Manual para Prenderbasepackage
Manual para Prenderbasepackage
Reference Manual
Version 7
© 2017 BRUKER OPTIK GmbH, Rudolf-Plank-Straße 27, D-76275 Ettlingen,
www.bruker.com
All rights reserved. No part of this manual may be reproduced or transmitted in any
form or by any means including printing, photocopying, microfilm, electronic sys-
tems etc. without our prior written permission. Brand names, registered trademarks
etc. used in this manual, even if not explicitly marked as such, are not to be consid-
ered unprotected by trademarks law. They are the property of their respective own-
er.
This manual is the original documentation for the OPUS spectroscopic software.
Table of Contents
1 Starting OPUS . . . . . . . . . . . . . . . . . . . . . 1
1.1 Us e r I n t e rf a c e . . . . . . . . . . . . . . . . . . . . . . . . 4
2 OPUS Basics . . . . . . . . . . . . . . . . . . . . . . 7
2.1 Ba si c S oftw are F ea tu r es . . . . . . . . . . . . . . . . 7
2.2 Workspace . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Ac c e s s C o n t rol . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Paths and Directories . . . . . . . . . . . . . . . . . 12
2.5 Us e r R i g h t s . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 .6 Bro wser . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 . 6 . 1 Ma k e i n v i s i b l e . . . . . . . . . . . . . . . . . 16
2 . 6 . 2 Po p - u p men u s . . . . . . . . . . . . . . . . . 17
2 .7 Wi zard . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2 . 7 . 1 Ma k e i n v i s i b l e . . . . . . . . . . . . . . . . . 20
2.7.2 Levels . . . . . . . . . . . . . . . . . . . . . . . 21
2.8 Spectrum Window . . . . . . . . . . . . . . . . . . . . 26
2.8.1 Enlarging or reducing the spectrum
w i nd o w . . . . . . . . . . . . . . . . . . . . . . 27
2.8.2 Fixing the size of the spectrum window 27
2.8.3 More than one s p e ct r u m i n t h e
spectrum window . . . . . . . . . . . . . . . 27
2.8.4 Pop-up menu in t he spectrum window . 28
2.8.5 Zooming in a spectral range . . . . . . . . 29
2 . 8.6 Zo o mi n g ou t a spectr al range . . . . . . . 30
2 . 8 . 7 Sc a l i n g a l l s p e c t r a . . . . . . . . . . . . . . 31
2.8.8 Maximizing each spectru m . . . . . . . . . 31
2 . 8.9 Sca l i n g o rdi na te . . . . . . . . . . . . . . . . 31
2.8.10 Shifting curve . . . . . . . . . . . . . . . . . . 32
2.8.11 Activating crosshair . . . . . . . . . . . . . 32
2.8.12 Deactivating crosshair . . . . . . . . . . . . 33
2.8.13 Changing color of th e spe ctrum . . . . . 33
2.8.14 Removing spectrum from spectrum
window . . . . . . . . . . . . . . . . . . . . . . 33
2.8.15 Adding annotations to spectrum . . . . . 34
2.8.16 Copying one spectrum . . . . . . . . . . . . 35
3 File . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.1 Load File . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.2 Unload F ile . . . . . . . . . . . . . . . . . . . . . . . . 65
3 .3 Unlo ad A ll Fi les . . . . . . . . . . . . . . . . . . . . . 66
3.4 Save File /Save File As . . . . . . . . . . . . . . . . 66
3.5 Un d o C h a n g e s . . . . . . . . . . . . . . . . . . . . . . 71
3.6 Send File . . . . . . . . . . . . . . . . . . . . . . . . . . 72
3.7 Send File to GRAMS . . . . . . . . . . . . . . . . . . 73
3.8 Send File to InStep . . . . . . . . . . . . . . . . . . . 74
3.9 Delete Data Blocks . . . . . . . . . . . . . . . . . . . 74
