IVENT101
IVENT101
IVENT101
Technical Reference
Manual
0473
VersaMed iVent101 Technical Reference Manual
Owner’s Record
The model number and serial number of your iVent101 are on the rear panel of your ventilator. Record the serial number in the
space provided below to have this information should you need to call for service or support.
Model Number:
Serial Number:
CAUTION
Servicing of this product in accordance with this Technical Reference manual should
never be undertaken in the absence of proper tools, test equipment and the most
recent revision to this service manual which is clearly and thoroughly understood.
Technical Competence
The procedures described in this Technical Reference manual should be performed
by trained and authorized personnel only. Maintenance should only be undertaken by
competent individuals who have a general knowledge of and experience with devices
of this nature. No repairs should ever be undertaken or attempted by anyone not
having such qualifications.
VersaMed strongly recommends using only genuine replacement parts,
manufactured or sold by VersaMed for all repair parts replacements.
Read completely through each step in every procedure before starting the procedure;
any exceptions may result in a failure to properly and safely complete the attempted
procedure.
Manufacture Address
VersaMed Medical Systems Ltd
Kadima 60920
POB 5011
Israel
Table of Illustrations
Section 1: Introduction
The information in this manual refers both the ivent101 by VersaMed,
and ClevAir by MPV TRUMA
The information in this manual describes the service, maintenance
and test procedures for the iVent101/ ClevAir ventilator system. It is
intended to assist in maintaining the optimal functional operation and
safety of the device.
This technical reference manual should only be used by authorized
VersaMed or MPV certified technicians. Should you require further
information on becoming a VersaMed or MPV certified technician,
please contact your local VersaMed or MPV representative.
The information in this manual is not intended as a sole source of
reference and has been intended for use in conjunction with the
iVent101/ ClevAir Operators Manual. Both manuals should be
referenced when performing any maintenance to the system.
The contents of this document are not binding. If any significant
difference is found between the product and this document, please
contact VersaMed for further information. Some earlier hardware or
software versions may not be fully supported by this manual.
VersaMed reserves the right to modify the product without amending
this document or advising the user.
1 Exhalation exhaust
2 Proximal pressure connector
(one limb configuration only)
3 Exhalation valve connector
(one limb configuration only)
4 Inspiratory port
5 LCD touch screen
6 Activity LED
7 Battery LED
8 Alarm LED
9 Expiratory port (Two limb configuration only)
10 Air inlet filter
1 O2 inlet connector
2 External DC connector
3 On/ Off button
4 Remote alarm connector
5 MMC/SD memory card slot (future)
6 SpO2 connector (future)
7 LAN connector (future)
8 USB connector
9 Fuse
10 RS232 connector
11 AC connector
12 AC cable strain relief
VersaMed and MPV TRUMA cannot be held responsible for incidents caused
by this device unless the installation, maintenance or modification are
made by an authorized and trained person (in particular, training for
handling of products sensitive to electro-static discharge must include a
section on the use of ESD protected devices), using original spare parts and
respecting quality assurance and traceability rules approved by VersaMed
or MPV TRUMA.
The iVent101/ ClevAir should be used with a bacteria filter to protect the
patient
When nebulizing medications with the iVent101, use HEPA filter connected
to the Patient wye in order to protect the exhalation valve.
Close the oxygen valve on the flow meter prior to disconnecting the oxygen
from the iVent101/ ClevAir ventilator.
Do not cover the iVent101/ ClevAir ventilator while it is in use. Make sure
that the unit is positioned so that its inlet ports are open to freely
circulating air.
During infant ventilation with a high sensitivity, false triggers may occur in
certain clinical conditions. In this case, it is recommended to decrease the
sensitivity (to set its number higher).
The iVent101/ ClevAir ventilator shall not be used with flammable agents
and flammable anesthetics.
Only medically pure oxygen should be used for operation of the device.
Anesthetics and potentially explosive gasses may not be used.
To avoid any potential fire hazard, keep all matches, lighted cigarettes
and other sources of ignition away from the device.
