H8 Service Mannual
H8 Service Mannual
H8 Service Mannual
SERVICE MANUAL
IVD
Shenzhen Lifotronic Technology Co., Ltd.
Intellectual Property
Lifotronic reserves the right to update product technology without prior notice;
Lifotronic reserves the right to modify product specifications without prior notice.
Lifotronic reserves the right to modify the manual without prior notice.
Statement
Lifotronic makes no guarantee for this document whatsoever, including (but not
limited to) implied merchantability and suitability for a particular purpose.
Lifotronic shall be responsible for the safety, reliability and performance of its
product only when:
The assembling, expansion, readjustment, improvement and repair are carried out
by Lifotronic authorized personnel;
The machine is used according to operation instructions;
Electrical devices are in compliance with national standards.
Lifotronic shall not be responsible for the safety, reliability and performance
of its product if:
1
The product has reached its lifetime limit;
Parts have been disassembled, stretched or readjusted;
The machine is not used properly according to User Manual.
Warranty
Repair Service
(1)Lifotronic will provide fee-based service if the product is not within the range of
Lifotronic’s warranty provisions;
(2)Within warranty period, fee-based service will be provided if the product defect
is caused by:
Misuse;
Man-made damage
Replacement with parts not approved by Lifotronic;
Repair of the machine by personnel not authorized by Lifotronic;
Grid voltage beyond equipment specifications;
Uncontrollable natural disasters.
2
Return of Product
CONTACT
Manufacturer:Shenzhen Lifotronic Technology Co., Ltd.
Register Address: 4th Floor, Building 15, 1008 Songbai Road, Nanshan
District, Shenzhen, China
Manufacturing Address: 4th Floor, Building 15, 1008 Songbai Road, Nanshan
District, Shenzhen, China
Postcode:518055
Tel:86-755-29060026
Fax:86-755-29060036
Service Line:86-755-29060197
Web:en.lifotronic.com
3
●The system is operated and used by Lifotronic or designated
agents trained professionals, doctor or lab assistant;
●If hospitals or institutions can not achieve a satisfactory repair /
maintenance plan, it may cause abnormal instrument failure, and may endanger
human health.
●Ensure to run the instrument under the conditions that are written in this
manual. If not, it may fail to run, result in inaccurate results, or even damage the
components and injure humans.
4
CONTENT
1.1Preparation
2.2 Methodology
3.1 Diagram
4.2.1 Overview
6
Chapter 5 Machinery& Maintenance
Analyzer
6.1 Tool
7
Chapter 1 Installation of Instrument
1.1 Preparation
1.1.1Destination
According to this section, the steps are introduced and can be successfully installed on
the client side.
1.1.2 Tool
■ Blade or pliers
■ Pipette (10-100ul and 100-1000ul range)
■ Cross-head Screwdrivers
8
Caution:
● Additional electrical interference may be introduced with the use of the
switchboard, resulting in erroneous analysis results. Please select a place close to the
power outlet to install the system and avoid using the cutting board.
● Please use a random power cord.Use of other power lines may damage the system
or lead to incorrect analysis.
The environment should be clean, no mechanical vibration, no pollution, no noise source and
the power interference.
Keep away from strong electromagnetic interference sources, so as not to affect the normal
operation of the equipment.
Do not get close to brush type motor, flashing fluorescent lamp and frequent switch of
electrical contact equipment.
Avoid direct exposure to strong light or before the source of heat and wind.
9
Indoor use.
2. After removing the packing belt, remove the top cover of the wood, you can see the
attached package in the box, and take out the attached package, as shown in the
figure.
3. Remove the packaging bag from the main machine, then 2 people lift the bottom of
the machine to the left and right side, and carry it to the workbench.
Caution:
10
In the process of moving the machine, Minimize inclination and avoid the big impact.
11
2.Cut the 2-d arm with the cutting pliers and clean the cut belt
Flow Direction
Caution:
• Have to follow the direction of the arrow indicated on the label to install the
column, not reverse assembly.。
• Do not oscillate or shake the column.
• When the pressure is increased at the time of testing (about 4.0MPa higher
than the initial pressure), a new line filter should be replaced immediately.If
the pressure is not reduced, a new column should be replaced.
