Manual - NP40
Manual - NP40
NP40
USER’S MANUAL
NP40 Power Quality Analyzer User’s Guide
Introduction
With dual DSP processor architecture, large-scale integrated circuits (FPGA), and embedded
system (uClinux), the NP40 Power Quality Analyzer can calculate a large number of electrical
parameters and process all data quickly. The Analyzer is designed specially for power
inspection and maintenance department, offering extensive and powerful measurements to
check power distribution system, detecting quality and electrical characteristics of power grid
rapidly and conveniently. The analyzer has large-screen color LCD display interface, and easy-
to-use keyboard.
Main features:
Waveform real-time display(4 voltages/4 currents)
Half cycle RMS measurement (voltage and current)
intuitive operation
variety of optional current clamps
Measure DC component
Measurement, calculation and display of harmonics and interharmonics can be
up to 50 times.
Transient capture
Vector, Trend, Bar Graph and events table display
Active power, reactive power, apparent power and energy, shift power factor and
true power factor
Three-phase unbalance (voltage and current)
Flicker
Inrush current
Detection and record of Dips& Swells, Voltage Rapid Change, Interruption .
Detect according to EN50160 or grid with user-defined limit.
data storage and screenshots (can be replayed or output to a PC)
Through the LAN interface PC can keep real-time remote communication
with the Analyzer, operate the Analyzer and read back measurement data.
Built-in 8G memory card.
Additional equipment
Introduction ............................................................................................................................... 1
The Analyzer and its accessories .............................................................................................. 2
CT Options ................................................................................................................................. 2
General Safety Information ...................................................................................................... 3
Chapter 1 Getting started ......................................................................................................... 6
1.1 Overview of The Analyzer .................................................................................................. 6
1.2 Charge the Battery and Prepare for Use ............................................................................... 6
1.3 Input Connections ................................................................................................................ 7
1.4 Rapid Overview of Measuring Modes .................................................................................. 7
1.5 Screen and Function Keys ..................................................................................................... 9
Chapter 2 Basic Operations ................................................................................................... 14
2.1 Tilt Stand and Hang Strap................................................................................................... 14
2.2 Power on/off ...................................................................................................................... 14
2.3 Display Brightness .............................................................................................................. 15
2.4 Update Firmware ................................................................................................................. 15
2.5 Input Connections ................................................................................................................ 15
2.6 Display Information ............................................................................................................. 16
2.7 Set Up the Analyzer ............................................................................................................. 17
2.8 Using Memory and PC ........................................................................................................ 20
Chapter 3 Application examples .......................................................................................... 22
3.1 Scope ................................................................................................................................... 22
3.2 Voltage/Current/Frequency ................................................................................................. 22
3.3 Dips and Swells .................................................................................................................. 23
3.4 Harmonics ........................................................................................................................... 26
3.5 Power and Energy ................................................................................................................ 28
3.6 Flicker ................................................................................................................................. 29
3.7 Unbalance ........................................................................................................................... 30
3.8 Transients............................................................................................................................ 31
3.9 Inrush currents .................................................................................................................... 32
3.10 Power Quality Monitoring ................................................................................................ 33
3.11 Logger............................................................................................................................... 37
Chapter 4 Service and support ............................................................................................. 39
4.1 Warranty ............................................................................................................................. 39
Chapter 5 Specifications........................................................................................................ 40
14
13
3
4
15 12
5
11
6
10
9
8
7
frequency range match the local line power range. To prevent decrease of batter capacity,
charge it at least twice a year.
Before measurements, set the Analyzer according to the line voltage, frequency and wiring
configuration of power system you want to measure, read section ”Set the Analyzer” for more
information.
A
B
C
N
GND
The Analyzer has 4 BNC-inputs for current clamps and 5 banana-inputs for voltages. For a 3-
phase system, make the connections as shown in Figure 1.
First put the current clamps around the conductors of phase A (L1), B (L2), C (L3) and N
(neutral). The clamps are marked with an arrow indicating the correct signal polarity.
