03 Solid-Phase Extraction

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

SOLID-PHASE EXTRACTION

SOLID PHASE EXTRACTION (SPE)


Solid phase extraction1 (SPE) is a sample preparation technique using a solid adsorbent contained most commonly in a cartridge device
(Figure 1), or on a disk to adsorb select species from solution. SPE is used to isolate a species in a sample or to clean-up a sample
before analysis.

Figure 1. Photograph of an SPE cartridge (top).


As the sample is slowly passed through the SPE cartridge or disk, the analyte and some of the sample matrix compounds may be
retained on the SPE material. Depending on the properties of the analyte and the SPE sorbent, a wash solvent can be chosen to
selectively remove (elute) components from the SPE sorbent while retaining others. The ultimate goal is to remove interferents present
in the matrix from the analyte, producing a solution containing primarily analyte. The following different scenarios can be utilized to
achieve this goal.
1. Bind and elute strategy: The analyte(s) are retained and matrix components do not retain. The analyte subsequently is washed and
eluted with a strong solvent. (Figure 2A)
2. Removal/trapping strategy: The analyte(s) are not retained and elutes, while the matrix components are retained. (Figure 2B)
The captured analyte elution solvent can be further reduced in volume to concentrate the analyte prior to analysis.
SPE is used to:
simplify complex sample matrices
purify compounds of interest
reduce the ion suppression in mass spectrometry applications
fractionate complex mixtures for analysis by classification
concentrate analytes present at low levels.

10/24/2019 1 https://chem.libretexts.org/link?155906
Figure 2. SPE bind and elute strategy (A) where the analyte is selectively retained on the adsorbent, while interferents pass through (a-
b). Next, the analyte is eluted (c); and removal/trapping strategy (B) where the analyte is not retained and elutes, while the interferents
are retained on the sorbent.

HOW DOES SPE COMPARE TO LIQUID-LIQUID EXTRACTION?


Liquid-liquid extraction (LLE) separates an analyte through partitioning between two immiscible solvents. LLE suffers from several
disadvantages that can be overcome using SPE, including the need for expensive glassware (separatory funnels), the large and costly
amount of organic solvents needed for separation, as well as the less than quantitative recoveries that can stem from low values of KP
or incomplete phase separations. For a review of LLE, refer to Section 2 of this module.
The general steps involved in performing SPE are:
1. Pre-treat the sample (e.g. dilution, adjustment of pH)
2. Condition the cartridge (run water or solvent through it)
3. Load the sample
4. Elute the fractions
SPE methods start with a conditioning step (Step 2 above). This consists of rinsing the SPE cartridge with a solvent similar to the
matrix of the sample you want to load. For an aqueous sample, we would choose an aqueous solvent such as water for conditioning.
Once the cartridge has been conditioned, the sample is then loaded on the cartridge (Step 3; note Steps 3 and 4 can be viewed in Figure
3). The amount of sample loaded will vary depending on the purchased cartridge. The cartridges can be loaded on a vacuum manifold
to apply a vacuum or centrifuged to speed up the separation process. This separation can also be done using gravity, but will take more
time. Once the sample has been collected on the adsorbent, the cartridge can be rinsed with 100% H2O to remove un-retained
compounds. More rinses can then be done with increasing percentages (ex.: 5%, 10%, 20%, 100%) of an organic solvent, such as
isopropanol, to remove compounds in terms of decreasing polarity.
Q1. Explain the difference between liquid-liquid extraction and solid-phase extraction.
Figure 3 shows an SPE procedure on a grape Kool-Aid sample. In this example, a polar solvent is passed through a less polar adsorbent
material. This is referred to as reversed-phase (RP) SPE. For this example, let’s say we are interested in determining the amount of red
and blue dye in the sample. (Please note that the examples in this module involve separating dyes, as the separation can be easily
visualized. However, SPE also can be used to extract and separate colorless species.)
Q2. List one or two species that you might find in the Kool-Aid matrix. Do you expect these species to be water soluble?
Q3. Consider Figure 3. In steps 1 and 2, grape Kool-Aid is passed through the cartridge, followed by water. Which species are retained
on the adsorbent and which species are eluted with the water?

