PyPhotometry Open Source Python Based Hardware and
PyPhotometry Open Source Python Based Hardware and
com/scientificreports
Fiber photometry is the process of recording bulk neural activity by measuring fluorescence changes in
activity sensitive indicators such as GCaMP through an optical fiber. We present a system of open source
hardware and software for fiber photometry data acquisition consisting of a compact, low cost, data
acquisition board built around the Micropython microcontroller, and a cross platform graphical user
interface (GUI) for controlling acquisition and visualising signals. The system can acquire two analog and
two digital signals, and control two external LEDs via built in LED drivers. Time-division multiplexed
illumination allows independent readout of fluorescence evoked by different excitation wavelengths
from a single photoreceiver signal. Validation experiments indicate this approach offers better signal
to noise for a given average excitation light intensity than sinusoidally-modulated illumination.
pyPhotometry is substantially cheaper than commercial hardware filling the same role, and we
anticipate, as an open source and comparatively simple tool, it will be easily adaptable and therefore of
broad interest to a wide range of users.
Fiber photometry has emerged as an important tool for behavioural neuroscience, which allows for the measure-
ment of neuronal activity from genetically-defined neuronal populations or axonal projections in superficial or
deep brain structures in behaving animals1–11. The rapid development of novel genetically encoded fluorescent
indicators for intracellular calcium12–16 and chloride17, neurotransmitter release18–20, and membrane voltage21–23
suggest that experimental applications for fiber photometry will continue to grow rapidly.
In fiber photometry experiments, excitation light of one or more wavelengths is transmitted to the brain
structure of interest through an optic fiber where it excites one or more fluorescent indicators. Light emitted by
the indicators is transmitted back through the optic fiber, separated from the illumination light and fluorescence
from other indicators using optical filters, and converted to electrical signals by high sensitivity photoreceivers
or cameras. These electrical signals are digitised and constitute the data generated by the experiment. Excitation
light of different wavelengths may be differentially modulated to allow the fluorescence evoked by each to be
demultiplexed from a single photodetector signal, for example to independently measure fluorescence evoked
from the genetically encoded calcium indicator GCaMP by 470 nm and 405 nm excitation light. As GCaMP is
approximately isosbestic at 405 nm (i.e. fluorescence is independent of calcium concentration) this gives a cal-
cium sensitive signal and a calcium insensitive signal that can be used to control for movement artefacts5.
Turn-key commercial systems exist for fiber photometry data acquisition that are convenient but expensive.
Alternatively acquisition can be controlled using generic hardware such as that from National Instruments, but
this requires a substantial investment of time to setup as well as proprietary hardware and software licences. We
sought to develop an open source acquisition system that offered the convenience of commercial hardware at low
cost. As well as democratising access to new experimental methods, open source hardware can help to improve
reproducibility as the entire signal acquisition and processing pipeline is open, and facilitate new applications as
researchers can modify tools themselves24–27.
pyPhotometry is a system of hardware and software consisting of an acquisition board and graphical user
interface (GUI). The system implements the following functionality: (1) Digitisation of 2 analog voltage signals
1
Department of Experimental Psychology, Oxford University, Oxford, UK. 2Champalimaud Neuroscience Program,
Champalimaud Center for the Unknown, Lisbon, Portugal. Correspondence and requests for materials should be
addressed to T.A. (email: [email protected])
(at 15-bit resolution) and 2 digital signals. (2) Two constant current LED driver circuits with a 0–100 mA output.
(3) Control of data acquisition and online visualisation of signals via the GUI. (4) Streaming of acquired data
to disk in a compact binary format. (5) Time-division multiplexed illumination to prevent crosstalk between
fluorescence signals and bleed-through of ambient light, with online demultiplexing and visualisation. (6) User
documentation at https://pyphotometry.readthedocs.io.
We report the system design and rationale, validation experiments characterising system performance, and
data showing its application to recording calcium signals from VTA dopamine neurons.
Results
Acquisition board. The acquisition board is built around the Micropython Pyboard, an ARM Cortex micro-
controller that is programmed in Python (Fig. 1). Programming the firmware in a high level language allows it to
be simple and compact (<200 lines of code), facilitating rapid development. The acquisition board has 2 analog
and 2 digital inputs, each a BNC connector.
Analog signals are acquired using the Pyboard’s analog to digital converters (ADCs). These are 12-bit ADCs
with a 0–3.3 V range. Oversampling is used to increase the resolution to 15-bit, i.e. for each sample the ADC is
read 64 times and the values averaged to give an extra 3 bits of resolution28,29. This gives a resolution of ~10−4 volts
per division of the digitised signal.
