Testomat ECO GB 190308
®
Testomat ECO
Online analysis instrument
for residual total hardness
(water hardness)
Contents
Contents
Contents............................................................................................... 2
Important safety information ............................................................. 4
Intended use ......................................................................................... 4
Qualification of the staff ........................................................................ 4
Warning notices in these instructions ................................................... 5
Further documents ................................................................................ 5
Pay particular attention to.................................................................. 5
General instructions .............................................................................. 5
Properties of the measured water ......................................................... 5
Installation ............................................................................................. 6
Operation .............................................................................................. 6
After switch-off and longer downtime .................................................... 6
Cleaning ................................................................................................ 6
De-installation ....................................................................................... 6
Disposal ................................................................................................ 6
Scope of delivery ................................................................................ 7
Performance specifications ............................................................... 7
®
Indicators for Testomat ECO instruments ........................................... 8
Application instructions ......................................................................... 9
Installation ......................................................................................... 10
®
Operating Testomat ECO in the pressure range 0.3 to 1 bar ........... 10
®
Installing Testomat ECO ................................................................... 10
Connecting the water inlet and outlet ................................................. 11
Water inlet ........................................................................................... 11
Water outlet ......................................................................................... 11
Connecting the power supply and devices ......................................... 12
®
Block diagram Testomat ECO ........................................................... 12
®
Internal design Testomat ECO .......................................................... 13
Connecting the mains voltage............................................................. 14
Connecting the plant components ...................................................... 15
Connecting the inputs and outputs ..................................................... 16
Commissioning ................................................................................. 17
Inserting the indicator bottle ................................................................ 17
Extracting the indicator ....................................................................... 17
Opening the water inlet ....................................................................... 17
Instrument settings and data input ...................................................... 18
Functions of the operating and display elements ......................... 18
®
Switching Testomat ECO on/off ........................................................ 18
Display functions ................................................................................. 18
Operating elements and function keys ............................................... 20
Operating system ................................................................................ 21
Entering basic program data ........................................................... 22
Selecting the indicator and the bottle size .......................................... 22
Selecting the operating mode ............................................................. 22
Selecting the time control.................................................................... 22
Setting the analysis interval (interval pause) ...................................... 23
Selecting the quantity control .............................................................. 23
Selecting the type of water meter ....................................................... 23
Selecting the quantity control/time priority .......................................... 24
2
Contents
3
Important safety information
Intended use
®
Testomat ECO is used for the automatic determination and monitor-
ing of residual total hardness (water hardness) in water with a pH
level above 4. The required measuring range is determined by the
indicator selection and the respective user programming.
Always adhere to the performance limits stated in the section enti-
tled “Technical data“.
Always observe the application areas/application limits of the indi-
cators and the requirements of the medium being measured.
To ensure correct and intended usage, always read and understand
these instructions, especially the section entitled “Important safety
information”, prior to use.
The instrument is not used as intended if
it is used in areas not specified in these instructions
it is used in areas which do not correspond to the ones described
in these instructions.
4
Pay particular attention to
Further documents
®
Testomat ECO is a plant component. Therefore, always observe the
® ®
maintenance manual of Testomat 2000 /ECO and the documenta-
tion of the plant manufacturer.
®
using Heyl Testomat 2000 indicators in the pH-range 4 – 10.5!
5
Pay particular attention to
Installation
Always completely disconnect the relevant plant part before in-
Operation
Ensure that the maximum electrical load capacity of the relay out-
puts is never exceeded.
Immediately switch off Testomat ECO and contact service staff if
®
Cleaning
Only use a dry, lint-free cloth for cleaning.
De-installation
Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a
defective instrument (irrespective of the warranty period) if it has
been de-installed and returned to us with a description of the error.
Disposal
Dispose of the instrument in accordance with national regulations.
6
Scope of delivery
Scope of delivery
®
1 Testomat ECO
1 plasic bag with screw cap with a hole and an insert for the
screw cap of the indicator bottle
1 operating instructions
Performance specifications
®
Testomat ECO is used for the automatic determination and monitor-
ing of residual total hardness (water hardness) in water with a pH
level above 4. The required measuring range is determined by the
indicator selection and according to the user programming.
