Image Quality Quick Test
Image Quality Quick Test
SP
© Siemens AG 1997
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPR2-230.037.01.08.02 Doc. Gen. Date: 03.03
Replaces: SPR2-230.037.01.07.02
0-2 Revision
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer 0
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Page
Document revision level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2
Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2
1 _______Introduction ___________________________________________________ 1 - 1
System ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Notes and symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Required measurement devices and auxiliary materials . . . . . . . . . . . . . . . .1-3
Test requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
ADR curves for fluoroscopy operating mode . . . . . . . . . . . . . . . . . . . . . .2-1
ADR curves for pulsed fluoroscopy operating mode . . . . . . . . . . . . . . . . . .2-3
ADR curves for DR operating mode. . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
3 _______Resolution ____________________________________________________ 3 - 1
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Edge enhancement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Noise reduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Motion detector (factory only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Page
SIREMOBIL Iso-C with keyboard. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Checking the newly programmed ADR curves . . . . . . . . . . . . . . . . . . . . . 8 - 1
System ID 1
Image quality acceptance test performed completely in the factory and documented by:
Name (print): __________________ Dept.: __________________
Signature: __________________ Date: __________________
Safety information 1
CAUTION When performing the work steps and tests, the product-specific
safety information contained in the documents, as well as the
general safety information TD00-000.860.01... must be observed.
CAUTION CAUTION used with the safety alert symbol indicates a risk of
danger that could result in slight or moderate physical injury and/
or damage to property.
NOTICE NOTICE used without the safety alert symbol indicates a risk of
danger that, if disregarded, could result in an undesirable result
or state other than death, physical injury or property damage.
• Service PC
Test requirements 1
• The entire system must be functioning properly; in addition, make sure that the following
are installed:
- Grid,
- X-ray I.I. cover, and
- Sirephos cover (without DHHS spacer).
• If the nominal values are not attained, the system must be adjusted or repaired according
to the service instructions.
• Mark the appropriate boxes y/n with "y" for yes or "n" for no.
Prerequisites 2
• The standard settings are programmed for fluoroscopy, pulsed fluoroscopy and DR (see
appendix).
• Attach the 2.1 mm Cu precision radiation filter for prefiltration to the X-ray port on the
tube assembly.
• Select "mid" as the dose rate; this corresponds to a dose rate of
X-ray I.I. 23 (9"): 0.185 µGy/s ±15% with survey format (see User-Setup).
X-ray I.I. 27 (11"): 0.14 µGy/s ±15% with survey format (see User-Setup).
Evaluation 2
• Select
• Select the image intensifier format; the LED of the button is off.
• Select . Select the image intensifier full image format; the LED of the button is
off.
• Select Sirematic HC 1 ADR curve; the LED of the mA button is off.
• Radiation on.
• Read the indicated kV and mA values on the control panel.
• Radiation off.
• Enter the values in Tab. 2.
• Select Sirematic HC 2 ADR curve; the LED of the mA button is on.
• Radiation on.
• Read the indicated kV and mA values on the control panel.
• Radiation off.
• Enter the values in Tab. 2.
• The actual values documented in the factory must be reproduced at the application site.
Permissible deviations: Tube voltage (kV) ± 1kV, tube current (mA) ± 10 %.
• Prerequisites:
- Select FLUO and "Mid" as the dose rate level.
- Select Sirematic HC 1 ADR curve; the LED of the mA button is off.
- Set edge enhancement for optimum recognition of resolution.
- Press the button for noise reduction;
the LED of the button must be off.
- Set the contrast (LUT 3 or 4) for the monitors to optimum recognition of resolution.
- Factory: use resolution test type 41 A.
- Application site: use resolution test type 41 A.
Attach the resolution test directly to the grid in the center of the image intensi-
fier. It should be located at an angle of approximately 45º to the grid lines.
- Factory:
Attach 25 mm AL measurement stand to the image intensifier.
- Application site:
If a 25 mm measurement stand is present (with 0.4 mm notch), attach it
close to the image intensifier.
