Manual Ingles Elster A1800
Manual Ingles Elster A1800
Manual Ingles Elster A1800
Technical manual
TM42-2410G
CONTENTS
Notes:
Disclaimer of Warranties and Limitation of Liability vii
Safety Information
Installation, operation, and maintenance of this product can present
potentially hazardous conditions (for example, high voltages) if safety
procedures are not followed. To ensure that this product is used safely, it is
important that you:
Review, understand, and observe all safety notices and recommendations
within this manual.
Do not remove or copy individual pages from this manual, as this manual is
intended for use in its entirety. If you were to remove or copy individual
pages, cross references and safety notices may be overlooked, possibly
resulting in damage to the equipment, personal injury, or even death.
Inform personnel involved in the installation, operation, and maintenance
of the product about the safety notices and recommendations contained in
this manual.
Within this manual, safety notices appear preceding the text or step to
which they apply. Safety notices are divided into the following four
classifications:
+ Q
- +
- Q
L1 L 2 L 3
COM 0 2
T1 T2 T3 T4 T5 T6 T7 T8 EOI LC TC TST
5,000 imp/kWh
TYPE A1800 5,000 imp/kVarh
MODEL 1830RALN
3 x 58/100...277/480V 60 Hz
0.2S
1(10)A
2006 Made in USA
ELSTER SAMPLE CT A
SERIAL # 01234567
VT V
S LE
ZE3JF R00LM-01 *01234567*
R .2-062 0 02AG-000000
imp/kWh(kVARh)
Introduction 2
Standards Compliance
IEC
The A1800 ALPHA meter meets or exceeds the following IEC standards for
electricity metering.
Table 1-1. IEC standards supported by the A1800 ALPHA meter
IEEE/ANSI
The A1800 ALPHA meter meets or exceeds the following IEEE/ANSI
standards for electricity metering, and it is intended for use by commercial
and industrial utility customers.
Table 1-2. IEEE/ANSI standards supported by the A1800 ALPHA meter
DLMS
The A1800 ALPHA meter meets or exceeds the following IEC standards for
DLMS protocol support (when the meter is enabled for DLMS protocol
support). See “Communication protocols” in chapter 2 for more
information.
Table 1-3. IEC standards supported by the A1800 ALPHA meter for DLMS protocols
DIN
The A1800 ALPHA meter meets or exceeds the following DIN standards for
electricity metering.
Table 1-4. DIN standards supported by the A1800 ALPHA meter
Benefits
Reliability
The A1800 ALPHA meter, part of the ALPHA line of meters, uses the
patented ALPHA meter technology for measurement and accurate
calculation of energy quantities. With over 3 million ALPHA polyphase
meters in operation throughout the world, the A1800 ALPHA continues the
tradition of reliable electronic meters.
The power supply in the meter operates from any available phase. A three-
phase, four-wire A1800 ALPHA meter maintains operation if the neutral
line and any one or two of the line voltages become disconnected. The
meter can also operate using the auxiliary power supply, which can power
the meter from an independent power source in the situation where main
power is unavailable.
The A1800 ALPHA meter can use its internal crystal oscillator or the power
line frequency to maintain time and date functions. The crystal oscillator
can be used when the power line frequency is known to be too unstable for
accurate timekeeping.
The A1800 ALPHA meter has been designed to function to provide long
battery life. Because of the low current drain, the service life of the lithium
battery can exceed the life of the meter.
The A1800 ALPHA meter uses nonvolatile memory to store billing and
other critical data. The data is preserved even if the power fails.
Maintainability
The A1800 ALPHA meter is easy to maintain. Meter register functions and
communication interfaces are fully integrated on a single, surface-mount
technology circuit board.
The meter firmware resides in flash memory, allowing the firmware to be
upgraded in the field.
Adaptability
The A1800 ALPHA meter allows configuration for custom TOU rates
(tariffs), offering a broad range of demand and TOU operations.
Most common services and mounting configurations are supported, and
functional upgrades are easily performed as new situations arise. The wide
operating range allows installation at any of the common meter voltages.
Additionally, the factory-configurable optical port accommodates ANSI and
IEC standards. The 16-segment character liquid crystal display (LCD)
improves readability and provides flexibility for displaying meter
information. As an added feature, the main meter circuit board provides
selectable, independent, serial remote interfaces for RS-232 or RS-485
communication.
Introduction 5
Economy
The A1800 ALPHA meter saves both time and money. It can increase
personnel productivity because of the following features:
no user calibration required (factory calibrated)
reduced testing times
fewer styles to learn and maintain
dual serial communications interfaces on the main meter circuit board
automated data retrieval
system service verification
on-site instrumentation displays
tamper restraint and quality monitoring (TRueQ™) tests
event logging
Security
The A1800 ALPHA meter is tamper-resistant. Passwords may be specified
that prevent unauthorized access to meter data. The standard TRueQ
feature or the optional instrumentation profiling (or both) can be used to
detect possible tampering of energy measurements.
All A1800 ALPHA meters provide auditing capabilities that can be used to
indicate potential meter tampering like terminal cover open detection and
per phase outage recording.
The A1800 ALPHA meter can be ordered with a partially-transparent
terminal cover, making it easier to see obvious tampering.
Accuracy
The A1800 ALPHA meter meets or exceeds requirements of IEC standards.
Configuration IEC 62053-22 IEC 62053-21 IEC 62053-231
Class 0.2 S Class 0.5 S Class 1.0 Class 2.0 Class 2.0 Class 3.0
direct connect*
transformer-rated
*Actual reactive energy accuracy is substantially better than required by the standard.
The meter precisely measures demand and energy across a wide range of
voltage and current despite variations in temperature and power factor.
The low current sensor burden may also improve the accuracy of external
current transformers when measuring light loads.
Introduction 6
Meter types
Different meters within the A1800 ALPHA meter family have specific
capabilities (see Table 1-5 and Figure 1-2).
Note: Throughout this manual, the term A1800 ALPHA is used to
describe any meter in the meter family. When necessary, a specific
meter designation (for example, A1830 meter) will be used to indicate
that the description applies to only one meter in the meter family.
Table 1-5. Meter designations of the A1800 ALPHA meter family
1 Same wide temperature range rating as the meter (-40 °C to +85 °C inside meter cover; -40 °C to +60 °C outside)
Introduction 7
Interchange
metering
A1880
ily
fam
Large C & I
r
ete
A1860
Am
A1830
PH
Mid C & I
AL
00
A1810
8
A1
Light C & I
A1805
Residential
Introduction 8
Meter series
The meter series is used to indicate the options and capabilities available
for the A1800 ALPHA meter.
Series 110 and later
Optional DLMS/COSEM protocol support available
Series 150
Optional Read without Power feature available
Series 200
Optional virtual option board
The meter can add and activate protocols while the meter is in the
field. See “Virtual option board” in chapter 2 for more information.
Faster instrumentation value refresh times
For certain instrumentation quantities, the meter now supports
refresh times of about ½ second. See “System instrumentation”
on page 4-1 for more information.
New TRueQ test sets
The meter now supports 3 separate test sets: the standard test
set, the EN 50160 test set, and the IEEE test set. See “TRueQ
tests” on page 4-14 for more information.
Interposing relay control
Configurable demand reset times
The meter now supports a specified time and date that a reset
occurs. See the Metercat documentation or online help for
information on configuring the demand reset time.
Optional support for disabling the Metercat Register Edit task
The meter can be factory configured to prevent Metercat from
editing the meter register. See the Metercat documentation or
online help for information on disabling Metercat’s Register Edit
task.
Support for European Summer Time as well as Daylight Saving
Time
The meter can now adjust for seasonal time changes based on
local custom. When programming the meter for the seasonal time
change, Metercat obtains the configuration directly from Microsoft
Windows, and the meter will change the time based on the
programming.
Introduction 9
Metercat software
Beginning with Metercat release 2.3, the following features are supported:
Enhanced TRueQ test set
Up to 8 metered quantities
Independent TOU switch times for energy and demand
Beginning with Metercat release 3.0, Metercat supports the DLMS
protocol. And beginning with Metercat release 3.2, the following features
are supported
Interposing relay functionality support
Virtual option board (DLT-645 version 1.1) support (requires Metercat
release 3.2.2.1)
Alpha Keys
Alpha Keys™ software allows A1800 ALPHA meters to be upgraded so they
provide additional functionality. Upgrading with Alpha Keys software means
that the meter does not have to be returned to the factory and new meters
do not have to be purchased to gain functionality.
The following features can be added to the meter by using Alpha Keys:
load profiling
instrumentation profiling
transformer and line loss compensation 2
4-quadrant metering
DLMS/COSEM protocol support2
Notes:
2 PRODUCT DESCRIPTION
Physical description
The A1800 ALPHA meter is designed for indoor mounting. The cover
assembly of the A1800 ALPHA meter exceeds the environmental
requirements of IEC 62053-11. The case of the A1800 ALPHA meter
provides an IP54 degree of protection for the meter.
The physical components of the A1800 ALPHA meter consist of the
following:
terminal cover
long terminal cover (see Figure 2-1)
short terminal cover (see Figure 2-2)
medium terminal cover
partially-transparent long terminal cover
meter cover assembly
inner cover assembly
base electronic assembly
Figure 2-1. Front view of the A1800 ALPHA meter (long cover)
RESET button
(sealable)
* (ALT) button
Optical port
(IEC configuration)
Utility information
card area
Terminal cover
screws (sealable)
The terminal cover and meter cover assembly are manufactured using a
UV-protected polycarbonate plastic. The terminal cover is available in
either the long version or the short version. The meter cover assembly has
a clear plastic window that allows the meter LCD and nameplates to be
viewed.
Figure 2-2. Front view of A1800 ALPHA meter with short terminal cover (transformer rated)
The A1800 ALPHA meter can be sealed using any or all of the following
methods:
Seal location Purpose
Meter cover screws Prevents access to the meter except for the main
(certification) connections, relay connections, communication
interface connections, and nameplate. Also can
prevent reprogramming and recalibration of the
meter.
Terminal cover Prevents non-utility access to the main
screws (utility) connections, relay connections, and utility
information card
RESET push button Prevents unauthorized manual demand resets
The four cover screws can be individually sealed (Figure 2-1). The two
terminal cover screws limit access to the main terminals and auxiliary
wiring connections only. Therefore, only the terminal cover seals must be
broken to access these connections. The two meter cover screws are
located on the lower front of the meter under the terminal cover. Sealing
these screws seals the main enclosure and limits access to the metering
circuit board and sensing elements.
For maximum protection of the metering components, seal all four screw
seals.
Product description 13
Figure 2-3. A1800 ALPHA meter with cover removed (transformer rated)
Meter cover
screw (sealable)
Meter cover
Terminal cover screw (sealable)
screw (sealable)
TOU battery
Battery connector
Terminal cover removal
detector switch
Terminal connectors
Terminal cover
Auxiliary power supply screw (sealable)
Optical port
The A1800 ALPHA meter provides an optical port that can be ordered with
either an ANSI-compliant or IEC-compliant interface (see Figure 2-4). To
use Elster meter support software to read or program the meter through
the optical port, an optical probe is required. This probe connects from the
serial port of the computer to the optical port on the meter.
Figure 2-4. Comparison of ANSI-compliant and IEC-compliant optical port interface
Elster recommends use of the UNICOM III probe to reliably read the A1800
ALPHA meter. For information on ordering the UNICOM III probe, visit
www.elsterelectricity.com or contact your local Elster representative.
LCD
The A1800 ALPHA meter is equipped with a 16-segment character liquid
crystal display. See “Indicators and controls” in chapter 3 for details.
Nameplate
Elster installs the nameplate at the factory. See Appendix C, “Nameplate
and style number information,” for details on the nameplate.
Communications
The A1830 ALPHA meter provides remote communications interfaces on
the main meter circuit board for RS-232 or RS-485 serial communication.
Physical outputs exist for both RS-232 and RS-485 interfaces; however,
only one can be used at any given time. No configuration is necessary to
switch between an RS-232 and RS-485 selection. Additionally, the A1860
ALPHA meter provides a second, independent serial communication port
that supports either RS-232 (see Figure 2-6) or RS-485 (see Figure 2-7).
See Chapter 5, “Outputs,” for more information on the RS-232 or RS-485
ports.
Figure 2-6. A1860 ALPHA meter with RS-232 as second communication port
RS-232 connector
(optional)*
RS-485 terminals
RS-232 connector
Figure 2-7. A1800 ALPHA meter with RS-485 as second communication port
RS-485 connector
(optional)*
RS-485 terminals
RS-232
connector
Battery
The terminal block has a battery well and connector for the optional TOU
battery.
Terminal configurations
The A1800 ALPHA meter supports the following terminal configurations:
10 A transformer-rated (sequential)
10 A transformer-rated (symmetrical)
120 A direct connect-rated (sequential)
120 A direct connect-rated (symmetrical)
Communication protocols
The A1800 ALPHA meter supports the ANSI C12 standards for both local
and remote communication.
Additionally, the A1800 ALPHA meter can also support the DLMS
standards for both local and remote communication. When equipped with
the optional DLMS communication support, the meter supports both ANSI
C12 and DLMS communication. When establishing a communication
session, the meter automatically determines the protocol being used and
responds in the same protocol.
For series 110 meters and later, DLMS support can be enabled by
purchasing the appropriate Alpha Key and upgrading the A1800 ALPHA
meter with the Alpha Keys support software. 1 See “Alpha Keys” in chapter
1 for more information.
For a list of the DLMS OBIS identifiers supported by the A1800 ALPHA
meter, see the A1800 ALPHA meter communication options product guide
(PG42-1012).
Port addressing
The ANSI C12 protocol does not have a specific address but the RS-485
ports do need to be addressed. Additionally, DLMS has an address in
addition to the RS-485 port addresses. When supporting DLT645, the
DLT645 protocol also has a specific address in addition to the use of the
RS-485 addresses.
