SK 600 Service Manual
SK 600 Service Manual
SK 600 Service Manual
Infusion Pump
维修手册
Service Manual
Intellectual Property Rights
The intellectual property right of this product and its Service Manual belongs to Shenzhen Shenke Medical
Instrument Technical Development Co., Ltd. (hereinafter short as Shenke).
Shenke reserves the right to final interpretation of this Service Manual. Without prior approval from Shenke
in writing, this Service Manual shall not be photocopied, modified or translated, fully or partially, by any
individual or organization.
I
Foreword
Introduction
This Service Manual provides detailed introduction of the product and its components, as well as the
procedures for installation, removal, testing and troubleshooting, intended to help the service technicians solve the
common problems effectively. This Service Manual does not provide depth introduction on the product structure
and design principle. In case of any problems you cannot solve, please contact our After-sale Service Division.
This Service Manual describes the product based on most complete configurations, so that some parts might
be unsuitable to your product. If you have any questions, please contact our After-sale Service Division.
Before repairing the product, please read this Service Manual thoroughly until full understanding of the
contents herein, so as to ensure correct repair and avoid causing damage to the product or human body.
Target Readers
This Service Manual is intended for the biomedical engineers, authorized repairmen or service
representatives responsible for maintenance of this product.
Version Information
With the change of software or technical specifications, the version number of this Service Manual is subject
to updating without prior notice. Version information of this Service Manual is as follows:
■ Version: 1.0
■ Date Released: 2011-12
© 2011 All rights reserved. Shenzhen Shenke Medical Instrument Technical Development Co., Ltd.
II
Contents
Chapter 1 Safety ........................................................................................................................................................ 1
1.1 Safety Information ................................................................................................................................. 1
1.1.1 DANGER.................................................................................................................................... 2
1.1.2 WARNING ................................................................................................................................ 2
1.1.3 CAUTION ................................................................................................................................. 2
1.14 NOTES .......................................................................................................................................... 3
1.2 Symbols ................................................................................................................................................. 4
Chapter 2 Principle of Design.................................................................................................................................... 1
2.1 Overview ............................................................................................................................................... 1
2.2 System Breakdown ................................................................................................................................ 1
2.3 Hardware Configuration ........................................................................................................................ 3
.2.3.1 Mainboard .................................................................................................................................. 3
2.3.2 Pump Body ................................................................................................................................ 4
2.3.3 Power Board .............................................................................................................................. 5
2.3.4 Air Bubble Pressure Board ........................................................................................................ 6
2.3.5 Battery Cover Plate .................................................................................................................... 7
2.3.6 Door Opening Detection Board ................................................................................................. 7
2.3.7 Infusion Rate Sensor .................................................................................................................. 7
Chapter 3Test and Maintenance................................................................................................................................. 1
3.1 Introduction ........................................................................................................................................... 1
3.1.1 Test Report ................................................................................................................................. 1
3.1.2 Recommended Interval .............................................................................................................. 1
3.2 Visual Inspection ................................................................................................................................... 2
3.3 Power-On Self-Test (POST) .................................................................................................................. 2
3.4 Infusion Flow Test ................................................................................................................................. 3
3.5 Pressure Sensor Test .............................................................................................................................. 4
3.6 Battery Test ............................................................................................................................................ 5
3.7 Electrical Safety Test ............................................................................................................................. 5
3.7.1 Electric Leakage Test of Enclosure ............................................................................................ 6
3.7.2 Electric Leakage Test of Patient................................................................................................. 7
3.8 Cleaning and Sterilizing......................................................................................................................... 7