If you notice that the iVent101/ ClevAir is damaged in any way, its life
supporting function can no longer be guaranteed. Stop ventilation with
the defective device immediately and use an alternate form of
ventilation.
*The iVent101/ ClevAir with the Two limb configuration complies with
the requirements of ISO 10651-2: 2004.
** The iVent101/ ClevAir with the One limb configuration complies with
the requirements of ISO 10651-6: 2004.
Activity Green
Battery • Yellow: Charging
Yellow blinking: slow charge or the battery in cool-
down mode
• Blue: In conjunction with the battery status
indicators (see Understanding the Power Sources,
page 19) indicates charging completion.
Alarm Red
On/Off Light
1.7.4. Nameplate
The nameplate is attached underneath the ventilator and contains the
following symbols and information:
Ventilator Name
Ivent101 or ClevAir
Manufacturer:
VersaMed or
MPV TRUMA
Date of manufacturing
Class 2 – Double
Isolation (electrical)
Type BF – Body Floating
S/N
Read this book prior to
use.
CSA/UL
After powering up the ventilator for the first time charge the ventilator
until the icon appears. Refer to section 6.7: Battery Gas Gauge
Update Procedure.
Note: When not connected, wrap the power cord around the battery
screws on the back panel, as shown in Figure 2.2.
Note: When the ventilator is not in use for over a month it is
recommended to disconnect the battery and store it separately from
the device.
Disposable one limb and two limb patient circuits are intended for a
single patient use only.
Reusable one limb and two limb patient circuits should be handled
according to their manufacturer’s specifications.
The patient circuit has to be inspected every day to:
• Make sure there are no cracks or holes in the hose.
• Be certain that all the connections are secure and free from leaks.
Before connecting the patient to the ventilator with a new patient circuit,
you must perform a complete Ventilation Circuit Test. To perform the test,
refer to section 5.1: Patient Tube Testing Procedure.
Note: The air inlet filter must be replaced for every 1 month of operation.
Verify that the air inlet filter is firmly attached by turning all the way
clockwise.
To replace the Air Inlet filter refer to section 10.2.1: Inlet Filter
Replacement.
Verify that oxygen source is closed and that there is no pressure in the
tube prior to connecting the oxygen tubing to the ventilator.
• Last OVT – The last time that OVT procedure was performed. On
the right side the status of the last OVT is displayed: succeeded or
failed.
Note: The ‘Setup’ folder name is case sensitive. Verify that the name is
written correctly.
5. Copy the encrypted file containing the configuration for the target
device to an USB flash drive (Disk on Key), into the Setup folder.
6. Insert the USB flash drive with the “Setup Configuration” folder into
the USB Port (see Figure 1.2).
7. On the Services screen tap the Advanced… button. The Advanced
Services screen is displayed.
8. Tap the Setup Configuration button.
9. The Setup Configuration box is displayed.
The device verifies that a USB flash drive is connected. It starts
loading the configuration files from the USB flash drive.
The setup procedure takes several minutes during which time the
device should be left undisturbed. When the setup procedure is
complete the device restarts and the Main screen will
redisplayed.
During restart the device will emits continuous beeping noise
from the buzzers.
Required Equipment:
• One Limb or Two limb patient circuit according to the
configuration.
• 22mm sealing cap
Test Procedure:
1. Ensure that device is powered up and in Standby mode.
2. Connect a patient circuit to the Inspiratory port (refer to Figure 1.1).
3. From the Main screen tap the (Patient Tube Testing) button.
The Patient Tube Testing screen is displayed.
If, after doing all the above, the Patient Tube Testing fails:
1. Replace the patient circuit and repeat the test.
2. Replace the exhalation valve and repeat the test.
3. If fails: perform a Ventilator Ventilation Tests, refer to section 5.2:
Ventilator Verification below.
• Every time that “VVT Required” alarm pops up on the Main screen.
• Planned maintenance, following the replacement of the pneumatic
unit at 15,000 hours.
