HPLC OQ PQ OperatingInstructions
HPLC OQ PQ OperatingInstructions
HPLC OQ PQ OperatingInstructions
Chromeleon
Operational Qualification/
Performance Qualification
for HPLC Instruments
Operating Instructions
Version: 8.8
Date: Nov. 2013
Doc.: HPLC_OQ_PQ_OperatingInstructions_V8_8.docx
Doc. No. 4828.3250A
OQ and PQ Operating Instructions
OQ and PQ Operating Instructions
Contents
1 How to Use This Manual ................................................................................... 1
2 Introduction ....................................................................................................... 3
2.1 Defining the Limits .......................................................................................................................3
2.1.1 Operational Qualification (OQ) ...............................................................................................3
2.1.2 Performance Qualification (PQ) .............................................................................................3
2.1.3 System Suitability Check (SSC; also: System Suitability Test, SST).....................................3
2.2 Basic Requirements for Successful OQ and PQ ......................................................................4
5 Chromeleon 7 .................................................................................................. 33
5.1 Chromeleon 7 Terminology .......................................................................................................33
5.2 Creating the Sequences for the Qualification Checks ...........................................................33
5.3 Performing the Checks ..............................................................................................................37
5.4 Supported Instruments ..............................................................................................................38
5.5 Evaluating the Test Sequences ................................................................................................38
5.6 Selecting Special Test Procedures in Chromeleon 7 .............................................................39
7 Procedures ...................................................................................................... 63
7.1 Baseline Noise, Drift, and Lamp Intensity of the UV Detector ...............................................63
7.1.1 Theory ..................................................................................................................................63
7.1.2 Performing and Evaluating the Check ..................................................................................63
7.2 Wavelength Accuracy of the UV Detector ...............................................................................64
7.2.1 Performing the Check ...........................................................................................................64
7.2.2 Evaluation of the Check for UV Detectors ............................................................................64
7.2.3 Evaluation of the Check for Photodiode Array Detectors .....................................................64
7.2.4 Evaluation of the Check for Two-Channel Detectors ...........................................................65
7.2.5 Evaluation of the Check for Single Wavelength and VWD-3400RS Detectors ....................65
7.3 Linearity of the UV Detector ......................................................................................................65
7.3.1 Theory ..................................................................................................................................65
7.3.2 Performing and Evaluating the Check ..................................................................................66
7.4 Precision of Injection Volume ...................................................................................................67
7.4.1 Theory ..................................................................................................................................67
7.4.2 Performing the Check ...........................................................................................................67
7.5 Carry-Over by the Autosampler ................................................................................................68
7.5.1 Theory ..................................................................................................................................68
7.5.2 Performing the Check ...........................................................................................................68
8 Troubleshooting .............................................................................................. 81
8.1 General Notes .............................................................................................................................81
8.2 Failure of Individual Checks .....................................................................................................81
8.2.1 UV Detector ..........................................................................................................................81
8.2.2 Autosampler .........................................................................................................................82
8.2.3 Pump ....................................................................................................................................82
8.2.4 RF2000 Fluorescence Detector ...........................................................................................82
8.2.5 RI Detector ...........................................................................................................................83
8.2.6 ELS Detector ........................................................................................................................83
8.2.7 Corona detector ....................................................................................................................83
8.2.8 Electrochemical detector ......................................................................................................83
Note: Indicates general information to help obtain optimum performance of the instrument.
The information contained in this manual should not be construed as a commitment by Dionex. Dionex
assumes no responsibility for any errors that may appear in this document. This document is believed
to be complete and accurate at the time of publication. In no event shall Dionex be liable for incidental
or consequential damages in connection with or arising from the use of this document. We appreciate
your help in eliminating any errors that may appear in this document.
Tip: The descriptions in this manual refer to sequence templates version 8.8 or later. For
sequence templates of a later version than 8.8, the Release Notes for all sequence
templates with a version later than 8.8 complement this manual. The version number
of the sequence template is indicated in the name of the report template.
2 Introduction
The increasing number of standards and official regulations provide evidence that it is extremely
important to monitor the used instruments and to make sure that they work as intended if you want to
achieve reliable analytical results. To make the results transparent, quality management according to
ISO 9000 and following monitors and documents the quality of the equipment at different times.
This is the purpose of the Operational Qualification (OQ) and Performance Qualification (PQ)
procedures described in the sections below.
2.1.3 System Suitability Check (SSC; also: System Suitability Test, SST)
The purpose of SSC is to prove and document that the necessary limits are met for a specific
measuring application. The specific conditions required for that application, e.g., solvents, column
material, and temperature, must be taken into account. The check can be developed by the supplier
on request. However, it is not part of the test procedures below.
Do not use SSC limits that are more restrictive than the limits used for Performance Qualification.
3 Test Procedures
3.1 General Test Procedure
The following required materials and parts are provided in the Performance Qualification kit (part no.
4832.5000A).
The Standards kit (part no. 3323.0010) contains the seven required caffeine and pyrene standards.
The standard at position RA8 (or 8) contains water as solvent. Due to legal shipping restrictions, the
pyrene standard is shipped in solid form. Before you can use the standard, dissolve the solid pyrene in
1 mL of methanol (HPLC-grade). Complete the following steps:
• Unscrew the cap from the 2 mL vial labeled 3 µg Pyrene.
• Add about 1 mL of methanol (HPLC-grade), which is about half the vial volume.
• Screw the cap onto the vial. Make sure that the cap seals tight.
• Shake the vial for about 10 seconds to dissolve the solid pyrene.
• Place the vial at the appropriate position in the autosampler.
Note: The pyrene standard is used for checking the wavelength accuracy of all UV detectors (except
the Dionex VWD-3100 and VWD-3400 detectors, the AD25 detector, and any other supported single
wavelength detectors; see section 3.2). Concentration deviations of ±30% do not affect the test
results.
The kit with part no. 3325.0010 contains the five required standards for qualification of an RI detector.
Note that an additional flexible temperature sensor is required in addition to the Column Thermostat
PQ Kit when qualifying an ACC-3000(T) or ECD-3000RS column compartment. The temperature
sensor is available as Temperature Sensor Type K for Thermometer P600/P700, part no. 6820.0010.
For qualifying and electrochemical detector, you need one simulator cell (QualifierRS type) for each
potentiostat that you want to test; simulator cells are available under part no. 6070.4200. Please note
that the simulator cells have a limited life time (for details, refer to the certificate shipped with the
simulator cell).
Injection valve
1 Fitting, finger-tight, 33 mm
(6000.0011)
2 PEEK tubing, approx.
10 cm (2252.6001)
3 Single-part fitting, finger-
tight (2200.5502)
4 Union (754.ZU1M)
Thermostatted
column 5 Restriction tubing
(5040.3000)
Detector
If the pump units of a Dionex dual gradient pump are shared on two different systems, proceed
as described under "System" above. Qualification is comparable to a standard gradient pump
(e.g., Dionex LPG-3400SD).