The digital inputs are read at the same sampling rate as the analog signals and their intended use is to enable
post hoc synchronisation of the photometry data with other systems such as behavioural hardware. We typically
use one digital input to receive synchronisation pulses from the behavioural hardware and the other to receive
a TTL each time a behavioural event of particular interest (e.g. reward delivery) occurs. This allows the average
response to the event of interest to be visualized online (see below), and the response to other behavioural events
to be calculated offline by using the sync pulse times to align behavioural and photometry data. We typically use
sync pulses with random inter-pulse intervals as this creates a unique and unambiguous match between pulse
times recorded on different systems (see https://pycontrol.readthedocs.io/en/latest/user-guide/synchronisation/).
The digital inputs connect directly to pins on the microcontroller so in principle could be used as outputs, e.g. for
triggering closed loop stimulation, though this is not currently supported by the firmware.
The LED driver circuits are voltage controlled current sinks implemented using an op-amp and MOSFET,
adapted from Figure 200 of 30. Their outputs are M8 connectors compatibles with commonly-used connector-
ised LEDs from Doric Lenses or Thorlabs. The LED driver circuit is linear over a 1–100 mA current range and
responds to control voltage transients in ~1 uS (see electrical characterisation).
The acquisition board draws power from the Pyboard’s USB connection which is also used to stream data to
the computer. Capacitors on the 5 V rail improve transient response and smooth the load presented to the power
supply, with a current limiting IC to restrict the inrush current when the board is powered up.
Assembly of the acquisition board requires only standard through-hole and surface-mount soldering tech-
niques. To make a complete photometry system the acquisition board must be paired with LEDs, photoreceivers,
filter cubes and other optical components. A complete parts list for the setup we use for green/red two colour
experiments (e.g. GCaMP/TdTomato) is provided in Table 1. Assembled acquisition boards can be purchased
from the Open Ephys store for €350 (http://www.open-ephys.org/store/pyphotometry).
Graphical user interface. The GUI is used to control data acquisition, visualise signals and record data to
disk (Fig. 2). It is written in Python and organised into separate modules for communication with the acquisition
board, layout of the GUI window, and plotting. The GUI is built using PyQt, a GUI programming toolkit which
enables rich GUIs to be implemented compactly- the complete GUI is ~600 lines of code. The GUI provides con-
trols for connecting to acquisition boards, setting acquisition parameters, LED currents, the data directory and
subject ID, and for starting and stopping data acquisition and recording.
The GUI window has three plots for displaying data implemented using PyQtGraph - a plotting library
designed for fast online interactive plotting in GUIs. An Analog signal plot displays a scrolling view of the two
analog input signals, and is the primary visualisation of the photometry data. A Digital signal plot displays a
scrolling view of the two digital input signals. An Event triggered plot shows a recency weighted event triggered
average of analog signal 1 triggered on rising edges of digital signal 1. This allows the average response to an event
of interest to be visualised online during data acquisition.
Data format. pyPhotometry can save data either as binary data files with a .ppd file extension or as comma
separated value files with a .csv file extension.
The binary file format is designed to be straightforward to import into analysis programs while generating
files that are no larger than necessary. The user guide provides detailed information on importing .ppd files, and
functions for importing .ppd files into Python and Matlab are provided in the tools folder of the code repository
(https://bitbucket.org/takam/pyphotometry/src/default/tools/). Files consist of header information and data. The
header is a JSON object that contains the subject ID, start date and time, and acquisition parameters. JSON is
a lightweight data interchange format designed to be readable by humans and easy for machines to parse, and
is supported by most programming languages31. The remainder of the data file contains the analog and digital
signals encoded as binary data. Each 2 byte chunk encodes one 15-bit analog signal sample and one digital signal
sample, resulting in very compact data files – e.g. a 1 hour recording at 130 Hz sampling rate results in a ~1.9 MB
file.
The comma separated value files generated by pyPhotometry are text files where each line contains one sample
from each of the two analog inputs and two digital inputs, separated by commas. When data is stored as a .csv file,
acquisition settings are saved as a JSON object in a separate text file with a .json file extension. Saving data as .csv
files results in files sizes approximately 4 times larger than binary files.
modulation at different frequencies can be applied to different wavelengths of excitation light, and the fluores-
cence evoked by each independently measured by lock in amplification at the appropriate frequency – a form
of frequency-division multiplexing. This is typically used to independently measure the fluorescence evoked in
GCaMP by excitation light at 470 nm and 405 nm.