• Following any ventilator repair.
• Following software update.
• During troubleshooting of the device.
Note: Make sure sensors calibration has pass successfully prior to the
Pneumatic Unit test.
• Blower Motor
• Blower pressure sensor
• Patient pressure sensor
• Exhalation valve
The test procedure takes approximately 15 seconds to complete.
Test Procedure:
1. From the Pneumatic Unit screen tap the arrow near the Pressure
Test.
The Pressure Test VVT screen is displayed.
Pressure Test
8. Blower pressure test – the measured pressure should be within
pass criteria + 5cmH20 of the target pressure.
If the measured pressure falls outside this range the test fails.
9. Patient pressure test – the measured pressure should be within
+ 5cmH20 of the target pressure.
If the measured pressure falls outside this range (+ 5cmH20) the
test fails.
10. Exhalation test – the ventilator performs Easy Exhale.
If the measured patient pressure is found to be greater than 70%
of the target pressure the test fails.
Test Procedure:
The Flow test screen is displayed automatically when the Pressure
test is completed.
Test Procedure:
The Leak Test VVT screen is displayed automatically when the flow
test is completed.
Leak Test
2. Disconnect the TSI from the test setup and set aside.
Procedure
3. Connect the appropriate type of tubing system to the device.
4. Connect an Rp 20 resistor and 2L test lung to the wye.
5. Tap Start when the test lung is connected.
6. The ventilator delivers flow to the patient system and energizes the
hold pump and solenoid to seal the exhalation valve.
7. The Ventilator closes the POV and monitors the patient pressure and
inspiratory and expiratory flow sensors.
Leak test • If a negative flow is detected by the inspiratory flow sensor the test
pass fails, indicating a leak through the check valve.
criteria
• If the flow is detected by the expiratory flow sensor the test fails,
indicating a leak through the exhalation valve or problem in the hold
circuit.
8. When the test is completed, the VVT results for the Pneumatic Unit are
displayed on the right panel of the screen.
Test Procedure:
1. On the VVT screen tap the arrow near Power.
The AC power Connected Test screen is displayed.
Test Procedure
The External Connected VVT Test screen is displayed automatically
when the AC connected test is completed.
Test Procedure
The External Battery Disconnect Test VVT screen is displayed
automatically when the external battery test is completed.
Test Procedure
1. Before starting this test, make sure a sufficiently charged internal
battery is inserted to the iVent101.
2. Follow the onscreen instructions: disconnect the AC power cable from
the ventilator.
3. Tap Start.
4. The iVent101/ ClevAir verifies that the AC power is disconnected, and
disables the external DC rail.
With External AC disconnected and external DC disabled, the
system switches to the internal battery. The test fails if the
internal battery does not engage.
Note: Since all other power sources are disabled or disconnected,
failure in this test may cause the iVent101 to enter backup mode (touch
screen turns off, a continuous beep sounds) until power becomes
available again.
Test Procedure:
1. The Internal connected test VVT screen is displayed automatically
when the AC disconnected test is completed.
Test Procedure
The Charger Test VVT screen is displayed automatically when internal
connected test is completed.
Calibration Procedure
1. Review the patient configuration setting (One or Two limb) and ensure
that that the appropriate hardware configuration is in use.
2. From the Calibration screen tap Sensors. The Sensors Calibrations
screen is displayed.
Calibration Procedure
1. From the Factory Calibration screen tap PU Devices. The first
calibration screen, the Turbine Factory Calibration screen is
displayed.
Calibration Procedure
1. The POV Factory Calibration screen appears automatically when the
Turbine Factory calibration is competed. Tap Start.
Calibration Procedure
1. The PU Resistance Factory Calibration screen appears
automatically when the POV Factory calibration is completed. Tap
Start.
Calibration Procedure
1. Review the patient configuration setting (One or Two limb) and ensure
that that the appropriate hardware configuration is in use.
2. From the Calibration screen tap Sensors. The Sensors Calibrations
screen is displayed.
