AIA360 - ServiceTrainingGuide - REV12 PDF
AIA360 - ServiceTrainingGuide - REV12 PDF
AIA360 - ServiceTrainingGuide - REV12 PDF
TOSOH AUTOMATED
IMMUNOLOGY ANALYSER
AIA-360
Contents list:
1. TRAINING PROGRAM 6
2. LEARNING OBJECTIVES 7
3. AIA GENERAL MEASUREMENT SEQUENCE 8
4. AIA-360 CARACTERISTICS 9
5. AIA-360 SYSTEM CONFIGURATION 10
6. AIA-360 DISASSEMBLY 11
7. AIA-360 ASSEMBLY 12
8. AIA-360 ELECTRONICS 13
8.1. POWER SUPPLY CONFIGURATION 13
8.2. MAIN BOARD 14
8.3. SENS BOARD 15
8.4. DRIVER BOARD 15
8.5. DETECTOR BOARD 16
8.6. EKI BOARD 16
8.7. LCD TOUCH PANEL BOARD 16
8.8. THERMAL PRINTER 16
8.9. KEY BOARD 16
9. ADJUSTING INSTRUMENT 17
9.1. GENERAL MECHANICAL ADJUSTMENTS 17
9.1.1. Turntable adjustment 17
9.1.2. Wash probe adjustment 19
9.1.3. Seal breaker adjustment 20
9.1.4. Sample height adjustment 22
9.1.5. Cup reader position adjustment 22
9.1.6. BCR position adjustment 23
9.2. SAMPLE NOZZLE MECHANICAL ADJUSTMENT 24
9.2.1. Diluent suction position 24
9.2.2. Washing position 25
9.2.3. Drain port position 25
9.2.4. Cup dispensing position 25
9.2.5. Sample suction position for sample cup 26
9.2.6. Sample suction position for 75mm tube 27
9.3. ADJUSTMENTS CHECK 27
9.4. MANUAL ADJUSTMENTS 27
9.5. ELECTRONIC ADJUSTMENT 28
9.5.1. Fluid level sensing 28
9.5.1.1. Definition sampling and level sensing 28
9.5.1.2. Why level sensing is so important 28
1. TRAINING PROGRAM:
AIA-360 MEASUREMENT PRINCIPLE EXPLANATION
AIA-360 CARACTERISTICS
AIA-360 SYSTEM CONFIGURATION
AIA-360 DISASSEMBLY/ASSEMBLY
AIA-360 ELECTRONICS
AIA-360 ADJUSTMENTS
GENERAL ADJUSTMENTS
SAMPLE NOZZLE ADJUSTMENT
FLUID LEVEL SENSING ADJUST
TEMPERATURE ADJUSTMENT
CLOG DETECTION ADJUSTMENT
DETECTOR ADJUSTMENT
SYSTEM SOFTWARE
AIA-360 ANALYTICAL TROUBLESHOOTING
AIA-360 ERROR TROUBLESHOOTING
AIA-360 GENERAL VERRIFICATION
AIA-360 MAINTENANCE
AIA-360 UPGRADES AND TECHNICAL BULLETINS
AIA-360 INSTALLATION
DEMO TOSOHCATALOG AND ISAP DVD
TRAINING EVALUATION AND EXAM
2. LEARNING OBJECTIVES:
Preface: This manual is made for training purposes and can be used for
troubleshooting, maintenance, installation and upgrades. It consists out of 17
modules of practical and theoretical explanations. At the conclusion of this
training, the participant will be able to ….
4. AIA360 CARACTERISTICS:
Waste circuit with waste pump LP401 connected to the overflow basin through
valve SV204 with overflow sensor and to the wash probe through valve SV402 for
wash aspiration during B/F separation.
Diluent circuit with diluent pump LP201 through valve SV202 connected to the
diluent port.
Wash solution circuit with wash solution pump LP202 to clean sample probe.
Sample syringe drive PM201 is responsible for sample pickup and dispensing. BF
syringe drive responsible for wash solution dispensing to the wash probe during
B/F separation.
Substrate circuit with substrate syringe driver PM501 for substrate dispensing.
6. AIA360 DISASSEMBLY:
The following procedure shows a way to disassemble the main parts to have
access to the incubator, in case cleaning needs to be performed during
maintenance or in case parts from the incubator needs to be replaced. A
second method will also be explained.
PROCEDURE:
Mark the turntable when S101 is OFF (this will make the adjustments
easier after reassembly)
Remove sampling needle
Remove the detector
Remove the detector support
Remove COMPLETE Wash probe unit (3 screws) and explain
importance
Remove COMPLETE substrate unit (2 screws)
Remove the CCD camera.
Remove Seal breaker together with support (4 screws)
Remove incubator cover and wheel clamping plate
Remove the instrument frame fixed to the chassis
Remove home sensor incubator.
Remove the turning table, and explain how to remove the sample
holders.
7. AIA360 ASSEMBLY:
This module is a practical session for trainees. Trainees will be asked to
assemble the instrument following the reverse sequence explained in
previous module, only leaving the panels removed! After assembly the
instrument needs to be started up in the test mode: Press ‘MENU’ button on
the key path while switching on the main power and wait until test mode
appears on the screen.
