Adams and Adams
Adams and Adams
Jing Liu, Yan Qiao, Chun Xian Guo, Sierin Lim, Hao Song, Chang Ming Li*
geometric area of 2.0 cm2) was conducted in 0.2 mg/ml GO solution by using 0.3
time was optimized by using 1, 2, 4, 6, 8, 10, 12, 14, 16 min. CVs and EIS
hexacyanoferrate (10 mM K3[Fe(CN)6] + 1M KCl) and the results are shown in Fig.
S1. As electrodeposition time increased, the peak separation of CVs (Fig. S1a) and
1
charge transfer resistance (Fig. S1b) decreased until deposition time reached 10 min
and then they remained almost stable, indicating that at this time point the graphene is
Figure S1. CVs (a) and EIS (b) measurements of graphene/carbon cloth anodes
prepared by different electrodeposition time, the inset of Fig. S1b is charge transfer
resistance under different electrodeposition time.
2
Figure S2. Raman spectra (a), and FESEM images of plain carbon cloth (b) and
graphene/carbon cloth (c).
consecutivelysetup MFCs. When the output current density of the first MFC (MFC
3
A) dropped sharply, the graphene/carbon electrode was collected and used as anode in
deliver the current density as high as that in the first MFC. This result indicates that
the stability of graphene/carbon electrode is very good, and can be repeatedly used.
4
Figure S5. FESEM images of P. aeruginosaadhered plain carbon cloth anode (a),
and graphene/carbon cloth anode (b).
5
Figure S6. UVVis spectra of anodic supernatant for carbon cloth anode based MFC
(MFC/cloth) and graphene/carbon cloth anode based MFC (MFC/graphene)
In FTIR spectrum (Fig. S2a) the strong band at 1600 cm-1 indicates the stretch
vibration of the C=C and C=N in the polycyclic aromatic ring. The broad strong peak
at 3400 cm-1 should be assigned to –OH stretching. There is no C=O bond contained
in the compound as no stretch vibration peak showed around 1700 cm -1. The
characteristic bands locate over the fingerprint range of 1400-400 cm -1. For
comparison, the FTIR spectrum of pure PYO (Sigma) treated with culture medium
was recorded and shows similar bands with that of the anodic supernatant (Fig. S2b).
Figure S7. FTIR transmittance spectra of anodic electrolyte (a) and PYO (b).
6
Standard curve of PYO
10 μg/ml) and subjected to UV-Vis spectrum analysis. The standard curve was draw
according to the absorption at 311 nm, at which wavelength the PYO shows a strong
absorption peak.
Figure S8. UV-Vis spectra of PYO, the inset is standard curve of PYO according to
7
Figure S9. Proliferation curves of bacteria in anodic chamber of MFC during
discharge.
Glucose assay
Glucose concentration of anodic supernatant was measured using glucose assay kit
60μg/ml, 80μg/ml), both standard solution and test solution was processed according
to the protocol. The absorbance at 540 nm and the standard curve is shown below, and
the glucose concentration in the anodic chamber was calculated (Table S1).
(μg/ml)
Absorbance 0.154 0.302 0.445 0.601 0.006 0.006
8
Figure S10. Standard curve of glucose concentration.