User'S Manual: PC Usb Digital Oscilloscope
User'S Manual: PC Usb Digital Oscilloscope
OSCILLOSCOPE
USER’S MANUAL
Hantek 6052BE/6082BE
www.hantek.com
Content
General Safety Summary ............................................................................... 1
Chapter 1 Getting Start .................................................................................. 3
1.1 System Requirement ......................................................................................................... 4
1.2 Install Software ................................................................................................................. 5
1.3 Install Drive....................................................................................................................... 8
1.4 General Check ................................................................................................................. 13
1.5 Probe Compensation ....................................................................................................... 13
1.6 Self Calibration ............................................................................................................... 15
1.7 Function Check ............................................................................................................... 15
1.8 Accessories...................................................................................................................... 17
Chapter 2 Operating Basics ......................................................................... 18
2.1 The User’s Interface ........................................................................................................ 19
2.2 The Menu System ........................................................................................................... 21
2.3 The Horizontal System .................................................................................................... 24
2.4 The Vertical System ........................................................................................................ 24
2.5 The Trigger System ......................................................................................................... 25
2.6 Input Connectors ............................................................................................................. 27
Chapter 3 Functions ..................................................................................... 28
3.1 Setup the Oscilloscope .................................................................................................... 29
3.2 Set Vertical System ......................................................................................................... 30
3.3 Set Horizontal System ..................................................................................................... 39
3.4 Set Trigger System .......................................................................................................... 41
3.5 Save/Load........................................................................................................................ 44
3.6 Utility Function ............................................................................................................... 45
3.7 Measure Signal ................................................................................................................ 55
3.8 The Display System ........................................................................................................ 61
3.9 Zoom In/Out and Drag Waveforms ................................................................................ 67
3.11 Acquisition Modes......................................................................................................... 70
3.12 Print And Print Preview ................................................................................................ 70
Chapter 4 Application Example .................................................................... 72
4.1 Simple Measurement ....................................................................................................... 73
4.2 Pass/Fail Test .................................................................................................................. 74
4.3 Capturing a Single-Shot Signal ....................................................................................... 76
4.4 The Application of the X-Y Operation ........................................................................... 77
Chapter 5 Appendix ..................................................................................... 82
Appendix A: Specifications .................................................................................................. 83
Appendix B: General Maintenance ....................................................................................... 85
Hantek 6052BE/6082BE
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Connect the probe output to the measurement
device before connecting the probe to the circuit under test. Disconnect the probe
input and the probe reference lead from the circuit under test before disconnecting
the probe from the measurement device.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Use Proper Probe. To avoid shock hazard, use a properly rated probe for your
measurement.
Do Not Operate With Suspected Failures. If suspected damage occurs with the
device, have it inspected by qualified service personnel before further operations.
USER’S MANUAL 1
Hantek 6052BE/6082BE
Safety Information
In order to ensure the correct using and the best efficient service, please carefully read the
user’s manual.
System Symbols
These symbols may appear in this manual or on the instrument.
Caution: “Caution” identifies conditions and actions that may damage the
product or other properties.
System Terms
The following terms may appear on the instrument:
Danger: The term “Danger” is used in the manual to indicate that when you read this
mark, personal injury may be caused to immediately.
Waning: The term “Danger” is used in the manual to indicate that when you read this
mark, personal injury may not be caused to you immediately, but you need to be
cautionary.
Notice: The term “Noise” is used in this manual to indicate that damages may be
caused on this product or other properties.
USER’S MANUAL 2
Hantek 6052BE/6082BE
In addition to the list of general features on the next page, this chapter describes how to
do the following tasks:
System Requirements
Install software
Install Drive
General Check
probe compensation
self calibration
Accessories
USER’S MANUAL 3
Hantek 6052BE/6082BE
Operating System
Windows XP/Vista/Win 7
Processor
Upwards of 1.00G processor
Memory
256M byte
Disk Space
500M disk free space
Screen resolution
800 x 600
Recommended Configuration
Operating System
Windows XP SP3 System
Processor
2.4G Processor
Memory
1G Byte Memory
Disk Space
80G Disk Space
Screen resolution
1024 x 768 or 1280 x 1024 resolution
DPI Setting
Normal Size (96DPI)
USER’S MANUAL 4
Hantek 6052BE/6082BE
Caution: You must install the software before using the oscilloscope.