3 .10 Exte rnal P rog ram . . . . . . . . . . . . . . . . . . . . 76
3.11 VisualBasic Script . . . . . . . . . . . . . . . . . . . 82
3.12 Load Data Point Table . . . . . . . . . . . . . . . . 84
3 .13 Scan OP US Fil es . . . . . . . . . . . . . . . . . . . . 85
3.14 Find OPUS Files . . . . . . . . . . . . . . . . . . . . . 87
3 .15 Clon e E ntry /Cl one Origi nal . . . . . . . . . . . . . 89
3.16 Add Comment . . . . . . . . . . . . . . . . . . . . . . 90
3.17 RS 232 Communication . . . . . . . . . . . . . . . . 91
3.18 Copy Data Block . . . . . . . . . . . . . . . . . . . . . 96
3.19 Change Data Block Type . . . . . . . . . . . . . . . 97
3.20 C o m p a re P a ram e t e rs . . . . . . . . . . . . . . . . . 98
3.21 LabJournal . . . . . . . . . . . . . . . . . . . . . . . 100
3.21.1 Lab journal view . . . . . . . . . . . . . . 1 02
4 Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
4.1 Ed i t P a r a met e r . . . . . . . . . . . . . . . . . . . . . 110
4 .2 Repl ay . . . . . . . . . . . . . . . . . . . . . . . . . . 112
4.3 Attach Video Image . . . . . . . . . . . . . . . . . 113
4.3.1 Further setting options . . . . . . . . . . 114
4.3.2 Spectrum file in the browse r . . . . . . 1 16
4.3.3 Report view of graph d a t a b l o c k . . . . 116
4.4 Add Information . . . . . . . . . . . . . . . . . . . . 117
4.5 Setup Information Mask . . . . . . . . . . . . . . 1 21
4 .6 Import Structure . . . . . . . . . . . . . . . . . . . . 124
4.7 Edit/Create Structure . . . . . . . . . . . . . . . . 125
4.8 Attach Structure . . . . . . . . . . . . . . . . . . . . 129
4.9 Convert 3-D Jcamp . . . . . . . . . . . . . . . . . . 1 30
4.10 Create Spectrum from Struc t u re . . . . . . . . . 134
4.11 Setup Filelist . . . . . . . . . . . . . . . . . . . . . . 136
4.12 C ut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
4.13 Copy /Paste . . . . . . . . . . . . . . . . . . . . . . 140
4 .14 Copy to Image Fi le . . . . . . . . . . . . . . . . . . 140
5 View . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
5 .1 Toolb ars . . . . . . . . . . . . . . . . . . . . . . . . . . 141
5 .2 Sta tus Bar . . . . . . . . . . . . . . . . . . . . . . . . 142
5 .3 Bro wser . . . . . . . . . . . . . . . . . . . . . . . . . . 142
5 .4 Wi z ard Ba r . . . . . . . . . . . . . . . . . . . . . . . . 142
5.5 Information View . . . . . . . . . . . . . . . . . . . . 143
6 Window . . . . . . . . . . . . . . . . . . . . . . . . 145
6.1 Ne w S p e c t r u m W i n d o w . . . . . . . . . . . . . . . 145
6.2 New 3D Window . . . . . . . . . . . . . . . . . . . . 146
6.3 New Report Window . . . . . . . . . . . . . . . . . 146
6 .4 Ne w Reg is te re d W ind ow . . . . . . . . . . . . . . . 148
6.4.1 TextView . . . . . . . . . . . . . . . . . . . . 1 49
6.5 Cascade . . . . . . . . . . . . . . . . . . . . . . . . . 150
6.6 Ti l e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
6.7 Ar r a n g e I c o n s . . . . . . . . . . . . . . . . . . . . . . 152
7 Measure . . . . . . . . . . . . . . . . . . . . . . . . 153
7.1 Ad va n ce d Me a su rement . . . . . . . . . . . . . . 1 57
7 . 1 . 1 Ba s i c . . . . . . . . . . . . . . . . . . . . . . . 1 61
7 . 1 . 2 Ad v a n c e d . . . . . . . . . . . . . . . . . . . . 1 64
7 . 1 . 3 Op t i c . . . . . . . . . . . . . . . . . . . . . . . 1 66
7.1.4 Acquisition . . . . . . . . . . . . . . . . . . . 1 66
7.1.5 FT . . . . . . . . . . . . . . . . . . . . . . . . . 166