10/24/2019 2 https://chem.libretexts.org/link?155906
Figure 3. SPE of a grape Kool-aid sample. Looking left to right, the following are eluted through the cartridge: grape Kool-aid, 100%
water, 95% water/5% isopropanol, 90% water/10% isopropanol, 80% water/20% isopropanol, and 100% isopropanol. (Photos are
courtesy of Cynthia Larive.)
Q4. After the water is passed through the cartridge in Figure 3 (step 2), water/isopropanol mixtures with increasing amounts of
isopropanol are passed through the cartridge. Is the polarity of the solvent increasing or decreasing?
If you said “decreasing” you are correct. This is termed making the solvent “stronger”, i.e. more like the adsorbent, which in this case
is non-polar.
Q5. Consider the two scenarios at the beginning of this section for retaining and eluting analyte (dye molecules) and interferences.
Which scenario is being used in Figure 3?

TYPES OF SPE
In the Kool-Aid example, we used a SPE cartridge where the adsorbent was less polar than the solvent. We called this reversed phase
SPE. The first step in using SPE is to select the correct type of adsorbent for the type of separation that is needed. The type of SPE
sorbent is chosen based on the chemical properties of the species we wish to separate. It is important to have an understanding of the
sample matrix, and the analytes of interest, when selecting the type of SPE. For example, if the analyte of interest is charged or
ionizable, you would select an ion exchange cartridge, either anion or cation depending on the charge of your analyte.
The types of SPE fall into several categories including, reversed phase, normal phase, ion exchange (anion/cation), as well as
mixed-mode phases, which have the properties of more than one type of SPE material. Figure 4 describes the properties of these
phases.

10/24/2019 3 https://chem.libretexts.org/link?155906
Figure 4. Various types of SPE with details about their separation mechanisms.

ADSORPTION (REVERSED PHASE)


Reversed phase SPE techniques are optimal for analytes with moderate to low polarity, and separate analytes based on hydrophobicity
with the most polar compounds eluting first. Because many analyses involve analytes dissolved in an aqueous sample (e.g. surface
waters, wastewater, urine or plasma), SPE using reversed phase sorbents are commonly employed.

Figure 5. Red and blue dyes of Kool-Aid separated using reverse phase SPE.2
Let’s think back to the Kool-Aid example. Here we want to again separate the red and blue dyes that form the purple colored Kool-Aid
solution. Figure 5 shows a cartoon similar to the process in Figure 3.
The process of separating the red and blue dyes within the purple Kool-Aid sample will start by loading an aliquot of sample onto a
reversed phase cartridge that has been pre-conditioned with water. The first fraction collected will contain mostly water and
compounds of high polarity and the dye will be retained in the adsorbent. Rinse steps with increasing concentrations of isopropyl
alcohol (IPA) will be introduced to elute the red and blue dyes separately. Once the dyes have been collected, the IPA is increased to
elute non-polar species within the sample.
Q6. Figure 5 depicts the separation of dyes within the Kool-Aid sample. The red dye appears to elute before the blue dye with only a
small addition of IPA. Please explain why the red dye elutes first.