7
Performance specifications
Parameter/Indicator type
Water hardness
TH 2005 TH 2025 TH 2100 TH 2250
°dH 0.05 - 0.50 0.25 – 2.50 1.0 – 10.0 2.5 – 25.0
(resolution) (0.01) (0.05) (0.2) (0.5)
ppm
0.89 – 8.93 4.5 – 44.8 18 - 179 45 - 448
CaCO3
(0.2) (0.9) (3.8) (10)
(resolution)
8
Performance specifications
Application instructions
Switching on/off
Wait at least 5 seconds before switching the instrument on and
then off again at the main switch.
Ambient conditions/Installation site
Always protect the instrument against moisture and humidity. It
should never come into contact with condensation or splash wa-
ter.
Malfunctioning/Repairing a defective instrument
The repair of a defective instrument – irrespective of the warranty
period - is only possible after the instrument has been dismantled
and returned to us with a description of the error. Furthermore,
please inform us of the indicator type being used and the meas-
ured medium. Do not carry out any actions at the instrument
which are not described in these instructions; failure to adhere to
these instructions will negatively affect any warranty claims that
you make thereafter. Before you return the instrument for repair,
write down a description of the error (failure effect). Repair work
repair work, remove the bottle and ensure that the measuring
chamber has been flushed out and is empty. Prior to dismantling,
always (irrespective of the warranty period) is only possible when
the instrument is dismantled and returned to us with a description
of the error.
After a protective circuit (fuse) has been tripped, attempt to elimi-
nate the cause of malfunctioning (e.g. replace a defective valve)
before reactivating the protective circuit. Frequent tripping is al-
ways due to an error which, in certain circumstances, may also
cause damage to the instrument.
Operating safety instructions
Careful handling of the instrument increases both its operational
reliability and service life! Therefore, carry out a visual inspection
at regular intervals as described below:
- Are the hose connections of the dosing pump free of leaks?
- Is there any air inside the dosing hoses?
- Are all the water connections free of leaks?
- Are the doors of the instrument closed properly?
- Is the instrument heavily soiled?
Maintenance and servicing instructions
(For more information, please refer to the section entitled
®
“Maintenance” and the "Maintenance manual of Testomat 2000 /
®
Testomat ECO )
9
Installation
Installation
Risks resulting from incorrect installation!
! WARNING ®
Install Testomat ECO at a location where it is protected against
dripping or splash water, dust and aggressive substances – e. g. in
a switch cabinet or on a suitable wall.
10
Installation
Water inlet
The measuring water is taken from the main water line of the water
®
treatment plant and fed to the inlet connection of Testomat ECO .
The instrument is equipped with a plug connector for plastic hoses
6/4 x 1 (external diameter 6 mm/ internal diameter 4 mm, wall thick-
ness 1 mm) as standard.
®
Install the connection for the branch line of Testomat ECO direct-
ly at the main water line directly after the water treatment plant
Water outlet
The feed water flows through the measuring chamber to the drain via
the outlet hose.
®
Connect the outlet connection of Testomat ECO to an outlet hose
(internal diameter 12 mm)
Lay this hose without backwater development and any syphoning
effect, e.g. via an open funnel, to the drain.
11
Installation
12
Installation
Terminal strip for inputs Stop, IN2, output OUT, +12 V for turbine
Mains switch
Dosing pump
Measuring chamber
13
Installation
Loosen both fastening screws and open the upper door. The
terminal box is now accessible.
Pierce the required rubber cable glands with a screwdriver and
insert the cable through the bush into the terminal box (1)
Subsequently pull back the cable until the bush has been turned
over (2)
Terminal Typ
description
Function Comment
et
14
Installation
No. Terminal
description
Type Function Comment
15
Installation
No. Terminal
Type Function Comment
description
22 OUT -
OUT Current interface 0/4 - 20 mA Galvanically separated *
23 OUT +
16
Commissioning
Commissioning
®
Extracting the indicator
Switch the instrument on and press the "STANDBY" key
During operation, the pump (DOSIClip) automatically extracts
the indicator
To ensure that indicator is available for the initial analyses, the
intake hose and the transport hose must be filled with indi-
cator from the pump up to the measuring chamber.