Otherwise:
Attach the 17µm Cu strip directly to the grid next to the resolution test, and
place an additional 1.2 mm Cu filtration in the beam path, close to the tube.
- Switch radiation "ON." Determine the resolution of the LIH image and enter it in Tab. 1.
- If overframing occurs at the edge of the image intensifier, eliminate this with
collimation.
- Switch radiation "OFF." Determine the resolution of the LIH image and enter it in Tab.
1.
- In each case, check that the minimum contrast is detectable, i.e. whether the line pairs
indicate visible modulation.
• Evaluation
- Determine the line pairs/mm and enter the value in Tab. 1.
Fac- Application
NOTICE The dynamic test in conjunction with the Plexi capillary test is
used to display small differences in contrast.
• Measurement setup:
- Remove the 2.1 mm radiation filter from the beam path.
- Attach the dynamic test without the image intensifier fixing but with heart contour
diaphragm and Plexi capillary test directly in front of the I.I. input (Plexi capillaries close
to image intensifier).
• Prerequisites:
- Select FLUO ( button) standard settings (see appendix)
- Select Sirematic HC1. The LED of the mA button is off.
- Select full format by pressing the button.
- Select high noise reduction; the LED of the button is off.
- Edge enhancement ( button) off.
- Select LUT 1.
- Switch on radiation and evaluate the live image.
• Evaluation of the monitor image
- Nominal value
The unmarked Plexi capillaries in Fig. 1a must be visible.
- Cross off Plexi capillaries in Fig. 1b that are not visible.
.
Nomi- Fac- Applica-
Group
3 mm
wide
2 mm
wide
1 mm
wide
Subtraction mode 4
Measurement setup:
• Attach the dynamic test without the image intensifier fixing but with the heart contour
diaphragm and Plexi capillary test directly in front of the image intensifier input (Plexi
capillaries close to image intensifier).
Prerequisites
• Select SUB.
• Use the standard settings.
(Set the dose rate level at "High" in the user setup)
Noise reduction for mask: k = 32; check subtraction k-factor, phase A
Noise reduction for fill: k = MD2; check subtraction k-factor, phase B in Technical Setup)
• Select full format by pressing the button.
• Edge enhancement off; select = the softest display.
Procedure
• Release the scene (radiation on)
• After 3 seconds, the mask is automatically set.
The message "Inject" appears on the monitor.
• Start the Plexi capillary test by pressing the rubber ball.
• Shut off radiation after 3 seconds.
• Skip back one image for the white lines.
• Continue flipping back through images until the maximum contrast of the dark lines is
visible.
NOTICE Do not evaluate the first white line. Start the evaluation with the
first black line.
• Enter the results in table 1 and table 2. If a line is not recognizable, identify it with x.
• Evaluate the 3 mm black capillary line (the large line) in fields 2L, 1, and 5R on monitor B.
No obvious differences in brightness should be visible between the fields.
No obvious differences in brightness Application
Factory
are visible in fields 2L, 1, and 5R. site
y/n
y/n
Roadmap mode 4
Measurement setup:
• Attach the dynamic test without the image intensifier fixing but with the heart contour
diaphragm and Plexi capillary test directly in front of the image intensifier input (Plexi
capillaries close to image intensifier).
Prerequisites
• Select Roadmap.
• Use the standard settings.
• Set the dose rate level at "High" in the user setup)
Noise reduction for Phase A: k = 16; check subtraction k-factor, phase A
Noise reduction for Phase B: k = 8; check subtraction k-factor, phase B in Technical
Setup)
Noise reduction for Phase C: k = MD2; check in User Setup or Technical Setup)
• Select full format.
• Edge enhancement off.
Procedure
• Release the scene (radiation on; phase A)
• After >= 2 seconds, the mask is automatically set. The message "Inject" appears on the
monitor.
• Do not shift the Plexi capillaries. (radiation remains on; phase B)
• Shut off radiation after 3 seconds.