pyPhotometry uses a different approach to achieve these ends, based on time-division rather than
frequency-division principles. Independent readout of fluorescence evoked by different excitation light wave-
lengths is achieved by alternately switching on LEDs 1 and 2, and acquiring samples of signal 1 when LED 1 is
on and of signal 2 when LED 2 is on (Fig. 3c). Additionally, baseline subtraction is used to render measurements
insensitive to ambient light levels and other low frequency noise sources; for each sample the ADC is read twice;
once with both LEDs off to obtain a baseline measurement, and again with the respective LED on. The baseline is
subtracted from the sample such that only the difference in light intensity between the LED on and off conditions
influences the signal. The acquisition sequence is:
a b c LED 1
Signal 1 baseline
Photoreceiver Signal 1 sample
3 Photoreceiver
0 2 4 6 8 0 2.5 5 7.5
Time (ms) Time (ms)
d e f
2.5 time division
10 0 0.015
Normalised overlap
10
Continous
10 −2 0.01
Signal (V)
1.5
1 10 −3
0.005
10 −4
0.5
10 −5 0
0 0 4 8 12 16 20
0 25 50 75 100 5 10 15 20
Integration time (ms) Average LED current (mA)
LED current (mA)
Figure 3. Time-division illumination. (a) Optical setup for testing time-division illumination and comparison
with sinusoidal illumination. (b) Photoreceiver voltage response to 4 ms illumination light pulse. (c) Timing
of events and photoreceiver voltage waveforms for time-division acquisition sequence at 130 Hz sampling rate.
Black lines show timing of LED illumination and ADC reads of baseline and sample for signals 1 and 2. Blue
and red lines show the photoreceiver voltage waveform due to fluorescence evoked by illumination of LEDs
1 and 2. (d) Baseline subtracted signal as a function of LED current for in-phase illumination, anti-phase
illumination and continuous illumination of fluorescent target. (e) Numerical evaluation of the integration time
required for orthogonality between sinusoidal modulations at 211 and 531 Hz. Orthogonality was quantified
as the standard deviation of the overlap between the two sinusoids normalised by the average overlap of one
sinusoid with itself - where overlap between two signals was defined as their product integrated over the time
window. (f) Comparison of noise on signals obtained using sinusoidal illumination with lock-in amplification
(orange) and time-division illumination with baseline subtraction (blue). Noise was quantified as the coefficient
of variation (standard deviation divided by mean) of the signal, as a function of the average LED current.
This acquisition sequence is used for two acquisition modes, which differ with respect to the analog inputs
read to generate signals 1 and 2. The first mode, termed ‘1 colour time-division’, uses the same analog input (i.e.
a single photoreceiver) for signals 1 and 2. This mode is designed for independently recording the fluorescence
evoked at a single emission wavelength by two different excitation wavelengths, for example for measuring the
fluorescence from GCaMP due to illumination at 470 and 405 nm.
The second mode termed ‘2 colour time-division’ reads signals 1 and 2 from analog inputs 1 and 2 respectively,
i.e. from two separate photoreceivers. This acquisition mode is designed to be used with two fluorophores with
different excitation and emission spectra (e.g. GCaMP/tdTomato), allowing separation of both the excitation and
emission spectra to be exploited to minimise crosstalk.
Validation of time-division illumination. To determine the timing of the acquisition sequence for
time-division illumination we characterised the time course of photoreceiver voltage signals in response to excita-
tion light transients (Fig. 3a,b). The photoreceiver response peaked 0.76 ms after the onset of the excitation light
and returned to baseline 2.5 ms after the light was turned off. The transient response of the photoreceiver voltage
is determined by the photoreceiver’s bandwidth, as both the LED currents and fluorescent lifetimes are on a much
faster timescale, ~1 μs for the LED current (Fig. 4c), and ~1 ns for the fluorescence lifetime of DiI32 and GCaMP33.
Using a photoreceiver with a faster transient response would therefore allow time division illumination with a
faster sampling rate.
Based on these timings we implemented the acquisition sequence shown in Fig. 3c. Light pulses from each
LED were 0.75 ms in duration and occurred at 130 Hz, with the two LEDs in anti-phase. Acquisition from the
ADCs used 64x oversampling at a 256 KHz oversampling rate such that reading a sample took 0.25 ms. The base-
line for each signal was read immediately before the respective LED was turned on, and the signal sample was
read immediately before the LED was turned off.