PROCEDURE:
Leave the printer uninstalled as this will allow you to adjust the wash
probe positions and detector center position.
The incubator cover can also be left off for adjustment purposes.
Do not install the detector yet, only the bracket, as you will have to adjust
the detector position.
Do not tighten the screws from the periphery components around the
turntable, as they have to be adjusted later
8. ELECTRONICS:
This module shows an overview of the different PCB’s in the instrument and
their functions. The most important test points, LED‘s and POT meters are
explained. The general electronic diagram (see App 4) is important to
understand the power supply distribution and the connectivity between the
different PCB’s. This diagram is indispensable during troubleshooting
procedures to help you locate and repair electronic and mechanical errors.
PWR-V2 CHECKPOINTS:
FUSES A TP V LED V FUNCTION
FB1 1.6 1 +5 1 +5 LCD
FB2 5.0 2 +5 2 +5 Main PCB/Printerinterfaceboard
FB3 5.0 3 +24 3 +24 Input 24V/main fuse
Valve/motor/printerhead/Fan1/Fan2
4 GND 4 +24
ADO-MAIN-V2 SWITCHES
SWITCHES FUNCTION
SW1 Main CPU reset
SW2 Slave CPU reset
SW3 Main CPU mode setting
SW4 Slave CPU mode setting
ADO-SENS CHECKPOINTS
F A FUNCTION TP TYPE FUNCTION LED TYPE FUNCTION
FB1 1.6 +15V PROT 1 AGND Analog ground 1 EKI Brightness ~ Voltage
ADO-DRV-V2R0
JUM POS SETTING FUNCT TP FUNCT LED FUNCT
1-10 1-2 Open all Microstep 1 GND 1 +5V Vcc and digital
9. ADJUSTING INSTRUMENT:
In this module the participants will practically execute all the mechanical
adjustments, the mechanical adjustments for the AIA360 are easy and can
be performed without any tools or calculations. Electronic adjustments for
level sensing, clog detection and detector are explained in following modules
in order to better understand the adjustments and measurement principles
that are used in all AIA instruments.
9.1. GENERAL MECHANICAL ADJUSTMENTS
9.1.1. Turntable adjustment:
Remarks:
Be aware if you remove the turntable or turntable motor all positions for sample
needle, seal breaker, will need to be readjusted as the pitch sensor S101
position has been altered. Therefore mark the Turntable before removing it!
The Turntable reagent cup position has to be adjusted by the sensor S101 and
is the reference or detector reading position. The detector reading position
alignment with the turntable can also be done by moving the detector bracket
manually; this avoids changing the pitch sensor S101.
Ones the detector position is adjusted by the pitch sensor all other positions for
sample needle, seal breaker, cup reader, substrate, etc. will have to be
adjusted or checked manually towards the cup hole positions on the turntable.
Select MENU, SPECIAL MENU, MECH CHECK and select FREE PULSE
MOTOR.
Remove the detector (do not disconnect the cable)
Refit the Turntable corresponding to the marks you placed on the turntable and
the chassis while S101 was OFF. If no marks available then refit the Turntable
and turn it manually until a cup holder is positioned to the Nord position.
Verify the sensor S101 and when the LED is ON loosen the encoder disk
screws and adjust the slit position by holding the turntable fixed until LED is
OFF. Now fix the screws.
Place a target cup in a position on the turntable and move the turntable until the
target cup is positioned under the detector reading position and the S101 LED
is OFF!
Fine adjust the detector position with the detector bracket screws until the
target cup is perfectly under the detector reading position. Fix the detector
bracket fixing screws.
Remark: Now the reference is set and the encoder disk position cannot be
changed in further adjustments!
If unable to adjust by moving the wash probe, you have to readjust the
turntable pitch sensor by repeating 8.1.1. And adjust the detector position
again!!
Select MECH Check, WASHER, SET HOME and TURNTABLE, SET HOME.
Replace the target cup with a pierced STD cup and press ‘SAMPLE FEED’ 12
times until the target cup is located under the wash probe.
Select MECH CHECK and press FREE PULSE, now lower the wash probe by
hand so that the tip or white nozzle slightly contacts the bottom of the cup (the
gap should be 0!). See also T116
The gap
Screws to set
the seal breaker Other screws
vertical that can be
used for
adjustment
Thicker ring
If above conditions are not for filled (see above example), loosen the two
screws securing the bracket of the sample sensors and move the unit closer to
the turntable. Repeat this procedure to check again.
Select MECH CHECK, select TURN TABLE and press OK. Now select
BARCODE READ and when OK is pushed the barcode reader will read the
label. If not change the angle of the barcode reader by unscrewing the
barcode reader while pushing the OK button until the reader reads correctly.