• Qualifying a Dionex dual gradient pump (both pumps on the same system)
To qualify (within the same system) both pump units of a Dionex dual gradient pump with the
same autosampler, restriction tubing, and detector, you have to use an external motorized
switching valve, such as, a valve in the Dionex TCC-100, TCC-3100, TCC-3200(B) or TCC-
3000RS/SD. (→ Section 3.5.4 for a list of supported valves.) Refer to the figures below for
information how to connect the fluidics of the entire system.
to the waste
to the waste
Restriction tubing
Figure 2: Fluid connection for testing the Dionex dual gradient pump, using a 6-
port/2-position valve
The valve is in position A or 1, depending on the valve type
to the waste
to the waste
Restriction tubing
Detector
to the waste
Figure 3: Fluid connection for testing the Dionex dual gradient pump, using a 6-
port/2-position valve
The valve is in position B or 2, depending on the valve type
Restriction tubing
Detector
to the waste
to the waste
Figure 4: Fluid connection for testing the Dionex dual gradient pump, using a 10-
port/2-position valve
The valve is in position A
Pump unit (L = left, R =
right)
Autosampler
to the waste
Restriction tubing
Detector
to the waste
to the waste
Figure 5: Fluid connection for testing the Dionex dual gradient pump, using a
10-port/2-position valve
The valve is in position B
• Qualifying an Agilent system that is controlled via Agilent Instrument Control Framework
(ICF)
As described above under "System", the motor switching valve of the column oven is not required
to qualify Agilent systems. However, if an Agilent system includes a column oven with motor
switching valve, the created instrument methods will automatically set the valve position depending
on the valve type (corresponds to the first position in the Instrument Method Editor window). For
this case, make sure that the valve is not switched dry.
• Manual injection valve
Verify that the injection valve is fitted with a 10 µl sample loop.
• Autosampler
• Position the standards as shown in the tables on page 5 and the following pages.
• For the Thermo Scientific Accela autosampler, solvent reservoir BT1 (bottle) must be used.
• Column compartment
When qualifying the column compartments, the temperature sensor of the thermometer must be
securely attached to the heating block.
• When qualifying the Dionex TCC-100 or TCC-3x00(SD/RS) column compartments, the
temperature sensor must be fixed as shown in Figure 6.
• When qualifying the Dionex ACC-3000(T) column compartment and ECD-3000RS, install the
temperature sensor as shown in Figure 7. Be sure to use the type K temperature sensor (and
not the sensor from the Column Thermostat PQ Kit). Install the temperature sensor behind the
left capillary clip, from a vertical point of view in the center of the oven, and 2 cm away from
the right edge of the heat-conductive pad.
• When qualifying the Thermo Scientific Accela autosampler, the temperature sensor must be
positioned inside the oven near the internal temperature sensor.
• When qualifying the supported Shimadzu column compartments, loosen a fastening screw,
insert the sensor between the screw and the metal block, and carefully retighten the screw.
3.3.2 Configuration
• Agilent Instrument Control Framework (ICF)
The OQ/PQ templates can only be used to qualify "pure" Agilent systems (that include Agilent
modules only).
• RF2000 fluorescence detector
Before you connect the RF2000 fluorescence detector with Chromeleon, set the ZWAVE
parameter to 1. You can set this value only on the instrument. It is not possible to set the value
from Chromeleon. Complete the following steps:
• Disable the keyboard interlock by simultaneously pressing <Shift> and <CE>.
• Press <func> repeatedly until the ZWAVE command appears on the display.
• Press <1> on the number keypad and confirm with <Enter>.
• Enable remote operation. To do so, press <func>, until the RS232 command appears on the
display. Confirm with <Enter>, and then press <func>. On the display, the reading is
CONNECT. Confirm with <Enter>. You can now connect the instrument with Chromeleon.
• AD25 UV detector
On the Signals page for the AD25 in the Server Configuration program, change the unit to mAU
(instead of AU) and the factor to 1000 (instead of 1.00).
• ECD-3000RS EC detector
After you have connected the simulator cell(s) to the potentiostats, you must click "Read Smart
Cells" on the "Dectector" page of the ECD-3000RS driver in the Server Configuration program, and
then select the detected bays. In addition, check that the DC Mode (nA) option is selected under
"Mode and Range". Otherwise, no template for qualifying detector noise will be offered. Confirm
your selection by clicking "OK".
• WPS-3000(T)PL and WPS-3000(T)PLRS autosamplers
For successful qualification of the Dionex WPS-3000(T)PL and WPS-3000(T)PLRS autosamplers,
make sure that the Upgrade Kit for a 250 µl syringe is installed.
• WPS-3000T(B)FC Analytical autosampler
The "WPS-3000TFC/WPS-3000TBFC" check box must be activated in the Server Configuration on
the Options page of the autosampler:
• WPS-3000TBPL Analytical autosampler
In order to ensure a successful qualification, the Dionex WPS-3000TBPL Analytical autosampler
must be equipped with the Standard or Large Volume configuration. In addition, the WPS-
3000TBPL Analytical check box must be activated in the Server Configuration on the Options page
of the autosampler:
Tip: When changing the temperature sensor, you may have to adapt the
calibration values and sensor type setting of the thermometer. To do so, follow
the Operating Instructions shipped with the instrument. Otherwise, the
thermometer may show the wrong temperature. This is especially important
when qualifying the column compartment of the ACC-3000(T), which is
qualified using a type K temperature sensor.
Tip: Sequence templates created with Chromeleon < 6.50 can be used in
Chromeleon 6.50 or later only after you have deleted the
STH_manual.connect line from the COLUMN_OVEN program file. If you
created new sequence templates from the Chromeleon 6.50 or later master
templates, you do not need to adapt the program file manually.
Tip: Option C (qualifying the column compartment with manual data acquisition) is
not supported for Agilent, Shimadzu, and Waters instruments.
3.4 Preparations
Note: Even when qualifying only the detector (but not the pump and autosampler),
make sure that there is sufficient eluent (water) in the pump's eluent supply
and a restriction capillary is connected (see paragraph "Pump in this section).
• Autosampler
Before you start the check, rinse the autosampler thoroughly with water. To do so, inject 250 µl of
water at least five times. (If the allowed maximum injection volume of the autosampler is smaller,
inject five times the largest possible volume). Make sure that the fluid components and the syringe
are free of air bubbles. (Note: Although methanol is used as solvent for the first OQ and PQ check,
rinse the autosampler with water, as water is the solvent for all successive checks. Automatically
rinsing the system after the wavelength accuracy check ensures that the fluid system is sufficiently
prepared.)
• Pump
When qualifying an RI detector, rinse the entire HPLC system with water. If you want to check the
wavelength accuracy for the UV detector, disconnect the fluid components of the RI detector from
the HPLC system before you rinse channel A with methanol. When qualifying Dionex VWD-
3400RS detectors and all single wavelength detectors, or if the wavelength accuracy is not
checked, use water to rinse channel A. In this special case, you need not disconnect the fluid
components of the RI detector from the HPLC system. For gradient pumps, use water + 0.1 % Vol.
acetone to rinse channel B. For a ternary high-pressure gradient system, use water to rinse
channel C.
(a)
(b)
Figure 9: Performing OQ/PQ: (a) Step 1: OQ/PQ Setup; (b) Step 2: Instruments OQ/PQ
The PQ_OQ directory on the Chromeleon CD has the following subdirectories (→ Figure 10):
AS_AP_TEMPLATES,ED_TEMPLATES, HPLC_TEMPLATES, IC_TEMPLATES and
NANO_CAP_LC_TEMPLATES, Demo, and Reports.