The device IDs are used by the RS-485 hardware interface scheme for
addressing each meter. By giving each meter a different address based on
the last two digits of the serial numbers, each meter can be configured in a
manner that allows the meter to be installed directly in the field and
minimizing the possibility for duplicate addresses within the same RS-485
loop. The only exception to this is the main meter of any loop that needs to
be reprogrammed to have an address of 1.
Table 2-1. Port addressing scheme
System architecture
The A1800 ALPHA meter main circuit board contains all the electronics
that make up the meter registers and communication interfaces. See for
the meter circuit board block diagram. The circuit board as shown in
contains the following:
meter engine
microcontroller
EEPROM
resistive dividers for the 3 phase voltages
load resistors for the 3 current sensors
power supply
high frequency crystal oscillator
32 kHz low power timekeeping crystal oscillator
optical port components
liquid crystal display (LCD) interface
RS-232 and RS-485 communication interfaces
option board interface
pulse outputs
Precision
reference
LCD Low
power
crystal
Resistive
divider
Power fail
Resistive
2X line freq
divider
A
Resistive B
divider
Meter engine C
Current Clock
Phase C current
sensor
Power supply
Meter engine
Multiplication and other calculations are performed using a custom
integrated circuit, called the meter engine. The meter engine contains the
digital signal processor (DSP) with built-in analog-to-digital (A/D) converters
capable of sampling each current and voltage input. The A/D converters
measure the voltage and current inputs for a given phase. The DSP
multiplies the signals appropriately, using the factory-programmed
calibration constants.
Product description 21
Microcontroller
The microcontroller performs many different functions, for example:
communicates with the DSP and EEPROM
provides for serial communication over the optical port
provides for serial communication over the remote ports
generates optical output pulses
controls the LCD
controls any option boards
The microcontroller and the meter engine communicate with each other
constantly to process voltage and current inputs. When the microcontroller
detects a power failure, it initiates the shutdown and stores billing and
status information in EEPROM.
EEPROM
The A1800 ALPHA meter uses electrically erasable programmable read
only memory (EEPROM) for nonvolatile storage of manufacturing data,
meter configuration data, and energy measurement values. The A1800
ALPHA meter is provided with either 128 KB or 256 KB of main board
memory. See “Style number information” in appendix C for information
regarding how to identify the amount of main board memory on your meter.
The EEPROM provides storage of all information needed to ensure the
integrity of the demand or energy calculations, including the following:
configuration data
billing data
all TOU data
log and profiling data
meter status
constants
energy usage
maximum demand
cumulative demand
Product description 22
Billing data
Metered energy and demand quantities
All A1800 ALPHA meters are capable of measuring delivered and received
kWh energy and kW demand. The A1800 ALPHA meters can also measure
reactive and apparent energy and demand. The meter engine samples the
voltage and current inputs and sends these measurements to the
microcontroller. In the meter engine, each pulse is equal to one Ke defined
as one of the following:
secondary rated Wh per pulse
secondary rated varh per pulse
secondary rated VAh per pulse
The following list shows the available metered quantities for the A1800
ALPHA meter. Basic metered quantities (indicated by * in the following list)
can be selected as a source for relay outputs. The remaining metered
quantities are calculated from 2 or more basic metered quantities.
Metered energy and demand quantities
kVAh delivered (Q1 + Q4)
kVAh Q1
kVAh Q2
kVAh Q3
kVAh Q4
kVAh received (Q2 + Q3)
kVAh sum (delivered + received)
kvarh (Q1 - Q4)
kvarh (Q1 + Q4)*
kvarh (Q2 - Q3)
kvarh (Q2 + Q3)*
kvarh (Q3 - Q2)
kvarh delivered (Q1 + Q2)*
kvarh net
kvarh Q1*
kvarh Q2*
kvarh Q3*
kvarh Q4*
kvarh received (Q3 + Q4)*
kvarh sum (delivered + received)*
kWh delivered*
kWh net
kWh received*
kWh sum*
Product description 23
The meter can store up to two average power calculations, which can be
configured in Elster’s meter support software. Average power factor is
calculated every second. Upon a demand reset, the values used in this
calculation are set to zero and the AvgPF will be set to 1.000.
Demand calculations
Demand is the average value of power over a specified time interval. The
A1800 ALPHA meter supports three different methods for demand
calculation:
rolling interval
block interval
thermal time constant
An interval is the time over which demand is calculated. The length of a
demand interval is programmable using Elster meter support software, but
the value must be evenly divisible into 60 minutes. Common demand
interval lengths are 15 or 30 minutes.
Rolling interval
Rolling demand interval is defined by two parameters:
the demand interval length - specified in minutes and may be any
value that is evenly divisible into 60
subinterval length - also specified in minutes and may be any value
that is evenly divisible into the interval length
Both of these values are configurable by Elster meter support software.
The demand is calculated at the end of each subinterval, resulting in
overlapping demand intervals (or a “rolling” demand).
For example, the A1800 ALPHA meter can be configured for a 15-minute
demand interval length and a 5-minute subinterval length. In this case, the
demand is calculated every 5 minutes based on the 3 previous
subintervals (see Figure 2-10).
Product description 24
15-minute interval
15-minute interval
15-minute interval
0 5 10 15 20 25
Time (minutes)
Block interval
Block demand interval is a special case of rolling interval demand in which
the subinterval is the same size as the interval (see Figure 2-11).
Figure 2-11. Block demand intervals
0 15 30 45 60
Time (minutes)
Maximum demand
Maximum demand (also referred to as indicating demand or peak
demand) is the highest demand value that occurs in a billing period. The
demand for each demand interval is calculated and compared to an earlier
maximum demand value. If the new interval demand exceeds the previous
maximum demand, then the new demand is stored as the maximum
demand (see Figure 2-12). When a demand reset occurs, the maximum
demand is reset to zero. The demand for the first full interval after a
demand reset becomes the maximum demand.
Product description 25
Interval 7
demand Interval 8
Interval 6 (9.9 kW) demand
demand (9.5 kW)
(9.2 kW)
Demand forgiveness
Demand forgiveness is the time during which demand is not calculated or
stored after a qualified power outage. Demand forgiveness has two
programmable settings:
outage time: the number of minutes a power outage must last to
qualify for demand forgiveness (0 to 15 minutes)
time: the number of minutes that demand is not calculated or stored
(0 to 255 minutes) following a qualified power outage; zero disables
demand forgiveness
TOU data
All A1800 ALPHA meters store the total (single-rate) data for energy and
demand. TOU meters can store the total data and the data for up to 4
rates. TOU rates can be based on any combination of day (up to 4 day
types), time (up to 132 switch times), or season (up to 12 seasons). The
switch points for energy and demand may be configured independently of
each other.
All selected metered quantities are stored according to the TOU rate. The
meter stores the energy, demand, and average power factor for each rate.
Always Positive
The A1800 ALPHA meter can be ordered with the optional Always Positive
feature which records reverse energy as positive energy flow on a phase-
by-phase basis. Always Positive can be used to deter power theft or
minimize the effects of improper meter wiring. The following equation
shows how total active power is calculated using theft-resistant
measurement:
𝑃𝑡𝑜𝑡𝑎𝑙 = |𝑃1 | + |𝑃2 | + |𝑃3 |
Event
History
Billing
TrueQ
Config LP
Voltage sag
Manufacturing
Self read IP
Etc.
LP*
IP*
In most cases, the 256 KB option is sufficient to meet data logging and
profiling requirements. In some cases (for example, if extensive
instrumentation profiling is desired), more memory may be required. When
the data storage cannot be met with the 256 KB main memory option,
extended memory can be used to add shared memory to the A1800 ALPHA
meter.
Product description 29
Event log
All A1800 ALPHA meters have an event log. The A1800 ALPHA meter
stores the date and time that events occur. Elster meter support software
is used to define and program the number of event log entries that the
meter will record. Events that can be included in the event log are as
follows:
power fail start and stop (2 event log entries)
date and time change information (2 event log entries)
date and time of demand resets (1 event log entry)
date and time of event log reset (1 event log entry)
date and time of test mode activity (2 event log entries)
start and stop time when the current TOU rate is overridden by the
alternate TOU rate schedule (2 event log entries)
start and stop time of per phase outage (2 event log entries)
date and time of terminal cover removal (1 event log entry)
date and time of main cover removal (1 event log entry)
The meter will detect and log the removal of either the terminal cover or
main cover even when the meter is not powered (provided the TOU battery
is functioning).
After the maximum number of entries has been stored, the meter will begin
overwriting the oldest entries. The event log can be disabled through Elster
meter support software.
History log
All A1800 ALPHA meters have a history log that stores table information
and procedure ID for configuration-altering writes to the meter. The A1800
ALPHA meter records a sequential listing of records, along with the date
and time. The meter records this information as an audit trail, maintaining
a history of programming changes made to the meter.
After the maximum number of entries has been stored, the meter will begin
overwriting the oldest entries. The history log can be disabled through
Elster meter support software.
Self reads
All A1800 ALPHA meters can support self reads. A self read captures the
current period billing data and stores it in memory. The A1800 ALPHA
meter can store up to 35 self reads can be stored depending on memory
requirements for logs, data, etc. This data can be retrieved later for
analysis or billing. If the meter has recorded the maximum number of self
reads, the next self read will overwrite the oldest copy.
Self reads are events that can be triggered by any of the following:
scheduled calendar events
every demand reset
communication procedure
Self reads are different from previous billing data copies. The previous
billing data copy stores only one copy of billing data at a time and only
when a demand reset occurs. See “Demand reset data area” in chapter 3
for more information.
Product description 30
Load profiling
For meters with load profiling capabilities (designated with an “L” suffix),
the A1800 ALPHA meter is capable of recording 8 channels of information.
Load profiling sources
kVAh delivered (Q1 + Q4)
kVAh Q1
kVAh Q2
kVAh Q3
kVAh Q4
kVAh received (Q2 + Q3)
kVAh sum
kvarh (Q1 - Q4)
kvarh (Q1 + Q4)
kvarh (Q2 - Q3)
kvarh (Q2 + Q3)
kvarh (Q3 - Q2)
kvarh delivered (Q1 + Q2)
kvarh net
kvarh Q1
kvarh Q2
kvarh Q3
kvarh Q4
kvarh received (Q3 + Q4)
kvarh sum
kWh delivered
kWh net
kWh received
kWh sum
Load profiling has its own, separate interval length that is configured
independently from the demand interval length. The length of the load
profiling interval must adhere to the following rules:
the length must be between 1 and 60 minutes
the time must be evenly divisible into an 60 minutes
Table 2-2 show the number of days of load profiling available. These values
are estimates and may vary depending on the firmware used in the meter.
Data in Table 2-2 are based on the following settings:
load profiling at 15-minute intervals
no instrumentation profiling
the meter is programmed for 6 metered quantities, 2 average power
factors, and 4 coincident values
Product description 31
The first number shows the number of days of load profiling, assuming all
other logs and self reads record the maximum number of entries. The
second number shows the number of days of load profiling, assuming all
other logs and self reads record the minimum number of entries.
Table 2-2. Estimated days of load profiling storage per number of channels
1 2 3 4 5 6 7 8
min/max min/max min/max min/max min/max min/max min/max min/max
256 KB 594/714 317/381 242/291 178/214 151/182 124/149 110/133 95/114
1 MB 3177 1696 1294 954 812 664 592 509
Instrumentation profiling
In meters with instrumentation profiling (designated with an “N” suffix), the
meter has two sets of instrumentation profiling. Each set can record up to
16 channels from the sources listed below. Also, instrumentation profiling
can use the sources listed in “Load profiling” on page 30 for more
extensive load profiling.
Instrumentation profiling sources:
frequency
per phase current
per phase voltage
per phase watts
per phase VA
per phase voltage angle with respect to line 1 voltage
per phase fundamental (1st harmonic) current magnitude
per phase fundamental (1st harmonic) voltage magnitude
per phase 2nd harmonic current magnitude
per phase 2nd harmonic voltage magnitude
per phase voltage % total harmonic distortion (THD)
per phase current % THD
per phase harmonic current (sum of 2nd through 15th)
per phase current angle with respect to line 1 voltage
per phase vars (vectorial)
per phase 2nd harmonic voltage %
Product description 32
Item Description
Minimum The meter samples the selected quantity over the
instrumentation interval. The minimum value of all the
samples is recorded.
Maximum The meter samples the selected quantity over the
instrumentation interval. The maximum value of all the
samples is recorded.
Average The meter samples the selected quantity over the
instrumentation interval. The average value of all the
samples is recorded.
End The meter samples the selected quantity over the
instrumentation interval. The last value of all the samples is
recorded.
Each set of instrumentation profiling has its own, separate interval length
that is configured independently from the demand interval length. The
length of the instrumentation profiling interval must adhere to the following
rules:
the length must be between 1 and 60 minutes
the time must be evenly divisible into an 60 minutes
Product description 33
TRueQ Log
The A1800 ALPHA meter has a TRueQ log that records TRueQ test failures.
Elster meter support software is used to define and program the number of
TRueQ log entries that the meter will record. Elster meter support software
is also used to define which tests can record failures in the TRueQ log.
The A1800 ALPHA meter can record the following data associated with the
TRueQ test:
the date and time when the TRueQ monitor first detects a qualified
failure and the identifier of the TRueQ test (1 TRueQ log entry)
the date and time when the TRueQ monitor no longer detects a failure
and the identifier of the TRueQ test (1 TRueQ log entry)
Note: See “TRueQ event counters and timers” in chapter 4 for
information on qualification time
For each TRueQ log entry, the meter also records an instrumentation
measurement related to the TRueQ test.
When the maximum number of entries has been stored, the meter will
begin overwriting the oldest entries.
See “TRueQ monitoring” in chapter 4 for more information.
User-defined tables
User defined tables offer specific data retrieval options for A1800 ALPHA
meters. User defined table configuration may be requested at the time of
purchase, and the specific configuration may be programmed at the
factory. An AMR system can then be configured to retrieve the user defined
table information from the meter instead of individual table reads. This
reduces the total communications time.
