Xray With KV and Mas
Xray With KV and Mas
Xray With KV and Mas
2018
Distance in cm cm
50 5 49.75 kVp 1.223 mGy
50 8 50.00 kVp 1.977 mGy
50 10 49.78 kVp 2.459 mGy
60 5 59.90 kVp 1.796 mGy
60 8
60 10
70 5
70 8
70 10
90 5
50 5
50 5
50 5
50 5
50 8
50 8
50 10
50 10 49.83 kVp 2.461 mGy
60 5 59.84 kVp 1.788 mGy
60 5 59.78 kVp 1.796 mGy
60 5 59.92 kVp 1.795 mGy
60 5 59.91 kVp 1.795 mGy
Tube Leakage Test
To get time value with R/F
90 8
detector
90 8 Tube (Left) 0.001 mR 0.155 mR/h
Collimator (Left) 0.000 mR 0.036 mR/h
Tube (Right)
Collimator (Right)
Tube (Front) 0.000 mR 0.055 mR/h
Collimator (Front) 0.000 mR 0.113 mR/h
Tube (Back) 0.000 mR 0.084 mR/h
Collimator (Back) 0.000 mR 0.033 mR/h
Tube (Top) 0.000 mR 0.031 mR/h
Collimator (Top) 0.001 mR 0.185 mR/h
Area Surveillance
kVp mAs
80 8 To get time value with R/F detector
Work load 0
0 8
0 8
0 8
0 8
0 8
0 8
0 8
0 8
0 8
0 8
0 8
0 8
DOSE RATE TIMER PULSE Frame/pulse
mGy/min f/s
15-10-201 14:08:48
15-10-201 14:05:38
15-10-201 14:05:54
15-10-201 14:06:02
15-10-201 14:06:20
15-10-201 14:06:44
15-10-201 14:06:53
15-10-201 14:07:07
15-10-201 14:07:15
15-10-201 14:09:34
15-10-201 14:09:41
15-10-201 14:09:48
15-10-201 14:10:20
0 mmAl
0 mmAl
0 mmAl
2.43732 mmAl
2.44099 mmAl
2.43146 mmAl
2.53148 mmAl
0 mmAl
0 mmAl
0 mmAl
0 mmAl
0 mmAl
0 mmAl
0 mmAl
0 mmAl
2.46193 mmAl
2.4423 mmAl
2.44046 mmAl
2.43998 mmAl
[Periodic Quality Assurance shall be carried out at least once in two years and also
after any repairs having radiation safety implications]
GOVERNMENT OF INDIA
RADIOLOGICAL SAFETY DIVISION
ATOMIC ENERGY REGULATORY BOARD,
NIYAMAK BHAVAN, ANUSHAKTINAGAR
MUMBAI-400 094
Page 9 of 15
[Periodic Quality Assurance shall be carried out at least once in two years and also
after any repairs having radiation safety implications]
90 kV Collimator #DIV/0!
8 mAs mGy in 1 hr
I hereby undertake that all the information provided above is correct
and in accordance with the detailed Quality Assurance report enclosed herewith.
Date: Signature
Place: 0 Name of Service Engineer Mr. Siddhartha Sankar Maity
Name of Service Agency R K Consultant, Kolkata
Seal/Stamp of Service Agency
# Quality Assurance Tests Report shall be signed by Institution’s Representative and duly stamped by the User’s Institution
Page 10 of 15
[Periodic Quality Assurance shall be carried out at least once in two years and also
after any repairs having radiation safety implications]
Operating Parameters:
Operating Parameters:
FFD (cm) 60
Tolerance:
Focus Stated Value of focal Measured Value of focal 1. + 0.5 f for f < Remark
spot size (mm x mm) spot size (mm x mm) 0.8 mm
Select 1 2.0 2. + 0.4 f for 0.8 ≤ f ≤ 1.5 mm Select
Select 3 4.0 3. + 0.3 f for f > Select
1.5 mm
Page 11 of 15
[Periodic Quality Assurance shall be carried out at least once in two years and also
after any repairs having radiation safety implications]
4. ACCURACY OF OPERATING POTENTIAL & TOTAL FILTRATION
FFD (cm) 0
Measured kVp
Applied mAs mAs Station mAs Station Average Deviation
kVp Station 1 2 3 kVp ( kVp )
5 8 10
50 49.75 50.00 49.78 49.84 -0.16
60 59.90 0.00 0.00 19.97 -40.03
70 0.00 0.00 0.00 0.00 -70.00
90 #REF! -- -- #REF! #REF!
Tolerance: For kVp ± 5 kVp
Operating parameters
FFD (cm) 0 kV 50
mAs Output (mGy) mGy/
Range Average COL Remark
1 2 3 mAs
5 1.223 0.000 0.000 0.408 0.08150
8 1.977 0.000 0.000 0.659 0.08235 0.3360 Select
10 2.459 0.000 2.461 1.640 0.16400
Tolerance in COL : < 0.1
6. OUTPUT CONSISTENCY
FFD (cm) 0
Applied Output (mGy)
mAs Average COV Remark
kVp 1 2 3 4 5
50 5 1.223 0.000 0.000 0.000 0.000 0.245 2.236 Select
60 5 1.796 1.788 1.796 1.795 1.795 1.794 0.002 Select
Tolerance : COV < 0.05
Not Applicable
Page 12 of 15
[Periodic Quality Assurance shall be carried out at least once in two years and also
after any repairs having radiation safety implications]
8. HIGH CONTRAST SENSITIVITY
Not Applicable
Not Applicable
10. RADIATION LEAKAGE LEVELS FROM X-RAY TUBE HOUSING AT 1m FROM THE FOCUS
mGy in
Maximum radiation leakage mR in 1 hr 1 Result
hr
Tube housing #DIV/0! #DIV/0! Pass
Collimator #DIV/0! #DIV/0! Pass
Acceptable Tolerance Limit: Leakage Radiation level at 1 metre from the focus should be
≤ 114 mR in one hour
or ≤ 1 mGy in one hour
Page 13 of 15
[Periodic Quality Assurance shall be carried out at least once in two
years and also after any repairs having radiation safety implications]
Equipment setting
kVp 80 mAs 8 mA 8
Measurement:
Permissible Limit :
[Periodic Quality Assurance shall be carried out at least once in two
years and also after any repairs having radiation safety implications]