CONTENTS
Ⅲ.Technical Specifications........................................................................................................................................6
VII. Safety...................................................................................................................................................................... 19
VIII. Maintenance.........................................................................................................................................................20
IV.Troubleshooting.................................................................................................................................................. 21
Packing List..................................................................................................................................................................23
1
Respectful Users,
Firstly, we heartily acknowledge that you have bought Model. PLX100 High Frequency Mobile
X-ray Equipment. To guarantee that you can correctly operate the machine and give a full play to all
its functions to make it be of great aid to you, please carefully read the instructions manual before
using the machine.
1.Purpose
The equipment is a combined high frequency X-ray video camera that can be conveniently moved to wards,
ICU, operating rooms and etc for taking film.
2. Product Features
2.1 Two-button adjustment by kV and mAs, display by LCD and memory by parameters after power
failure.
2.2 The users can depend on themselves to revise the parameters of the human anatomy program that
is set with 50 parameters.
2.3 It is featured with protection for shortage of high voltage, over-flowing of tube current and output
over-loading as well as indication for failure.
2.4 The structural design of human engineering is simple and convenient for operation and overhaul.
3. Product Category
The equipment belongs to walking mode of type B under category 1. It can not be used under the
circumstance that has mixture of flammable anesthesia gas and air or mixture of O2 and N2O.
4. Working System
Continuous operation with intermittent loading.
5 .Environmental Conditions
5.1 Temperature:+10℃~+40℃.
5. 2 Relative Humidity:30%~75%.
5. 3 Atmospheric Pressure:700 hpa~1060hpa.
5.4 No corrosive gas and dust
6.Power Supply Conditions
6.1 Single-phase simple alternating current: non-sine distortion is less than 2% of its amplitude value.
6.2 Power Supply Voltage:220V±22V.
6.3 Power Supply Frequency: 50Hz±0.5Hz.
6.4 Power Supply Internal Resistance: no more than lΩ.
6.5 Input power : 6KVA
2
6.6 good grounded
1. Primary Structure
Figure 1
3
P ER L O N G PLX100 High Frequency Mobile X-ray Equipment
Vol/Fre 220V/50Hz T ype Class I.Type B.
InputPower 6kVA MaxOutput 100kV
Work Conti nuous Operati on With Intermittent Loading
Product No. Date Month Year
Figure 2
2.1.2 Refer to Figure 3 for the position of the nameplate of the equipment.
Figure 3
4
P ER LO NG
X-ray Monoblock
X-Ray Tube XD6C-0.55,4/125 Total Filtration ≥ 2.5mmAl
Focus 1.5 Max Output Vol 110kV
Product No Date Month Year
Nanjing Brolove Radial-Video Equipment Co.,Ltd.
Figure 4
5
3. Sign Description
4. Working Principles
4.1 High Voltage Circuit
The high voltage needed by X-ray tube is rectified from commutating current 220V to DC 300V which is
added to inverter, producing 40 kHz inverting frequency square wave. Then it is increased to the set
voltage value by passing the transformer in the X-ray generator’s components and the voltage doubling
circuit. Voltage regulation is conducted by the means of phase shifting control mode..
4.2 Filament Circuit
The voltage needed by X-ray tube for lighting up filament is provided by filament power panel. After the DC
24V is adjusted, DC voltage adjustable within a certain range is produced, then the 1kHz square wave
inverted from the DC voltage through the circuit is delivered to the filament transformer to energize the
filament after being isolate.
4.3 Control Circuit
X-ray parameter’s control over high voltage circuit and filament circuit is conducted by CPU through control
panel and display respectively, thus achieving control over kV and mAs. When 40~110kV is displayed, mAs
adopts digital closed-loop control. Besides, the control panel has additional functions of saving kV and mAs
parameters, state check and control output.
Ⅲ.Technical Specifications
1. X-ray Tube
Model XD6C-0.55,4/125 Maximum anode thermal 35KJ
capacity
Target angle 150 Bias between the focus <1mm
and datum axis
Max continuous power 230W Nominal anode power 4KW
6
input
Nominal value of the 1.5 Nominal X-ray tube 125KV
focus voltage
Target material Tungsten Equivalent filtration of ≥0.5mmAl
X-ray tube
Load characteristics Refer to figure 6 Filament emission Refer to figure 6
characteristics
Anode heating and Refer to figure 7
cooling characteristics
Uf(v) I a( m A )
