Reference Electrodes For Monitoring of Cathodic Protection On Buried Pipelines R11 Approved November 2018
Reference Electrodes For Monitoring of Cathodic Protection On Buried Pipelines R11 Approved November 2018
Reference Electrodes For Monitoring of Cathodic Protection On Buried Pipelines R11 Approved November 2018
CEOCOR Commission 2
External Corrosion and Cathodic Protection
Working Group E
Chaired by Brian S Wyatt: Corrosion Control, UK
Rev11: Approved by WG E and CEOCOR Board Brussels
November 2018
Page 1 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
TABLE OF CONTENTS
Page
Table of
Contents.......................................................................................................... 2
1.0 Scope.................................................................................................... 3
Appendix A 26
Appendix B 30
Target values for Maximum Agreed Errors (MAEs) and Parameters as Basis
of Purchase Specification
Page 2 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
1.0 Scope
Page 3 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
This is available from the CEOCOR web site1.
The terms and symbols used in this Document correspond to those used in EN
12954 and EN 13509, with the exception that the ENs refer to copper/saturated
copper sulphate and this document (to be consistent with terminology for other
reference electrodes) uses the term copper/copper sulphate (saturated).
1
CEOCOR document ‘Remote Monitoring and Control of cathodic Protection’ 2104 http://ceocor.lu
Page 4 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
3.0 Definition and Types of Reference Electrodes
EN ISO 15589-1 defines a reference electrode as “an electrode whose open circuit
potential is constant under similar conditions of measurement, used to measure the
structure to electrolyte potential”.
2
It is acknowledged that such measurements are liable to errors due to field gradients and unstable
pipe/soil potentials due to fluctuating stray current. The key issue is that the reference electrode’s own
electrode potential should not add unknown or significant additional errors or uncertainties to the measured
values by virtue of its own instability.
3
For example, AS 2832.1: 2015 Cathodic protection of metals. Part 1: Pipes and cables paragraph 3.2
Page 5 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
temperature.
The electrode itself should have a low internal impedance in terms of its
construction and its installation in the ground. Of particular importance is the
interface of the porous plug of the electrode which must be in intimate contact with
the ground and not affected by damage, impaction, contamination, or by poor soil
contact.
Other materials sometimes used as ‘reference electrodes’ (e.g. zinc used in sea
water) follow reactions which are not actually reversible. These are more accurately
known as pseudo-reference electrodes; they do not have a thermodynamic
equilibrium as there is no common component in the two adjacent phases. The
pseudo-electrode is simply a single body component that is placed in the same
electrolyte as the pipeline. There are advantages and disadvantages in using
pseudo-electrodes in soils. They are simple and rarely can be contaminated. They
can, when suitably selected, provide relatively stable electrode potential, although
they are affected by polarisation and are often only stable in limited ranges of soil
chemistry. Although they are generally not as initially accurate as reference
electrodes; they may be more stable over long periods (within a wider range) than
reference electrodes. This document addresses reference electrodes and not
pseudo-reference electrodes.
Reference electrodes are used in two discrete ways in buried pipeline applications.
4
CEOCOR document ‘Remote Monitoring and Control of Cathodic Protection’ 2104
5
CEOCOR document ‘Application of Coupons and Probes for Cathodic Protection Monitoring Purposes’ 2013.
Page 6 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
electrodes may also be used to deliver a pipe/soil potential for remote potential
monitoring or for automatic potential control of d.c. power supplies. In some
applications the reference electrode can be installed for permanent use within or
below an underground box to facilitate access to it; this permanent reference
electrode can then be relatively easily removed or calibrated when necessary.
The reference electrodes most widely used in buried pipeline CP applications are
copper/copper sulphate (saturated); Cu/CuSO4 (sat). These are available in a
variety of constructions for both permanent and portable applications.
𝐶𝑢2+ + 2𝑒 − ⇌ 𝐶𝑢(𝑚𝑒𝑡𝑎𝑙𝑙𝑖𝑐)
0 𝑅𝑇
𝐸 = 𝐸𝐶𝑢/𝐶𝑢 2+ + ln 𝑎𝐶𝑢2+
2𝐹
Page 7 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
Where:
E0Cu/Cu2+ is the standard electrode potential of the copper / copper sulphate
electrode, 0.337V at 25°C
R is the Universal Gas Constant = 8.315 JK-1mol-1
T is absolute temperature (K)
a is the activity of copper ions (Activity is dimensionless. The activity of a solute is
the actual concentration / 1M x the activity coefficient. The activity coefficient is
typically considered to be 1 for simplicity but is often not. For example, the activity
coefficient in 0.1M H2SO4 is 0.265. 10M HCl it is 10.44 and 1M NaCl is 0.657)
F is the Faraday Constant = 9.649x 104 C mol-1.