We assessed the linearity of measurements using this time-division acquisition sequence by measuring the
acquired signal as a function of LED current for in-phase illumination of the fluorescent target - i.e. using signal
1 timings for acquisition and LED 1 timings for the light pulses (Fig. 3c). In-phase illumination produced a linear
relationship between LED current and acquired signal (Fig. 3d). We assessed crosstalk between the two signals by
LED driver
50 50
0.2 0.2
0.0 0.0 0 0
0 2000 4000 0 50 100 −1 0 1 2 0 2 4
DAC value DAC value Time (us) Time (us)
Analog inputs
d e
0 0.5 1 1.5 2
1.5 f
Deviation from 1
1.0 0
linearity (mV)
−1
0 0.5 1 1.5 2
0.5 g
Noise standard 0.25
deviation (mV)
0.0 0
0 0.5 1 1.5 2 0 0.5 1 1.5 2
Input voltage (V) Input voltage (V)
Figure 4. LED driver and analog input characterisation. (a–c) LED driver (a) LED current as a function of the
value written to the Pyboard DAC in 12 bit mode, points show average measurement across 4 driver circuits
tested, lines show linear fit. Left panel – full range of DAC values, right panel – low range of DAC values. The
linear fit is the same on both panels. (b) Standard deviation of LED current across tested driver circuits. (c)
Current waveform in response to 1 ms command voltage pulse, top panel - full pulse, bottom panels –rising and
falling edges. Line shows average of 32 waveforms, shaded area shows standard deviation (shaded area is hard
to see as standard deviation is very small). (d–f) Analog inputs. (d) Voltage measured by Pyboard analog inputs
as a function of input voltage. Points show average of 8 inputs across 4 Pyboards, line shows linear fit. (e) Error
between measured voltage and input voltage. Points show mean and error bars show standard deviation across
inputs. (f) Deviation from linearity of individual inputs, points show the average residuals from separate linear
fits to each input, error bars show the standard deviation of the residuals across inputs. (g) Standard deviation of
noise in the measured voltage, point show the mean across inputs.
measuring the acquired signal as a function LED current for anti-phase illumination – i.e. using signal 1 timings
for acquisition but LED 2 timings for the light pulses. The acquired signal was 0 independent of the intensity of
out of phase illumination (Fig. 3c) indicating no detectable crosstalk. Finally we measured the acquired signal as a
function of LED current for continuous illumination – i.e. using signal 1 timings for acquisition with the LED on
continuously. Again, the acquired signal was 0 independent of LED current, indicating that baseline subtraction
successfully removed sensitivity to steady state light intensity.
We assessed how the time-division illumination used by pyPhotometry compared with the more commonly
used sinusoidal modulation in terms of (i) bandwidth (the frequency range of signals that can be measured) and
(ii) signal to noise.
The bandwidth achievable using sinusoidal illumination with lock-in amplification is determined by the inte-
gration time needed for the modulations of different signals to be orthogonal. Over short integration times the
modulations will not be orthogonal and readout will be contaminated by noise due to varying overlap between
the target signal’s modulation and the modulation of the other signal. We numerically evaluated the strength of
this ‘overlap’ noise as a function of window duration for two signals of equal amplitude modulated at 211 and
531 Hz (as used in5). Overlap noise decreased smoothly with integration time (Fig. 3e) such that the required
integration time depended on the noise level deemed acceptable. To achieve an overlap noise standard deviation
<10−3 of the target signal size required a widow duration of ~10 ms. Using time division illumination we can
acquire samples from two channels at 130 Hz, corresponding to a time window of 7.7 ms per sample. The time
division illumination used by pyPhotometry therefore achieves comparable signal bandwidth to that achieved by
the frequency-division methods used in the literature. Both approaches achieve a bandwidth substantially larger
than that of GCaMP6f.
Intrinsic noise due to e.g. thermal fluctuations in the photoreceivers, may differentially affect signals acquired
using time-division illumination with baseline subtraction and sinusoidal illumination with lock-in amplifica-
tion. We assessed this experimentally by measuring the coefficient of variation of signal recorded using both
approaches from a fluorescent target (DiI solution) as a function of average illumination intensity (Fig. 3f). For
time-division illumination the pyPhotometry acquisition board was used to acquire signals and control the LED.
For sinusoidal illumination a National Instruments USB-6212 BNC board was used to acquire signals and gener-
ate the sinusoidal modulation signal, and a Doric Lenses LED driver was used to drive the LED. The optical setup
was otherwise identical for the two approaches (see methods). Signals acquired using both methods were low pass
filtered at 20 Hz to ensure their bandwidth was equivalent.
Noise was substantially lower for signals acquired using time-division illumination, such that a given signal to
noise level was achieved at approximately 50% of the average LED current required with sinusoidal illumination.
This likely reflects the fact that each LED is on for only 10% of the time during time-division acquisition, such that
the illumination light intensity when the signal is measured is 10x higher than the average illumination intensity.
Electrical characterisation. We performed electrical measurement to characterise the accuracy of the LED driver
outputs and the analog inputs.
Linearity of the LED driver was assessed by measuring the LED current as a function of the value written to
the Pyboard DAC. LED current was linear in the DAC value across the full range of values (Fig. 4a). The stand-
ard deviation of LED currents across the 4 driver circuits tested was <1% of the mean current over a range of
1–100 mA mean current (Fig. 4b). The transient response of the LED driver was assessed by measuring the cur-
rent waveform in response to a 1 ms control voltage pulse (Fig. 4c). Rise and fall times of the LED current were
of order 1 µs.