USER’S MANUAL 5
Hantek 6052BE/6082BE
USER’S MANUAL 6
Hantek 6052BE/6082BE
USER’S MANUAL 7
Hantek 6052BE/6082BE
Example: Hantek6052BE
1. Connect the A-Type Plug of USB cable to your PC’s USB port.
USER’S MANUAL 8
Hantek 6052BE/6082BE
USER’S MANUAL 9
Hantek 6052BE/6082BE
USER’S MANUAL 10
Hantek 6052BE/6082BE
USER’S MANUAL 11
Hantek 6052BE/6082BE
11. New hardware wizard installs software and finish new hardware search wizard.
USER’S MANUAL 12
Hantek 6052BE/6082BE
Perform this function to match the characteristics of the probe and the channel input.
This should be performed whenever attaching a probe to any input channel at the first
time.
From the “Probe” menu, select attenuation to 1:10. Set the switch to “X10” on the
probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, insert
the tip onto the probe firmly to ensure a proper connection.
Attach the probe tip to the Probe Compensator and the reference lead to the
ground connector, select CH1, and then press the “AUTOSET” button into the menu or
the toolbar.
USER’S MANUAL 13
Hantek 6052BE/6082BE
Correctly Compensated
Over compensated
Under Compensated
If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe
for the fattest square wave being displayed on the oscilloscope.
Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the
insulated cable, and do not touch the metallic portions of the probe head while it is
connected with a voltage source.
USER’S MANUAL 14
Hantek 6052BE/6082BE
The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5v or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the “Utility -> Calibration” option and follow the directions on the screen.
The self calibration routine takes about several minutes.
Perform this functional check to verify that your oscilloscope is operating correctly.
1. Set the attenuation switch on the probe as 10X and connect the probe on the
oscilloscope with CH1. Aim the slot in the probe connector at the faucet on BNC of CH1
and insert, then, turn right to lock the probe. Finally, attach the tip of probe and ground nip
to the Connector of Probe compensator.
USER’S MANUAL 15
Hantek 6052BE/6082BE
2. Set the CH1 probe attenuation of the oscilloscope to X10. (The default is X1).
3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click
USER’S MANUAL 16
Hantek 6052BE/6082BE
1.8 Accessories
All the accessories listed below are standard accessories for the oscilloscope:
Probe x 2 (1.5m), 1:1, (10:1) Passive Probes
A User’s Guide
An USB Cable
USER’S MANUAL 17
Hantek 6052BE/6082BE
Input Connectors
USER’S MANUAL 18
Hantek 6052BE/6082BE
Click the software icon on the desk after you finished the software setting and equipment
connecting. Then a user interface will be showed as follows:
USER’S MANUAL 19
Hantek 6052BE/6082BE
USER’S MANUAL 20
Hantek 6052BE/6082BE
USER’S MANUAL 21
Hantek 6052BE/6082BE
USER’S MANUAL 22
Hantek 6052BE/6082BE
USER’S MANUAL 23
Hantek 6052BE/6082BE
USER’S MANUAL 24
Hantek 6052BE/6082BE
1. Select channel:
2. VLTS/DIV: Set the selected channel voltage range.
The user can also drag the trigger icon to adjust the trigger level and the trigger level will
be displayed on the bottom right corner.
USER’S MANUAL 25
Hantek 6052BE/6082BE
2. Trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE.
3. Trigger Source: Selects the trigger source to CH1, CH2, ALT, EXT.
4. Trigger Slope: Selects the edge trigger slope to Positive or Negative slope
USER’S MANUAL 26
Hantek 6052BE/6082BE
Other Connector:
USER’S MANUAL 27
Hantek 6052BE/6082BE
Chapter 3 Functions
Set Oscilloscope
Measure Signal
Utility Function
Save/Load
USER’S MANUAL 28
Hantek 6052BE/6082BE
The oscilloscope will change the current settings to display this signal.
