-6F the first nominal unit is Henrys.
will set a nominal of 1uF.
:IMP:NOMINAL 0.47e-5
will set a nominal of 4.7uF
Response:
None.
General Purpose Interface Bus (GPIB) 6–31
IMPEDANCE MODE
IMPEDANCE MODE
TRANSFORMER MODE
:TRANsformer
Enter Transformer mode / path.
Parameters
None.
Response
None.
:TRAN:TEST:Primary-LQ
Select Primary L+Q measurement.
Parameters
None.
Response
None.
:TRAN:TEST?
Transformer test query.
Parameters
None.
Response
0 P-LQ
1 Ratio
2 P-RDC
3 S-RDC
4 P-LE
5 S-LE
6 C
6–34 General Purpose Interface Bus (GPIB)
TRANSFORMER MODE
:TRAN:TEST:RATIO
Select ratio measurement.
Parameters
None.
Response
None.
:TRAN:TEST:Primary-RDC
Select Primary Rdc measurement.
Parameters
None.
Response
None.
:TRAN:TEST:Secondary-RDC
Select Secondary Rdc measurement.
Parameters
None.
Response
None.
:TRAN:TEST:Primary-LEakage
Select Primary Leakage measurement.
Parameters
None.
Response
None.
General Purpose Interface Bus (GPIB) 6–35
TRANSFORMER MODE
:TRAN:TEST:Secondary-LEakage
Select Secondary Leakage measurement.
Parameters
None.
Response
None.
:TRAN:TEST:Capacitance
Select interwinding capacitance measurement.
Parameters
None.
Response
None.
:TRAN:TRIGger
Trigger a specified type of measurement and return results.
Parameters
None.
Response
The measurement result as displayed. Multiple results (L + Q, Leakage) will be separated
by commas.
Example: 5.0E + 0
would indicate a turns ratio of 5:1.
6–36 General Purpose Interface Bus (GPIB)
TRANSFORMER MODE
Parameters Parameters
The required test frequency in Hertz. None
The unit suffix ‘Hz’ is optional. Suffix
multipliers K, M, G may be used. Response
Example: :TRAN:FREQ 1k Return the current test frequency in Hz
as floating point number.
:TRAN:FREQ 1000 Hz
Example: 2.50E2
:TRAN:FREQ 1E3
for a test frequency of 250Hz.
are all equivalent commands and set
the test frequency to 1kHz.
Response
None
Response
None.
General Purpose Interface Bus (GPIB) 6–37
TRANSFORMER MODE
Parameters: Parameters:
MAX Maximum speed. None.
FAST Fast speed.
Response:
MED Medium speed.
Returns the test speed as an integer
SLOW Slow speed. according to the table:
Example: :TRAN:SPEED SLOW 0 Maximum
will select slow speed for 1 Fast
measurements.
2 Medium
Response: 3 Slow
None. Example: 1
indicates that Fast measurements are
selected.
Response:
None.
6–38 General Purpose Interface Bus (GPIB)
TRANSFORMER MODE
:TRAN:FUNC:NS/NP :TRAN:FUNC:MAJOR?
Select the display of Turns Ratio Ns/Np Major term query.
measurement.
Parameters
Parameters
None.
None.
Response
Response
0 Ns/Np
None
1 Np/Ns
2 Ns
Example: 0
indicates that the transformer test is set
to Ns/Np.
General Purpose Interface Bus (GPIB) 6–39
TRANSFORMER MODE
:TRAN:FUNC:NP/NS :TRAN:FUNC:NP?
Select the display of Turns Ratio Np/Ns Np query.
measurement.
Parameters
Parameters
None.
None.
Response
Response
Returns the value of Np as floating
None point number.
:TRAN:FUNC:NS <real>
Select the display of Turns Ratio Ns measurement.
Parameters
<real> is decimal numeric data to specify the number of turns of Np.
Example: :TRAN:FUNC:NS 1
will set Np to 1 turn.
Response
None
6–40 General Purpose Interface Bus (GPIB)
TRANSFORMER MODE
Response:
None.
INSULATION MODE
:INSulation
Select Insulation mode / path.
Parameters
None
Response
None
:INS:PRImary-SECondary
Select Primary to Secondary insulation measurement.
Parameters
None
Response
None
:INS:PRImary-GND
Select Primary to ground insulation measurement.
Parameters
None
Response
None
:INS:SECondary-GND
Select Secondary to ground insulation measurement.
Parameters
None
Response
None
6–42 General Purpose Interface Bus (GPIB)
INSULATION MODE
Parameters Parameters
<integer> is decimal integer data None
which can be 100, 200 or 500.
Response
Example: :INS:LEV 500
Returns the test level in volts as integer
will set the insulation test level to
number 100, 200 or 500.
500V.
Response
None
:INS:DISPlay <disc>
Select display µA or MΩ.
Parameters
UA or MOHM.
Example: :INS:DISP UA
will set the analyzer to display µA.
Response
None
:INS:TRIGger
Trigger the insulation test and return results.
Parameters
None
Response
The test result in engineering format.
Example: 5E + 7
would indicate an insulation resistance of 50MΩ.
General Purpose Interface Bus (GPIB) 6–43
BINNING MODE
:BINning
Select one of the previously worked binning modes / binning command path.
Parameters:
None.
Response:
None.
:BINning:SET
Select BIN SET mode.
Parameters:
None.
Response:
None.
:BINning:SORT
Select BIN SORT mode.
Parameters:
None.
Response:
None.
:BINning:COUNT
Select BIN COUNT mode.
Parameters:
None.
Response:
None.
6–44 General Purpose Interface Bus (GPIB)
Parameters: Parameters:
The following discrete parameters are None.
valid:
Response:
ABS Absolute limits.
0 Absolute limits.
PERC Limits as a percentage of
nominal. 1 Percentage limits.
Example: :BIN:LIMIT PERC Example: 1
will set the test limits to a percentage indicates that components will be
of the nominal value. tested against limits that are a
percentage of the nominal value.
Response:
None.
General Purpose Interface Bus (GPIB) 6–45
Parameters: Parameters:
The required high limit. <real> is None.
decimal numeric data. No suffix is
allowed. Response:
Example: :BIN:HI-LIM 10.0 Returns high limit value as a floating
point number.
will set a high limit of 10% when
percentage limits are selected. Example: 0.50E + 0
indicates a high limit of +5% when
Response:
percentage limits are selected.
None.
Parameters: Parameters:
<real> is decimal numeric data. No None.
suffix is allowed.
Response:
Example: :BIN:LO-LIM -10.0
Returns low limit value as a floating
will set a low limit of -10% when
point number.
percentage limits are selected.
Example: -.50000000E+01
Response:
indicates a high limit of -5% when
None. percentage limits are selected.
6–46 General Purpose Interface Bus (GPIB)
:BINning:DEL-ALL
Reset all the bin counters in BIN COUNT mode.
Parameters:
None.
Response:
None.
:BINning:SAVE <integer>
In BIN SET mode, save the present setup in non-volatile memory.
Parameters:
The memory store to use in the range 1 to 99.
Example: :BIN:SAVE 2
will save the current bin limits to memory store number 2.
Response:
None.
General Purpose Interface Bus (GPIB) 6–47
:BINning:LOAD <integer>
Load a set of bin limits from non-volatile memory.
Parameters:
The memory store to use in the range 1 to 99.
