0% found this document useful (0 votes)
64 views

User'S Manual: PC Usb Digital Oscilloscope

This document is the user manual for the Hantek 6052BE/6082BE PC USB digital oscilloscope. It provides instructions on installing the software and drivers, performing basic setup and calibration of the oscilloscope, and an overview of the oscilloscope's user interface and functions. The manual also includes chapters on operating basics, functions, application examples, specifications, and maintenance.

Uploaded by

Slaven Ivanovic
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
64 views

User'S Manual: PC Usb Digital Oscilloscope

This document is the user manual for the Hantek 6052BE/6082BE PC USB digital oscilloscope. It provides instructions on installing the software and drivers, performing basic setup and calibration of the oscilloscope, and an overview of the oscilloscope's user interface and functions. The manual also includes chapters on operating basics, functions, application examples, specifications, and maintenance.

Uploaded by

Slaven Ivanovic
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 87

PC USB DIGITAL

OSCILLOSCOPE

USER’S MANUAL

Hantek 6052BE/6082BE

www.hantek.com
Content
General Safety Summary ............................................................................... 1 
Chapter 1 Getting Start .................................................................................. 3 
1.1 System Requirement ......................................................................................................... 4 
1.2 Install Software ................................................................................................................. 5 
1.3 Install Drive....................................................................................................................... 8 
1.4 General Check ................................................................................................................. 13 
1.5 Probe Compensation ....................................................................................................... 13 
1.6 Self Calibration ............................................................................................................... 15 
1.7 Function Check ............................................................................................................... 15 
1.8 Accessories...................................................................................................................... 17 
Chapter 2 Operating Basics ......................................................................... 18 
2.1 The User’s Interface ........................................................................................................ 19 
2.2 The Menu System ........................................................................................................... 21 
2.3 The Horizontal System .................................................................................................... 24 
2.4 The Vertical System ........................................................................................................ 24 
2.5 The Trigger System ......................................................................................................... 25 
2.6 Input Connectors ............................................................................................................. 27 
Chapter 3 Functions ..................................................................................... 28 
3.1 Setup the Oscilloscope .................................................................................................... 29 
3.2 Set Vertical System ......................................................................................................... 30 
3.3 Set Horizontal System ..................................................................................................... 39 
3.4 Set Trigger System .......................................................................................................... 41 
3.5 Save/Load........................................................................................................................ 44 
3.6 Utility Function ............................................................................................................... 45 
3.7 Measure Signal ................................................................................................................ 55 
3.8 The Display System ........................................................................................................ 61 
3.9 Zoom In/Out and Drag Waveforms ................................................................................ 67 
3.11 Acquisition Modes......................................................................................................... 70 
3.12 Print And Print Preview ................................................................................................ 70 
Chapter 4 Application Example .................................................................... 72 
4.1 Simple Measurement ....................................................................................................... 73 
4.2 Pass/Fail Test .................................................................................................................. 74 
4.3 Capturing a Single-Shot Signal ....................................................................................... 76 
4.4 The Application of the X-Y Operation ........................................................................... 77 
Chapter 5 Appendix ..................................................................................... 82 
Appendix A: Specifications .................................................................................................. 83 
Appendix B: General Maintenance ....................................................................................... 85 
Hantek 6052BE/6082BE

General Safety Summary


Review the following safety precautions carefully before operate the device to avoid any
personal injuries or damages to the device and any products connected to it. To avoid
potential hazards use the device as specified by this user’s guide only.

‹ To Avoid Fire or Personal Injury

‹ Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.

‹ Connect and Disconnect Properly. Do not connect or disconnect probes or test


leads while they are connected to a voltage source.

‹ Connect and Disconnect Properly. Connect the probe output to the measurement
device before connecting the probe to the circuit under test. Disconnect the probe
input and the probe reference lead from the circuit under test before disconnecting
the probe from the measurement device.

‹ Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.

‹ Use Proper Probe. To avoid shock hazard, use a properly rated probe for your
measurement.

‹ Avoid Circuit or Wire Exposure. Do not touch exposed connections and


components when power is on.

‹ Do Not Operate With Suspected Failures. If suspected damage occurs with the
device, have it inspected by qualified service personnel before further operations.

‹ Provide Proper Ventilation. Refer to the installation instructions for proper


ventilation of the device.

‹ Do not operate in Wet/Damp Conditions.

‹ Do not operate in an Explosive Atmosphere.

‹ Keep Product Surfaces Clean and Dry.

USER’S MANUAL 1
Hantek 6052BE/6082BE

Safety Information
In order to ensure the correct using and the best efficient service, please carefully read the
user’s manual.
System Symbols
These symbols may appear in this manual or on the instrument.

Warning: “Waring” identifies conditions and actions that pose hazards to


the users.

Caution: “Caution” identifies conditions and actions that may damage the
product or other properties.

DANGER: Refer to Protective Chassis grand Earth(grand)


High Voltage the manual Condutor Terminal Terminal

System Terms
The following terms may appear on the instrument:
Danger: The term “Danger” is used in the manual to indicate that when you read this
mark, personal injury may be caused to immediately.
Waning: The term “Danger” is used in the manual to indicate that when you read this
mark, personal injury may not be caused to you immediately, but you need to be
cautionary.
Notice: The term “Noise” is used in this manual to indicate that damages may be
caused on this product or other properties.

USER’S MANUAL 2
Hantek 6052BE/6082BE

Chapter 1 Getting Start


The oscilloscope is small, lightweight, no external power required, portable oscilloscopes.
The oscilloscopes is ideal for production test, research and design and all of the
applications involving analog circuits test and troubleshooting, as well as education and
training.