Next make the voltage connections: start with Ground and then succession N, A (L1), B (L2), C
(L3). For correct measuring results, always connect the Ground input.
For single phase measurements, use voltage input A (L1) or current input A (L1) or voltage
inputs Ground.
This section provides an overview of all measuring modes. The screen information and the use
of function keys of the Analyzer will be explained in detail in the following sections.
SCOPE mode
Scope mode shows voltage/current by means of waveforms and numerical values, with cursor
and zoom functions.
MENU
Below measurements are available with MENU key:
Measurement mode Screen type Form of measuring results
MONITOR
Table screen
Trend screen
Waveform screen
Phasor screen
The phase relation between voltages and currents are shown in a vector diagram. Phasor screen
is available under Unbalance mode. Below figure explains this:
Scope Waveform and Phasor display are useful to check if voltage leads and current clamps are
connected correctly. In the vector diagram, the phase voltages and currents A (L1), B (L2) and
C (L3) should appear in sequence when observing them in clockwise direction as shown in
below figure:
Screen types
① Table screen: gives an instantaneous overview of important numerical measuring
values.
② Trend screen: this type of screen is related to a table screen. Trend screen shows
the course over time of measuring values from the table.
③ Waveform screen: shows voltage and current waveforms as displayed on an
oscilloscope. Channel A (L1) is reference channel.
④ Phasor screen: shows the phase relation between voltages and currents in a vector
diagram. The vector of reference channel A (L1) points to the positive horizontal
direction.
⑤ Bar Graph screen: shows the density of each measuring parameters as a
percentage by means of a Bar Graph.
Limits settings: For recall, save and definition of the limits for power quality
monitoring.
Setup interface
Press 【SETUP】 key to access to setup interface, use up/down arrow keys to select options to
be adjusted, then press left/right arrow keys to adjust. Settable options:
Wiring configuration: after selecting wiring configuration, press 【ENTER】
key to access to wiring configuration interface. Press【F4】to confirm selection.
Nominal frequency: adjust nominal frequency, after selecting nominal frequency,
switch between preset 50Hz and 60Hz with left/right arrow keys.
Nominal voltage: adjust nominal voltage, after selecting nominal voltage, press
【ENTER】key to access to voltage configuration interface and press 【F5】
key to confirm your selected nominal voltage.
Language: select language then press left/right arrow keys to switch between
Polish and English.
Zooming: select the zooming setup and press 【ENTER】key to access to set
up interface. Current clamp type and zoom of voltage/current are selectable.
Press 【F5】to confirm the setting.
Available function keys:
F2 User preference: access to user preference interface, configure data, time, LAN
interface etc.
F3 Calibration: access to calibration interface, calibrate the Analyzer.
F4 Monitor limits: configure limits value parameters for power quality monitoring.
F5 Confirm, access to menu interface.
User preference
User preference allows you to customize time and LAN interface, press 【F4】 to access
selected parameter preference menu, and 【F5】 to return to SETUP entry menu.
Date and time: chose Date and time, then press 【F4】to select and access the time
setting interface, use the up/down arrow keys to chose parameters, and the left/right
arrow keys to set up each parameter, press 【F5】 to confirm the selection and make
the setting active.
LAN interface configuration: Press up/down arrow keys to select LAN SETUP, then
press 【F4】 to access to setting interface of LAN interface, 【F1】 Tab keys to
switch configuration parameters. When DHCP or fixed IP is highlighted, press
【ENTER】 key radio button to select this option. Access to IP address with 【F1】
Tab key, then press up/down keys to change figure, and left/right keys to move the
cursor. Make the change active by pressing 【F4】confirmation key, press 【F5】to
cancel the change and return to previous menu.
Calibration interface
Press 【F3】 to access calibration interface which shows current firmware version, correct
password is requested for processing calibration, do not calibrate the Analyzer arbitrarily but
have it calibrated by specified institution or the manufacturer. Press 【F5】to return to setting
interface.
Monitor limits
The Analyzer presets a set of limits according to EN50160 standard, and reserves two user-
defined options, which the users can modify under EN50160 standard set of limits and save as
user-defined set of limits.