Example: :BIN:LOAD 1
will load the set-up currently stored in memory number 1.
Response:
None.
6–48 General Purpose Interface Bus (GPIB)
:BINning:TRIG
In BIN SORT or BIN COUNT mode, trigger a measurement and return results.
Parameters:
None.
Response:
In BIN SORT mode both the measurement result and the bin number are returned.
Example: 14.235E-6 , 5.820 ; 3
where the first two numbers are the measurement and the number after the semicolon is the
sorted bin.
In BIN COUNT mode only the result bin is returned.
Example: 3
indicating that the component met the characteristics of bin 3.
:BINning:DEL-LAST
Decrement by 1 in the most recent bin counter in BIN COUNT mode.
Parameters:
None.
Response:
None.
:BINning:RES?
Return the counts from all the bin counters.
Parameters:
None.
Response:
The cumulative counts of all the bins 0 to 8, the reject bin and the total number of
components tested are returned in comma delimited form.
Example: 4, 3, 2, 6, 3, 7, 8, 2, 5, 1, 34
indicating a total of 34 components tested with 1 reject and bins 0 through 8 containing 4,
3, 2, 6, 3, 7, 8, 2, 5 components respectively.
General Purpose Interface Bus (GPIB) 6–49
SEQUENCE MODE
:SEQuence
Select Sequence mode / path.
Parameters
None
Response
None
Parameters Parameters
The program will be stored in the The program stored in the analyzer
analyzer with the program number with the program number <integer>, 0
<integer>, 0 to 65535. to 65535, is sent to the controller if it
exists.
<contents> conforms to the IEEE488.2
and SCPI ‘Definite Length Block Example: :SEQ:PROG? 23
Response Data’.
will download program number 23 to
the controller.
Response
If the program number already exists, Response
an Execute Error will be generated.
If the program number does not exist
an Execute Error will be generated.
:SEQ:DELete <integer>
Delete a sequence program.
Parameters
The program stored in the analyzer with program number <int>, 0 to 65535, will be deleted
without warning if it exists.
Example: :SEQ:DEL 65
will delete the analyzer sequence program number 65 (if it exists). NO WARNING WILL
BE GIVEN!
Response
If the program number does not exist an Execute Error will be generated.
6–50 General Purpose Interface Bus (GPIB)
SEQUENCE MODE
Parameters Parameters
The program stored in the analyzer None
with program number <int>, 0 to
65535, will be loaded and run if it Response
exists. Returns the result of the sequence
Example: :SEQ:RUN 43 executed as an integer as follows:
Response Example: 0
If the program number does not exist would indicate that the last sequence
an Execute Error will be generated. program run failed.
:SEQ:LIST?
Returns a list of the sequence programs stored in the analyzer.
Parameters
None
Response
The program numbers separated by commas.
:SEQ:CLR
Clears all the sequence programs stored in the analyzer.
WARNING: This GPIB command will clear ALL the sequence data in the analyzer, use with
caution!
Parameters
None
Response
None
General Purpose Interface Bus (GPIB) 6–51
HANDLER MODE
:HANdler
Select handler mode / path.
Parameters
None
Response
None
:HAN:TEST:RDC :HAN:TEST?
Select Rdc measurement. Impedance test query.
Parameters Parameters
None None
Response Response
None 2 Turns Ratio
1 Rdc
0 AC
:HAN:TEST:AC
Select AC measurement.
Parameters
None
Response
None
:HAN:TEST:RATIO
Select Turns Ratio measurement.
Parameters
None
Response
None
6–52 General Purpose Interface Bus (GPIB)
HANDLER MODE
Parameters Parameters
The required test frequency in Hertz. None
The unit suffix ‘Hz’ is optional. Suffix
multipliers K, M, G may be used. Response
Example: :HAN:FREQ 1k Return the current test frequency in Hz
as floating point number.
:HAN:FREQ 1000 Hz
Example: 2.50E2
:HAN:FREQ 1E3
for a test frequency of 250Hz.
are all equivalent commands and set
the test frequency to 1kHz.
Response
None
HANDLER MODE
:HAN:BIAS <disc>
Set the bias condition.
Parameters:
ON Turn on bias.
OFF Turn off bias.
INT Select internal bias drive.
EEXT Select external bias drive.
Example: :HAN:BIAS INT
:HAN:BIAS ON
will select internal bias and turn it on.
Response:
None.
:HAN:BIAS? :HAN:BIAS-STATus?
Bias level query. Returns the current bias status.
Parameters Parameters:
None None.
Response Response:
Returns the bias as a floating point Returns bias status in two integers
number. delimited by a comma:
First integer: 0 Bias OFF.
1 Bias ON.
Second integer: 0 Internal bias.
1 External bias.
Example: 1,0
would indicate that internal bias is
turned on.
6–54 General Purpose Interface Bus (GPIB)
HANDLER MODE
Parameters: Parameters:
MAX Maximum speed. None.
FAST Fast speed.
Response:
MED Medium speed.
Returns the test speed as an integer
SLOW Slow speed. according to the table:
Example: :HAN:SPEED MAX 0 Maximum
will select maximum speed for 1 Fast
measurements.
2 Medium
Response: 3 Slow
None. Example: 1
indicates that Fast measurements are
selected.
Parameters: Parameters:
The following parameters are valid: None.
AUTO Auto-ranging.
Response:
HOLD Hold current range.
Returns the measurement range as an
1 to 7 Range 1 to 7 integer according to this table:
Example: :HAN:RANGE 1 0 Auto-ranging.
:HAN:RANGE AUTO 1-7 Current measurement range.
will select range 1 and auto-ranging Example: 0
respectively.
indicates that auto ranging is selected.
Response:
None.
General Purpose Interface Bus (GPIB) 6–55
HANDLER MODE
Parameters: Parameters:
The following parameters are valid: None.
SER Series equivalent circuit.
Response:
PAR Parallel equivalent circuit.
Returns the equivalent circuit flag
Example: :HAN:EQU-CCT SER according to this table:
will select series equivalent circuit. 0 Parallel.
1 Series.
Response:
Example: 0
None.
indicates that parallel equivalent circuit
is selected.
6–56 General Purpose Interface Bus (GPIB)
HANDLER MODE
:HAN:FUNC:L, C, Z, Q, D, R
Select major and minor terms.
Parameters
Selecting first measurement:
:HAN:FUNC:L Inductance.
:HAN:FUNC:C Capacitance.
:HAN:FUNC:Z Impedance.
Selecting second measurement:
:HAN:FUNC:Q Quality factor.
:HAN:FUNC:D Dissipation factor.
:HAN:FUNC:R Resistance.
Note that selecting Z as the first measurement will force the second measurement to be
Angle. This does not change the equivalent circuit flag setting.
Example: :HAN:FUNC:L;Q
will select L+Q measurements.
Response
None
General Purpose Interface Bus (GPIB) 6–57
HANDLER MODE
:HAN:FUNC:MAJOR? :HAN:FUNC:MINOR?
Major term query. Minor term query.
Parameters: Parameters:
None. None.
Response: Response:
Returns the measurement type Returns the measurement type
according to this table: according to this table:
0 Inductance. 0 Q-Factor.
1 Capacitance 1 D-Factor.
2 Impedance. 2 Resistance.
Example: 2 Example: 1
indicates that the major term is indicates that the minor term is
impedance (Z). dissipation factor (D). Note that if the
first measurement is polar (Z), this
query will return the last non-polar
setting.