4.1 Overview ............................................................................................................................................... 1
4.2 Replacing Components .......................................................................................................................... 1
4.3 Technical Alarm and Diagnosis ............................................................................................................. 1
4.4 Troubleshooting Guideline .................................................................................................................... 3
4.4.1 Power-On / Power-Off Trouble .................................................................................................. 3
1
4.4.2 Screen Malfunction .................................................................................................................... 3
4.4.3 Alarm Fault ................................................................................................................................ 4
4.4.4 Keypad Fault .............................................................................................................................. 4
4.4.5 Motor Fault ................................................................................................................................ 4
4.4.6 Pressure Sensor Fault ................................................................................................................. 4
4.4.7 Air Bubble Sensor Fault ............................................................................................................ 4
4.4.8 Communication Error ................................................................................................................ 5
4.4.9 Power Source Fault .................................................................................................................... 5
4.4.10 Infusion Rate Sensor Fault ......................................................................................................... 6
chapter 5 Repair and Disassembly ............................................................................................................................. 1
5.1 Tools ...................................................................................................................................................... 1
5.2 Preparations for Disassembly ................................................................................................................ 1
5.3 Disassembly Steps ................................................................................................................................. 2
5.3.1 Separation of Front and Rear Enclosure .................................................................................... 2
5.3.2 Disassembly of Power Board ..................................................................................................... 4
5.3.3 Removal of Front Enclosure Set ................................................................................................ 4
5.3.4 Removal of Mainboard .............................................................................................................. 7
5.3.5 Removal of Pump Body............................................................................................................. 8
5.3.6 Removal of Battery Plug Set.....................................................................................................11
chapter 6 Components ............................................................................................................................................... 1
6.1 Machine Components ............................................................................................................................ 1
6.2 Pump Body Assembly............................................................................................................................ 3
2
Chapter 1 Safety
This chapter describes the basic safety considerations that the user must take notice and abide by when
repairing or operating the infusion pump. The same or similar safety information or other safety information
related to actual operation will appear in other chapters.
DANGER
To indicate immediate danger. If not avoided, it might cause death, severe
human injury or property loss.
WARNING
To indicate the potential hazard or unsafe operation. If not avoided, it might
cause death, severe human injury or property loss.
CAUTION
To indicate the potential hazard or unsafe operation. If not avoided, it might
cause slight human injury, product problem, damage or property loss.
NOTES
To emphasize the important considerations, provide description or explanation
for better repair of this product.
1.1.1 DANGER
DANGER
The infusion pipe must be flatly installed to the bottom of the groove on infusion
pump.
1.1.2 WARNING
WARNING
Only the service people authorized by Shenke are allowed to disassemble the
infusion pump. The repair or upgrading of the infusion pump must be done by
the repairmen that are trained and authorized by Shenke.
To avoid fire or explosion, do not use the infusion pump in an environment with
inflammable or explosive substances like anesthetics.
The infusion pump is of Class II electric protection. The PE end of the included
Class I power cord shall not be used for protective earthing or functional
earthing.
Disposal of packaging materials must comply with the local laws or the waste
treatment regulations of the hospital. The packaging materials must be put at a
place out of the children’s reach.
1.1.3 CAUTION
CAUTION
The electromagnetic field will affect the performance of infusion pump.
Therefore, the other devices near the infusion pump must comply with
associated EMC requirements. Mobile phone, X-ray or MRS device are possible
sources of interference, as they can emit electromagnetic radiation of high
intensity.
Avoid sterilizing the infusion pump at high pressure or exposing it to chemical
substances.
Before turning on the power supply of infusion pump, please confirm if the
power voltage and frequency comply with the specifications marked on the label
or defined in the Instruction Manual.
During repair, take care to prevent the infusion pump from damage caused by
falling, knock, severe vibration or other external forces.
2
1.14 NOTES
NOTES
For details on the use of infusion pump and other information, please refer to the Instruction
Manual.
This Service Manual describes the product based on the most complete configurations and
functions. Maybe your infusion pump does not have some configurations or functions.
Please do not insert the equipment not designated by Shenke into the data interface.