In addition to the list of general features on the next page, this chapter describes how to
do the following tasks:

‹ System Requirements

‹ Install software

‹ Install Drive

‹ General Check

‹ probe compensation

‹ self calibration

‹ Accessories

USER’S MANUAL 3
Hantek 6052BE/6082BE

1.1 System Requirement


To run the oscilloscope software, the needs of computer configuration are as follows:

Minimum System Requirements

Operating System
Windows XP/Vista/Win 7

Processor
Upwards of 1.00G processor

Memory
256M byte

Disk Space
500M disk free space

Screen resolution
800 x 600

Recommended Configuration

Operating System
Windows XP SP3 System

Processor
2.4G Processor

Memory
1G Byte Memory

Disk Space
80G Disk Space

Screen resolution
1024 x 768 or 1280 x 1024 resolution

DPI Setting
Normal Size (96DPI)

USER’S MANUAL 4
Hantek 6052BE/6082BE

1.2 Install Software

Caution: You must install the software before using the oscilloscope.

1. While in Windows, insert the installation CD into the CD-ROM drive.


2. The installation should start up automatically. Otherwise in Windows Explorer, switch
to the CD-ROM driver and run Setup.exe.

3. The software Installation is started. Click 'Next' to continue.

USER’S MANUAL 5
Hantek 6052BE/6082BE

4. Choose a destination directory. Click ‘Next’ to continue.

5. Check the setup information. Click Next to start copying of files.

6. This Status dialog is displayed during copying of files.

USER’S MANUAL 6
Hantek 6052BE/6082BE

7. Updating your system configuration.

8. The installation is complete.

USER’S MANUAL 7
Hantek 6052BE/6082BE

1.3 Install Drive

Example: Hantek6052BE

1. Connect the A-Type Plug of USB cable to your PC’s USB port.

2. Connect the B-Type Plug of USB cable to Hantek6052BE USB port.

3. New hardware is found.

USER’S MANUAL 8
Hantek 6052BE/6082BE

4. New hardware search wizard starts.

5. New hardware search wizard starts to search the drive.

USER’S MANUAL 9
Hantek 6052BE/6082BE

6. New hardware wizard installs software

7. Finish new hardware search wizard.

USER’S MANUAL 10
Hantek 6052BE/6082BE

8. New hardware is found.

9. Found new hardware Wizard.

USER’S MANUAL 11
Hantek 6052BE/6082BE

10. New hardware search wizard starts to search.

11. New hardware wizard installs software and finish new hardware search wizard.

USER’S MANUAL 12
Hantek 6052BE/6082BE

12. Your new hardware is installed and ready to use.

1.4 General Check

Please check the instrument as following steps after receiving an oscilloscope:

Check the shipping container for damage:


Keep the damaged shipping container or cushioning material until the contents of the
shipment have been checked for completeness and the instrument has been checked
mechanically and electrically.

Check the accessories:


Accessories supplied with the instrument are listed in "Accessories" in this guide. If the
contents are incomplete or damaged, please notify the franchiser.

Check the instrument:


In case there is any mechanical damage or defect, or the instrument does not operate
properly or fails performance tests, please notify the franchiser.

1.5 Probe Compensation

Perform this function to match the characteristics of the probe and the channel input.
This should be performed whenever attaching a probe to any input channel at the first
time.

‹ From the “Probe” menu, select attenuation to 1:10. Set the switch to “X10” on the
probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, insert
the tip onto the probe firmly to ensure a proper connection.

‹ Attach the probe tip to the Probe Compensator and the reference lead to the
ground connector, select CH1, and then press the “AUTOSET” button into the menu or
the toolbar.

‹ Check the shape of the displayed waveform.

USER’S MANUAL 13
Hantek 6052BE/6082BE

Correctly Compensated

Over compensated

Under Compensated
‹ If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe
for the fattest square wave being displayed on the oscilloscope.

‹ Repeat if necessary.

WARNNING: To avoid electric shock while using the probe, be sure the perfection of the
insulated cable, and do not touch the metallic portions of the probe head while it is
connected with a voltage source.

USER’S MANUAL 14
Hantek 6052BE/6082BE

1.6 Self Calibration

The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5v or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the “Utility -> Calibration” option and follow the directions on the screen.
The self calibration routine takes about several minutes.

1.7 Function Check

Perform this functional check to verify that your oscilloscope is operating correctly.

‹ Connect the oscilloscope


You should connect the A-Type Plug of USB cable to your PC USB port and connect the
B-Type Plug of USB cable to oscilloscope USB port.

‹ Input a signal to a channel of the oscilloscope


The oscilloscope is equipped with two channels plus external trigger.

Please input signal in the following steps:

1. Set the attenuation switch on the probe as 10X and connect the probe on the
oscilloscope with CH1. Aim the slot in the probe connector at the faucet on BNC of CH1
and insert, then, turn right to lock the probe. Finally, attach the tip of probe and ground nip
to the Connector of Probe compensator.

USER’S MANUAL 15
Hantek 6052BE/6082BE

2. Set the CH1 probe attenuation of the oscilloscope to X10. (The default is X1).

3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click

the button. A square wave will be displayed within a several seconds.

(Approximately 1 kHz, 2V, peak- to- peak).

4. Inspect CH2 with the same method. Repeat steps 2 and 3.

USER’S MANUAL 16
Hantek 6052BE/6082BE

1.8 Accessories
All the accessories listed below are standard accessories for the oscilloscope:
‹ Probe x 2 (1.5m), 1:1, (10:1) Passive Probes

‹ A User’s Guide

‹ An USB Cable

‹ A PC software of the oscilloscope

USER’S MANUAL 17
Hantek 6052BE/6082BE

Chapter 2 Operating Basics

‹ The User’s Interface

‹ The Menu System

‹ The Horizontal System

‹ The Vertical System

‹ The Trigger System

‹ Input Connectors

USER’S MANUAL 18
Hantek 6052BE/6082BE

2.1 The User’s Interface

Click the software icon on the desk after you finished the software setting and equipment
connecting. Then a user interface will be showed as follows:

1. The Main Menu


All setting can be found in the main menu.
2. The Toolbar
3. Displays the trigger information
Displays the edge trigger slope, source and level.
4. The Horizontal Panel
The user can change Time/Div, format in the panel.
5. The vertical Panel
The user can turn on/off the CH1/CH2.Also the user can change the CH1/CH2 volt/div,
coupling and probe attenuation.
6. The Trigger Panel
In this panel, the user can change the trigger mode, sweep, source and slope.
7. Displays the system time.
8. Displays the main time base setting.
9. Marker shows Edge trigger level.