7 . 1.6 D i sp l ay . . . . . . . . . . . . . . . . . . . . . 1 67
7.1.7 Background . . . . . . . . . . . . . . . . . . 168
7 . 1.8 Ch e ck S i gn a l . . . . . . . . . . . . . . . . . 169
7 . 1 . 9 Be a m P a t h . . . . . . . . . . . . . . . . . . . 1 73
7.1.10 Spectral Range Select i o n . . . . . . . . 174
7.1.11 Raman Measurements . . . . . . . . . . . 1 77
7.2 Setup Measurement Paramete rs . . . . . . . . . 178
7 . 2 . 1 Ba s i c . . . . . . . . . . . . . . . . . . . . . . . 1 80
7 . 2 . 2 Ad v a n c e d . . . . . . . . . . . . . . . . . . . . 1 81
7 . 2 . 3 Op t i c . . . . . . . . . . . . . . . . . . . . . . . 1 83
7.2.4 Acquisition . . . . . . . . . . . . . . . . . . . 1 85
7.2.5 FT . . . . . . . . . . . . . . . . . . . . . . . . . 193
7 . 2.6 D i sp l ay . . . . . . . . . . . . . . . . . . . . . 1 94
7.2.7 Background . . . . . . . . . . . . . . . . . . 194
7 . 2.8 Ch e ck S i gn a l . . . . . . . . . . . . . . . . . 194
7 . 3 Rou ti ne Mea su remen t . . . . . . . . . . . . . . . . 1 95
7.4 Repeated Measurements . . . . . . . . . . . . . . 196
7.5 Rapid Scan T ime Resolved M e a s u r e m e n t . . 198
7.6 Direct Command Entry . . . . . . . . . . . . . . . 199
7.7 Optic Setup and Service . . . . . . . . . . . . . . 201
7 . 7 . 1 Op t i c a l B e n c h . . . . . . . . . . . . . . . . . 2 01
7 . 7 . 2 De v i c e s / O p t i o n s . . . . . . . . . . . . . . . 2 03
7.7.3 Optic Communication . . . . . . . . . . . 2 06
7.7.4 Interferometer/AQP . . . . . . . . . . . . . 2 07
7 . 7.5 Exp o rt Op ti on s . . . . . . . . . . . . . . . . 2 08
7 . 7 . 6 Se r v i c e . . . . . . . . . . . . . . . . . . . . . 2 09
7.7.7 Control Panel . . . . . . . . . . . . . . . . . 211
7.8 Optics Diagnostics . . . . . . . . . . . . . . . . . . 212
7.8.1 Status display for non-Ethernet
based spectrometers . . . . . . . . . . . . 216
7 . 9 Te mp e rat u r e C o n t rol . . . . . . . . . . . . . . . . . 218
7.10 Interleaved Time Resolved Measurement . . 220
7.11 Sample Wheel Measurement . . . . . . . . . . . 222
7 .12 Twister Control/Ca libration . . . . . . . . . . . . 2 29
7.12.1 Calibrating Twister . . . . . . . . . . . . . 2 30
8 Manipulate . . . . . . . . . . . . . . . . . . . . . . 247
8.1 Baseline Correction . . . . . . . . . . . . . . . . . 249
8.1.1 Standard Baseline Correction . . . . . . 249
8.1.2 Interactive Baseline Correction . . . . 251
8.1.3 Baseline Correction Methods . . . . . . 254
8.1.4 Examples of Baseline Correction . . . 257
8.2 Sp e c t ru m S u b t rac t i o n . . . . . . . . . . . . . . . . 259
8.3 AB <-> TR Conversion . . . . . . . . . . . . . . . 2 64
8.4 St rai gh t Li ne G en e ration . . . . . . . . . . . . . . 266
8.5 Spectrum Calculator . . . . . . . . . . . . . . . . . 267
8.6 Cu t . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 271
8.7 No r mal i z a t i o n . . . . . . . . . . . . . . . . . . . . . . 273
8.8 Make Compatible . . . . . . . . . . . . . . . . . . . 276
8 .9 Convert Spe ctra . . . . . . . . . . . . . . . . . . . . 279
8 .10 Smo oth . . . . . . . . . . . . . . . . . . . . . . . . . . 282
8.11 Derivative . . . . . . . . . . . . . . . . . . . . . . . . 286
8.12 F r equ e n cy C al i b rati on . . . . . . . . . . . . . . . . 2 88
8.12.1 Frequency Calibration Theory . . . . . 290
8.13 Raman Correction . . . . . . . . . . . . . . . . . . . 292
8.13.1 Raman Correction Theo ry . . . . . . . . 295