Also here you can use a functional test in the examination menu: BCR READ
25. Fill up the turntable with 24 reagent test cups and execute the test.
Overflow
Position
Drain
Position
Diluent
Position
Wash
needle
Position
This position is
not used
Select PARAM SET and move the sampling arm by hand to adjust the THETA-
axis and R-axis visually to the center of the STD cup and position the Z-axis
(around 2mm above the cup bottom)
Select SAMPLER TESTCUP BOTTOM and press OK. The sampler motors are
excited and the fine adjustment screen is displayed.
Fine adjust THETA-axis, R-axis and Z-axis by pressing – or + and press each
time OK. In the end the sampling arm returns to home position and the new
pulses are displayed. Press OK again to accept the new adjustment.
Select PARAM SET and move the sampling arm by hand to adjust the
THETA-axis, R-axis visually to the center of the cup and position the Z-axis a
little up from the bottom.
TOSOH EUROPE N.V. Marc Leupe 26
AIA360_ServiceTrainingGuide_REV12
Select PARAM SET and move the sampling arm by hand to adjust the Z-axis a
little up from the bottom (Theta en R-axis are the same for Sample Cup and
don’t need to be adjusted)
Select SAMPLER TUBE BOTTOM and press OK. The sampler motors are
excited and the fine adjustment screen is displayed.
Press OK to accept the new adjustment.
Example:
SAMPLER, THETA-AXIS, DILUENT POSITION and press ok. Here you can
use SET HOME and record the value.
Position the cursor to ABS PULSE and press OK. In the input screen
displayed enter the pulse value and press OK.
Press EXIT to return to MECH CHECK
Write down the new pulses and input the new pulses: MECH CHECK ,
PARAM MODIFY, select MODIFY
Change in Air Pressure: During the down movement the sample probe blows
air and the pressure is monitored. When reaching the liquid surface the
pressure is rising. A predefined pressure threshold will trigger the Zaxis motor
to stop. (used in AIA-21)
Advantage: Plastic disposable tips and probes can be used, less influence
from noise.
Disadvantage: Inaccurate due to obstructions in the probe, air leaks, slow
Change in Capacity: During the down movement the change in capacity of the
sample probe is measured by an electronic circuit. When reaching the liquid the
capacity will change in such a way that it will reach the threshold value and
trigger the Zaxis motor to stop. This method is used in all other AIA
instruments
Advantage: fast detection, accurate
Disadvantage: Influence to noise, use of expensive disposable conductive tips
670
650
640
Noise because of
vibration
630
620
0 2000 4000 6000 8000 10000 12000 14000
Z a xis pulses (P L S )
The instrument performs 5 level sensing’s on the sample cup and the sample
tube. The calculated volume and surface detection AD’s are printed out. An
excellent test to verify the level sensing stability and precision!
The AD values or volumes should be within following ranges:
120 < Scup (µL) < 250
400 < Tube (µL) < 600
CV (volumes) < 4%
If not verify the sample nozzle Z position for sample cup and tube and make
sure the nozzle reaches the bottom(see 9.2.5 and 9.2.6). Example:
Remark: The temperatures can also be adjusted by setting the SET DEG
parameter. Go to SPECIAL MENU, MECH CHECK, PRAM MODIFY. Enter the
correction value to 1.SET DEG at HT401,HT501,HT101 until temperature is within
the standard range. When changing the value at SET DEG restart the instrument
and check the temperature again.
REF +
PID REGULATOR HEATER
SENSOR
2. Pressure in the sampling tip is monitored and measured 9 times during the
suction of specimen. The measured pressures are compared with a calculated
reference curve. If difference between both curves is too high following below
formula, UC flag will be triggered:
9
∑ (Pi –Piref)² >= TH
i=1
The Piref values are determined by different human specimens taking into account
the instrument variations. TH is defined for each specific volume.
Reference pressures are determined for 4,10,20,25,75,100 µL and saved in the
instrument parameter list.
This method is not used in the AIA360.
Following chart shows the AD pressures for a serum compared with the ref curve
and this for a specific volume:
Method 1:
S= C-B/10
Taking the reference as the Blocked Sample Probe Curve would not trigger any
clog. Therefore we multiply the slope with the Blocked Suction Ratio
CLOGGING ADJUSTMENT:
On the SENS board turn VR1 (gain) and VR2 (offset) full clockwise.
Turn VR1 10 turns in CCW direction and check the following: MECH CHECK,
SAMPLER, 10AD, 2CLOG AD.
An AD value for pressure is displayed and needs to be adjusted to 450-550
using VR2, this is the offset adjustment (value A). Vacuum in the sample
needle will produce a lower AD value.
Select EXAM MENU and select IMPASS SENS
When READY is displayed attach a clogging tool (this can be a rubber or Para
film) to the sample nozzle and press OK. Values A,B en C are measured 3
times to verify the stability of the measurements and to verify if the needle was
clogged correctly (check stability of the C values)
The new parameter CLOG LEVEL1 is calculated and a report is printed out.
The new CLOG1 saved in the parameter list after COMPLETE is displayed.