Save Setup
The oscilloscope software saves the current setup before you close the oscilloscope
software. The oscilloscope recalls this setup when you run the software next time. You
can use the “Save Setup” menu to permanently save setting.
Load Setup
The oscilloscope can recall the last setup before the oscilloscope software was running,
any saved setups, or the factory setup. You can use the “Load Setup” menu to
permanently recall setup.
Factory Setup
The oscilloscope software is set up for normal operation when it is shipped from the
factory. This is the factory setup. Push the “Factory Setup” menu to recall this setup.
USER’S MANUAL 29
Hantek 6052BE/6082BE
USER’S MANUAL 30
Hantek 6052BE/6082BE
Change Volt/DIV
USER’S MANUAL 31
Hantek 6052BE/6082BE
You cam left click and drag the mouse on the green knob or the yellow knob to change the
voltage.
You can set the coupling to DC, AC or GND. If you set the coupling to DC, it blocks the AC
component of the input signal.
Select the attenuation factor for the probe. To check the probe attenuation setting, toggle
the probe menu to match the attenuation factor of the probe.
USER’S MANUAL 32
Hantek 6052BE/6082BE
Caution: The attenuation factor changes the vertical scale of the oscilloscope so that the
measurement results refect the actual voltage levels at the probe tip.
Invert
The invert function turns the displayed waveform 180 degrees, with respect to the ground
level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
Turn on Invert:
USER’S MANUAL 33
Hantek 6052BE/6082BE
The oscilloscope is set to full bandwidth and will pass the high frequency component in the
signal if the “BW Limit” was turned off.
USER’S MANUAL 34
Hantek 6052BE/6082BE
The oscilloscope will reject the frequency component higher than 20MHz if the “BW
Limit” was turned on.
When the “BW Limit” was turned on, a “B” sign will be displayed at the bottom of display
screen.
Filter definition:
None: without the Filter, read out the actual value directly.
Lower: with low limitation filter, the waveform will look more thin than in “None” mode.
Higher: with high limitation Filter, the waveform will look more thin than in “Lower” mode.
USER’S MANUAL 35
Hantek 6052BE/6082BE
Source A/B:
Operate:
Four Types:
A+B Add source A and source B
A-B Subtract source B from source A
A×B source A Multiply source B
A/B Division source A by source B
In this function, use the addition, subtraction, multiplication and division function to
operate and analyze the waveform.
Select the operate type in the Operate menu. Select source A and B. Then adjust the
vertical scale and offset to view the math channel clearly. The mathematic result can be
measured by the measure and the cursor.
USER’S MANUAL 36
Hantek 6052BE/6082BE
USER’S MANUAL 37
Hantek 6052BE/6082BE
Caution: If you turn on the “Reference” channel, the load file window will appear.
USER’S MANUAL 38
Hantek 6052BE/6082BE
Select the horizontal Time/Div (scale factor) for the main or the window time base.
USER’S MANUAL 39
Hantek 6052BE/6082BE
In the “Format” item, set the waveform display format (Y-T, X-Y and Roll).
Y -T: Shows the relative relation between vertical voltage and horizontal time.
Roll: In Roll Mode, the waveform display updates from right to left.
X -Y: Show CH1 value at X axis; CH2 value at Y axis.
In Roll mode, the waveform display rolls from right to left. No trigger or horizontal offset
control of waveforms is available during Roll Mode, and it’s only available when set to
1s/div or slower.
Caution: If the time/div bigger than 1s, the format will change to Roll mode automatically.
Click “Zero” to set the trigger point to the horizontal center of the screen.
Horizontal position changes the displayed waveform position, relative to the trigger point.
The user can drag on screen to change the horizontal position.
USER’S MANUAL 40
Hantek 6052BE/6082BE
The trigger determines when the oscilloscope starts to acquire data and display a
waveform. When a trigger is set up properly, it can convert unstable displays or blank
screens into meaningful waveforms.
If the oscilloscope wants to acquire a waveform, it collects enough data so that it can draw
the waveform to the left of the trigger point. The oscilloscope continues to acquire data
while waiting for the trigger condition to occur. The oscilloscope continues to acquire
enough data so that it can draw the waveform to the right of the trigger point after it
detects a trigger.