3
1.2 Symbols
Confirm Set
Stop
Eliminate alarm
Clear Select
ON OFF
Fragile! Handle with care during transport Keep upright during transport
Device complies with EC Directive 93/42/EEC and the basic requirements in its Appendix I. So
it is registered with CE Mark.
4
Chapter 2 Principle of Design
2.1 Overview
This infusion pump is intended for accurate continuous infusion of the drug to a single elderly person, child
or newborn baby patient.
The system is broken down by mechanical components, hardware and software. The composition diagram is
as below:
System
Mainboard
CN3
FPC1
P_Gate PYL_1 Infusion
Rate
Flash2
Air Bubble Power Board Sensor
CN2
Qpyl
Pressure Board
1
P_TX1 P_RX1
P1 CN1 PS52
Ultrasonic Ultrasonic
emitting receiving
Battery Cover
Plate
2
2.3 Hardware Configuration
Schematic Diagram:
Stepper Motor
Battery
Serial
Battery
Power Board Communication Mainboard
Cover
Plate
AC 3.3V/5V Power
Supply
Air Bubble
Pressure Board
The system core is mainboard and power board. LCD display and its power source are controlled by the
mainboard. The measuring and status information of each module is processed by the power board before sending
to the mainboard. Finally under the control of mainboard, the information is displayed on the screen. The
mainboard is also assigned to control the buzzer and status indicator. All modules are powered by the power board.
The power board is also assigned for driving of the motor, measuring of each sensor and processing of status
information.
.2.3.1 Mainboard
The mainboard serves for information integration of the whole system. It is designed to control the LCD and
the functions such as keypad input, buzzer sound and parameter saving.
3
2.3.1.1 Schematic Diagram
SRAM
MCU Buzzer
Screen
Run_LED
Keypad
5V 3.3V Serial Port 1: Power Board Serial Port 2: ZIGBEE wireless module
The mainboard communicates with the power board or ZIGBEE wireless module on power board via the
extended serial port of logic gate circuit.
The mainboard provides the display information to LCD and detects the key operation, realizing the user
interface;
The mainboard controls the buzzer for its alarm sound as well as the key stroke sound.
The mainboard controls the run indicator via adaption from power board.
SRAM supports the temporary storage of data and run of the program. EEPROM is used as the memory on
the machine.
The pump device is the provider of dynamic force to the whole system and it is the source of power to
deliver the liquid. During operation, the stepper motor drives the cam shaft to rotate, so that the pump impeller
reciprocates up and down by a specific sequence and rule as waves to squeeze the intravenous infusion pipe,
pushing the liquid in the infusion pipe to flow at a preset speed
4
Schematic Diagram
Infusion pipe
Pump impeller
The power board receives the function information from air bubble pressure sensor and then sends it to the
mainboard. It also drives the stepper motor to run according to the information from the mainboard. The power
board is also assigned to convert the power supply (from city grid and battery) into the power required for each
board. It also supports battery charging.
5
2.3.3.1 Schematic Diagram
5V
3.3V
LDO
AC
Input
AC/DC
MCU
Serial Port
Power ON/OFF
5V Output
7.4V Motor
Logic Gate Circuit Test
Circuit
Recharging Battery
12V circuit
12V Drive circuit
Stepper
Motor
Energizing
18.9V Circuit
Voltage Rise
Circuit
18V Output
Enable Off
Enable input
When AC power is input to the power board, 12V DC voltage is first obtained via a safety certified power
module. This DC voltage is used as the main input for DC/DC conversion and recharging circuit. In this case, we
can not only recharge the lithium battery, but also obtain 5V, 3.3V and 18.9V DV voltage via conversion by 5V
DC/DC, 3.3V LDO and 18.9V DC/DC circuit.
Except the power supply functions, the MCU on power board is also assigned for processing the
measurement and status information from the sensors before sending it to the mainboard. It also controls the run
of stepper motor according to the commands from the mainboard.
The air bubble pressure board provides the functions to detect the air bubble, occlusion and door opening. It
detects the air bubble by checking the attenuation of ultrasonic wave in different materials, while it detects the
occlusion by using the pressure sensor.