• More than 2 days no testing, it is suggested that install the Column in the
opposite direction, to perform the activation process, and seal it with plug then
stored in the refrigerator2 ~ 8 ℃, prevent dry chromatography column from
losing efficacy;
• Please amend the chromatography column and filter at ordinary times stored in
2 ~ 8 ℃ refrigerator, just turn off the instrument when there’s no testing.
12
1.4.4 Installation of Printing Paper
1) Turn on the wrench in the upper right corner of the thermal printer and open
the printer door.
2)In the direction shown in the following figure, the printer paper is put into the
paper cavity, and the paper head is outside the outlet of the paper.
3) Close the door of Printer.
4) Check the position of the printing paper and make sure the printing paper is
aligned with the paper.(the printing paper is 57mm wide and 5CM in diameter)
13
out the reagent filter components respectively from the attachment and
insert into A,B,C three kinds of eluent, and tighten the cap.The last,
1.5.2 From the attachment, the L liquid reagent tube assembly is removed,
and one end is connected to the L liquid joint on the back of the
1.5.3 The waste liquid sensor component is removed from the attachment
and one end is connected to the waste liquid junction on the back of the
instrument, and the other end is inserted into the waste bottle.
14
power adapter and power cord;
2) check and ensure that the instrument has properly installed the column and filter;
3) check and ensure that the 2-d arm fixed screw is completely removed;
4) check and ensure that ABC elution fluid, hemolytic agent and waste liquid junction
pipe are not folded, and the connection is positive.
Thermal LCD
Printer Screen
High Pressure
2D Arm Module
Pump
Negative
Sampling System
Pressure pump
System
LED Lamp
Test Module Panel
Waste Pump Control Board (Colorimetric
15
Pool)
Photoelectric
100ul Plunger Pump Switchboard
SN Name Functional Specification
1 Control Board Core Board+MCU+Pump&Valve Motor Drive+Optocoupler
Detection +Temperature Pressure Monitoring;
2 Pinboard Main control board all signal transfer, centralize the line;
3 Interface Board The whole machine external USB, Ethernet, power supply;
4 LED lamp Provide a light source of 415nm+500nm of testing module;
panel
5 Photoelectric Electrical signal converted by Light signal of
Switchboard
415nm+500nm of testing module;
2.2 METHODOLOGY
17
①----High Pressure Pump Components ②----Colorimetric Pool Module ③----Pressurizer
④----Online Deairing Machine
18
①----Power On/Off ②----Power Input
③----USB Connector ④----Ethernet
⑤----Hemolytic Agent Joint ⑥----Waste Liquid Joint
⑦----Waste Liquid Sensor
19
Chapter 3 Liquid Line of Instrument
3.1 Diagram
20
Filter SV11 P1(Waste Liquid Diaphragm Pump) Waste Bottle
Waste Bottle
21
3.4 High Pressure Pump Cleaning Route
Waste Bottle
front and back end of high pressure pump, there should be leakage
happened to
pump.
22
3.5 Sample Probe Inside Cleaning Route
The cleaning of the inner wall of the sampling needle is to insert the
sampling needle into the dilution pool and then use the 5ML plunger
pump to inject the hemolytic agent into the sampling needle, to clean
the sampling needle, and the waste liquid is pumped away by the
diaphragm pump.
23
3.6 Sample Probe Outside Cleaning Route
24
3.7 Sample Needle Sampling and Proofing
The Sample Needle will aspirate 5uL sample from the venous whole
blood with the power generated by 100uL plunger pump and dispense it
25
3.8 Sucking Up The Diluted Sample Already
26
6 Way Valve Position A Mode: 6 Way Valve Position B Mode:
Use 5mL Plunger Pump to aspirate Use High Pressure Pump to dispense
750uL Diluted Sample to 6 Way Valve. 5 uL Diluted Sample into
Chromatography Column for test.
27
Supplement: Two kinds of functions of Hemolytic
Agent(Reagent L); one for cleaning diluting cup and
diluting sample with the same flow direction of liquid
route, the other one for cleaning the sample probe and
High Pressure Pump.