Example:
Good: Not good:
9.5.4. Detector:
9.5.4.1. Theory:
The detector is capable of measuring two fluorescent signals: Sample signal
(450nm-480nm) and reference signal (575nm-625nm). Both signals are amplified
and offset adjusted by VR1 (offset R), VR2 (gain R), VR3 (offset S) and VR4 (gain
S), the signals are sent after V/F conversions to the main board were the signals
are processed:
After incubation and washing the reagent cup contains a certain amount of alkaline
phosphates (ALP) in relation to the concentration of the parameter we have to
measure. Just after incubation period a substrate solution 4MUP is added to the
reagent cup. ALP is an enzyme that is able to transform 4MUP into 4MU. The
speed of the production of 4MU is called the RATE and this will be the parameter
used in relation to the concentration of the test we have to measure. Because 4MU
appears to be fluorescent, we are able to measure the signal and calculate the
RATE!
C ↑ ALP ↑ S ↑ 4MU ↑
The substrate contains a second component DA used as reference and is used to
check the substrate quality (aging), lamp intensity and quantity during assaying
and to compensate lamp intensity and substrate quantity while measuring results
as results are calculated relatively towards the substrate reference background:
S/R
Conclusion: The signal S measured by the detector is a measure for 4MU and the
reference R measured by the detector is a measure for DA.
So after the incubation and wash, 200µL substrate is dispensed in the reagent cup
and the S and R are measured. In the AIA360 S and R are measured 80 times,
each measurement takes 1,2 sec. The complete measurement cycles takes
around 100 sec. (see also aia360.dat file / results/ view Ref, Sam to view all the
measurement data). The following values are measured for 80 points, So and Ro
are the background values measured at daily maintenance in an empty STD cup.
As the So,Ro values vary in time due to detector aging we measure the S and R
relatively to So and Ro:
Si-So/Ri-Ro
Overview of the SoffL,RoffL and S80000nM, R0nM values to adjust the detector:
FITTING EQUATION:
3 known 4MU concentrations: STD1= 0 ,STD2= 10000 and STD3= 80000 are used
to determine the graph:
One STD cup is filled with 200µL substrate and measured under the detector
with lamp on and the following values are obtained and saved as INI values in
the detector parameter list:
Ssubini = substrateIniSam (not used)
Rsubini = substrateIniRef
These values are not used for the detector calibration but are used to calculate
the IniRatio which is a value indicating the detector condition, this parameter is
also saved as INI value in the detector parameter list:
IniRatio= Rsubini-RoffL/ Roini-RoffL (for AIA360 this value is +/- 100)
Each 4MU solution 1,2 and 3 is dispensed 3 times in a STD cup, 9 STD cups
are needed in total and for each solution the mean and CV is calculated:
S1,R1: S2,R2: S3,R3
These values are used to calculate the detector calibration equation or
coefficients a, b and c.
The CV‘s are calculated and checked with the consistency parameters in the
system parameters:
CalibChk0nM ,10000nM,80000nM
New software versions are published on the ISAP. Copy the 4 files onto a new
smart media 32Mbyte or on an USB stick depending on the Main board type.
Software includes the following files:
Cu_main.mot/Touchpanel 3309 bit maps
Installation.mot/Installation 277
M_appl.mot /Main 1920 (includes text)
S_appl.mot/Slave 261
The new software version in smart media format can also be ordered by part nr:
021139 but no guarantee that it will contain the latest version!
WHAT?
It is possible to backup all data to the smart media card (results, testfiles,
parameters, system spec, detector, error list) and view it on a PC. A list of all
parameters can be seen in App 5.
WHEN?
Exchange the main board, parameters can be restored and
instrument does not need recalibration
HOW TO BACKUP?
Turn on system power while pressing MENU to enable the test mode and go
to MENU , select SPECIAL MENU
Select EXAM MENU , insert the SM card or USB stick and select 25 DATA
BACKUP , or DATA TESTFILE backup and press OK
After backup the file aia-360.dat or aia-360.testfile is created.
HOW TO RESTORE?
Insert the SM card or USB stick and select 26 DATA RESTORE in the EXAM
MENU , press OK
After checking info, press OK again and select the items to restore. You can
separately restore the information you want , like detector, test files,…After
restoring , reboot the instrument
PROCEDURE:
PROCEDURE:
Two files supplied on the ISAP DVD in the folder software need to be
translated:
Message.c (text file)
Touchpanel.bmp (bitmap file)
Translation of the text file can be done with the AIA360ResourceEditornew
which is also available on the ISAP. Start up this program and open the
message.c file, in the left column you will see all text messages in English, the
translator has to fill in all lines in the right column, save the file and sent it to
Tosoh headquarters service department
Translation of the bit map file can be done with paint pro. Be aware not to alter
the size of the file. Save the file and sent it back to Tosoh headquarters
service department.
Within reasonable time you will receive an e-mail with the compiled beta
version.
This beta version needs to be tested within the country with the evaluation
procedure.
11. ANALYTICALTROUBLESHOOTING:
11.1. DEFINITION AND EXPLANATION:
These problems are for sure the most difficult to solve or to locate. They can be
instrument, reagent, operator and environment related. Logic thinking, analyzing
and using the method of elimination will enable you to find the problem or
combination of problems. It is therefore indispensable to understand the meaning
of immunology and the reactions that take place in the reagent cup to be able to
interpret the results. This understanding will allow the engineer to distinguish
between instrument or reagent problem. If the problem is instrument related,
further analysis of the results will help to locate your defective part or to locate
external influences like temperature or contamination.