Calibration Procedure
Tube Compliance 1. Review the patient configuration setting (One or Two limb) and ensure
Calibration – Continues that the appropriate hardware configuration is in use.
2. From the Calibration screen tap Circuit Compensation. The Tube
Compliance Service Calibration screen is displayed, as shown in the
Figure 6.8 below.
Calibration Procedure
1. Review the device and ensure that the O2 sensor is installed and
enabled. You can enable the O2 sensor through the General
Configuration screen.
2. From the Calibration screen select O2.
The O2 100% Service Calibration screen is displayed.
This procedure may take 5 hours for the standard battery or 10 for the
extended batteryÅ.
Å
Average completion time at 24C. Time may vary with environmental conditions (primarily in terms of
temperature as well as battery capacity).
Notes:
1. It is recommended to perform CDC more frequently if the battery is
not in use on daily basis.
2. Failure to perform recommended maintenance, might lead to
inaccurate indication of the remaining battery capacity, and
accelerated degradation of the battery’s performance.
This procedure may take 15 hours for the standard battery or 28 for the
extended batteryÅ.
Å
Average completion time at 24C. Time may vary with environmental conditions (primarily in terms of
temperature as well as battery capacity).
Parameter Value
Screen Maximum
brightness
Patient circuit Two limbs
configuration
4. Tap the Start button.
5. Disconnect the iVent101 from the charging power source and
continue to ventilate.
6. The Battery icons at the bottom of the Main screen will indicate the
remaining capacity. (Refer to section Understanding the Power
Sources, page 19, for an explanation of these icons).
7. The AC Disconnect Alarm and the Internal Battery in use Alarm are
displayed. Tap Silence to accept these alarms.
8. When the Low Battery Alarm message is displayed, tap Silence to
accept this alarm.
9. When the Empty Battery Alarm message is displayed, tap Silence to
accept this alarm.
10. Wait until a constant high tone is heard, alerting that the battery is
fully depleted. The touch screen may also become blank.
Note: Ventilation will be stopped at this stage.
11. Reconnect the charging power source to the iVent101. The charge
indicator should be yellow during charging. Wait several hoursÅ until
Å
When the ventilator is charged in a room temperature. The time may change depending on the battery
capacity and temperature.
3. Visually inspect the ventilator interior and exterior for any obvious
problems such as missing or broken parts, loose assemblies,
disconnected wires, connectors or tubes. Make repairs, as necessary,
prior to performing the performance verification procedure.
7.3.1.3. Calibration
Performs Sensors Calibration as described in section 6.5.2, and Circuit
Compensation as described in section 6.5.3.
7.3.1.7. O2 System
Only if the O2 system is used you need to perform this test.
The purpose of this test is to verify the accuracy of the oxygen
percentage measurements. This accuracy of the oxygen
measurement should be + 3% absolute up to 30% FiO2, + 7% absolute
between 30 - 60% FiO2.
1. Set parameters according to table below:
Patient Ventilation Rate Vt. Peep P. Sensitivity High Compliance Resistance
Type Mode Limit PIP (on TTL) (on TTL)
Alarm
Adult SIMV Vctrl 12 500 5 60 2 60 0.02 20
2. Connect the ventilator to the oxygen supply through flow meter
3. Connect an external oxygen analyzer to the ventilator inlet.
4. Start ventilation and adjust oxygen flow to 2 L/min
5. Wait for 2 minutes.
6. Record FiO2 measured by iVent-101 and external oxygen analyzer.
7. Repeat the test for the oxygen flow 4 L/min. Record the results of the
measurements.
8. Disable the O2 sensor: Menu>General Config>Enable O2, unmark, Press
Accept).
3. Tap Accept twice to close the Alarms Settings and Settings screen.
Verify that PIP alarm is activated five seconds after Accept was
tapped.
4. Close the Alarm popup window and restore the alarm setting to its
default value. Verify that the alarm stops automatically after one
breath.