Use the up/down arrow keys to move the highlighted row, then press 【ENTER】 key radio
button to select highlighted row. Selected by the radio button means this set of limits is
currently active.
Press 【F4】to access to limits edition interface and edit current highlighted set of limits, a
survey of monitor limits adjustments:
Limits Adjustments
Voltage 2 probability percentages (100% and adjustable): each with adjustable
upper and lower limits.
Harmonics For 2-25 harmonics and THD 2 probability percentages (100% and
adjustable): each with adjustable upper limit.
Flicker (*) Weighing curve (lamp type). 2 probability percentages (100% and
adjustable): adjustable percentage with adjustable upper limit.
Dips (*) Threshold, hysteresis, allowed number of week.
Swells (*) Threshold, hysteresis, allowed number of week.
Interruption (*) Threshold, hysteresis, allowed number of week.
Rapid Voltage Voltage tolerance, steady time, minimum step, minimum rate, allowed
Change (*) number of week.
Unbalance 2 probability percentages (100% and adjustable): adjustable
percentage with adjustable upper limit.
Frequency 2 probability percentages (100% and adjustable): each with adjustable
upper and lower limits.
(*): Setups are also valid for measuring mode.
Using memory
The Analyzer has a built-in 8G TF card to store data the users saved. The Analyzer can save
screenshot of current screen, and data set of current measurement too.
SAVE interface
The 【SAVE】 key accesses save interface, select screenshot or data set as save type, the
name of the file to be saved is editable.
Press 【F1】 to switch save type. Use left/right arrow keys to move the cursor, and up/down
keys to change name, press【F4】SPACE to erase current selected character, the cursor moves
to its next position. Press 【F5】to save the setting and return to previous interface.
MEMORY interface
The MEMORY button accesses to the save list interface, which shows the save time, name and
type of saved files. Use up/down arrow keys to select specified row to save to. After accessing
to the save interface, insert a U disk and wait for a few seconds, then press F2 to copy selected
file to the U disk when “TO USB” characters become lighted, press ENTER to confirm when a
copy finished window is popped up before pulling out the U disk.
Insert the U disk to a PC and use upper computer software to view the copied file.
Available function keys:
F2 Copy file to U disk after inserting a U disk and the characters of the key become
lighted.
F3 View selected saved file.
F4 Delete selected saved file.
F5 Return to previous interface.
Using PC
The Analyzer is equipped with a LAN interface for communication with a PC. With supplied
upper computer software you can do remote control of the Analyzer on a PC. In addition, you
can also view the data and screenshot copied from a U disk with the upper computer software.
On USER Preference interface, set the LAN interface of the Analyzer correctly before
connecting the Analyzer to the internet with a network cable. Run the upper computer software,
input the IP address of the Analyzer to make the connection, an operation interface analoging
the Analyzer will be displayed for a successful connection. The saved file also can be
downloaded from analyzer.
Table screen
The number of columns in the Table screen depends on the power system configuration. The
figures in the Table screen are present values that may update constantly. Changes in these
values over time are recorded as soon as the measurement is turned on. The recording is visible
in the Trend screen.
Available function keys:
F1 Under 3-phase Y type wiring configuration, switch between voltage readings of each
phase (A/L1, B/L2, C/L3,N)or phase-phase (AB, BC, CA).
F4 Access to the Trend screen.
F5 Switch between HOLD and RUN.
Trend
All values in the Table screen are recorded, but the Trends from each row in the table are
displayed one at a time. Press function key F1 to switch between parameters.
The traces build up from the right side. Readings in the header correspond to the most recent
values plotted on the right.
Available function keys:
F1 Switch parameters that the current Trend screen displays, the header shows content
being displayed.
F4 Return to the Table screen.
F5 Switch between RUN and HOLD.
Tips and Hints
Voltage and frequency should be close to the nominal values of for example 120V, 230V,
480V, 60Hz or 50Hz.