6
Schematic Diagram
To
Door opening detection
Power board
MCU
5V Input
Pressure information output
Pressure amplifying circuit
The central processor on ultrasonic pressure board will process the signals from the ultrasonic circuit and
door opening sensor, and then send them to MCU on mainboard.
The pressure amplification circuit will amplify the weak voltage signals from pressure sensor before sending
them to the power board.
The main function of battery cover plate is to adapt the AC power cord. Connecting 2 pcs of 2A 250V fuse in
series on AC power cord can protect the downstream circuits.
A micro switch is placed on the door opening detection board. Any change in handle status can affect the
closing of micro switch. In this case, we can know the status of handle from detecting the closing state of the
micro switch.
Schematic Diagram:
7
The infusion rate sensor is fixed on the liquid filter. Measuring method: When the liquid droplet drips down
through the filter, it will impede the receiving of infrared light in the sensor. By applying the pulse signals
generated from the infrared light of different intensity, the dripping frequency of liquid drops is exchanged into
the flow rate of infusion. And by using this signal for feedback of actual infusion condition of the pump, we can
detect if there is any liquid leakage.
8
Chapter 3Test and Maintenance
3.1 Introduction
To ensure long reliable run of the infusion pump, the service people shall carry out periodical inspection,
maintenance and testing. This chapter provides the basic testing methods for infusion pump. In this chapter, we
have also recommended the suitable testing intervals and tools. The service people are required to repair and test
the infusion pump as needed, and select the suitable testing tools.
The testing and inspection methods provided in this chapter are mainly used for verifying if the infusion
pump reaches the performance requirements. If the testing result does not meet the requirements, it indicates that
some function module of the infusion pump is failed. In this case, immediate repair or replacement is required. If
you have any other questions, please contact our After-sale Service Division.
CAUTION
All the tests must be done by qualified service people.
Be sure to take care when setting or changing the contents in [Advanced Setting]
menu; otherwise it might cause loss of data.
Before testing, the service people shall ensure if the testing tools and connection
wires are suitable. The service people shall have good skills on use of these testing
tools.
After finishing the test, our service people are required to complete the following test report and send it back
to the After-sale Service Division.
Test Equipment
Name Model / Number Effective Date
Test Records
S/N Test Item Test Point Test Result
1
2
Judgment
If OK: Tester: Test Date:
The visual inspection is intended for thorough inspection of the appearance of infusion pump. The infusion
pump passes the visual inspection if there is no obvious physical damage. The inspection items include:
2. Press key. The infusion pump will give out “DI...” startup music.
3. The system gives outs a “DU” sound (indicating the alarm sound self-test is passed); all LEDs on the panel
light up and then the LED on top left is dark (indicating the alarm lamp self-test is passed).
2
4. The startup image disappears and the system enters into main interface, indicating the normal startup is
finished.
Select 5ml/h, 1000ml/h and 2000ml/h as the test points. Perform 3 measurements at each point and take the
maximum error.
Note: In the test a), b) and c) above, it is needed to adjust the position of pipe clamp on infusion set or replace the
infusion set before each test.
Liquid Level
Input
Output
Electronic balance
Computer
3
3.5 Pressure Sensor Test
Testing tools:
Infusion bottle
Infusion pipe
Precision pressure gauge: calibrated
Testing procedures:
1. Connect the infusion pump, infusion pipe and precision pressure gauge as shown below.
2. Before infusion, the reading on pressure gauge shall be zero. If not zero, disconnect the liquid line to
reset the reading on pressure gauge to zero. After that, reconnect the liquid line properly.
3. Under advanced setting mode, you may press key (CLEAR) and key (SET)at the same
time to enter the pressure calibration interface, as shown below.