Nonchalant removing used test cups after analyzing while beads will fall in
incubator.
Leaving sample cups and patients in the instrument after analyzing
Not closing the front cover during analyzing
Closing the substrate cap to firmly: can cause error 2017: substrate purge failure.
Using wrong substrate replacement solution, should be 70% ethanol
1 cup = 1 test
1 cup = 1 specific analyte
1 cup = 12 magnetic beads with immobilized antibody/antigen + lyophilized
conjugate: enzyme labeled (ALP) antibody/antigen
Designed for high molecular weight analyses (i.e. TSH, PRL, FER).
Generally using higher sample volumes (20µL-100µL)
Antigen in serum is sandwiched between the antibody which is chemically
bound to the bead and alkaline phosphates labeled antibody in the conjugate.
Fluorescence (enzyme activity) is directly proportional to the concentration of
the test to measure.
Curves are usually linear and most of the analyses use only two calibrators,
with the exceptions of BMG, CKMB, HGH, IRI and LH which require 6
calibrators.
Calibration curve is expressed as rate versus concentration.
Designed for low molecular weight analyses (i.e. TT3, T4, FT4, E2, PROG).
Generally using lower sample volumes (10µL-50µL)
Antigen in the serum and the labeled antigen in the conjugate compete for
antibody sites on the bead. The antibodies are chemically bound to the bead.
Fluorescence (enzyme activity) is reversely proportional to the concentration of
the analyze to measure.
Limited amount of antibody is on the bead.
Analyses require 6 calibrators.
Calibration curve is expressed as B/B0 versus log of concentration ( the
percentage of the rate of the calibrator to the rate of the zero calibrator)
AIA PRINCIPLE
Rate Detector
CPU
Rate= d4MU/dt
1.Substrate diluted by
remaining wash solution
1.-Wash probe height (not deep enough)
-Insufficient vacuum pressure waste pump
1.-Adjust wash probe depth
-Proc BF prime with Hitachi
after suction (related with -Obstructed suction valve SV402 -Check vacuum and valves with vacuumcheckups
error 2016) -Dirty wash nozzle procedure and tool 940062
-Check valve and clean with syringe tool 940081
2.Substrate quality 2. -Old substrate -Clean wash nozzle
(related DL error) - No substrate (alcohol, related DL error) -Verify the Rsubanal with Rsubdaily in aia360.dat
viewer
3. Insufficient detector 3. -Detector Lamp low (related DL error) -Perform EXAM, SUCTION WASHER to verify if all
reading - Detector lens dirty 10 cups are empty (see also chapter 13)
-Piercing quality 2.-Use new substrate and always perform daily
4. Beads aspiration maintenance.
,losing beads 4. -Dirty wash nozzle 3.-Check Rsub value with 200µL substrate and
5. Insufficient reaction - Wrong fitted wash nozzle (stylus too deep inside wash compare with Rsubini
during incubation nozzle) -Clean detector lens
- Wrong depth (too deep) -check and adjust seal breaker
5. - Insufficient diluent take-up: sample probe not deep 4.-Clean wash nozzle
enough in diluent port or diluent pump supply problem -verify stylus level versus wash nozzle and adjust
- Temperature incubator/Wash too low (should be at same level)
-adjust depth wash nozzle
5.-Adjust diluent position for diluent pickup.
-Check diluent supply pump and valve SV202
-Adjust temperature wash and incubator
Main cause: The sensor S402 was not triggered at the end of the wash cycles
(turntable drain position should be filled up with wash) during daily maintenance or
wash replacement when the wash probe was moving down to see if there was liquid.
S402 measures conductivity between dispensing stylus and outer metal tube of the
wash probe. LED 4 on the sens board lights up when a certain level of conductivity is
reached (threshold: cannot be adjusted)
Caused By Solution
Air bubbles in wash solution Checks were bubbles occur and check
tubing damage at connectors. Check if
tubing is present in the wash solution
Internal leaking valve SV401 Check Valve with 940081 tool, clean
valve or replace
External leak at valve tubing Fasten connectors
Leaking syringe seal (dirt at piston) Replace seal 021098
Wrong concentration wash solution Replace wash solution
Leaking wash solution at wash heater Fasten with extra tire wrap or perform
connections T082 and check Z- adjustment wash
probe (see T116)
Sensor cable Check cable and sens board LED4
Caused By Solution
Valve SV402 obstructed and no suction possible Replace or clean valve SV402
Waste pump LP401 malfunction Check vacuum with tool 940062
Air leaks at connectors or damaged tubing Replace tubing and fasten
(pinched tubing) connectors
Clean wash probe or replace
Wash probe obstructed by dirt or beads nozzle
Dismantle probe completely and
fasten firmly the Teflon flanged
Internal leakage probe causing false conductivity part.