8.14 B l ack B o d y . . . . . . . . . . . . . . . . . . . . . . . 296
8.14.1 Black Body Theory . . . . . . . . . . . . . 2 97
8.15 Interferogram to Spectrum . . . . . . . . . . . . . 299
8 .16 Inverse FT . . . . . . . . . . . . . . . . . . . . . . . . 309
8.17 Post Zerofilling . . . . . . . . . . . . . . . . . . . . . 310
8.18 Fourier Self-Deconvolution . . . . . . . . . . . . 3 13
8.18.1 Fourier Self-Deconvolution Theory . . 316
8.19 S y mm e t ri c F T . . . . . . . . . . . . . . . . . . . . . . 319
8.20 Kramers-Kronig-Transformat i o n . . . . . . . . . 320
8 .21 Spli t Interferog rams . . . . . . . . . . . . . . . . . 322
8.22 Spectrum from Interferogram s . . . . . . . . . . 324
8.23 E xtrap o l ati on . . . . . . . . . . . . . . . . . . . . . . 325
8.23.1 Extrapolation to 0cm-1 . . . . . . . . . . 3 26
8.23.2 Extrapolation to (infin i t y ) . . . . . . . . 328
8.24 Extended ATR Correction . . . . . . . . . . . . . 329
8 .25 1/cm <-> µm, nm . . . . . . . . . . . . . . . . . . . 335
8.25.1 Frequency Conversion Theory . . . . . 337
8.26 Averaging . . . . . . . . . . . . . . . . . . . . . . . . 341
8.26.1 Averaging Theory . . . . . . . . . . . . . . 3 43
8.27 Merge Spectral Ranges . . . . . . . . . . . . . . . 344
8.28 Atmospheric Compensation . . . . . . . . . . . . 349
8.28.1 Atmospheric compensation as part of
the measurement process . . . . . . . . 354
8.29 Straylight Correction . . . . . . . . . . . . . . . . . 357
8.29.1 Create new co rrecti on sp ectra . . . . . 3 59
8.29 .2 Corre ctio n A lgorithm . . . . . . . . . . . . 3 62
8.29.3 Straylight correction as part of the
measurement process . . . . . . . . . . . 364
8.30 Noise Generation . . . . . . . . . . . . . . . . . . . 367
8.31 M ovi ng Mea n . . . . . . . . . . . . . . . . . . . . . . 368
8.32 Make Mo notone . . . . . . . . . . . . . . . . . . . . 369
8.33 R ad i ometri c cal i b rati on . . . . . . . . . . . . . . . 371
8.33.1 Introduction . . . . . . . . . . . . . . . . . . 371
8.33.2 Theoretical background information . 373
8.33.3 Standard procedur e of radiometric
c a l i b ra t i o n . . . . . . . . . . . . . . . . . . . 3 76
8.33.4 Customizing Manipulat e menu . . . . . 3 76
8.33.5 Generating radiomet ric calibration file 377
8.33.6 Performing radiometri c calibr ation . . 380
8.33.7 Radiometric calibratio n n o t p a s s e d . . 383
9 Evaluate . . . . . . . . . . . . . . . . . . . . . . . . 385
9 .1 Curve Fi t . . . . . . . . . . . . . . . . . . . . . . . . . 386
9 . 1.1 Se tti ng up mo del . . . . . . . . . . . . . . . 3 88
9.1.2 Calculating Model . . . . . . . . . . . . . . 3 94
9.1.3 Saving Model . . . . . . . . . . . . . . . . . 396
9 . 1 . 4 Cu r v e F i t T h e o ry . . . . . . . . . . . . . . . 3 97
9.2 Integration . . . . . . . . . . . . . . . . . . . . . . . . 400
10 Display . . . . . . . . . . . . . . . . . . . . . . . . . 539
1 0.1 Back /Forward . . . . . . . . . . . . . . . . . . . . . 540
1 0.2 Sta cked . . . . . . . . . . . . . . . . . . . . . . . . . . 540
1 0.3 Scale Al l /S cale Y . . . . . . . . . . . . . . . . . . . 542
11 Print . . . . . . . . . . . . . . . . . . . . . . . . . . 549
1 1 . 1 P ri n t S p e ctra . . . . . . . . . . . . . . . . . . . . . . 550
11.2 Setup Analysis Report Templa t e . . . . . . . . 556
11.2.1 Dialog . . . . . . . . . . . . . . . . . . . . . . 5 57
1 1 . 2 . 2 Re p o r t T y p e s . . . . . . . . . . . . . . . . . 5 58