USER’S MANUAL 19
Hantek 6052BE/6082BE

10. Displays the CH2 information


Readout show the coupling of the channels.
Readout show the vertical scale factors of the channels.
A “B” icon indicates that the channel is bandwidth.
11. Displays the CH1 information
Readout show the coupling of the channels.
Readout show the vertical scale factors of the channels.
A “B” icon indicates that the channel is bandwidth.
12. The marker shows the reference points of the displayed waveforms.
If there is no marker, the channel is not displayed.
13. The same as 13.
14. A window that shows the display waveform in buffer position.
15. Marker shows horizontal trigger position.
16. Trigger status indicates the following.
Auto: The oscilloscope is in auto mode and is acquiring waveform waveform in the
absence of triggers.
Trig’D: The oscilloscope has seen a trigger and is acquiring the post trigger data.
WAIT: All pretrigger data has been acquired and the oscilloscope is ready to accept a
trigger.
Stop: The oscilloscope has stopped acquiring waveform data.
Run: The oscilloscope is running.
Play: The oscilloscope is displaying the record waveform.

USER’S MANUAL 20
Hantek 6052BE/6082BE

2.2 The Menu System

The Main Menu:

1. File: Load or save CSV, waveform and setup

2. View: Change the user interface

3. Channel: Set the vertical system

4. Trigger: Change trigger Mode, Sweep, Source, Slope and so on.

USER’S MANUAL 21
Hantek 6052BE/6082BE

5. Horizontal: Sets time base parameters and format parameters.

6. Cursor: Set cursor measure type

7. Measure: Set measurement parameters

8. Display: Change wave display type and color.

9. Acquire: Start, Stop, Autoset and other operation setting

USER’S MANUAL 22
Hantek 6052BE/6082BE

10. Utility: Utility setting

11. Help: Turn on help file

USER’S MANUAL 23
Hantek 6052BE/6082BE

2.3 The Horizontal System

Click “Horizontal” in main menu.


The following figure shows the horizontal parameters setting.

1. Time/DIV: leads the setting of the time base parameters


2. Format: leads the setting of the horizontal format parameters
Of course, the user can also set the parameters in the right sidebar.

2.4 The Vertical System

Click “Channel” in main menu.


The following figure shows the vertical parameters setting.

‹ CH1: Leads to the setting of the Ch1 parameters


‹ Ch2: Leads to the setting of the Ch2 parameters
‹ Math: Leads to the setting of the Math parameters
‹ Reference: Leads to the setting of the Reference parameters

You can also set parameters in the right sidebar.

USER’S MANUAL 24
Hantek 6052BE/6082BE

1. Select channel:
2. VLTS/DIV: Set the selected channel voltage range.

3. Coupling: Set the selected channel to DC, AC and GND.

4. BW Limit: Reject the frequency component higher than 20MHz.

2.5 The Trigger System

Click “Trigger” in main menu.


The following figure shows the trigger system control.

The user can also drag the trigger icon to adjust the trigger level and the trigger level will
be displayed on the bottom right corner.

USER’S MANUAL 25
Hantek 6052BE/6082BE

1. Trigger Mode: Sets the trigger mode.

2. Trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE.

3. Trigger Source: Selects the trigger source to CH1, CH2, ALT, EXT.

(Note: Hantek6082BE: CH1, CH2, ALT, EXT., EXT./10)

4. Trigger Slope: Selects the edge trigger slope to Positive or Negative slope

USER’S MANUAL 26
Hantek 6052BE/6082BE

2.6 Input Connectors

CH 1/CH 2: Input connectors for waveform display.


EXT.: Input connector for an external trigger source. Use the Trigger Menu to select the
Ext. source.

Other Connector:

GND.: Ground terminal


USB PORT: Connect the B-Type Plug of USB cable to this port.
CAL.: Probe compensation output.

USER’S MANUAL 27
Hantek 6052BE/6082BE

Chapter 3 Functions

‹ Set Oscilloscope

‹ Set Vertical System

‹ Set Horizontal System

‹ Set Trigger System

‹ Measure Signal

‹ Utility Function

‹ The Display System

‹ Save/Load

‹ Print

USER’S MANUAL 28
Hantek 6052BE/6082BE

3.1 Setup the Oscilloscope

Use “AUOSET” to display a signal automatically.


Auto setup functions one time each time you push the “AUTOSET” button. The function
obtains a stable waveform display for you. It automatically adjusts the vertical scale,
horizontal scale and trigger settings. Auto setup also displays several automatic
measurements in the graticule area, depending on the signal type.

Connect a signal to the CH1 input:

1. Connect a signal to the oscilloscope as described above.

2. Click the “Acquire -> Autoset” button.

The oscilloscope will change the current settings to display this signal.

Save Setup

The oscilloscope software saves the current setup before you close the oscilloscope
software. The oscilloscope recalls this setup when you run the software next time. You
can use the “Save Setup” menu to permanently save setting.

Load Setup

The oscilloscope can recall the last setup before the oscilloscope software was running,
any saved setups, or the factory setup. You can use the “Load Setup” menu to
permanently recall setup.

Factory Setup

The oscilloscope software is set up for normal operation when it is shipped from the
factory. This is the factory setup. Push the “Factory Setup” menu to recall this setup.

USER’S MANUAL 29
Hantek 6052BE/6082BE

3.2 Set Vertical System

Click “Channel” in main menu.

3.2.1 Change CH1/CH2


Click “CH1/CH2”in “Channel” menu.