The HPLC_TEMPLATES directory contains all master sequences required for OQ or PQ of a common
HPLC configuration. This directory has a SPECIAL HPLC TEMPLATES subdirectory for special
checks (→ section 4). When creating the sequence templates, the wizard provides only those
sequences that match the timebase. For IC and BioLC systems, the wizard provides the sequences
from the IC_TEMPLATES directory. for systems with an electrochemical detector, the sequences from
the ED_TEMPLATES directory are provided, and for nano, cap, and micro systems, the wizard
provides the sequences from the NANO_CAP_MIC_LC_TEMPLATES directory. For HPLC systems,
the wizard provides the sequences from the HPLC_TEMPLATES directory. The Dionex PDA-100 and
PDA-3000 detectors are included in the master sequences of the IC_TEMPLATES and
HPLC_TEMPLATES directories. These OQ/PQ operating instructions refer only to the sequences of
the HPLC_TEMPLATES directory.
(a)
• In the menu, click "OQ Setup" or "PQ Setup". A wizard guides you through copying of the
sequences. Click Next > to go to the next step.
• Select the timebase for which you want to perform OQ or PQ and enter the name of the computer
on which the timebase is installed.
• Select a unique name under which the sequence directory that contains all sequence templates
for this instrument is saved.
• A list of sequences is displayed. The list is adapted to the instrument configuration of the selected
timebase as defined in the Chromeleon Server Configuration program.
Tip: If you use the TSP UV1000 UV detector, you can select the sequences only if the UV
lamp is installed in the detector.
Note: When using Dionex detectors (UVD 170S, UVD 340S, UVD 170U, or UVD
340U) with non-analytical flow cells, you must enter the specifications listed
under 4.2 and 6.2.5 manually in the report, as automatic detection of the flow
cell is not possible or not implemented.
The serial number is entered automatically for the following devices: all devices of the Agilent
1100/12x0 series, all devices of the Dionex UltiMate 3000 series, all P680 and P580 pumps, the
Dionex ASI-100 autosampler, the Dionex PDA-100, AD25, UVD 340U, and UVD 170U detectors,
and the supported Shimadzu devices. For all other devices, enter the serial number in column K
from line 141 on. (The fields have a yellow background.) To delete the value in the related box, on
the Edit menu, click Clear Values. This removes the Chromeleon variable from the cell and clears
the audit.xxx entry for the cell on the status bar. When qualifying instruments that are not listed in
the table below, enter the model name in column H (cells with a yellow background), deleting the
existing audit.xxx entry as before. From line 159 on, enter the limits in the column with the related
model name.
• Enable the write protection (on the Edit menu, click Layout Mode), and then enter the
• SAVE the report. To do so, click Save Report Definition on the Workspace menu.
• To check the linearity of the UV detector, adapt the amounts in the QNT file of the sequence to the
actual amounts of the used standards.
Instrument Name
Devices names of the pump's eluent channels %A, (%B), (%C), (%D)
Device name of the external thermometer Thermometer
Signal name of the external thermometer TemperatureOVEN
Device name of the virtual channel VirtualChannels_01
Tips:
When you start the batch, the following warnings may appear:
{SOLVENT_CHANGE (91)} SOLVENT_CHANGE (91): Warning P0001: The program start time is undefined.
{SOLVENT_CHANGE - Sampler} Missing inject command.
{Pump} Eluent %A changed from Methanol to Water. Is this correct?
{OQ_COLUMN_OVEN (64) – TemperatureOven} Setting of property ‘Average’ overrides channel type default.
[Warning] {LONG_GRADIENT (161) - Pump} Ramp step duration (0.000166666666666667 min) is out of range. Minimum
supported duration is 0.01 min. Minimum duration will be used.
Warning: Sampler Program script for sample No. x contains no "Method" property assignment.
Warning: ECDRS Cells should be turned on before starting a queue. Otherwise, loss of data will occur.
For systems that are controlled by the Agilent Control Framework, the following warning may occur
(repeatedly):
{GRADIENT - LCSystem} Automatically resolving method inconsistencies. Inconsistencies: Parameters for Thermostat are set,
but not supported in current configuration.
Sequence templates created with Chromeleon < 6.50 can be used in Chromeleon 6.50 or later for
qualifying a thermostatted column compartment only after you have deleted the STH_manual.connect
line from the COLUMN_OVEN program file.
Tip: When you use a manual injection valve, make sure that no air is injected with the
samples. Always inject at least five times the sample loop volume, that is, at least 50
µl.
3.7 Duration
If the column compartment and the non-UV detectors are not included in the check, the entire check
takes approximately 3.5 hours. The additional duration for the other checks is as follows:
• 2 more hours when checking Dionex P680 or UltiMate 3000 pumps with mixing chamber extension,
an UltiMate 3000 LPG-3400M(B) pump, or an UltiMate 3000 LPG-3400BM pump (LONG_GRAD
sequence instead of the STD_GRAD or MICRO_GRAD sequence)
• 2 more hours when checking Dionex P680 DGP or UltiMate 3000 DGP pumps (standard
configuration)
• 4 more hours when checking Dionex P680 DGP or UltiMate 3000 DGP micro pumps, or pumps
with mixing chamber extension
• 3 more hours when checking the thermostatted column compartment
• 1.5 more hours when checking an RI detector
• 0.5 more hours when checking an ELS detector
• 1 more hour when checking an ECD detector
• 1 more hour when checking a fluorescence detector
• 2 more hours when checking a ternary high-pressure gradient system (channels C and B)
After the wavelength accuracy of the UV detector has been checked, that is, after approximately 15
min. or 3 h 15 min., you are prompted to change the solvent for channel A from methanol to water. If
necessary, connect the fluid components of the RI detector to the system. If an autosampler is
installed, OQ/PQ will then run automatically.
Note: For Agilent systems that are controlled via the Agilent Instrument Control Framework
(ICF), the procedure for changing the eluent takes several samples. Check the Audit
Trail for log commands that describe the necessary manual tasks. After the "Solvent
change step 1" sample, the batch is aborted. You are asked to change the eluent to
channel A and restart the batch. No more manual tasks are required afterward.
Exemption:
• It is not necessary to change the solvent manually when qualifying systems with a Dionex
VWD-3400RS or a single wavelength detector as UV detector.
When qualifying the ERC RefractoMax524 RI detector (micro variant), you will be prompted to
fluidically disconnect the detector from the system at the end of the RI Linearity sequence, as the
qualification sequence uses flow rates outside the detector specification.
Notes on compatibility:
When incompatible versions are used (HPLC OQ/PQ versus Chromeleon versus Agilent ICF), the
below error or a similar error may occur during Ready Check, depending on the instrument
configuration. Therefore, only use supported version combinations (see section ).
The sequence templates for these tests are available in the SPECIAL_HPLC_TEMPLATES directory
(see Figure 10). Start the OQ/PQ Setup from this directory as described in section 3.5.2.
4.2 Dionex VWD-3x00 Detectors: Noise and Drift with Dummy Flow
Cells
For qualifying the Dionex VWD-3100 and VWD-3400 detectors with the dummy flow cell, two
sequences will be offered.
• UV_NOISE_DRIFT_VWD3x00: Noise and drift measured at a wavelength of 254 nm
• UV_NOISE_VWD3X00_230nm: Noise measured at a wavelength of 230 nm
These sequences can be used only for the above detectors, and the test procedure requires that the
flow cell be changed twice. . If you want to perform both tests in a row, you do not have to exchange
the flow cell in-between the sequences (however, you still have to manually confirm the related
messages).
The following table shows the drift and noise limits for dummy flow cells at a certain wavelength. The
specifications have to be entered into the report manually.