The Edge trigger determines whether the scope finds the trigger point on the rising or the
falling edge of the signal. Select Edge trigger mode to trigger on Rising edge or Falling
edge.
USER’S MANUAL 41
Hantek 6052BE/6082BE
Source: You can use the trigger source options to select the signal that the oscilloscope
uses as a trigger. The source can be any signal connected to a channel BNC, or to the
EXT. BNC.
50%: Sets the trigger level to the center of the waveform amplitude.
The user can also change the trigger setting on trigger panel in sidebar.
USER’S MANUAL 42
Hantek 6052BE/6082BE
3.4.2 HF Rejection
The user can turn on “HF Rejection” to eliminate trigger higher-frequency (20M above).
USER’S MANUAL 43
Hantek 6052BE/6082BE
3.5 Save/Load
Save:
1. Save CSV
Save waveform data as a CSV file
2. Save Image
Save the software display window as a .bmp or .jpg file
3. Save Setups
Save the current oscilloscope setups to file
4. Save Waveform
Save waveform data as a TEST, WORD or EXCEL file.
Load:
1. Load CSV
Load the waveform that had saved as a CSV file
2. Load Setups
Load the instrument that had saved
USER’S MANUAL 44
Hantek 6052BE/6082BE
Click the “Menu -> Utility” to get into the “Utility” menu.
FFT
Record
Pass/Fail
Factory Setup
Language
USER’S MANUAL 45
Hantek 6052BE/6082BE
To use the FFT mode, you need to click “Menu->Utility->FFT” to open the FFT setup
window.
You can select the Source channel, Window algorithm, FFT number, and FFT Zoom
factor. It displays only one FFT spectrum at a time.
Source:
Selects the channel used as the FFT source
Window:
Selects the FFT window type
USER’S MANUAL 46
Hantek 6052BE/6082BE
Scale:
Selects the vertical scale units
Vertical:
Selects the vertical scale
Horizontal:
Selects the FFT zoom factor
Sets horizontal scale, If the information is more than the quantity displayed to the display
window, more information is displayed by using scroll bar.
You can click the “Start” button to start FFT math, also you can click this button to stop
FFT math.
Click “Save” button if you want to save FFT math result. You can save the data into Word
file, Excel file or Text file.
USER’S MANUAL 47
Hantek 6052BE/6082BE
USER’S MANUAL 48
Hantek 6052BE/6082BE
The Record window will display. The following picture shows the Record Interface.
This function can record input waveform form CH1 or CH2. The maximum record length is
1000 frames.
USER’S MANUAL 49
Hantek 6052BE/6082BE
Source:
Select record source channel. (CH1 or CH2)
End Frame:
Set the number of record times. The max frames are 1000.
Time Interval:
Record:
“Start” button:
Start to record frames. After you start to record waveforms, this button changes to
“Save” button:
Time Interval:
Start Frame:
USER’S MANUAL 50
Hantek 6052BE/6082BE
End Frame:
Current Frame:
Displays the current frame of play back. You can also change this number to watch
Repeat:
“Play” button:
Click this button to start playing back waveform. It can stop playing back if you
“Load” button:
Note: When it plays back waveform, the other channel will be turned off.
3.6.3 Pass/Fail
USER’S MANUAL 51
Hantek 6052BE/6082BE
The Pass/Fail function monitors changes of signals and outputs pass or fail signals by
comparing the input signal with the pre-created mask.
Control Setting
Source:
Output:
USER’S MANUAL 52
Hantek 6052BE/6082BE
Mask Setting
Vertical:
Horizontal:
“Create” button:
“Save” button:
“Load” button:
Information Display
Operation
USER’S MANUAL 53
Hantek 6052BE/6082BE
When you click the Factory Setup in Utility menu, the oscilloscope displays the CH1
and CH2 waveforms and removes all other waveforms.
The oscilloscope set up for normal operation when it is shipped from the factory and can
be recalled at anytime by user.