Wrong adjusted wash probe X,Y or touching the
cup border (false conductivity) Readjust the wash probe X,Y
Valve SV204 unable to close and causing Verify valve SV204 with syringe
insufficient suction tool 940081
Caused by Solution
Insufficient substrate (air bubbles) Check were air bubbles occur
Valve SV501 leaking or malfunctioning replace or clean SV501
Leaking substrate syringe due to seal Replace plunger seal
Obstructed substrate tubing caused by not Use tool 940081 to flush tubing with
performing daily check alcohol
Valve SV402 unable to close and causing
continuous suction Replace or clean valve SV402
Tubing damaged and connector leaks Replace tubing and fasten connectors
Substrate bottle cap too fastened and causing
vacuum in the bottle Check cap
Caused By Solution
Fasten connectors, check valve
Leaking tubing connector at valve SV501 leaking
Leaking substrate syringe Replace seal plunger
Leaking or disconnected tubes at BF wash Check tubing, Perform TB82 to
probe prevent leakage at wash heater
Condensation, Humidity Check instrument location
Insufficient wash suction in reagent cup (see
error 2016) causing overflow Check vacuum at waste pump
Disconnect sensor and check sensor
Faulty sensor board status
Caused by Solutions
Leaking or damaged wash heater tubing Fasten or replace tubing
Leaking valve SV401 Replace valve
Condensation, Humidity Check instrument location
Disconnect sensor and check sensor
Faulty Sensor board status
Loose connector at valve SV401 Tight connectors
Caused by Solution
Air Bubbles on Sample Surface Check Sample, Operator
Noise caused by EMI Check shielding, and environment
Noise caused by Vibrations Check bearings
Check Damper
Check Tension Belt
Check Feed Screw Z-movement
Noise by elec. fluctuations Check Grounding
Wrong Offset value Recalibrate Offset Value
Wrong LevelCalPulse Verify this parameter =0
Caused by Solution
No sample Check sample and operator
Wrong Zmaxtube Readjust Zmax parameter
Disconnected Sample probe Check wire from EKI board
Wrong setting OFS value Recalibrate Hand touch value
Crashed sample probe stylus Replace
Wrong adjusted sample probe Adjust X and Y
Defective EKI board Replace EKI board
12.18. DL flag
Main Cause: When Rsub/Rsubini < 0,5 (SubIniRefChk)
Caused By Solution Verification
Insufficient substrate in substrate Replace substrate Check Rsub value by
bottle or substrate empty placing manually a
cup with 200µL
substrate under the
detector.Perform
Detector Light
reading, the check the
Rsubini in the Exam
Menu Detector
Calibration Info
Calibrate detector
Insufficient wash probe suction Check if instrument is Check cups liquid
causes substrate dilution and low not causing 2016, quantity after
Rsubanal perform the suction test analyzing, should be
manually with Hitachi equal for all cups and
cup around 200µL
Some cups have bad piercing Check seal breaker Verify all cups after
analyzing for piercing
quality
Dirty detector lens clean Lens
Wrong adjusted detector reading Adjust reading position Remove detector and
position verify reading position
with pitch sensor
S101=OFF
12.19 HB Flag
Main Cause: 4MUt=0 < 1500 or during daily maintenance
Caused By Solution Verification
Old substrate Replace substrate During daily
maintenance the 4MU
is printed out. Also
you will see a higher
S1 value.
Drain Tubing:
Check if there is
or was liquid
inside
Good Bad
Values background S0, R0 too high Front cover was Close front cover
(>600) open, detector
problem
Value S1 too high (this will cause Old substrate, Replace substrate,
HB)verify also 4MU value detector problem. replace detector.
Decontaminate
substrate lines with
0.1N HCL
Value R1 too low (< 25000) (check Substrate quantity, Perform 8: Detector
the Rsubini in detector calibration substrate quality. light measurement
data) with 200µL
substrate, use new
substrate
Remark: Enter the TEST mode to get access to the EXAM menu. Most important
tests that can influence the result are mentioned
Verify if the wash nozzle is fitted correctly: The metal stylus inside the white nozzle and
the nozzle should be at the same level.
Verify if wash dispensing and suction tubing is connected correctly
Place the wash probe back and check if tubing and cables
Check manually
the up and down
movement
(spring) Loose this screw
to remove the
wash probe
Remove the blue cover behind the bottles and check for leakage around the
pumps and valves.
Valves
Drain Pump,
output to
waste bottle
Level sensing
test: Probe will
touch the liquid
and will move
up again. Then
the volume is
calculated.
14. MAINTENANCE:
Maintenance for the AIA-360 is normally foreseen ones a year. Additionally every 20000
tests or once every 3 years the diluent, wash and waste pumps need to be replaced. This
maintenance procedure is the result of practical experience and is build up in 6 steps.
An additional maintenance for customers is recommended every 4000 tests and consists
out of a visual inspection, clean or replace the wash nozzle.
For the official Tosoh maintenance procedure please see App 7. There are 2
maintenance kits available: 910287 Yearly Maintenance and 910228 3-yearly
maintenance kit (with pumps)
First interrogate the customer about existing problems or if he has special remarks about
the functionality of the instrument.
