Micros
Micros
Zacharias Jansen put several lenses in a tube (first compound microscope) and the object near the end
of tube appeared to be greatly enlarged, much larger than any simple magnifying glass could achieve.
The first microscope was more of a novelty than a scientific tool since maximum magnification was only
around 9X and the images were somewhat blurry. No Jansen microscopes survived.
Antony van Leeuwenhoek (1632-1723), Father of Microscopy, a Dutch draper and scientist, and one of
the pioneers of microscopy who in the late 17th century became the first man to make and use a real
microscope. Van Leeuwenhoek achieved greater success than his contemporaries did by developing
ways to make superior lenses, grinding and polishing five hundred and fifty lenses to make his new lens
tube that had a magnifying power of 270X and could view objects one millionth of a meter (other
microscopes of the time were lucky to achieve 50X magnification). Van Leeuwenhoek made many
biological discoveries using his microscopes. He was the first to see and describe bacteria, yeast
plantsjotted down as ‘Little Animalcules’ although the microscopes were invented quite before.
Ernst Ruska, was conferred on Nobel Prize in Physics in 1986, for his fundamental work in electron
optics, and for the design of the first electron microscope.
Microorganisms cannot be visualized with naked eyes, as limit of resolution with the naked eye of
human being is 200 um. The size of microorganisms generally ranges between 0.2-8u (most of the
bacteria and fungi). Viruses are ultramicroscopic, pathogenic viruses of veterinary and medical
importance ranges from 17 nanometer to 300 nanometer in size (nanometer- i.e., 10-9 meter).
The bacteria and fungi can be visualized by compound microscope, whereas viruses can be visualized
only with the help of electron microscope.The limit of resolution of compound microscope is200nm and
that of electron microscope is more than 1 nm. The magnification achieved with the help of compound
microscope is 1500X and that of electron microscope is more than 1,000,000X
Depending upon the principle on which magnification is based the microscopy is categorized into:
In electron microscope beam of electrons is used instead of light waves to produce the image. There are
two types:
a. Bright field Microscopy: The microscopic field is brightly lighted/illuminated and the microorganisms
appear dark as they absorb some of the light.Staining with dyes increases the light absorbing ability that
results into good contrast and colordifferentiation. Microscopes of this type produces generally 1000X
and 2000X magnification, above which the image becomes blur with bad resolution.
Resolving power:
The ability to distinguish two closely placed spots as two separate spots. Mere greater
magnification without ability to distinguish fine details is not useful. Large image without greater
resolution may be fuzzy or unclear.
The limit of resolution of human unaided eye is 200um. i.e., humans cannot see objects smaller
than 200um. The resolving power of compound microscope is 0.2 um. And that of electron
microscope is 1-10nm.
The greatest resolution in optical microscopy is obtained with shortest wavelength of visible light
and an objective with maximum NA.
Resolution can be calculated as 0.55/2x1.30=0.21u i.e. smallest details that can be seen with the
light microscope are those having 0.2um.
Numerical aperture:
NA of a microscope objective is a measure of its ability to gather light. The more light (higher NA)
the better the resolving power of the lens. Numerical aperture of the objective lens is an
important consideration in optics as it dictates the angle at which light enters it. Light gathering
ability of microscope objective is quantitatively expressed in terms of the numerical aperture
(NA). High NA produces image of high resolution.
NA= nSinθ
n-is the refractive index of the medium between the object and objective
Refractive index of oil is 1.56 and that of air is 1.003,Water -1.33 Glycerol -1.47
θ- Half aperture angle formed by the two most divergent rays of light which enter the objective.
Therefore,
NA of dry objective: 1.003 x sin 32° = <1, (Highest practical numerical aperture of a dry lens is
0.95.)
NA of oil immersion objective: 1.56x sin 58° = 1.56X0.85=1.33
Immersion media increase the NA of an objective or a condenser by bringing the beams with
higher incidence angle into the light path.
Magnification:
Magnification means number of times the image of a specimen is amplified. 10x i.e., the size of
the image is increased by ten times.
Magnifying power of an objective is determined by the dividing the optical tube length by the
focal length of lens. Optical tube length is the length of the microscope body tube
between the nosepiece opening, where the objective is mounted, and the top edge of the
observation tubes where the eyepieces are inserted. In most microscopes, it is fixed i.e., 160mm.
Low power dry objective: 160/16 = 10x
High power dry objective: 160/4 = 40x
Oil immersion objective: 160/1.7 = 94x or approximately 100x
Eye pieces of magnifying power of 5x, 10X & 20X are available.Generally an eye piece having
magnification of 10X is used. The optical microscopes are equipped with three objective lens,
capable of a different degree of magnification referred as Low power (10X), High power(40X) and
Oil immersion(100X).The magnification of an objective is engraved on barrel.
The total magnification of the system is determined by multiplying the magnifying power of an
objective by that magnifying power of aneye piece.
In optical microscopy with the best optics, the highest useful magnification achieved is 1400x .