Low FiO2:
1. Connect the flow meter of the concentrator to the O2 connector on
the ventilator back panel (refer to section 2.4: Connecting the
Ventilator to an Oxygen Source).
2. On the concentrator adjust the flow to 5L per minute (refer to the
manufacturer’s instructions if needed). Wait a few seconds until the
FiO2 value stabilizes.
3. From the Main screen record the measured value for the O2.
4. Tap Setting – Alarms to open the Alarm Settings screen. Set the FiO2
alarm low settings to value above the measured value.
5. Tap Accept twice to close the Alarms Settings and Settings screen.
Verify that Low FiO2 alarm is activated one minute after Accept is
tapped.
6. Close the Alarm popup window and restore the alarm setting to its
default value. Verify that the alarm stops automatically after four
breaths.
Apnea:
1. Tap Settings, and set the Rate to 2bpm.
2. Tap Accept. Verify that Apnea alarm is activated 20 seconds after
Accept is tapped, and the iVent101/ ClevAir switches to Apnea backup
ventilation.
3. Wait one minute and then squeeze the Test Lung three times to
simulate two subsequent patient initiated breaths. Verify that the
Apnea alarm stops and that the previous parameters are
automatically restored.
4. Restore the Rate setting to its default value.
Section 8: Troubleshooting
This section details how to troubleshoot common problems and error
codes that may be encountered while operating the iVent101/ ClevAir
ventilator system.
10. Download the logs to a USB flash drive using the “Copy Log Files”
feature within the “Advanced…” services dialog.
11. If a laptop PC is available, copy the logs to the laptop, and view them
using the Logs Translator tool.
8.1.2. Diagnostics
1. Read the content of the logs, focusing on the time of the reported event:
a. If no laptop PC is available, use the iVent101 internal log viewer
(Menu – Log Book).
b. If a laptop PC is available, use the Log Translator to read all the
alarms logs.
2. Record occurrences of alarms – type of alarms, timing, repetition,
frequency, and specific numbers in specific log entries that may assist
you in diagnosing the problem.
3. Refer to the following sections within this troubleshooting section to
analyze the problem.
HIGH Block in the cooling vents 1. Verify that the cooling vents are clear
TEMPERATURE of the ventilator, high and clean.
external heat, or internal 2. Shutdown the device for at least 20
malfunction. minutes.
3. Power up the device. If the problem
persists at restart replace the PCB
board (section 10.2.11).
4. If the problem occurs after 15 minutes
replace the Pneumatic unit (section
10.2.12).
LOST LAST USER Internal malfunction 1. If the problem persists and appears
CONFIGURATION with other system alarms perform SW
update (section 4.3.3).
2. Replace the PCB board (section 10.2.11).
3. Touch panel
malfunction
4. Main board.
1. Check the Internal connections.
Blank screen after 1. Internal cable
switch on connections 2. Replace the top panel (section 10.2.8).
3. Replace Main PCB (section 10.2.11).
2. Touch panel
malfunction
3. Main board.
electrical shock. At all times follow the accepted safety practices for
electrical equipment when performing any repairs.
Lever
handle
Lever
handle
5. Remove the two cap head screws located at the upper-rear of the top
enclosure using a 3mm hex key (refer to Figure 10.10 below).
6. Lift the top enclosure up and away from the main enclosure and Main
PCB. Use ESD grounding facilities for protection.
7. Reach under the top enclosure and disconnect the display ribbon
cable, backlight inverter cable, and the LED’s cable from the Main
PCB. Disconnect the ground cable from the main board (refer to
Figure 10.10 below).
8. Lift the top enclosure free and set it aside into ESD safe packaging.
9. Replace the top enclosure and reconnect the display flat cable,
backlight inverter cable, and the LED’s cable to the Main PCB.
10. Secure the top enclosure back in place by screwing the caps head
screws in the lower and top enclosure, using a 3mm hex driver with a
torque of 0.7Nm.