14.1. CLEANING:
Remove the covers: Back, front and middle and general clean the instrument
with Wipes to clean. You can use 940074 Tool Wipes to clean, 940083 Tool
cleaning Decont. lingettes
Remove the dust and clean ventilators.
Check the instrument for leakage: at the S701 and S702 sensors, at the
bottom of the bead collector, at the wash probe area, at the wash heater input
and output, at the pump chassis plate.
Clean the Turntable especially at the turn table drain position
Clean the seal breaker
Dismantle and clean wash probe completely if very dirty, illustration:
Clean all the parts and refit the probe, make sure to tighten all parts firmly to
prevent air leaks.
TOSOH EUROPE N.V. Marc Leupe 83
AIA360_ServiceTrainingGuide_REV12
Sample probe and flush with 70% alcohol: use the new tool 940081: Tool
Syringe for M5 F&M
Every 4000 tests the customer or the service engineer needs to clean or
replace the washing nozzle. The procedure to remove the panels and the
washing probe is explained in T029. Technical bulletin T057 explains the new
type of substrate holder that makes the removal of the panels easier.
Instruments from #1701 have that option.
Customer should regular check the turn table drain position for beads and dirt.
Instruct the customer to report any leaks or abnormal dirt that has been
observed.
16. INSTALLATION:
For detailed installation procedures please see service manual or ISAP. The following is a
quick installation procedure. The installation program in the software can also be used to
install the instrument. Do not forget to verify needle and general adjustments during
installation!
Unpack the instrument and remove the 3 screws securing the sample arm at the back
of the instrument. Attach the metal plate over the hole.
Remove the screws on the back panel securing the turntable
Connect waste, wash and diluent tubing and level sensing wiring, respecting the
colors.
Install the waste, diluent and wash bottle holder
Install printer paper.
Make up fresh diluent, wash and substrate.
Remove the white middle panel and start the instrument in the test mode. In
SYSTEM set YES for ‘rate printout’ parameter
Perform all sample probe adjustments as explained in chapter 9.2.
Verify all other adjustments in particular wash probe X,Y and Z
Flush diluent 2 times, wash 6 times and substrate 2 times.
Switch off and on instrument and perform a daily maintenance
Check the background and substrate values.
Perform a precision test on a blood sample prior to calibration! (e.g.: run 5 times a
patient for FT4, TSH or any other test and evaluate the CV on the rates)
If rate values are ok, switch off the instrument and enter the test mode.
Program or enter all assay specification for the tests to calibrate.
Perform the calibrations
Perform the ‘Installation Validation Sheet’ to check if instrument is ok
(ISAP/AIA_SERVICE/AIA360/INSTALLATION/…)
Training checklist.
Training Evaluation.
Exam.
2 Sample dispensingDispensing
unit syringe drive 5 PM201 Two-phase stepping motor Syringe suction Dispensing syringe home 9 S201 Transmission Home (Up)
Dispensing nozzle Z-axis drive 6 PM202 Two-phase stepping motor Down Dispensing nozzle Z-axis home 10 S202 Transmission Home (Up)
Dispensing nozzle R-axis drive 7 PM203 Two-phase stepping motor Outward Dispensing nozzle R-axis home 11 S203 Transmission Home (Front)
Dispensing nozzle θ-axis drive 8 PM204 Two-phase stepping motor CW Dispensing nozzle θ-axis home 12 S204 Transmission Home (Front)
Surface detection (Sample) 13 S205 Conductive A/D processing
Block detection 14 S206 Pressure sensor
A/D processing
Nozzle washing control valve 9 SV201 Two-way solenoid valve Open
Diluent pump 10 LP201 Liquid pump ON Surface detection (Drain) 15 S207 Electrode Liquid present
W ash solution pump 11 LP202 Liquid pump ON
Diluent port valve 12 SV202 Two-way solenoid valve Open
W ash solution port valve 13 SV203 Two-way solenoid valve Open
Diluent drain valve 14 SV204 Two-way solenoid valve Open
3 Seal breaker Seal breaker Z-axis drive 15 PM301 Two-phase stepping motor Up-Down Seal breaker Z-axis home detection 16 S301 Transmission Home (Up)
Seal break check 17 S302 Transmission Seal break
4 W ashing unit B/F probe (up/down) 16 PM401 Two-phase stepping motor Down B/F probe 1 home position detection 18 S401 Transmission Home (Up)
Overflow detection 19 S402 Electrode Overflow
W ashing syringe drive 17 PM402 Two-phase stepping motor Syringe suction W ashing syringe home detection 20 S403 Transmission Home
W ash solution temperature regulation 18 HT401 C a r t r i d g e heater ON Incubator temperature detection 21 T401 Thermistor A/D processing
Vacuum source 19 LP401 Liquid pump ON
B/F discharge control valve 20 SV401 Three-way solenoid valve Syringe→Discharge
B/F discard control valve 21 SV402 Two-way solenoid valve Open
5 Substrate unit Substrate syringe drive 22 PM501 Two-phase stepping motor Syringe suction Syringe pump substrate home detection 22 S501 Transmission Home (Up)