In environments where chloride ions can migrate into the copper sulphate solution,
the redox reaction is impaired, and the potential varies. This is evidenced by the
clear blue copper sulphate electrolyte becoming opaque and green/white in colour
and the copper rod turning dark green/black in colour. With 5000ppm chloride
contamination of the electrolyte the same source5 indicates that the error is ca. -
20mV (-0.850V measured as -0.830V) and with 12000ppm the error is ca. -120mV
(-0.850V measured as -0.730V).
6
Shreir’s Corrosion, 2010 Elsevier ISBN 978 0 444 52787 5
7
Ansuini FJ, Dimond JR, Factors affecting the Accuracy of Reference Electrodes, MATERIALS PERFORMANCE, Vol. 33,
No. 11, pp 14-17 (1994) November
Page 8 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
released into soil it can become attached to organic material and clay, sand, etc. in
the top layers of soil and may not move very far. When copper and copper
compounds are released into water, the copper that dissolves can be carried in
surface waters either in the form of copper compounds or as free copper or, more
likely, copper bound to particles suspended in the water. Even though copper binds
strongly to suspended particles and sediments, there is evidence to suggest that
some water-soluble copper compounds do enter groundwater. Copper that enters
water eventually collects in the sediments of rivers, lakes, and estuaries.8
Silver/Silver Chloride electrodes have not been used until relatively recently (mid
2000’s) as permanent reference electrodes for buried pipelines; however, they can
be usefully applied in high salinity soils where Cu/CuSO4 electrodes would
become contaminated and in brackish water crossings or estuarine crossings.
Their use has been reported in preference to Cu/CuSO4 electrodes for permanent
electrodes for buried pipelines following long term stability testing.9
𝐴𝑔𝐶𝑙 + 𝑒 − ⇌ 𝐴𝑔 + 𝐶𝑙 −
8
US Environment Protection Agency, Office of Solid Waste and Emergency Response. OSWER Directive 9285.7-68. Rev
February 2007
9
Wyatt B S, Permanent Buried Reference Electrodes: Technical Note on Significant Inaccuracies and Medium Term Trials,
CEOCOR Florence 2013
Page 9 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
Silver/Silver Chloride/Potassium Chloride electrodes are available with a range of
concentrations of potassium chloride electrolyte; they have different electrode
potentials dependent on the activity (concentration) of chloride ions used.
In the UK, a favoured concentration is 0.5M KCl and this electrode has an
electrode potential of approximately +0.240V with respect to the Standard
Hydrogen Electrode; this is within 5mV of the potential of the previously commonly
used Saturated Calomel Electrode (Hg/HgCl2/KCl(sat)), now in reduced use in
laboratories due to the health/environment threat from mercury and banned in
Europe for these reasons. It is similarly close in electrode potential to the
Ag/AgCl/sea water electrode when used in pure sea water as above.
Silver chloride-based electrodes are interfered with by reactions with other ions of
lower solubilities than silver chloride. The most likely ions to be encountered are
bromide, which is present in the sea at concentrations of 65mg/l. The effect of the
presence of bromide has been reported as increasing the potential of the electrode
by approximately 10mV in concentrations similar to those found in seawater,
although the effect varied from electrode to electrode. It is possible to make
silver/silver bromide electrodes or silver / silver iodide electrodes both of which
would not be affected by the presence of bromide.10
Electrodes made with silver salts are photoelectric. The silver salt decomposes to
silver and chlorine as electrons are emitted from the surface in response to incident
light. To avoid this, silver/silver chloride electrodes are packaged such that light
cannot penetrate them.