:HAN:FUNC:NS/NP
Select the display of Turns Ratio Ns/Np measurement.
Parameters
None.
Response
None
:HAN:FUNC:NP/NS
Select the display of Turns Ratio Np/Ns measurement.
Parameters
None.
Response
None
6–58 General Purpose Interface Bus (GPIB)
HANDLER MODE
Response
None
:HAN:CLRLCD
Clear the LCD Display.
Parameters
None
Response
None
:HAN:XCURSOR <integer>
Set X cursor pixel position.
Parameters
An integer in the range 0 to 319.
Example: :HAN:XCURSOR 20
will set the X cursor position 20 pixels from the left of the LCD screen.
Response
None
General Purpose Interface Bus (GPIB) 6–59
HANDLER MODE
:HAN:YCURSOR <integer>
Set Y cursor pixel position.
Parameters
An integer in the range 0 to 239.
Example: :HAN:YCURSOR 60
will set the Y cursor position 60 pixels from the top of the LCD screen.
Response
None
:HAN:DISP-SMALL <string>
Display string in small characters at current cursor position.
Parameters
Any alphanumeric character may be used within the string. Upper case characters only.
Example: :HAN:DISP-SMALL THIS IS A LOWER CASE STRING
will display ‘this is a lower case string’ at the current cursor position.
:HAN:DISP-LARGE <string>
Display string in large characters at current cursor position.
Parameters
Any alphanumeric character may be used within the string. Upper case characters only.
Example: :HAN:DISP-LARGE THIS IS AN UPPER CASE STRING
will display ‘THIS IS AN UPPER CASE STRING’ at the current cursor position.
6–60 General Purpose Interface Bus (GPIB)
HANDLER MODE
TELECOMS MODE
:TELecom
Select telecommunications transformer test mode / path.
Parameters:
None
Response:
None
Parameters: Parameters:
The required frequency in Hertz. None.
The unit descriptor ‘HZ’ is optional.
Response:
Example: :TEL:FREQ 1E3 HZ
The test frequency in engineering
set the telecom test frequency to 1kHz. format.
Example: +.80000000E+03
Response:
indicating a test frequency of 800Hz.
None.
:TELecom:NETwork :TELecom:NETwork?
Set the damping network state. Query the damping network state.
Parameters: Parameters:
The required state. None
ON or OFF
Response:
Example: :TEL:NET ON
0 = Network is OFF
turns on the damping network.
1 = Network is ON
Response:
None.
6–62 General Purpose Interface Bus (GPIB)
TELECOMS MODE
:TELecom:BLOCK
Turn on and off the blocking capacitor in the damping network.
Parameters:
The required state.
ON or OFF
Example: :TEL:BLOCK OFF
turns off the damping network.
Response:
None.
Parameters: Parameters:
The required value in Ohms. None.
Example: :TEL:Z0 600
Response:
sets the line impedance to 600Ω.
The line impedance in engineering
format.
Response:
Example: +.60000000E+03
None.
indicating a line impedance of 600Ω.
TELECOMS MODE
Parameters: Parameters:
The required capacitor value in Farads. None.
Example, :TEL:CT 1E-9
Response:
sets the termination capacitor to 1nF.
The termination capacitor value in
engineering format.
Response:
Example, +.470000000E-10
None.
indicating a termination of 47pF.
Parameters: Parameters:
The required capacitor value in Farads. None.
Example: :TEL:CD 47E-10
Response:
sets the damping capacitor to 4.7nF.
The damping capacitor value in
engineering format.
Response:
Example: +.470000000E-9
None.
indicating a damping capacitor value of
470pF.
Parameters: Parameters:
The required resistor value in Ohms. None.
Example: :TEL:RD 560
Response:
sets the damping resistor to 560Ω.
The damping resistor value in
engineering format.
Response:
Example: +.15000000E+03
None.
indicating a damping resistor value of
150Ω.
6–64 General Purpose Interface Bus (GPIB)
TELECOMS MODE
MULTI-FREQUENCY MODE
:MULTI
Select multi-frequency mode.
Parameters:
None.
Response:
None.
:MULTI:SET
Switch to the multi-frequency set-up page.
Parameters:
None.
Response:
None.
:MULTI:RUN
Switch to the multi-frequency run page.
Parameters:
None.
Response:
None.
:MULTI:TEST :MULTI:TEST?
Select the frequency step to edit. Return the number of the step that is
currently being edited.
Parameters:
Parameters:
The frequency number in the range 0 to
7. None.
Example: :MULTI:TEST 0
Response:
will select the top frequency for editing
The frequency number in the range 0 to
7.
Response:
Example: 7
None.
would indicate the last frequency is
selected for editing.
6–66 General Purpose Interface Bus (GPIB)
MULTI-FREQUENCY MODE
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: MEAS:FREQ 1k
Returns the current test frequency in
will set the selected frequency to 1kHz
engineering format.
Response: Example: +.10000000E+04
None. for a test frequency of 1kHz.
Parameters: Parameters:
The required higher limit. None.
example: :BIN:HI-LIM 10.0
Response:
will set a high limit of 10% when
The high limit value in engineering
percentage limits are selected.
format.
Response: Example: +.50000000E+01
None. indicates a high limit of +5% when
percentage limits are selected.
General Purpose Interface Bus (GPIB) 6–67
MULTI-FREQUENCY MODE
Parameters: Parameters:
The required lower limit. None.
Example: :BIN:LO-LIM -10.0
Response:
will set a low limit of -10% when
The low limit value in engineering
percentage limits are selected.
format.
Response: Example: -.50000000E+01
None. indicates a high limit of -5% when
percentage limits are selected.
Parameters: Parameters:
The required limit. None.
Example: :BIN:MINOR 1.0
Response:
will set a low limit of 1.0 for the minor
The minor limit value in engineering
test.
format.
Response: Example: .10000000E+01
None. indicates a minor limit of 1.0.
Parameters: Parameters:
The required nominal value, no unit is None.
required: the measurement mode unit is
used. Response:
Example :MULTI:NOM 33e-9 Returns the nominal in engineering
format.
will set a nominal value of 33nF.
Example: +.68000002E-07
Response:
would indicate a nominal of 68nF if
None. the measurement is capacitance.
6–68 General Purpose Interface Bus (GPIB)
MULTI-FREQUENCY MODE
Response:
None.
:MULTI:DEL
Removes the current frequency.
Parameters:
The frequency number in the range 0 to 7
Example: MULTI:DEL 0
will delete the top frequency.
Response:
None.
General Purpose Interface Bus (GPIB) 6–69
MULTI-FREQUENCY MODE
:MULTI:SORT <disc>
Sorts the current frequency list into the required order.
Parameters:
The required sort order.
UP Ascending frequency.
DOWN Descending frequency.
Example: MULTI:SORT UP
will sort the frequencies in ascending order.
Response:
None.
:MULTI:TRIGger
Starts a run of multi-frequency measurements.
Parameters:
None.
Response:
None.
:MULTI:RES? <integer>
Query the result of the selected frequency step.
Parameters:
The frequency number in the range 0 to 7
Response:
The first and second result separated by a comma, if the result is being checked against
limits (absolute or percentage) the PASS/FAIL flag will prefix the result.
