Cobas c311
Cobas c311
kadar zat-zat yang terkandung dalam darah Metode: Spektrofotometer Prinsip: melakukan prosedur
pemeriksaan kimia klinik secara otomatis mulai dari pemipetan sampel, penambahan reagen,
inkubasi serta pembacaan serapan cahayanya, yang dapat dilakukan dengan cepat dan bisa
digunakan untuk memeriksa sampel dengan jumlah yang banyak secara bersamaan.
1.Perpraring the analyzer
Always take suitable safety measures as d
escribed in the Operator’s manual
a.
Check that the container for highly concentrated waste is clean and empty. b.
Check that the cell covers, pipetter probes, ISE sipper nozzle, and the rinse nozzles are clean. If
trouble occurs, take the necessary remedial action.
2.
Switch on instrument. When the automatic power up is specified in the analyzer software, the pipe
function will be executed automatically. b.
When the analyzer is in sleep mode, it starts up automatically by using the sleep pipe function
(Power Up Pipe with starting time)
3.
System Overview
–
Workflow Guide
Daily Maintenance a.
Daily maintenance
Select
Execute
4.
OK
5.
Reagen preparing
a.
Select the preventive action check box on the system overview screen.
Reagent preparing
Yes The reagent load list will be printed for reagents with yellow, purple, or red alarm. b.
Load reagent onto the instrument according to the reagent load list. c.
Before replacing a reagent bottle, put the analyzer in Maintenance mode or Shutdown status. b.
After replacement turn the maintenance switch back to Operation mode (or power up the analyzer if
it was in Shutdown status). c.
Perform a Reagent Prime if an ISE Ref. Or ISE IS bottle has been replaced:
Reagent preparing
Reagent prime d.
Sellect all e.
Execute f.
7.
Load calibrator and QC samples onto the analyzer according to the calibrator and QC load list. d.
Start
8.
End of routine
a.
Load wash solutions for ISE wash: Pos. W1: ISE cleaning solution SysClean Pos. W2: ISE activator
SmpCln Position 1: Multiclean b.
Utility: Maintenance c.
Select: Execute
9.
Application, Calibration and QC set up Loading application parameters from cobas link
a.
Utility
Application
Download b.
Search d.
Download
OK f.
10.
Calibration
Install b.
Add c.
OK d.
Edit f.
Update
OK
11.
QC
Install
Download b.
Search d.
Download
OK f.