11. Press the screw hole covers back into positions.
Ground
cable Backlight
inverter
cable
LED’s
cable
Display
flat cable
Mounting
Mounting screws
screws
3. Place the device back onto its lower enclosure and remove the screw
(M3 x 8mm) retaining the main enclosure at the exhalation housing.
4. Lift the main enclosure up and away from the device, and while doing
so, reach inside and disconnect the cooling fan harness.
Fans connector
J23
Ground
Gas Gauge
cable Power Speaker
supply cable
Pump/Solenoid
cable cable
Backup battery
Motor cable
POV cable cable
O2 sensor
PU cable
Counter
cable
Ground
Low High
Inspiration
tube
Blower
pressure tube Inlet pressure
tube
Patient
pressure tube LOW
HI
Exhalation
tube
11. Reconnect all of the electrical harnesses to the upper side of the
PCB and the grounds cables to the left and right side of the PCB.
Refer to Figure 11.10 above.
Note: Take care not to place undue stress on the PCB when
reattaching the electrical connections.
12. Secure the main enclosure back in place as described is section
10.2.10.
13. Secure the top enclosure back in place as described in section
10.2.8.
Hold Solenoid O2
sample
tube
Exhaust tube
O2 inlet tube
11. Set the replacement pneumatic unit back into the lower
enclosure and screw it in position with the four screws
removed earlier with a torque of 1 Nm.
12. Tight the Main PCB mounting bracket
13. Reattach the solenoid, with strap, to the Pneumatic Unit.
Refer to Figure 10.17 above.
14. Re-establish the following tubing connections to the
pneumatic unit: box pressure line, O2 inlet tube, O2
sample tubes (two tubes), exhaust line and HPB to
Exhalation Valve. Refer to Figure 10.17 above.
HPB Pressure
port PU Ambient
O2 sample
tube O2 sample
tube
Exhaust tube
Backup
Battery
Connector J19
Back-up
Battery
11. Reconnect all silicone tubes, in right way, to pump and solenoid.
Refer to Figure 10.17
Pump
Power Supply
Power Supply
Mounting Screws
13. Remove the 3 counter sunk screws from inside the battery enclosure
using a small Philips screwdriver.
14. Carefully remove the power supply module and harness from the base.
15. Remove the replacement base assembly from its packaging, and
inspect it for any signs of damage.
16. Set the power supply in position in the new base and fasten it in place
by screwing the screws removed earlier.
17. Reconnect the Power supply harness the base.
Formex box
Power Supply
Harness
Power Supply
18. Fix Formex cover box (as open sheet) in place on the base top using the
adhesive areas.
19. Reassemble the back panel assembly, Refer to section 10.2.19.
20. Reconnect the AC power entry harness.
21. Put back the pneumatic unit in place and assemble, Refer to section
10.2.11 on page 164.
22. Put back the main PCB and reconnect its tubing as described in section
10.2.11, on page 161.
23. Secure the main enclosure back in place as described in section
10.2.10, on page 159.
24. Secure the top enclosure in place as described in section 10.2.8, on
page 157.
25. Reinstall the power pack into place as described in section 10.2.6, on
page 155.
