CKT3800-150A User's Guide
CKT3800-150A User's Guide
CKT3800-150A
Dry Block Temperature Calibrator
1.1 Attention
1.1.1 Please be sure to use this instrument as described in this manual. Otherwise, the
protective function of the instrument may be affected. See the warning and attention section for
security information.
1.1.2 The following definitions apply to the term "warning" and "attention".
1.1.3 "Warning" indicates conditions and actions that may harm the user.
1.1.4 "Attention" indicates the conditions and actions that may damage the instrument used.
Warning
Summary
Do not use this instrument for other applications other than calibration. The instrument is
designed for temperature calibration. Any other use may cause unpredictable damage to the
user.
Do not place the instrument under the cabinet or other objects. The top needs to be set aside
Special attention should be paid to the use of this instrument at high temperature for a long
time. It is not recommended that no one should be monitored at high temperature, and there
In addition to vertical placement, no other bearing operation instrument is allowed. Tilting the
Beware of burns
Never touch a thermostat at work.
Never use an instrument near combustible materials.
The use of this instrument at high temperature requires attention.
In constant temperature above 50 ℃, the screen will display the high temperature warning icon
and text. No matter whether the instrument is working or not, please do not remove the plugin
2
to avoid personal injury or fire.
Don't shut down the instrument when temperature higher than 100 ℃ . This can lead to
dangerous situations. Choose the set point is below 100 ℃, and close the output, and let it cool
before shut down the instrument.
2 Introduction
Easy to use and convenient, Easy and efficient temperature calibration., be extensive applied
in machinery, chemical industry, food, medicine and other industries.
At present, there is a disadvantage of slow heating and slow temperature in the field of dry
type calibration furnace in China, so it will take a long time for users to calibrate. The latest
generation of dry well furnace is designed with the most advanced heating principle in the world,
which has the characteristics of fast heating, fast temperature and fast cooling, which greatly
improves the existing calibration efficiency.
The temperature control circuit is measured with high precision sensor, which ensures the
precision is higher than that of other domestic manufacturers, and the technology meets the
international standard. The world's first touch operation is simple and fast.
·Multiple types of inserted into the pipe and can meet different size, the number of sensor test and
·Have been inserted into the depth of the same industry leading;
·5.0 -inch TFTLCD touch-screen, 16 bit true color image, simple and intuitive to use;
·With load short circuit, load circuit, sensor protection, and other functions.
3 Reference
3
2 3 4
5
8
6
9 7
4
3.3 Ready star
Follow these steps to quickly use:
Set the target temperature.
As shown in figure 4.1.1, click the setting temperature input box under the main interface,
pop up the temperature window, enter the target temperature, click "confirm" button, return to the
main interface, and the temperature setting is successful.
2. Start work
Click the "start" button on the main interface. The main interface work status icon turns green,
the instrument starts work, and the main output indicator flashes at a specific time interval.
3. Stop work
Click the "stop" button on the main interface and the output will stop.
4 operation instructions
5
4.2 Output Settings
The system adopts PID control to control the temperature of the thermostat, and a set of
setting parameters is preset when the system is delivered.Users can customize the Settings to meet
site usage.
Set the PID correlation parameter in the output setting interface, and the output setting
interface is shown in figure 5.2. Click the restore default button to restore all Settings to the
default value.
Note: after setting, press the "save" button to save the setting value, otherwise it is deemed to have
been abandoned
.
Figure4.3 Output Settings Interface
6
PID cycle setting
The adjustment operation period of the instrument is in seconds; Setting range: 0~100,
system preset to 3. This parameter has great influence on the adjustment quality, and the
appropriate numerical value can be used to solve the overshoot and oscillation, and obtain the best
response speed. It is recommended to modify the default values.
The proportional coefficient determines the size of the proportional band. The smaller the
proportional band, the stronger the regulation effect (equivalent to enlarging the amplification
factor); Conversely, the larger the ratio, the weaker the regulation. It is recommended to modify
the default values.
The integral time determines the integral action intensity. The integral time is short, the
integral action is strong, the time of eliminating static difference is short, but the overstrong
integral action may cause the temperature to be stable with a large oscillation. On the contrary, the
integral time is long and the integral function is weak, but the time to eliminate static difference is
longer. It is recommended to modify the default values.
Differential time determines the intensity of differential action. The differential time is long
and the differential action is strong. It is sensitive to temperature change and can reduce the
temperature overshoot. But strong differential action may increase the temperature oscillation
amplitude and lengthen stable time.
Power limit
Limit the main output power. The percentage indicates that the larger the value of 0~100, the
larger the output power is when heating, and the faster the heating, but may not be conducive to
the service life of the heating module. The system defaults to 90%.
7
4.3 system setting (part of the following functions need to
be authorized by the manufacturer)
System Settings include language setting, temperature setting, alarm setting, screen
brightness setting, temperature correction. The temperature correction button is the key to enter
the temperature correction interface. The system setup interface is shown in figure4.4.
8
message if the temperature is lower than the alarm limit. Set the range of 0 ℃ ~ 660 ℃, and
cannot be larger than the upper limit alarm.If don't need,set it to -1.
9
Click the "file" button in the menu interface to enter the file list button, as shown in figure
4.6.The file is shown in the form of an icon, and the text is not displayed on the icon. Users can
edit and save 10 files, and each file can record data of 6 sensors. Click the file icon to enter the
corresponding file editing interface.
3
7
10
1. File name: must be entered before save, up to 16 English characters, will be displayed on
the file list. User manual input is required;
2. Sensor information display area: store the information of the sensor, including the number,
index number and data unit; Click the corresponding attribute location, and the user can edit the
information.
3. Test temperature area: record the temperature value of the test. User manual input is
required;
4. Test data; Click the corresponding position input;
5. File operation button; Click the delete button to clear the file, click the save button to save
file 1;
6. Sensor flip button: a single file can record the data of 6 sensors, and the left and right page
can switch to display different sensor data.
7. Temperature data turning button: a single file can record the data of 10 temperature points,
and the corresponding data of different temperature points can be displayed by turning the page up
and down.
Note: click return and the system will also save the data.
11
4.5 View System Information
5 Technical indicators
Note: this technical indicators need to 23 + / - 5 ℃ environment, valid for stable under 10
minutes after the products reach the set temperature.
·Temperature:-20℃ ~ 150 ℃
·Temperature units: ℃ or ℉
·Temperature meter accuracy: 0.1%.
·Display resolution: 0.1 ℃ maximum display digits 5
·Speed: 30 ℃ to 100 ℃ for 20 minutes(Stabilizing time, 15 minutes); 30 ℃ to 150 ℃
40minutes(Stabilizing time, 40 minutes)
·Temperature stability: +/- 0.1 ℃
·Level temperature field: + / - 0.2 ℃
·The vertical temperature field: from soaking piece of hole bottom 30 mm range deviation is not
12
more than 1 ℃
·Insert depth: 160mm.
·The number and aperture of the sensor can be inserted: the standard is 4 holes, respectively, 6mm
8mm 10mm and 12mm.
Note: the outer diameter of the soaking zone is 30mm, and the insertion depth and outer diameter
of the sensor should be specified.
7 Maintenance
13