The voltages and currents in the table can be e.g. used to check if power applied to a 3-phase
induction motor is in balance. Voltage unbalance causes high unbalanced currents in stator
windings resulting in overheating and reduced motor life. Each of the phase voltages should
not differ more than 1% from the average of the three. Current unbalance should not exceed
10%. In case of too high unbalance, use other measuring modes to further analyze the power
system.
A Crest Factor close to 2.0 indicates high distortion. CF=2.0 can e.g. be found if you measure
the current drawn by rectifiers that only conduct at the sine wave top.
Rapid voltage changes are quick transitions of the RMS voltage between two steady-states.
Rapid voltage changes are captured based on steady voltage tolerance, steady time, minimum
compensation and minimum rate detected. When voltage change exceeds dip or swell threshold,
it is considered as Dip or Swell instead of Rapid Voltage Change. The events list shows voltage
step change and transient time. The detailed events list shows maximum voltage change over
nominal voltage. Voltage change trend is shown in Fig. 3-3-4.
In addition to the voltage, current is also recorded. This allows you to see cause and effect of
deviations. Function key 【F2】 accesses event tables where voltage events are listed in
sequence.
Trend
Available function keys:
F1 Switch between voltage and current trends, the header shows parameters being
displayed.
F2 Access to Events tables
F3 Access to Cursor
F4 Selection of Cursor move or Zoom
F5 Switch between RUN and HOLD.
Event criteria such as threshold, hysteresis and others are preset, but they may be adjusted. The
adjustment menu is reached via the 【SETUP】 key and limits setup.
Events tables
The Events table lists all threshold crossings of phase voltages. Thresholds according to
international standards or user –definable thresholds can be used. The Events table records
major event characteristics: start time, duration, voltage magnitude, event type, and occurrence
phase etc.
The below abbreviation is used in the events tables,
3.4 Harmonics
Harmonics measure and record harmonics and interharmonics up to the 50th. Related data such
as DC components, (THD) Total Harmonic Distortion, and K-factor are measured. Harmonics
are periodic distortions of voltage, current or power sinewave. A voltage or waveform can be
considered as a combination of various sinewaves with different frequencies and magnitudes.
The contribution of each of these components to the fundamental signal is measured. Readings
can be given as a percentage of the fundamental (%f), or as a percentage of full signal RMS
(%r).Results may be viewed in a Bar Graph display. Harmonics are often caused by non-linear
loads such as DC power supplies in computers, TV’s and adjustable speed motor drives.
Harmonics can cause transformers, conductors, and motors to overheat.
The left/right arrow keys are used to position the Cursor on particular bar. The screen header
will show for that bar phase identifier, harmonic component ratio, frequency and phase angle.
If not all bars are shown on the screen, you can bring the next set within the viewing area by
moving the Cursor off the left or right end of the screen.
Available function keys:
F1 Selection of harmonics type: voltage, current.
F2 Selection of bar set to be displayed: L1, L2, L3, N or all
F3 Interharmonics display on/off
F4 Open the table screen
F5 Switch between RUN and HOLD.
Table Screen
The table screen lists all the harmonic parameters, including Harmonic Voltage, Harmonic
Current, Interharmonic Voltage and Interharmonic Current. Select next page with up/down
keys.
Available function keys:
F3 Take fundamental voltage as percentage reference (%f) or total harmonic voltage as
percentage reference (%r)
F4 Open the bar graph screen of harmonic
F5 Switch between RUN and HOLD.
Table
The table displays power data for each phase and in total: real or active power (kW), apparent
power (kVA, the product of rms voltage and current), reactive power (kVAR, the reactive
component of apparent power caused by phase shift between AC current and voltage in
inductors and capacitors), power factor (TPF, the ratio of real power to apparent power for the
total rms including harmonics), displacement power factor (DPF, the ratio of real power to
apparent power for fundamental), and the rms values of voltage and current.
A popup table with energy usage by phase and in total can be activated by pressing 【F3 】 –
Energy function key. The table shows real energy (kWh), apparent energy (kVAh) and reactive
energy (kVARh). The energy measurement starts when Power&Energy is started. The readout
can be reset with function key F5.