Perfor- Patient
Service PU Power Sensor Factory O2 Touch Alarm
mance Tube OVT
Performed VVT VVT Cal Cal Cal Screen Cal Tests
Tests Comp
Inlet Filter
- -- -- -- -- -- -- -- -- --
Replacement
Exhalation
Valve -- -- -- -- -- -- -- -- -- D
Replacement
One /Two Limb
changeover
-- D -- D D -- -- -- -- D
O2 Sensor
Replacement
-- -- -- -- -- -- D -- -- --
Mains AC fuse
Replacement
-- -- D -- -- -- -- -- -- --
Power Pack
Replacement
-- -- D -- -- -- -- -- -- --
Top Panel
Replacement
-- -- D -- -- -- -- D -- --
Filter Housing
Replacement
- D D -- -- -- -- -- -- --
Main
Enclosure - D D -- -- -- -- -- -- --
Replacement
Main PCB
Replacement D D D D D D D D D D
Pneumatic D
System D D D D D D D -- D
Replacement
Back-Up D
Battery - D D -- -- -- -- -- --
Replacement
RTC Battery D
Replacement D D D -- -- -- -- -- --
Speaker D
Replacement - D D -- -- -- -- -- D1
Perfor- Patient
Service PU Power Sensor Factory O2 Touch Alarm
mance Tube OVT
Performed VVT VVT Cal Cal Cal Screen Cal Tests
Tests Comp
O2 sensor D
harness - D D -- -- -- D -- --
Replacement
Pump/Solenoid D
harness D D D -- -- -- -- -- --
Replacement
Power supply D
Replacement
D D D D D D - -- D
Back Panel D
Replacement
D D D D -- - D -- --
Base D
Replacement
D D D D D D D D D
1. Perform short alarm test according to the Operator’s Manual
2 1
9
10
11.2.4. Accessories
Note: You can select files to convert by dragging the files from the
explore window and drop it to the List of selected files or folders
panel.
3. Click on the Browse button to define the path for the translated log
file.
Note: If you selected a single file to convert, you can check the
“Convert to default filename”. In this case the generated file is located
in the setup folder.
4. Select the dictionary matching your installed version (it is possible
you’ll have only one dictionary available).
Figure A. 6. SW version
6. Click on the Generate button to generate the selected logs:
∗ Note:
A Configuration Screen
Lock screen...............................................59
AC connected test...................................... 88 Configuration screens
AC Disconnected test ............................... 92 Date .............................................................57
AC Power........................................................ 48 Time .............................................................57
AC Power Disconnect test ...................... 92 Cooling Fans .................................................53
Air Inlet Filter................................................. 37 Copy Log File.................................................71
Alarm................................................................ 58
D
Alarm Sound Levels................................... 58
Alarms Tests .............................................. 124 Date ..................................................................57
B Dual limb Patient Circuit ..........................23
E
Back Up Battery ............................. 140, 166
Backlight Inverter PCB.............................. 53 Enclosure............................................. 37, 157
Backup Battery............................................ 50 Exhalation Valve...42, 44, 137, 140, 150,
Battery 153, 154
Backup ....................................................... 50 External Dc Power ......................................48
Integrated................................................. 20 External Disconnected Test....................90
Battery Gas Gauge ................................. 115 External Test .................................................89
C F
Calibration Factory Calibrations ...............................100
factory calibrations ........................... 100 Fan.....................................................................53
FiO2........................................................... 111 Filter
Outlet valve calibration ................... 103 Air Inlet........................................................26
pneumatic resistance calibration104 Filter ................................... 26, 137, 139, 149
Sensors ................................................... 106 Filter Housing.............................................159
Service calibration ............................. 106 FiO2 ....................................................................44
Troubleshooting.................................. 118 Flow Sensor ...................................................42
Tube Compensation.......................... 108 Flow Test.........................................................81
Tube resistance................................... 108 G
Turbine calibration............................. 102
CDC Procedure ......................................... 115 Gas Gauge PCB............................................50
Charge-Discharge-Charge cycle...... 115 General Configuration............................56
Charger Test ................................................. 93 H
Circuit Compensation............................ 108
Cleaning and Routine Planned Heated Humidification .............................25
Maintenance ........................................ 137 High Pressure Box .............................. 40, 44
Configuration screen HME...................................................................25
O2 sensors................................................ 60 I
Sound Levels ........................................... 58
Indicator Lights............................................53
Inlet Filter ................. 26, 37, 137, 139, 149 Dual limb....................................................23
Integrated Battery ..............................20, 49 One limb.....................................................24
Charge..................................................... 115 Patient Configuration................................67
Interface PCB................................................ 52 Patient Pressure Transducer .................42
L Patient Tubing Test ....................................75
PCB
Labels............................................................... 13 Interface.....................................................52