Product description 34
89
22
204 224*
307
150 5
Approximate dimensions in millimeters
170
89
204
224*
273
89
22*
213* 224*
240
202
150
Figure 2-18. A1800 ALPHA meter, bottom view (direct connect and transformer rated)
170 170
6.2
Ø 10 5.4
Direct connect meter Transformer rated meter
Approximate dimensions in millimeters.
Product description 36
85
77
3.5
1.4
27 32
9.5
Viewing area
Approximate dimensions in millimeters
5 2
As shown in Figure 3-2, the LCD is divided into different display regions.
Figure 3-2. LCD regions
Quantity identifier
Low battery indicator
Phase Alternate mode
Error/warning indicator
indicators (3) indicator
+ Q
-P +P
Energy direction - Q
L1L2 L3 Comm. port
indicator indicator
COM 0 1 2
Display quantity
Power/energy
units identifier
The A1800 ALPHA meter can be ordered with a backlight option for the
LCD. The LCD can be illuminated by pressing one of the push buttons,
making it easier to read the LCD in no-light or low-light conditions. The
backlight option must be specified at the time of ordering. See “Using the
backlight” on page 43 for more information.
Note: LCD backlight is not available in Read without Power mode.
Operating instructions 38
Quantity identifier
This 7-digit region identifies the displayed quantity as defined and
programmed with Elster meter support software. An identifier can be
assigned to most display quantities in the display sequence. See Appendix
B, “Display table,” for more information.
Display quantity
This 8-digit display on the LCD shows either metered quantities or other
displayable information, depending upon how the A1800 ALPHA meter has
been programmed.
The displayable digits are definable using Elster meter support software for
both energy and demand readings. From 3 to 8 digits with up to 4 decimal
places can be used. These digits are also used to report error codes for the
following error conditions:
operational errors (E1, E2, or E3)
system instrumentation and service test errors (SE)
warnings (W1 or W2)
communication codes (COM 0, COM 1, COM 2)
For instrumentation values and tests, numeric values may be replaced by
or mixed with alphabetic characters to better define the value. See
Appendix B, “Display table,” for more information.
Phase indicators
Each phase indicator (L1, L2, and L3) corresponds to a line voltage (Line 1,
Line 2, and Line 3, respectively) present on the A1800 ALPHA meter
connections. The state of the indicators correspond to the following:
If the indicators are on, then all expected line voltages are present.
If an indicator is blinking, then that expected line voltage is either
missing or below the defined threshold for voltage sag detection.
If an indicator is off, the line is not expected for the configured meter
type.
See “Voltage sags” in chapter 4 for more details on momentary voltage sag
detection and the phase indicators.
Error indicator
The error indicator flashes when any error condition is present or remains
on if a warning condition is present. When the error indicator is on, the LCD
will also display the appropriate error or warning code. See “System service
error codes” in chapter 4 and “Codes and warnings” in chapter 6 for
details.
Note: This indicator also turns on during the LCD all-segments test.
Code Port
COM 0 Optical port
COM 1 Remote port 1
COM 2 Remote port 2
Display indicators
The 12 display indicators () are used to more precisely identify the
information displayed on the meter’s LCD.
These identifiers may be shown individually or in combination to describe a
particular displayed quantity.
The manufacturer’s nameplate details the meaning of the display
indicators. See Appendix C, “Nameplate and style number information.”
Tariff indicators
The tariff indicators (T1, T2, T3, and T4) indicate the current tariff. If the
displayed quantity is a TOU item (for example, tariff 1 total kWh), the
corresponding indicator (T1) turns on. If the quantity’s tariff is active at the
time, the tariff indicator flashes.
Note: The active tariff indicators also turns on during the LCD all-
segments test.
EOI indicator
The end-of-interval (EOI) indicator is used to verify the timing of the
demand interval. Ten seconds before the end of the demand interval, the
EOI indicator will be turned on and remain on until the end of the interval.
For rolling demand, the EOI indicator turns on for 10 seconds before the
end of each subinterval.
Push buttons
The following push buttons are located on the front of the A1800 ALPHA
meter:
RESET (sealable)
If sealed, the RESET button is only accessible after breaking the seal; the
button is always accessible.
If the A1800 ALPHA meter is ordered with the backlight display option,
then either the RESET button or the button can be used to illuminate the
display. See “Using the backlight” on page 43 for more information.
A1800 ALPHA meter push buttons
Figure 3-4. A1800 ALPHA meter push buttons
* (ALT) button
RESET button
(sealable)
RESET button
To activate the RESET button, it may be necessary to break the seal that
locks the RESET button in the inactive position. After the seal is broken,
rotate the push button 90 ° in either direction and press the push button
(see Figure 3-5). Pressing the RESET button performs a demand reset (see
“Demand reset” on page 48 for a description on what happens during a
demand reset). The RESET button performs differently depending on the
A1800 ALPHA operating mode, as shown in Table 3-2.
If the A1800 ALPHA meter is ordered with the backlight display option,
then the RESET button can be used to illuminate the display. See “Using
the backlight” on page 43 for more information.
Table 3-2. RESET button behavior
Mode Description
Normal Performs a demand reset
Alternate Returns to normal mode and performs a demand
reset
Test Resets test value and remains in test mode
Read without Power Sets the meter to perform a demand reset
immediately on power restoration
Operating instructions 42
To seal the RESET button, rotate the RESET button 90 ° back to the
inactive position and apply the seal.
Figure 3-5. RESET button positions
Using the RESET button to lock the service will not perform a demand
reset unless it is pressed a second time.
button
Pressing the button normally initiates the alternate mode (see
“Operating modes” on page 44 for more information about the A1800
ALPHA operating modes). The button performs differently depending on
the operating mode, as shown in Table 3-3.
If the A1800 ALPHA meter is ordered with the backlight display option,
then the Þ button can be used to illuminate the display. See “Using the
backlight” on page 43 for more information.
Operating instructions 43
The A1800 ALPHA meter can be ordered with the backlight always turned
on. With this option, the LCD backlight will always be illuminated, and the
RESET and Þ buttons will operate as specified in Table 3-2 and Table 3-3,
respectively.
Figure 3-6. Using the backlight on the A1800 ALPHA meter LCD (default operating mode)
Backlight off
Any button is
pressed
Backlight on
Button pressed
Yes, * while LCD lit? Yes, RESET
No
Yes
Operating modes
The A1800 ALPHA meter operates in one of the following modes:
Read without Power mode (optional)
normal mode
alternate mode
test mode
Read without Power mode (optional)
As part of its function, the meter performs self tests to make sure it is
operating normally. The self test ensures that the A1800 ALPHA meter is
functioning properly and that its displayed quantities are accurate. If the
self test indicates an error, the LCD displays the error indicator. In addition,
the meter can be programmed to “lock” the error code on the display. The
meter attempts to function normally, however, the meter data may be
suspect. See “Meter self test” in chapter 4 for more information on self
tests, and see chapter 6 for more information on errors.
Operating instructions 45
Normal mode
Normal mode is the default operation mode for the A1800 ALPHA meter. It
is generally used to display billing data on the LCD. The meter is fully
operational in this mode, and it will process and store data while the LCD
scrolls through the normal display list quantities.
The LCD test will always appear immediately after power is connected to
the A1800 ALPHA meter or after a power restoration from an outage.
Typically, the normal mode display cycle begins with an LCD test which
turns on all of the display segments. This is recommended because it
provides a quick way to determine if the LCD is functioning properly. The
LCD test can be disabled using Elster meter support software. The normal
display cycle will scroll through all programmed display quantities before
beginning the cycle again.
While in normal mode, the LEDs transmit pulses proportional to metered
energy. See “LED pulse outputs” in chapter 5 for details on the LEDs.
Alternate mode
Alternate mode can be programmed with Elster meter support software to
display a second set of quantities on the LCD. Alternate mode is most
often used for displaying non-billing data, but it can be programmed to
display any of the available quantities. This mode is activated in one of the
following ways:
pressing the button on the A1800 ALPHA meter
after power up for one cycle of the alternate display list
This feature can be disabled using Elster’s meter support software.
The meter is fully operational while in alternate mode. While in alternate
mode, the alternate display indicator is turned on. Additionally, the LEDs
transmit energy pulses (see “LED pulse outputs” in chapter 5).
There are several different ways to exit alternate mode. Whenever exiting
the alternate mode, the meter returns to normal mode.
Table 3-4. Exiting alternate mode
Method Description
Wait for the end of the If the meter is scrolling through the alternate
alternate display list display list automatically, the meter exits
alternate mode after the last item is displayed.
Press the RESET Exits alternate mode and performs a demand
button reset.
Wait for the timeout If the LCD remains on a quantity, the meter
exits alternate mode after 2 minutes of
inactivity. If the LCD remains on a pulse line
cumulative counter, the meter will exit the
alternate mode at midnight.
Power failure occurs Exits alternate mode; when power is restored,
the meter's display is in normal mode.
At midnight Exits alternate mode at the next midnight
crossing.
Operating instructions 46
Test mode
The A1800 ALPHA meter enters test mode by a command through the
optical port. While in test mode, the test mode indicator (TST) will flash on
the meter’s LCD.
Test mode displays test readings without affecting the present energy
usage and billing data values in the A1800 ALPHA meter. Shorter demand
intervals may be used in test mode to reduce demand test time and will
not interfere with billing data.
When normal mode is resumed, readings taken during test mode will be
discarded and present energy usage and billing data values will be
restored. The status of the meter (including billing data, profiling data,
errors, and warnings) before the meter entered test mode is restored.
While in test mode, the optical port transmits test pulses proportional to
metered energy (see “LED pulse outputs” in chapter 5).
Test mode is entered using Elster meter support software. The meter exits
test mode under any of the following conditions:
Table 3-5. Exiting test mode
Method Description
Test mode expires Automatically after a programmable timeout
has expired (between 1 and 255 test mode
intervals)
Send an exit command Using Elster meter support software, send
an exit command over the optical port.
Automatically after 24 Automatically after a programmable timeout
hours (1-255 test mode intervals).
Power failure occurs Exits test mode; when power is restored, the
meter's display is in normal mode.
The meter has limited operational capability while in Read without Power
mode. While in read without mode, only the following items can be
performed:
manually reading the LCD
performing a billing read or diagnostic using an optical probe over the
optical port1
manually performing a demand reset by pressing the RESET button
(see “RESET button” on page 41)
displaying the alternate display list by pressing the button (see “
button” on page 42)
Other functions, including data altering communication, meter
programming, service tests, and instrumentation, are not available while
the meter is in Read without Power mode. If an unavailable item is
programmed in a display list, the LCD will briefly display dashes (-) in the
display quantity area before scrolling to the next item.
The meter exits Read without Power mode when any of the following
actions occur:
power is restored to the meter
the maximum lifetime usage of the Read without Power battery is
exceeded
the specified timeout expires
The timeout period can be extended by 30 seconds if the user either
presses a button or starts an optical communication session toward the
end of the timeout period. The timeout period is extended by 30-second
increments up to a maximum of twice the specified Read without Power
timeout period.
1 Billing data is always available in Read without Power mode. Data stored in shared memory may not be available while in Read
without Power mode (for more information on shared memory, see “Logs and data sets” in chapter 2).
Operating instructions 48
Demand reset
A demand reset can be performed one of three ways:
pressing the RESET button
issuing a command over the optical or remote ports
as a scheduled calendar event
Regardless of how the demand was reset, the meter performs many
different functions, including the following:
the present billing data is copied to the demand reset data area
the billing data’s present maximum demand is added to the
cumulative demand, and then the billing data’s present maximum
demand is reset to zero
the billing data’s dates and times of the maximum demands are reset
to zero
the billing data’s present coincident values are reset to zero
all demand calculations are reset to zero and a new demand interval is
started
previous interval demands are reset to zero
present interval demands are reset to zero
all average power factor calculations are restarted
pulse line cumulative counters are cleared
current conditions for certain errors or warnings are cleared
Notes:
4 METER TOOLS
System instrumentation
System instrumentation is a collection of displayable items designed to
assist in evaluating a service by providing real time analysis of the
conditions present at the A1800 ALPHA installation. Instrumentation
quantities should not be confused with billing quantities because they are
intended for an entirely different purpose.
System instrumentation quantities are measured instantaneously while
billing quantities are measured and averaged over a number of minutes.
Instrumentation quantities are generally provided on a per phase basis,
while billing quantities represent a combination of all present phases. This
can result in discrepancies between similar billing and instrumentation
data, and this is to be expected.
The instrumentation measurements are near instantaneous. Using Elster
meter support software, instrumentation quantities may be placed in
normal, alternate, or test mode display sequences. The alternate mode
display sequence is recommended because it is generally not necessary
for these quantities to be displayed at all times.
Most instrumentation quantities are true root mean square (rms)
measurements over an even number of line cycles, but others are
compound quantities. Compound quantities require multiple
measurements at slightly different times with the results calculated from
these multiple measurements. Instrumentation quantities can also round
or restrict the quantity to a desirable value under certain system
conditions. See Table 4-1 for more information about how the
instrumentation quantities are obtained.
Table 4-1. Description of system instrumentation quantities
𝑘𝑣𝑎𝑟 = √𝑘𝑉𝐴2 − 𝑘𝑊 2
The result is then signed based on the kvar direction.
Phase voltages and currents True rms values measured by meter engine
Phase voltage angle relative to Line 1 voltage Each voltage angle is measured relative to line 1
voltage zero crossings and rounded to 30°
1 Beginning with series 200 meters, the values update approximately every ½ second. Prior versions, these quantities updated
approximately every second.
Meter tools 52
2
𝑘𝑉𝐴𝑣𝑒𝑐𝑡 = √𝑠𝑦𝑠𝑡𝑒𝑚 𝑘𝑊 2 + 𝑠𝑦𝑠𝑡𝑒𝑚 𝑘𝑣𝑎𝑟𝑣𝑒𝑐𝑡
System power factor (vectorial) System kW divided by system kVA (vectorial)
System power factor angle (vectorial) The arccosine of system power factor (vectorial)
Phase current angle relative to line 1 voltage Each current angle is measured relative to line 1
voltage zero crossings
Phase power factor Phase kW divided by phase kVA, both measured
simultaneously. Phase power factor is set to 1.00 if
phase current is less than the absolute minimum
current (twice starting amps).