Mobj Meyepiece NAobj Mtot Museful Magnification
10x 10x 0.35 100 175-350 low
40x 10x 0.70 400 350-700 ok
100x 10x 1.40 1000 700-1400 ok
100x 15x 1.40 1500 700-1400 empty
Magnification beyond a good resolution is of no value. NA and wavelength of light limit microscope
resolution. Mere enlargement of image does not necessarily resolve new features. The magnifying
power of the lens is limited. After a certain point, the magnification results in a blurred image and is
termed empty magnification. (The airy disk on retina/camera should not exceed twocell/pixel sizes).
Dark field microscopy is accomplished by equipping the light microscopy with a special kind of
condenser that transmits a hallow cone of light from the source of illumination. Not all the light rays
directed through the condenser enters the objective. The microscopic field is essentially dark, if the
transparent medium contains microbes; the diffracted light enters the objective and reach the eye. The
object is illuminated against dark background i.e., microbes appears bright against dark background.
Application includes examination of unstained microorganisms (Spirochetes) suspended in a fluid
specially wet mounts and hanging drop preparations for bacterial motility assessment.
c. Phase contrast microscopy
Phase contrast microscopy is accomplished by equipping the light microscopy with a phase contrast
objective and a phase contrast condenser. Such type of special arrangement of optical systems makes it
possible to distinguish unstained structures within the cell, which differs slightly in their refractive
indices.
Light passing through one material and into another material with slightly different refractive index and
thickness will undergo a change in phase. The differences in phases are translated into variations in
brightness of the structure and can be detected by eye in unstained preparation.
d. Fluorescence microscopy
Fluorescent dye used are the chemical substances, which absorbs light of a particular wavelength, and
energy, emits light of longer wavelength and lesser energy. This phenomenon is known as fluorescence
and application of this phenomenon is the basis for fluorescence microscopy.
In this microscopy, the microorganisms are stained with a fluorescent ye and then illuminated with blue
light; the blue light is absorbed by the fluorescent dye and green light is emitted by the dye used.
After focusing with blue light,the exciter filter removes all but blue light, where as the barrier filters
blocks the blue light and allows green light or other light emitted by fluorescing specimen, to pass
through and reach eye.
It is possible to chemically combine fluorescent dyes with antibodies i.e., Labeled antibodies. When
specimen stained with labeled antibodies and observed under fluorescence microscopy, such type of
microscopy is termed as Immuno-fluorescence.
e. Confocal Microscopy
Laser beam is used to illuminate a variety of planes in the specimen and computer compiles a 3D
master image. With confocal microscopy, the living cells can be viewed.
Electron Microscopy
1931: German scientists Max Knoll (1897–1969) and his pupil Ernst Ruska (1906–1988) build the
first experimental TEM in Berlin.
1933: Ernst Ruska builds the first electron microscope that is more powerful than an optical
microscope.
1935: Max Knoll builds the first crude SEM.
1941: German electrical engineers Manfred Von Ardenne and Bodo von Borries patent an
"electron scanning microscope" (SEM).
Examination of viruses and the ultrastructure of microbial cells is only possible with electron
microscope. Instead of light source and optics for illumination and creation of image, electron beam and
magnetic fields are used. The electron microscope provides tremendous magnification with much higher
resolution using very short wavelength of the electron beam to magnify the specimen. By employing 60-
80 Kv electrons the wavelength is only 0.05 A It is possible to resolve object as small as 10A The
resolution power of electron microscope is 100 times more than optical microscope. It produces useful
magnification of 4,00,000X to 10,00,000X.
A high voltage electron gun is used to produce a beam of electrons. Electrons are emitted from a
tungsten filament. Electrons are accelerated with an electric field (80 kV or 200 kV, for example)
towards the specimenin vacuum chamber (Air can scatter electrons).
Electrons do have poor penetrative power hence, the specimen for electron microscopy should be
extremely thin and dry. The specimenis introduced into the instrument at a point between the magnetic
condenser and the magnetic objective (Comparable with stage in light microscopy). The magnified
image will be seen on fluorescent screen through an airtight window or can be recorded on photo plate
with the help of inbuilt camera.
Various techniques such as shadow casting, negative staining, freeze etching, ultrathin sectioning,
autoradiography are available for use with electron microscopy for characterizing cellular structures.
In SE, the specimen is subjected to a narrow electron beam scan over the surface of specimen
resulting into shower of secondary electrons and other radiations from the specimen surface. The
intensity of the secondary electrons depends upon the shape and chemical composition of irradiated
specimen. The secondary electrons collected by a detector which generates electronic signals which
are scanned to produce a striking and sharp 3D image (compared to the flat images produced by
TEMs) on a screen, similar to a cathode-ray TV screen. Surface topography can be well studied with
SE.
Advantages:
Almost for all kinds of samples, conducting and non-conducting (stain coating
needed);
Based on surface interaction. No requirement of electron-transparent sample.
Produces very sharp, 3D images. Imaging at all directions through x-y-z (3D) rotation of
sample is possible.
