Article Type: Research Article: Elsa Y. León-Marroquín, Daniel J. Mulrow, Rao Khan, and Arash Darafsheh
Article Type: Research Article: Elsa Y. León-Marroquín, Daniel J. Mulrow, Rao Khan, and Arash Darafsheh
Article Type: Research Article: Elsa Y. León-Marroquín, Daniel J. Mulrow, Rao Khan, and Arash Darafsheh
Accepted Article
Spectral analysis of the EBT3 radiochromic films for
Darafsheh1,b,†
1
Department of Radiation Oncology, Washington University School of Medicine, St. Louis,
MO 63110
2
Department of Chemistry, Washington University in St. Louis, St. Louis, MO 63110
a
Electronic mail: [email protected]
b
Electronic mail: [email protected]
*
These authors contributed equally to this work.
†
Senior authorship is shared between these individuals.
This article has been accepted for publication and undergone full peer review but has not
been through the copyediting, typesetting, pagination and proofreading process, which may
lead to differences between this version and the Version of Record. Please cite this article as
doi: 10.1002/mp.13330
This article is protected by copyright. All rights reserved.
Abstract
Accepted Article Purpose: To investigate the spectral response of the EBT3 radiochromic films to different
beam qualities for radiation therapy dosimetry. Dose, dose rate, and inter-batch dependencies
Methods: Pieces of EBT3 films placed between layers of solid water phantoms were
irradiated with 6 and 15 MV photon beams, 6 and 10 MV-FFF photon beams, and 6 and 20
MeV electron beams at dose levels between 0.4-50 Gy. Net absorbance was measured as a
function of wavelength from the spectra acquired in the wavelength range of 400-800 nm
Results: No significant change was observed in the absorption spectra of the EBT3 films
irradiated with the same amount of dose using different beam qualities. Also, no spectral
change with dose rate was observed. The measured net absorbance per Gy was independent
of beam quality in the 1-50 Gy dose range. Slight differences in the spectral shape and
absorption band positions were observed in film samples from different batches. The net
absorbance spectra showed two absorption bands centered around 634-636 nm (primary) and
583-585 nm (secondary). However, depending on the film batch, for doses above a certain
level the primary absorption band appears to “split” into two bands centered around ~624-
Conclusions: The spectral shape of the EBT3 radiochromic films irradiated with photons
(including FFF) and electron beams is beam quality and dose rate independent; however it
assurance (QA), especially to compare planned and measured dose distributions [1-6]. Such
films have been used for in vivo and small field dosimetry as well as in high dose-gradient
regions and other challenging scenarios owing to their high spatial resolution, water
optical density (OD) or absorbance (A) [17]. The absorbance measured using an optical
system can then be correlated to the radiation dose. In clinical dosimetry, a high-resolution
flatbed scanner [18, 19], e.g. EPSON 10000XL, is typically used to establish the dose
response relationship [2, 3, 7], commonly called a calibration-curve [17]. The measured
optical density is a convolution of the densitometer light source emission spectrum, the film
absorption spectrum, and the spectral sensitivity of the densitometer’s detector [20].
Accurate knowledge of the absorption spectrum of the radiochromic films and its
behavior with dose helps to better understand their response to radiation [21]. The optical
absorption spectra of the EBT, EBT2, and EBT3 radiochromic films have been studied for
their optical properties [22-30]. In most of the existing literature it has been shown that the
irradiation of these films results in two principal absorption bands centered at 636±3 nm and
585±2 nm. However, some groups have reported a third band with a lower intensity centered
at 560±6 nm [24-27]. The absorption spectrum has been modeled as a weighted sum of
Lorentzian functions corresponding to the electronic and vibrionic transitions of all polymer
monochromator sequentially creates narrow band light from a broadband light source and the
detector records the intensity of the transmitted light corresponding to the narrowband probed
wavelength.