Examples: 1, +.68898363E-07, +.72168059E-04
would indicate a pass result on a 68nF capacitor.
+.68898363E-07, +.72168059E-04
would be the result if limits were not being checked.
6–70 General Purpose Interface Bus (GPIB)
GRAPH MODE
:GRAPH
Select graphing mode.
Parameters:
None.
Response:
None.
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: :GRAPH:ST 1k
Returns the start frequency in
will set the start frequency to 1kHz.
engineering format.
Response: Example: +.10000000E+05
None. for a start frequency of 10kHz.
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: :GRAPH:SP 100k
Returns the stop frequency in
will set the stop frequency to 100kHz.
engineering format.
Response: Example: +.125000000E+06
None. for a start frequency of 125kHz.
General Purpose Interface Bus (GPIB) 6–71
GRAPH MODE
Parameters: Parameters:
The required scale type: None.
ON Logarithmic scale.
Response:
OFF Linear scale.
The current scale type:
Example: GRAPH:LOGF ON
1 Logarithmic scale.
will select the logarithmic frequency
0 Linear scale.
scale.
Example: 0
Response:
would indicate that the linear
None. frequency scale is selected.
Parameters: Parameters:
The required scale type: None.
ON Logarithmic scale.
Response:
OFF Linear scale.
The current scale type:
Example: GRAPH:LOGY ON
1 Logarithmic scale.
will select the logarithmic scaling of
0 Linear scale.
the Y-axis (available for Z, Y only).
Example: 1
Response:
would indicate that logarithmic scaling
None. of the Y-axis will be used if Z or Y is
selected.
6–72 General Purpose Interface Bus (GPIB)
GRAPH MODE
Parameters: Parameters:
The following discrete parameters are None.
valid:
Response:
ABS Absolute plot.
0 Absolute plotting.
PERC Plot as a percentage of
nominal. 1 Percentage plotting.
Example: :GRAPH:LIMIT ABS Example: 1
will select plotting of the absolute indicates that the graph will be plotted
measurement result. with the results calculated as a
percentage of the nominal value.
Response:
None.
:GRAPH:MarKer?
Returns the first and second measurement from the current marker position.
Parameters:
None.
Response:
The results in engineering format, separated by a comma.
Example: +.10666955E-06, +.11760951E+01
General Purpose Interface Bus (GPIB) 6–73
GRAPH MODE
Parameters: Parameters:
The required start value. None.
Example: :GRAPH:MAJ-LO 10.0
Response:
will set the minimum to 10.
The current minimum in engineering
format.
Response:
Example: +.95000006E-04
None.
would indicate that the Y-axis will
start at 95µF for example.
6–74 General Purpose Interface Bus (GPIB)
GRAPH MODE
Parameters: Parameters:
The required maximum value. None.
Example: :GRAPH:MAJ-HI 1000.0
Response:
will set the end point to 1k.
The current maximum in engineering
format.
Response:
Example: +.105000006E-03
None.
would indicate that the Y-axis will stop
at 105µF for example.
Parameters: Parameters:
The required minimum value. None.
Example: :GRAPH:MIN-LO 0.0
Response:
will set the minimum to zero.
The current minimum in engineering
format.
Response:
Example: +.100000000E-01
None.
would indicate that the Y-axis will
start at 1Ω for example.
General Purpose Interface Bus (GPIB) 6–75
GRAPH MODE
Parameters: Parameters:
The required maximum. None.
Example: :GRAPH:MAJ-HI 100.0
Response:
will set the end point to 100Ω for
The current maximum in engineering
example.
format.
Response: Example: +.100000000E+02
None. would indicate that the Y-axis will stop
at 10Ω for example.
GRAPH MODE
Parameters: Parameters:
The following values are valid: None.
st
1 Plot 1 measurement.
Response:
nd
2 Plot 2 measurement.
1 1st measurement.
Response: 2 2nd measurement.
None. Example: 2
would, for example, indicate that the
Angle measurement would be
displayed if the selected measurements
were Z+Angle.
Response:
None.
General Purpose Interface Bus (GPIB) 6–77
GRAPH MODE
:GRAPH:SET
Go to the graph mode set-up page.
Parameters:
None.
Response:
None.
:GRAPH:VIEW
Redraw the graph. This command is useful for viewing the graph after switching between the
two measurements, without plotting the graph again fully. Otherwise the display would be left
in the set-up page.
Parameters:
None.
Response:
None.
:GRAPH:FIT
Fit the Y-axis scale to the current measurement data.
Parameters:
None.
Response:
None.
:GRAPH:TRIG
Start plotting a graph with the current settings.
Parameters:
None.
Response:
None.
6–78 General Purpose Interface Bus (GPIB)
GRAPH MODE
:GRAPH:PEAK
Move the marker to the highest point on the current graph.
Parameters:
None.
Response:
None.
:GRAPH:DIP
Move the marker to the lowest point on the current graph.
Parameters:
None.
Response:
None.
:GRAPH:PRINT
Print the current graph on an Epson compatible printer.
Parameters:
None.
Response:
None.
General Purpose Interface Bus (GPIB) 6–79
GRAPH MODE
Parameters Parameters
The required test type: None.
1 Impedance
Response
2 Transformer
The required test type:
Example: :GRAPH:TEST 2
1 Impedance
will select transformer measurements.
2 Transformer
Response
None.
Parameters Parameters
The required sweep type: None.
1 Frequency
Response
2 Level
3 Internal Bias The selected sweep parameter:
4 External Bias (BOOST) 1 Frequency
Example: :GRAPH:TYPE 1 2 Level
3 Internal Bias
will select a frequency sweep.
4 External Bias
6–80 General Purpose Interface Bus (GPIB)
RESONANCE MODE
:RESOnance
Enter resonance mode / path.
Parameters:
None.
Response:
None.
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: :RESO:ST 1k
Returns the start frequency in
Would set the search to start at 1kHz.
engineering format.
Response: Example: +.10000000E+05
None. For a start frequency of 10kHz.
Parameters: Parameters:
The required frequency in Hertz. The None.
unit suffix ‘Hz’ is optional.
Response:
Example: :RESO:SP 1k
Returns the stop frequency in
Would set the search to stop at 1kHz.
engineering format.
Response: Example: +.10000000E+05
None. For a stop frequency of 10kHz.
General Purpose Interface Bus (GPIB) 6–81
RESONANCE MODE
Parameters: Parameters:
The following parameters are valid: None.
SER Series resonance.
Response:
PAR Parallel resonance.
Returns the equivalent circuit state
Example: :RESO:EQU-CCT SER according to this table:
will select the series resonance search. 0 Parallel.
1 Series.
Response:
Example: 0
None.
indicates the parallel resonance
search circuit is selected.
:RESOnance:TRIG
Begin a resonance search.
Parameters:
None.
Response:
Returns the resonant frequency, capacitance, inductance, resistance and Q all separated by
commas.
Example: +.77534195E+06, +.47321000E-05, +.89043000E-08,
+.19562000E-02, +.221748E+02
indicating a resonant frequency of 775.342kHz with equivalent series values at resonance
of 4.7321µF, 8.904nH, 1.956mΩ and a Q value of 22.175.
6–82 General Purpose Interface Bus (GPIB)
DEMAG MODE
:DEMAG
Select demagnetization mode / path.
Parameters:
None.