Figure 18: Temperature sensor inserted into the vial (example showing a Dionex
autosampler)
• Dionex autosampler: Rotate the carousel until the carousel cover closes completely.
• Thermo Scientific Accela autosampler: Firmly close the autosampler door.
When the nominal temperature is reached, the external thermometer is used to record the sample
temperature over a period of 60 minutes. At the end of the check, you are prompted to remove the
temperature sensor. Afterward, the carousel rotation is restarted.
Tip: Do not perform any autosampler commands during the test. Movements of the needle
arm or carousel may damage the thermometer or autosampler.
Note: For Dionex instruments with a firmware version ≥ 4.07: If the check is interrupted,
the carousel rotation is not automatically turned on afterward. In this case, repeat the
check or use the "Tray_Rotation_On" program to turn it on again. Simply add the
program to the batch (object type: Program) and start the batch.
For Dionex instruments with firmware version < 4.07: These instruments do not
support carousel rotation. When you issue a command that relates to carousel
rotation, the following error message is displayed in the Audit trail:
[Error] 13:50:25 0.000 {Sampler} Unknown property. Perform a driver and/or firmware
update.
4.3.4 Duration
The test takes approximately 75 minutes.
4.4.1 Introduction
For determining the linearity of the Agilent series G1321 fluorescence detector, you must select the
"FLUORES_LINEARITY" sequence.
In addition to the detector to be qualified, this test requires an autosampler and an HPLC pump.
To determine the linearity of the Agilent 1100/12x0 series G1321 fluorescence detector, you need the
standards listed in the table below.
Note on eluent change: For Agilent systems that are controlled via the Agilent Instrument
Control Framework (ICF), the procedure for changing the eluent takes several samples. Check
the Audit Trail for log commands that describe the necessary manual tasks. After the "Solvent
change step 1" sample, the batch is aborted. You are asked to change the eluent on channel A
and restart the batch (the system still includes the restriction tubing from the general qualification
tests performed earlier). After the "Solvent change step 3" sample, the batch is aborted again.
You are asked to install the required column and restart the batch.
4.5.1 Introduction
In order to fully qualify the detector, select the CORONA_(VEO)_NOISE_DRIFT_SNR and
CORONA_(VEO_)RESP_CALIB sequences.
In addition to the detector to be qualified, this test requires an autosampler and an HPLC pump.
Qualification is performed by using the digital signal of the detector (USB port, unit: pA).
Note: Before qualifying the detector, make sure that the other modules in the system (for
example, the pump and autosampler) have been qualified successfully. Qualification
of the other system modules requires a supported UV detector. In this case, note the
information on Corona detectors in section 3.4.1.
The system is not automatically equilibrated. All the steps below must be completed manually.
• Ensure that the gas flow is on for a sufficient period of time (> 5 min - see Operating Instructions)
and that the gas pressure is within the required range (Corona, Corona ultra series: 35 psi ± 1 psi;
TM
Corona Veo series: 55-65 psi) while the pump flow is still turned off.
• Change the eluent from channel A [water/methanol 80:20 (v/v)] and purge the system for at least
5 minutes at a flow rate of 5 mL/min.
• Make sure that the capillary at the column outlet is not connected to the detector at this point, but
goes directly to the waste. Install the required column and equilibrate the system at a flow rate of
1 mL/min for at least 15 minutes.
• Connect the capillary from the column outlet to the detector inlet and equilibrate the system for
another 30 minutes.
• Finally, start the batch with the two qualification sequences.
5 Chromeleon 7
5.1 Chromeleon 7 Terminology
Tip: Please note that Chromeleon 7 terminology is different from the terminology used in
Chromeleon 6.80. For details, refer to the 'Glossary - Chromeleon 7.0,' which is
available in the Documents folder of your Chromeleon 7 installation.
• To start the wizard, in the "Tools" menu of the Chromeleon Console, click "Instrument
Qualification".
• Select the qualification type: Installation (qualification of the installation), Operational (qualification
in the working environment), or Performance Qualification (qualification during routine operation).
• On the third wizard page, you can select special test procedures for each module (see also 5.6 and
4). This option is available only for the following modules:
VWD-3100
VWD-3400RS
Corona (all supported variants)
WPS-3000T (+ Dostmann thermometer with channel TemperatureOVEN)
WPS-3000T (+ Dostmann thermometer with channel TemperatureOVEN)
Accela autosampler (+ Dostmann thermometer with channel TemperatureOVEN)
Agilent G1321A/B
• Select the sequences that you need for the tests that you want to perform. Mandatory tests, such
as "Warm up", are shown in the list, but the selection cannot be changed.
• On the last wizard page, select a unique name under which the OQ and/or PQ sequence directory
for this instrument is saved.
• Finally, the Instrument View dialog box open, showing the Queue tab. As soon as you start the
queue, Chromeleon runs the sequences.
Note: A qualification sequence will be offered for the supported column compartments and
ovens only if the signal name for the Dostmann thermometer is "TemperatureOven"
(also see section 3.3.2 – "Qualifying the Column Compartment").
As an alternative, click "Print" on the Sequence Editor toolbar to print the report.
Tip: The test for determining the sample temperature accuracy for the WPS, ACC-3000T,
or Accela autosampler is available only if the instrument configuration includes a
Dostmann thermometer with a TemperatureOVEN temperature channel.
The tests are prepared and performed as described in 4. Exception: CM7 includes a new driver for the
Dostmann thermometer that allows you to record temperature data directly as a signal channel. It is no
longer required to set up a virtual channel. All you need to do is go to the "Signals" tab page in the
driver configuration and change the signal name to TemperatureOVEN:
6.2.1 Pumps
The used test procedures are described in detail in sections 7.8 and 7.9.
DGP-3600RS / Precision
DGP-3600SD(N) Ripple ≤ 0.5 % ≤ 0.5 %
LPG-3400BM / Gradient Step gradient channel A/B ≤ 1.0 % ≤ 2.0 %
DGP-3600BM Accuracy Steps: 10, 50, 90% channel B
Flow rate: 1 mL/min
Gradient SD ≤ 0.3 % SD ≤ 0.5 %
Precision
Ripple ≤ 0.5 % ≤ 0.5 %
LPG-3400XRS Gradient Step gradient channel A/B ≤ 1.0 % ≤ 2.0 %
Accuracy Steps: 10, 50, 90% channel B
Flow rate: 1 mL/min
Gradient SD ≤ 0.2 % SD ≤ 0.5 %
Precision
Ripple ≤ 0.5 % ≤ 0.5 %
Thermo Scientific Flow Precision Flow rate: 0.3 mL/min RSD ≤ RSD ≤
Accela Pump Determined using the retention time 0.05 % or 0.1 % or
precision (standard deviation and SD ≤ SD ≤
relative standard deviation) of 0.01 min 0.02 min
caffeine. The greater value is the
valid limit.
Gradient Step gradient channels A/B and C/D ≤ 2.0 % ≤ 2.0 %
Accuracy Steps: 10, 50, 90% channel B
Gradient Flow rate: 1 mL/min SD ≤ 0.5 % SD ≤ 0.5 %
Precision
Ripple ≤ 0.5 % ≤ 0.5 %
Agilent 1100/12x0: Flow Precision Flow rate: 0.3 mL/min RSD ≤ RSD ≤
G1310A/B Determined using the retention time 0.07 % or 0.07 % or
G1311A/B/C precision (standard deviation and SD ≤ SD ≤
G5611A relative standard deviation) of 0.02 min 0.02 min
G1312A/B/C caffeine. The greater value is the
valid limit.
