4263B-06

S E R V I C E N O T E

SUPERSEDES: NONE

AGILENT 4263B LCR Meter

Serial Numbers:MY40100101 / MY40100233

Measurement accuracy may be out of specification

To Be Performed By: Agilent-Qualified Personnel

Parts Required: Not used

Situation:
Units given serial number above may not meet its measurement accuracy specification due
to mis-calibration at factory. All units in the range given above should be re-calibrated at
service center immediately.
Solution / Action:
Perform all adjustments. And do performance test and functional check to verify instrument
operation.

DATE: January 2001

ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:

MODIFICATION RECOMMENDED
ACTION IMMEDIATELY STANDARDS:
CATEGORY: ON SPECIFIED FAILURE LABOR 2.0 Hours
AGREEABLE TIME
LOCATION CUSTOMER INSTALLABLE SERVICE RETURN USED RETURN
CATEGORY: ON-SITE INVENTORY: SCRAP PARTS: SCRAP
SERVICE CENTER SEE TEXT SEE TEXT
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: January 2002

AUTHOR: HAM ENTITY: 3355 ADDITIONAL INFORMATION: Using O2G for the Repair Class
and enter the Service Note # in the Repair Details SIS-Entry

© 2000 AGILENT TECHNOLOGIES


PRINTED IN U.S.A.

1 of 1