Response:
None.
Parameters Parameters
The required frequency in Hz. None.
Example: :DEMAG:FREQ 100
Response
will set a de-magnetization frequency
Returns the demagnetization frequency
of 100Hz.
in engineering format.
Response Example: +.10000000+03
None. for a frequency of 100Hz.
Parameters Parameters
The required level in volts or amps. None.
Example: :DEMAG:LEV 2.0V
Response
:DEMAG:LEV 0.02A
Returns the demagnetization level in
will select drive levels of 2V and engineering format.
20mA respectively.
Example: +.10000000+01
for a level of 1V.
General Purpose Interface Bus (GPIB) 6–83
DEMAG MODE
:DEMAG:TRIG
Start demagnetization
Parameters
None.
Response
None.
6–84 General Purpose Interface Bus (GPIB)
CALIBRATE MODE
:CAL
Select calibrate mode / path.
Parameters:
None.
Response:
None.
:CAL:OC-TRIM <integer>
Perform open circuit trimming.
Parameters:
The required trim type.
1 Spot trim.
2 Up to 10kHz.
3 Up to 100kHz.
4 All frequency.
Example: :CAL:OC-TRIM 4
would perform an open circuit trim across the whole frequency range of the unit.
Response:
None.
General Purpose Interface Bus (GPIB) 6–85
CALIBRATE MODE
:CAL:SC-TRIM <integer>
Perform short circuit trimming.
Parameters:
The required trim type.
1 Spot trim.
2 Up to 10kHz.
3 Up to 100kHz.
4 All frequency.
Example: :CAL:SC-TRIM 1
would perform a short circuit trim at the current frequency.
Response:
None.
:CAL:HF-CAL
Perform HF lead compensation.
Parameters:
None.
Response:
None.
:CAL:SELF-CAL
Perform self-calibration; disconnect all BNCs from the instrument terminals before using this
command.
Parameters:
None.
Response:
None.
6–86 General Purpose Interface Bus (GPIB)
CALIBRATE MODE
:CAL:RES?
Returns the result of the most recent trim or calibration performed.
Parameters:
None.
Response:
The trim flag:
1 Calibration passed.
0 Calibration failed.
Example: 1
would indicate that the last trim or calibration was successful.
General Purpose Interface Bus (GPIB) 6–87
ROOT COMMANDS
:TRIGger
Trigger a measurement in the current mode.
Parameters:
None.
Response:
The measurement result depending on the mode.
Response:
None.
6–88 General Purpose Interface Bus (GPIB)
ROOT COMMANDS
Parameters: Parameters:
The required mode: None.
2 2-Terminal.
Response:
4 4-Terminal.
The current setting:
Example: :TERM 4
2 2-Terminal.
will select 4 terminal measurement.
4 4-Terminal.
Response: Example: :TERM 4
None. will select 4 terminal measurement.
General Purpose Interface Bus (GPIB) 6–89
ROOT COMMANDS
Parameters: Response:
The required mode: The set-up condition:
ON Show set-up. 1 Set-up displayed.
OFF Hide set-up. 0 Set-up hidden.
Example: :SETUP OFF Example: 1
will turn off the set-up display. would indicate that the set-up is
visible.
Response:
None.
Response:
None.
6–90 General Purpose Interface Bus (GPIB)
ROOT COMMANDS
:MODE?
Query the currently selected operating mode.
Parameters:
None.
Response:
The current mode:
0 Main menu
1 Impedance
2 Transformer
3 Calibrate
4 Insulation
5 Binning
6 Sequence Edit
7 Sequence Run
8 Handler
9 Telecoms
10 Multi Freq
11 Graph
12 Resonance
13 Demag
14 Cal Status
15 Settings
Example: 1
would indicate that Impedance Mode is selected.
General Purpose Interface Bus (GPIB) 6–91
ROOT COMMANDS
:DUMP-BMP
Returns the display as a windows compatible bitmap. The data conforms to IEEE 488.2 or
SCPI ‘Indefinite Length Arbitrary Block Response Data’.
Parameters:
None.
Response:
None.
Parameters Parameters
The required demagnetization state: None.
ON Enable demagnetization function.
Response
OFF Disable demagnetization function.
The selected demagnetization status.
Example: :DE-MAG ON
0 OFF
will turn on the demagnetization function.
1 ON
Response
None.
Parameters Parameters
The required low magnetization state: None.
ON Enable low magnetization function.
Response
OFF Disable low magnetization function.
The selected low magnetization state:
Example: :LO-MAG ON
0 OFF
will turn on the low magnetization
1 ON
function.
Response
None.
The following examples are written for Microsoft QuickBasic 4.5 running on a PC with a
National Instruments GPIB controller. The programs are short and can be readily converted to
another language/platform as their function is primarily to illustrate the use of the instrument
GPIB commands.
Example 1:
Simple identification query, use this program to establish that the GPIB configuration is
correct.
Example 2:
Simple measurement program. This program triggers a single AC measurement and
displays the result.
Example 3:
Simple querying example. This program interrogates the instrument and display the current
values for a number of AC measurement settings.
Example 4:
Multi-frequency example for AC tests. This program sets up a 4-measurement multi-
frequency test and displays the results from a single trigger.
Example 5:
Performs a graphical sweep of impedance from 20-400kHz and finds the lowest impedance
value. It also takes a screenshot of the graph to a file.
General Purpose Interface Bus (GPIB) 6–93
6.3.1 Example 1
' **************************************************************
'
' Program 1 : Simple GPIB operation check Version 1.0
'
' Platform : QuickBasic 4.5
'
' Description :
'
' This program will ask the instrument to identify itself.
' It assumes the instrument is called 'WK' in the National
' Instruments configuration.
'
' **************************************************************
6.3.2 Example 2
' **************************************************************
'
' Program 2 : Simple Measurement Version 1.0
'
' Platform : QuickBasic 4.5
'
' Description :
'
' This program will set-up and run a single Z+Angle measurement
' on a component.
' This program assumes that the GPIB configuration is correct
' enough to be able to run example program 1 correctly.
'
' **************************************************************
' $INCLUDE: 'QBDECL.BAS' ' National Instruments include file.
CLS ' Clear the screen.
6.3.3 Example 3
PRINT "AC Drive level ="; level; "V" ' Print the AC level.
PRINT "MAX"
END SELECT
6.3.4 Example 4
6.3.5 Example 5
7. SPECIFICATION
7.4.1 Rdc
0.2Ω to 500kΩ 0.5%
7.4.2 L, R, Z, C
Refer to the accuracy chart
100H
10H
1%
1H
100mH
Inductance
10mH
1mH
2%
100µH
0.1%
10µH
0.2%
1µH 0.5%
1%
2%
0.1µH
100 1k 10k 100k 1M
Frequency (Hz)
Specification 7–7
7.5.2 Display
High contrast black and white LCD module 320 x 240 dot with CFL back lighting and manual
contrast control. Visible area 115 x 86mm. Viewing angle 45°.
Installation Category
II (in accordance with IEC664)
7–8 Specification
Temperature Range
Storage -40°C to +70°C
Operating 0°C to 40°C
Full Accuracy 15°C to 35°C
Relative Humidity
Up to 80% non-condensing.
Pollution Degree
2 (mainly non-conductive)
Altitude
Up to 2000m
7.5.8 Safety
Designed to meet the requirements of EN61010-1.