To assess the accuracy of voltage measurement using the Pyboard ADCs we used the GUI to read both ADCs
at 1 KHz while we presented constant voltage inputs. We tested 2 ADCs on each of 4 Pyboards for a total of 8
analog inputs. The measured voltage was very close to linear across the measured voltage range (Fig. 4d), with
variation in measured voltage across inputs <1% of the mean across the range tested (Fig. 4e). We also assessed
the extent to which individual inputs deviated from linearity by fitting a least squares linear fit to the measure-
ments from each input separately and plotting the mean and standard deviation of the residuals across inputs
(Fig. 4f). Deviation from linearity was <1 mV across the measured input voltage range and had a characteristic
shape across inputs. We measured the noise amplitude of the measured signal. Noise amplitude increased with
input voltage but was <0.06% of the measured voltage across the range of voltages tested (Fig. 4g).
These results confirm the acquisition board is able to control LED current with high accuracy and temporal
precision, and acquire analog signals with good linearity and low noise.
Dopamine neuron recordings. Having characterised the performance of the hardware we tested its appli-
cation to recording neuronal activity in vivo. Calcium transients were recorded from GCaMP6f expressing VTA
dopamine neurons in response to unpredictable reward delivery, which is known to activate a large proportion of
these neurons. For these recordings water restricted mice nose-poked in a reward port where 6 μL water rewards
became available on a random interval 20 s schedule.
Signals were recorded from cell bodies in VTA using the ‘2 colour time-division’ acquisition mode to record a
GCaMP signal and a movement control signal from co-expressed tdTomato (Fig. 5a,b). Similar to previous stud-
ies5,7,8, reward delivery consistently produced a marked, fast transient increase in GCaMP6f fluorescence, while
the tdTomato signal was unaffected.
We also recorded signals from dopamine axons in nucleus accumbens (NAc) using the ‘1 colour time division’
acquisition mode to record a calcium sensitive fluorescence signal evoked from GCaMP by 465 nm light and a
calcium insensitive signal evoked by 405 nm isosbestic illumination (Fig. 5c,d). Unpredictable reward delivery
again produced sharp calcium transients in the calcium sensitive signal while the isosbestic signal was unaffected.
These recordings demonstrate the ability of the system to acquire measures of bulk activity from cell bodies
and axons in deep-brain structures in behaving animals.
Discussion
We have developed an open source acquisition board and GUI for fiber photometry data acquisition. The system
is compact, convenient, and cheap, and we have characterised its performance in detail.
The system can use time-division illumination to independently measure fluorescence evoked by different
excitation wavelengths, combined with baseline subtraction to render measurements insensitive to ambient light.
We compared the bandwidth and signal to noise performance of this approach to the more commonly used
sinusoidal modulation and lock-in amplification. We found that time-division illumination offered comparable
bandwidth to sinusoidal modulation schemes used in the literature, but better signal to noise as a function of
illumination light intensity, allowing approximately 50% lower light intensity for a given noise level. Reducing
illumination intensity is desirable as it reduces photo-bleaching of fluorophores.
Components of the system lend themselves to adaptation for other applications. The LED driver circuit may
be useful in applications where a linear voltage controlled current source with fast transient response is required.
The circuit can be modified to handle higher currents by using a MOSFET with higher power dissipation and
changing some resistor values. For instance, we use a modified version of the circuit as an analog LED driver for
optogenetics.
Photoreceiver
signal
0.1V
0.1V
0.1V
0.1V
Figure 5. VTA dopamine neuron recordings. (a) Fluorescent signals acquired in VTA from GCaMP6f (green)
and tdTomato (red) expressed in dopamine neurons, recorded using the ‘2 colour time division’ acquisition
mode. Black triangles indicate times of unpredictable reward delivery. (b) VTA response aligned on reward
delivery, line shows mean and shaded area shows standard error. (c) Signals acquired from GCaMP6f expressing
dopamine neuron terminals in nucleus accumbens using the ‘1 colour time division’ acquisition mode with
465 nm illumination (green) and 405 nm isosbestic illumination (purple). (d) NAc terminal response aligned on
reward delivery.
The GUI code may by useful as a template for applications which require a Python based GUI for data acqui-
sition and plotting. Some pyPhotometry GUI code is shared by pyControl (https://pycontrol.readthedocs.io), a
system of open source hardware and software for controlling behavioural neuroscience experiments, also built
around the Micropython microcontroller.