Press the "manual" key several times until the intake hose
and the transport hose are filled with indicator up to the measur-
ing chamber.
If necessary, manually tighten the hose connectors of the intake
and transport hose slightly in case of bubble formation.
17
Functions of the operating and display elements
Instrument settings and data input
Please read the following information before carrying out settings
and entering data for operating the instrument.
Functions of the operating and dis-
play elements
®
The Testomat ECO display shows operating statuses and measured
values. The input keys for programming (cursor block) and the func-
tion keys are located underneath the display.
(2)
18
Functions of the operating and display elements
Display functions
1 2 2a 2b 3 4
╘╛
19
Functions of the operating and display elements
Programming
keys
(cursor block)
ECO®
Function keys
20
Functions of the operating and display elements
Operating system
Testomat instrument
(display) in display mode SELECTING FUNCTIONS (example: "select operating mode")
Enter the quantity interval in the same way after selecting the menu
item "VOLUME". Select the four digits and confirm via "ENTER". It is
possible to enter values between 1- 9999 litres.
21
Entering basic program data
Type TH2025
The selection shown on the left is displayed.
Type TH2100
Select the size of the indicator bottle
Type TH2250
(A 500 ml bottle " * " is factory set)
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
Select the indicator type
(The indicator type TH2005 " * " is factory set)
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
22
Entering basic program data
TIME 10m In the menu, select => BASIC PROGRAM => INTERVAL => TIME
Volume 0000l Use the cursor keys to select the interval in minutes
(10 minutes are preset)
Press "ENTER" to complete all the entries
Time-controlled In the menu, select => BASIC PROGRAM => INTERVAL =>
VOLUME INTERVAL * VOLUME
Press "ENTER" to confirm the selection
Enter the respective flow rate in litres
Time 10m
Press "ENTER" to confirm the entry
VOLUME 0350l
1 L/IMPULSE
Selecting the type of water meter
2,5 L/impulse
It is necessary to connect a water meter to input IN for quantity-
5 L/impulse
dependent analysis triggering.
10 L/impulse
100 L/impulse * Select the menu => BASIC PROGRAM => WATER METER
500 L/impulse Select the water meter constant (litre/impulse)
1000 L/impuls
(100 litres/impulse " * " is factory set)
Imp/L 000.0
000.0 Or enter the reciprocal value in Imp/l in the last field
when using a turbine with a high impulse number
Press "ENTER" to confirm the selection/entry
23
Entering basic program data
24
Entering basic program data
The limit value outputs are strictly allocated to the limit values!
Monitoring of two
limit values 1 LV1 = Limit value 1 2 LV2 = Limit value 2
If the limit value LV1 has been exceeded, the limit value control dis-
play 1 lights up RED and the relay output LV1 reacts as pro-
grammed in the switch function. If the limit value has not been ex-
ceeded, the display lights up GREEN. The same applies for the limit
value LV2.
LV 2: 0,15 °dH Enter the values for "LV 1" or "LV 2"
000l
Press "ENTER" to confirm the entry
Hysteresis
The respective limit value output only switches after the first, second
Suppression of or third bad analysis (suppression of the first or the second measured
bad analyses
value). This increases the reliability of the analysis evaluation, e.g. if
the sampling line has not been flushed sufficiently. The hystereses of
the two outputs LV1 and LV2 can be set independently of each other.
With a hysteresis of "2", the next analysis is immediately carried out
when the limit value of this analysis has been exceeded for the first
time. The respective output is only switched after the limit value of
this analysis has been exceeded for a second time. With a hysteresis
of "3", the respective output only switches when the limit value of this
analysis has been exceeded for a third time. This setting is only reac-
HYSTERESIS LV1 1 tivated once the value has fallen below the limit value!
(The basic setting for LV1 and LV2 is 1)
25
Entering basic program data
26
Entering basic program data
Function off
In the menu, select => BASIC PROGRAM => BOB OPERATION
FUNCTION ON * Select "FUNCTION ON"
Press "ENTER" to confirm the entry
27
Description of the signal inputs/outputs
Stop input
28
Description of the signal inputs/outputs
Connecting a turbine
Besides the connections 20 and 21, an additional power supply is
required for connecting a turbine. Connect this power supply (+ 12 V)
to terminal 24.