• Release radiation again. (Phase C)
• Start the Plexi capillary test by pressing the rubber ball.
• Shut off radiation after another 3 seconds.
NOTICE Do not evaluate the first white line. Start the evaluation with the
first black line.
• Enter the results in table 3 and table 4. If a line is not recognizable, identify it with x.
Prerequisites 5
- Attach the dynamic test without the image intensifier fixing but with the heart contour
diaphragm and Plexi capillary test directly in front of the image intensifier input (Plexi
capillaries close to image intensifier).
- Select FLUO.
- Contrast setting for monitors - step 1 (LUT = 1).
- Edge enhancement = 0
Edge enhancement 5
Noise reduction 5
- Press the button for selecting the noise reduction factor ; the LED of the button
must light up (low noise reduction).
- Switch fluoroscopy on briefly and assess the noise impression of the image during
radiation.
- Press the button for selecting the noise reduction factor ; the LED of the button is
off (high noise reduction).
- Switch on fluoroscopy briefly and assess the impression of the image with radiation on.
The noise should be less.
Fac- Applica-
- Noise reduction functioning o.k.? y/n y/n
(Image noise is less with the LED of the
key off).
• Prerequisites
- Select FLUO.
- Select low noise reduction (MD 1); the LED in the button is on.
- In addition, place a screwdriver on the center of the dynamic test.
- Radiation "ON."
- When activating the key for image rotation, the live image shows only a slight trailing
effect.
The image noise is more clearly visible in the moving parts of the image.
- Radiation "OFF."
- Select low noise reduction (MD 2); the LED in the button is off.
- Radiation "ON."
- When activating the key for image rotation, the live image shows a greater trailing
effect than for low noise reduction
The image noise is more clearly visible in the moving parts of the image.
- Radiation "OFF."
- Remove the screwdriver form the image intensifier.
• Evaluation of the monitor image
Fac-
- Motion detection o.k.? y/n
(Trial effect vs. image noise)
Prerequisites
• For SIREMOBIL Iso-C with 2 monitors, both must be set for approximately the same
(synchronous) brightness and contrast (LUT, brightness and contrast setting).
Preparations
• Attach the dynamic test without the image intensifier fixing:
- with heart contour diaphragm,
- without Plexi capillary test to the image intensifier.
• Select FLUO: dose rate level "Mid."
• Select Sirematic HC1 ADR curve; the LED of the mA button is off.
• Select LUT 1 for both monitors (if existing).
• Select edge enhancement at maximum recognition of bright/dark transitions.
• Switch X-ray image intensifier to full format.
• Open the collimator to maximum aperture.
• Prefilter with copper until 106 kV to 109 kV is displayed.
Switch on fluoroscopy briefly (approx. 6 mm Cu is required).
• Radiation ON.
• Evaluate the brightness of the fluoro image.
• Radiation OFF.
Evaluation 2
• Attach an additional 3 mm Cu at the radiation output.
• Radiation ON.
• Evaluate the brightness of the fluoro image.
• The brightness with approx. 8.1 mm Cu prefiltration and dynamic test should be
approximately equal to the brightness with 11.1 mm Cu and the dynamic test.
Factory
Brightness with ADR and iris diaphragm control the
y/n
same?
Image artifacts 7
• Cross off all image artifacts which have been detected during setting and image quality
tests in the table provided in the image quality test certificate.
• If any image artifacts are detected which are not listed in the table, these must be
described under "Other artifacts."
• To evaluate the respective artifacts, there are three rating numbers which indicate the
extent of each artifact:
• Hum:
Inhomogeneity caused by electro-magnetic fields in imaging systems. This artifact may
significantly influence the diagnostic value of the images depending on the degree.
Optimally, this artifact should not occur at all. However, extremely low levels can be
tolerated. Hum artifacts appear as periodic, horizontal bright and dark structures in the
image; they appear briefly and are not specific to one location.
• Interference stripes:
Very high frequency electro-magnetic fields appear in the image as bright or dark,
sometimes very short, transverse marks in the image. They appear briefly. Stripes
caused by contaminants on the surfaces of lenses, etc. should be included in this group.