6.2.2 Autosamplers
The used test procedures are described in detail in sections 7.4 to 7.7.
PDA-3000 / Baseline Noise Measuring wavelength: 254 nm. 0.03 mAU 0.10 mAU
PDA 100 / Drift 1.0 mAU/h 1.0 mAU/h
6 6
PDA 100U Lamp Intensity >13 x 10 >10 x 10
counts/s counts/s
Wavelength Nominal wavelength 1: 272.1 nm ± 1.0 nm ± 1.0 nm
Accuracy Nominal wavelength 2: 333.3 nm
(maxima of pyrene)
Linearity Absorption range: up to 1.5 AU r≥ r≥
99.90 % 99.90 %
RSD ≤ 5 % RSD ≤ 5 %
Thermo Scientific Baseline Noise Measuring wavelength: 254 nm. 0.30 mAU 0.50 mAU
Accela PDA Drift 2.0 mAU/h 2.0 mAU/h
(analytical flow cell Lamp Intensity Not Not
with 1 cm path checked checked
length) Wavelength Nominal wavelength 1: 272.1 nm ± 2.0 nm ± 2.0 nm
Accuracy Nominal wavelength 2: 333.3 nm
(maxima of pyrene)
HPLC_OQ_PQ_OperatingInstructions_V8_8.docx – Version: 8.8 of Nov. 2013 Page 53 of 121
OQ and PQ Operating Instructions
Note: When a detector of the Dionex UltiMate 3000 series (VWD-3x00 or DAD / MWD-
3000) is qualified, the flow cell type is automatically detected and specifications are
automatically entered in the report.
When a detector of the Dionex Summit series (UVD) is qualified, the specifications
must be entered manually.
The used test procedures are described in detail in sections 7.1 to 7.3.
(1)
OQ limits with optimum measuring conditions, recommended PQ limits.
(2)
The manufacturer specification of ± 2 nm for the excitation and emission wavelengths can be
checked only by using a special flow cell and a mercury lamp. For OQ and PQ, the instrument
should preferably be checked with the components used for the measurements.
Note: When a detector of the Dionex UltiMate 3000 series (FLD-3100 or FLD-3400) is
qualified, the flow cell type is automatically detected and specifications are
automatically entered in the report.
(1)
OQ limits with optimum measuring conditions, recommended PQ limits.
7 Procedures
7.1 Baseline Noise, Drift, and Lamp Intensity of the UV Detector
7.1.1 Theory
Drift and baseline noise are important factors for UV detectors. Increased baseline noise considerably
reduces the sensitivity, as it is not possible to distinguish between low-level signals and noise. With
increased drift, it is more difficult to integrate the signals correctly because the less stable the baseline
is, the more inaccurate is integration.
The baseline noise of the detector mainly depends on the condition of the lamp. There is a
considerable increase in noise if an old lamp with poor light intensity is used. This is also true when
the flow cell is dirty. In addition, make sure that the measuring and ambient conditions are constant
and that the flow cell is free from gas bubbles.
To measure the drift of a UV detector, also make sure that the measuring and ambient conditions are
constant. In addition, it is very important that the lamp has been turned on for several hours. In the
detector environment, avoid drafts and direct sunlight.
Figure 30: Lamp drift directly after the lamp has been turned on (bottom
chromatogram) and after it has been turned on for six hours (top chromatogram)
The lamp intensity decreases while the lamp is in operation. In addition, lamps age when turned on
and off frequently.
To calculate noise, the measuring signal is split into 20 intervals of 1 minute each. For each interval,
Chromeleon calculates a regression based on measured values, using the method of least squares.
Parallel to the regression line, two lines are drawn through the values with maximum distance from this
regression line. The noise is the distances of these lines. The calculated values are averaged for all 20
intervals to establish the final value. To calculate the drift, Chromeleon calculates a regression line
from all data points within a range of 1 to 21 minutes based on the method of least squares. The slope
of the regression line is the calculated drift.
271.8 333.2
50,0
40,0
317.8
30,0
20,0
10,0
0,0
nm
-10,0
249,4 260,0 270,0 280,0 290,0 300,0 310,0 320,0 330,0 340,0 348,2
250
200
Peak Area / Peakfläche [mAU*min]
150
100
50
0
0 50 100 150 200 250 300 350
Figure 32: Linearity of the detector signal depending on the peak area
Tip: Depending on which injection module you use, it may happen that the peak height of
the sample with the highest concentration exceeds 1500 mAU. This is usually not
within the linearity range of UV detectors. Thus, the limits for the regression coefficient
and the relative standard deviation may not be met. In this case, reduce the injection
volume for all samples used for the linearity check so that the peak height of the
sample with the highest concentration is in the linearity range of the tested detector,
that is, usually below 1500 mAU. As an exception, a linearity range of up to 2500 mAU
is specified for the supported Shimadzu detectors and the Dionex VWD-3100 and
VWD 3400RS detectors with an analytical flow cell installed. The Dionex VWD-3100
and VWD-3400RS detectors with a semi-micro or micro flow cell installed have a
specified linearity range of up to 1700 mAU. If an autosampler is used when the
Dionex PDA-100 or PDA-3000 detectors are tested, only 8 µl of sample will be
injected by default. However, when qualifying the Dionex VWD-3100 and VWD-
3400RS detectors using an autosampler, 13 µl of sample will be injected.
Areawater ,corr : Area of the caffeine peak in the water sample (sample 9 – sample 6)
AreaConc:2000µg/mL : Peak area of the highly concentrated caffeine sample (sample 8)
Areawater ,CarryOver: Peak area of the water injection (sample 9: solvent and caffeine peaks) after
the carry-over sample (sample 8)
Areawater: Peak area of the water injection (sample 6: solvent peak) before the carry-
over sample (sample 8)
AreaReference : Peak area of the reference sample (sample 7)
cReference : Caffeine concentration of the reference solution (conc.: 10 µg/mL)
cConc:2000µg/mL : Caffeine concentration of the carry-over solution (conc.: 2000 µg/mL)
Note: The detection parameter settings for automatic peak integration aim to reliably
and precisely determine the integration line. However, as the peak shape of
water injections may be very small and noisy, it is not always possible to
ensure a correct automatic integration. In this case, we recommend to
manually correct peak integration.
40
35
30
20
15
10
0
0 20 40 60 80 100
7.7.1 Theory
The sample temperature accuracy mainly depends on the cooling and heating accuracy of the
autosampler, the insulation of the sample compartment, and the thermal transfer from the carousel to
the vial.
Figure 34: Theoretical (broken line) and real gradients (STD_GRAD standard sequence for
gradient pumps)
7.9.3 Performing the Checks for the Dionex P680 and UltiMate 3000 Pumps
Five sequences are provided for qualifying the Dionex LPG-3400A(B), LPG-3400RS, LPG-
3400SD(N), DPG-3600A(B), HPG-3200A, HPG-3200M, HPG-3200RS, HPG-3200SD ,HPG-3400A,
HPG-3400M, HPG-3400RS, HPG-3400SD, and P680 pumps: STD_GRAD, MICRO_GRAD,
LONG_GRAD, STD_GRAD_P680DGP_Left, and LONG_GRAD_P680DGP_Left.