The connection configuration is shown in the diagram on the left.
29
Description of the signal inputs/outputs
Type 4 - 20 mA *
Select the desired current range
Press "ENTER" to confirm the selection
Measured value
Function Current = x 16 mA + 4 mA
4 - 20 mA Maximum value
30
Description of the relay outputs
31
Description of the relay outputs
Error messages
NOTE The error message can only be deleted once the error has been
eliminated!
32
Information menu "i"
Call (1)
Use the key to open the information menu "i".
Call (1)
Opening the information menu Request options: Operating values, program values
for checking or requesting
settings and operating values.
Selection (2)
- Operating values or
- Program values
Please refer to the section entitled "Entering basic program data" for
further programming and setting details of the individual menu items".
33
Program menu "M"
Flush (5)
Press the "ENTER" key to start
the flushing of the sampling line
via the internal valves. Press
the "ENTER" key again to
cancel this function.
Availability of functions
NOTE All manual functions can only be selected during an analysis
pause. Analyses are not carried out during manual operation. All
signal inputs and outputs are locked.
34
Program menu "M"
Language (9)
GERMAN Select the desired language for the display.
English *
Francais
Diagnosis (10)
Italiano
Polski
It is possible to request a list of current statuses of the signal inputs
Nederlands
and outputs. Active statuses are marked with an *. (See "Structure of
Espanol
the basic program").
The current interface can be checked under "OUT output". Press the
"Enter" key to toggle between minimum and maximum current.
Changeover occurs between 000 and 200 at 0-20 mA!
Basic program
Access this menu item by pressing the "M" key (menu). Here you can
carry out basic programming of the instrument and select various
service functions.
In the basic program, the following abbreviations are used in the re-
spective menu items:
s = seconds; m = minutes; h = hours; T = days; l = litres
35
Program menu "M"
Call the basic factory program by simultaneously pressing and holding down the "M" and "i" keys
while switching on the instrument. CAUTION, the most recent programming will be lost!
36
Error messages/Troubleshooting
Error messages/Troubleshooting
Displayed message (flashes at Instrument result functions Possible causes Remedies
selected display)
Ff. DOSING PUMP - Continuous alarm - Dosing pump is defective Replace dosing pump
- Standby - No dosing message from Check cable to the dosing
the dosing pump pump for correct connec-
tion
Ff. OUTLET TO DRAIN - Continuous alarm - Water remains in the meas- Check water outlet
- Standby uring chamber Connector at the outlet
CANCEL WITH HORN KEY valve oxidised
Replace valve block
LOW INDICATOR LEVEL - Continuous alarm - Minimum indicator quantity Check indicator level and,
- Continue measurements not reached if necessary, refill
Without BOB: 50 ml (10 %), (enter the filling quanti-
ty!)
With BOB: According to
CANCEL WITH HORN KEY calculation
Mf. SOILING - Continuous alarm - Sight-glass windows are Clean sight-glass windows
- Continue measurements soiled
37
Maintenance
Further information
Maintenance
Required maintenance measures
NOTE Regular maintenance is necessary to ensure trouble-free opera-
tion of the instrument!
38
Maintenance
Cleaning measures
! ATTENTION Ensure tension-free mounting of the sight glass windows. Tighten the
screws equally alternating both sides. Otherwise, the sight glass win-
dows may break.
39
Maintenance
Service instructions
The surface of the instrument has not been treated. Therefore, soiling
caused by indicators, oil or grease should be avoided. However, if the
housing becomes soiled, please clean the surface with isopropanol
(never use other solvents).