In this case, they are specific to one location and appear constantly. Stripes cannot be
tolerated.
• Ghosting: These artifacts are object contours displayed twice, with the second contour
generally shifted laterally. They are caused by reflections in long, poorly adjusted video
cables. Clearly visible ghosting cannot be tolerated.
• Background structures are stationary, grid-shaped patterns primarily in dark regions of
the image. They are also referred to as "standing noise."
• Pixel errors are image pixels without image information. They are visible on the monitor
as bright and black dots the size of pixels.
Some types of pixel errors can be tolerated, while others cannot. The TV camera is
carefully inspected with respect to pixel errors at the TV test bay prior to shipment. TV
cameras can be turned over to the customer only if the number of pixel errors meets
factory specifications. These pixel errors can be tolerated and must be documented in
the IQ test certificate.
Image artifacts 7
Required value for the assessment of the artifact: Only 1 and 2 can be tolerated.
Factory Start-up
Type of artifact Rating of the artifact *1
1 2 3 1 2 3
Hum
Interference stripes
Ghost images (reflections)
Background structures
Pixel errors
Remarks:___________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
Other artifacts 7
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
_______________________________________________________________________
The customer can change the organ programs himself using the operating instructions.
The programmed parameters can be read out again by the customer by calling up "User
Setup."
Prerequisites
The test of the standard programmed ADR curves (Sirematic HC1 and Sirematic HC2)
was already performed. Refer to chapter 2 of these instructions.
• Select "Mid" as the dose rate level. This corresponds to a dose rate of
➪ X-ray I.I. 23 (9"): 0.185 µGy/s ±15% with survey format.
➪ X-ray I.I. 27 (11"): 0.14 µGy/s ±15% with survey format.
• If required, temporarily set the dose rate levels in the organ programs to "Mid."
Procedure
• Select FLUO.
• Attach a 2.1 mm Cu precision radiation filter for prefiltration to the radiation output.
• After selecting the respective fluoro organ program and activating the mA key, the
programmed ADR curves can be selected.
• Switch fluoro ON.
NOTICE Note the programmed dose rate level. To check the newly pro-
grammed ADR curves, temporarily set the dose rate level to "Mid"
(see above).
Since the Multispot 2000 must be optimally set for the film used and the development
required at the user's site, only a functional test is performed at the factory.
The functional check must be performed for all possible image subdivisions, in each case
with positive and negative image display.
Definition: Positive image display means that the hardcopy image is identical with the
monitor display.
Prerequisites 9
Multispot 2000 1/4, 008 550 035 042 020 660 035 042
full format
Multispot 2000 1/4, 008 550 016 019 020 660 016 019
4-on-1 image
Multispot 2000-2 008 550 021 025 020 660 021 025
Tab. 1
NOTICE In the case of deviations from the nominal value 2 (film density
40%), B must be corrected (adjusted).
The setting must be performed for all possible image subdivisions, in each case with the
image display (positive and/or negative) requested by the customer. Mark the settings not
performed with n/a in the relevant table.
Definition: Positive image display means that the hardcopy image is identical with the
monitor display.
• Film type used: ______________ Emulsion number: ______________
Multispot 2000-2 9
59
/ 1
A B
3 1
Fig. 1
Memory test image on the monitor [Film = Positive image display] (Fig. 1)
Application site:
Film density values for Multispot 2000-2 or Multispot 2000 1/4 with full-field image
and positive image display:
Application site:
Film density values for Multispot 2000 1/4 with 4-on-1 image and negative image
display:
Application site:
Film density values for Multispot 2000 1/4 with 4-on-1 image and positive image dis-
play:
Prerequisites 9
• If overframing occurs at the edge of the image intensifier, eliminate this with collimation.
• Switch radiation ON, and let it stabilize.
• Switch radiation OFF.
• Transfer (store) LIH image on monitor B.
• Select negative image display of the Multispot.