Available function keys:
F3 Access the Energy popup screen.
F4 Access the Trend screen.
F5 Switch between RUN and HOLD.
Trend
The figures in the table are instantaneous values that update constantly. While changes in these
values over time are recorded in the Trend. The traces build up from the right side. The
readings in the header correspond to the most recent measurements plotted on the right.
Available function keys:
F1 Switch parameter being displayed.
F4 Return to the Table screen.
F5 Switch between RUN and HOLD.
3.6 Flicker
Flicker quantifies the luminance fluctuation of lamps caused by supply voltage variations. The
Analyzer design strictly meets the IEC61000-4-15 Flicker Meter model. The Analyzer converts
duration and magnitude of voltage variations into an ‘annoyance factor’ caused by the resulting
flicker of a 60W lamp. A high flicker reading means that most people would find the
luminance changes irritating. The voltage variation can be relatively small. The measurement is
optimized to lamps powered by 120V/60Hz or 230V/50Hz. Flicker is characterized per phase
by the parameters shown in a table. The Trend screen shows changes of instantaneous flicker
sensation level over time.
Table
Flicker is characterized by: short term severity Pst (measured over 10 minutes) and a long term
severity Plt (measured over 2 hours). The Analyzer records severity measured over 1 minute
too for fast feedback. Pst and Plt are parameters showing flicker severity over a certain period
of time. Momentary flicker is shown in the PF5 submenu and is reached via function key
【F4】. Flicker PF5 is displayed as a fast Trend plot.
Available function keys:
F4 Access PF5 Trend screen.
F5 Switch between RUN and HOLD.
3.7 Unbalance
Unbalance displays phase relations between voltages and currents. Measuring results are based
upon the fundamental frequency component (50 or 60Hz, use symmetrical components). In a 3-
phase power system, the phase shift between voltages and between currents should be close to
120°. Unbalance mode offers a measurement table and a Phasor display.
Table
The table screen shows all relevant numerical values: negative voltage unbalance percentage,
zero sequence voltage unbalance percentage, negative current unbalance percentage, zero
sequence current unbalance percentage, fundamental phase voltage, frequency, fundamental
phase current, angle between voltage and current for each phase relative to the reference phase
(A/L1)and angles between voltage and current for each phase.
Available function keys:
F4 Access Phasor screen.
F5 Switch between RUN and HOLD.
Phasor
Shows the phase relation between voltages and currents in a vector diagram divided in 30
degree sections. The vector of the reference channel A (L1) points to the horizontal direction.
Additional numerical values are given: negative voltage and current unbalance percentage, zero
sequence voltage and current unbalance percentage, fundamental phase voltage and current,
frequency, phase angles. With function key 【F1】, you can choose readings of all phase
voltages, phase currents or voltage and current in one phase.
Available function keys:
F1 Selection of signals to be displayed: V displays all voltages; A displays all currents.
L1, L2, L3 give simultaneous display of phase voltage and current.
F4 Return to unbalance table screen.
F5 Switch between RUN and HOLD.
should not exceed 2%, current unbalance should not exceed 10%. In case of too high unbalance,
use other measuring modes to further analyze the power system.
Each phase voltage or current can be split-up into three components: positive sequence,
negative sequence, and zero sequence.
The positive sequence component is the normal component such as present in balanced 3-phase
systems. The negative sequence component results from unbalanced phase-to-phase currents
and voltages. This component for instance causes a ‘braking’ effect in 3-phase motors, which
will result in overheating and life reduction. Zero sequence components may appear in an
unbalanced load in 4-wire power systems and represent the current in the neutral wire.
Unbalance exceeding 2% is considered as too high.
3.8 Transients
The Analyzer can capture waveforms at high-resolution during a variety of disturbances. The
Analyzer will give a snapshot of the voltage and current waveforms at the precise time of the
disturbance. This allows you to see the waveforms during transients.