Pressure: 100.0kPa
Pressure sensor: 54
Initial pressure: 0
Coefficient: 0.63
Stop
4. Press key (START) to start the infusion. Verify that the maximum error of the pressure sensor at
any pressure point within the measuring range shall not exceed ±20Kpa against the readings on pressure
gauge.
5. Use the infusion pump to infuse the pressure gauge until the pressure inside the gauge reaches 60kPa.
Then, repeat Step 4.
6. Use the infusion pump to infuse the pressure gauge until the pressure inside the gauge reaches 100kPa.
Then, repeat Step 4.
7. Use the infusion pump to infuse the pressure gauge until the pressure inside the gauge reaches 140kPa.
Then, repeat Step 4.
4
If one of the pressure errors exceeds ±20Kpa, please contact our technical support people.
The battery performance might be deteriorated with the time going on. It is suggested to recharge and
discharge the battery every 3 months, and check the battery performance periodically.
Please check the battery performance according to the steps below:
1. Connect the infusion pump to AC power source and recharge continuously for 8~14 hours.
2. Disconnect the AC power source. Then, continuously run the infusion pump at an infusion rate of
25ml/h, until the pump is stopped because the battery power is too low.
If the time from start of infusion to stop of the infusion pump lasts 200 minutes or longer, it
indicates that the battery is in good condition.
If the time from start of infusion to stop of the infusion pump lasts 60~200 minutes, it indicates
that the battery is close to the end of its service life.
If the time from start of infusion to stop of the infusion pump lasts shorter than 60 minutes, it
indicates that the battery has reached the end of its service life and shall be replaced.
3. After finishing the battery inspection, it is needed to recharge the battery for next use.
NOTES
The battery might be damaged or have trouble if the lasting time of power
supply is too short after it is fully recharged. The lasting time of battery power
depends on the infusion pump configuration and operating frequency, e.g. long
use of LCD backlight.
The battery shall be replaced or properly recycled if there is obvious damage
(bulge, deformation or leakage) or the battery cannot store the electric power.
WARNING
The electrical safety testing is intended for finding out the abnormal electrical
hazards by using the accredited tester. Failure to discover these hazards might
cause injury to the patients or operators.
The electrical safety testing may be done by using the commercially provided
safety analyzer and other testing devices. The service people must guarantee the
suitability, functional integrity and safety of these testing devices, while having
good skills on their use.
The electrical safety testing shall be according to the latest version the following
standard: EN 60601-1.
Please also observe the local laws on electrical safety testing, if any.
In the patient ward, all the equipments that are powered by city grid and
5
connected to the medical instrument must comply with IEC 60601-1 standard.
Meanwhile, electrical safety testing must be done to these equipments according
to the testing interval of the tester.
The electrical safety testing is intended for finding the possible hazards that might cause electrical safety risk
to the patients, operators and repairmen. Please carry out the electrical safety testing under normal environment
(including: temperature, humidity and atmospheric pressure).
The electrical safety testing described in this chapter is based on 601 safety analyzer. Different safety
analyzers might be used in different regions. Please ensure the suitability of your electrical safety testing program.
Tested
Equipment
Testing tools:
Safety analyzer
Isolation transformer
If the electrical safety testing is failed, please contact our technical support people.
Always use the materials and methods listed in this chapter to clean or sterilize the infusion pump. We do
not provide warranty for the damage or accident caused by use of other materials or methods.
The chemicals or methods listed by us are only used as the means to control the infection. We will not
assume any liability for their effectiveness. For infection control method, please consult the infection
prevention department or epidemiological experts of the hospital.
Please ensure that the infusion pump and its accessories are placed in dust-free environment. To avoid
damage to the infusion pump, be sure to observe the following regulations:
Dilute the cleanser and disinfectant as instructed by the manufacturer, or use them at a concentration as
low as possible.
Never immerse the infusion pump into the liquid.
Never pour the liquid onto the infusion pump or accessories.
Never allow the liquid to enter the pump body.