An in-depth survey of literature reveals that the spectral analysis of radiochromic films
irradiated with electron beams, Flattening Filter-Free (FFF) photon beams, and photon beams
of energies other than 6 MV have not been reported and they have not been rigorously
investigated from spectral perspective. The aim of this work was to investigate the spectral
response of the EBT3 radiochromic films to megavoltage photon and electron beam
irradiations with various beam qualities for radiation therapy dosimetry. We used 6 MV, 6
MV FFF, 10 MV-FFF, and 15 MV clinical photon beams, and 6 MeV and 20 MeV electron
beams to characterize the absorption spectra of the EBT3 films in a wide range of clinical
scenarios. In this work a broadband light, which better mimics the light sources typically used
for film dosimetry, is passed through the sample, allowing all wavelengths to interact with the
sample simultaneously. The spectrometer records all light signals simultaneously and
consequently the acquired spectra are sensitive to optical processes that may lead to
wavelength shifting. The impact of dose, dose rate, and inter-batch dependencies on the
Films from three batches of EBT3 Gafchromic (Ashland Inc., Bridgewater, NJ), batch #1 –lot
number 11091602 (expiration date November 2018), batch #2 –lot number 06281701
(expiration date June 2019), and batch #3 –lot number 06181801 (expiration date June 2020),
were cut into 5×5 cm2 pieces. Film pieces were marked on their corner to indicate their
orientation. Films were used, handled, and stored in accordance with the manufacturer’s
recommendations and the recommendations outlined in the AAPM TG-55 report [17].
2. B. Spectral response
Spectral response of the films was characterized by the absorbance ( ) and net absorbance
()
() = − ()
, (1)
where () is the transmitted intensity of the non-irradiated film sample and () is the
()
() = − ()
, (2)
where () is the transmitted intensity of the irradiated film sample. It should be noted that
dark current spectra ( ()) were obtained by collecting spectra with the spectrometer’s
input aperture covered, and were subtracted from each raw spectrum ( ()) acquired, i.e.
Durham, NC) over 400-800 nm spectral range with 2.5 nm spectral resolution. A deuterium-
tungsten halogen lamp (DH-2000, Ocean Optics) was used as a light source. The spectra were
acquired while both deuterium and halogen sources were simultaneously turned on, to
provide adequate light intensity over the visible spectrum. A set of lenses were arranged to
optically couple the light to a fused silica optical fiber (FT600EMT, Thorlabs Inc., Newton,
NJ) with ~0.6 mm core diameter and 0.39 NA. The light output from the source was
the film samples. Another identical lens was used to collect the light passing through the film
and couple that to the fiber. Both lenses have the following properties: focal length = 40 mm,
diameter = 25.4 mm, and thickness along central axis = 6.4 mm. The film samples were
placed in a holder, ensuring they were perpendicular to the optical beam direction to avoid
introducing any possible systematic uncertainty in the measurement of the net absorbance due
to their orientation. The fiber was carefully placed near the focal point of the second lens to
collect maximum amount of light that passes through the film. One film sample was used for
each irradiation scenario. Since the non-uniformity of the EBT3 films has been shown to be
minimal [31, 32], the spectra were measured over 5 different regions of interest (ROIs) on the
film, as schematically shown as light spots in Fig. 1(b). A distance of ~5 mm from the edges
of the film samples, affected by the strain artifacts due to the cutting process, was avoided in
the measurements. Five spectra were used for data analysis for each spot, each of which was
manufacturer’s specification of the light source (output variations ≤ 0.1% per hour at 254 nm
for deuterium and ≤ 0.1% per hour at 700 nm for tungsten halogen).
Similarly the measurement uncertainty arising from the non-uniformity of the film in the
absorbance spectra was determined by acquiring 50 spectra in each of the five ROIs (Figure
1(b). Therefore, a total of 250 spectra for each film sample were obtained and their standard
deviation was calculated and used as error bars in the plots. This resulted in <1% intra-batch
Spectral response of the irradiated films was measured 48 hours post-irradiation. For dose,
dose rate, and inter-batch studies of the irradiated films, ROI was a single spot with ~1 cm
diameter in the film’s center. For that spot, 50 spectra were collected, each with 100 ms
acquisition time, and the median of those spectra was used for data analysis. In order to
compare the film response to different beam qualities and dose levels, the net absorbance per
2. C. Irradiation setup
The irradiation setup for photon and electron beams is shown in Figs. 2(a) and 2(b),
respectively. For photon irradiations, source to surface distance (SSD) was 100 cm, field size
at isocenter was 10×10 cm2, and the film samples were positioned at a depth of 5cm in a solid
phantoms were placed underneath the film samples in all experiments. For each electron
dose, dmax (1.4 cm for 6 MeV and 2.6 cm for 20 MeV beam). A Varian TrueBeamTM linear
Film pieces from batch #3 were irradiated using 6 MV, 6 MV-FFF, 10 MV-FFF, and 15 MV
photon beams, and 6 MeV and 20 MeV electron beams. The PDD values at 10 cm for the
photon energies used in this work are 66.4, 63.5, 73.5, and 76.7%, respectively. The values of
R50 for the electron energies used in this work are 2.3 and 8.2 cm, respectively. Eight
different doses levels 0.5, 1.0, 2.0, 5.0, 10.0, 20.0, 30.0, and 50.0 Gy were delivered for each
beam quality. A nominal dose rate of 600 MU/min was used for 6 MV, 6 MV-FFF, and 15
MV photon beams, and 6 MeV and 20 MeV electron beams beam. For the 10 MV-FFF
photon beam, 2400 MU/min dose rate was used. The film response defined as ratio of Net
Absorbance and delivered doses (Gy) was determined for each beam quality.