The MICRO_GRAD sequence is used to qualify a pump with MicroFlow kit installed (e.g., an HPG-
3200M or HPG-3400M). The gradient program corresponds to the program of the STD_GRAD
sequence. However, the mixing volume of a pump with MicroFlow kit is lower than the volume of a
pump with standard mixing chamber, and therefore the detector signal is detected earlier. This fact is
considered for the evaluation of the check.
The LONG_GRAD sequence is used if a mixing chamber extension (600 µl or 1200 µl for UltiMate
3000 pumps, or 1250 µl for the P680 pump) is installed. A higher mixing chamber volume increases
the equilibration time of the gradient. The gradient program has been adapted accordingly and
evaluation of the check considers this.
For DGP pumps (P680, UltiMate 3000 DGP-3600A(B), UltiMate 3000 DGP-3600RS and UltiMate
3000 DGP-3600SD(N)), qualification is performed for both pump units. The sequences STD_GRAD
and LONG_GRAD are used to check the right pump unit. The sequences
STD_GRAD_P680DGP_Left and LONG_GRAD_P680DGP_Left are used to check the left pump unit.
For the UltiMate 3000 LPG-3400M(B), LPG-3400BM, DGP-3600M(B) and DGP-3600BM micro
pumps, only the LONG_GRAD and LONG_GRAD_P680DGP_Left sequences are available.
Tip: With the column compartment of the Waters Alliance 2690 Separation Module, the set
temperature can be changed during a sample only when the autosampler is injecting.
That is why 1 µl of water is injected for qualifying the column compartment of the
Waters Alliance 2690 Separation Module.
b) According to ASTM with noise evaluation at 397 nm (only for FLD-3100 and FLD-3400RS)
Average Signal Value 397 nm − Average Signal Value450 nm
SNR(ASTM) =
Noise397 nm
To determine the noise, the measuring signal is split into 40 intervals of 30 seconds each. For each
interval, Chromeleon calculates a regression line, based on the method of least squares. The noise
value is the distance between two parallel lines and the regression line through the lowest and highest
values. For the calculated values, the 40 interval values are averaged.
7.11.3 Performing and Evaluating the Check for Dionex RF2000 and RF1002
Fluorescence Detectors
The Fluorescence sequence is used to determine the noise and the signal height. When testing the
Dionex RF2000, make sure that the detector's ZWAVE parameter is set to 1 (→ 3.3.2). Water is
pumped through the flow cell at a flow rate of 1 mL/min. The excitation wavelength is 350 nm; the
emission wavelength is 397 nm.
To determine the noise, the measuring signal is split into 30 intervals of 30 seconds each. For each
interval, Chromeleon calculates a regression line, based on the method of least squares. The noise
value is the distance between two parallel lines and the regression line through the lowest and highest
values. For the calculated values, the 30 interval values are averaged.
7.12.3 Performing and Evaluating the Check for Dionex RF2000 and RF1002
Fluorescence Detectors
The Fluorescence sequence is used to determine the wavelength accuracy of the emission spectrum.
Water is pumped through the flow cell at a flow rate of 1 mL/min. For an excitation wavelength of
350 nm, the emission wavelength changes in 1 nm increments from 380 nm to 410 nm. The relative
signal maximum is compared to the theoretical maximum.
OQ_FLUORESCENCE #2 Fluorescence_Detector_Wavelength Emission
40,0
mV EM:380 nm
Em = 381
Em = 382
Em = 383
Em = 384
Em = 385
Em = 386
Em = 387
Em = 388
Em = 389
Em = 390
Em = 391
Em = 392
Em = 393
Em = 394
Em = 395
Em = 396
Em = 397
Em = 398
Em = 399
Em = 400
Em = 401
Em = 402
Em = 403
Em = 404
Em = 405
Em = 406
Em = 407
Em = 408
Em = 409
Em = 410
35,0
30,0
25,0
20,0
15,0
10,0
5,0
0,0
min
-5,0
0,00 0,50 1,00 1,50 2,00 2,50 3,00 3,50 4,00 4,50 5,00 5,50 6,00 6,50 7,00 7,50 8,00
Figure 35: Chromatogram for defining the relative maximum of the emission spectra
between 380 nm and 410 nm
7.13.1 Theory
The linearity of a fluorescence detector mainly depends on the optical and electronic systems, the
sample concentration and the eluent. With electronic systems, non-linearity is caused by dark current
and dark current drift. Dark measurements can be used to compensate the influence of these factors.
Contamination in the flow cell or optics and extremely high sample concentrations or the eluent may
cause stray light that influences the detector linearity. In addition, adsorption from the sample on the
cell walls can be found with very low sample concentrations. This effect also influences detector
linearity.
The influence of the sample in a suitable concentration range is fully used in this test procedure to
determine the detector linearity. Consider that the resulting deviations of the linear behavior are only
important with extremely high or very low sample concentrations. Therefore, the test results reflect the
influence of the detector itself on linearity.
For all checks, an eluent of water/methanol 80:20 (v/v) is pumped through the detector at a flow rate of
1 mL/min. For checks 4 and 5, six additional caffeine injections of 10 µL each are performed. The
standard concentration of the caffeine solution is 25 µg/mL (Corona and Corona ultra series), and
5 µg/mL (Corona Veo series). The settings for recording the detector signal are as follows:
• Filter: None (Corona, Corona plus)
Corona (Corona ultra)
4 (Corona ultra RS)
5 s (Corona Veo series)
• Nebulizer temperature: off (Corona, Corona plus)
25°C (Corona ultra (RS))
35°C (Corona Veo series)
• Data collection rate: 10 Hz
• Power function: 1
• Gain: 100 pA
• Offset 0%
To calculate noise, the measuring signal is split into 20 intervals of 1 minute each. For each interval,
Chromeleon calculates a regression line, based on the method of least squares. Parallel to the
regression line, two lines are drawn through the smallest and largest values. The noise is the distance
of these lines. The calculated values are averaged for all 20 intervals to establish the final value.
To calculate the height of the largest spike, the measuring signal is also split into 20 intervals of 1
minute each. For each interval, Chromeleon calculates the signal average value, the minimum, and
the maximum signal. The height of the positive spikes within an interval is calculated according to the
following formula:
Spike Height
Int. X
( )
= Signal Maximum Int. X − Signal AverageInt. X-1 + Signal AverageInt. X+1 / 2
The height of the negative spikes within an interval is calculated according to the above formula,
however, the signal minimum is used instead of the signal maximum. The absolute greatest height of
a spike of all intervals corresponds to the height of the largest spike. The signal average values of the
first and last interval outside of the measuring signal are extrapolated.
To calculate the drift, Chromeleon calculates a regression line from all data points within a range of 2
to 22 minutes based on the method of least squares. The slope of the regression line is the calculated
drift.
The signal-to-noise ratio (SNR) is calculated as follows:
Peak Height Average
SNR =
Signal Noise
The relative standard deviation of the peak heights of the six injections indicates the precision of the
peak height.
8 Troubleshooting
8.1 General Notes
• A system pressure that is well above 130 bar for a flow rate of 1 mL/min (solvent: water) after the
restriction tubing has been connected indicates that a capillary is contaminated. Inspect and
exchange the capillaries (including the restriction tubing) to ensure that OQ and PQ are correctly
performed.
• If problems occur during the checks that cannot be solved observing the notes below, also refer
to the respective sections in the instruments' manuals.