7 70
6 60
100kV
5 50
50kV
4 40
3 30
2 20
1 10
Figure 6
7
Figure 7 Characteristic curve for heating and cooling of anode
2. X-ray Monoblock
2.1 Max continuous thermal dissipation: 60W
2.2 Max thermal capacity: 460kJ
2.3 Characteristic curve for heating and cooling: refer to figure 8
2.3 Datum loading conditions: 100kV 25mA
2.4 Stipulated max input energy for per hour: 800mAs
2.5 X-ray leakage radiation under the loading state: <1.0 mGy/h
2.6 Angle formed by anode target of the X-ray tube and the datum axis (through the perpendicular bisector
of the X-ray monoblock window.): 150
2.7 Tolerance of the focus center of the X-ray tube to the datum axis:≤ 1mm.
2.8 Total filtration: ≥ 2.5mmAl
3. Output Data of the Power
3.1 Max Output Power and X-ray Tube Voltage and X-ray Tube Current:
a) Max output power: 2.5 kW
8
b) X-ray tube voltage: 59kV
c) X-ray tube current: 42mA
d) Max rating capacity: 59kV、160mAs
3.2 Nominal Output Power and X-ray Tube Voltage and X-ray Tube Current:
Nominal output power: 2.5kW (100kV, 25mA, 0.1s).
3.3 Nominal tube voltage and the corresponding max tube current: 100kV, 25mA.
3.4 Nominal tube current and the corresponding max tube voltage: 50mA, 49kV.
4. Loading Factors
4.1 Combination of Loading Factors
4.2. 45 steps can be divided within 1.0~ 160 mAs,with the value for each step being shown as follows:
1.00 1.10 1.25 1.40 1.60 1.80 2.00 2.20 2.50 2.80 3.20 3.60 4.00 4.50 5.00 5.60
6.30 7.10 8.00 9.00 10.0 11.0 12.5 14.0 16.0 18.0. 20.0 22.0 25.0 28.0 32.0 36.0
40.0 45.0 50.0 56.0 63.0 71.0 80.0 90.0 100 110 125 140 160mAs
4.3 It is adjustable successively for 40kV~100kV, with 1kV of step length.
The front panel of the film cassette will attenuate the x-ray beam during exposure and its maximum
attenuation dosage is no more than 1.2mmAl. If there are other accessories being used, such as the
diagnostic bed placed between the X-ray Monoblock and the film, it will cause more attenuation; therefore,
the exposure parameters should be increased accordingly
5. Focus Center
The focus center of the X-ray tube monoblock keeps balance within the range of 550mm~1650mm. And the
machine is with secure braking within the range. The ascending pull should be less than 80N while the
descending pull should be less than 60N.
The horizontal rotation around the datum axis of the X-ray tube monoblock should be no more than ±900
while the vertical rotation around the datum axis should be no more than ±900 .
6.Distance between the Focus and Body (Focus-skin distance)
9
The range of the distance between focus and body: 500~1700 (mm)
Distance between focus and body is adjustable. The value reflected by pulling the focus-skin distance
meter (2M) to the body is used to indicate the distance between the focus and body.
Move the Monoblock to make a distance of 1 meter between the focus and body, then the x-ray field after
getting through the beam applicator is 340 * 340mm. Adjust the hand wheel of the beam applicator, it will
form random rectangles with areas not larger than 340*340mm. The longer the distance between the focus
and body, the larger the x-ray area formed. When taking normal radiography, keep a longer distance
between the focus of x-ray tube and the body of the patient, and meanwhile, move the patient as close as
possible to the film.
1. Dimension:120cm×62cm×130cm
2. Weight
Net Weight:150kg
1. General Description
Please first read carefully the operation instruction of the equipment before formally installing this unit. The
operation places for the operation, installation, and debugging of this unit must be equipped with the
stipulated power socket in accordance with the operation instruction. Meanwhile, there should be reliable
connecting location for protective grounding cables.
2. The Installation and Disassembly
2.1 X-ray Monoblock
When installing the unit, fix the X-ray monoblock on the locking mechanism and lock it (refer to Figure 11),
and clip the hose into the hooks at the side of the rocking arm. Open the protective cover of the 24-core
socket (refer to Figure 3), and then connect the 24-pin head to its socket (Note: the end with the hose is on
top), push it down and lock it by the side clips. The connection of the X-ray Monoblock and the equipment
finishes. When disassemble the unit, firstly, pull out and release the clips, and take off the 24-pin head, then
take the hose out of the hooks at the side of the rocking arm, and then loosen the locking mechanism and
take off the X-ray Monoblock, and at last, put it back into the packing box.