In order to study the influence of various dose on the film spectra, film pieces from batch #1
and batch #3 were irradiated at different dose levels using a 6 MV photon beam and 6 MeV
electron beam separately. Ten dose levels in the range 0.4-17.2 Gy were delivered for the 6
MV photon beam. For the 6 MeV electron beam, films were irradiated with nine dose levels
18.1 Gy at two different dose rates using 6 MV-FFF (600 MU/min and 1400 MU/min) and 10
Film pieces from all batches were irradiated using 6 MV-FFF and 10 MV-FFF photon beams
at 1400 MU/min and 2400 MU/min dose rates, respectively. The irradiations were done at the
3. RESULTS
The absorbance spectrum of non-irradiated EBT3 film samples from three different
batches, measured using Eq. (1), is presented in Fig. 3. These spectra show the typical two
absorption bands characteristic of EBT film models [22-29]. The highest absorption band
centered around 583-585 nm. Interestingly, a third absorption band centered at 548 nm with
the absorbance value 0.58 is also observed for non-irradiated film from batch #1, as shown in
Fig. 3. It can be seen in Fig. 3 that there exists an inter-batch difference in the absorbance
spectrum of the films. Relative to batch #1, an absorbance variation of 16% and 6%
corresponding to primary and secondary bands, respectively, was observed for batch #2.
Using same reference, the absorbance variation of 38% and 26% corresponding to primary
MV photons. Similar results were obtained for other beam qualities used in this work.
In order to determine the Net Absorbance (Net A), transmission spectra were divided by
the transmitted intensity of the non-irradiated film sample according to Eq. (2). The spectral
response of the films from batch #3 irradiated with different beam qualities and doses is
presented in Fig. 5. It can be seen that the net absorbance spectra remains identical for the
films irradiated with the same amount of dose across different beam qualities. In Fig. 5, it can
be seen that the net absorbance spectra of the irradiated films show only two peaks centered
at 636 nm and 585 nm for doses up to 10.0 Gy. However, at 20.0 Gy the absorption band at
636 nm appears to “split” into two bands. In section 3.C we will describe this phenomenon in
more detail.
Figure 6 shows the film response, net absorbance per gray ( / ), for different
beam qualities calculated for the primary (centered at 636 nm) and secondary (centered at
585 nm) absorption bands. The value of / decreases as the dose increases due to the
fact that the film absorbance increases non-linearly with the radiation dose. It is worth noting
that the primary and secondary absorption bands have different / values. Compared
to the primary absorption band, the / decreases by 52 %, 48%, 40%, 30%, 18%, 6%,
2.5%, and 0.7% (average) for 0.5 Gy, 1 Gy, 2 Gy, 5 Gy, 10 Gy, 20 Gy, 30 Gy, and 50 Gy,
measure the film response. Relative to 6MV, variations < 2% in the / were observed
for 6 MV-FFF, 10 MV-FFF, 15 MV, 6 MeV, and 20 MeV. However, when the secondary
band is used, variations < 3% in the / were observed for rest of the beam qualities
compared to the 6 MV case from 1-50 Gy. At low dose of 0.5 Gy, the / variations
were < 3.4%, < 5.3%, < 3.2%, < 4.3%, and < 6.5% for 6 MV-FFF, 10 MV-FFF, 15 MV, 6
MeV, respectively, compared to the 6 MV case [see Fig. 6(b)] measured on the secondary
absorption band.
Figures 7(a-f) show the spectral response of the films from batch #1 and batch #3 irradiated at
several dose levels in the range 0.4-20 Gy using a 6 MV photon and a 6 MeV electron beam.
An increase in the net absorbance is observed with increasing the dose for both photon and
electron irradiation. Remarkably, we observed that the shape of the net absorbance spectrum
is different for low and high dose levels. Additionally, inter-batch differences in the spectral
response were more evident. From Fig. 7, a dose threshold for low doses at ∼5 Gy for batch
#1 and ~10 Gy for batch #3 can be established. For low doses, two intense absorption bands
centered at 636 nm (primary) and 585 nm (secondary) are observed, see Fig. 7(c) and 7(f).