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
3.3 Other Electrodes
There are other electrodes offered in the market including Zn, Zn/ZnSO4, metallic
antimony electrodes which can be used in conjunction with an Ag/AgCl/0.5M KCl
reference electrode to measure pH, Redox electrodes (or reduction - oxidation
electrodes) for indication of microbial activity, etc. There are many other electrodes
used in the laboratory which are not commonly used by CP practitioners in the
field. These electrodes are not in extensive use for CP and are not intended to be
the subject of this document. This document does not address the particular
application of reference electrodes built within coupons; these are partially
addressed in the CEOCOR document on coupons.3
From EN ISO 12696: 2016 (CP of steel in concrete), the following equivalents are
established; CEOCOR have ‘rounded’ the numbers to maintain a ‘standard’ 0.240V
difference between vs NHE and vs SCE values; the quoted NHE values have been
assumed to be correct. Ag/AgCl/0.5M KCl and Cu/CuSO4 (sat) are added by
CEOCOR into the Table 1 below:
12
Force Technology, Park Allé 345, DK-2605 Brøndby, Denmark, ERE 20 electrode
Page 11 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
Table 1: Reference Electrode Potentials
13
Handbook of Analytical Chemistry. L. Meites, ed. McGraw Hill, NY, 1963
14
Friis E P, Anderson JET, Madsen LL, Bonander N, Moller P, Ulstrop J. Electrochim. Acta 43 1998 pp 1114 - 1122
15
Electrochemistry for Chemists 2nd Ed. Sawyer DT, Sobkowiak AJ, Roberts J Jnr. Wiley, NY, 1995
16
Electrochemical Methods: Fundamentals & Applications. Baird AJ, Faulkner LR, Wiley, NY, 2000
17
Application of Physical Methods to Inorganic and Bioinorganic Chemistry. Scott RA ed. Section Table 3. Wiley, NY, 2007
18
http://www.corrosion-doctors.org/References/Potential.html
19
Paper 243, NACE Corrosion 97, Arup H, Klinghoffer O, Mietz J
20
Handbook of Reference Electrodes, Inzelt G, Lewenstam A, Scholtz F, Springer Heidelberg, NY 2013 ISBN 978-3-642-
36187-6
21
Reference Electrodes (Half-Cells)", J Lichtenstein, Materials Performance, Oct., 2001
Page 12 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
Table 2: Temperature Variations of Reference Electrode Potentials
Cu/CuSO4 (sat)
Page 13 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
3.5 Electrode Construction issues
The porosity of the ‘porous plug’ will significantly affect the life and accuracy of both
permanent and portable reference electrodes but the effect is much more
significant for permanent electrodes which, when buried, cannot be easily
inspected, tested, cleaned and refilled with uncontaminated electrolytes. Not only
can contaminants such as chlorides (and other halides and sulphides) enter the
electrode and contaminate it but, for example for Cu/CuSO4 (sat), the copper ions
will also diffuse through the porous plug until the solution is no longer saturated and
the potential changes in accordance with the Nernst equation above.
The porous plug pore size and path length will determine the diffusion rates of
species through the plug. The concentration gradient across the porous plug will
also influence the rate of diffusion. As such the diffusion rate (loss) of chlorides
from a 0.5M KCl electrolyte in a Ag/AgCl/KCl electrode (see below) will be less
than from a 4M KCl electrode, if all other factors are identical. However, the
logarithmic relationship between potential and electrolyte concentration offsets the
potential change arising from diffusion.
Junction potentials also form across the porous plug due to differing diffusion rates
between cations and anions.
Most permanent electrodes and some of those portable electrodes which include
fixed concentration electrolytes utilise gelling agents to limit the diffusion of
electrolyte through the porous lug. Historically permanent Cu/CuSO4 (sat)
reference electrodes have used ‘gelling agents’ of wood dust (saw dust) and
‘plaster of Paris’ (gypsum plaster, or calcined gypsum, CaSO4.0.5 H2O, which re-
hydrates to CaSO4.2H20 when mixed with water) have been used. The latter has
also been used in a variety of Ag/AgCl/ KCl portable and permanent electrodes.
There are commercially available gelling agents such as carbopol polyacrylic acid
gelling agents neutralised with NaOH, acrylic acid, carboxy polymethylene,
carboxymethyl cellulose, methyl cellulose, algin, agar, polyethylene oxide, polyvinyl
alcohol, and mixtures thereof, as suitable gelling agents. Of these, agar and
agarose have been widely used for laboratory electrodes but suffer from drying and
microbial activity.20 Hydroxyethyl cellulose and methyl cellulose are both reported
to have been used commercially.20
22
In the case of measurements of coupon or probe potentials, the reference electrode is placed near the
coupon or the probe or may be integrated into the probe or coupon.