To avoid damage to the enclosure, never wash the infusion pump by using the abrasive materials (e.g.
steel wool or silver polish) or any solvent like xylene or acetone.
WARNING
Before cleaning the infusion pump, be sure to turn off the power source and
disconnect the power cord from the socket.
The infusion pump shall be cleaned periodically. The cleaning interval shall be shortened in the areas with
severe pollution or great sandy wind. Before cleaning, please consult or check the regulations for cleaning of
medical equipment.
7
Below are the cleansers and disinfectants available for use:
Warm water
Diluted soap water
Diluted ammonia water
Sodium hypochlorite (bleach for wash use)
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
1. Turn off the power source and disconnect the power cord from socket.
2. Wipe the surface of infusion pump with the soft cloth wetted by warm water.
3 You may sterilize the surface of infusion pump by wiping it with the tampons that have absorbed
75% ethanol.
4 After cleaning or sterilizing, put the infusion pump in a cool and well-ventilated place for natural
drying.
The above operations are for reference only. The sterilizing effect shall be verified through appropriate
approach.
CAUTION
Never sterilize the infusion pump by using the gas (EtO) or formaldehyde.
8
Chapter 4 Troubleshooting
4.1 Overview
This chapter provides categorized description of the infusion pump troubles by fault position and symptom.
When eliminating the trouble, please refer to the corresponding fault sheet and check, locate and eliminate the
fault by order.
The recommended methods in this chapter can help you solve most of the equipment troubles, but exclusive
of any problem that might occur. In case of any fault not described in this chapter, please contact our After-sale
Service Division.
You may replace the circuit board set and other main components or parts of the infusion pump. Once you
have located the circuit board set with trouble, you may replace it according to the steps described in Chapter 5 –
Repair and Disassembly. Then, check if the trouble has been eliminated, or if the infusion pump can pass the
relevant testing. If the trouble is eliminated, it indicates that the original circuit board set is damaged. In this case,
please return it to our company for repair. If the trouble remains, please reinstall the original circuit board set and
try eliminating the trouble from other possible causes.
To know and obtain the components you need to replace, please refer to Chapter 6 – Components.
Before troubleshooting, please check if there is technical alarm on the infusion pump. If any, please eliminate
the technical alarm first and then eliminate the trouble according to the description in 4.4 – Troubleshooting
Guideline, so as to avoid unnecessary disassembly of the equipment.
Alarm Message
Note: The column “A” means if the trouble can be fully eliminated; the column “B” means if the sound and light
can be cleared; and the column “L” refers to the alarm level.
Alarm
A B L Cause Solution
Message
“Air” YES No High There is air bubble in the
infusion pipe Press key (STOP) to stop infusion and
silence the alarm. After eliminating the air
2
Alarm
A B L Cause Solution
Message
Malfunction of infusion Refer to 4.4 – Troubleshooting Guideline.
rate sensor
No AC No No Low No AC power supply Connect to AC power supply.
power connected.
supply
Low No No High Battery power is low. Connect AC power supply and recharge the
Voltage battery.
Battery aging or Refer to 4.4 – Troubleshooting Guideline.
malfunction in the battery
charging circuit.
Battery used No No High Battery power too low Connect AC power supply and recharge the
up battery.
Battery aging or Refer to 4.4 – Troubleshooting Guideline.
malfunction in the battery
charging circuit.
Startup No No Low The infusion pump is not Start the infusion pump within the preset time
Message started within the preset after correction installation of the infusion
time after correction pipe. For setting of the startup time, please
installation of the infusion refer to the Instruction Manual.
pipe.
3
4.4.3 Alarm Fault
4
to the power board is properly connected.
5
NOTES
● The modules and infusion rate sensor are powered by the power module. Fault with the power module
will cause many troubles to other components. Therefore, if the components cannot work normally,
take care to check if the working voltage of this component is normal. For the working voltage and
voltage testing point for each component, please refer to Chapter 2 – Principles of Design.