28
Chapter 4 Hardware Structure&Configuration
4.1 PCB Configuration Figure
Interface Board
Pin Board
4.2.1 Overview
Main control board is consist of both analog and digital, analog part mainly
responsible for the sensor signal extraction, amplification, conditioning, AD sampling,
and other functions, and the digital part mainly provides motor, pump, valve drive
signal, the whole machine external interface, platform based on embedded software ,
2 of 80 pin sockets used to connect to the core board.
4.2.2 Function
29
4.2.4 Sockets List
S PIN Function Description
N Numb
er
CN1 80 Used to connect Core Board CN1;
CN2 80 Used to connect Core Board CN2;
J105 3 Powering the printer;1:+5VDC;2:Hollow;3:GND;
J3 2 Used to upgrade the core board system program;The jumper cap can be used
after short connection;
SD1 13 MicroSD card rabbet;
J102 10 2 way USB signal sockets;
J106 2 Powering 6 way valve;1:+24VDC;2:GND;
J107 18 Ethernet interface socket;
J108 40 Be used to connect TFT;Note: 10-i02d000-00007 main control board PCBA
is 30pin;
J109 5 Be used to connect screen;
J111 32 Optocoupler Signal;
J112 32 Control signal of Optocoupler and 6 way valve
J113 32 Analog signal of Pressure、Temp-sensor、Double color LED light、A/D;
J114 32 Control Signal of Valve,Pump,Indicator,Printer,fan;
J115 32 Various Motor;
X,Z-axis-Motor,5mL-Plunger-Pump,100uL-Plunger-Pump-Motor,Auto-sampl
er;
J116 2 Peltier;
J117 4 Powering Main Control;1、2:GND;3、4:+24VDC;
6 2 way USB_DN1;
8 2 way USB_DP1;
7 2 way USB Power Supply Pin:+5VDC;
30
4 2 way USB Supply Ground Pin;
J111 插座定义:
1 X-motor Origin-optocoupler signal output;
2 X-motor Origin-optocoupler Power Supply
Pin :1V~2V;
3、4、7、8、11、12、 GND;
15、16、19、20、23、
24、27、28
5 The high-pressure pump blocks the optical
coupler signal output pin.;
6 The high-pressure pump blocks the optical
coupler Power Supply pin.:+5V;
9 5ML Plunger Pump Reset Optocoupler signal
output pin;
10 5ML Plunger Pump Reset Optocoupler Power
supply pin:1V~2V;
13 100uL Plunger Pump Reset Optocoupler
signal output pin;
14 100uL Plunger Pump Reset Optocoupler
Power supply pin:1V~2V;
17 Z-motor Origin-optocoupler signal output
pin;
18 Z-motor Origin-optocoupler Power Supply
Pin:1V~2V;
21 Tube-detection-optocoupler signal output
pin;
22 Tube-detection-optocoupler Power Supply
Pin +5V;
25 Storehouse door switchdetection-optocoupler
signal output pin;
26 Storehouse-door-switch-detection-optocouple
r Power Supply Pin+5V;
29 Auto-sample Front-optocoupler signal output
pin ;
30 Auto-sample Front-optocoupler Power
Supply Pin+5V;
31 The loudspeaker signal;
32 The loudspeaker signal;
31
1 Auto-sampler post-optocoupler signal output
pin;
2、3、7、11、15、19、 GND;
23、27、31、24
4 Auto-sampler post-optocoupler Power Supply
Pin:+5V;
5、6、8、10、12、14、 Empty pin;
16、18、20、17
9 Motion control module PUL-;
13 Motion control modulePUL+(+5V);
21 Control Signal Input Pin of 6 Way Valve;
25 Six - way valve A position feedback pin;
29 Six - way valve A position feedback pin;
28 Six-way valve COMM signal pin;
26 Float sensor signal pin.;
30 X axis code teeth optocoupler power supply
pin:1V~2V;
32 X axis code teeth optocoupler signal Input
Pin;
32
27 Chromatographic column temperature sensor contact 1
29 Chromatographic column temperature sensor contact 2
24 Ambient temperature sensor contact 1
28 Ambient temperature sensor contact 2
31 Communication I2C_ data with 500nm photoelectric
conversion board
32 Communication I2C_ clock with 500nm photoelectric
conversion board.