The CH1 Menu:

The CH2 Menu:

USER’S MANUAL 30
Hantek 6052BE/6082BE

The CH1/CH2 control panel in sidebar:

The Vertical function:

Turn ON/OFF: Turn on/off the channel


Volt/DIV: Select the channel voltage/div
Coupling: Select the channel coupling
Probe: Select the channel probe attenuation
Filter: Select software filter
Reset: Set the channel vertical position to zero
Invert: Turn on/off the invert function.
Bandwidth Limit: Limit the channel bandwidth to 20MHz to reduce noise.

Change Volt/DIV

You can click “Volt/Div” in “CH1/CH2” to select the voltage.

USER’S MANUAL 31
Hantek 6052BE/6082BE

You can also change “CH1/CH2” voltage in side bar.

You cam left click and drag the mouse on the green knob or the yellow knob to change the

voltage.

Set Channel Coupling

Click “Coupling” in “Channel” menu.

You can set the coupling to DC, AC or GND. If you set the coupling to DC, it blocks the AC
component of the input signal.

Probe Attenuation Setting

Select the attenuation factor for the probe. To check the probe attenuation setting, toggle
the probe menu to match the attenuation factor of the probe.

This setting remains in effect before you changed again.

Click “Probe” in Vertical Setup window to select the probe attenuation.

USER’S MANUAL 32
Hantek 6052BE/6082BE

The probe setting window in the sidebar:

Caution: The attenuation factor changes the vertical scale of the oscilloscope so that the
measurement results refect the actual voltage levels at the probe tip.

Invert

The invert function turns the displayed waveform 180 degrees, with respect to the ground
level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
Turn on Invert:

Turn off Invert:

USER’S MANUAL 33
Hantek 6052BE/6082BE

The following picture shows the waveform before inversion:

The following picture shows the waveform of inversion:

Set the Channel Bandwidth Limit

The oscilloscope is set to full bandwidth and will pass the high frequency component in the
signal if the “BW Limit” was turned off.

USER’S MANUAL 34
Hantek 6052BE/6082BE

The oscilloscope will reject the frequency component higher than 20MHz if the “BW
Limit” was turned on.

When the “BW Limit” was turned on, a “B” sign will be displayed at the bottom of display
screen.

Filter definition:
None: without the Filter, read out the actual value directly.
Lower: with low limitation filter, the waveform will look more thin than in “None” mode.
Higher: with high limitation Filter, the waveform will look more thin than in “Lower” mode.

3.2.2 Set Math

Click “MATH” in Channel menu to set MATH channel.


The MATH Menu:
‹ Turn On/Off: Turn On/Off the MATH Channel.
‹ Source A/B: Set the sources of the math channel.
‹ Operate: Set operates type of the math channel.
‹ Volt/DIV: Set the resolution of the math channel.
‹ Probe: Set the math channel probe attenuation.
‹ Reset: Set the math channel vertical position to zero.
‹ Invert: Turn on/off the invert function
The mathematic functions include addition, subtract, multiply and division for CH1 and
CH2.

USER’S MANUAL 35
Hantek 6052BE/6082BE

Source A/B:

Source A and Source B menu:

Operate:

Four Types:
A+B Add source A and source B
A-B Subtract source B from source A
A×B source A Multiply source B
A/B Division source A by source B

In this function, use the addition, subtraction, multiplication and division function to
operate and analyze the waveform.
Select the operate type in the Operate menu. Select source A and B. Then adjust the
vertical scale and offset to view the math channel clearly. The mathematic result can be
measured by the measure and the cursor.

USER’S MANUAL 36
Hantek 6052BE/6082BE

The Math Function Display:

3.3.3 Set Reference

Click “Reference” in “Channel” menu to set REF channel.

The Reference Channel Function:


Turn On/Off: Turn on/off the reference channel.
Volt/DIV: Channel the resolution of the reference channel.
Reset: Reset the reference channel vertical position to zero.
Load Reference: Load the reference waveform from the “.ref” file from your computer.
Save Reference: Save the current reference waveform to your computer as “.ref” format.
You can change the vertical scale of a waveform. The waveform display will contract or
expand relative to the reference level.
Click “Save Reference” to save the waveform to *.ref file. The saved source window
appears.

USER’S MANUAL 37
Hantek 6052BE/6082BE

The Reference Waveform Display Window:

Caution: If you turn on the “Reference” channel, the load file window will appear.

USER’S MANUAL 38
Hantek 6052BE/6082BE

3.3 Set Horizontal System

Click “Horizontal” in main menu.

3.3.1 Change Time/Div

The “Time/Div” Menu:

Select the horizontal Time/Div (scale factor) for the main or the window time base.

The Horizontal Panel

Click the blue knob can change Time/Div.

3.3.2 Change Format

Click “Format” menu.

USER’S MANUAL 39
Hantek 6052BE/6082BE

In the “Format” item, set the waveform display format (Y-T, X-Y and Roll).
Y -T: Shows the relative relation between vertical voltage and horizontal time.
Roll: In Roll Mode, the waveform display updates from right to left.
X -Y: Show CH1 value at X axis; CH2 value at Y axis.
In Roll mode, the waveform display rolls from right to left. No trigger or horizontal offset
control of waveforms is available during Roll Mode, and it’s only available when set to
1s/div or slower.
Caution: If the time/div bigger than 1s, the format will change to Roll mode automatically.

3.3.3 Change Horizontal Position

The “Zero” menu:

Click “Zero” to set the trigger point to the horizontal center of the screen.
Horizontal position changes the displayed waveform position, relative to the trigger point.
The user can drag on screen to change the horizontal position.