6
chapter 5 Repair and Disassembly
5.1 Tools
You might need the following tools during disassembly and replacement of the components:
Before disassembly of the infusion pump, please stop infusion to the patients and shut down the infusion
pump, while disconnecting all the accessories and external equipments.
CAUTION
Please eliminate the static electricity before disassembly. To avoid damage to the components,
please wear personal protective equipment like wrist strap or electrostatic gloves when
disassembling some components with electrostatic sensitive label.
During reinstallation, take care to connect the wires properly and place them in good order, so as
to avoid the short circuiting due to damage of the connection wires.
Take care to select the appropriate screws during reinstallation. Forcible tightening of the
incorrect screws might cause damage to the equipment. And during use after reinstallation, the
screws or components might fall off, thus causing unpredictable damage or injury to the product
or human body.
Disassemble the equipment in correct order. Incorrect order and forcible assembly might cause
irreversible damage to the equipment.
Before separating the components, please be sure that all connections have been plugged off.
During separation, take care not to break the connection wire or damage the connector.
Please store the removed screws and other parts by category for easy reach during reinstallation.
And avoid dropping, contaminating or losing them.
5.3 Disassembly Steps
If the machine is installed with infusion rate sensor, please pull out the infusion rate sensor socket from the
rear enclosure.
As shown, use the screwdriver to remove 4 screws M4×30 fixing the rear enclosure.
NOTES
● As the mating joint is applied with waterproof adhesive, you might have to use a
strong force to open the front and rear enclosure. Please use force with care.
2
As shown, separate the front and rear enclosures.
NOTES
● To avoid breaking the connecting wires or damaging the plug, please use
force with care when separating the front and rear enclosures.
3
5.3.2 Disassembly of Power Board
1. First, unplug the wires connected to the power board. Then, pull out the power board slowly.
2. Open the clamp and unplug the 14pin strip. Remove the power board.
1. Tear off the surface film on the left of front enclosure by using the forceps or other similar
4
tools.
NOTES
● To avoid scratching the front enclosure, please use force with care when tearing off
the surface film.
3. Use the screwdriver to remove 2 countersunk head screws M4×10 and 4 pan head screws
M4×8.
4. Use the screwdriver to loosen 1 screw M4×12. Then, take out the screw with forces.
5
NOTES
● To avoid dropping the screws into pump body, please use force with care when
removing the screws.
6
6. Remove the pump body assembly.
1. Use the screwdriver and hex socket wrench 5.5. to remove 4 ST3×6 screws, 2 M3 X 10
countersunk head screws and 2 M3 nuts fixing the mainboard.
7
3 Open the pin holder of mainboard and remove 14pin strips.
8
2 Unplug the wires connected to the air bubble pressure board.
3 Use the screwdriver to loosen 2 pan head screws M3×6. Then, remove the air bubble pressure
board.
4 Use the screwdriver to loosen 2 pan head screws M2×5. Then, remove the handle detection
board.
9
5 Use the screwdriver to remove 2 pan head screws M3×6 with washer. Then, remove the
bracket.
6 Use the screwdriver to remove 1 pan head screws M4×25 with washer. Then, remove the
ultrasonic pressure set.
10
7 Use the screwdriver to loosen 2 pan head screws ST2.9×9.5. Then, remove the ultrasonic
module.
First unplug the wires connected to the battery cover plate. Then, use the screwdriver to loosen 2 pan head screws
ST2×10. Remove the battery.
11
Blank Page
12
chapter 6 Components
13 80080001 Nut M3 2
2
6.2 Pump Body Assembly
Explosive View
3
Parts List
4 60003008 Belt 1
7 60002019 Magnet 2
12 60002018 Bearing 1
24 80080002 Nut M4 3
25 80080092 Gasket 1
4
29 60003178 Ultrasonic rod 1
34 60002035 Support 1
42 80080070 Washer 4 1
46 60002025 Guide 2