8.2.2 Autosampler
8.2.3 Pump
8.2.5 RI Detector
The possible causes for errors are diverse and are described in detail and using many examples in the
Operating Instructions for the instrument. Therefore, this manual does not give a list of errors.
9 PGM Files
As described in section 3.5, the OQ or PQ Setup Wizard adapts the program files to the device names
of the instruments installed on the timebase. This section lists the program files of the following Dionex
HPLC instruments:
; Pump DGP
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Methanol"
3DFIELD.MaxWavelength = 350.0
3DFIELD.MinWavelength = 250.0
DX_UV_VIS_1.Wavelength = 331
DX_UV_VIS_2.Wavelength = 333
DX_UV_VIS_3.Wavelength = 335
0.000
; Detector specific settings
DX_UV.Autozero
Wait DX_Sampler.Ready
Wait DX_UV.Ready
DX_Sampler.Inject
3DFIELD.AcqOn
DX_UV_VIS_1.AcqOn
DX_UV_VIS_2.AcqOn
DX_UV_VIS_3.AcqOn
DX_Pump_Pressure.AcqOn
1.500 3DFIELD.AcqOff
DX_UV_VIS_1.AcqOff
DX_UV_VIS_2.AcqOff
DX_UV_VIS_3.AcqOff
DX_Pump_Pressure.AcqOff
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
0.000
; Detector specific settings
DX_UV.Autozero
21.000 DX_UV_VIS_1.AcqOff
DX_Pump_Pressure.AcqOff
22.000 Log DX_UV.UVLampIntensity
Protocol "The lamp intensity is given at 254 nm in [counts/s]"
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
0.000
; Detector specific settings
DX_UV.Autozero
;==================================================================
; Macros used for pump modules
WriteReportMacro Name="PumpModule", Value="DX_PumpModule"
; Pump
WriteReportMacro Name="Pump", Value="DX_Pump"
WriteReportMacro Name="Pump_Pressure", Value="DX_Pump_Pressure"
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 2 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
0.000
; Detector specific settings
DX_UV.Autozero
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
;==================================================================
; Macros used for pump modules
WriteReportMacro Name="PumpModule", Value="DX_PumpModule"
; Pump
WriteReportMacro Name="Pump", Value="DX_Pump"
WriteReportMacro Name="Pump_Pressure", Value="DX_Pump_Pressure"
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 0 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
; Sampler settings
DX_Sampler.TempCtrl = On
DX_Sampler.ReadyTempDelta = 0.5 [°C]
DX_Sampler.Temperature.Nominal = 10.0 [°C]
0.000
Message"Make sure that the thermometer is located correctly in the vial (refer to
the manual for details)."
Message"Caution: To avoid damage to the system, do not perform sampler commands
while the test is running."
Wait DX_Sampler.Ready
Wait DX_Sampler.TemperatureReady
DX_Sampler.Inject
TemperatureOven.AcqOn
30.000 TemperatureOven.AcqOff
Message"The test is now complete. Please remove the thermometer."
log DX_Sampler.Temperature.Value
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
8.000 DX_Pump.%B.Value = 50.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_Pump.%B.Value = 99.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
12.000 DX_Pump.%B.Value = 99.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_Pump.%B.Value = 100.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
15.000 DX_Pump.%B.Value = 100.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_Pump.%B.Value = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
24.000 DX_UV_VIS_1.AcqOff
DX_Pump_Pressure.AcqOff
DX_Pump.Flow = 2.0
DX_Pump.%B.Value = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Pump.Pressure.LowerLimit = 0 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Sampler.AcquireExclusiveAccess
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
0.000
; Settings specific for complete systems
Wait DX_Sampler.Ready
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.Flow = 1.000
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
0.000
; Settings specific for complete systems
; Pump specific settings
Wait DX_PumpModule.Ready
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.Flow = 1.000
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
0.000
; Settings specific for complete systems
; Pump specific settings
Wait DX_PumpModule.Ready
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.Flow = 1.000
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
-40.000
; Detector specific settings
DX_RI.Autozero
0.000
; Pump specific settings
DX_Sampler.Inject
DX_RI_1.AcqOn
DX_Pump_Pressure.AcqOn
22.000 DX_RI_1.AcqOff
DX_Pump_Pressure.AcqOff
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_RI_1.Step = 0.2
DX_RI_1.Average = On
;============================================================================
; Macros used for pump modules
WriteReportMacro Name="PumpModule", Value="DX_PumpModule"
; Pump
WriteReportMacro Name="Pump", Value="DX_Pump"
WriteReportMacro Name="Pump_Pressure", Value="DX_Pump_Pressure"
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Sampler.AcquireExclusiveAccess
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
; Pump specific settings
DX_Pump.Flow = 1.000
DX_Pump_Pressure.Step = 0.01 [s]
DX_Pump_Pressure.Average = Off
DX_ELS_1.Average = On
DX_ELS_1.Step = 1.0 [s]
0.000
; Detector specific settings
DX_ELSD.Autozero
Wait DX_Sampler.Ready
Wait DX_ELSD.Ready
DX_Sampler.Inject
DX_ELS_1.AcqOn
DX_Pump_Pressure.AcqOn
21.000 DX_ELS_1.AcqOff
DX_Pump_Pressure.AcqOff
DX_Sampler.ReleaseExclusiveAccess
End
9.15 Baseline Noise, Signal Height, Spikes, and Drift of the Corona
Detector
;===========================================================================
; Noise and Drift for Corona Detectors
; --------------------------------------------------------------------------
; PGM-Version Jun 2011
;
; Column Holder: Shiseido Guard Holder, 20 mm for 2.0 and 4.0 ID cartridges
; Column: Shiseido 20 x 4.0 mm C18 MG 3 µm guard column
; Eluent A : Water/Methanol (80/20 Vol.%) (HPLC quality)
;
; Solvent degassed manually or via online degasser
;
; HPLC-System:
; ------------
; Pump specific settings
; Pump: Dionex LPG-3400(A/BM/RS/SD) Pump (UltiMate 3000)
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_Pump.Pressure.LowerLimit = 10 [bar]
DX_Pump.Pressure.UpperLimit = 300 [bar]
DX_Pump.%A.Equate = "Water"
DX_Pump.Flow = 1.000
DX_Pump.%B = 0.0
DX_Pump.%C = 0.0
DX_Pump.%D = 0.0
DX_RI_1.Step = 1.50
DX_RI_1.Average = On
DX_Pump_Pressure.Step = 0.01 [s]
DX_Pump_Pressure.Average = Off
Wait DX_Sampler.Ready
Wait DX_CAD.Ready
Wait DX_CAD.Nebulizer_WaitForTempReady
Wait DX_UV.Ready
0.000
; Pump specific settings
DX_Sampler.Inject
DX_CAD_1.AcqOn
DX_Pump_Pressure.AcqOn
22.000 DX_CAD_1.AcqOff
DX_Pump_Pressure.AcqOff
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock", Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
DX_RI_1.Step = 0.2
DX_RI_1.Average = On
Wait DX_CAD.Ready
Wait DX_CAD.Nebulizer_WaitForTempReady
Wait DX_UV.Ready
DX_Sampler.Inject
DX_CAD_1.AcqOn
DX_Pump_Pressure.AcqOn
2.500 DX_CAD_1.Acqoff
DX_Pump_Pressure.AcqOff
DX_Sampler.ReleaseExclusiveAccess
End
; Pump DGP
WriteReportMacro Name="PumpLeft", Value="$PumpLeft"
WriteReportMacro Name="Pump_Pressure_LeftBlock",
Value="$Pump_Pressure_LeftBlock"
WriteReportMacro Name="PumpRight", Value="$PumpRight"
WriteReportMacro Name="Pump_Pressure_RightBlock",
Value="$Pump_Pressure_RightBlock"
0.000 DX_ECDRS.Autozero
Wait DX_ECDRS.Ready
Sampler.Inject
DX_ECDRS_1.AcqOn
DX_ECDRS_2.AcqOn
21.000 DX_ECDRS_1.AcqOff
DX_ECDRS_2.AcqOff
End
10 Report
For an OQ report, refer to the following pages.