USER’S MANUAL 40
Hantek 6052BE/6082BE

3.4 Set Trigger System

3.4.1 Set Edge Trigger

Click “Trigger” in main menu:

The trigger determines when the oscilloscope starts to acquire data and display a
waveform. When a trigger is set up properly, it can convert unstable displays or blank
screens into meaningful waveforms.
If the oscilloscope wants to acquire a waveform, it collects enough data so that it can draw
the waveform to the left of the trigger point. The oscilloscope continues to acquire data
while waiting for the trigger condition to occur. The oscilloscope continues to acquire
enough data so that it can draw the waveform to the right of the trigger point after it
detects a trigger.
The Edge trigger determines whether the scope finds the trigger point on the rising or the
falling edge of the signal. Select Edge trigger mode to trigger on Rising edge or Falling
edge.

Mode: Select the trigger mode.

Sweep: Set the sweep mode to Auto, Normal or Single.


Auto: Acquire waveform even no trigger occurred
Normal: Acquire waveform when trigger occurred.
Single: Acquire waveform when trigger occurred then stop.

USER’S MANUAL 41
Hantek 6052BE/6082BE

Source: You can use the trigger source options to select the signal that the oscilloscope
uses as a trigger. The source can be any signal connected to a channel BNC, or to the
EXT. BNC.

CH1: Select CH1 as trigger signal


CH2: Select CH2 as trigger signal
ALT: Select CH1 and CH2 as trigger signal
EXT.: Select EXT. as trigger signal
EXT./10: Select attenuated EXT./10 as trigger signal (Only Hantek6082BE)

Slope: Set the slope to Rising (+) or Falling (-).

Rising: Trigger on rising edge


Falling: Trigger on falling edge

50%: Sets the trigger level to the center of the waveform amplitude.

The user can also change the trigger setting on trigger panel in sidebar.

USER’S MANUAL 42
Hantek 6052BE/6082BE

3.4.2 HF Rejection

Click “HF Rejection” in “Trigger” menu:

The user can turn on “HF Rejection” to eliminate trigger higher-frequency (20M above).

USER’S MANUAL 43
Hantek 6052BE/6082BE

3.5 Save/Load

Click “File” in main menu to save waveform, setups and screen.

Save:
1. Save CSV
Save waveform data as a CSV file

2. Save Image
Save the software display window as a .bmp or .jpg file

3. Save Setups
Save the current oscilloscope setups to file

4. Save Waveform
Save waveform data as a TEST, WORD or EXCEL file.

Load:
1. Load CSV
Load the waveform that had saved as a CSV file

2. Load Setups
Load the instrument that had saved

USER’S MANUAL 44
Hantek 6052BE/6082BE

3.6 Utility Function

Click the “Menu -> Utility” to get into the “Utility” menu.

The “Utility” Functions:

‹ FFT

‹ Record

‹ Pass/Fail

‹ Factory Setup

‹ Language

USER’S MANUAL 45
Hantek 6052BE/6082BE

3.6.1 Fast Fourier Transform Function


You can use the FFT mode to convert a time-domain signal into its frequency components
(spectrum). You can use the FFT mode for the following types of analysis:

To use the FFT mode, you need to click “Menu->Utility->FFT” to open the FFT setup
window.

The FFT setup window:

You can select the Source channel, Window algorithm, FFT number, and FFT Zoom
factor. It displays only one FFT spectrum at a time.

Source:
Selects the channel used as the FFT source

Window:
Selects the FFT window type

USER’S MANUAL 46
Hantek 6052BE/6082BE

The windows reduce spectral leakage in the FFT spectrum.

No. of Freq Bins:


Selects the FFT number

Scale:
Selects the vertical scale units

Vertical:
Selects the vertical scale

Horizontal:
Selects the FFT zoom factor
Sets horizontal scale, If the information is more than the quantity displayed to the display
window, more information is displayed by using scroll bar.

You can click the “Start” button to start FFT math, also you can click this button to stop
FFT math.
Click “Save” button if you want to save FFT math result. You can save the data into Word
file, Excel file or Text file.

Harmonic information window shows the information of Harmonics.

Measurement information window shows the information about measurement of FFT.

USER’S MANUAL 47
Hantek 6052BE/6082BE

The FFT display Windows:

USER’S MANUAL 48
Hantek 6052BE/6082BE

3.6.2 Record and Play Back

Click “Record” in “Utility” menu.

The Record window will display. The following picture shows the Record Interface.

This function can record input waveform form CH1 or CH2. The maximum record length is
1000 frames.

Record Setup window:

USER’S MANUAL 49
Hantek 6052BE/6082BE

Source:
Select record source channel. (CH1 or CH2)

End Frame:

Set the number of record times. The max frames are 1000.

Time Interval:

Set the time interval of record waveform.

Record:

Record counter, it shows the record frames.

“Start” button:
Start to record frames. After you start to record waveforms, this button changes to

“Stop” button. It stops recording waveforms.

“Save” button:

Save the record setup to file.

Play back setup window

Time Interval:

Set the time interval of play back.

Start Frame:

Set the start frame of play back.

USER’S MANUAL 50
Hantek 6052BE/6082BE

End Frame:

Set the end frame of play back.

Current Frame:

Displays the current frame of play back. You can also change this number to watch

the waveform one by one.

Repeat:

If it was checked, the waveform will be played back repeatedly.

“Play” button:

Click this button to start playing back waveform. It can stop playing back if you

started playing back.

“Load” button:

Click this button to load a record setup.

Note: When it plays back waveform, the other channel will be turned off.

3.6.3 Pass/Fail

Click “Pass/Fail” in “Utility” menu.

The Pass/Fail window appears:

USER’S MANUAL 51
Hantek 6052BE/6082BE

The Pass/Fail function monitors changes of signals and outputs pass or fail signals by
comparing the input signal with the pre-created mask.

Control Setting

Source:

Select the Pass/Fail channel

Output:

Select the Pass/Fail output condition

Stop When Output:

If it was checked, the Pass/Fail will stop when output.

USER’S MANUAL 52
Hantek 6052BE/6082BE

Mask Setting

Vertical:

Set the vertical limit range

Horizontal:

Set the horizontal limit range

“Create” button:

Click this button to create Pass/Fail area according to the mask

“Save” button:

Click this button to save the setups to file.