Page 14 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
steel in concrete, where the reference electrode is placed on the concrete surface, or
the measurement of the potential of an offshore structure, where the reference
electrode is positioned in the seawater.
The reference electrode should, preferably, be as close to the working electrode (the
steel), as possible. However, for practical reasons, this is often not the case. Thus, in
many situations, the reference electrode is some distance from the steel/electrolyte
interface. Consider, for example, the reference electrode inserted in the upper 10
centimeters of soil, used to measure the pipe/soil potential buried tens of meters
away.
There are two effects that arise under such conditions and that may markedly impair
the accuracy of the potential measurement. The first effects are “diffusion potentials”,
and the second “streaming potentials”.
In soil, the chemistry of the pore liquid will certainly differ from one location to another,
e.g. in terms of pH or concentrations of sulfate, chloride, etc. Particularly, a pH
gradient will give rise to pronounced diffusion potentials. In any case, this source of
error can easily exceed 10 mV or even 100 mV (strongly depending on the actual
conditions) and thus contribute a significant error in the pipe/soil potential
measurement.
In the case of movement of a liquid through a porous system, interaction between the
dissolved ionic species and the pore walls is likely to occur. This means that some
species are “held back” by this electrical interaction, while other species less strongly
affected. For instance, positively charged particles are decelerated, while negatively
charged particles are not (or less) affected. This gives again rise to a potential
difference that affects the measurement. These potential differences are usually
termed “streaming potentials”. Porous materials (soil, concrete, etc.) seldom exhibit
homogenous moisture distributions, and thus, streaming potentials arise. For the case
of concrete, these streaming potentials may easily exceed 100 mV.
As a conclusion, diffusion and streaming potentials are serious error sources for
potential measurements whenever the reference electrode cannot be placed in direct
vicinity of the working electrode. Their magnitude depends strongly on the actual
Page 15 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
conditions, but they can establish errors of several tens of mV or even exceed 100
mV.23
Particularly significant errors are likely when attempting to measure pipe/soil potential
values under concrete hard-standings or road crossings with portable reference
electrodes placed on the concrete surface.
3.7 Nomenclature
In all of the above, the strictly correct designation of, for example, the commonly
referred to Cu/CuSO4 (sat) reference electrode is actually Cu│CuSO4 (sat)
reference electrode. The Ag/AgCl/0.5M KCl reference electrode is more correctly
Ag│AgCl│0.5M KCl. The latter (using │) format is found in some scientific texts and
very occasionally in cathodic protection engineering texts by those trying to
impress. The former (using /) is in ubiquitous technical and engineering use and is
used in this document.
23
Angst U, Vennesland, Ø, Myrdal, R. Diffusion potentials as source of error in electrochemical measurements
in concrete, Materials and Structures 42 (2009) 365-375.
Page 16 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
4.0 Permanent Reference Electrodes
Reference electrodes are regularly buried for permanent monitoring close to buried
pipelines. The installation can be either by direct burial of the reference electrode
close to the pipe or, in some applications, the reference electrode can be installed
within or directly below a chamber or box with access at ground level, typically
above the pipeline. This is to facilitate calibration and replacement of the reference
electrode with the avoidance of excavation, resulting in reduced costs.
These four options result in different operational characteristics for the reference
electrode and may be important in determining the appropriate input impedance of
the measurement device and its influence on both the accuracy and life of the
Page 17 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
reference electrode.
These factors are seldom disclosed by manufacturers and there is little published
data on commercially available reference electrodes. These factors are likely to be
of particular importance in environments with freely draining soils such as sands
and gravels where the water table is above the electrode burial depth; if the
diffusion rate of the electrolyte from within the electrode body is too high, and the
volume within the body too low, significant errors will be established quickly.
A field study presented in CEOCOR 20137 presented field data for a range of
proprietary ‘high quality’ reference electrodes from reputable manufactures in
Europe and North America; many were significantly inaccurate after only a few
months of burial subjected to operating parameter a. above. One electrode type
exhibited errors of over 300mV, others had errors of around 200mV. ‘Old style’
large porous pot electrodes performed relatively well in the test period; only 3 of 11
Page 18 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
Cu/CuSO4 (sat) reference electrodes tested were stable to within 20mV during the
test period of 32 weeks. One type of Ag/AgCl/0.5M KCl reference electrode was
one of the most stable electrodes tested and was selected for permanent
installation.