10/24/2019 4 https://chem.libretexts.org/link?155906
ADSORPTION (NORMAL PHASE)
Normal phase SPE techniques are commonly used when the analyte of interest has low to high polarity, or is neutral.3 The cartridge
contains a polar adsorbent, such as silica. This separation will be based on polarity, with the least polar components eluting first. The
sample is usually in a non-aqueous matrix. For example, a researcher wishes to determine the concentration of polycyclic aromatic
hydrocarbons (PAH) in coffee samples. Many PAH are carcinogenic and can be introduced during the roasting process. The coffee
beans are ground up, dissolved and the oily residue extracted into the organic solvent 1-pentane. This extraction sample contains the
PAH but also potential interferents. Next, SPE is used to clean up the sample and/or concentrate the PAH. A “bind and elute” (Figure
2A) sample cleanup strategy is employed. A PAH, such as benzo(k)fluoranthene shown in Figure 6, is dissolved in the 1-pentane
phase, but the pi-electrons can interact with the polar functional group (e.g. silica) on the solid phase. When the sample is loaded onto
the SPE cartridge, the PAH are retained onto the stationary phase. Next, the solvent is changed and the PAH are eluted from the
cartridge.
Q7. In order to elute the PAH from the cartridge, how would you alter the polarity of the solvent?

Figure 6. Chemical structure of the PAH benzo(k)fluoranthene.


Q8. A pesticide has been extracted from a soil sample. The pesticide is relatively non-polar and soluble in organic solvents. However,
it contains a polar functional group. Describe a procedure to separate the pesticide from potential interferents (such as hydrocarbons) in
the sample.
Q9. If we wanted to use SPE in the preparation of soda and tea samples, which type of adsorption would we use (Normal or Reversed
phase) and why?

ION EXCHANGE (CATION/ANION)


Ion exchange SPE lets us separate compounds based on charge. Let us look back to the example of 4-aminophenol and acetaminophen
(Section 1 on liquid-liquid extraction). Notice that the 4-aminophenol has an –NH2 group that will be positively charged at an acidic
pH, while the acetaminophen will not. We would conclude that the best way to separate these two compounds would be by charge
using ion exchange.
There are two types of ion exchange SPE, cation exchange and anion exchange. For this example, we would select cation exchange
because the analyte we wish to retain is cationic (positively charged). This separation will be a basic 3-step approach to SPE (Figure
2A), starting with the first step of loading the sample onto the cartridge. The cation exchange cartridge will retain the positively
charged 4-aminophenol by electrostatic interaction. The first wash step will elute the negatively charged and neutral species such as
acetaminophen. To elute the 4-aminophenol that has been retained in the cartridge, a solvent with high ionic strength can be used. This
will introduce a large amount of positively charged ions to compete for binding of the substrate, causing the 4-aminophenol to elute
from the cartridge.
Q10. Given the example of cation exchange, can you explain a scenario where anion exchange SPE would be necessary?
Q11. Assuming you have two analytes of interest (refer to Figure 7), using ion exchange SPE, please explain how you could isolate the
two analytes separately.

10/24/2019 5 https://chem.libretexts.org/link?155906
Figure 7. Process of eluting two analytes of interest separately.2

MIXED MODE
In a perfect example of bind and elute SPE, only the analyte of interest would be retained while all interferents would be rinsed off
leaving a pure, concentrated sample with little to no loss. Unfortunately, this scenario is unlikely with the use of a single SPE cartridge.
Many times, conditions strong enough to remove impurities will also remove some analyte, reducing the recovery. Mixed-mode SPE
can enhance analyte retention and/or reduce matrix impurity contamination.
Mixed-mode SPE combines the capabilities of ion-exchange and reversed-phase SPE for an enhanced separation. Examples of mixed-
mode SPE include:
Reversed-phase/strong cation-exchange
Reversed-phase/strong anion-exchange
Reversed-phase/weak cation-exchange
Reversed-phase/weak anion-exchange
The ion-exchange component of the cartridge allows the appropriately charged compounds to be retained for further extraction. Once
the appropriately charged compounds are retained, the SPE can be rinsed to remove any impurities. The pH of the eluent can then be
adjusted to reduce the charge on the analyte molecules, which will release them from the ion-exchange component of the cartridge
causing them to elute. The mixed-mode SPE system also offers a reverse-phase mechanism. This will allow you to change the organic
concentration of the eluent to achieve improved selective elution.
Mixed-mode is commonly used in the separation of basic, acidic, and neutral compounds in a mixture. This is widely used in the
concentration, purification, and analysis of pharmaceutical compounds and antibacterial agents found in wastewater.