“Load” button:

Click this button to load the saved setups file.

Information Display

Fail: Displays the fail waveform number

Pass: Displays the pass waveform number

Total: Displays the l total Pass/Fail waveform number

Operation

USER’S MANUAL 53
Hantek 6052BE/6082BE

Click “Start” button to start the Pass/Fail test.


Click “Stop” button to stop the Pass/Fail test.

The Pass/Fail function display:

NOTE:Pass/Fall function is unavailable in X-Y mode and Roll mode.

3.6.4 Factory Setup

Click “Factory Setup” in “Utility” menu to load default setups.

When you click the Factory Setup in Utility menu, the oscilloscope displays the CH1
and CH2 waveforms and removes all other waveforms.
The oscilloscope set up for normal operation when it is shipped from the factory and can
be recalled at anytime by user.

USER’S MANUAL 54
Hantek 6052BE/6082BE

The Factory Setup function does not reset the following settings:
‹ Language option
‹ Date and time

3.6.5 Language

Click “Language” in “Utility” menu.

There are four languages in “Language” menu. The default language is English.

3.7 Measure Signal

3.7.1 Cursor

Click “Cursor” in main menu.

This method allows you to take measurements by moving the cursors.

1. Source

The user can set the source to CH1, CH2, FFT and MATH. But the FFT source can only

USER’S MANUAL 55
Hantek 6052BE/6082BE

use Trace type.


When you use cursors, be sure to set the Source to the waveform on the display that you
want to measure.

2. Type

There are four types of cursors: Cross, Trace, Vertical and Horizontal.

1) Cross

The Cross cursors appear as cross lines on the display and measure the vertical and
horizontal parameters.
The Cross cursor display window:

The Cross measure result displays on status bar

2) Trace

The Trace cursors appear as vertical lines on the display and measure the waveform
amplitude at the point the waveform crosses the cursor.

USER’S MANUAL 56
Hantek 6052BE/6082BE

The Trace cursor display window:

The Trace cursor measure result display on status bar

3) Vertical

The Vertical cursors appear as vertical lines on the display and measure the vertical
parameters.
The Vertical cursor display window:

USER’S MANUAL 57
Hantek 6052BE/6082BE

The Vertical cursor measure result display on status bar:

4) Horizontal

The Horizontal cursors appear as horizontal lines on the display and measure the
horizontal parameters.

The Horizontal cursor display window:

The Horizontal cursor measure result display on status bar:

3.7.2 Measure Menu

Click “Measure” in main menu.

The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time
measurements).

USER’S MANUAL 58
Hantek 6052BE/6082BE

1. Source

The user can use the “Source” menu to select a measure source.

2. Vertical

Maximum: Voltage of the absolute maximum level, measured over the entire waveform
Minimum: Voltage of the absolute minimum level, measured over the entire waveform
Peak to Peak: Peak- to-peak=Max-Min, measured over the entire waveform
Top: Voltage of the statistical maximum level, Measured over the entire waveform
Base: Voltage of the statistical maximum level, measured over the entire waveform
Middle: Voltage of the 50%level from base to top
RMS:T he Root Mean Square voltage over the entire waveform
Amplitude: Amp=Base-Top, measured over the entire waveform
Mean: The arithmetic mean over the entire waveform
Cycle Mean: The arithmetic mean over the first cycle in the waveform
Preshoot: Positive Overshoot =(Max-Top)/Amp*100% , measured over the entire
waveform

USER’S MANUAL 59
Hantek 6052BE/6082BE

3. Horizontal

Period: Time to take for the signal cycle to complete in the waveform
Frequency: Reciprocal of the period of the first cycle in the waveform
Rise Time: Time taken from lower threshold to upper threshold
Fall Time: Time taken from upper threshold to lower threshold
+Duty Cycle: Positive Duty Cycle = (Positive Pulse Width)/Period x 100%,Measured of
the first cycle in the waveform
-Duty Cycle: Negative Duty Cycle = (Negative Pulse Width)/Period x 100%, Measuredof
the first cycle in the waveform
+Pulse Width: Measured of the first positive pulse in the waveform. The time between the
50% amplitude points
-Pulse Width: Measured of the first negative pulse in the waveform. The time between the
50% amplitude points

USER’S MANUAL 60
Hantek 6052BE/6082BE

4. Clear Measure

Clear all measure items on display screen.

The Measure Display Window:

Caution: The results of the automatic measurements will be displayed on the bottom of
the screen. Maximum 8 results could be displayed at the same time. When there is no
room, the next new measurement result will make the previous results moving left, out of
screen.

3.8 The Display System

3.8.1 Display Type


Click “Type” in “Display” menu.

The following figure shows the parameters setting.

USER’S MANUAL 61
Hantek 6052BE/6082BE

If the Vectors type mode is selected, the waveform will be displayed as following figure.

If the Dots type mode is selected, the waveform will be displayed as following figure.

USER’S MANUAL 62
Hantek 6052BE/6082BE

3.8.2 Display Grid


Click “Display” in main menu.

The grid shows:

USER’S MANUAL 63
Hantek 6052BE/6082BE

Without the grid shows:

USER’S MANUAL 64
Hantek 6052BE/6082BE

3.8.3 Intensity

Click “Display->Intensity” In main menu


The following figure shows the intensity dialog. It shows the display parameters setting.

You can change the grid and waveform color intensity in this dialog.

USER’S MANUAL 65
Hantek 6052BE/6082BE

3.8.4 Waveform Color

Click “Display->Waveform Color” in main menu.

The following figure shows the waveform color dialog. It shows the waveform color setting.

If you click the color rectangle, the color dialog will be showed.

The waveform color was changed:

USER’S MANUAL 66
Hantek 6052BE/6082BE

3.9 Zoom In/Out and Drag Waveforms

The software will stop updating waveform after the user clicked “Stop” button, The user
can change the waveform display by adjusting the scale and position. When you change
the scale, the waveform display will increase or decrease in size. When you change the
position, the waveform will move up, down, right, or left.