Disadvantages:
Low resolution, SEMs are generally about 10 times less powerful than TEMs (so
can use them to see things about 10 nanometers in size).
Usually required surface stain-coating with metals for electron conducting.
In transmission electron microscopy (TEM), a beam of highly focused electrons are directed
toward a thinned sample (<200 nm). These highly energetic incident electrons interact with the
atoms in the sample producing characteristic radiation and particles providing information for
materials characterization. Information is obtained from both deflected and non-deflected
transmitted electrons, backscattered and secondary electrons, and emitted photons are focused
with the help of coils to form an image on screen or on a photographic plate (for making a
permanent record of the image).
In TEM, a giant electron gun uses electromagnetic coils and high voltages (typically from 50,000
to several million volts) to accelerate the electrons to very high speeds. The faster the electrons
travel, the smaller the waves they form and the more detailed the images they show up.
Advantages:
Disadvantages:
Scanning Tunneling microscopes, STMs were invented by Gerd Binnig and Heinrich Rohrer in 1981.
invent the STM and produce detailed images of atoms on the surface of a crystal of gold.
STMs are designed to make detailed images of the atoms or molecules on the surface of something like
a crystal. STMs have an extremely sharp metallic probe that scans back and forth across the surface of
the specimen. As it does so, electrons try to wriggle out of the specimen and jump across the gap, into
the probe, by an unusual phenomenon called "tunneling". The closer the probe is to the surface, the
easier it is for electrons to tunnel into it, the more electrons escape, and the greater the tunneling
current. The microscope constantly moves the probe up or down by tiny amounts to keep the tunneling
current constant. By recording how much the probe has to move, it effectively measures the peaks and
troughs of the specimen's surface. A computer turns this information into a map of the specimen that
shows up its detailed atomic structure.
Steady tunneling current is maintained between microscope probe and specimen. Up and down
movements of probe as it maintains constant current and is used to create image of the surface.
Advantage:
One of the major limitations with ordinary electron microscopes is that they produce amazing detail
using high-energy beams of electrons, which tend to damage the objects they are imaging. STMs avoid
this problem by using much lower energies.
Limitations:
Scan only conductors of electricity, can make images of conductors.
Atomic force microscopes (AFMs)
In 1985, G. Binnig, C. Quate and Christoph Gerber invented the first atomic force microscope (AFM) by
attaching a diamond to a piece of gold foil.
AFMs (atomic force microscopes), overcomes the limitations of STMs is that, they rely on electrical
currents (flows of electrons) passing through materials, so they can only make images of conductors.
AFMs don't suffer from this problem because, although they use still tunneling, they don't rely on a
current flowing between the specimen and a probe, can make atomic scale images of non conducting
materials such as plastics.
An AFM is a microscope is equipped with a little arm called a cantilever with a tip on the end that scans
across the surface of a specimen. As the tip sweeps across the surface, the force between the atoms
from which it's made and the atoms on the surface constantly changes, causing the cantilever to bend
by minute amounts. The amount by which the cantilever bends is detected by bouncing a laser beam off
its surface. By measuring how far the laser beam travels, we can measure how much the cantilever
bends and the forces acting on it from moment to moment, and that information can be used to figure
out and plot the contours of the surface.
In AFM, probe moves over specimen surface at constant distance. Up and down movement as probe
maintains constant distance. AFMs can make images of things at the atomic level and they can also be
used to manipulate individual atoms and molecules—one of the key ideas in nanotechnology.
Optical and electron microscopes can easily generate two-dimensional images of a sample surface, with
a magnification as large as 1000X for an optical microscope, and a few hundred thousands ~100,000X
for an electron microscope. However, these microscopes cannot measure the vertical dimension (z-
direction) of the sample, the height (e.g. particles) or depth (e.g. holes, pits) of the surface features.
AFM, which uses a sharp tip to probe the surface features by raster scanning, scan image, the surface
topography with extremely high magnifications, up to 1,000,000X, comparable or even better than
electronic microscopes. The measurement of an AFM is made in three dimensions, the horizontal X-Y
plane and the vertical Z dimension. Resolution (magnification) at Z-direction is normally higher than X-Y.
Image Lowest Resolvable
Technique Approx Lower Limit
Formed By Unit
1 μm
Optical Microscopy Light Rays Microns (μm)
(monochromatic light)
Transmission 2Ǻ
Electrons Angstroms (Ǻ)
Electron Microscopy (high resolution TEM)
Nanometers (nm) to 10 nm(100 Ǻ)
Scanning Electron Microscopy Electrons
Angstroms (Ǻ)
Atomic Force & Scanning Molecular
Tunneling Microscopies Mechanical Angstroms (Ǻ) 40 Ǻ(theoretical)
Probes
Reference:
Microbiology, Michael J Pelczar, Jr., E.C.S. Chan, Noel R Krieg, The Microscopic Examination of
Microorganisms 5th Edition, Pg.50-69
*****