While writing this manuscript we became aware of another open source project for fiber photometry called
PhotometryBox34. This supports generation of sinusoidally modulated control signals (sent to external LED driv-
ers), online demodulation of signals for visualisation (though for analysis demodulation is performed offline),
and recording of signals to disk. PhotometryBox uses a microcontroller for generating sinusoidal signals and
online demodulation, and a National Instruments board for recording data. The principal differences between
the systems are: pyPhotometry has built in LED drivers and is USB bus powered; PhotometryBox must record at
higher sampling rates (5 KHz) as demodulation is performed offline, resulting in larger data files. pyPhotometry
is controlled via a dedicated GUI custom-written for photometry while PhotometryBox uses generic electro-
physiology acquisition software (WinEDR) combined with physical controls. Ultimately both systems provide
open source photometry data acquisition at a fraction of the cost of commercial systems. We expect this prolifer-
ation of open source tools to both reduce research costs and facilitate the development of novel applications for
photometry.
Methods
Open source repositories. Design files for the system are at https://bitbucket.org/takam/pyphotometry and
https://bitbucket.org/takam/pyphotometry_hardware. User documentation is at https://pyphotometry.readthe-
docs.io.
Electrical calibration experiments. A Picoscope 2204A USB oscilloscope was used for all electrical meas-
urements and as a signal generator for testing the analog inputs.
LED current was measured via the voltage across the 4.7 ohm current sense resistors that form part of the
driver circuit, using the provided connection points on the board. LED current was measured at 8 DAC values
spaced from 500 to 4000 (the DAC takes values from 0 to 4095 in 12 bit mode), and a further 10 values spaced
from 10 to 100 to cover the low current range. A single linear fit was made to the full set of points.
Time-division illumination. Experiments characterising the time-division illumination mode and compar-
ing it with sinusoidally modulated illumination used a Doric Lenses CLED 560 nM LED for illumination, a Doric
Lenses FCM5 minicube to separate excitation and emission light (excitation filter 555–570 nM, emission filter
580–680 nM), and a Newport 2151 photoreceiver in DC coupled mode to detect the emitted light. The fluorescent
target was a solution of DiI in ethanol, shielded from ambient light and coupled to the minicube via 200 um core
0.48 NA optical fiber.
To evaluate signal to noise for time-division illumination (Fig. 3f) we used the pyPhotometry acquisition
board and GUI to both control the LED and acquire signal. For sinusoidal illumination the LED was controlled
using a Doric Lenses LEDD_2 driver in low power analog modulation mode. The sinusoidal voltage signal used
to modulate the LED was generated by a National Instruments USB-6212-BNC board controlled by WinEDR
software (John Dempster, University of Strathclyde). The modulation amplitude was equal to the average value,
i.e. the modulation depth was 100%. The USB-6212 was used to record both the LED modulation signal and
the photoreceiver signal at 10 KHz. Lock in amplification was performed offline using the following steps: (1)
Bandpass filtering the recorded modulation and photoreceiver signals around the modulation frequency. (2)
Applying a time lag to the modulation signal to phase align it with the photoreceiver signal. (3) Multiplying the
lagged modulation with the photoreceiver signal. (4) Low pass filtering the product at 20 Hz.
All filtering steps used 4th order zero phase filters, implemented by filtering in the forward and reverse direc-
tions using a 2nd order Butterworth filter. The same 20 Hz low pass filtering was applied to signals acquired using
time-division illumination to ensure noise comparison was made on signals of equivalent bandwidth.
Figure 3e used numerical simulation to assess the bandwidth achievable with the sinusoidally modulated illu-
mination and lock-in amplification used in5. Consider using lock-in amplification to measure the amplitude of
sinusoidal signal a(t) in the presence of sinusoidal signal b(t) at a different frequency. Lock-in amplification works
by multiplying the input with a sinusoidal reference signal synchronised to the target signal, followed by low pass
filtering. The lock-in output at time T is:
∞
D( T ) = ∫−∞ F(t − T )r(t)(a(t) + b(t))dt
where r(t) is the sinusoidal reference signal and F(t−T) is the impulse response of the low pass filter.
The quantity we plot in Fig. 3e which we term normalised overlap noise is:
−∞
∞
expectation∫ F(t − T )r(t )a(t )dt
−∞
where the expectation and standard deviation are over measurement time-point T. The numerator is the standard
deviation of noise on the lock-in output due to overlap between the sinusoidal input b(t) and the reference signal
r(t). The denominator is the expected contribution of the target signal a(t) to the lock-in output.
We evaluated this quantity numerically for sinusoids of equal amplitude at 211 and 531 Hz. A
Dolph-Chebyshev window with 100 dB side lobe attenuation was used as the filter impulse response function
F(t − T). We report the overlap noise as a function of the width of this integration window. This window function
was chosen because it performed best out of all those tested.