Multispot 2000-2 9
Image format Nominal resolution values [Lp/ Actual resolution values [Lp/mm]
mm] negative image display
negative image display
I.I. 23 (9") I.I. 27 (11") Factory Application site
Full format >1.2 >1.0
Preparations 10
Fig. 1 Side view of the test configuration Fig. 2 Test configuration viewed from above
Fig. 6 Dialog box for selecting patient position and the direction of radiation
Fig. 7
• Press the B button. This activates the "Start 3D Scan" button (Fig. 7).
• Now select scan protocol "Slow."
• Click on the "Start 3D Scan" button.
Fig. 8
Fig. 9
• Manually move the C-arm to the start position (into the limit switch).
Position: image intensifier located on main unit.
Fig. 10
Measurement run 10
During the scan, the following pop-up window will appear (Fig. 11).
Fig. 11
Fig. 12
Fig. 13 Syngo interface following successful data input and back projection.
Objective:
Determine an optimal slice plane for imaging the lead strip test pattern in a Syngo window.
• Select the plane perpendicular to the lead strip test pattern with the vertical bar
(see Fig. 14).
• Next, select a slice plane in the center of the lead strip test pattern (see Fig. 15).
• Next, by double clicking on the image "Slice parallel to lead strip test pattern," create a
full-field display of the lead strip test pattern (Fig. 17).
• Check the 3D high resolution (1LP/mm). For this purpose, it may be necessary to review
previous windows.
Fig. 1 Setting the I.I. laser light localizer correctly Fig. 2 Calibration phantom
• Take the required stand out of the "Calibration Phantom Service Case".
• Attach the calibration phantom to the stand.
• Adjust the laser light localizer to the tube assembly center (Fig. 1).
• Position the calibration phantom in the C-arm (Fig. 2).
With available light localizer:
• Using the laser light localizer, align the C-arm correctly to the markings on the calibration
phantom, first in the vertical (Fig. 3) and then in the horizontal (Fig. 4) C-arm position.
Without laser light localizer:
• Align the C-arm correctly to the markings on the calibration phantom, first in the vertical
(Fig. 3) and then in the horizontal (Fig. 4) C-arm position. Briefly perform fluoroscopy and
check the position of the C-arm relative to the calibration phantom (see also Fig. 5).
Fig. 5
Without 3D navigation 11
• After booting press Ctrl and D and the system switches into the 3D mode.
• Select Calibration Iso C 3D in the "Home Menu".
➪ The following window appears (Fig. 6).
Fig. 6
Fig. 7
• Click OK.
Angulation = 0 degrees
No. 3 100
Orbital = - 95 degrees
Angulation = 0 degrees
No. 4 100
Orbital = +95 degrees
The adjustment of the calibration
Intermediate step phantom must be checked once n.a.
again here
Angulation = 180 degrees
No. 5 50
Orbital = +95 degrees
NOTE The navigation shield (if fitted) must be removed from the image
intensifier before the calibration.
Fig. 8
• In the "Home Menu" service click the "Calibration Iso-C 3D" button.
Fig. 9
Fig. 10
Fig. 11
• Click OK.
Angulation = 0 degrees
No. 3 100
Orbital = - 95 degrees
Angulation = 0 degrees
No. 4 100
Orbital = +95 degrees
The adjustment of the calibration
Intermediate step phantom must be checked once n.a.
again here
Angulation = 180 degrees
No. 5 50
Orbital = +95 degrees
Angulation = 180 degrees
No. 6 50
Orbital = - 95 degrees
Angulation = 180 degrees
No. 7 100
Orbital = + 95 degrees
Angulation = 180 degrees
No. 8 100
Orbital = - 95 degrees
NOTE The relevant navigation shield must be fitted on the image intensi-
fier using knurled screws / Allen keys. The calibration must be
performed together with the navigation system and thus at the
same time with a technician of the navigation company. During
the calibration it must be possible for the markers on the image
intensifier navigation shield and the calibration phantom to be
acquired ("seen") by the cameras of the navigation system.