Phase 1st harmonic (fundamental) voltage magnitude The per phase magnitude of the fundamental voltage
Phase 1st harmonic (fundamental) current magnitude The per phase magnitude of the fundamental current
Phase 2nd harmonic voltage magnitude The per phase magnitude of the 2nd harmonic voltage
Phase 2nd harmonic current magnitude The per phase magnitude of the 2nd harmonic current
Phase 2nd harmonic voltage percentage Per phase, the 2nd harmonic voltage magnitude
divided by the fundamental voltage magnitude
Phase total harmonic current magnitude Per phase, the square root of the sum of the 2nd to
15th harmonic currents squared. In other words:
𝑖=15
𝑇𝐻𝐶 = √ ∑ 𝐻𝐶𝑖2
𝑖=2
Meter tools 53
√∑𝑖=15 2
𝑖=2 𝐻𝐶𝑖
𝑇𝐷𝐷 =
𝑀𝑎𝑥𝑖𝑚𝑢𝑚 𝑎𝑚𝑝𝑠
Where HCi represents the ith harmonic current.
Instrumentation accuracy
kW and kvar instrumentation quantities at test amps, test volts, and
applicable power factor angles have a ± 0.5 % accuracy (typically less than
±0.25 %). Applicable power factor angles are ±60 degrees around the
power factor angle for maximum output of each quantity. Maximum power
factor angles for watts are 0 degrees and 180 degrees. Maximum power
factor angles for vars are 90 degrees and 270 degrees. Outside of these
applicable power factor angles, accuracy decreases as the magnitude of
the quantity of interest decreases.
Figure 4-1. Applicable power factor angles for kW instrumentation values
90°
120° 60°
Maximum Maximum
180° 0°
rec kW del kW
240° 300°
270°
Meter tools 54
Figure 4-2. Applicable power factor angles for kvar instrumentation values
Maximum
del kvar
90°
150° 30°
180° 0°
210° 330°
270°
Maximum
rec kvar
The quantity identifier gives information about the quantity being displayed
on the A1800 ALPHA meter LCD, as indicated in Table 4-2.
Meter tools 55
+P
L1L2 L3
+P
L1L2 L3
+P
L1L2 L3
+P
L1L2 L3
+P
L1L2 L3
To indicate that a service voltage test is complete, the LCD displays the
following (an example is shown in Figure 4-8):
phase rotation (for example, L1-2-3 or L3-2-1)
voltage magnitude (for example, 120 or 240)
service type showing the number of wires and the service type, for
example:
1L is a single phase service
+P
L1L2 L3
+P
L1L2 L3
Smart autolock
When smart autolock is enabled through Elster meter support software,
the A1800 ALPHA meter will attempt to lock the service automatically once
it is determined to be valid. Both the voltage magnitude and phase angle
of the service are compared to a table of valid relationships stored within
the meter memory. The meter accepts the service that most closely
matches one of the stored values in the A1800 ALPHA meter.
The A1800 ALPHA meter periodically checks the service. Under certain
conditions, the smart autolocked service may lock on a different service.
This is useful because the meter may have been moved to a new service.
The service voltage test will be performed and the service may be changed
in response to the following events:
power up
exit of test mode
after a data-altering communication session
If a new, valid service is detected, the meter locks on the new service. If a
valid service cannot be detected, the meter responds in the following
manner:
the meter remains locked on the last known valid service
the LCD displays an error code
Meter tools 59
Manual lock
When configured through Elster meter support software for manual lock,
the A1800 ALPHA meter will detect and evaluate the service in the same
manner as it does when autolock is enabled. The identified service
information will also be shown on the LCD; however, the RESET button
must be pressed in order to lock the detected service (see “Using to lock
service” in chapter 3).
When the service type has been detected, the phase rotation, voltage
magnitude, and the service type will be displayed on the LCD. If the RESET
button is not pressed to accept the service, the LCD will alternate between
L1-2-3 ------ and the detected service information until the service has been
manually locked.
At midnight
If the service is locked, the meter checks the service at midnight. The
meter always does the following when the service voltage test is run at
midnight:
Meter tools 61
Manual lock
Smart autolock
Current state is locked
1 The meter initiates the service 4 The meter initiates the service
voltage test. test.
2 The phase indicator voltage 5 The phase indicator voltage
threshold levels are based on threshold levels are based on
the currently locked service. the currently locked service.
3 The meter attempts to match 6 The meter attempts to match
the service. the service.
If the service matches the If the service matches the
presently locked service, presently locked service,
then the LCD displays the then the LCD displays the
locked valid service. locked valid service.
If the service does not If the service does not
match the presently locked match the presently locked
service, then the LCD service, then the LCD
displays SE 555000. The displays a service test error.
meter restarts the service The meter restarts the
voltage test in diagnostic service voltage test in
mode (see “Restarting the diagnostic mode (see
service voltage test in “Restarting the service
diagnostic mode” on page voltage test in diagnostic
62). However, the lock mode” on page 62).
remains on the last valid However, the lock remains
service until a new valid on the last valid service until
service is detected. a new valid service is
detected.
As a TRueQ test
When the service voltage test is programmed as a TRueQ test, the service
test is performed only if the service is locked. TRueQ tests are available
only on meters with TRueQ capabilities. See “Service voltage test” on page
57 for more information.
If at any point a valid service is found and locked, the meter displays the
locked service on the LCD and continues to the next item in the display
sequence. Otherwise, the cycle restarts at step 1.
+P
L1L2 L3
If the test is not successful, a warning is set. Also, the LCD will indicate a
service error by displaying SE and a code, an example of which is shown in
Figure 4-11. See “System service error codes” on page 64 for more
information. The following conditions can cause the service current test to
fail:
current remains on one phase while no current is on any other phase
current on any single phase is below the programmed low current limit
current on any phase is greater than the programmed absolute
maximum
current is negative on any phase (reverse power)
power factor on any phase is less than the limit set for leading or
lagging power factor
If all phases are below the absolute minimum current threshold, the low
and missing current failure will not be reported. It is assumed that this is a
valid, no-load condition. In this case, the low and zero current warnings will
display if the condition exists.
Figure 4-11. Service current test error
+P
L1L2 L3
Meter tools 64
Error code
Service error condition (SE) Voltage phase
L1 L2 L3
Low nominal voltage on Line 1 1 0 0 0 0 0
Low nominal voltage on Line 2 0 1 0 0 0 0
Low nominal voltage on Line 3 0 0 1 0 0 0
High nominal voltage on Line 1 2 0 0 0 0 0
High nominal voltage on Line 2 0 2 0 0 0 0
High nominal voltage on Line 3 0 0 2 0 0 0
Unrecognized service 5 5 5 0 0 0
Bad phase angle on Line 1 8 0 0 0 0 0
Bad phase angle on Line 2 0 8 0 0 0 0
Bad phase angle on Line 3 0 0 8 0 0 0
Low voltage & bad phase angle on Line 1 9 0 0 0 0 0
Low voltage & bad phase angle on Line 2 0 9 0 0 0 0
Low voltage & bad phase angle on Line 3 0 0 9 0 0 0
High voltage & bad phase angle on Line 1 A 0 0 0 0 0
High voltage & bad phase angle on Line 2 0 A 0 0 0 0
High voltage & bad phase angle on Line 3 0 0 A 0 0 0
Meter tools 66
Error code
Service error condition (SE) Current phase
L1 L2 L3
Missing Line 1 current 0 0 0 1 0 0
Missing Line 2 current 0 0 0 0 1 0
Missing Line 3 current 0 0 0 0 0 1
Low Line 1 current 0 0 0 2 0 0
Low Line 2 current 0 0 0 0 2 0
Low Line 3 current 0 0 0 0 0 2
Missing & low current on Line 1 0 0 0 3 0 0
Missing & low current on Line 2 0 0 0 0 3 0
Missing & low current on Line 3 0 0 0 0 0 3
Low PF on Line 1 0 0 0 4 0 0
Low PF on Line 2 0 0 0 0 4 0
Low PF on Line 3 0 0 0 0 0 4
Reverse power on Line 1 0 0 0 5 0 0
Reverse power on Line 2 0 0 0 0 5 0
Reverse power on Line 3 0 0 0 0 0 5
Low PF & low current on Line 1 0 0 0 6 0 0
Low PF & low current on Line 2 0 0 0 0 6 0
Low PF & low current on Line 3 0 0 0 0 0 6
Reverse power & low on Line 1 0 0 0 7 0 0
Reverse power & low on Line 2 0 0 0 0 7 0
Reverse power & low on Line 3 0 0 0 0 0 7
Excess current on Line 1 current 0 0 0 8 0 0
Excess current on Line 2 current 0 0 0 0 8 0
Excess current on Line 3 current 0 0 0 0 0 8
Excess current & low PF on Line 1 current 0 0 0 C 0 0
Excess current & low PF on Line 2 current 0 0 0 0 C 0
Excess current & low PF on Line 3 current 0 0 0 0 0 C
Excess current & reverse power on Line 1 current 0 0 0 d 0 0
Excess current & reverse power on Line 2 current 0 0 0 0 d 0
Excess current & reverse power on Line 3 current 0 0 0 0 0 d
Meter tools 67
TRueQ monitoring
All A1800 ALPHA meters are equipped with the tamper restraint and power
quality (TRueQ) monitoring features that can monitor circuit parameters on
a cyclic basis, 24 hours a day throughout the billing period. TRueQ tests
may be turned on or off through Elster meter support software.
TRueQ tests will recognize any deviation beyond the thresholds. When
shipped, the meter is stored with default values for the thresholds. Using
Elster meter support software, these thresholds can be edited.
Most TRueQ tests are performed individually so that circuit parameters are
not being monitored continuously. Each subsequent test will begin
immediately after the previous one has ended. The momentary voltage sag
test, however, uses the per phase rms voltage calculation which is part of
the voltage sensing process within the meter engine. The rms voltages are
calculated once every 2 line cycles, so the momentary voltage sag test is
capable of recognizing any phase voltage deviation that remains below a
specified threshold for as few as 2 line cycles.
TRueQ timing
In addition to defining thresholds for each test, a minimum time may also
be defined. Once the monitored parameter falls outside the threshold and
remains there longer than the minimum time, the failure will be stored and
the cumulative count will increment by one. A cumulative timer will also be
activated and will run for as long as the event is detected. The cumulative
count and timer for each test can be retrieved through Elster meter
support software.
TRueQ log
All A1800 ALPHA meters record TRueQ events in the TRueQ log. Meters
with TOU capability will also record the date and time of any TRueQ failure
in the TRueQ log. See “TRueQ Log” in chapter 2 for more information about
the TRueQ log.
A qualified TRueQ failure causes the W2 020000 warning code to be
shown on the LCD. See “W2 020000: TRueQ test failure warning” in
chapter 6 for more details.
Voltage sags
A momentary sag in voltage can reset process control equipment and
computer systems. The momentary voltage sag monitor watches for
decreases in voltage that last for a measured number of cycles. This
monitor can detect any voltage decrease that falls below a programmed
threshold for as few as 2 line cycles. Threshold and duration are defined
using Elster meter support software.
Meter tools 68
TRueQ tests
TRueQ tests do not interfere with any meter functions related to energy
measurement. These tests run separately from the metering functions.
Depending on the test set selected using Elster meter support software,
different test sets support different tests:
default Elster TRueQ test set
enhanced TRueQ test set
IEEE voltage imbalance test set
EN 50160 (requires A1800 ALPHA meter firmware version 4.0 or later)
Tamper detect test set
Meter tools 69
During the low current and reverse power and power factor tests, there will
be no event detected if all measured line currents drop below the absolute
minimum current threshold. An event will be detected if any single phase
or two phases drop below the programmed threshold for the qualification
time. This eliminates false detection when the load is dramatically reduced
or turned off.
Name Service voltage test Test set Default Enhanced IEEE EN5160 Tamper detect
Description Per phase leading and lagging PF threshold values are defined in the system service
current test specifically for L1-L2-L3 phase rotation. This permits independent PF
thresholds for each service type. If the locked service has L3-L2-L1 phase rotation, then
the per phase leading and lagging thresholds are reversed (that is, the same threshold
for Line 1 minimum leading PF with L1-L2-L3 rotation is also used as the Line 1
minimum lagging PF for L3-L2-L1 rotation).
Also, if the Irms value for a given phase is less than zero current threshold, then no
power factor or reverse energy failure will be detected from that phase (to eliminate
false failures).
Stored value None
Meter tools 74
Formula 𝐼𝑓 𝑙𝑎𝑔𝑔𝑖𝑛𝑔 𝑃𝐹, 𝑡ℎ𝑒𝑛 (𝑃𝐹𝐿1 𝑜𝑟 𝑃𝐹𝐿2 𝑜𝑟 𝑃𝐹𝐿3 ) < 𝑠𝑝𝑒𝑐𝑖𝑓𝑖𝑒𝑑 𝑙𝑎𝑔𝑔𝑖𝑛𝑔 𝑃𝐹 𝑡ℎ𝑟𝑒𝑠ℎ𝑜𝑙𝑑 𝑜𝑟
𝐼𝑓 𝑙𝑒𝑎𝑑𝑖𝑛𝑔 𝑃𝐹, 𝑡ℎ𝑒𝑛 (𝑃𝐹𝐿1 𝑜𝑟 𝑃𝐹𝐿2 𝑜𝑟 𝑃𝐹𝐿3 ) < 𝑠𝑝𝑒𝑐𝑖𝑓𝑖𝑒𝑑 𝑙𝑒𝑎𝑑𝑖𝑛𝑔 𝑃𝐹
0.00 to 1.00 for the minimum lagging power factor with L1-L2-L3 rotation (per phase);
Variable
0.00 to 1.00 for the minimum leading power factor with L1-L2-L3 rotation (per phase)
0.20 for the minimum lagging power factor with L1-L2-L3 rotation (per phase)
Default value
0.20 for the minimum leading power factor with L1-L2-L3 rotation (per phase)
Configuration based on Specified threshold for leading and lagging power factors
This test checks the power factor for any deviation beyond the programmed threshold.