Page 19 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
5.0 Portable Reference Electrodes
Generally, the accuracy and life of portable reference electrodes is not adversely
affected by characteristics of the measurement devices (unless inappropriately low
in quality and price plus low input impedance digital or analogue voltmeters are
used). The life of portable reference electrodes is generally determined by
mechanical damage, wear and tear and maintenance. Some electrodes (notably
the copper/copper sulphate (saturated) reference electrode) can be regularly
cleaned and recharged with electrolyte.
All portable reference electrodes can be and should be regularly check calibrated
according to their use.
All test leads shall be clean, fully insulated, fitted with appropriate plugs and
connection clips for the measurement device and test facility and should be
undamaged. Significant errors can be introduced if corrosion products, moisture
and ‘dirt’ are in place at the various connections.
The same issues are applicable, at least in part to permanently installed electrodes.
It is essential that all cable connections to electrode elements are
insulated/encapsulated in a suitable manner for long term burial/immersion and that
all connections and terminations are protected from contamination and moisture.
The use of multiple input impedance voltmeters (100MΩ and 1000MΩ units are
manufactured specifically for CP, historically some general-purpose digital
voltmeters have been switchable between 10MΩ and 1000MΩ) can be useful in
detecting if electrode to soil resistance is introducing errors into the potential
measurements. Historically Wheatstone Bridge based instruments with low
impedance moving coil meters were used which allowed the measurement current
to be zeroed and there to be no significant error due to electrode/soil contact.
Page 20 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
6.0 Reference Electrode Parameters: Permanently Installed Electrodes
I. Permanently buried reference electrodes are subject to errors and shall not
be used as the sole source of data for cathodic protection performance
evaluation without routine checking and evaluation of measurement
accuracy.
Page 21 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
This is particularly important, but difficult to install after construction, at locations
where the pipeline is covered by concrete or tarmac. Even in green field locations,
if the electrode is buried at pipe level, such provisions are valuable.
The following are optimum target performance parameters for permanent reference
electrodes; it is recommended that these are used in purchase specifications for
the reference electrodes. It is noted that most reference electrode producers do not
disclose the full details of their products or warrant accuracy or life.
This would allow for a mutual increase in understanding of the very basic
performance aspects of reference electrodes.
These techniques have been presented within the CEOCOR Work Groups but not
yet formally published. They show promise and involve the recording of field
gradients in fluctuating interaction conditions and the concurrent measurement of
permanent electrode to a portable electrode placed above it. If the field gradient is,
at times, zero then the voltage gradient between the permanent and portable
electrodes should also be zero. Thus, the measured potential difference at
effectively zero voltage gradient can be used as a check calibration value. 24
24
Verification of Permanent Reference Electrodes. Orta A, Gomila A. CEOCOR Work Group presentation
Weimar 2014. Accessible from Lluis Gomila <[email protected]>
Page 22 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
7.0 Reference Electrode Parameters: Portable Electrodes
Portable reference electrodes used for buried pipeline applications are almost
exclusively examples of the Cu/CuSO4 (sat) reference electrode.
Portable reference electrodes are generally purchased as proprietary items and not
against performance specifications or with extensive data sheets. The optimum
target performance parameters for portable reference electrodes that should be
used in purchase specifications for these units are given in Annex B.
Page 23 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
8.0 Installation Procedures for Permanent Reference Electrodes
The ideal position for installing the permanent reference electrode depends on a
number of factors:
The measurement value of interest influences the correct position of the reference
electrode. The parameter to be assessed e.g. on-potential, the local field gradients,
an approach to an IR-free potential, impact on the optimum reference electrode
location. Correspondingly the optimum position of the reference electrode may be
at remote earth, or at some "relevant position" or within 100-500 mm from the
pipeline at invert level.
The presence, location and magnitude of field gradients generated by the pipeline,
the anode bed or third-party infrastructure may influence the selected reference
electrode location. The field gradients may have a time dependence and no single
location may be optimum for all conditions.
The environment may change; both ground water and sea water level change with
time and it may be necessary to measure pipe/soil potentials at different elevations
around the pipeline.
When permanent Cu/CuSO4 (sat) reference electrodes are buried at pipeline depth,
in soil, the distance to the pipeline must be sufficient to prevent electro-deposition
of copper onto the pipeline. Similarly, if permanent reference electrodes are used
with coupons they should, if Cu/CuSO4 (sat), be spaced to prevent this effect, or
alternative reference electrodes should be used.