2、4、6、8、20、22、 Empty pin
26、30、7、15
33
2 Auto-sampler Motor A1
4 Auto-sampler Motor A2
6 Auto-sampler Motor B1
8 Auto-sampler Motor B2
12 X-axis Motor A1
14 X-axis Motor A2
16 X-axis Motor B1
18 X-axis Motor B2
9、10、20 GND
1、3、5、7、22、24、 Empty pin
27、29、31
34
4.3.1 Function
4.4.1 Overview
35
The main function of LED lamp board is to carry dual wavelength LED light.
4.4.2 Function
Providing a light source of 415nm+500nm.
4.4.3 Diagram
4.4.4 Socket
J9 LED power supply 1:negative;2:positive;
4.5.1 Overview
The main function of the photoelectric switchboard is to convert the optical signal into
electrical signal, and the data is sent to the main control board through the IIC bus
after the ADC collection is amplified and adjusted.
4.5.2 Function
(1) Photoelectric conversion
(2) Magnification setting
(3) The ADC conversion
36
4.5.3 Real Picture
4.6.1 Function
The high pressure pump is a double piston tandem high pressure pump which is
necessary for conveying analysis.Have been working on the analysis of the process
usually, based on the cycle of the analysis , through the solenoid valve
switch ,transmitting three kinds of eluent of different salt concentration, gradient
(concentration gradient), the components of hemoglobin will be separated by making
use of analyzer.
4.6.2 Real Picture
4.6.3 Adjustment
Confirmed in high-pressure pump (PPS) in 1024, under the premise of the 8 MPa
tooling PEEK tubes, to connect a small tube in the side of column-liquid-out. Then in
the maintenance of interface, open the SV1 and the high-pressure pump, weighing
liquid within 5 minutes and 3 times in a row, remarked to X1, X2, X3, calculate the
mean X = (X1 + X2 + X3) / 3;Then, Y = 7.5*1024/X, and finally the high pressure
37
pump (PPS) is set as Y.
4.7 Colorimetric cuvette assembly( Test Module)
4.7.1 Function
The changes in the absorbance of hemoglobin in the samples after elution separation
of the chromatographic column will be measured.The light source adopts blue LED ,
the detector adopts 415nm/500nm dual wavelength, which improves the
anti-interference ability of the system.
4.7.2 Picture
4.7.3 Adjustment
Connecting the 8 MPa tooling PEEK tube ,coming into the senior maintenance - >
pump valve test, open the SV1, start the high pressure pump, under the premise of
stable pressure , switch to the Interface of Other Settings, set the value of DAC - LED
to 4000, by adjusting the gain of 415 (on a scale of 0-255 range adjustment), make
ADC415 values between 12000-13000, by adjusting the gain of 500 (on a scale of
0-255 range adjustment), make ADC500 values between 4000-5000;
Note: Under the premise of no bubble in the colorimetric cuvette ,to adjust the
gain .The larger the gain value, the smaller the ADC value. Conversely, the smaller
the gain, the larger the ADC value.
4.8 Two dimensional arm assembly
4.8.1 Function
The two - dimensional arm component mainly realizes the pretreatment of the sample,
including the procedure of puncture sampling and sample dilution.The 2-d arm can
automatically identify the sample type placed on the sample rack, and then the whole
blood sample is automatically diluted and then sent to the sample valve. For the
dilution sample, the sample is sent directly into the sample valve.
4.8.2 Picture
38
4.8.3 Adjustment
39
bottom of the EP tube 1 mm, by pulling EP up (a little well, not completely resisted by
the needle), to judge the position has been adjusted in the right place, click “save” and
confirm once again. If it is confirmed well, please be sure to take dilution pipe rack
and EP!
(7) Whole Blood
Under the premise of taking out the dilution pipe rack and EP in the 10 # position ,
click “next”, by + +, +, --, - button to adjust Y go to the travel limit with 0.5 mm left,
with a 0.5 mm feeler to confirm, click “save” and confirm again.
4.9 Constant Temperature Module
4.9.1 Function
In order not to make room-temperature changing affect the analysis, usually, the
column should be kept at a certain temperature with the oven.
4.9.2 Picture
4.10.1 Function
This device could eliminate the bubbles existed in the elution liquid that’s transmitted
by pump. In order to maintain the vacuum pressure of the degassing vessel, the
vacuum pump will work intermittently.