This monitor may be used alone to monitor rate-based conditions in conjunction with
the reverse power test and PF monitor to provide a more thorough analysis of power
factor fluctuations.
The leading and lagging thresholds are individually defined for each phase. These
settings may be different than those defined in the service current configuration.
Description
Per phase leading and lagging threshold values are defined specifically for L1-L2-L3
phase rotation. If the locked service has L3-L2-L1 rotation, then the per phase leading
and lagging thresholds are reversed (that is, the same threshold value for Line 1
minimum leading power factor with L1-L2-L3 rotation is also used as the Line 1 lagging
power factor with L3-L2-L1 rotation). Also, if the Irms value for a given phase is less than
the zero current threshold, then no power factor failure will be reported, regardless of
the calculated power factor value (to eliminate false failures).
Stored value None
Meter tools 75
Magnetic detection
Name Test set Tamper detect
(leading PF angle)
𝐼𝑓 [(𝑊𝐿1 > 0 𝑎𝑛𝑑 𝐼𝐿1
> 𝑀𝑖𝑛𝑖𝑚𝑢𝑚 𝑡𝑒𝑠𝑡 𝑝ℎ𝑎𝑠𝑒 𝑐𝑢𝑟𝑟𝑒𝑛𝑡 𝑎𝑛𝑑 𝐿𝑖𝑛𝑒 1 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑃𝐹 𝑎𝑛𝑔𝑙𝑒
< 𝑠𝑝𝑒𝑐𝑖𝑓𝑖𝑒𝑑 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑝𝑜𝑤𝑒𝑟 𝑓𝑎𝑐𝑡𝑜𝑟 𝑎𝑛𝑔𝑙𝑒 𝑡ℎ𝑟𝑒𝑠ℎ𝑜𝑙𝑑) 𝑜𝑟
Formula (𝑊𝐿2 > 0 𝑎𝑛𝑑 𝐼𝐿2 > 𝑀𝑖𝑛𝑖𝑚𝑢𝑚 𝑡𝑒𝑠𝑡 𝑝ℎ𝑎𝑠𝑒 𝑐𝑢𝑟𝑟𝑒𝑛𝑡 𝑎𝑛𝑑 𝐿𝑖𝑛𝑒 2 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑃𝐹 𝑎𝑛𝑔𝑙𝑒
< 𝑠𝑝𝑒𝑐𝑖𝑓𝑖𝑒𝑑 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑝𝑜𝑤𝑒𝑟 𝑓𝑎𝑐𝑡𝑜𝑟 𝑎𝑛𝑔𝑙𝑒 𝑡ℎ𝑟𝑒𝑠ℎ𝑜𝑙𝑑) 𝑜𝑟
(𝑊𝐿3 > 0 𝑎𝑛𝑑 𝐼𝐿3 > 𝑀𝑖𝑛𝑖𝑚𝑢𝑚 𝑡𝑒𝑠𝑡 𝑝ℎ𝑎𝑠𝑒 𝑐𝑢𝑟𝑟𝑒𝑛𝑡 𝑎𝑛𝑑 𝐿𝑖𝑛𝑒 3 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑃𝐹 𝑎𝑛𝑔𝑙𝑒
< 𝑠𝑝𝑒𝑐𝑖𝑓𝑖𝑒𝑑 𝑛𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑 𝑝𝑜𝑤𝑒𝑟 𝑓𝑎𝑐𝑡𝑜𝑟 𝑎𝑛𝑔𝑙𝑒 𝑡ℎ𝑟𝑒𝑠ℎ𝑜𝑙𝑑)]
90.0 degrees to -90.0 degrees (leading or negative angles are expected, but lagging or
Variable
positive angles are allowed for special conditions) (See special note below)
Default value -15 degrees (that is, 15 degrees leading)
Configuration based on Specified normalized PF angle threshold
When a magnet begins to saturate a current sensor core, it causes a leading phase
shift in the output of the current sensor. Most loads are unity to lagging, so a leading
power factor phase angle is unusual. This test looks for that shift at some (typically)
leading phase angle to detect the possible presence of a magnet tamper condition. If
however, there are legitimate leading power factor load conditions, the test should
either be disabled or have the threshold set to a more leading value.
Each phase is tested individually, and if any phase fails, the test will fail. Once any
phase is failing, then all phases must be passing for the PQM test failure to be over.
Description Each phase is initially checked to ensure that real energy (Wh) is in the Delivered
direction. If it is, then the rms current for that phase is checked to ensure that it is
greater than a user defined minimum test phase current threshold (as a percentage of
Class Amps, and defined in the Service Test component). If it is, then the normalized
power factor angle is tested to see if it is more leading than the normalized power
factor angle threshold (negative values are leading so less than means more leading).
Normalization of a power factor angle subtracts the nominal per phase power factor
angle (which may vary depending on the service type and phase rotation) from the
actual per phase power factor angle. This allows a single threshold angle to be used for
comparison purposes on all phases, and all service types and rotations.
Stored value The normalized power factor phase angle difference (always a positive value)
The specified normalized power factor angle threshold is expected to be leading
(negative), but lagging (positive) angles are allowed for special conditions where actual
loads are expected to be significantly lagging. The test will still fail if an actual
normalized PF angle is less (more leading) than the specified normalized power factor
Special note angle threshold, even if the angle and the threshold are both lagging (positive) angles.
But the user should beware that use of lagging angles would generally be a source of
false failures.
To prevent the monitor from creating a false alarm from legitimate leading PF phase
angles, the recommended qualification time is 15 minutes.
Security
All A1800 ALPHA meters include features that help prevent unauthorized
access to meter data and record events that may indicate meter
tampering.
Meter passwords
Access to the A1800 ALPHA meter is protected through the use of
passwords. When establishing communication with the meter, the meter
will request a password. If the correct password is not supplied, the meter
will not communicate or perform the commands that are issued.
Passwords help ensure that the meter data is protected and that the
programming cannot be altered without proper authorization.
The A1800 ALPHA meter uses three passwords to control access to the
meter. As shown in Table 4-6, each password allows different activities
that can be performed on the meter. For more information regarding
passwords, see the documentation that comes with the Elster meter
support software.
Table 4-6. A1800 ALPHA meter passwords
When communicating with the A1800 ALPHA meter remotely, the meter
supports the password encryption standards in accordance with ANSI
C12.21. In accordance with ANSI C12.18, the password is not encrypted
when communicating using the optical port.
The meter records the number of failed password attempts that were used
in trying to access the meter. An internal warning will be generated if 10
failed password attempts occur since the last demand reset. This warning
can be used to control a relay output or to trigger an alarm call.
Meter tools 89
Anti–tampering
All A1800 ALPHA meters provide auditing capabilities that can be used to
indicate potential meter tampering. These capabilities can record such
items as the following:
programming changes
power outages
number of days since last energy pulse
number of manually-initiated demand resets
number of days since last demand reset
reverse energy flow
history log
cover removal detection
Program protection
As a security feature, the A1800 ALPHA meter can be ordered with
program protection. Program protection prevents metrological parameters
from being altered. Some data and configuration parameters can be
altered while in program protection. These alterable items must be
specified at ordering and can include the following:
communication parameters
TRueQ parameters
time of day (TOU or load profiling configurations)
switch times (TOU configurations)
special dates list (TOU or load profiling configurations)
All other parameter changes require the meter to exit program protect
mode. The A1800 ALPHA meter supports one of two methods of exiting
program protect mode:
breaking the terminal cover seals and meter cover seals
pressing the RESET and buttons
The desired method to exit program protect must be specified at the time
of ordering
Breaking seals to temporarily disable program protection:
1 Break the terminal cover seals and remove terminal cover. The TC
indicator will turn on.
2 Break the meter cover seals and lift the meter cover.
3 At this point, you can perform any of the data or program altering
operations available using the Elster meter support software.
4 Close the meter cover and install the seals.
5 Install the terminal cover and seals. The TC indicator will turn off.
If programmed to do so, changes in the state of the terminal cover and the
meter cover are logged in the event log. See “Event log” in chapter 2 for
details.
Meter tools 90
Notes:
5 OUTPUTS
Relay outputs
The A1810 ALPHA meter does not include relay outputs. The A1830 ALPHA
meter and higher supports up to 4 relays.
Figure 5-1. A1800 ALPHA meter with RS-232 as second communication port
RS-232 connector
(optional)*
RS-485 terminals
RS-232
connector
RS-485 connector
(optional)*
RS-485 terminals
RS-232
connector
Ethernet connector(optional)*
RS-232 terminals
RS-485 terminals
For more information about relay outputs and communications, see the
instructional leaflet (IL) that comes with the option board.
The output relays on the main circuit board can switch up to 125 VAC or
180 VDC at up to 70 mA. See Appendix D, “Wiring diagrams.”
With the A1800 ALPHA meter, all relay outputs are fully programmable
using Elster meter support software. Sources for relay outputs are listed in
Table 5-1.
Outputs 96
½ ½ ½ ½ ½ ½
On
Pulse
Off
Pulse period Pulse period Pulse period
In toggle mode, a relay changes state for each energy pulse received from
the meter engine.
Outputs 97
10 msec. 10 msec.
10 msec.
On
Pulse
Off
Pulse period Pulse period Pulse period
Specifications
Relay C, when activated, supplies a pulse that will connect the interposing
relay. Relay D, when activated, supplies a pulse that will disconnect the
interposing relay.
The pulse duration is between 24 milliseconds and 32 milliseconds. Relay
C and Relay D can switch up to 125 VAC or 180 VDC at up to 70 mA.
Arming
It is possible to program the meter to delay activating Relay C until the
button is pressed. The meter is “armed” remotely; however, arming the
interposing relay control does not activate Relay C. Instead, Relay C can
only be activated by a user pressing the button on the meter.
It is also possible to clear an armed state remotely by sending a command
or by using Metercat.
LCD behavior
When the meter is enabled for control of an interposing relay, the meter
LCD indicates the state of the interposing relay:
When the state of the interposing relay is disconnected, the meter
displays a warning on the LCD (see “W1 001000: Interposing relay
disconnected” in chapter 6).
When the state of the interposing relay is connected, the meter LCD
operates as expected based on its operating mode.
If the meter is armed for activation, the meter locks the LCD with the
display ARMED / PUSH .This special display overrides all other
display modes.
Outputs 99
Figure 5-6. Display for the meter armed for interposing relay activation
Power failures
If the meter experiences a power failure, the state of the interposing relay
is maintained when power is restored to the meter. For example, if the
meter is armed for interposing relay activation when the power fails, the
meter will maintain the armed state when power is restored.
For meters with the optional Read without Power feature, because it is not
possible to change the state of the interposing relay during a power failure,
the meter will not display the armed state on the LCD.
Relay-related alarms
The A1800 ALPHA meter periodically performs a self test to determine if it
is operating properly. If any errors are detected, the meter can respond in
any or all of the following ways:
display an error or a warning (see “Codes” in chapter 2)
initiate a telephone call using a modem
trigger a relay
See Table 5-2 for errors, warnings, and events that can trigger a relay.
Outputs 100
Table 5-2. Errors, warnings, and events that can trigger a relay
Condition Description
Carryover error See “E1 000001: Carryover error” in chapter 6.
Clock error See “E3 030000: Clock error” in chapter 6.
Crystal oscillator error See “E1 000010: Crystal oscillator error” in chapter 6.
Demand overload warning See “W1 100000: Demand overload warning” in chapter 6.
EEPROM access error See “E1 010000: EEPROM access error” in chapter 6.
End of calendar warning See “W2 200000: End of calendar warning” in chapter 6.
Event log wrap event The event log has exceeded the maximum number of entries, and the
oldest records will be overwritten.
General configuration error See “E1 100000: General configuration error” in chapter 6.
History log wrap warning The history log has exceeded the maximum number of entries.
Depending on programming, the meter will either lock the history log or
start overwriting the oldest records. If the history log is locked, no
further changes to the meter are allowed until the history log has been
read.
Improper meter engine operation See “W1 000010: Improper meter engine operation warning” in chapter
warning 6.
Instrumentation profiling set 1 wrap Set 1 of the instrumentation profiling log is within 2 days of overflowing.
imminent event Data will be lost if the instrumentation profiling log is not read within 2
days.
Instrumentation profiling set 2 wrap Set 2 of the instrumentation profiling log is within 2 days of overflowing.
imminent event Data will be lost if the instrumentation profiling log is not read within 2
days.
Internal communication error See “E1 001000: Internal communication error” in chapter 6.
Low battery warning See “W1 000001: Low battery warning” in chapter 6.
Possible tamper warning This condition indicates possible tampering of the meter because a
specified number of invalid passwords used to access the meter has
been used (called “tamper detect warning” in this manual). This
condition does not generate an error or warning code on the LCD.
Potential indicator warning See “W1 010000: Potential indicator warning” in chapter 6.
Power fail data save error See “E2 200000: Power fail data save error” in chapter 6.
Pulse profiling wrap imminent event The pulse profiling log is within 2 days of overflowing. Data will be lost if
the pulse profiling log is not read within 2 days.
Rate override warning The current TOU rate is being overridden by the alternate TOU rate
schedule.
Reverse energy flow warning See “W1 000100: Reverse energy flow warning” in chapter 6.
Service current test failure warning See “W2 000002: Service current test failure warning” in chapter 6.
Service voltage test failure warning The service voltage test was unable to find a valid service or the
measured service does not match the locked service.
Outputs 101
Output specifications
The LEDs support up to 120 pulses per second. The pulse width is fixed at
8 msec.
Depending on the operating mode of the meter, the LEDs are programmed
at the factory to emit a pulse as follows:
Error codes
Error codes override any other item that is being displayed on the LCD.