SUMMARY
Use the table below to answer the following questions about an SPE method.
The compound punicalagin (Figure 8), is present in pomegranate juice. Punicalagin analysis in pomegranate juice can be a method of
quantifying how much pomegranate juice is in a juice blend. Note that punicalagin is colorless and it is anthocyanin compounds that
give pomegranate juices their red color.

Figure 8. Chemical structure of punicalagin.


Q12. Consider the various SPE modes we discussed. Choose and defend your choice for the most appropriate methods to extract
punicalagin from pomegranate juice. Suggest both an adsorbent material and eluents.
Table 2: Summary of SPE Modesa

10/24/2019 6 https://chem.libretexts.org/link?155906
Mode Adsorbent Retention Mechanism Sample Matrix Analyte Elution Scheme Example
Properties Applications

Reversed Non-polar Non-polar or hydrophobic interactions Aqueous samples Exhibits non- Disrupt reversed-phase interaction Drugs and
Phase C2 Biological fluids polar with solvent or solvent mixtures of metabolites in
C8 Aqueous functionalities sufficient non-polar character biological
C18 extracts Most organics Methanol, acetonitrile, fluids
Phenyl Environmental Alkyl, dichloromethane Environmental
water samples aromatic, Buffer/solvent mixtures pollutants in
Wine or beer alicyclic water
functional Aqueous
groups extracts of
tissues and
solids

Normal Polar Polar Interactions Non-polar Exhibits polar Polar interactions disrupted with a Clean-up of
Alumina H-bonding Organic extracts functionalities more polar solvent or solution organic extracts
Silica (Si- pi-pi of solids Hydroxyl, acetonitrile, methanol, of soils and
OH) dipole-dipole Very non-polar carbonyls, isopropanol sludge
Diol induced dipole solvents amines, combination of solvent /solvent Fractionation of
cyano Fatty oils, double bonds mixtures petroleum
amino hydrocarbons Hetero atoms hydrocarbons
florisil (O,N,S,P) PCB’s in
resonance transformer oil
Isolation of
compounds in
cosmetics

Ion Quaternary Electrostatic attraction of charged Aqueous or organic Cation exchange Electrostatic interactions disrupted Pharmaceutical
Exchange amine functional groups of the analyte to samples of low walt – for basic via: compounds and
(anion) oppositely charged functional groups concentration compounds such pH modification to neutralize drugs in
Sulfonic on the sorbent (<0.1M) as amines compound and/or sorbent biological
acid Biological fluids functional groups fluids
(cation) Solution-phase Anion exchange increase salt concentration Fatty acid
synthesis (>1M); or use a more selective removal from
– for acidic
reactions counter-ion to compete for ion- food or
compounds
exchange binding sites agricultural
samples
Clean-up of
synthetic
reactions
Organic acids
in urine
Herbicides in
soil

a
Table adapted from Sigma Aldrich website4

REFERENCES
1. For further information on SPE, see Chapter 7 “Obtaining and Preparing Samples for Analysis” in Analytical Chemistry 2.0
(accessed September 13, 2018)
2. Waters Corporation Solid Phase Extraction online Tutorial http://www.waters.com/waters/en_US/SPE-Method-
Development/nav.htm?cid=10083845 (accessed September 13, 2018)
3. Agilent Application Note: Sasso, Pat “PAHs in Chocolate and Peanuts with Agilent J&W Select PAH and Longer GC Columns”
http://www.agilent.com/cs/library/applications/5991-2299EN.pdf (accessed September 13, 2018)
4. Sigma Aldrich Bulletin 910 “Guide to Solid Phase Extraction”
http://www.sigmaaldrich.com/Graphics/Supelco/objects/4600/4538.pdf (accessed September 13, 2018)

10/24/2019 7 https://chem.libretexts.org/link?155906