The document is a user manual for the YH-T3 weighing indicator. It contains 7 chapters that cover: [1] main specifications of the indicator; [2] installation instructions including connecting the load cell and serial communication interface; [3] operating instructions for functions like turning on, weighing, calibration, and errors; [4] calibration procedures; [5] error indications; [6] battery charging instructions; and [7] maintenance guidelines.

Uploaded by

rudy wibisono
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3K views11 pages

T3 User Manual PDF

The document is a user manual for the YH-T3 weighing indicator. It contains 7 chapters that cover: [1] main specifications of the indicator; [2] installation instructions including connecting the load cell and serial communication interface; [3] operating instructions for functions like turning on, weighing, calibration, and errors; [4] calibration procedures; [5] error indications; [6] battery charging instructions; and [7] maintenance guidelines.

Uploaded by

rudy wibisono
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Weighing Indicator

YH-T3

User Manual
Content
CHAPTER 1 MAIN SPECIFICATION ………………………………………………………....…….....-2-

CHAPTER 2 INSTALLATION ……………………………...…………………...................................- 3-


2.1 FRONT AND BACK VIEW OF INDICATOR ……………………………………………...………- 3-
2.2 CONNECTING LOAD CELL TO INDICATOR ………………………………………..................- 4-
2.3 SERIAL COMMUNICATION INTERFACE (OPTIONAL) ………………………………………..- 4-

CHAPTER 3 OPERATION…………………………...………………………………..………………. - 5-
3.1 TURNING ON………………………………………..……………………………...……………… - 5-
3.2 OPERATION………………………………..…………………………………...………………….. - 5-

CHAPTER 4 CALIBRATION……………………………………………………………..……………...- 7-

CHAPTER 5 ERRORS INDICATION……………………………...………………………..................-9-

CHAPTER 6 CHARGEABLE BATTERY………… …………………………………………………... -9-

CHAPTER 7 MAINTENTANCE ……………………..………………………….……………………..-10-

YH-T3 -1-
Chapter 1 Main Specification

1. Analog:
Input Signal: 0 ~ 10mV
Converting Speed: 40 times/second
A/D code: 360000
Non-linearity: 0.03%
Drift: 0.03%
Stimulating Voltage: DC+5V
Load Cell: 1×350Ω

2. Display
Display: 0 ~ 999999
Interval: 1/2/5/10/20/50 (optional)

3. Serial Communication Interface (Optional)


Signal: RS232C
Transmission Distance: < 20m

4. Ambient Environment
Power Supply: AC 220v ( ~15% ~ +10% ); 50HZ (-2% ~ +2%)
Ambient Temperature: 0℃ ~40℃
Storing Temperature: -25℃ ~55℃
Ambient Temperature: ≤85%RH
Warm-up Time: 10~15minutes
Fuse: 500Ma

5. Weight: 2Kg ( excluding battery )

YH-T3 -2-
Chapter 2 Installation
2.1 Front and back view of the Terminal

Front View of the Terminal

Back View of the Terminal

YH-T3 -3-
2.2 Connecting load cell to Terminal
The 9 pin plug is for load cell , please connects the line as following graphs

End of Indicator End of load cell


h +V Stimulating Power Supply +
-V Shield

c d e f g c-V Stimulating Power Supply -


h i j k
k+IN Signal Output +
+V -IN +IN j -IN
Signal Output -
gShield

Connection of load cell

◆You must cut off power when connecting load cell, the connection must be reliable.
◆After connecting load cell, you should fasten plug with screw-driver.

◆Load cell and Terminal are static sensitive devices, you must adopt anti-static measures during

operation. You must mount lightening rod in thunderstorm frequently happening area.

2.3 Serial Communication Interface (Optional)


1. Terminal communicate with computer by serial communication interface .
2. Connection
Terminal communication interface adopts 5 pin plug, pin definition is as follows
5 pin ------TXD signal (serial communication data)
1 pin ------ ground line
2 pin ------ shield line
The communication interface is 2pin plug.
1 pin ---------signal (serial communication line)
2 pin ---------ground line for communication
3. Parameter
① Signal: RS232C
② Baud Rate: 1200/2400/4800/9600 (Ref. Chapter 4 Calibration)
③ Data format:=〈Weight data (including decimal)〉,ASCII Code

The low bit is prior to high bit and symbol. Negative symbol is”-“; Positive symbol is “0”.