Using Elster support software, error codes can be configured to “lock” the
display, preventing other items from being displayed, and the error
indicator turns on. There are exceptions to errors locking the display:
The normal and alternate display sequence can be viewed even when
an error code locks the display. See “ button” in chapter 3 for more
information.
Warning codes can be programmed to display an error code. When the
condition causing the warning code is clear, the error code is no longer
displayed. See “E3 300000: Display locked by warning” on page 106
for more information.
Communication codes are temporarily displayed on the LCD even when the
LCD is “locked” by an error code. After the communication code clears,
LCD returns to showing the error code.
Error codes are indicated on the LCD by a group code and a numerical
code. The group code makes it easier to identify the error on the LCD. The
numerical code indicates the specific condition that has occurred. See
Figure 6-1 for a sample error code displayed on the meter LCD. Table 6-1
through Table 6-3 describe the different error conditions and their codes.
Codes 104
+P
L1L2 L3
Condition Code
Carryover error 0 0 0 0 0 1
Crystal oscillator error 0 0 0 0 1 0
Table CRC error 0 0 0 1 0 0
Internal communication error 0 0 1 0 0 0
EEPROM access error 0 1 0 0 0 0
General configuration error 1 0 0 0 0 0
Condition Code
Security configuration error 0 0 0 0 0 2
Password table CRC error 0 0 0 0 2 0
Encryption key table CRC error 0 0 0 2 0 0
ROM fail error 0 2 0 0 0 0
Power fail data save error 2 0 0 0 0 0
Condition Code
Clock error 0 3 0 0 0 0
Display locked by warning 3 0 0 0 0 0
Error codes of the same group are displayed in combination (E1 001010,
for example), indicating that more than one error condition has been
detected. If errors exist in more than one group, the meter will continually
cycle through the different groups. Any problems must be corrected before
normal operation can continue. In some cases, the meter may need to be
reprogrammed or returned to the factory for repair or replacement.
Codes 105
Since shipping can take several days, this error will likely be seen on
meters shipped without a connected battery.
The meter battery may need to be replaced, and the error will need to be
reset through Elster meter support software. If the error code is still shown
after using Elster meter support software, the meter must be returned to
the factory for repair or replacement.
1 Billing data is always stored in nonvolatile memory. Depending on meter configuration, other data may be stored in RAM, which
uses a battery to preserve memory. If the battery fails, this data would be lost.
Codes 106
Warning codes
Warning codes indicate conditions of concern that do not yet affect the
integrity of billing data. When the condition is present, a warning code is
automatically inserted as the last item in the normal and alternate display
sequences. When the condition clears, the warning code, is removed from
the display sequence. Elster meter support software can be used to select
individual warnings that will lock the display as an error. See “Error codes”
on page 103 for more information.
Warning codes are indicated on the LCD by a group code and a numerical
code. The group code makes it easier to identify the error on the LCD. The
numeric code indicates the specific condition that has occurred. See
Figure 6-2 for a sample warning code displayed on the LCD. Table 6-4
through Table 6-6 describe the different warning conditions and their
codes.
Figure 6-2. Sample warning code
+P
L1L2 L3
Condition Code
Low battery warning 0 0 0 0 0 1
Improper meter engine operation 0 0 0 0 1 0
warning
Reverse energy flow warning 0 0 0 1 0 0
Interposing relay disconnected 0 0 1 0 0 0
Potential indicator warning 0 1 0 0 0 0
Demand overload warning 1 0 0 0 0 0
Condition Code
Service current test failure warning 0 0 0 0 0 2
Demand threshold exceeded warning 0 0 0 2 0 0
Line frequency warning 0 0 2 0 0 0
TRueQ test failure warning 0 2 0 0 0 0
End of calendar warning 2 0 0 0 0 0
Condition Code
Read without Power battery discharged 0 0 0 0 0 3
Read without Power mode active 0 0 0 0 3 0
Codes 108
Communication codes
Communication codes temporarily override any other item that is being
displayed on the LCD (including error codes). Communication codes are
indicated on the LCD by a port code and a numerical code. The port code
identifies the affected port. The numerical code indicates the status of the
communication session. See Figure 6-3 for a sample communication code
displayed on the meter’s LCD. See Table 6-7 for the communication codes
that can be displayed.
Figure 6-3. Sample communication code
+P
L1L2 L3
Condition Code
CRC error C 0 0 1 0 1
Syntax error C 0 0 0 0 3
Framing error C 0 0 1 0 4
Timeout error C 0 0 1 0 5
Notes:
7 INSTALLATION &
REMOVAL
Preliminary inspection
The A1800 ALPHA meter is calibrated and tested at the factory, and it is
ready for installation. Follow proper installation and removal procedures for
personal safety and protection of the meter.
Before installing and applying power to the A1800 ALPHA meter, a quick
inspection of the meter itself is recommended. Check for some of the
following items:
no broken or missing parts
no missing or broken wiring
no bent or cracked components
no evidence of overheating
check the nameplate to make sure it is appropriate for the service
Physical damage to the outside of the A1800 ALPHA meter could indicate
potential electronic damage in the inside of the meter. Do not connect
power to a meter that is suspected to have unknown internal damage.
Contact your local Elster representative if you suspect your meter may be
damaged.
Installation & removal 114
Make sure to install the correct meter for the service type, maximum
current, and capacity required. Always verify that the maximum meter
voltage and current ratings are equal to or greater than the maximum
service voltage and current. Installing inappropriate meters can damage
equipment.
To use the A1800 ALPHA meter effectively and safely, follow this
procedure:
1 Make sure that the meter hanger, located on the base of the A1800
ALPHA meter, is in the desired position. Sliding the hanger down to the
hidden position will hide the top supporting screw.
2 Use at least an M6 screw for the top supporting position and hang the
meter on it, making sure it is level. The meter will operate correctly in
any position, but failing to mount the meter in a proper vertical
position will place the other mounting holes at the wrong place on the
mounting panel.
3 Use at least an M6 screw in each of the bottom supporting screws to
secure the A1800 ALPHA meter enclosure; the mounting holes are 7.1
mm (0.28 inches) in diameter.
Before wiring the meter into the power circuit, use authorized utility
procedures to install proper ground connections on all appropriate VT and
CT circuits and on the meter ground terminals. Also, be certain that CTs on
energized lines are securely short-circuited either with circuit-closing test
switches or with properly installed conductors. Dangerous voltages can be
present. Personal injury, death, or equipment damage can result from
wiring an ungrounded meter or mishandling improperly grounded metering
transformer circuits.
A1800 ALPHA meter terminals are designed for optimum use with copper
wiring. For direct connect-rated meters, aluminum wiring can be used but
if so, it is extremely important to use proper aluminum wiring practices.
Aluminum wiring compound or wiring paste (grease) should be used when
attaching the bottom-connected terminals. Tighten the connections, allow
them to relax for a few minutes, then tighten them again. This will
minimize the cold-flow effects of aluminum cable. Failure to observe
correct practices for installing aluminum wiring could lead to overheating
of the terminals, equipment failures, or damaging fires. Where possible,
Elster recommends copper-compatible meter terminals and aluminum
wire. Such adapters also can provide for use or larger aluminum
conductors that can be otherwise used in the terminals of the A1800
ALPHA meter.
Screw mounts
Before installing the battery, the A1800 ALPHA meter must have been
energized for at least 1 minute within the preceding 60 minutes. This
ensures that the supercapacitor is properly charged and that the battery is
not immediately drained upon installation. If this is not done, then the
battery may be damaged and the meter may not function correctly. While
the meter is powered, verify that the LCD is active and functioning.
To install the battery:
1 If the meter has not been energized for at least 1 minute during the
previous 60 minutes, energize the meter for 1 minute.
If the meter has been energized for at least 1 minute during the
previous 60 minutes, proceed to step 2.
2 De-energize the meter.
3 Remove the terminal cover screws and seals.
4 Remove the terminal cover to expose the battery well.
Installation & removal 117
TOU battery
5 Slide the battery leads into the connector to the right of the battery
well.
6 Place the battery firmly in the battery well.
7 Replace the terminal cover.
8 Energize the meter and verify that the LCD becomes active and
functioning properly.
Verify that the low battery symbol on the meter LCD is not displayed.
See “Indicators and controls” in chapter 3 for details.
9 Replace the terminal cover screws and seals.
10 Reprogram the meter or clear the errors (as necessary).
Troubleshooting
Not following this procedure can cause the meter to function improperly. In
case a battery has been installed correctly and the meter is not
functioning properly (for example, display is blank but the meter is
powered), use the following procedure.
1 De-energize the meter and let it sit without power for 48 to 72 hours.
This provides sufficient time for the supercapacitor to discharge and
for the microcontroller to shut down.1
2 Energize the meter for at least 1 minute. The microcontroller should
power up correctly and the supercapacitor will charge. Verify that the
LCD becomes active and functioning correctly.
3 De-energize the meter and insert the battery, following the instructions
earlier in this section.
If the meter still does not function properly, then it should be returned to
the factory.
1 If the battery was installed with the polarity reversed, the battery should not be damaged. If the battery was installed without
having the meter properly energized, then the battery will lose approximately 8.5 % of its service life each day.
Installation & removal 118
Initial setup
After installing and powering the A1800 ALPHA meter, verify the following:
The system service voltage test (if enabled) shows the valid service for
this installation. The phase rotation, service voltage, and service type
should be indicated on the LCD. Other validation information can be
obtained using the system instrumentation display quantities.
All potential indicators (from L1 to L3 depending on the wiring) are
present and are not blinking. A blinking indicator means that the
phase is missing the required voltage or is below the programmed
minimum voltage threshold value.
The LEDs are blinking and the energy direction indicators on the LCD
show the correct energy flow direction.
Required meter seals are in place.
Any information (such as registration and location of the meter) has
been recorded.
If the meter is not working correctly after it has been installed, then check
for improper installation or wiring. If the installation and wiring are correct,
then verify these other areas:
Introduction
What is Loss Compensation?
The Handbook for Electricity Metering defines loss compensation as
follows:
A means for correcting the reading of a meter when the metering point
and the point of service are physically separated resulting in
measurable losses including I2R losses in conductors and
transformers, and iron-core losses. These losses may be added to, or
subtracted from the meter registration.1
For example, it may be desirable to measure the energy usage on the low
voltage side of a distribution transformer that serves an industrial
customer even though the end-point customer actually owns the
transformer and is responsible for any transformer losses. In this case, the
utility billing point is actually the high voltage side of the transformer. Using
loss compensation, the meter on the low voltage side of the transformer
can actively adjust the energy registration to account for the losses in the
transformer.
Availability
The loss compensation functionality is available only on the following CT-
connected A1800 ALPHA meter configurations (with the loss compensation
key enabled):
2-element
3-element
1 Edison Electric Institute, Handbook for Electricity Metering, 10th edition, Washington, DC: Edison Electric Institute, 2002, p. 16.
Loss compensation 122
Item Equation
VAphase If bank of 3 transformers:
𝑉𝐴𝑝ℎ𝑎𝑠𝑒 = 𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000
If one 3-phase transformer:
(𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000)
𝑉𝐴𝑝ℎ𝑎𝑠𝑒 =
3
Vsec rated For 2 element, 3-wire delta applications:
𝑉sec 𝑟𝑎𝑡𝑒𝑑 = 𝑉sec 𝐿−𝐿
For 3 element, 4-wire wye applications:
𝑉𝑠𝑒𝑐 𝐿−𝐿
𝑉𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 =
√3
Vpri rated For 2 element, 3-wire delta applications:
𝑉𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 = 𝑉𝑝𝑟𝑖 𝐿−𝐿
For 3 element, 4-wire wye applications:
𝑉𝑝𝑟𝑖 𝐿−𝐿
𝑉𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 =
√3
Isec rated All applications:
√3 × 𝑉𝐴𝑝ℎ𝑎𝑠𝑒
𝐼𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 =
𝑉𝑠𝑒𝑐 𝐿−𝐿
Ipri rated All applications:
√3 × 𝑉𝐴𝑝ℎ𝑎𝑠𝑒
𝐼𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 =
𝑉𝑝𝑟𝑖 𝐿−𝐿
Note: For a bank of three single phase transformers, the following
calculations should be performed independently for each transformer
and then summed to obtain the total losses.
LWFe Take directly from power transformer test report
Loss compensation 124
Item Equation
LWCu Take directly from power transformer test report
LVAFe %𝐸𝑋𝐶
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
LVACu %𝑍
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
LVFe √𝐿𝑉𝐴𝐹𝑒 2 − 𝐿𝑊𝐹𝑒 2
LVCu √𝐿𝑉𝐴𝐶𝑢 2 − 𝐿𝑊𝐶𝑢2
Step 2.
If it is desired to compensate for line losses then calculate the full load
watt line loss and the full load var line loss values (see next section for
details on line loss calculation).
Parameter Description
LiWTOT Total full load watt line loss (line loss watt)
LiVTOT Total full load var line loss (line loss var)
Step 3.
Calculate the per element % correction factors, the meter voltage, and the
meter current. These are the values that must be entered into the loss
compensation software to configure the meter properly.
If the meter is on the primary side of the power transformer, then
Vrated = Vpri rated and Irated = Ipri rated.
If the meter is on the secondary side of the power transformer, then
Vrated = Vsec rated and Irated = Isec rated.
Parameter Equation
%LWFe 𝐿𝑊𝐹𝑒 × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
%LWCu (𝐿𝑊𝐶𝑢 + 𝐿𝑖𝑊𝑇𝑂𝑇 ) × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
%LVFe 𝐿𝑉𝐹𝑒 × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
%LVCu (𝐿𝑉𝐶𝑢 + 𝐿𝑖𝑉𝑇𝑂𝑇 ) × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
Meter current 𝐼𝑟𝑎𝑡𝑒𝑑
𝐶𝑇𝑅
Meter voltage 𝑉𝑟𝑎𝑡𝑒𝑑
𝑉𝑇𝑅
Loss compensation 125
Step 1.
Calculate line resistance and line reactance
The equations below should be applied individually to the primary and the
secondary conductors.
