0% found this document useful (0 votes)
580 views2 pages

03.QSP-QC-003-Transmitter Calibration Procedure With RTD & TC

This document provides a calibration procedure for temperature transmitters (RTD/TC). It outlines the purpose, scope, reference documents, equipment, standards, and step-by-step procedure used to calibrate temperature transmitters at various temperature points to ensure accuracy. Temperature transmitters are calibrated using temperature baths and a multifunction calibrator by comparing the readings of the unit under test and a master thermocouple and calculating any error. The error must be within specified tolerance limits according to industry standards.

Uploaded by

Mangesh Mohite
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
580 views2 pages

03.QSP-QC-003-Transmitter Calibration Procedure With RTD & TC

This document provides a calibration procedure for temperature transmitters (RTD/TC). It outlines the purpose, scope, reference documents, equipment, standards, and step-by-step procedure used to calibrate temperature transmitters at various temperature points to ensure accuracy. Temperature transmitters are calibrated using temperature baths and a multifunction calibrator by comparing the readings of the unit under test and a master thermocouple and calculating any error. The error must be within specified tolerance limits according to industry standards.

Uploaded by

Mangesh Mohite
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 2

GAUGES BOURDON (I) PVT.

LTD
CALIBRATION PROCEDURE FOR TEMPERATURE TRANSMITTER
Rev. No.: 00
DEPT.:- QC-RTD-THERMOCOUPLE DOC. NO.: QSP/QC/002 Issue Date : 01.01.2015
Rev. Dt. : ---
PURPOSE :- This Test determines the accuracy of the Temperature Transmitter (RTD/TC).
SCOPE :- This procedure is used for calibration of Temperature Transmitter (RTD/TC).
REFERENCE DOCUMENTS:
a) IEC-60584 – Thermocouples-1-EMF specifications and tolerances/Thermocouples Tolerances
b) DIN -43760 – Electric Thermometer
c) ASTM E 644-04 – Industrial Resistance thermometers
EQUIPMENTS AND STANDARDS :-
1. PRT (PT 100) Sensor (4-wire) -Make-Isotech Range: -200 to 165°C
2. Cold Temperature Bath -Make-Nivtech Range: -45 to Ambient
3. Water Temperature Bath -Make-Ganesh Range: 30 to 95°C
4. Oil Temperature Bath -Make-Ganesh Range: Ambient to 150°C
5. Multifunction Calibrator (Calsys-50) -Make-AOIP Range: As per manual
6.mA/mV Calibrator -Make-Fluke Range:
PROCEDURE :-
1. Select the required temperature on the Temperature Bath & wait to stabilize the temperature.
2. Place the Master PRT/Thermocouple & UUT with Transmitter Thermocouple for calibration in the Temperatu
bath.
3. Connect the leads to the readout ensuring proper wire Connection.
4. Measure & Record the corresponding milliVolts of the Master & the UUT Thermocouple.
5. Measure the Ambient Temperature & the corresponding milliVolts for both.
6. Add the milliVolts value in the final readings of both Thermocouples.
7. This will be the reading of UUT & Master Thermocouple at 100 Degree Celsius.
8. Compare the measured readings of the UUT & the Master Thermocouple.
9. Error calculation, Error = [UUT TC Reading – Master TC Reading.]
10. This error should lie in specified tolerance limits as per IEC 60584
Special Note - Calibrate UUT Thermocouple as per the below chart & customers’ requirement for
different temperature points.

Sr. Thermocouple Type Calibration Points in Degree Celcisus


No. Minimum Maximum
1. J 100 500
2. K 100 500
3. E 100 500
4. T 100 200
5. N 100 500
6. R 500 1000
7. S 500 1000
8. B 800 1000

Prepared & Reviewed By Approved & Issued By