10
Figure 9
2.2 Beam Applicator
2.2.1 Disassembling and Installation
Under general circumstance, the position of the beam applicator can be referred to figure 10 when is
leaves the factory. If it is necessary to disassemble the beam applicator, the 2-M3 screw on the back cover
in figure 11 should first be removed. After removing the back cover, the fixed screw on the beam applicator
terminal of the beam applicator connection should be screwed off to take away the beam applicator
connection. Afterwards, anticlockwise rotate the four locking nuts on the two sides of the beam applicator in
figure 10. After all theses steps, the beam applicator can be taken away from the outer window of the X-ray
monoblock
When installing the unit, insert the plug of the power cable into the electricity – inputting socket. Please
note: the groove of the plug is upward and it corresponds to the plug, please insert the plug in the correct
way, or you cannot insert it in the socket. Then tighten the fixed nail-cover of the plug; insert the 3-pin plug
of the power cable into the net electricity source after all the parts have been well connected. When
disassembling the unit, pull out the power cable first to disconnect it with the net electricity source, then
screw off the fixed nail-cover and draw it out, and roll it up and put it back into the packing box.
When installing the unit, insert the plug of the hand-switch into the hand-switch socket. Note: the groove of
the plug is upward, please insert the plug in the correct way, or you cannot insert it into the socket. Then
tighten the fixed nail-cover. When knocking down the unit, loosen the fixed nail-cover and draw it out, roll it
When installing the unit, screw off the nut in the outboard of the grounding mark, and connect one end of
the grounding cable to it through the stud, then tighten the nut, and then connect the other end to the
protective ground of the room. When disassembling the unit, screw off the nut, take off the grounding cable
and tighten the nut, then separate the other end from the protective ground. At last, roll it up and put it into
When the X-ray monoblock is in the lowest position, gently push the down X-ray Monoblock with hand, and
turn the two chromized handles towards the direction of arrow at the same time. (refer to Figure12 and
12
figure 13) Then, the Monoblock will be elevated, moving freely up and down while keeping balance. (Note:
chromized handles serve as the switch of the safety mechanism. It is impossible to move the X-ray
monoblock if it is not elevated.) When the X-ray monoblock reaches low positions during its downward
movement, if a “kluck” sound is heard, the safety mechanism is locked by the chromized handles again. If
the movement is performed according to the above statements, the X-ray monoblock will recover to its
upward movement by pulling the chromized handles upward to the standard position with hands.
镀Chromized
铬把手 Handle
Figure 12 Position of the Chromized Handle Figure 13 Amplified figure for the chromized handle
3.2.1 The X-ray monoblock moves upward and downward together with the supporting arm. At random
position, the focus keeps balance, and the machine is with secure braking. When it is needed to fix at a
certain point, you can use the hand wheel in Figure 13 to tighten it. The unit can be moved freely and it
3.2.2 Loose the small handle in Figure 13 and the X-ray monoblock can make horizontal rotating around
3.3.3 The X-ray monoblock can make vertical rotating around the datum axis 90°
4.1 Explanation for the control panel, signs, and the functions of the keys (refer to Figure 15 and Figure
16).
13
Figure 14
Figure 15
4.2.1 After the mechanical parts and the X-ray monoblock is installed and adjusted, the power supply
4.2.2 Put the breaker (see figure 3) on “ON” location. Press for starting, then the screen of LCD
display will become bright, displaying the name of Najing Brolove Radial- Video Equipment Co., Ltd.
14
4.3 Test
After the machine is installed, before use for the first time or when the machine has been out of use for
more than 3 months, TEST must be conducted to clear away any possible gas existing in the X-ray tube.
Press down No.11 key , and the screen will display “Mode”. The normal mode displayed on the
screen is as follows:
b) Press No. 4 “up” key or No.5 “down” key corresponding to “mode” to choose TEST mode.
There are “Normal” and “Test” options of the mode, meaning normal mode and test mode correspondingly.