40
Testomat ECO® spare parts and accessories
Art. No. Pressure controller Art. No. Instrument spare parts list
40125 Controller / filter receiver, complete 31582 Fuse GS-M 5x20E 4A
40120 Controller / filter receiver 37245 Basic circuit board T-ECO complete
40129 Controller plug T2000, complete 230 V
11225 Flow controller valve body, complete 40332 Control circuit board T-ECO complete .
11230 Retaining pin 3x38 90° 40091 Plug in circuit board driver/receiver SE-
T2000 (6)
11217 Inlet filter 19.5dx25
40190 Cable sleeve 5-7, grey
11218 Spring for inlet filter
40191 Cable sleeve 7-10, grey
40121 Inlet connector
31713 Ribbon cable 10 pole with EMI filter
40153 Plug in connector G 1/4" -6
clamp
40157 Angled plug-in connector G 1/8"
40096 Ribbon cable 26 pole with EMI filter
Measuring chamber clamp
40173 Sight-glass window with seal, T2000 40060 Cable loom 2V for T2000
40170 Sight-glass window 30x3 40062 Cable loom 2P for T2000
40176 Sight-glass holder, countersunk and 40200 Cable loom complete with power switch
thread and cover
33253 Screw M3x40, A2, DIN 965 31622 Fuse, soldered T0.16A
40032 Latch fastener TL-17-201-52 31592 Fuse, soldered T1.016A
11210 Plug for measuring chamber Spare parts requirement for 2 - 3 years of
T2000/Eco operation
40022 Measuring chamber T2000, complete 40173 Sight-glass window with seal, T2000
Measuring chamber holder 11217 Inlet filter 19.5dx25
40029 Measuring chamber holder, complete 40124 Gasket set T2000
ET 31622 Fuse, soldered T0.16A
40050 Magnetic stirrer, processed 31592 Fuse, soldered T1.016A
40156 Srew in connector 3/8" -10, processed
40056 Solenoid valve, 2/2-way
40181 Rear guide bar for measuring chamber
5 x 60
Dosing pump DosiClip®
40001 Dosing pump DosiClip ET
40011 Suction hose, complete
40016 Pressure hose, complete
40040 Valve set
32046 Plastic cover CNH 45 N
Bottle connection / Suction device
40131 Screw cap with bottle insert T2000
40130 Screw cap GL32 - hole
40135 Bottle insert for screw cap with push-fit
suction tube
41
Testomat ECO® spare parts and accessories
Accessories
42
Check List Testomat ECO®
Dear customers and service technicians,
This check list cannot replace your expertise or extensive experience in fault resolution. It is intended to support fast and systematic error diagnosis
and error documentation. This list does not claim to be complete. We are therefore always grateful for any advice and information you may be able
to provide. General user instructions can be found on the rear of this check list.
Block 2 / Error message and error history Please mark appropriately (X)
What does your instrument’s error history display?
(“i” and “Enter” key => operating instructions)
( Error history text )
Does an error message appear on the display? Yes No
For example, “Mf. analysis”, “Low water press.”, etc.
(See operating instructions: “Error messages / Troubleshooting”)
( Error message text )
Block 3 / Visual inspection and functional test Please mark appropriately (X) If applicable, values / comments
Is the instrument connected to the mains voltage specified on the rating plate? Yes No
Does a message appear on the display? Yes No
Does the instrument display a plausible measured value? Yes No Measured value:
(possible manual measurement ___________ value)
Are the measuring chamber and sight-glass windows clean? Yes No
Are the measuring chamber and the water-carrying hoses free of leaks? Yes No
Is the indicator’s expiry date still valid? Yes No Expiry date:
(See expiry date printed on the indicator bottle)
Has the correct indicator type been programmed? Yes No Type:
(TH 2025 => 0.25 to 2.5 °dH = factory setting)
Is the water pressure within the specified range (400 ml/min)? Yes No Plant pressure:
(See the type plate on the instrument)
Does the drain hose prevent the risk of backwater? Yes No
(No “syphoning effect”!!)
Is the drain hose free of blockages? Yes No
(Microorganisms caused by contamination, etc.)
Does the set flushing time/quantity of flush water ensure that only freshwater Yes No Flushing time:
is measured?
Are the hoses at the dosing pump free from air bubbles? Yes No
(Operate the pump manually / Carry out a manual analysis)
CARRYING OUT A (MANUAL) ANALYSIS
Does the water column rise evenly up to the overflow borehole when filling the Yes No
measuring chamber (5 mm below the upper edge of the measuring chamber)?