Parameter Description
RL Line resistance ()
XL Line reactance ()
Deq Geometric mean distance between phase conductors (in
meters)
DL1,L2 Distance between Line 1 and Line 2 (in meters)
DL2,L3 Distance between Line 2 and Line 3 (in meters)
DL3,L1 Distance between Line 3 and Line 1 (in meters)
Parameter Equation
RL
Item Equation
XL If using GMR:
𝑓 𝐷𝑒𝑞
𝐿 × 0.2794 × ( ) × log ( )
60 𝐺𝑀𝑅
If using Xa:
𝑓
𝐿 × {𝑋𝑎 + [0.2794 × ( ) × log 𝐷𝑒𝑞 ]}
60
where:
𝐷𝑒𝑞 = 3√𝐷𝐿1,𝐿2 × 𝐷𝐿2,𝐿3 × 𝐷𝐿3,𝐿1
Step 2.
Calculate the line losses.
Item Description
LiWTOT Total full load watt line loss (line loss watt)
LiVTOT Total full load var line loss (line loss var)
Vpri L-L Primary line-to-line voltage of power transformer
Vsec L-L Secondary line-to-line voltage of power transformer
Ipri rated Rated primary current of power transformer
Isec rated Rated secondary current of power transformer
Note: Vpri L-L, Vsec L-L, Ipri rated, and Isec rated are the same values as used in
calculation of transformer losses (see previous section).
When compensating for both transformer and line losses:
Item Equation
2
LiWsec 𝐼𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 × 𝑅𝐿 𝑠𝑒𝑐 × 𝑛
2
LiVsec 𝐼𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 × 𝑋𝐿 𝑠𝑒𝑐 × 𝑛
2
LiWpri 𝐼𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 × 𝑅𝐿 𝑝𝑟𝑖 × 𝑛
2
LiVpri 𝐼𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 × 𝑋𝐿 𝑝𝑟𝑖 × 𝑛
Note: In the special case that you are compensating only for line loss
(no transformer losses), then the values for Ipri rated and Isec rated must be
directly specified by the user. Typically, these two values will be
inversely proportional to the rated secondary and primary voltages of
the power transformer. That is,
𝐼𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑 𝑉𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑
=
𝐼𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 𝑉𝑝𝑟𝑖 𝑟𝑎𝑡𝑒𝑑
Loss compensation 127
Step 3.
If compensating for both transformer and line losses, return to Step 3 of
the previous section using the above calculated line losses to help
calculate the %LWCu and %LVCu values.
If compensating only for line losses use the following equations to
calculate the per element % correction factors, the meter voltage and the
meter current for entry in the loss compensation software.
If the meter is on the primary side of the power transformer,
Irated = Ipri rated.
If the meter is on the secondary side of the power transformer,
Irated = Isec rated.
Vrated is the nominal voltage seen on the high side of the instrument
transformer supplying voltage to the meter.
Parameter Equation
%LWFe 0
%LWCu 𝐿𝑖𝑊𝑇𝑂𝑇 × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
%LVFe 0
%LVCu 𝐿𝑖𝑉𝑇𝑂𝑇 × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
Meter current 𝐼𝑟𝑎𝑡𝑒𝑑
𝐶𝑇𝑅
Meter voltage 𝑉𝑟𝑎𝑡𝑒𝑑
𝑉𝑇𝑅
Loss compensation 128
Calculation example
The following example can be used as a guideline. This is based on the
sample transformer data for loss compensation shown in chapter 10 of the
Handbook for Electricity Metering (10th edition). 2
Application notes:
The application is a bank of three single-phase power transformers.
The metering occurs on the low (secondary) side of a power
transformer, and losses will be added to the measured energy.
There is a delta connection on the secondary of the power transformer
and thus a 2-element meter will be used to measure the service.
Losses are being compensated for the power transformer only (no line
losses).
Meter data:
Parameter Value
Elements 2
2 Edison Electric Institute, Handbook for Electricity Metering, 10th edition, Washington, DC: Edison Electric Institute, 2002,
Chapter 10, “Special Metering,” pp. 249-88.
Loss compensation 129
Step 1.
Calculate the quantities
Because the metering is on the secondary side of the power transformer,
all references to rated voltage and rated current refer to the secondary
rated values.
Parameter Description
VAphase Bank of three transformers:
𝑉𝐴𝑝ℎ𝑎𝑠𝑒 = 𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000
3333 × 1000 = 3,333,000
Vrated Secondary side, 2-element delta application
𝑉𝑠𝑒𝑐 𝐿−𝐿 = 2520
Irated Secondary side application
√3 × 𝑉𝐴𝑝ℎ𝑎𝑠𝑒
𝐼𝑠𝑒𝑐 𝑟𝑎𝑡𝑒𝑑 =
𝑉𝑠𝑒𝑐 𝐿−𝐿
√3 × 3,333,000
= 2290.84
2520
Line 1
Calculations
Parameter Value
LWFe 9650
LWCu 18935
LVAFe %𝐸𝑋𝐶
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
1.00
3333 × 1000 × ( ) = 33,330
100
LVACu %𝑍
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
8.16
3333 × 1000 × ( ) = 271,973
100
LVFe √𝐿𝑉𝐴𝐹𝑒 2 − 𝐿𝑊𝐹𝑒 2
√33,3302 − 96502 = 31,902
LVCu √𝐿𝑉𝐴𝐶𝑢 2 − 𝐿𝑊𝐶𝑢2
√271,9232 − 18,9352 = 271,313
Loss compensation 130
Line 2
Calculations
Parameter Value
LWFe 9690
LWCu 18400
LVAFe %𝐸𝑋𝐶
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
1.06
3333 × 1000 × ( ) = 35,330
100
LVACu %𝑍
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
8.03
3333 × 1000 × ( ) = 267,640
100
LVFe √𝐿𝑉𝐴𝐹𝑒 2 − 𝐿𝑊𝐹𝑒 2
√35,3302 − 96902 = 33,975
LVCu √𝐿𝑉𝐴𝐶𝑢 2 − 𝐿𝑊𝐶𝑢2
√267,6402 − 18,4002 = 267,007
Line 3
Calculations
Parameter Value
LWFe 9340
LWCu 18692
LVAFe %𝐸𝑋𝐶
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
0.91
3333 × 1000 × ( ) = 30,330
100
LVACu %𝑍
𝑘𝑉𝐴𝑟𝑎𝑡𝑒𝑑 × 1000 × ( )
100
8.12
3333 × 1000 × ( ) = 270,640
100
LVFe √𝐿𝑉𝐴𝐹𝑒 2 − 𝐿𝑊𝐹𝑒 2
√30,3302 − 93402 = 28,856
LVCu √𝐿𝑉𝐴𝐶𝑢 2 − 𝐿𝑊𝐶𝑢2
√270,6402 − 18,6922 = 269,993
Loss compensation 131
Step 2
Compensate for line losses, if needed.
Per the stated assumptions, there is no compensating for line losses:
LiWTOT = 0
LivTOT = 0
Step 3
Now the per element % correction factors may be calculated:
Parameter Value
%LWFe 𝐿𝑊𝐹𝑒 × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
28,680 × 100
= 0.2484
2520 × 2290.84 × 2
%LWCu (𝐿𝑊𝐶𝑢 + 𝐿𝑖𝑊𝑇𝑂𝑇 ) × 100
𝑉𝑟𝑎𝑡𝑒𝑑 × 𝐼𝑟𝑎𝑡𝑒𝑑 × 𝐸𝑙𝑒𝑚𝑒𝑛𝑡𝑠
56,027 × 100
= 0.4853
2520 × 2290.84 × 2
For a 2-element meter, watts and vars are compensated every two line
cycles according to the following equations:
Compensation Equation
2 2 2 2
W 𝑅 × (𝐼𝐿1 𝑚𝑒𝑎𝑠 + 𝐼𝐿3 𝑚𝑒𝑎𝑠 ) + 𝐺 × (𝑉𝐿1 𝑚𝑒𝑎𝑠 + 𝑉𝐿3 𝑚𝑒𝑎𝑠 )
2 2 4 4
var 𝑋 × (𝐼𝐿1 𝑚𝑒𝑎𝑠 + 𝐼𝐿3 𝑚𝑒𝑎𝑠 ) + 𝐵 × (𝑉𝐿1 𝑚𝑒𝑎𝑠 + 𝑉𝐿3 𝑚𝑒𝑎𝑠 )
where:
Term Description
R Per element resistance
G Per element conductance
X Per element reactance
B Per element susceptance
Ix meas Per phase rms current
Vx meas Per phase rms voltage
load compensation terms for line 3 will be in error because they will be
calculated based on one-half the line 1 voltage.
The same situation would result if line 1 experiences a loss of voltage.
button. The push button that activates the alternate mode. It also can
be used to control the scrolling of display quantities in the different
operating modes.
Alpha Keys. A system combining hardware and software to upgrade
existing A1800 ALPHA meters. Keys allow addition of new functionality to
an existing meter for an additional fee.
alternate mode. The operating mode in A1800 ALPHA meters used to
display a second set of display quantities on the LCD. It is generally
activated by pressing the Q button on the meter. A typical use of the
alternate mode is to display non-billing data as programmed by Elster
meter support software.
AvgPF. see average power factor.
average power factor. Calculated once every second, when the meter is
not in test mode, using the following formula:
𝑘𝑊ℎ
𝐴𝑣𝑔𝑃𝐹 =
√𝑘𝑣𝑎𝑟ℎ2 + 𝑘𝑊ℎ2
billing data. The measured quantities recorded and stored by the meter for
use in billing the consumer. May also be referred to as tariff data.
bit. Short for binary digit. It is the smallest information unit used in data
communications and storage.
coincident. Information regarding one parameter occurring at the same
time as another. For example, coincident kvar demand is the kvar demand
occurring during the interval of peak kW demand.
communication session count. The number of data-altering
communications occurring since the A1800 ALPHA meter was last
programmed or a clear of the values and status.
complete LCD test. A display showing 8 in all the display areas and all
identifiers on the LCD turned on. This confirms that all segments are
operating properly.
continuous cumulative. A display technique used with demand
calculations and similar to cumulative demand except continuous
cumulative demand is updated constantly.
CTR. see current transformer ratio.
cumulative. A display technique used with demand calculations. Upon a
demand reset, the present maximum demand is added to the sum of the
previous maximum billing period demand values.
current transformer ratio. The ratio of the primary current to the secondary
current of a current transformer. For example, 400 A to 5 A would have a
current transformer ratio of 400:5 or 80:1.
Glossary 136
external display multiplier. Used when the transformer factor is larger than
can be stored within the A1800 ALPHA meter. When programmed with
Elster Metering meter support software for an external display multiplier,
display quantities read from the meter LCD must be manually multiplied by
this value to yield proper readings.
factory default. Operating parameters that are programmed into the meter
at the factory and assure that the meter is ready for correct energy
measurement when installed.
four quadrant metering. See Figure A-1 for an illustration of energy
relationships for delivered and received real power (kW), apparent power
(kVA), and reactive power (kVAR).
Figure A-1. Four quadrant metering quantity relationships
kvar Delivered
Lag Lag
kVA Delivered
kVA Received
kW Delivered
kW Received
Q2
Q2 Q1
Q1
Q3
Q3 Q4
Q4
Lead Lead
kvar Received
pulse relay. A relay used with the meter to provide output pulses from the
meter to an external pulse collector. Each pulse represents a specific
amount of energy consumption.
rec. see received.
received. Used to specify the energy received by the utility at an electric
service.
recurring dates. Holidays or other special dates that occur on a predictable
basis.
self read. The capturing of current billing data and storing it in memory.
Self reads are scheduled events that can be triggered by the specific day of
month, every set number of days, or command by Elster meter support
software. See also previous billing data.
tariff data. See billing data.
TOU. see time-of-use.
TOU meter. A meter that records energy usage and demand data on a
time-of-use basis.
test mode. The test mode stores billing data in a secure memory location
while the meter measures and displays energy and demand data for
testing purposes. The TEST identifier will flash while the test mode is
active. When test mode is exited, the accumulated test data is discarded
and the original billing data is restored.
timekeeping. The ability of the meter to keep a real time clock, including
date and time.
time-of-use. A billing rate that records energy usage and demand data
related to specific times during the day. See also timekeeping.
transformer-rated. A meter designed to work with current or voltage
transformers. The maximum current of a transformer-rated A1800 ALPHA
meter is typically 20 A.
voltage transformer ratio. The ratio of primary voltage to secondary voltage
of a transformer. For example, 12,000 V to 120 V would have a voltage
transformer ratio of 100:1.
VTR. see voltage transformer ratio.
watthour constant. A meter constant representing the watthours per
output pulse on the LED. Historically, the constant represents the energy
equivalent to one revolution of an electromechanical meter.
Glossary 140
Notes:
B DISPLAY
Display format
Displayable items are described in “Display list items” on page 143. The
A1800 ALPHA meter supports up to 64 quantities for display on the LCD.
The LCD can be divided into different regions, as described in Table B-1.
See “Indicators and controls” in chapter 3 for more detailed information on
the LCD regions.
Figure B-1. A1800 ALPHA meter LCD
Quantity identifier
Low battery indicator
Phase Alternate mode
Error/warning indicator
indicators (3) indicator
+ Q
-P +P
Energy direction - Q
L1L2 L3 Comm. port
indicator indicator
COM 0 1 2
Display quantity
Power/energy
units identifier
Item Description
Quantity identifier Identifies the displayed quantity. Using Elster
meter support software, an identifier can be
assigned to most quantities. For instrumentation
quantities, the identifiers are fixed.
Alternate display Indicates that the meter is currently displaying
indicator items in the alternate display list (see “ button”
in chapter 3)
Active COM port Indicates that a communication session is in
indicators progress and the communication port that is
being used: either COM 0, COM 1, or COM 2
Power/energy units Indicates the unit of measurement for the
identifier quantity currently displayed on the LCD.