11.2.3 Setting options for report data . . . . . 558
1 1 . 2 . 4 Re p o r t S t r u c t u r e . . . . . . . . . . . . . . . 5 59
11.2.5 How to set up a templa t e ? . . . . . . . . 559
11.2.6 Examples of a single sample report
te mpl ate . . . . . . . . . . . . . . . . . . . . 5 62
11.2.7 Special features in case of a multi
sa mp l e rep o rt . . . . . . . . . . . . . . . . . 5 63
1 1 . 2 . 8 Mu l t i E v a l u a t i o n . . . . . . . . . . . . . . . 5 67
11.3 Generate Analysis Report . . . . . . . . . . . . . 568
11.3.1 Report Layout . . . . . . . . . . . . . . . . . 569
11.3.2 Standard Template . . . . . . . . . . . . . 570
1 1 .3 .3 Re p o rt O utp u t . . . . . . . . . . . . . . . . . 5 71
1 1.4 Qui ck Prin t . . . . . . . . . . . . . . . . . . . . . . . 572
1 1.5 New Layout . . . . . . . . . . . . . . . . . . . . . . . . 572
11.5.1 PLE View . . . . . . . . . . . . . . . . . . . . 5 73
1 1 . 5 . 2 Po p - u p men u . . . . . . . . . . . . . . . . . 5 78
11.5.3 Resizing Drawing Area . . . . . . . . . . 5 79
11.5.4 Adding elements, modifying spectrum
frame . . . . . . . . . . . . . . . . . . . . . . . 5 85
11.5.5 Adding text and modi f y i n g t e x t f ra m e 597
11.5.6 Adding table and modifying table
frame . . . . . . . . . . . . . . . . . . . . . . . 6 01
1 1 . 6 O p e n La yo u t . . . . . . . . . . . . . . . . . . . . . . . 605
1 1.7 Pri nt /Print P revi ew . . . . . . . . . . . . . . . . . . 606
1 1 . 8 P ri n t S etu p . . . . . . . . . . . . . . . . . . . . . . . . 606
12 Macro . . . . . . . . . . . . . . . . . . . . . . . . . . 607
1 2 .1Sc ri p t Re c o rde d H is to ry . . . . . . . . . . . . . . 608
1 2.2 Insert MyInstru ment . . . . . . . . . . . . . . . . . 609
1 2.3 Macro C onverter . . . . . . . . . . . . . . . . . . . . 610
1 2.4 Run Macro . . . . . . . . . . . . . . . . . . . . . . . . 611
12.5 Macro Debuggen . . . . . . . . . . . . . . . . . . . 611
1 2.6 Edit Macro . . . . . . . . . . . . . . . . . . . . . . . . 612
12.6.1 Macro editor vie w . . . . . . . . . . . . . . 612
1 2 .6 .2 Co r re ct in g m a c ro s y n t a x e rr o rs . . . . 613
12.6.3 Adding special command lines . . . . . 613
12.6.4 Adding OPUS commands . . . . . . . . . 614
12.6.5 Adding variables . . . . . . . . . . . . . . . 6 15
12.6.6 Adding macros to the toolb ar . . . . . . 6 16
12.6.7 Adding macros to a m e n u . . . . . . . . 617
12.7 Compile Macro . . . . . . . . . . . . . . . . . . . . . 618
12.8 New Procedure . . . . . . . . . . . . . . . . . . . . . 618
1 2 . 8 . 1 Po p - u p men u . . . . . . . . . . . . . . . . . 6 25
12.9 Edit Procedure . . . . . . . . . . . . . . . . . . . . . 633
12.10 Run Procedure . . . . . . . . . . . . . . . . . . . . . 634
13 Validation . . . . . . . . . . . . . . . . . . . . . . 635
13.1 Setup OVP . . . . . . . . . . . . . . . . . . . . . . . . 6 3 7
13.1.1 Measure Laser Wavenumber . . . . . . 6 40
13.1.2 Setting up OVP tests . . . . . . . . . . . . 6 44
1 3.2 Run OV P Tests . . . . . . . . . . . . . . . . . . . . 676
13.2.1 Running OVP test using the Instrument
S t a t u s d i a l o g . . . . . . . . . . . . . . . . . 6 80
13.2.2 Start ing OVP tests for mul ti p l e tes t
channels sequentially . . . . . . . . . . . 6 8 3
13.3 Measuring with test channel-specific laser
w a ve n u mbe r . . . . . . . . . . . . . . . . . . . . . . . 684
1 3 . 4 O V P i n un a tte n d e d mode . . . . . . . . . . . . . . 686