Dopamine recordings. Mice were anaesthetised with isoflurane (3% induction, 0.5–1% maintenance), treated
with buprenorphine (0.1 mg/kg) and meloxicam (5 mg/kg), and placed in a stereotactic frame. The skull was
exposed and holes drilled to allow virus injection and implantation of fiber optic cannula. Mice were given addi-
tional doses of meloxicam each day for 3 days after surgery, and were monitored carefully for 7 days post-surgery.
For VTA cell body recordings (Fig. 5a,b), GCaMP and tdTomato were expressed in VTA dopamine neurons
using AAV1-Syn-Flex-GCaMP6f-WPRE-SV40 (titer 6.2 × 1013) and AAV1-CAG-Flex-tdTomato-WPRE-bGH (titer
3.1 × 1013) viruses (Penn Vector Core) in male B6.SJL-Slc6a3tm1.1(cre)Bkmn/J mice. The viruses were mixed and
diluted in a ratio of 20% GCaMP6f, 10% TdTomato, 70% saline. 500 nL per hemisphere of the diluted virus was
injected at 1 nL/second at AP: −3.3, ML: ±0.4, DV: −4.5 mm relative to bregma. Recordings were made through
a 200 um 0.53 NA fiber optic cannula implanted at AP: −3.3, ML: +0.4, DV: −4.3 mm relative to bregma.
For the NAc terminal recordings (Fig. 5c,d) GCaMP was expressed in VTA dopamine neurons using AAV
1-Syn-Flex-GCaMP6f-WPRE-SV40 (titer 6.2 × 1013). The virus was diluted 1:10 with saline and 1 μL was injected
per hemisphere at AP: −3.3, ML: ±0.4, DV: −4.2 mm relative to bregma. Optic fiber cannula were implanted at
AP: 1.4, ML 0.8, DV: 4.1 relative to bregma.
Prior to recording, mice were put on a water restriction schedule where on training days they received 0.5–
1.5 mL water from rewards received in the task and on non-training days 1 hour of unrestricted access in their
home cage. Mice maintained a typical body weight of >90% pre-restriction levels. Experiments were carried out
in accordance with the Oxford University animal use guidelines and performed under UK Home Office Project
Licence P6F11BC25.
VTA cell body recordings were acquired using the ‘2 colour time division’ acquisition mode using 470 and
560 nm wavelength LEDs respectively for the GCaMP and tdTomato excitation light and 500–540 and 600–
680 nm emission filters for the GCaMP and tdTomato signals. NAc terminal recordings were acquired using the
‘1 colour time division’ acquisition mode using 470 and 405 nm wavelength LEDs respectively for the GCaMP and
isosbestic signals.
Acquired signals were bandpass filtered between 0.01 and 20 Hz using a fourth order zero phase filter. The full
set of optical components used is listed Table 1 as well as in the hardware repository. The Newport photoreceivers
were used in DC coupled mode for all recordings.
Data Availability
All data and analysis code used in the study are in the file ‘manuscript code and data.zip’ in the downloads section
of the code repository: https://bitbucket.org/takam/pyphotometry/downloads/.
References
1. Adelsberger, H., Garaschuk, O. & Konnerth, A. Cortical calcium waves in resting newborn mice. Nat Neurosci 8, 988–990 (2005).
2. Cui, G. et al. Deep brain optical measurements of cell type–specific neural activity in behaving mice. Nat Protoc 9, 1213–1228 (2014).
3. Cui, G. et al. Concurrent activation of striatal direct and indirect pathways during action initiation. Nature 494, 238–242 (2013).
4. Gunaydin, L. A. et al. Natural Neural Projection Dynamics Underlying Social Behavior. Cell 157, 1535–1551 (2014).
5. Lerner, T. N. et al. Intact-brain analyses reveal distinct information carried by SNc dopamine subcircuits. Cell 162, 635–647 (2015).
6. Wimmer, R. D. et al. Thalamic control of sensory selection in divided attention. Nature 526, 705–709 (2015).
7. Kim, C. K. et al. Simultaneous fast measurement of circuit dynamics at multiple sites across the mammalian brain. Nat Methods 13,
325–328 (2016).
8. Soares, S., Atallah, B. V. & Paton, J. J. Midbrain dopamine neurons control judgment of time. Science 354, 1273–1277 (2016).
9. Girven, K. S. & Sparta, D. R. Probing deep brain circuitry: new advances in in vivo calcium measurement strategies. ACS Chem
Neurosci 8, 243–251 (2017).
10. Saunders, B. T., Richard, J. M., Margolis, E. B. & Janak, P. H. Dopamine neurons create Pavlovian conditioned stimuli with circuit-
defined motivational properties. Nat Neurosci 21, 1072 (2018).