7.5.9 EMC
Complies with EN50081-1, EN50082-1 generic emissions and immunity standards by meeting
with the requirements of EN55022, IEC801.2, EN801.3 & IEC 801.
7.5.10 Mechanical
Height 150mm (6")
Width 440mm (17⅜")
Depth 520mm (20½")
Weight 11kg (24lb 4oz)
8. THEORY REFERENCE
8.1 Abbreviations
B Susceptance (= 1/X) R Resistance
C Capacitance X Reactance
D Dissipation factor (tan δ) Y Admittance (= 1/Z)
E Voltage Z Impedance
G Conductance (= 1/R) ω 2π x frequency
I Current
L Inductance Subscript s (s) = series
Q Quality (magnification) factor Subscript p (p) = parallel
8.2 Formulae
E
Z = (all terms complex)
I
I 1
Y = =
E Z
j
Z s = R + jX = R + jωL = R -
ωC
Zs = (R 2
+ X2 )
RX
Zp =
(R 2
+ X2 )
j
Yp = G + jB = G + jωC = G -
ωL
Yp = (G 2
+ B2 )
GB
Ys =
(G 2
+ B2 )
1 1
where XL = ωL XC = BC = ωC BL =
ωC ωL
ωL S 1
Q = = (series R, L, C values)
RS ωC S R S
RP
Q = = ωC P R P (parallel R, L, C values
ωL P
GP
D = = ωL P G P (parallel G, L, C values)
ωC P
8–2 Theory Reference
RS
D = = ωC S R S (series R, L, C values)
ωL S
1
Note : The value Q = is constant regardless of series/parallel convention
D
X S = Z sinθ B P = Y sinθ
Note that, by convention, +ve angle indicates an inductive impedance or capacitive admittance.
If capacitance is measured as inductance, the L value will be –ve.
If inductance is measured as capacitance, the C value will be –ve.
D = tan δ where δ = (90 – θ)˚ admittance measurement.
1
Q = where δ = (90 – θ)˚ impedance measurement.
tan δ
Maintenance, Support and Services 9–1
9.1 Guarantee
The equipment supplied by Wayne Kerr Electronics is guaranteed against defective material
and faulty manufacture for a period of twelve months from the date of dispatch. In the case of
materials or components employed in the equipment but not manufactured by us, we allow the
customer the period of any guarantee extended to us.
The equipment has been carefully inspected and submitted to comprehensive tests at the factory
prior to dispatch. If, within the guarantee period, any defect is discovered in the equipment in
respect of material or workmanship and reasonably within our control, we undertake to make
good the defect at our own expense subject to our standard conditions of sale. In exceptional
circumstances and at the discretion of the service manager, a charge for labour and carriage
costs incurred may be made.
Our responsibility is in all cases limited to the cost of making good the defect in the equipment
itself. The guarantee does not extend to third parties, nor does it apply to defects caused by
abnormal conditions of working, accident, misuse, neglect or wear and tear.
9.2 Maintenance
9.2.1 Cleaning
The body of the equipment can be cleaned with a damp lint-free cloth. Should it be required,
weak detergents can be used. No water must enter the equipment. Do not attempt to wash down
internal parts.
9.2.2.2 Tests
1) DISCONNECT THE INSTRUMENT FROM THE AC POWER SUPPLY!
2) Inspect the unit and associated wiring for damage e.g. dents or missing parts which might
impair the safety or function of the equipment. Look for any signs of overheating or
evidence that objects might have entered the unit.
3) Ground Bond: Ensure that 25A DC can flow from exposed metal parts of the unit (not
BNC connector outers) to ground with an impedance of less than 100mΩ.
4) Insulation Test: Connect the Live and Neutral of the power cable together and test the
insulation between this point and the ground at 500V DC. Readings greater than 1MΩ are
acceptable.
9–2 Maintenance, Support and Services
USA
Wayne Kerr Electronics Inc.
165L New Boston Street
Woburn MA 01801-1744
Tel: 781 938 8390
Fax: 781 933 9523
email: [email protected]
www.waynekerrtest.com
UK
Wayne Kerr Electronics
Vinnetrow Business Park
Vinnetrow Road
Chichester
West Sussex PO20 1QH
Tel: +44 (0)1243 792200
Fax: +44 (0)1243 792201
email: [email protected]
email: [email protected]
www.waynekerrtest.com
Asia
Microtest
14F-6, No.79, Hsin Tai Wu Road, Sec. 1,
Hsi-chih, Taipei 221, Taiwan, R.O.C.
Tel: +886-2-2698-4104
Fax: +886-2-2698-0716
Email: [email protected]
www.waynekerrtest.com
When returning the instrument please ensure adequate care is taken with packing and arrange
insurance cover against transit damage or loss. If possible re-use the original packing box.
INDEX
Abs % soft key..................................................................... 4–31, 5–19, 5–21, 5–38, 5–41, 5–45
AC meas ....................................................................................................................... 4–30, 4–32
Addresses ..................................................................................................................................... 2
Adjustment of variable components ................................................................................... 4–30
ALC .................................................................................. 4–34, 5–4, 5–6, 6–11, 6–14, 6–27, 7–3
Asterisk ...................................................................................................... 4–12, 4–27, 4–28, 5–7
Auto range ................................................................................................................... 4–33, 5–18
Automatic level control .................................................. 4–34, 5–4, 5–6, 6–11, 6–14, 6–27, 7–3
Bar graph display ..................................... 4–7, 4–30, 4–31, 5–33, 6–15, 6–30, 6–31, 6–32, 7–7
Battery operation .................................................................................................................... 3–1
Bias ................................................. 2–1, 4–2, 4–3, 4–6, 4–9, 4–11, 4–33, 5–44, 6–14, 6–26, 7–4
Bin handler ............................................................................................................................ 5–43
connector pin assignment .................................................................................................... 4–8
external ................................................................................................................................ 4–6
interface ............................................................................................................................... 4–6
Binning mode ...................................................................................................... 5–19, 6–43, 7–2
count .................................................................................................... 5–25, 6–16, 6–43, 6–48
count parameters................................................................................................................ 5–25
set......................................................................................................... 5–22, 6–16, 6–43, 6–44
set parameters .................................................................................................................... 5–22
sort ....................................................................................................... 5–24, 6–16, 6–43, 6–48
sort parameters................................................................................................................... 5–24
BNC
connectors............................................................................................................ 3–3, 5–1, 7–7
LEDs .................................................................................... 5–3, 5–6, 5–11, 5–16, 5–17, 5–18
Cal status page ...................................................................................................................... 5–55
Calibration
data................................................................................................................... 4–9, 4–13, 4–17
handler ..................................................................................................................... 4–23, 4–25
HF lead compensation ....................................................................................................... 4–13
self ............................................................................................. 4–13, 4–22, 4–23, 6–22, 6–85
status .......................................................................................... 5–54, 6–10, 6–11, 6–22, 6–86
trimming .................. 3–2, 4–11, 4–13, 4–26, 4–28, 5–5, 5–11, 5–14, 5–15, 5–16, 5–34, 5–53
Cleaning ................................................................................................................................... 9–1
Connections
active.................................................................................... 5–3, 5–6, 5–11, 5–16, 5–17, 5–18
aux AC out........................................................................................................................... 4–2
aux control out ..................................................................................................................... 4–2
aux in ................................................................................................................................... 4–2
BNC ............................................................................................................................. 3–3, 5–1
for transformer/telecoms measurements.............................................................................. 5–3
front panel............................................................................................................................ 5–1
for handler mode ................................................................................................................................4–23, 5–33
GPIB .................................................................................................................................... 4–2
handler ................................................................................................................................. 4–2
measurement ........................................................................................................................ 7–7
parallel printer...................................................................................................................... 4–2
power cable .................................................................................................................. 1–2, 3–1
protocol ........................................................................................................................ 3–3, 5–1
for handler mode................................................................................................................................ 4–23, 5–33
for insulation mode...................................................................................................................................... 5–12
rear panel.............................................................................................................................. 4–2
safety interlock..................................................................................................................... 4–2
trigger in............................................................................................................................... 4–2
trimming................................................................. 4–16, 4–19, 4–23, 4–24, 5–11, 5–14, 5–16
two-, three, and four-terminal .............................................................................................. 5–2
Contrast control ...................................................................................................................... 4–9
Control keys........................................................................................................................... 4–12
Data entry keypad................................................................................................................. 4–16
key sequence examples ...................................................................................................... 4–15
DC bias current ............................. 2–1, 4–2, 4–3, 4–6, 4–9, 4–11, 4–33, 5–44, 6–14, 6–26, 7–4
DC drive level ........................................................................ 2–1, 4–30, 4–32, 5–10, 6–14, 6–25
Demag mode .......................................................................................................................... 5–51
Drive level ................................................................................................................................ 7–3
AC ........................................................................................ 2–1, 4–32, 5–3, 5–10, 6–14, 6–25
AC (for measurement of iron-cored and ferrite inductors).................................................. 5–5
DC ...................................................................................... 2–1, 4–30, 4–32, 5–10, 6–14, 6–25
insulation mode.................................................................................................................. 5–18
source impedance............................................................................................................... 4–34
telecoms mode ........................................................................................................... 5–36, 7–4
Examples
graph mode......................................................................................................................... 5–47
impedance mode ................................................................................................................ 4–28
insulation mode.................................................................................................................. 5–15
key sequences..................................................................................................................... 4–15
Multi freq mode ................................................................................................................. 5–38
programs ............................................................................................................................ 6–91
telecoms mode ................................................................................................................... 5–34
transformer mode ................................................................................................................. 5–7
External bin handler............................................................................................................... 4–6
External trigger ............................................................................................. 4–2, 4–3, 4–6, 4–27
Fixture
four-wire ........................................................ 4–23, 4–24, 4–25, 4–26, 5–2, 5–12, 5–14, 5–33
Four-wire scanner/fixture ................................ 4–23, 4–24, 4–25, 4–26, 5–2, 5–12, 5–14, 5–33
Frequency
coarse/fine steps......................................................................... 4–14, 4–32, 5–36, 5–51, 5–52