Zoom In/Out
The user can click “Zoom In/Out” in Acquire menu, then left or right click the mouse
button on display screen to Zoom in/out the waveform. Also the user can change
Time/Div in Horizontal menu or in Horizontal panel to zoom in/out the waveform.

Drag
The user can modify the waveform position after clicked “Drag” in “Acquire” menu
following the following steps.

USER’S MANUAL 67
Hantek 6052BE/6082BE

3.10 Interpolation
At the time 40ns/div or faster, user can use the different interpolation mode to get
waveforms of different smoothness.
The Step Interpolation:

The Linear Interpolation:

USER’S MANUAL 68
Hantek 6052BE/6082BE

The Sin(x)/x Interpolation:

Caution: The default interpolation mode is Sin(x)/x.

USER’S MANUAL 69
Hantek 6052BE/6082BE

3.11 Acquisition Modes


Acquisition

When you acquire a signal, the oscilloscope converts it into a digital form and displays a
waveform. The acquisition mode defines how the signal is digitized and the time base
setting affects the time span and level of detail in the acquisition.

There are two acquisition modes: Normal and Average.

Normal: In this acquisition mode, the oscilloscope samples the signal in evenly spaced
intervals to construct waveform.
Average: In this acquisition mode the oscilloscope acquires several waveforms, averages
them, and display the resulting waveform. You can use this mode to reduce random noise.

3.12 Print And Print Preview

Click “Print” in “File” menu to set the printer to print the current waveform.

USER’S MANUAL 70
Hantek 6052BE/6082BE

1. Click “print” in “File” menu to set the current waveform.


2. Click the “Print Preview” in “File” menu to get into the Preview windows.
Click the “Print Preview” in “File” menu to get into the Preview window.
In “Print Preview” window, use the “Zoom In” button and the “Zoom Out” button to
change the size of the waveform graph. Click the “Close” button to turn this window off
and click the “Print” button to print the report.

The Print report:

USER’S MANUAL 71
Hantek 6052BE/6082BE

Chapter 4 Application Example

‹ Sample Measurement

‹ Pass/Fail Test

‹ Capturing a Single-Shot Signal

‹ The Application of the X-Y

‹ Taking Cursor Measurement

USER’S MANUAL 72
Hantek 6052BE/6082BE

4.1 Simple Measurement

To acquire and display a signal, please do the steps as follows:


1. Connect signal to CH1 by using probe
2. Click the button on toolbar or “Acquire” -> “Auto Setup” on menu.

To measure the frequency and “Vpp”, you can do these steps as follows:

1. Click the “Measure”-> “Horizontal”-> “Frequency” button, the frequency of the


signal display on the bottom of the waveform interface.

2. Click the “Measure”-> “Vertical”-> “Peak-to-Peak” button, the “Vpp” of the signal
will also display on the bottom of the waveform interface.

3. To clear the measurement on the waveform interface, click the “Measure”-> “Clear
Measure” button.

USER’S MANUAL 73
Hantek 6052BE/6082BE

4.2 Pass/Fail Test

The Pass/Fail function monitors changes of signals and outputs pass or fail signals by
comparing the input signal with the pre-created mask.

Control Setting:

Source:

Select the Pass/Fail channel.

Output:
Select the Pass/Fail output condition.

Stop When Output:

Mask Setting

USER’S MANUAL 74
Hantek 6052BE/6082BE

Vertical:

Set the vertical limit range.

Horizontal:

Set the horizontal limit range.

“Create” button:
Click this button to create Pass/Fail area according to the mask.

“Save” button:

Click this button to save the setups to file.

“Load” button:
Click this button to load the saved setups file.

Information Display:

Fail: Displays the fail waveform number.


Pass: Displays the pass waveform number.
Total: Displays the total Pass/Fail waveform number.

Operation:

Click “Start” button to start the Pass/Fail test.


Click “Stop” button to start the Pass/Fail test.

USER’S MANUAL 75
Hantek 6052BE/6082BE

Caution: Pass/Fail function unavailable in X-Y mode and Roll mode.

4.3 Capturing a Single-Shot Signal

To capture a single event, it needs to gather some pre-test knowledge of the signal in
order to set up the trigger level and slope correctly. For example, if the event is derived
from 3.3V COMS logic, a trigger level of 1.2 or higher Volts should work on a rising edge.
Do these steps as follows:
1. Set the probe and the channel attenuations to X 10.
2. Set up the trigger in the Trigger Menu, or in the Trigger Setting window.
(1) Adjust the Trigger Mode to Edge.
(2) Set the Trigger Sweep to Single.
(3) Set the Trigger Source to CH1.
(4) Set the Trigger Slope to “+” which means you select the rising edge.
(5) Adjust the Volts/Div and the time base in a proper range for the signal.
(6) Drag the trigger level sign on the waveform display screen to proper position. It ally
higher a little above the normal level.
(7) Click START button to start capturing. When the trigger conditions are met, data
appears on the display representing the data points that the oscilloscope obtained with
one acquisition.
This function helps to capture the signal occurrence easily, such as the noise with large
amplitude; set the trigger level higher a little above the normal level and press and wait.
When noise occurs, the instrument will record the waveform before and after the trigger.

USER’S MANUAL 76
Hantek 6052BE/6082BE

4.4 The Application of the X-Y Operation

X-Y Plot acts to analyze correlation of data of two channels. Lissajous diagram is
displayed in the screen when you use X-Y Plot, which enables to compare frequencies,
amplitudes and phases of counterpart waveform against the reference waveform. This
makes it possible to compare and analyze frequency, amplitude and phase between input
and output.
Do these steps as follows:
1. Set the probe attenuation to “x10”. Set the switch to “x10” on the probes.
2. Connect the CH1 probe to the input of the circuit, and connect the CH2 probe to the
output of the circuit.