For installation post construction, ‘no dig’ or ‘minimum dig’ techniques can be used
for installation including hand and machine screw auger, water jetting auger and
vacuum auger techniques, to prepare a hole, sometimes lined, to electrode depth.
These techniques may dictate the use of reference electrodes of limited diameter.
Care should be taken when positioning electrodes so that they are not ‘choked’
inside tubes by debris or animals. If geotextiles are used to prevent sand of similar
ingress into tubes it shall be selected and tested to not be hydrophobic or to form
an electrolytically resistive barrier between the electrode and the soil.
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
more satisfactory in some applications to the provision of a plastic tube between
the ground level and the electrode. This is particularly important, but difficult to
install after construction, at locations where the pipeline is covered by concrete or
tarmac. Even in green field locations, if the electrode is buried at pipe level, such
provisions are valuable. These provisions for checking depend primarily on the
presence of IR-errors and are especially important in the case of field gradients
generated by third parties.
Page 25 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
9.0 Pre-Commissioning and Commissioning Procedures
Page 26 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
ANNEX A
The above scheme is one developed and reportedly in use by a large operating
pipeline company with many field CP operatives using many reference electrodes.
For those companies with fewer reference electrodes deployed it would be possible
to delete one layer of complexity as is detailed below in A-2
Page 27 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
A-2 Scheme for laboratory calibration/control of Reference Electrodes
Key:
1 : ‘Secondary Standard Reference Electrode (SSRE)’ (saturated calomel electrode
or Ag/AgCl/0.5M KCl electrode), 1. This electrode is only used in the laboratory and
is itself traceable back to a Primary Standard Reference Electrode
6 : Sponge
7 : Cap/stopper
Page 28 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
A-3 Calibration/Control Procedure
The difference of potential which should be read on the voltmeter is given by the
following formula:
E = E2 – E1,
With:
• E1 : potential of the Secondary Standard Electrode’
• E2 : potential of the ‘Reference or Master Site Reference Electrode’ which is
subject to this laboratory Calibration/Control
Page 29 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
ANNEX B
Annex B provides target values for Maximum Agreed Errors (MAE). This
terminology is based on the one used in Annex A.
Page 30 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
identifying the unique serial number shall be maintained as a record related
to the installation.
iv. The reference electrode design and quality should be such that during the
service life of the reference electrode its potential should be expected to
remain within +/- 30mV of their declared electrode potential, or some larger
Maximum Agreed Error defined by the user according to its needs and
practices. This shall be measured with respect to Master Working
Reference Electrode (MWRE) temporarily placed close to the permanent
electrodes, at intervals determined by the pipeline/structure operator and CP
assessment specialists advising them, as appropriate to demonstrate
adequate functionality and accuracy for the purpose for which the
permanent reference electrode is utilised in accordance with EN ISO 15589-
1. It is noted that many reference electrodes fail to meet this +/-30mV
requirement during service.
i. If the portable MRE is supplied complete i.e. ‘wet’ and filled with the
electrolyte, they shall be calibrated with respect to a Secondary Standard
Reference Electrode (SSRE) according to Annex A by the manufacturer
before despatch. The reference electrodes shall be within a Maximum
Agreed Error of their advertised potential, as defined by the purchaser
according to the needs and practices of the intended user. Typical values
are +/- 5mV of their advertised potential. Each electrode shall have a unique
serial number. The calibration record shall be made available to the
purchaser.
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR
programme and inspection and testing regime to prove that the electrodes
have this accuracy.
iv. Portable MREs shall be maintained clean externally, with their cables and
any terminals/leads in good condition. Cu/CuSO4 (sat) reference electrodes
shall be regularly visually inspected for their electrolyte level, colour (clear
bright blue), the presence of excess suspension of hydrous CuSO 4.5H2O in
de-ionised water, and the electrode condition which should be clean, bright
copper surface with no black or green deposits).
vii. The life of the portable reference electrode will likely be determined by
mechanical damage; wear to the porous plug or failure of seals especially
rubber seals. Some manufacturers offer replaceable porous plugs and seals.
A reasonable life expectancy is 5 years. In reality, most portable electrodes
are lost, and so having at least two spares is essential.
25
EN ISO 15589-1 Cathodic Protection of Pipeline Systems – Part 1: On Land Pipelines Annex D.3.2
Page 32 of 32
CEOCOR Commission 2
Reference Electrodes
BSW SF MB DC FB Rev 11 221118 Final Approved
© CEOCOR