demag mode ....................................................................................................................... 5–51
for measurement of iron-cored and ferrite inductors ........................................................... 5–6
for measurement of very small inductors............................................................................. 5–5
high frequency operation ..................................................................................................... 3–3
impedance mode ................................................................................................................ 4–32
lead set recommendations .................................................................................................... 3–2
lin/log representation in graph mode ................................................................................. 5–44
measurement ...................................................................................................................... 4–32
multi freq mode.................................................................................................................. 5–37
quantized steps in graph mode........................................................................................... 5–47
range .............................................................................................. 2–1, 4–32, 5–4, 5–36, 5–51
resonant.............................................................................................................................. 5–49
sorting in multi freq mode ................................................................................................. 5–41
supply................................................................................................................................... 3–1
supply frequency rejection ................................................................................................ 4–34
telecoms mode ................................................................................................................... 5–36
used for all freq trim .......................................................................................................... 4–25
used for spot trim............................................................................................................... 4–18
used for trimming .......................................................................................... 4–16, 4–18, 5–55
Front panel ............................................................................................................................ 4–16
connections .......................................................................................................................... 5–1
for handler mode ................................................................................................................................4–23, 5–33
LEDs .................................................................................... 5–3, 5–6, 5–11, 5–16, 5–17, 5–18
Fuse ratings ..................................................................................................................... 3–1, 7–7
GPIB
address ....................................................................................................................... 5–52, 6–2
connector pin assignment .................................................................................................... 4–5
Graph mode........................................................................................................................... 5–49
example.............................................................................................................................. 5–47
quantized frequency steps.................................................................................................. 5–47
set parameters .................................................................................................................... 5–47
Guarantee ................................................................................................................................ 9–1
Guard resistance ..................................................................................................................... 5–4
Handler mode........................................................................................................................ 5–33
calibration .......................................................................................................................... 4–23
HF lead compensation..................................................................................................................................4–25
O/C Trim and S/C Trim ...............................................................................................................................4–23
connections ........................................................................................................................ 4–24
HF lead compensation ...................................................................................... 4–16, 4–20, 4–25
Hide scale........................................................................................................... 4–30, 4–31, 6–30
Impedance mode ................................................................................................................... 4–29
example.............................................................................................................................. 4–28
parameters.......................................................................................................................... 4–30
In-circuit measurements ........................................................................................................ 5–4
guard resistance ................................................................................................................... 5–4
Insulation mode..................................................................................................................... 5–19
parameters.......................................................................................................................... 5–19
Keypad ................................................................................................................................... 4–16
key sequence examples...................................................................................................... 4–15
Maintenance .................................................................................................................... 1–2, 9–1
Measurement
2- / 4-terminal .......................................................................................................... 4–12, 5–53
connections
non-Wayne Kerr test leads and fixtures .........................................................................................................3–3
Wayne Kerr test leads and fixtures.................................................................................................................3–3
frequency ............................................................................................... 2–1, 4–32, 5–36, 5–51
for measurement of iron-cored and ferrite inductors ..................................................................................... 5–6
for measurement of very small inductors....................................................................................................... 5–5
noise ................................................................................................................... 4–33, 5–3, 5–5
of a component................................................................................................................... 4–29
example ....................................................................................................................................................... 4–28
of air-cored coils .................................................................................................................. 5–5
of ferrite inductors ............................................................................................................... 5–6
of iron-cored inductors......................................................................................................... 5–6
of transformers ..................................................................................................................... 5–6
of very small capacitors ....................................................................................................... 5–5
of very small inductors......................................................................................................... 5–5
ranges ........................................................................................... 4–33, 4–35, 5–4, 5–18, 5–46
results ................................................................................................................................... 4–4
terms................................................................................................................................... 4–30
Messages
calibration ................................................................................................................ 4–21, 4–22
calibration data lost............................................................................................................ 4–22
cannot set level................................................................................................................... 4–34
current too large ................................................................................................................. 5–11
delete bin counts are you sure? ........................................................................................ 5–24
delete frequency are you sure?......................................................................................... 5–41
delete last result are you sure? ......................................................................................... 5–23
error message from an invalid keypad entry...................................................................... 4–13
handler factory cal lost default values used..................................................................... 4–25
HF lead compensation........................................................................................................ 4–20
HF lead compensation data lost ......................................................................................... 4–21
high voltage on......................................................................................................... 4–19, 5–14
nearest available................................................................................................................. 4–13
nearest drive level .............................................................................................................. 4–34
no action taken ................................................................................................................... 5–31
resonance not found extrapolate? .................................................................................... 5–50
self-calibration reminder.................................................................................................... 4–21
sequence run mode connection prompt.............................................................................. 5–32
sequence run mode failure ................................................................................................. 5–33
settings lost ........................................................................................................................ 4–17
unit not available...................................................................................................... 5–12, 5–22
Mode
binning ............................................................................................................................... 5–19
count............................................................................................................................................................ 5–25
count parameters.......................................................................................................................................... 5–24
set ................................................................................................................................................................ 5–22
set parameters .............................................................................................................................................. 5–20
sort............................................................................................................................................................... 5–24
sort parameters............................................................................................................................................. 5–22
demag................................................................................................................................. 5–51
parameters.................................................................................................................................................... 5–51
graph .................................................................................................................................. 5–49
example ....................................................................................................................................................... 5–47
quantized frequency steps ............................................................................................................................5–47
handler ............................................................................................................................... 5–33
impedance .......................................................................................................................... 4–26
example ........................................................................................................................................................4–28
parameters ....................................................................................................................................................4–30
insulation ........................................................................................................................... 5–19
example ........................................................................................................................................................5–15
parameters ....................................................................................................................................................5–17