Substrate temperature regulation 23 HT501 C a r t r i d g e heater ON Incubator temperature detection 23 T501 Thermistor A/D processing
Substrate discharge control valve 24 SV501 Three-way solenoid valve Syringe→Discharge
6 Detector unit Detector light source LED lighting 25 LE601 Ultraviolet LED ON/OFF Detector data (V/F input) Sample 24 V/F
Detector data (V/F input) Reference 25 V/F
Detector drive 26 DM601 DC motor Left-Right Detector position (L) 26 S601 Transmission Position OK
Detector position (R) 27 S602 Transmission Position OK
7 Panel
W arning 27 LE701 Red LED ON
SV201
V7
SV402 23
Turntable S201 LP202
⑯ Dispensing nozzle θ-axis Dispensing syringe drive
PM204 S204 PM201 Liquid feed pum p (W ash solution)
Dispensing nozz;e R-axis Seal breaker
PM203 S301 PM301
S203 S302
PM102
Cup reader
S105
Agitator Dispensing nozzle Z-axis 33 W ash
S202 solution
W ashing port
V6 S103
Surface
SV204 detection
28 ⑲ 27
overflow line S104
sample detection LP201
22 20 I/O specs
Liquid feed pum p (Diluent)
Incubator temperature regulation 33
DI Diluent
Heater HT101
Therm istor T101
1.5kΩ
W aste bottle
R
W aste pum p
LP401 4.7kΩ 1/2W
ADC
DO
RS-232C
COM
Capture
Board
App2: Flow diagram
(DSC_ 40_ BW
App5: Parameters
Annual Inspections
Location Inspection Item Description Corrective Action
1 Table (Carousel) □Cup holder cleaning Wipe the cup contact surface using alcohol.
□Sample holder cleaning Wipe the blood tube and the sample cup contact
surface using alcohol.
□Hot plate cleaning Clean the contact surface of the cup bottom using
alcohol.
□Position check Check the positional relations with units Position adjustment
according to the adjustment procedures.
2 Seal breaker □Punch cleaning Remove the punch and clean the blade
using alcohol.
□Positional deviation check Check the position according to the adjustment Adjust the position with a jig.
procedures. The breaker shall punch a seal Replace the breaker if the blade is dull.
evenly without making printing unclear.
3 Dispensing arm □Metal nozzle cleaning Clean the nozzle using alcohol.
(Nozzle)
□Nozzle tube cleaning Check the suction hole of the nozzle using
a mirror to ensure that the nozzle is clean
and there is no blockage.
□Sample suction position check Check that the nozzle comes to the center at Position adjustment
each position:
・Nozzle washing position
・Cup discharge position
・Sample suction position
□Greasing Grease the Z-axis feed screw. Clean with CRC and apply AFC grease
Grease the R-axis LM guide. Clean with CRC and apply AFC grease
4 Dispensing syringe □Leakage test Attach an end-blocked chip to the nozzle and Tube and joint check
reduce 100 ul for pressure fluctuation check. and syringe replacement
□Greasing Grease the feed screw. Clean with CRC and apply AFC grease
□Suction AD Verify the suction AD before and after syringe seal
replacement
□Seal replacement Replace the seal with a new one. Seal replacement
5 Washing syringe □Greasing Grease the feed screw. Clean with CRC and apply AFC grease
□Seal replacement Replace the teflon seal with a new one. Seal replacement
6 Substrate syringe □Greasing Grease the feed screw. Clean with CRC and apply AFC grease
□Seal replacement Replace the seal with a new one. Seal replacement
7 B/F washing □Washing probe cleaning Remove the probe and eliminate any blockage
or dirt from the probe tip.
□Wash probe tip replacement Replace the teflon tip on the wash nozzle
□Suction check Place the washing probe in a vessel and Check to see that the tubing is not blocked.
prime the wash solution to see that there is Replace the liquid pump
no liquid suction problem. and the washing probe.
□Position check Check that the washing probe lowers to Position adjustment
the center of the cup.
8 Substrate dispenser □Point cleaning Remove the substrate dispenser and clean
the dispenser tip.
9 Detector □Lens cleaning Remove the detector and wipe the lens using
cleaning alcohol (70%).
10 Liquid feed □Filter replacement for diluent and Replace the filter with a new one.
wash solution tanks
□Air purge from wash solution line Replace once with alcohol.
□Substrate line cleaning Clean the substrate line with diluted nitric acid.
□Diluent pump replacement Replace the diluent pump with a new one.
□Wash solution pump replacement Replace the wash solution pump with a new one.
□Waste pump replacement Replace the waste pump with a new one.
11 Temperature check □C heck temperature of main incubator Adjust temperature if not in range
□C heck substrate line incubator Adjust temperature if not in range
□Verify wash solution preheater Adjust temperature if not in range
12 Carousel base plate □C lean the plate around the carousel Look for lost beads
After the above procedures, initialize the system to see that no abnormalities occur.
If a corrective action affects data, perform calibration and control measurement as required.
Test AIA360