USER’S MANUAL 54
Hantek 6052BE/6082BE
The Factory Setup function does not reset the following settings:
Language option
Date and time
3.6.5 Language
There are four languages in “Language” menu. The default language is English.
3.7.1 Cursor
1. Source
The user can set the source to CH1, CH2, FFT and MATH. But the FFT source can only
USER’S MANUAL 55
Hantek 6052BE/6082BE
2. Type
There are four types of cursors: Cross, Trace, Vertical and Horizontal.
1) Cross
The Cross cursors appear as cross lines on the display and measure the vertical and
horizontal parameters.
The Cross cursor display window:
2) Trace
The Trace cursors appear as vertical lines on the display and measure the waveform
amplitude at the point the waveform crosses the cursor.
USER’S MANUAL 56
Hantek 6052BE/6082BE
3) Vertical
The Vertical cursors appear as vertical lines on the display and measure the vertical
parameters.
The Vertical cursor display window:
USER’S MANUAL 57
Hantek 6052BE/6082BE
4) Horizontal
The Horizontal cursors appear as horizontal lines on the display and measure the
horizontal parameters.
The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time
measurements).
USER’S MANUAL 58
Hantek 6052BE/6082BE
1. Source
The user can use the “Source” menu to select a measure source.
2. Vertical
Maximum: Voltage of the absolute maximum level, measured over the entire waveform
Minimum: Voltage of the absolute minimum level, measured over the entire waveform
Peak to Peak: Peak- to-peak=Max-Min, measured over the entire waveform
Top: Voltage of the statistical maximum level, Measured over the entire waveform
Base: Voltage of the statistical maximum level, measured over the entire waveform
Middle: Voltage of the 50%level from base to top
RMS:T he Root Mean Square voltage over the entire waveform
Amplitude: Amp=Base-Top, measured over the entire waveform
Mean: The arithmetic mean over the entire waveform
Cycle Mean: The arithmetic mean over the first cycle in the waveform
Preshoot: Positive Overshoot =(Max-Top)/Amp*100% , measured over the entire
waveform
USER’S MANUAL 59
Hantek 6052BE/6082BE
3. Horizontal
Period: Time to take for the signal cycle to complete in the waveform
Frequency: Reciprocal of the period of the first cycle in the waveform
Rise Time: Time taken from lower threshold to upper threshold
Fall Time: Time taken from upper threshold to lower threshold
+Duty Cycle: Positive Duty Cycle = (Positive Pulse Width)/Period x 100%,Measured of
the first cycle in the waveform
-Duty Cycle: Negative Duty Cycle = (Negative Pulse Width)/Period x 100%, Measuredof
the first cycle in the waveform
+Pulse Width: Measured of the first positive pulse in the waveform. The time between the
50% amplitude points
-Pulse Width: Measured of the first negative pulse in the waveform. The time between the
50% amplitude points
USER’S MANUAL 60
Hantek 6052BE/6082BE
4. Clear Measure
Caution: The results of the automatic measurements will be displayed on the bottom of
the screen. Maximum 8 results could be displayed at the same time. When there is no
room, the next new measurement result will make the previous results moving left, out of
screen.
USER’S MANUAL 61
Hantek 6052BE/6082BE
If the Vectors type mode is selected, the waveform will be displayed as following figure.
If the Dots type mode is selected, the waveform will be displayed as following figure.
USER’S MANUAL 62
Hantek 6052BE/6082BE
USER’S MANUAL 63
Hantek 6052BE/6082BE
USER’S MANUAL 64
Hantek 6052BE/6082BE
3.8.3 Intensity
You can change the grid and waveform color intensity in this dialog.
USER’S MANUAL 65
Hantek 6052BE/6082BE
The following figure shows the waveform color dialog. It shows the waveform color setting.
If you click the color rectangle, the color dialog will be showed.
USER’S MANUAL 66
Hantek 6052BE/6082BE
The software will stop updating waveform after the user clicked “Stop” button, The user
can change the waveform display by adjusting the scale and position. When you change
the scale, the waveform display will increase or decrease in size. When you change the
position, the waveform will move up, down, right, or left.