4.10.2 Picture
40
4.10.3 Adjustment
Negative pressure will not affect the degassing effect of vacuum, and the engine won't
be at fault, negative pressure is set to 65%, less than 65% negative pressure pump
work on 5 seconds, if more than 65% negative pressure pump does not work.
4.11 Sampling Frame Assembly
4.11.1 Function
The sampling frame assembly is used to transfer and place the test tube rack ,to
realize automatic loading of the sample.
4.11.2 Picture
4.11.3 Adjustment
The parameter of returning sample defaults to 1500, in the upper right corner of the
screen, click .Then test tube rack would retreat from the sample storehouse. On
41
the premise of Without skidding, test tube rack can completely back, until breaking
away from origin opto-coupler of the test tube rack. If the test tube rack cannot be
separated from the origin opto-coupler, it will need to be adjusted about the tightness
of the installation of the sample motor.
Safety Precautions
Electronic Precautions
Environment Protection
Biological Warning
verification step after the repair steps, the maintenance engineer must
42
Cross screwdriver
Open end wrench
Hexagonal screwdriver in no.5.
5.2.3 Removal
(1) Power off and unplug the main power cord;
(2) Open the left door;
(3) Open the top cover.
(4) No.5 internal hexagonal screw M4*20 (4 pieces) of the main pump head and the
auxiliary pump head are unscrewed by the hexagonal screwdriver, as shown in the
figure below:
(5) Remove the fixed base cleaning block from the straight line bearing seat, as shown
in the figure below:
43
(7) Remove the diaphragm (note: the direction of diaphragm )
Installation
The installation step is the reverse operation of the disassembly step.
Verification
a.Check the assembly of all parts and put them in place;
b. Check whether the connection of gas and liquid pipe is correct;
c. Start-up verification.
5.3.3 Removal
(1) Open the left door;
(2) Unscrew the grip joint on the filter holder of the pump side head.
(3) Unscrew the filter installation seat with the open spanner, as shown below.
44
(4) Remove the filter mounting block.
Prefilter
(5) Use a paper clip or an art cutter to take out the pre-filter.
Installation
The installation step is the reverse operation of the disassembly step.
Verification
a.Check the assembly of all parts and put them in place;
b. Check whether the connection of gas and liquid pipe is correct;
c. Start-up verification.
45
5.4.1 Purpose
The malfunction of two - dimensional arm in the working process , it’s necessary to
be replaced when maintenance.
5.4.3 Removal
(1) Remove cover and right door.
(2)Unscrew the 2 countersunk head screws of the fixed towing chain on the right
vertical board, as shown in the figure below.
(3)Unscrew 6 composite screws from the right panel, and remove the 2-d arm
components to tidy up the pipeline, as shown in the picture below:
Installation
The installation step is the reverse operation of the disassembly step.
Verification
a.Check the assembly of all parts and put them in place;
b. Check whether the connection of gas and liquid pipe is correct;
c. Start-up verification.
46
5.5 The Replacement of Puncture Probe
5.5.1 Purpose
The malfunction of puncture probe in the working process, it needs to be replaced
when maintenance.
(3) Cut off the ribbon, take out the needle from the silicone tube.
Installation
The installation step is the reverse operation of the disassembly step.
Verification
a.Check the assembly of all parts and put them in place;
47
b. Check whether the connection of gas and liquid pipe is correct;
c. Start-up verification.
(3) Pull out the locking plate and take down the swab assembly.
48
Reassembly And Installation
The installation step is the reverse operation of the disassembly step.
Verification
a.Check the assembly of all parts and put them in place;
b. Check whether the connection of gas and liquid pipe is correct;
c. Start-up verification.
49
b. Power off, to remove the cover of the instrument, and to short out the J3 socket of
the main control board, and insert the micro SD card into the card slot of the board
card SD1. As shown in the picture below.
c.starting up, waiting for about the 80S, the yellow light of the core board starts
flashing, indicating that the download is successful.
50
d. waiting for 5seconds , turn off power. Operation completed.
6.3.2 Method and steps of upgrading QT
a. Make sure that the main control board J3 is disconnected (make sure the SD card is
not inserted in the SD1 slot).As shown.