The position of these bands is maintained for all beam qualities at doses < 5 Gy for batch #1
and doses < 10 Gy for batch #3. However, for high doses ( > 5 Gy for batch #1 and > 10 Gy
for batch #3), the primary absorption band appears to “split” into two bands; the spectral
distance of these split bands increases with dose, as it can be seen in Fig. 7(a) and 7(b). For
example at ~15 Gy dose the separation between the split peaks is ~22 nm for batch #1, see
Our results show that the position of the secondary absorption band for batch #1 slightly
varies (red-shifted) with dose. For example, ~4 nm red-shift is seen at ~15 Gy, see figure
7(c). However, the position of the secondary absorption band for batch #3 remains unchanged
with dose.
The spectral response of the films from batch #1 irradiated with 6 MV-FFF and 10 MV-FFF
photon beams at two different dose rates are shown in Figs. 8(a) and 8(b), respectively. No
dose rate dependency was found in the net absorbance spectra of the films.
3. E. Inter-batch dependence
Spectral response of the film samples from three different batches, irradiated with a 6 MV-
FFF and a 10 MV-FFF photon beam, are presented in Fig. 9. Inter-batch variations in net
absorbance were observed. For low doses (<5 Gy for batch #1, <8 Gy for batch #2 and <10
Gy for batch #3), the inter-batch variations were ~8%, on average, for both beam qualities in
batch #2 and batch #3 compared to batch #1. For high doses (>5 Gy for batch #1, >8 Gy for
batch #2 and >10 Gy for batch #3), the spectral differences are not uniform, see Fig. 9.
Inter-batch differences on the position of the absorption bands were also observed. The
red/blue-shift was up to 4 nm for films irradiated with 10 MV-FFF photon beam at 1.81 Gy
It can also be seen that the threshold dose to cause the “splitting” appearance of the
primary band is higher for batch #3 compared to that in batch #1 and batch #2. For example,
in Fig. 9(e,f), the “splitting” at 8.41 Gy can be seen in batch #1, while it is not visible for
It should be mentioned that for irradiated EBT3 films, Villarreal-Barajas and Khan
reported inter-batch differences of ~1% in film response measure using a flatbed scanner
[33]. It was suggested that the difference may be associated to the changes in the composition
and thickness of the film active layer during the manufacturing process between batches.
However, in our work, we found ~7-17% difference in net absorbance of irradiated EBT3
films measured using a spectroscopy system which may manifest as inter-batch variation in
4. DISCUSSION
The spectral shape of the films irradiated with a given dose using different beam qualities
remains similar for films from all three batches studied. For low doses, the net absorbance
spectra of the irradiated films shows two peaks centered at 636 nm and 585 nm. For high
doses, however, the absorption band at 636 nm appears to flatten or “split” into two bands.
The absorption band at 585 nm can be slightly red-shifted, as we observed in films irradiated
at ~15 Gy for batch #1. Previous studies have not reported this “splitting” observation [19-
26]. However, it should be noted that in those studies a monochromatic light source was used
to measure the absorption spectra. In the present study we used a broadband light source,
similar to the light source used in the study by Callens et al. [34]. However, they did not use
an un-irradiated film as a reference, i.e. in the denominator in Eq. (2), to obtain the net
absorption spectra. Instead, they created a “background” reference spectrum from an un-
irradiated film by replacing the values in 500-680 nm wavelength range by a fourth order
polynomial. In the present study, an actual un-irradiated film spectrum was used as the
“splitting” observed in this work. This difference may account for why we are observing
“splitting” in the primary absorption band. Nevertheless, in both works a broadening and
flattening of the primary absorption band can be seen as the radiation dose increases.