13.5 OVP Trend Chart . . . . . . . . . . . . . . . . . . . 688
13.5.1 Displaying trend chart . . . . . . . . . . . 690
13.5.2 Pop-up menu in trend ch a rt . . . . . . . 6 9 3
13.6 Automatic Accessory Recogni t i o n . . . . . . . . 6 9 6
13.6.1 Activating automatic accessory re-
c o g n i t i o n . . . . . . . . . . . . . . . . . . . . 696
13.6.2 Registering new accesso ri es . . . . . . 6 97
13.6.3 Performing OVP tests . . . . . . . . . . . 7 00
13.6 .4 Automa ti c Pe forman ce Te st . . . . . . . 7 00
14 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . 709
14.1 Logout . . . . . . . . . . . . . . . . . . . . . . . . . . . 710
1 4.2 Setup JCAMP <-> OPUS C onversion . . . . . 711
14.3 User Management . . . . . . . . . . . . . . . . . . . 713
14.4 Change User Passw o r d . . . . . . . . . . . . . . . 721
14.5 Setup User Macro List . . . . . . . . . . . . . . . . 724
1 4.6 Regi ster OPU S . . . . . . . . . . . . . . . . . . . . . 726
14.7 Customize Toolbars . . . . . . . . . . . . . . . . . . 728
1 4 . 7 . 1 Mo v i n g t o o l b a r . . . . . . . . . . . . . . . . 7 29
1 4 .7.2 Ad d i ng to o l ba r . . . . . . . . . . . . . . . . 7 29
1 4 . 7 . 3 De l e t i n g t o o l b a r . . . . . . . . . . . . . . . 7 29
14.7.4 Customizing toolbar . . . . . . . . . . . . 7 30
1 4 .7.5 Crea ti ng T oo l bar . . . . . . . . . . . . . . . 7 33
14.7.6 Adding or deleting icons . . . . . . . . . 736
14.7.7 Copying icons . . . . . . . . . . . . . . . . . 7 37
1 4 .7.8 De p i cti n g i co n s . . . . . . . . . . . . . . . . 7 37
14.7.9 Grouping icons in the toolbar . . . . . . 738
14.7.10 Creating user-defined icons . . . . . . . 739
14.8 Customize Menus . . . . . . . . . . . . . . . . . . . 741
14.8.1 Adding or deleting menu command s . 744
14.8.2 Moving or copying menu commands . 745
14.8.3 Moving or copying menus and men u
commands in toolbar . . . . . . . . . . . . 745
14.8.4 Creating sub-menus . . . . . . . . . . . . 7 45
14.8.5 Grouping menu commands . . . . . . . . 747
1 4.9 Shortcuts . . . . . . . . . . . . . . . . . . . . . . . . . 748
14.9.1 Assigning shortcuts . . . . . . . . . . . . . 7 48
14.9.2 Deleting shortcuts . . . . . . . . . . . . . . 7 50
14.9.3 Resetting shortcuts . . . . . . . . . . . . . 7 50
1 4 . 1 0 Us e r S e t t i n g s . . . . . . . . . . . . . . . . . . . . . . 751
14.10.1 Instrument Test (only in case of non-
OVP supported instruments) . . . . . . 761
1 4 . 1 1 Open P l a n n e r . . . . . . . . . . . . . . . . . . . . . . 766
14.1 1.1 V iew Opti ons . . . . . . . . . . . . . . . . . 7 67
14.11.2 Creating planner items . . . . . . . . . . 769
15 Help . . . . . . . . . . . . . . . . . . . . . . . . . . . 777
1 5.1 Help Top ics . . . . . . . . . . . . . . . . . . . . . . . 777
1 5.2 Onl ine Docs . . . . . . . . . . . . . . . . . . . . . . . 777
15.3 F T-IR T u to ri al . . . . . . . . . . . . . . . . . . . . . . 778
15.4 A b o u t OP U S . . . . . . . . . . . . . . . . . . . . . . . 778
16 A p p e n di x A . . . . . . . . . . . . . . . . . . . . . 779
B C
It shows the serial number of your OPUS copy and the registered
user name. The basic OPUS package integrates these additional
licensed packages as an all-in-one application.
C
D
H G
I
Figure 4: OPUS user interface
A Pull-down menus
Allow access to all OPUS functions.
B Toolbar
Enables quick access to frequently used functions.
C Browser
Contains the spectrum file loaded together with the data
blocks.