Care must be taken that the correct navigation system is selected
in the calibration window.
• In the "Home Menu" service click on the "Calibration Iso-C 3D" button.
➪ The following window appears (Fig. 12).
Fig. 12
Fig. 13
• Select the navigation system to be calibrated (e. g. 1st Navigation System, see Fig. 13)
(see software description setting for network nodes).
• Select the C-arm position. The selected C-arm position is displayed normally, all other
C-arm positions are displayed with a gray background.
• Bring the C-arm into this position.
• Select the number of images.
• Click "Go".
• Press the left foot switch, the measuring run is released by this. Keep the foot switched
pressed down (Fig. 14).
Fig. 14
• Click OK.
NOTE For the selected navigation system the calibration must be per-
formed in all 4 C-arm positions both with 50 and with 100 images
(8 calibrations).
Angulation = 0 degrees
No. 3 100
Orbital = - 95 degrees
Angulation = 0 degrees
No. 4 100
Orbital = +95 degrees
The adjustment of the calibration
Intermediate step phantom must be checked once n.a.
again here
Angulation = 180 degrees
No. 5 50
Orbital = +95 degrees
NOTE The calibration of the navigation system for the SIREMOBIL Iso-C
3D is performed by Medivision with its own calibration phantom.
Also the check of the total accuracy of the 3D navigation.
• In the "Home Menu" service click the "Calibration Iso-C 3D" button.
➪ The following window appears (Fig. 12).
Fig. 15
Fig. 16
• Select the C-position. The selected C-arm position is displayed normally, all other C-arm
positions are displayed with a gray background.
• Bring the C-arm into this position.
• Select the number of images.
• Click on "Go".
• Press the left foot switch, the measuring run is released by this. Keep the foot switch
pressed down (Fig. 14).
Fig. 17
• Click OK.
Angulation = 0 degrees
No. 3 100
Orbital = - 95 degrees
Angulation = 0 degrees
No. 4 100
Orbital = +95 degrees
The adjustment of the calibration
Intermediate step phantoms must be checked once n.a.
again here
Angulation = 180 degrees
No. 5 50
Orbital = +95 degrees
NOTE For the navigation system of the MEDIVISON with fixed mounted
image intensifier navigation shield:
The calibration of the navigation system for the SIREMOBIL Iso-C
3D is performed by the MEDIVISION company with its own calibra-
tion phantom. Also the check of the total accuracy of the 3D navi-
gation.
• Fasten the reference clamp to the phantom carefully (the phantom is the customer’s
property).
• Position the phantom in the isocenter of the SIREMOBIL Iso-C 3D.
• Perform 3D acquisition and transfer image data record to the navigation system.
• According to Fig. 15 approach in three planes in each case a marker cone with a pointer
up into the tip of the marker cone.
Marker
Fig. 18 Fig. 19
• If all three tips are within the imaged marker cone on the navigation screen of the
navigation company, then the accuracy of the 3D navigation is proven (Fig. 16).
Remarks:
Poor image quality after cali- Calibration process Check 3D limiting resolution
bration and recalibrate, if necessary
Grid holder on the I.I. is Clean the I.I. and the grid hol-
dirty der
No registration possible during Markers are not detec- Contact engineer of navigation
calibration with marker ring ted in the starting posi- company
Calibration does not start tion
• If the SIREMOBIL Iso-C covers were removed, perform the protective conductor test in
accordance with ARTD part 2 (CD-ROM).
You can call up Technical Setup for the memory by pressing CTRL + T.
Various test patterns can be called up from the first menu.
Standard programming 13
The following default values are programmed in organ program 1 when the SIREMOBIL
Iso-C is delivered:
Organ program 1
Duration of the phase B1 (*3) n.a. n.a. n.a. 0 sec n.a.
Disk tranfer rate phase B1 (*3) n.a. n.a. n.a. 0 n.a.
Disk tranfer rate phase B2 (*3) n.a. n.a. n.a. 3 n.a.