11. Menegas W., Akiti K., Amo R., Uchida N., Watabe-Uchida M. Dopamine neurons projecting to the posterior striatum reinforce
avoidance of threatening stimuli. Nat Neurosci (2018).
12. Miyawaki, A. et al. Fluorescent indicators for Ca2+ based on green fluorescent proteins and calmodulin. Nature 388, 882–887
(1997).
13. Tian, L. et al. Imaging neural activity in worms, flies and mice with improved GCaMP calcium indicators. Nat Methods 6, 875–881
(2009).
14. Akerboom J. et al. Genetically encoded calcium indicators for multi-color neural activity imaging and combination with
optogenetics. Front Mol Neurosci, 6 (2013).
15. Chen, T.-W. et al. Ultrasensitive fluorescent proteins for imaging neuronal activity. Nature 499, 295–300 (2013).
16. Lin, M. Z. & Schnitzer, M. J. Genetically encoded indicators of neuronal activity. Nat Neurosci 19, 1142–1153 (2016).
17. Grimley, J. S. et al. Visualization of Synaptic Inhibition with an Optogenetic Sensor Developed by Cell-Free Protein Engineering
Automation. J Neurosci 33, 16297–16309 (2013).
18. Jing, M. et al. A genetically encoded fluorescent acetylcholine indicator for in vitro and in vivo studies. Nat Biotechnol 36, 726–737
(2018).
19. Patriarchi T. et al. Ultrafast neuronal imaging of dopamine dynamics with designed genetically encoded sensors. Science, https://doi.
org/10.1126/science.aat4422 (2018)
20. Sun, F. et al. A Genetically Encoded Fluorescent Sensor Enables Rapid and Specific Detection of Dopamine in Flies, Fish, and Mice.
Cell 174(481–496), e19 (2018).
21. Jin, L. et al. Single Action Potentials and Subthreshold Electrical Events Imaged in Neurons with a Fluorescent Protein Voltage
Probe. Neuron 75, 779–785 (2012).
22. St-Pierre, F. et al. High-fidelity optical reporting of neuronal electrical activity with an ultrafast fluorescent voltage sensor. Nat
Neurosci 17, 884–889 (2014).
23. Gong, Y. et al. High-speed recording of neural spikes in awake mice and flies with a fluorescent voltage sensor. Science 350,
1361–1366 (2015).
24. Marder, E. The haves and the have nots. eLife 2, e01515 (2013).
25. Chagas, A. M. Haves and have nots must find a better way: The case for open scientific hardware. PLOS Biol 16, e3000014 (2018).
26. Siegle, J. H. et al. Open Ephys: an open-source, plugin-based platform for multichannel electrophysiology. J Neural Eng 14, 045003
(2017).
27. Liberti, W. A., Perkins, L. N., Leman, D. P. & Gardner, T. J. An open source, wireless capable miniature microscope system. J Neural
Eng 14, 045001 (2017).
28. Horowitz P., Hill W. The art of electronics. Cambridge Univ. Press; 2015.
29. Silicon Labs: Improving ADC resolution by oversampling and averaging 2013.
30. Regan T., Munston J., Zimmer G., Stokowski M. Current Sense Circuit Collection; Making Sense of Current 2005,
31. T. Bray E: TheJavaScript Object Notation (JSON) Data Interchange Format 2017.
32. Berezin, M. Y. & Achilefu, S. Fluorescence Lifetime Measurements and Biological Imaging. Chem Rev 110, 2641–2684 (2010).
33. Akerboom, J. et al. Optimization of a GCaMP Calcium Indicator for Neural Activity Imaging. J Neurosci 32, 13819–13840 (2012).
34. Owen, S. F. & Kreitzer, A. C. An open-source control system for in vivo fluorescence measurements from deep-brain structures.
Journal of neuroscience methods 311, 170–177 (2019).
Acknowledgements
We thank Laura Grima for providing the dopamine neuron recording shown in Figure 5a,b, and for comments
on the manuscript. The work was funded by a Wellcome Senior Research Fellowship to Mark Walton
(202831/Z/16/Z).
Author Contributions
T.A. designed and built the hardware and software, designed experiments, analysed data and wrote the
manuscript. M.E.W. designed experiments and edited the manuscript.
Additional Information
Competing Interests: T.A. has a consulting contract with Open Ephys Production Site who sell assembled
pyPhotometry acquisition boards. M.E.W. has no competing interests.
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and
institutional affiliations.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International
License, which permits use, sharing, adaptation, distribution and reproduction in any medium or
format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Cre-
ative Commons license, and indicate if changes were made. The images or other third party material in this
article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the