D Spectrum window
Shows the spectrum loaded
E Overview window
Contains the full frequency range of the spectrum selected
F OPUS view
The OPUS view you work with is indicated by a tab:
Each tab
has a different color to be able to distinguish between them.
If several views are open simultaneously, click the respective
tab or use the buttons on the left and right to change
between the different views.
G Instrument status
Gray: no spectrometer connected
Green: spectrometer connected
Yellow: warning
Red: error
H Online help
The online help is context sensitive.
I Status bar
Indicates which background task is running.
Drag & Drop OPUS supports drag & drop, a method to move
graphical elements by using the mouse. You can
drag single elements of a graphical user interface
from one particular position and drop them at a dif-
ferent position.
Drag & drop simplifies e.g. the Copy and Paste pro-
cess of files. In this case you have to select the file
first, press the left mouse button while moving the
file to the respective dialog box (or window). If you
release the mouse button, the file will be loaded into
the dialog box (or window) selected. Double clicking
the spectra file name allows to move spectra files to
a specific dialog.
2.2 Workspa ce
These settings are stored in one file which has the extension *.ows.
To m o d i f y t h e w o r k s pa c e o r c r e a t e a n e w
one...
...you need the right to do so. On the Setup menu, click User Set-
tings. Activate the respective check boxes on the Rights tab.
With OPUS 7.5 and higher the path structure of OPUS adapts to
the one used in Microsoft Windows. Chapter 2.4 describes the
default path structure used in OPUS.
User path
The different user groups have the following user rights in OPUS:
The user rights are defined by the User settings command on the
OPUS Setup menu (see chapter 14.10). The Administrator has to
ensure that the Operator has only access to special commands
within the OPUS workspace assigned to the Operator.
2.6 B rowser
The OPUS browser shows the spectrum files together with the cor-
responding data blocks.
To e n l a r g e o r r e d u c e t h e s i z e o f t h e
b r o w s e r w i n d o w. . .
To h a v e t h e b r o w s e r w i n d o w d i s p l a y e d a s
drop-down list
Left click the upper edge of the browser and draw it underneath the
toolbar. The OPUS browser window will be displayed as drop-down
list, including all the files loaded.
To undo this view click the upper edge of the drop-down list and
move the list to the original browser window position, i.e. to the left
side of the spectrum window.
B The blue document symbol indicates that the file has not yet
been manipulated. In case of manipulated files the docu-
ment symbol turns from blue to red.
The number next to the file name indicates the number of
copies made of the file loaded. You can load a file several
times into the OPUS browser window. In this case the copies
will be incremented accordingly.
C All data blocks which are part of the spectrum file are dis-
played. The data blocks displayed in figure 9 indicate a trans-
mission spectrum, single-channel spectrum, interferogram
and single-channel background spectrum.
Colored data block symbols indicate that the spectra or inter-
ferograms are displayed within the spectrum window. For fur-
ther information refer to chapter 2.12.
The icon allows to make the browser invisible. In this case the
browser is displayed as a blue bar below the toolbar on the left
side.
To m a k e t h e b r o w s e r v i s i b l e a g a i n . . .
...position the cursor onto the blue bar below the toolbar. Now, the
icon points to the right. As soon as you move the cursor outside
the browser on the OPUS interface, the browser will be hidden
again.
To h a v e t h e b r o w s e r p e r m a n e n t l y d i s-
played again...
To e d i t s p e c t r u m d a ta o p e n t h e p o p - u p
menu...
...by right clicking the name of the spectrum file. The pop-up menu
contains the following options:
A
B
D
E
To c h a n g e t h e s p e c t r u m d i s p l a y o p e n t h e
pop-up menu...
...by right clicking one data block of the spectrum file. For further
details on this menu refer to chapter 2.8.4.
To a d a p t t h e O P U S i n t e r f a c e o p e n t h e p o p -
up menu...
2.7 Wiz a rd
Note: If the wizard is not displayed, make sure that the Wiz-
ard Bar option is checked on the View menu.
A C
Use this menu to open the setup dialog for the command
selected as well as the online help.
C Up to 4 levels are available to edit spectra data.
D An instrument test to qualify the spectrometer performance
for daily routine work can be performed.
E This window contains a brief description of the OPUS com-
mand currently selected.
The icon allows to make the wizard invisible. If you click this
icon, the wizard is hidden in a blue bar below the toolbar.
To m a k e t h e w i z a r d v i s i b l e a g a i n . . .