3. Click button.
4. Adjust the vertical scale and offset to display approximately the same amplitude
signals on each channel.
5. Select X-Y format at Horizontal window. The oscilloscope will displays a Lissajous
pattern representing the input and the output characteristics of the circuit.
6. Adjust the scale and offset of the horizontal and vertical to a desirable waveform
display. The following picture shows a typical example.
7. Apply the Ellipse Method to observe the phase difference between the two channels.

Signal in X-Y Format:

USER’S MANUAL 77
Hantek 6052BE/6082BE

Instruction of the Ellipse Method:

Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals.
From the formula above:
θ = ±arcsine (A/B) or ±arcsine (C/D)
θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I
and III quadrant .If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π)
or (π~3π/2).

4.5 Taking Cursor Measurements

Use cursors to make time and amplitude measurements on a waveform quickly.

Measure the Frequency or Time of the Sine Waveform


Do these steps:
1. Click “Cursor”-> “Source”, select CH1 (select CH2 if you want measure CH2).
2. Click “Cursor”-> “Type”, select Vertical.
3. Push left mouse button, and the vertical lines appear.
4. Drag the mouse button to the point you want to measure.

5. Release the left mouse button, the frequency difference and time difference will be
shown at the status bar.

USER’S MANUAL 78
Hantek 6052BE/6082BE

Measure the Frequency and Time:

Read the details showing in the status bar.

Measure the Amplitude of the First Waveform Peak of the Waveform


Do these steps:
1. Click “Cursor”-> “Source” select CH1 (select CH2 if you want measure CH2).

2. Click “Cursor”-> “Type”, select Horizontal.

3. Push left mouse button, and the Horizontal lines appear.

4. Drag the mouse button to the point you want to measure.

5. Release the left mouse button, the voltage difference will be shown at the status bar.

USER’S MANUAL 79
Hantek 6052BE/6082BE

Measure the Amplitude:

Read the details showing in the status bar.

Trace the Amplitude of a fixed position on X-axis in a Waveform

Do these steps:

1. Click “Cursor->Source”, select CH1 (select CH2 if you want trace CH2).

2. Click “Cursor->Type”, select Trace.

3. Click the cursor at the position that you want traced of the wave in the waveform
window.

USER’S MANUAL 80
Hantek 6052BE/6082BE

Trace the Amplitude:

Read the details showing in the status bar.

Caution: Click “Cursor->Type” to select “Cross”, you can measure time and amplitude
at one time.

USER’S MANUAL 81
Hantek 6052BE/6082BE

Chapter 5 Appendix

‹ Appendix A: Specifications

‹ Appendix B: General Maintenance

USER’S MANUAL 82
Hantek 6052BE/6082BE

Appendix A: Specifications

Specifications Table:

Input
Real-time sampling:
Max. sample rate 150MS/s using one channel(Hantek6052BE)
250MS/s using one channel(Hantek6082BE)
Channels 2
50 MHz analog bandwidth(Hantek6052BE)
Bandwidth
80MHz analog bandwidth(Hantek6082BE)
Vertical resolution 8 bits/channel
Gain range 10mV ~5V/div @ x1 probe(10mV, 20mV, 50mV, 100mV, 200mV,
500mV, 1V, 2V, 5V/div, 1,2,5 sequence)
100mV ~ 50V/div @ x10 probe
1V ~ 500V/div @ x100 probe
10V ~ 5KV/div @ x1000 probe
Range 8 divisions
Offset level +/-4 divisions
Coupling AC, DC,GND
Offset increments 0.02 div
Impedance 1M ohm
DC accuracy +/-3%
Input protection 35Vpk (DC + peak AC < 10 kHz, without external attenuation)
Display Mode Y-T, X-Y and Roll
Timebase
4ns/div ~ 1h/div(4ns, 10ns, 20ns, 40ns, 100ns, 200ns, 400ns,
1us, 2us, 4us, 10us, 20us, 40us, 100us, 200us, 400us, 1ms, 2ms,
Timebase range
4ms, 10ms, 20ms, 40ms, 100ms, 200ms, 400ms, 1s,2s, 4s,
10s,20s,40s,10m,20m,40m,1h /div 1-2-4 sequence)
Real-time sampling: 4ns/div ~ 400ms/div.
Acquisition mode
Roll mode: 1s/div ~ 1h/div
Range 10 divisions
Buffer Size 10K ~ 64KB /CH (Hantek6052BE, Hantek6082BE)
Trigger
Type Edge trigger
Mode Auto, Normal and Single
Autoset Yes
Source CH1, CH2, ALT, EXT, EXT/10
Range 10 divisions
Trigger level +/-4 divisions

USER’S MANUAL 83
Hantek 6052BE/6082BE

Settabillity 0.02 div increments


Math
Vp-p, Vmax, Vmin, Vtop, Vbase, Vmid, Vrms, Vamp, Vmean,
Vcyc mean, Positive overshoot, Negative overshoot;
Measurements
Period, Frequency, Rise time, Fall time, Positive duty cycle,
Negative duty cycle, Positive Width, Negative Width
Math Addition, Subtraction, Multiplication, Division
FFT Rectangular, Hanning, Hamming, Blackman Window
Physical
Interface USB2.0
No external power source required.
Power
Bus-powered from USB
Dimensions 190 x100 x 35(mm)

USER’S MANUAL 84
Hantek 6052BE/6082BE

Appendix B: General Maintenance

General Care
Do not store or leave the oscilloscope where the device will be exposed to direct sunlight
for long periods of time.

Caution

To avoid damages to the device or probes, do not expose them to sprays, liquids or
solvents.

To avoid damages to the surface of the device or probes not use any abrasive or chemical
cleaning agents.

Cleaning
In order to keep the instrument clean ,you need to check often whether the channels are
dusty or not .Please clean the instrument surface and the channel connecting point as
per the following cautions(Make sure the device disconnect form all power sources):

1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth.

2. Use a soft cloth dampened with water to clean the device.

USER’S MANUAL 85

You might also like