Zoom In/Out
The user can click “Zoom In/Out” in Acquire menu, then left or right click the mouse
button on display screen to Zoom in/out the waveform. Also the user can change
Time/Div in Horizontal menu or in Horizontal panel to zoom in/out the waveform.
Drag
The user can modify the waveform position after clicked “Drag” in “Acquire” menu
following the following steps.
USER’S MANUAL 67
Hantek 6052BE/6082BE
3.10 Interpolation
At the time 40ns/div or faster, user can use the different interpolation mode to get
waveforms of different smoothness.
The Step Interpolation:
USER’S MANUAL 68
Hantek 6052BE/6082BE
USER’S MANUAL 69
Hantek 6052BE/6082BE
When you acquire a signal, the oscilloscope converts it into a digital form and displays a
waveform. The acquisition mode defines how the signal is digitized and the time base
setting affects the time span and level of detail in the acquisition.
Normal: In this acquisition mode, the oscilloscope samples the signal in evenly spaced
intervals to construct waveform.
Average: In this acquisition mode the oscilloscope acquires several waveforms, averages
them, and display the resulting waveform. You can use this mode to reduce random noise.
Click “Print” in “File” menu to set the printer to print the current waveform.
USER’S MANUAL 70
Hantek 6052BE/6082BE
USER’S MANUAL 71
Hantek 6052BE/6082BE
Sample Measurement
Pass/Fail Test
USER’S MANUAL 72
Hantek 6052BE/6082BE
To measure the frequency and “Vpp”, you can do these steps as follows:
2. Click the “Measure”-> “Vertical”-> “Peak-to-Peak” button, the “Vpp” of the signal
will also display on the bottom of the waveform interface.
3. To clear the measurement on the waveform interface, click the “Measure”-> “Clear
Measure” button.
USER’S MANUAL 73
Hantek 6052BE/6082BE
The Pass/Fail function monitors changes of signals and outputs pass or fail signals by
comparing the input signal with the pre-created mask.
Control Setting:
Source:
Output:
Select the Pass/Fail output condition.
Mask Setting
USER’S MANUAL 74
Hantek 6052BE/6082BE
Vertical:
Horizontal:
“Create” button:
Click this button to create Pass/Fail area according to the mask.
“Save” button:
“Load” button:
Click this button to load the saved setups file.
Information Display:
Operation:
USER’S MANUAL 75
Hantek 6052BE/6082BE
To capture a single event, it needs to gather some pre-test knowledge of the signal in
order to set up the trigger level and slope correctly. For example, if the event is derived
from 3.3V COMS logic, a trigger level of 1.2 or higher Volts should work on a rising edge.
Do these steps as follows:
1. Set the probe and the channel attenuations to X 10.
2. Set up the trigger in the Trigger Menu, or in the Trigger Setting window.
(1) Adjust the Trigger Mode to Edge.
(2) Set the Trigger Sweep to Single.
(3) Set the Trigger Source to CH1.
(4) Set the Trigger Slope to “+” which means you select the rising edge.
(5) Adjust the Volts/Div and the time base in a proper range for the signal.
(6) Drag the trigger level sign on the waveform display screen to proper position. It ally
higher a little above the normal level.
(7) Click START button to start capturing. When the trigger conditions are met, data
appears on the display representing the data points that the oscilloscope obtained with
one acquisition.
This function helps to capture the signal occurrence easily, such as the noise with large
amplitude; set the trigger level higher a little above the normal level and press and wait.
When noise occurs, the instrument will record the waveform before and after the trigger.
USER’S MANUAL 76
Hantek 6052BE/6082BE
X-Y Plot acts to analyze correlation of data of two channels. Lissajous diagram is
displayed in the screen when you use X-Y Plot, which enables to compare frequencies,
amplitudes and phases of counterpart waveform against the reference waveform. This
makes it possible to compare and analyze frequency, amplitude and phase between input
and output.
Do these steps as follows:
1. Set the probe attenuation to “x10”. Set the switch to “x10” on the probes.
2. Connect the CH1 probe to the input of the circuit, and connect the CH2 probe to the
output of the circuit.