Transients are fast spikes on the voltages waveform. Transients can have so much energy that
sensitive electronic equipment can be affected or even damaged. A waveform is captured each
time that the voltage exceeds adjustable limits. A maximum of 100 events can be captured. The
sample rate is 20kS/s.
Waveform display
Cursor and Zoom can be used to investigate details of captured waveforms.
Available function keys:
F2 Playback captured Transients waveforms.
F3 Access to Cursor.
F4 Selection of Zoom or Cursor.
F5 Switch between RUN and HOLD.
Trend display
Use the arrow keys in the Start menu to adjust the trigger limits: expected inrush time, nominal
current, threshold and hysteresis. The maximum current determines the vertical height of the
current display windows. Threshold is the current level that triggers the trend capture. The
inrush time is the time between trigger and the time that the current falls to the value indicated
by Hysteresis and is indicated on the trend display between two vertical markers. The screen
header displays the rms of all phases during the inrush time. If the Cursor is on, the rms
measuring values at the Cursor are displayed.
Available function keys:
F1 Switch parameters to be displayed.
F3 Access to Cursor.
F4 Selection of Zoom or Cursor.
F5 Switch between RUN and HOLD.
The bars for Dips/Interruptions/Rapid Voltage Changes/Swells are narrow and indicate the
number of limits violations that occurred during the observation period. The allowed number is
adjustable (for instance to 20 Dips/week). The bar turns to red if the adjusted limit is violated.
You can use a pre-defined set of limits or define your own. An example of a pre-defined set is
that according to the EN50160 standard.
The table below gives a survey of the aspects of Power Quality Monitoring:
Parameter Available Bar Limits Averaging
Graphs Interval
V rms 3, one for each phase Probability 100%: upper 10 minutes
& lower limits
Probability x %: upper &
lower limits
Harmonics 3, one for each phase Probability 100%: upper 10 minutes
limit
Probability x %: upper
limit
Flicker 3, one for each phase Probability 100%: upper 2 hours
limit
Probability x %: upper
limit
Dips/Interruptions/Rapid 4, one for each Allowed number of 1/2 cycle rms
Voltage Changes/Swells parameter covering events based
all 3 phases
Unbalance 1, covering all 3 Probability 100%: upper 10 minutes
phases limit
Probability x %: upper
limit
Frequency 1, measured on Probability 100%: upper 10 seconds
Reference Voltage & lower limits
Input A/L1 Probability x %: upper &
lower limits
Events table
The events table shows the events that occurred during the measurement with time of start,
phase and duration. Events recording situation:
V rms events: an event is recorded each time that a 10 minute aggregated RMS value
violates its limits.
Harmonics events: an event is recorded each time a 10 minute aggregated harmonic
or THD violates its limit.
Dips/Interruption/Rapid Voltage Change/Swells events: an event is recorded each
time one of the items violates its limit.
Unbalance and Frequency events: an event is recorded each time that a 10 minute
aggregated RMS value violates its limit.
Available function keys:
F3 Access Trend Screen
F4 Switch between selected and all events.
F5 Return to previous menu.
The below abbreviation is used in the events table,
CHG changing of fast voltage
DIP voltage dip
INT voltage interrupt
SWL voltage swell
Hx the harmonic order of beyond limit value
Hz Frequency
Trend
Press【F3】 in Events Table to access Trend Screen. Press left/right arrow keys to access
horizontal time base zoom.
Available function keys:
F5 Return to previous menu.
number
CHG L1 event number L2 event number L3 event
number
Unbalance the transfinite
the transfinite
percentage of
percentage of
x% limit value 100% limit value
unbalance unbalance
frequency x% limiting
the transfinite The max Fre. The mini Fre. of
deviation frequency percentage of x% of 100% limit 100% limit value
limit value value
*the transfinite percentage of x% limit value: the percentage between x% measured value and
setting limit value
* the transfinite percentage of 100% limit value: the percentage between 100% measured value
and setting limit value
3.11 Logger
Logger function is used to record a group of measurement data as your selected parameter, the
interval selected from 1s to 1hour. When each interval ends, the maximum, minimum and
average of selected parameters are recorded to memory and then start next interval recording.