Display 142
Item Description
Display indicators Indicates whether the meter is currently doing
the following:
accumulating in tariff (T1 - T4)
has reached the end of an interval (EOI)
compensating for transformer line loss (LC)
indicating that either the terminal cover or
the meter cover has been removed
is operating in test mode (see “Test mode”
in chapter 3)
Display quantity Shows metered quantities or other displayable
information. From 3 to 8 total digits with up to 9
decimal places can be used. These digits are
also used to report the following:
operational errors
system instrumentation and service test
errors
warnings
communication codes
Display identifiers More precisely identifies the information
presented on the LCD.
Energy direction Indicates the directions of active (P) and reactive
indicators (Q) energy flow (positive energy flow is energy
delivered to the consumer load, while reverse
energy flow is energy received from the
consumer load)
Error indicator Indicates either of the following:
flashes when any error flag is set
remains on if a displayable warning flag is
set and no error exists
Low battery If the indicator is turned on, the battery warning
indicator flag has been set.
Phase indicators L1, L2, and L3 (Line 1, Line 2, and Line 3,
respectively) correspond to a phase voltage
present on the A1800 ALPHA meter connections.
If the indicators are on, then all phase
voltages are present.
If an indicator is blinking, then that phase
voltage is either missing or below the
defined threshold for voltage sag detection.
Display 143
Character Represents
Blank (space)
. Decimal
- hyphen; represents testing in progress
* asterisk; represents all 16 character segments on
: colon; separates time units (hh:mm), etc.
a Any alphanumeric character displayable on the LCD.
dd Numeric character; represents day (01 to 31)
Display 144
Character Represents
H Indicates the day type is holiday
hh Numeric character; represents time in hours (01 to 24)
mm Numeric character; represents time in minutes (00 to 59)
MM Numeric character; represents month (01 to 12)
x Any numeric character.
ss Numeric character; represents time in seconds (00 to 59)
YY Numeric character; represents two digit year (00 to 99)
LCD test
The A1800 ALPHA meter tests the LCD by displaying all the identifiers, as
shown in Figure B-2. The meter tests the LCD for 3 seconds after power up.
Figure B-2. LCD all segment test
+ Q
-P +P
- Q
L1L2 L3
COM 0 1 2
Meter configuration
Display description Display quantity Quantity ID Units ID
Program ID xxxxxxxx PRG ID
Pulse ratio (P/R) x.xxxxxxx P/R
Pulse output ratio [imp/kWh] xxxxx.xxx imp/kWh
Current transformer (CT) ratio xxxxxxxx CT
Voltage transformer (VT) ratio xxxxxxxx VT
Demand interval - normal mode xxxxxxxx INTERV
Demand interval - test mode xxxxxxxx INTERVT
Watthours per pulse (Ke) xxxxxxxx Wh/Imp
Meter Kh xxxxxxxx Kh
Transformer factor (CT × VT) xxxxxxxx CTxVT
External multiplier xxx.xxxxx ExtMult
Demand overload value xxxxx.xxx DmdOvld
Status
Display description Display quantity Quantity ID Units ID
Communication session count (port 1) xxxxxxxx Com1No
Communication session count (port 2/optical) xxxxxxxx Com2No
Days since demand reset xxxxxxxx ResDays
Days since input pulse xxxxxxxx ImpDays
Number of manual demand resets xxxxxxxx RstPress
Number of all demand resets xxxxxxxx DmdRes
Power outage count xxxxxxxx Outages
Initial remote baud (port 1) xxxxxxxx COM1bps
Initial remote baud (port 2) xxxxxxxx COM2bps
Transformer Loss Comp Status xxxxxxxx
TRueQ Status (On/Off) On
Off
Outage Log
Program Change Date (port 1) MM:dd:YY
Display 146
Metered quantities
A1800 ALPHA meters can measure two quantities. Meters with the
optional 4-quadrant metering can measure eight quantities. The A1800
ALPHA meter can display the available metered quantities for each meter
type.
To indicate a self read quantity, the LCD will use the last two characters of
the quantity identifier to indicate the last self read number (01 to 35).
Display description Display ID Display quantity Quantity ID Units ID
Current billing, Previous billing, Previous season, Last self read
Total energy xxxxxxxx Deliver kWh/kVAh/kvarh
Receive
Q1
Q2
Q3
Q4
Display 147
Cumulative demand
The A1800 ALPHA meter records either the cumulative or continuous
cumulative demand.
Display description Display ID Display quantity Quantity ID Units ID
Cumulative demand xxxxxxxx CumDmd kW/kVA/kvar
Tariff 1 cumulative demand T1 xxxxxxxx CumDmd kW/kVA/kvar
Tariff 2 cumulative demand T2 xxxxxxxx CumDmd kW/kVA/kvar
Tariff 3 cumulative demand T3 xxxxxxxx CumDmd kW/kVA/kvar
Tariff 4 cumulative demand T4 xxxxxxxx CumDmd kW/kVA/kvar
Display 150
System instrumentation
The A1800 ALPHA meter can display system instrumentation quantities.
See “System instrumentation” in chapter 4 for a listing of the
instrumentation quantities that can be displayed.
Display description Display quantity Quantity ID Units ID
Line frequency xx.xxHz L123
Line 1 voltage (secondary) xxx.xxx V L1
Line 1 voltage (primary) xxx.xxxkV
Line 2 voltage (secondary) xxx.xxx V L2
Line 2 voltage (primary) xxx.xxxkV
Line 3 voltage (secondary) xxx.xxx V L3
Line 3 voltage (primary) xxx.xxxkV
Line 1 current (secondary) xxx.xxx A L1
Line 1 current (primary) xxx.xxxkA
Line 2 current (secondary) xxx.xxx A L2
Line 2 current (primary) xxx.xxxkA
Line 3 current (secondary) xxx.xxx A L3
Line 3 current (primary) xxx.xxxkA
Line 1 power factor xx.xx L1 COS
Line 2 power factor xx.xx L2 COS
Line 3 power factor xx.xx L3 COS
Line 1 power factor angle xxx.xx° L1
Line 2 power factor angle xxx.xx° L2
Line 3 power factor angle xxx.xx° L3
Line 1 voltage phase angle xxx.xx°V L1
Line 2 voltage phase angle xxx.xx°V L2
Line 3 voltage phase angle xxx.xx°V L3
Line 1 current phase angle xxx.x°A L1
Line 2 current phase angle xxx.x°A L2
Line 3 current phase angle xxx.x°A L3
Line 1 kW (primary) xxxx.xxxx L1 kW
Line 1 kW (secondary) xxx.xxx MW
Line 2 kW (primary) xxxx.xxxx L2 kW
Line 2 kW (secondary) xxx.xxx MW
Line 3 kW (primary) xxxx.xxxx L3 kW
Line 3 kW (secondary) xxx.xxx MW
Line 1 kvar (primary) xxxx.xxxx L1 Kvar
Line 1 kvar (secondary) xxx.xxx Mvar
Line 2 kvar (primary) xxxx.xxxx L2 Kvar
Line 2 kvar (secondary) xxx.xxx Mvar
Line 3 kvar (primary) xxxx.xxxx L3 Kvar
Line 3 kvar (secondary) xxx.xxx Mvar
Display 151
Communication codes
The A1800 ALPHA meter indicates the status of a communication session
by displaying it on the LCD. See “Communication codes” in chapter 6.
Display 154
Notes:
C NAMEPLATE & STYLE
NUMBER INFORMATION
Nameplate
The nameplate provides important information about the meter. The
nameplate can be configured to meet the needs of the utility company;
however, Figure C-1 is an illustration of an A1800 ALPHA nameplate for
transformer rated meters.
Figure C-1. Sample nameplate
T1 T2 T3 T4 T5 T6 T7 T8 EOI LC TC TST
LED pulse
set ings
Meter type 5,000 imp/kWh
TYPE A1800 5,000 imp/kVarh
Meter model MODEL 1830RALN s150
Number of elements
Voltage rating Accuracy Class 0.2
and frequency 3 x 58/100...240/415V, 60 Hz
0.2S
1(10)A
Nominal (max) 2006 Made in USA Isolation Class 2
current symbol
ELSTER SAMPLE
SERIAL # 01234567 Year and place
of manufacture
STYLE
Elster style number ZE3JF4R00LM-01 *01234567*
R1.2-06210102AG-000000
U ility symbol
Firmware version Meter serial number
and barcode
Code Description
R Meter release
01 Firmware version number
00 Firmware revision number
YYWW Manufacturing date code (year and week)
DD Meter engine code set version
XX Slot 1 option board firmware version, if installed (numbers
are omitted from the nameplate if no option board is
installed)
ZZ Slot 2 option board firmware version, if installed (numbers
are omitted from the nameplate if no option board is
installed)
AAAAAA Manufacturer order number
Nameplate & style number information 156
SERVICE
Symmetrical wiring J
Sequential wiring K
Sequential wiring T
Sequential wiring N
COMMUNICATION OPTIONS
1 Only for non-standard instances of transformer-rated meters where Imax is greater than 10 A. All other transformer-rated meters
are to use 10 A as the standard selection.
2 Voltage link options available for direct connect-rated meters only.
3 Select this option for direct connected-rated meters without voltage links.
4 Contact Elster for availability.
Nameplate & style number information 158
FIELD NUMBERS 1 2 3 4 5 6 7 8 9 10 11
MISCELLANEOUS OPTIONS
METERING OPTIONS
FIELD NUMBERS 1 2 3 4 5 6 7 8 9 10 11
SERVICE
MISCELLANEOUS OPTIONS
METERING OPTIONS
5 Additional virtual option boards available with these models. Available: DL/T 645.
6 4-wire RS-485 implementation
7 2-wire RS-485 implementation
Nameplate & style number information 159
FIELD NUMBERS 1 2 3 4 5 6 7 8 9 10 11
SERVICE
METERING OPTIONS
Watt, var, 4-quadrant metering, loss compensation, load profiling, instrumentation profiling, TRueQ LP
Nameplate & style number information 160
Notes:
D WIRING DIAGRAMS
See the wiring diagram on the nameplate of each meter for specific
terminal assignments. All connections are equipped with combination-head
screws that accept either a slotted or Phillips screwdriver.
Direct connected
Figure D-1. 3-element, 4-wire wye or delta
L3 L1
L2
1 3 4 6 7 9 10 12
L1 L
L2 O
L3 A
N D
DC001.vsd
L1 L3
L2
or
L3 L1
N
1 3 4 6 7 9
L
L1
O
L2 or N
A
L3
DC002.vsd
D
Wiring diagrams 162
CT-connected meters
Figure D-3. 2-element, 3-wire delta, transformer connected
1 3
1 2 3 5 7 8 9
L 1 L
I O
2
N A
E 3 D
CT001.vsd
Figure D-4. 2-element, 3-wire sequential, transformer connected meter used in single phase transformer application
1 3
1 2 3 5 7 8 9
L 1 L
I O
2
N CT006.vsd
A
E D
Wiring diagrams 163
1 3
1 2 3 4 5 6 7 8 9 11
1
L 2 L
I O
N3 A
EN D
CT002.vsd
Figure D-6. 3-element, 4-wire current transformer, sequential connection meter used in single phase application
1 3
1 2 3 4 5 6 7 8 9 11
1
L L
I O
N A
EN D
CT007.vsd
Wiring diagrams 164
1 3
1 2 3 4 5 6 7 8 9 11
1
L 2 L
I O
N 3 A
EN D
CT003.vsd
1 3
1 2 3 4 5 6 7 8 9 11
1
L
L
2 O
I
N3 A
E D
N
CT004.vsd
Wiring diagrams 165
Figure D-9. 3-element, 4-wire instrument transformer, sequential connection meter used in a single phase application
1 3
1 2 3 4 5 6 7 8 9 11
1
L
L
I O
N A
E D
N
CT008.vsd
1 3
1 2 3 4 5 6 7 8 9 11
1
L L
I 2 O
N 3 A
E D
N
CT005.vsd
Wiring diagrams 166
Notes:
E TECHNICAL
SPECIFICATIONS
Absolute maximums
Voltage Continuous 528 VAC
Surge voltage withstand Test performed Results
Oscillatory (IEC 61000-4-12) 2.5 kV, 60 seconds
Fast transient (IEC 61000-4-4) 4 kV
Impulse voltage test (IEC 60060-1) 12 kV @ 1.2/50 µs ≥ 450 Ω (8 kV with
option boards)
AC voltage (insulation) test 4 kV, 50 Hz for 1 minute
Current Continuous at Imax
Temporary (0.5 seconds) at 2000 % of Imax (transformer rated)
½ cycle at 30 × Imax (direct connect rated)
Operating ranges
Voltage
Nameplate nominal range 58 V to 415 V
Operating range 49 V to 528 V
Auxiliary power supply range For AC power: 57 V rms to 240 V rms (nominal)
For DC power: 80 V to 340 V (nominal)
Current 0 A to 10 A (transformer rated)
0 A to 120 A (direct connect rated)
Frequency Nominal 50 Hz or 60 Hz ± 5 %
Temperature -40 °C to +85 °C inside meter cover
-40 °C to +60 °C outside
Humidity range 0 % to 100 % noncondensing
Technical specifications 168
Operating characteristics
Power supply burden Less than 3 W
Per phase current burden Less than 0.1 VA (transformer rated & direct connect rated)
Per phase voltage burden1 0.008 W at 120 V 0.03 W at 240 V 0.04 W at 480 V
Accuracy
Active energy 0.2 % (IEC 62053-22) 0.5 % (IEC 62053-22) 1.0 % (IEC 62053-21)
Reactive energy 2.0 % (IEC 62053-23) Actual accuracy is better than 0.5 % for 2.0 %
accuracy meters
1 Conforms to IEC 62053-61 (Electricity Metering Equipment, Power Consumption and Voltage Requirements)
Elster Solutions © 2015 by Elster. All rights reserved.
208 S Rogers Lane Information herein is subject to change without notice.
Raleigh, NC 27610-2144 Product specifications cited may change. Contact your
Elster representative for the most current product
information.