3. Click button.
4. Adjust the vertical scale and offset to display approximately the same amplitude
signals on each channel.
5. Select X-Y format at Horizontal window. The oscilloscope will displays a Lissajous
pattern representing the input and the output characteristics of the circuit.
6. Adjust the scale and offset of the horizontal and vertical to a desirable waveform
display. The following picture shows a typical example.
7. Apply the Ellipse Method to observe the phase difference between the two channels.
USER’S MANUAL 77
Hantek 6052BE/6082BE
Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals.
From the formula above:
θ = ±arcsine (A/B) or ±arcsine (C/D)
θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I
and III quadrant .If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π)
or (π~3π/2).
5. Release the left mouse button, the frequency difference and time difference will be
shown at the status bar.
USER’S MANUAL 78
Hantek 6052BE/6082BE
5. Release the left mouse button, the voltage difference will be shown at the status bar.
USER’S MANUAL 79
Hantek 6052BE/6082BE
Do these steps:
1. Click “Cursor->Source”, select CH1 (select CH2 if you want trace CH2).
3. Click the cursor at the position that you want traced of the wave in the waveform
window.
USER’S MANUAL 80
Hantek 6052BE/6082BE
Caution: Click “Cursor->Type” to select “Cross”, you can measure time and amplitude
at one time.
USER’S MANUAL 81
Hantek 6052BE/6082BE
Chapter 5 Appendix
Appendix A: Specifications
USER’S MANUAL 82
Hantek 6052BE/6082BE
Appendix A: Specifications
Specifications Table:
Input
Real-time sampling:
Max. sample rate 150MS/s using one channel(Hantek6052BE)
250MS/s using one channel(Hantek6082BE)
Channels 2
50 MHz analog bandwidth(Hantek6052BE)
Bandwidth
80MHz analog bandwidth(Hantek6082BE)
Vertical resolution 8 bits/channel
Gain range 10mV ~5V/div @ x1 probe(10mV, 20mV, 50mV, 100mV, 200mV,
500mV, 1V, 2V, 5V/div, 1,2,5 sequence)
100mV ~ 50V/div @ x10 probe
1V ~ 500V/div @ x100 probe
10V ~ 5KV/div @ x1000 probe
Range 8 divisions
Offset level +/-4 divisions
Coupling AC, DC,GND
Offset increments 0.02 div
Impedance 1M ohm
DC accuracy +/-3%
Input protection 35Vpk (DC + peak AC < 10 kHz, without external attenuation)
Display Mode Y-T, X-Y and Roll
Timebase
4ns/div ~ 1h/div(4ns, 10ns, 20ns, 40ns, 100ns, 200ns, 400ns,
1us, 2us, 4us, 10us, 20us, 40us, 100us, 200us, 400us, 1ms, 2ms,
Timebase range
4ms, 10ms, 20ms, 40ms, 100ms, 200ms, 400ms, 1s,2s, 4s,
10s,20s,40s,10m,20m,40m,1h /div 1-2-4 sequence)
Real-time sampling: 4ns/div ~ 400ms/div.
Acquisition mode
Roll mode: 1s/div ~ 1h/div
Range 10 divisions
Buffer Size 10K ~ 64KB /CH (Hantek6052BE, Hantek6082BE)
Trigger
Type Edge trigger
Mode Auto, Normal and Single
Autoset Yes
Source CH1, CH2, ALT, EXT, EXT/10
Range 10 divisions
Trigger level +/-4 divisions
USER’S MANUAL 83
Hantek 6052BE/6082BE
USER’S MANUAL 84
Hantek 6052BE/6082BE
General Care
Do not store or leave the oscilloscope where the device will be exposed to direct sunlight
for long periods of time.
Caution
To avoid damages to the device or probes, do not expose them to sprays, liquids or
solvents.
To avoid damages to the surface of the device or probes not use any abrasive or chemical
cleaning agents.
Cleaning
In order to keep the instrument clean ,you need to check often whether the channels are
dusty or not .Please clean the instrument surface and the channel connecting point as
per the following cautions(Make sure the device disconnect form all power sources):
1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth.
USER’S MANUAL 85