Note: it is suggested to copy the QT program (G-08-I02D000-00005 I02D application
software download program) and the lower computer (G-08-I02D000-00004
I02DMCU motion control unit download program) to the U disk at the same time.
b. Insert the USB port with the QT program (g-08-i02d000-00005 I02D application
software download program) into the USB port of the switch board, and confirm that
the instrument is turned off.
Copy the update. Sh /I02D_QT/ script.tar.gz to the H8 directory in the U disk, as
shown in the figure below.
51
c. Insert the USB flash disk into the USB port on the back of the instrument, and then
turn it on, and the screen will appear on the screen.
d. click on the screen to show the interface shown below (the image is dark and the
word "successful" is displayed), which means that it is finished and the USB flash
drive is taken out.
e.Restart the device.If the above upgrade is successful, the application (QT) interface
52
appears, as shown in the figure below.
b.Click "system" -> "advanced maintenance", enter the account and password to enter
the advanced maintenance interface.
Account:lifo
Password:661166
53
C. Click "other Settings" on the left sidebar.
d.Click the "lower computer" in the software upgrade to update the data and
disappear after the update. (the update time is about 1min, and the LED lights in the
MCU chip will be flashing as normal after the update.)
54
e.Restart the device to see if the "information" interface is corresponding to the new
version of software.
55
List 7- 1Error Codes and Error Messages
56
pump motor 0X0404 The target position of 100 μ l plunger pump motor
error exceeds the maximum limit
0X0405 100μl plunger pump motor reset sensor error
0X0406 The timeout of 5ml plunger pump motor waiting
movement
0X0407 100 μ l plunger pump motor controller is not
initialized
0X0502 Attempt to operate the running Z-axis motor
0X0503 Z-axis motor code plate lost step
X-axis motor 0X0504 The target position of Z-axis motor exceeds the
error maximum limit
0X0505 Z-axis motor reset sensor error
0X0506 The timeout of Z-axis motor waiting movement
0X0507 Z-axis motor controller is not initialized
0X0602 Attempt to operate the running Z-axis motor
0X0603 Z-axis motor code plate lost step
0X0604 The target position of Z-axis motor exceeds the
Z-axis motor maximum limit
error 0X0605 Z-axis motor reset sensor error
0X0606 The timeout of Z-axis motor waiting movement
0X0607 Z-axis motor controller is not initialized
57
7.3 Analyzing&Processing of Malfunction
ADC value 16383 1. The channel gain 1. Increase the gain value.
value is too small. 2. Reduce the LED value.
2.LED value setting is 3. Replace the data acquisition board.
too high. 4. Performing to prime A liquid.
3. Failure of When necessary, the colorimetric
acquisition board. pool was taken out in the process of
4.There are bubbles in priming, and A few strokes were used
the color pool. to help the bubbles to escape rapidly.
ADC415unstable Having bubbles in the Open SV1, open the high pressure
58
between12000 and colorimetric cuvette pump and observe whether the ADC
13000 value is stable.When necessary, take
out the colorimetric cuvette in the
process,Use fingers flip the liquid
tube lightly to help the bubbles
escape quickly.
List 7- 3 Troubleshooting
59
About Reagents &Testing
List 7- 4Troubleshooting
Existing bubbles in the line A or B or C eluent is used Check the connection tube
connected reagents or the connection tube is and reagent dosage, then
not inserted below the perform to prime A\B\C
liquid. solution to remove the
bubbles.
Abnormal Graphs 1. Liquid road blocked. 1. Clean the system after
2. Column column or replacing the filter.
filter failure. 2. Replace the column or
3. Failure of reagent. filter.
4. Sample analysis was 3. Reagents.
conducted when the 4. After the temperature of
temperature of the the column is stable,
column was not stable. sample analysis is
Bubbles occur. conducted.
5.Bubble removal
60
18% The whole blood sample is size.
not enough. 2. Replace the whole
3. Column column or filter blood sample or adopt the
failure. diluted blood model.
4. Failure of reagent. 3. Replace the column or
The system didn't build up filter.
the pressure. 4. Reagent.
5.Bubble removal
Error Code e1 1. The sample size is too little 1.Mix it well and repeat
or not absorbed. the test. If the retest value
2.No samples were placed in is still abnormal, it is
the sample. recommended to test with
the method of diluting
blood.
2.Reposition the sample
and retest it.
61