The report was generated for the following system configuration:
Operational Qualification
• Additional Information
Customer: Customer's Name
Operator: Operator's Name
Operator's Jobtitle
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Remark:
For instrument qualification not always all accessories are necessary - see HPLC_OQ_PQ_OperatingInstructions for details.
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
• Limits
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
271.9
50.0 333.3
40.0
317.9
30.0 261.0
20.0
251.1
10.0
0.0
nm
-10.0
250 260 270 280 290 300 310 320 330 340 350
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
Remark: Noise and drift are measured dynamically with a floated cell. The limits are different from
published specifications, because they are valid for static conditions only (empty cell).
1 0.024
2 0.018
3 0.021
4 0.019
5 0.024
6 0.027
7 0.020
8 0.024
9 0.020
10 0.017
11 0.023
12 0.016
13 0.026
14 0.016
15 0.018
16 0.022
17 0.022
18 0.018
19 0.025
20 0.026
Average: 0.021 mAU
Limit: 0.030 mAU
Result: ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Detector Noise
0.10
0.09
0.08
0.07
0.06
0.05
0.04
0.03
0.02
0.01
0.00
0 2 4 6 8 10 12 14 16 18 20
Segment No.
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
Back Pressure Device Capillary (L:15 m; ID:0,18 mm)
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
min
Caffeine Caffeine
$UV_VIS_1 $UV_VIS_1
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
1.00
0.60
0.20
-0.20
1 2 3 4 5 6 7 8 9 10
-0.60
-1.00
Sample No.
1.00
0.75
0.50
0.25
0.00
-0.25 1 2 3 4 5 6 7 8 9 10
-0.50
-0.75
-1.00
Sample No.
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
• Chromatogram Overlay
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
• UV Detector Linearity
Accessories Name
Back Pressure Device Capillary (L:15 m; ID:0,18 mm)
Sample 2 Caffeine in Water 10 µg/ml
Sample 3 Caffeine in Water 60 µg/ml
Sample 4 Caffeine in Water 140 µg/ml
Sample 5 Caffeine in Water 220 µg/ml
Sample 6 Caffeine in Water 300 µg/ml
Solvent A Water (HPLC-Grade)
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
• Calibration Curve
100
50
0 ppm
0 50 100 150 200 250 310
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
Remark: Noise and drift are measured dynamically with a floated cell. The limits are different from
published specifications, because they are valid for static conditions only (empty cell).
1 25.3 173.299
2 26.3 231.224
3 36.8 282.414
4 44.8 839.815
5 27.2 4.386
6 30.3 214.676
7 24.7 204.259
8 19.0 953.907
9 30.5 494.897
10 22.8 344.469
11 29.8 47.102
12 23.4 923.453
13 21.6 4.237
14 19.8 221.655
15 25.2 181.431
16 22.7 3.639
17 23.7 185.468
18 25.5 545.040
19 23.4 163.138
20 28.5 417.141
Average: 26.6 nRIU 321.8 nRIU/h
Limit: 50.0 nRIU 500.0 nRIU/h
Result: ok ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Detector Noise
75.00
60.00
45.00
30.00
15.00
0.00
0 2 4 6 8 10 12 14 16 18 20
Segment No.
Amount of Drift
1000.00
800.00
600.00
400.00
200.00
0.00
0 2 4 6 8 10 12 14 16 18 20
Segment No.
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
• RI Detector Linearity
Accessories Name
Back Pressure Device Capillary (L:15 m; ID:0,18 mm)
Sample 9 Glycerine in Water 5 mg/ml
Sample 10 Glycerine in Water 10 mg/ml
Sample 11 Glycerine in Water 15 mg/ml
Sample 12 Glycerine in Water 25 mg/ml
Sample 13 Glycerine in Water 35 mg/ml
Solvent A Water (HPLC-Grade)
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
30.0
20.0
10.0
0.0
0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0
RI RI RI RI
Correlation Coefficient
99.999 %
Limit: 99.900 %
Result: ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
• Injector Linearity
Accessories Name
Back Pressure Device Capillary (L:15 m; ID:0,18 mm)
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
• Calibration Curve
40.0 Caffeine External UV_VIS_1
Area [mAU*min]
30.0
20.0
10.0
µl
0.0
0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0
99.99979 % 0.238 %
Limit: 99.99000 % 0.500 %
Result: ok ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
Back Pressure Device Capillary (L:15 m; ID:0,18 mm)
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
2.50
2.00
1.50
1.00
2 - Caffeine - 0.567
0.50 1 - 0.200
3 - 1.150
-0.00
min
-0.50
0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.25
min mAU*min
Caffeine Caffeine
UV_VIS_1 UV_VIS_1
Limit: 0.100 %
Result: ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
1 0.118
2 0.109
3 0.099
4 0.074
5 0.101
6 0.109
7 0.123
8 0.069
9 0.089
10 0.066
11 0.073
12 0.097
13 0.149
14 0.080
15 0.076
16 0.074
17 0.151
18 0.052
19 0.129
20 0.110
21 0.091
22 0.112
23 0.109
24 0.079
25 0.128
26 0.113
27 0.101
28 0.113
29 0.108
30 0.073
Average: 0.10 mV
Limit: 0.30 mV
Result: ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
RF Detector Noise
0.16
0.14
0.12
0.10
0.08
0.06
0.04
0 5 10 15 20 25 30
Segment No.
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
The emission wavelength is changed from 380 nm to 410 nm in steps of 1 nm per 15 sec.
The maximum of the emission spectrum is determinated as maximum of the signal
of this chromatogram.
30.0
20.0
10.0
-5.0 min
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
200
150
100
50
min
-50
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
200
150
100
50
min
-50
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
200
150
100
50
min
-50
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories Name
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
200
150
100
50
3
2
0 1
min
-50
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Operational Qualification
Accessories
Thermometer SN: 43077
• Additional Information
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Chromeleon (c) 2013 Thermo Fisher Scientific Inc. PQ_OQ_Report_8_8 / COLUMN OVEN
Version 6.80 SR13 Build 3966 Printed: 14.10.2013 3:17 PM
Smp: Oven Temperature Runtime: 14.04.2004 11:29:06 Page 2 of 2
Seq: DQ_PQ_OQ_FOQ\DEMO_for_Manual\HPLC_OQ_DEMO_RUNS\OQ_COLUMN_OVEN
Temperature Temperature
80 80.4 0.4 ok
60 60.2 0.2 ok
30 30.3 0.3 ok
10 10.1 0.1 ok
______________________________ ______________________________
Reviewer's signature // Date Operator's signature // Date
Chromeleon (c) 2013 Thermo Fisher Scientific Inc. PQ_OQ_Report_8_8 / COLUMN OVEN
Version 6.80 SR13 Build 3966 Printed: 14.10.2013 3:17 PM