multi freq ........................................................................................................................... 5–43
example ........................................................................................................................................................5–38
parameters ....................................................................................................................................................5–41
run ................................................................................................................................................................5–43
set .................................................................................................................................................................5–41
repetitive ...................................................... 4–12, 4–14, 4–15, 4–27, 5–19, 5–23, 5–31, 5–53
resonance ........................................................................................................................... 5–50
sequence run ...................................................................................................................... 5–31
single shot .................................................... 4–12, 4–14, 4–15, 4–27, 5–11, 5–19, 5–23, 5–53
telecoms ............................................................................................................................. 5–37
example ........................................................................................................................................................5–34
parameters ....................................................................................................................................................5–36
transformer......................................................................................................................... 5–10
example ..........................................................................................................................................................5–7
parameters ......................................................................................................................................................5–8
Multi freq – run .................................................................................................................... 5–41
Multi freq – set ...................................................................................................................... 5–38
Multi freq mode .................................................................................................................... 5–43
example.............................................................................................................................. 5–38
parameters.......................................................................................................................... 5–41
Navigation keys ..................................................................................................................... 4–10
O/C trim................................................................................................................................. 4–16
O/C trim (pri)........................................................................................................................ 4–17
O/C trim (pri–sec)................................................................................................................. 4–18
Power
cable............................................................................................................................. 1–2, 3–1
connections ............................................................................................................................................1–2, 3–1
fuse rating.......................................................................................................................................................1–2
ground conductor ...........................................................................................................................................1–2
removal precautions............................................................................................................. 4–9
supply........................................................................................................... 1–2, 1–3, 3–1, 4–1
frequency........................................................................................................................................................3–1
voltage setting ................................................................................................................................................3–1
Printer.............................................................................................................. 2–1, 4–2, 4–4, 7–7
connector ..................................................................................................................... 4–2, 4–3
pin assignment................................................................................................................................................4–5
enable/disable code.................................................................................................... 4–3, 4–13
output ................................................................................................................................... 4–4
printing analyzer cal status/settings................................................................................... 4–13
printing from graph mode.................................................................... 5–43, 5–47, 6–21, 6–78
Rack mounting ........................................................................................................................ 3–1
Rdc meas .................................................................................................. 4–30, 4–32, 6–14, 6–23
Rear panel ................................................................................................................................ 4–8
Connections ......................................................................................................................... 4–2
Repetitive mode ............................................... 4–12, 4–14, 4–15, 4–27, 5–19, 5–23, 5–31, 5–53
Resonance mode .................................................................................................................... 5–50
extrapolated search ............................................................................................................ 5–50
S/C trim.................................................................................................................................. 4–16
S/C trim (pri) ......................................................................................................................... 4–17
S/C trim (sec) ......................................................................................................................... 4–19
Safety ........................................................................................................................................ 1–3
Safety checks............................................................................................................................ 9–1
Safety interlock................................................................................................................ 4–2, 7–4
connections .......................................................................................................................... 4–2
not required .......................................................................................................................... 4–3
Save nom ................................................................................................................................ 4–31
Scanner
four-wire ........................................................ 4–23, 4–24, 4–25, 4–26, 5–2, 5–12, 5–14, 5–33
Self calibration ............................................................................................................ 4–22, 4–23
Sequence edit mode
copying a program ............................................................................................................. 5–28
deleting a program ............................................................................................................. 5–29
loading an existing program............................................................................................... 5–26
new program ............................................................................................................ 5–25, 5–27
program label ..................................................................................................................... 5–28
saving a program................................................................................................................ 5–28
Sequence run mode ............................................................................................................... 5–31
program development ........................................................................................................ 5–31
running a program.............................................................................................................. 5–32
selecting a program ............................................................................................................ 5–31
Service .............................................................................................................................. 1–2, 9–2
Settings page .......................................................................................................................... 5–54
parameters .......................................................................................................................... 5–54
Show scale .......................................................................................................... 4–30, 4–31, 6–30
Single shot mode.............................................. 4–12, 4–14, 4–15, 4–27, 5–11, 5–19, 5–23, 5–53
Soft keys ................................................................................................................................. 4–10
Speed ........................................................................................ 4–27, 4–33, 5–5, 6–14, 6–26, 7–3
affecting accuracy ................................................................................................................ 7–4
graph marker ...................................................................................................................... 5–46
Static electricity....................................................................................................................... 1–3
Supply frequency..................................................................................................................... 3–1
Support and service ................................................................................................................ 9–2
Telecoms mode ...................................................................................................................... 5–37
connections .......................................................................................................................... 5–3
example .............................................................................................................................. 5–34
parameters .......................................................................................................................... 5–36
Transfer standard capacitor ................................................................................................ 4–16
Transformer mode ................................................................................................................ 5–10
connections .................................................................................................................. 5–3, 5–6
example................................................................................................................................ 5–7
parameters............................................................................................................................ 5–8
Trigger ......................................................... 4–7, 4–8, 6–1, 6–14, 6–20, 6–22, 6–24, 6–69, 6–87
bin handler interface .................................................................................................... 4–8, 7–7
external bin handler ............................................................................................................. 4–6
key............................................................................ 4–6, 4–12, 4–27, 5–11, 5–32, 5–42, 5–50
query local condition ............................................................................................... 6–22, 6–87
set local condition.................................................................................................... 6–22, 6–87
Trigger in connector ..................................................................................... 4–2, 4–3, 4–6, 4–27
Trimming......................................... 3–2, 3–3, 4–11, 4–20, 4–26, 4–28, 5–53, 6–22, 6–84, 6–85
default values..................................................................................................................... 4–13
for turns ratio measurements ............................................................................................... 5–9
frequency ............................................................................................................................. 5–5
insulation mode.................................................................................... 5–11, 5–14, 5–15, 5–16
level ..................................................................................................................................... 5–5
measurement of very small inductors .................................................................................. 5–5
options ............................................................................................................................... 4–18
status .................................................................................................... 5–54, 6–10, 6–11, 6–86
telecoms mode ................................................................................................................... 5–34
transformer mode................................................................................................................. 5–7
Wayne Kerr test leads and fixtures....................................................................................... 3–3