LCD Touch screen ...................................... 52 Pneumatic Unit .......................................52
Leak Test ........................................................ 83 PCB
LEDs.................................................................. 53 Backlight Inverter...................................53
Lock screen configuration...................... 59 Main .............................................................51
Log File ............................................................ 71 Planned Maintenance
Low Flow O2 connector ................................ 27 Every 12 Months..................................140
Low Flow O2 connector with locking nut Every 6 Months ....................................139
....................................................................... 28 Planned Maintenance............................141
Low Oxygen Supply................................... 26 Pneumatic Resistance Calibration ...104
M Pneumatic System .....................................33
Pneumatic Unit ...............................141, 164
Main Alarm Speaker.................................. 53 Pneumatic Unit PCB...................................52
Main Enclosure ......................................... 160 Pneumatic Unit Test ..................................79
Main PCB ..............................................51, 161 POV....................................................................38
Manifold.......................................................... 37 Power Pack .......................................140, 156
MDI.................................................................... 25 Power Sources .............................................19
Membrane.......137, 140, 150, 153, 154 Power Supply.............................................172
Memory Card................................................ 52
R
O
Real Time Clock Battery ..............140, 168
O2 Calibration ........................................... 111
O2 Connector S
Low Flow O2 connector ........................... 27 Sensors Calibration.................................106
Low Flow O2 connector with locking Service calibration...................................106
nut ........................................................... 28 Service Notice............................................128
O2 Sensor.................................................... 154 Service Notices Numbers.....................130
O2 sensors Service Screen..............................................61
Enabling..................................................... 60 Copy Log File............................................71
On Off Switch ............................................... 47 Patient Configuration...........................67
One limb ..................................................24, 67 Setup Configuration .............................71
One Limb ........................................................ 41 Software Update....................................65
One Way Valve ............................................ 41 Technical Info ..........................................64
Outlet valve calibration ........................ 103 Software Update.........................................65
OVT.................................................................... 75 Sound Levels
P Alarm...........................................................58
Button .........................................................58
Patient Circuit............................................... 22 Strut ...................................................40, 41, 42
Warranty
This Product is sold by VersaMed under the warranties set forth in the following
paragraphs. Such warranties are extended only with respect to the purchase of this
Product directly from VersaMed or VersaMed’s Authorized Dealers as new
merchandise and are extended to the Buyer thereof, other than for the purpose of
resale.
For a period of twelve (12) months from the date of original delivery to Buyer or to
Buyer’s order, but in no event for a period of more than two years from the date of
original delivery by VersaMed to a VersaMed Authorized Dealer, this Product, other
than its expendable parts, is warranted against functional defects in materials and
workmanship and to conform to the description of the Product contained in this
Technical Reference manual and accompanying labels and/or inserts, provided that
the same is properly operated under the conditions of normal use, that regular
periodic maintenance and service is performed and that replacements and repairs
are made in accordance with the instructions provided. This same warranty is made
for a period of thirty (30) days with respect to expendable parts.
The foregoing warranties shall not apply if the Product has been repaired other than
by VersaMed or in accordance with written instructions provided by VersaMed, or
altered by anyone other than VersaMed, or if the Product has been subject to abuse,
misuse, negligence, or accident.
VersaMed’s sole and exclusive obligation and Buyer’s sole and exclusive remedy
under the above warranties is limited to repairing or replacing, free of charge, at
VersaMed’s option, a Product, which is telephonically reported to the nearest
VersaMed Customer Service Center and which, if so advised by VersaMed, is
thereafter returned with a statement of the observed deficiency, not later than seven
(7) days after the expiration date of the applicable warranty, to the VersaMed
Customer Service and Distribution Center during normal business hours,
transportation charges prepaid, and which, upon VersaMed examination, is found not
to conform with above warranties. VersaMed shall not be otherwise liable for any
damages including but not limited to incidental damages, consequential damages, or
special damages.
There are no express or implied warranties which extend beyond the warranties
hereinabove set forth. VersaMed makes no warranty of merchantability or fitness for
a particular purpose with respect to the product or parts thereof.