Use No. 4 “up” key or No.5 “down” key corresponding to choose the proper “mode”. After
selecting the test mode, press No.10 key to enter into the initial interface, Press down No.10 key
again to enter into TEST operation, and the words of “KV” and “TEST” will be displayed on the
screen as follows:
Choose certain kV value by No. 4 “up” key or No.5 “down” key corresponding to “KV”, then press
down No.10 key , and No.15 light will be on and the screen will be displayed as follows:
Press down hand switch, No.14 light will be on, indicating the machine start emitting X-ray. After the
15
The above procedures a), b), and c) consist one process of TEST. After per Test of 1 minute, the machine
needs 5 minutes to have a break. By No. 4 “up” key or No.5 “down” key corresponding to “KV”,
Gradually increase the kV value by 5 KV each step, and repeat operation procedure c) until the KV value
Before moving the equipment, remember to first place the X-ray monoblock in the lowest position and in
the state of self-locking. Then, push the equipment with great care to clinical position needed for
photography. Step on the pedal brake to fix the equipment. (Note: When the equipment is being moved,
please step on the pedal emphatically to make the front wheel rising if uneven floor, doorsill, elevator door
or other obstacles are confronted, this will enable the equipment to traverse the obstacles favorably..
Gently push down the X-ray Monoblock with hand, and turn the chromized handles towards the direction of
arrow at the same time (refer to Figure 13 and Figure 14), the X-ray monoblock will be elevated to the
needed height and can keep balance. Re-reminding: The chromized handles serve as the switch for the
safety mechanism, hence, the X-ray monoblock will not make upward movement if it is not elevated.
2.2 Adjust the X-ray monoblock and fix it at proper position. Place the X-ray film cassette at certain parts of
2.3 Put breaker on “ ON” location (see figure 3) .Connect the power supply: Press key , then the power
supply is on and the machine conducts self-checking. The screen displays as follows:
16
3. Parameter Selecting
3.1 Press down No.10 key to enter into the normal mode, the screen will show as follows:
3.2 According to clinical needs, you can choose kV value with No. 4 “up” key or No.5 “down” key
corresponding to “KV”, and mAs value with No. 6 “up” key or No.7“down” key corresponding to
“mAs”. You can also set some typical KV and mAs combination and store them in “No.” When they are
used, select them with No,.8 “page up” and No.9 “page down” corresponding to “No.”.
3.3 Beam applicator adjustment: Press the collimator lamp knob to turn on the Collimator Lamp in the
beam applicator, adjust the size of the “front-back” or “left-right” knob to obtain proper X-ray field in
4. Photography
4.1 Press down key to confirm. Store the selected “kV”, “mAs” to the corresponding “No.”. The
indicator light of in the operation panel will be on several seconds later, indicating that the machine
is ready.
4.2 Press down the remote control or the hand switch to emit X-rays, then the yellow indicator light of
in the operation panel lights on and buzzer sounds, the photography starts.
4.3 The preset time is over (or release the the remote control or hand switch) , and the yellow indicator light
goes out and the buzzer stops buzzing, the photography finishes.
5. Power off
Press key to shut down the equipment, Put air switch on ”OFF” location(see figure 3) .then disconnect
6. Notes
6.1 Keep a distance of about 10m from the machine when applying remote control to conduct photography.
Release the remote control (or manual control Switch) button if the radiograph has to be stopped in
17
midstream
6.2 The interval between two exposures: 6 minutes. (Explanation: If the time interval between two exposure
times is not enough, it may melt the anode target of the X-ray tube, even may burst the regulator and
damage the X-ray Monoblock, which will make the transformer oil ejecting from the Monoblock.)
6.3 When adjusting the beam applicator to get the required radiation area, 30 seconds after being
energized, Collimator Lamp will automatically switch off power. If illumination is needed, turn on the
Collimator Lamp again by pressing down the light switch in the beam applicator.
6.4 After pressing down the confirm key ,the indicator lamp will light on. If the hand switch
or the remote switch is not pressed down within 60 seconds to conduct the exposure, the machine will
automatically switch to the screen before confirmation. If the exposure is needed again, re-press down the
6.5 When applying manual control, the operator should take necessary measures to protect against
radiation: wear x-ray protective coats and take radiography in the significant zone of occupancy (see Figure
6.6 When the left, right or the right front parts are to be taken photos, do remember to loose the small handle
showed in the figure firstly in accordance with the Figure14. Afterwards, turn the X-ray monoblock to the
required position and lock the small handle showed in the figure. Under this circumstance, you can take
18
VII. Safety
1. General Description
The unit belongs to radiology unit and must only be used in compliance with the instructions indicated in this
manual and must not be used for other purpose than medical radiography. The unit must only be used by
personnel with the necessary knowledge in the field of radiation and with the necessary training in use of
x-ray units. Take special notes of the following items when put the unit into operation:
a) The unit must not be used when there are any electrical and mechanical faults.
b) Must be assembled with ground connection (the olive green one) when operating the unit.
c) Under no circumstances work in places saturated with vapours and/or flammable gases or explosives.
d) Any modification to the unit must be permitted by the manufacturer or its authorized service personnel.
e) Regular check and maintenance is necessary for the unit.
f) Operations must be performed according to the user’s manual. Do not modify the system without
permission.