The whole process lasts as you selected duration time and the record parameter are all
selectable for users.
Press 【MENU】key and select Logger Menu. Press 【ENTER】key to access to logger
setting interface. User can view the spare memory, set logger parameter, interval, duration time
and name of saved file. When finished the setting, press 【F5】to start the logging by timing
or immediately.
The record file is saved in SD card as CSV format, which can be opened as EXCEL table by
Office 2007 or advanced version in PC. User can make the record data into the graph as
demand. For example, select first line of logger time and three lines of maximum, minimum,
average of L1 Vrms, then insert the scatter lots with smooth line, it will generates the graph as
below picture 3-11-1.
Fig. 3-11-1
Table Screen
LUMEL S.A. -37-
NP40 Power Quality Analyzer User’s Guide
The table screen displays all the real time measurement data as selected parameters. With
left/right key to refer to next page for view of the data.
Available function:
F5 Switch between RUN and Hold.
F4 Save records. The prompt of success saved will pop up after pressing this key. Press
【ENTER】to confirm and the record will automatically stop.
Chapter 5 Specifications
5.1 Frequency Measurement
Nominal frequency Measurement range Resolution Accuracy
50Hz 42.50~57.50 Hz 0.01Hz ±0.01Hz
60Hz 51.00~69.00 Hz 0.01Hz ±0.01Hz
Note: measured on Reference Voltage Input A/L1.
5.2 Voltage Input
Numbers of input 4 (3 phase + neutral) DC coupling
Max continuous input 1000Vrms
voltage
Range of nominal voltage 50 to 500V
Max pulse peak voltage 6kV
Input impedance 4MΩ/5pF
Voltage
Vpk ±6000Vpk 1V ±15%
Vrms 10~1000Vrms 1V ±2.5%
Minimum test time 50 s
Sampling rate 20kS/s
LUMEL S.A.
-43-
NP40 Power Quality Analyzer User’s Guide
Memory
Flash memory 128M
TF card Standard 8G
Standard
Measurement method IEC61000-4-30 S degree
Measurement performance IEC61000-4-30 S degree
Power quality monitoring EN50160
Flicker IEC61000-4-15
Harmonic IEC61000-4-7
Environment
Working temperature 0°~ 40°
Storage temperature -20°~ 60°
Humidity 90% relative humidity
Safety
Complied with IEC61010-1
Safety Degree: 600V CAT IV 1000V CAT III
Pollution Degree: 2
Mechanical
Dimension 262× 173 × 66
Weight 1.6 kg
Power
Adapter input 90~264V
Adapter output 9V 2.2A
Battery Rechargeable lithium ion 7.4V 4.4Ah
Battery operating time >7 hours
Battery charge time 6 hours
5.10 The specification of additional equipment
Model Range Turns ratio Accuracy Size mm
KLC8C-5A (clamps) 5A 10mV/A 0.2% Φ8
CTC0080 (clamps) 50A 10 mV/A 0.2% Φ8
CTC0130(clamps) 100A 10 mV/A 0.2% Φ13
CTC1535(clamps) 1000A 1 mV/A 1.0% Φ52
PY-3000A 3000A 65 mV/1000A 1.0% (+2% Φ162
(Rogowski coils) position error)
PY-5000A 5000A 50 mV/1000A 1.0% (+2% Φ143
(Rogowski coils) position error)
Both of ways can make the connection available. After clicking ‘Connect’, the below
interface will appear,
Remote control the device through pressing buttons in left area of above interface, the right
area is the stored list files, user can click ‘Download’ button to download selected files to
the computer.
(1) Open the picture file and take the below screenshot for example
(2) Open the date file, and take the below picture saved under voltage/current
measurement for example
(3) Open log file with .csv format, user can directly open EXCEL file under windows or open it
using this software.
Export department:
tel.: (+48 68) 45 75 139, 45 75 233, 45 75 321, 45 75 386
fax.: (+48 68) 32 54 091
NP40-09
e-mail: [email protected]