To develop further insights into the observed “split” anomaly, we measured Net
Absorbance for a small subset of films using a Cary 5000 UV-Vis-NIR (Agilent, Santa Clara,
CA) spectrometer, which uses a monochromatic light source for absorbance measurement. As
can be seen in Fig. 10, the broadening or splitting is not observed in the spectra when a
monochromic light source is used. This may provide some explanation for the
splitting/broadening appearance of the spectra for high dose levels films presented in this
Further work is required to explain this observed difference in the spectra measured with
different systems. In all of our measurements, we assured that there is enough signal in the
transmission spectra used for absorbance measurement. We hypothesize that the film is
absorbing light in the near-UV and blue region, and weakly emitting light at a higher
wavelength that due to the logarithmic nature of the optical density, would create
flattening/split behavior in the observed absorption spectrum. The choice to use a broadband
light source in our work was to simulate the light source used in a flatbed scanner. We
repeated our measurements with the deuterium-tungsten light source turned off and still
observed the splitting/flattening effects using the halogen light source. Another possibility is
that the transmitted light from the broadband source may be Raman scattered in the second
PMMA layer of the film resulting in light that would have been normally absorbed by the
quantify the polymerization due only to the received ionizing radiation on the films. This
“saturation” effect in the primary absorption band (centered at 636 nm) is in qualitative
agreement with previous observations that the film sensitivity decreases with increasing the
dose. It has been reported in the literature that when a flatbed scanner is used as a readout
system for radiochromic films, the sensitivity of the film measurements depends on the color
channel and the dose level of the films [35]. For example, for EBT3 films, it is shown that the
red channel has the highest sensitivity when the dose level is < 6 Gy; when the dose level is >
6 Gy the green channel has highest sensitivity [28]. This result is confirmed by our work,
when the EBT3 film response is measured using a spectrometer and broadband light source.
The variations of the / between primary and secondary absorption bands is higher
up to ~5 Gy. For doses >5 Gy, the / values for both absorption bands are very
similar. However, another study reports that the optimal dose range to use the red channel is
up to 10 Gy when the EBT3 film is analyzed using a flatbed scanner [31]. These variation in
the optimal range reported for the EBT3 film is related with the inter-batch differences found
in our study. Since the “saturation” effect in the primary absorption band (centered at 636
The changes in the spectral response of irradiated films from different batches could be
associated to the slight changes in the active layers of the films during fabrication processes.
These differences suggest that a per batch film response characterization may still be required
monochromatic optical system (e.g. laser) working around the absorbance bands of the films
to characterize the films [36, 37] can lead to an improved sensitivity and accuracy as well as a
greater dynamic range in the film response measurement. However, caution must be
exercised in selecting such optical systems due to the dose-dependent behavior of the spectral
5. CONCLUSIONS
The spectral response of the EBT3 radiochromic films to megavoltage photon and electron
beam irradiations was investigated for radiation therapy dosimetry. The spectral response of
the EBT3 radiochromic films and the / are beam quality and dose rate independent
for megavoltage radiation. This independence means that the results of the spectral response
characterization for a particular beam quality are valid for any beam quality irradiation
The spectral response is batch and dose dependent; at high doses the primary absorption
band appears to splits into two bands when a broad band light source is used to acquire the
spectra. This dose dependence is important to take into account when choosing light sources
and optical readout system for film dosimetry. The batch dependence of the response
suggests that a batch-based film response characterization is required to ensure accuracy for
clinical dosimetry.
The results obtained in this study show the importance of the spectral response
from the CONACYT Mexico for her post-doctoral studies at the Washington University
School of Medicine.
FIG. 2. Geometry of the irradiation setup for (a) all photon and (b) 20 MeV electron beam
irradiation.
FIG. 4. Absorption spectra of EBT3 films irradiated at various dose levels for 6 MV photon
beam.
FIG. 5. Spectral response of the EBT3 radiochromic films (batch #3) irradiated with photon
(a) 6 MV, (b) 6 MV-FFF, (c) 10 MV-FFF, (d) 15 MV, and electron (e) 6 MeV, and (f) 20
MeV beams.
FIG. 6. Net absorbance per gray of the EBT3 radiochromic films for different beam qualities.
(a) Primary absorption band and (b) Secondary absorption band.
FIG. 7. Spectral response of the EBT3 radiochromic films for different dose levels. For batch
#1: (a) 6 MV photon beam; (b) 6 MeV electron beam; (c) Absorption band positions for films
irradiated at 2 Gy and 15 Gy using a 6 MeV electron beam. For batch #3: (d) 6 MV photon
beam; (e) 6 MeV electron beam; (f) Absorption band positions for films irradiated at 2 Gy
and 15 Gy using a 6 MeV electron beam.
FIG. 8. Spectral response of the EBT3 radiochromic films at two different dose rates. (a) 6
MV-FFF photon beam, and (b) 10 MV-FFF photon beam.
FIG. 9. Spectral response of the EBT3 radiochromic film for batch #1, batch #2, and batch
#3: (a) 6 MV-FFF photon beam; (b) 10 MV-FFF photon beam; (c,e) 6 MV-FFF beam plotted
at a different scale for better visualization; (d,f) 10 MV-FFF beam plotted at a different scale
for better visualization.
FIG. 10. Spectral response of the EBT3 radiochromic film from batch #3 irradiated with 6
MV photons measured using a broadband light source and a monochromatic light source.