For example: While Terminal displays –500.00Kg, serial output is 00.005-


While Terminal displays 500.00Kg, serial output is 00.0050

YH-T3 -4-
Chapter 3 Operation
3.1 Turning on
3.1.1 Turn on the power, the Terminal performs self-checking and displays current voltage of
battery and go into weighing mode.
3.2 Operation

3.2.1

Press for zero return, Terminal returns to zero within the zero range 4%FS. To make sure the
stabilization light is on when zero operation.

3.2.2

At weighing mode, press to deduct displaying weight while it is positive and stabilization light is

on. Terminal displays net weight “0”, tare light is on, press to deduct tare , tare light is off,
indicator displays gross weigh.

3.2.3

At weighing mode, press to enlarge the resolution of weight to 10 times. Press again to
get back.

3.2.4

At weighing mode, press to lock the weight value. Holding light is on. .The weight value will be

hold even if you take object off or put it on. Press again to get back

3.2.5 Manual accumulation


At weighing mode, while the weight value is large or equal to 20 counts and the value is stable ,

press to perform manual accumulation operation. Meanwhile, Terminal first shows the total
accumulating times (two steps) :[n=] (that means the times displaying), after 1 second later , it
displays accumulating times [ ****]; and then to display the total value of accumulation (two
step):[ totl =] (that means to display the total value of accumulation), after 1 second later, it
displays accumulating value [******], accumulating light is on.
Note: the max. accumulation times are 99999 (under the condition of accumulating amount
≤999999); Prior to the delete operation ,the accumulating amount is recorded even power
off .

YH-T3 -5-
3.2.6 Automatic accumulation

At weighing mode, press + together, Terminal performs automatic accumulation, and


accumulating light flashes. At automatic accumulation mode, while the weight value is large or
equal to 20 counts and the value keeps stable for 1-2 seconds, indicator performs one time
automatic accumulation, displaying accumulation times and amount.
Note: The accumulating amount is not recorded under automatic accumulating mode when
power off.

3.2.7

At weighing mode, press to delete the accumulation times and amount.

3.2.8 Kg/lb diversion

At weighing mode, press + together, Terminal performs unit diversion; while the unit is lb,
the final decimal of the displaying value is lighted.
Note: once you operate accumulating and fail to delete the accumulation amount at one of the unit,
unit diverting is not allowed ,unless you delete the accumulation amount
3.2.9 preset tare

At weighing mode, press + together, Terminal performs tare presetting.


The procedure is as follows:
Step Operation display Explanation
1 [ ****** ] Weighing display
2 [P00000] Remind to input preset tare
Press +

3 Input preset tare, eg “6000” [ 6000]


Press 6 times to change the

lowest digit to “6”, and then press


3 times to set the value as “6000”, and

press to confirm

4 [ ******] Turn back to weighing mode,


displaying the net weight of deducting
tare

3.2.10 Automatic dormancy


Terminal keeps at stable mode for 2 minutes and no any operating and weighing, it will turn to
automatic dormancy, meanwhile the final decimal is lighted. Any operating and weighing will
arouse it to working mode.

YH-T3 -6-
3.3 At calibration and preset tare mode, some of the keys perform as follows

1. : perform “1” adding, press to add “1” on the lowest digit .

2. : perform moving, press to move the lowest digit to left.

3 : perform inputting, press to input setting value to the indicator

4 : perform exiting, press to exit calibration

Chapter 4 Calibration
4.1 Turn on the Terminal and connect load cell correctly . the indicator warms up for 15-30 minutes
at “no load”

At weighing mode, press + together for calibration

4.2 Calibration steps


Step Operation Display Explanation
1 [ ****** ] Weighing display
Setting of interval: Terminal displays the
2 Press + [d *] original interval, which can be updated
by operator
Input updated interval
Eg: “1”
Parameter range:1,2,5,10,20,50,

3 Press once to change the [d 1] Press to confirm the updated one is


value to “1”
as same as the original one
Press for confirmation

Decimal setting:
Terminal displays the original decimal
4 [dP *]
position, which can be updated by
operator.
Parameter range:0~5,
Input updated decimal
0 means no decimal,
Eg: “3”
1~5 separately means the decimal
Press three times to position from 1~5
5 [dP 3]
change the value to “3”
Press to confirm the updated one is
Press for confirmation
as same as the original one.

YH-T3 -7-
F.S setting: Terminal displays the original
6 [F *****]
F.S,, which can be updated by operator.