2. Electrical Safety
2.1 Make sure the voltage, input power of the electric net and the inner resistance of the power supply that
the unit is to be connected complies with the foreseen power supply requirements.
2.2 Please pull out the power plug when cleaning
2.3 Power off and pull out the power plug after using.
3. Breaker
3.1 The instrument apply breaker to ensure electrical safety. See its location in figure 3.
3.2 Do not put the breaker on by force ,if the breaker break off twice automatically during using. The correct
method is : cut off the power supply immediately ,and then contact with us or specialist authorized by our
company.
4. Mechanical Safety
Warning: The alarming plate in the machine’s rocking arm specifies that:
a) Before mounting X-ray Monoblock, you are never allowed to move the chromized handles as shown in
Figure 13 or Figure 14, otherwise the rocking arm will rebound and cause injury
b) When there exists no X-ray monoblock on the rocking arm, you are never allowed to move the
chromized handles as shown in Figure 13 or Figure 14, nor to forcibly move the rocking arm upwards,
otherwise the rocking arm will rebound to cause machine damage and person injury
c) When disassembling the X-ray monoblock, make sure the rocking arm is in the lowest position and a
“click” from the chromized handle is heard. Meanwhile, it is impossible to move the X-ray monoblock
upward slightly. This indicates that the machine is in the self-locking state and the disassembling of the
5.2 It is advisable for the operators to apply exposure mode of “remote control” to carry out the radiography.
If the exposure mode of “manual control” is used, the operations should be carried out within the significant
5.3 When taking normal radiography, the focus-skin distance should be longer (at least more than 200mm),
6. Battery
The remote control of this machine requires an A23 battery. If the remote control will be left unused for a
long period of time, please remove the battery to prevent the battery liquid from damaging the remote
control.
VIII. Maintenance
1.Maintenance Requirements
Daily checking Check whether the signal, display and indicator light are normal;
Check whether the switch, remote control works normally.
20
Check whether the X-ray monoblock leaks oil;
Weekly checking Check whether there is abnormal sound when exposure of X-ray monoblock
occurs.
IV.Troubleshooting
supply;
Press power button, no power Power wire break off Check and repair the power wire
supply input Breaker is at OFF location Put the breaker at “ON” location
Electrical malfunction
If any abnormal things occur to the machine, turn off the machine and check the power supply, then, restart
21
Ⅹ。Transportation and Storage
1. Transportation
1.1 Transit and protection should be well prepared according to the icons and/or symbols on the package
If the machine is correctly operated and properly maintained, it can be used for a period of 8 years. If the
tube is broken, it can be sent back to our company for exchange with a new one. After the unit has served
for 8 years, it may break down and cannot be repaired any more, then, the unit has to be disposed properly.
The transformer oil in the x-ray generator has to be recycled by industrial oil recycling department; and the
other components made from steels and coppers have to be disposed by the designated departments.
1. Manufacturer’s Responsibility
The manufacturer is responsible for unit safety, reliability and performance related quality under the
following circumstances:
1. All the installation, adjustment, changes and maintenance are conducted by our staff or the staff
authorized by us.
2. The electric equipment used for this unit conforms to the technique requirements in this manual.
3. The operators are well-trained technicians and they strictly operate the unit according to the
requirements in this manual.
2. Three Guarantees
a) Under the circumstance that the users abide by the terms and conditions mentioned above, we will
repair the unit or replace the unit parts free of charge, even refund if the unit can not work properly
owing to quality reasons within one year from delivery.
b) If there is any damage resulted from ignorance of the above mentioned terms and conditions or the unit
has expired the one-year guarantee period from delivery, we will repair or change the unit parts and
accessories for users with moderate charge of the material and labor costs.
22
c) We are responsible for the regular maintenance of the unit with proper use within 8 years.
The electrical connection figure is attached at the end of the operation instruction for the user’s reference.
Note: After the purchase of the equipment, we ensure that we provide you with necessary
information for equipment repair, provided that we have confirmed your technicians can repair the
unit effectively.
Packing List
Packing List
Serial
Name of the parts Number Remark
No.
1 Main Frame 1
2 X-ray generator 1
3 Controller 1
4 Beam applicator 1
5 Instruction manual 1
23
(6) #8 Inner hexagon spanner 1
24