(If not: check the water pressure, water throughput/flow controller)
Does the indicator pump dose correctly when starting an analysis? Yes No No. of dosing strokes:
(LED at the pump illuminates!)
Have the indicator and water been mixed properly in the measuring cham- Yes No
ber after the dosing process?
Check the magnetic stirring bar! =>see maintenance manual “Adjust mode”
PROGRAMMING DATA / OPERATING CONDITIONS
Have the limit values been set correctly? (Within the measuring Yes No Limit values:
range/according to the performance limit of the plant?)
Is the Testomat instrument always supplied with mains voltage – except Yes No See the “General instructions for operat-
during maintenance work/emergency situations? ing Testomat 2000® and Testomat®
(Temporary shutdown only via the “Standby” key or the “Stop” input!) ECO”
Please refer to “Error messages / Troubleshooting” in the operating instructions for further information on error messages and possible causes
of faults.
Further functional tests (e.g. overflow detection and amplification setting => “Special function Adjust mode”) and service instructions can be found
in the maintenance manual.
After completing these checks, experience shows that it can be assumed that the checked functions (Block 3) are in effective working order if you
have answered all the questions with “Yes”. We recommend you to carry out these checks during each inspection or if faults occur.
43
Instrument settings
Caution!
Your settings may be deleted if repairs are carried out. Therefore, note down your instrument settings in the table below before
sending the instrument to our service team for repairs. Please enclose a copy of the table with the instrument. If you have noted
down the settings, they can be easily re-entered by your service staff once any repairs have been completed.
Menu Setting
MODE OF OPERATION INTERFACES
Time-controlled Type 0-20 mA
Volume interval Type 4-20 mA
Type RS232
DISPLAY UNIT
Display in °dH SPRACHE/LANGUAGE
Display in °f DEUTSCH
Display in ppm CaCO3 English
Display in mmol/l Français
Italiano
TYPE OF REAGENT Polski
500ml-bottle Nederlands
100ml-bottle Espanol
TH2005 Water hard.
TH2025 Water hard.
TH2100 Water hard.
TH2250 Water hard.
LIMIT VALUES
Limit val. 1:
Limit val. 2:
FLUSH TIMES
Flush
BOB-OPERATION
Function off
Function on
FUNCTION LV1
Duration
Impulse
Interval
Two point
Time:
FUNCTION LV2
Duration
Impulse
Interval
Time:
HYSTERESIS LV1
Analysis (1,2,3)
HYSTERESIS LV2
Analysis (1,2,3)
FUNCTION STOP
Normally open contact
Normally closed contact
Technical data
Technical data
Protection class: I
Degree of protection: IP 65
Water temperature: 10 – 40 °C
®
* When using Testomat ECO at a pre-pressure of 0.3 bar, it must
be ensured that the flow rate through the measuring chamber is at
least 400 ml/min.
** See also the notes on pages 5, 6, 7
We reserve the right to make technical changes without notice in
the interest of constantly improving our products!
Our manuals are updated regularly. If you have an older version
(see version at the back of the manual) you will find the current
manual on our website www.heyl.de on the download page.
45
Technical data
Conformity Declaration
46
Product overview Testomat 2000®-Instruments
Testomat 2000® • Water hardness 0.05-25 °dH • Universal for water treatment
• Carbonate hardness 0,5-20 °dH plants
Testomat 2000® CLF • Free Chlorine 0-2.5 mg/l • DPD-method for swimming pool
and drinking water control
Testomat 2000® CLT • Total Chlorine 0-2.5 mg/l • DPD-method for swimming pool
and drinking water control
Testomat 2000® CrVI • Chromate 0-2.0 mg/l • process control of waste water in
• Chrome-VI 0-1.0 mg/l galvanic industry
Testomat 2000 THCL® • Total Chlorine 0-2.5 mg/l • DPD-method for swimming pool
• Water hardness 0.25-2.5 °dH and drinking water control
• combination system for hardness
and chlorine
47
Gebrüder Heyl
Analysentechnik GmbH & Co. KG
Orleansstraße 75b
D 31135 Hildesheim
www.heyl.de Testomat_ECO_GB_190308.doc