Input undated F.S.


Eg:”6000”

Press six times to change If there is no need for zero and F.S

the value to “6”, and then calibration, press to step 13,


7 [ 6000]
press three times to move otherwise you must input F.S value to
step 8 and 9.
to “6000”, press for
confirmation
Zero calibration: Terminal self-checks
the zero position, provided indicator
[ nloAd ]
8 didn’t calibrated zero before, it displays
the zero inner code, otherwise it go to
step 9
[ LoAd] F.S calibration: Zero position is
confirmed, you can load weights for the
9
preparation of F.S calibration, that should
be >1/2 of the F.S
Finish of weight loading, press [ ******] Terminal displays the inner code of the
10
loaded weight
[ ****** ] At the stabilization of the value, Terminal
11
will display the weight of loaded weight
Input the actual weight of [ 3000]
loaded weight Press to step 13, if the inputted
Eg:”3000” weight is as same as the loaded weight.

12 Press three times to


change the value to “3”, and

press for confirmation

[bt *] Baud rate: Terminal displays the original


13 baud rate parameter, which can be
updated by operator
[ ****** ] Calibration is over, exit to weighing mode
14

YH-T3 -8-
Chapter 5 Error Indication

ERRORO The zero inner code is over 260000 when calibration.

ERROR1 The loaded inner code is less ; load cell capacity is more .

ERROR2 The loaded inner code is more ; load cell capacity is less.

ERROR 3 Calibration exceeding

HHHHHH The zero lever is too high or load on the platform .

LLLLLLL The zero lever is to high or no pan on the platform.

Chapter 6 Chargeable Battery


6.1 Turing on the AC power, the Terminal will charge the battery automatically. So if you don’t use
battery frequently, you should take battery out.
Note: red end is +, black end is -. Wrong connection will destroy Terminal.
Note: The built-in battery should be fully charged before it is used for the first time.

6.2 Only when you turn off the AC power, and push start key, battery works.
Displaying [LouoL] means the insufficient of voltage, it needs charge.

6.3 When you use the battery first time, you should charge the battery for 20 hours in order to
prevent low voltage resulted from the self leakage of the battery.

6.4 If you don’t use battery for a long time, you should charge the battery for 10-12 hours for each
2 month to prolong using life of battery.

6.5 The battery is easily exhausted products. And it is not granted free guarantee.

YH-T3 -9-
Chapter 7 Maintenance

7.1 To guarantee Terminal clarity and using life, the indicator shouldn’t be placed directly under
sunshine and should be set in the plain space.

7.2 The Terminal can’t be placed into the place where the dust pollution and vibration are serious.

7.3 Load cell should connect with Terminal reliably, and the system should be well connected into
ground. The Terminal must be protected from high electrical field and high magnetic field.
◆ In order to protect the operator, Terminal and relevant device, you should mount lightning rod
in thunderstorm frequently happening area.
◆ Load cell and Terminal are static sensitive device, you must adopt anti static measures.

7.4 It is strictly forbidden to clean the case of Terminal with intensive solvents (for example:
benzene and nitro oils)

7.5 Liquid and conducting particle should not be poured into the Terminal, otherwise the electronic
components will be damaged and electric shock is likely to happen.

7.6 You should cut off power supply of Terminal and relevant device before you pull-in and out the
connecting line of Terminal and external device.
◆ You must cut off power supply of Terminal, before pulling out connecting line of load cell.

7.7 During operation, if trouble occurs, operator must pull off the power supply plug immediately,
and user should return this Terminal to our company for repair. Non-weighing manufacturer
should not repair it, or by yourself, otherwise further destruction may happen.

7.8 The storage is not granted the free repair guarantee, because it is easily exhausted products.
◆ In order to prolong using life, please charge the cell fully before using it. If you don’t use the
Terminal for a long time, you must charge the cell every two month and for eight hours/each
charging time.
◆ Moving or installation must be carefully taken and must avoid strong vibration, impact and
bump in order to protect the storage cell from being damaged.

7.9 From invoice date, the Terminal has a one-year free repair period. If any non-artificially
obstacle about the Terminal happens under correct using conditions within this period, the
user is allowed to send the product with its guarantee card (of the correct number) back to our
corporation for free repair. The Terminal shouldn’t be taken apart, otherwise free guarantee will
be cancelled.

YH-T3 - 10 -