Evolution 350 Uv Vis Spectrophotometer Specifications PS52998
Evolution 350 Uv Vis Spectrophotometer Specifications PS52998
Evolution 350 Uv Vis Spectrophotometer Specifications PS52998
––No routine service costs for system qualification and verification following
lamp replacement
•C
hoose an optional Calibration and Validation Carousel (CVC) to automate all
your routine performance verification tasks. Just insert the CVC, select your
method and walk away. The report is ready for inspection and filing in your
log when you return.
Evolution 350 UV-Vis Spectrophotometer Specifications
Optical Design Modified Ebert
Double beam with sample and reference cuvette/accessory positions
Spectral Bandwidths Selectable 0.5, 1.0, 1.5, 2.0, 4.0 nm
Light Source Xenon flash lamp
Typical lifetime: >5 years; longer if not using live display
Warranty period: 3-year source replacement warranty
Detector Detector dual-matched silicon photodiodes
Grating Holographic, 1200 lines/mm, blazed at 240 nm
Beam Separation 210 mm
Scan Ordinate Modes Absorbance, % Transmittance, % Reflectance, Kubelka-Munk,
Software support Log(1/R), Log(Abs), ABS × Factor, Intensity, 1st–4th Derivative
Range 190–1100 nm
Comprehensive and powerful
Accuracy ±0.20 nm (546.07 nm Hg emission line), ±0.30 nm, 190–900 nm
software completes your
Wavelength Repeatability Standard deviation of 10 measurements <0.05 nm
Evolution 350 System.
Scanning Speeds Variable, up to 6000 nm/min
• Insight Software offers Data Interval 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm
sophisticated tools for data Range >4 A
acquisition, analysis and reporting Display ±6 A
in general research and quality Accuracy – 1A: ±0.004 A
Instrument* 2A: ±0.004 A
control environments. 3A: ±0.006 A
Photometric Noise 0A: <0.00018 A
• Insight Security Software offers 1A: <0.00022 A
advanced security options and 2A: <0.00050 A
500 nm, 2.0 nm SBW, RMS
complete tools for achieving
Repeatability 1A: ±0.0025 A
21 CFR Part 11 compliance.
Drift <0.0005 Abs/hour, 500 nm, 2.0 nm SBW, 2 hour warm-up
• Insight Bio Software adds Stray Light 198 nm: 2.4 A KCl
220 nm: 3.5 A NaI
pre-programmed methods for 340 nm: 4.0 A NaNO2
performing routine assays, such Baseline Flatness ±0.0015 A (200–800 nm)
as nucleic acid tests and protein 2.0 nm SBW, smoothed
Dimensions (W × D × H) 61 × 53 × 38 cm
concentrations. Additional
Weight 22 kg
kinetics capabilities include
Electrical Supply 100–240 V, 50–60 Hz
temperature-based rate and
*M
easured at 546 nm using neutral density filters traceable to NIST. When testing instrument performance, the specification used for pass/fail
DNA melting experiments. determination is the sum of the instrument specification listed here and the uncertainty in the calibration data for the filter, listed on the