...position the cursor on the blue bar below the toolbar. The
icon now points to the right. As soon as you move the cursor out-
side the wizard on the OPUS interface, the wizard will be hidden
again.
To h a v e t h e w i z a r d p e r m a n e n t l y d i s p l a y e d
again...
2.7.2 Levels
Acquire
Manipulate
C
B
E D
Evaluate
C
B
E D
Report
A Activate the check box to store the spectrum file, and note the
following:
• If you load a spectrum file (by means of the wizard) from a
particular directory, this file will be stored into this directory.
• If you measure a spectrum by means of the wizard, this
spectrum will be stored into the C:\Users\<User name>\My
Documents\Bruker\OPUS_<version>\MEAS directory. This
directory is generated after a measurement has been per-
formed.
B Activate the check box to print the spectrum file.
C Activate the check box to unload the spectrum file, i.e. to
remove the spectrum from the spectrum window.
D Activate the check box to measure another sample.
E Click Go, to perform the options selected between A - D.
Note: The default settings for the spectral range are from
4,000 to 400 wavenumbers and from 0 to 1.5 extinc-
tion units.
2.8.1 E n l a r g i n g o r r e d u c i n g t h e s p e c-
trum window
1 Position the cursor exactly between overview and spectrum
window.
➣The cursor changes to .
2 Now, re-size the spectrum window.
When working with more than one spectrum in the spectrum win-
dow simultaneously, it is recommended to activate on the Settings
menu On New File tab the Replace in current window option
button.
In this case the window currently loaded will not only be displayed
in the spectrum window but also in the browser (see
chapter 14.10).
The overview window below the spectrum window shows the com-
plete spectral range. The spectral range depicted on the white
background is exactly the one displayed in the spectrum window.
position the cursor on the interface line between white and gray
background, the cursor changes to . While pressing the left
mouse button, you can move the display limits of the spectral range
selected.
2.8.10 S h i ft i n g c u r v e
2.8.10.1 U n d o s h i ft c u r v e
1 Right click the spectrum displayed in the spectrum window.
2 From the pop-up menu that opens, select the Shift curve
command.
3 Select the Reset option.
2 . 8 . 11 Activating crosshair
2.8.14 R e m o v i n g s p e c t r u m f r o m s p e c-
trum window
1 Right click the spectrum displayed in the spectrum window.
2 From the pop-up menu that opens, select the Remove from
Display command.
➣The spectrum window is only removed from the spectrum
window but not unloaded from the browser. By drag & drop
you can reload the spectrum file from the browser to the
spectrum window.
2.8.15 A d d i n g a n n o ta t i o n s t o s p e c t r u m
1 Right click the particular position of the spectrum displayed
in the spectrum window.
2 From the pop-up menu that opens, select the Add Annota-
tion command.
➣By default, exactly at the cursor position an arrow is
inserted indicating the wavenumber of this data point. On
the browser, the data block is added to the spectrum
file.
➣If you position the cursor onto the arrow head of the anno-
tation, the cursor changes to . Press the left mouse but-
ton to move the arrow head.
2.8.15.1 E d i t i n g a n n o ta t i o n s
1 Right click the particular annotation.
2 From the pop-up menu that opens, select the Properties
command.
3 Enter the new text for the annotation.
4 If required, deactivate the Center text line or With arrow
checkboxes.
5 Click OK to confirm.
34 OPUS Reference Manual Bruker Optik GmbH
Spectrum Window Chapt. 2
2.8.15.2 R e m o v i n g a n n o ta t i o n s
1 Right click the particular annotation.
2 From the pop-up menu that opens, select the Remove com-
mand.
To change the display limits of the spectral range enter the values
into the particular entry field. If required, activate the Compressed
wavenumbers checkbox. The wavenumbers are displayed in a
more compressed manner in the spectrum window.
2.8.19.2 Axes
If the spectrum window contains more than one spectrum and the
spectra are displayed in stacked mode, you can define separate
axes for each spectrum if you select the X between option.
2.8.19.3 General
If you activate the Popup info check box and select an option, the
path and file name, or path name only, will be displayed as soon as
you position the cursor on the spectrum.
To m a k e t h e i n f o r m a t i o n v i e w v i s i b l e
again...
...position the cursor on the blue bar. The icon now points to
the right. As soon as you move the cursor outside the information
view on the OPUS interface, the view will be hidden again.
To h a v e t h e i n f o r m a t i o n v i e w p e r m a n e n t l y
displayed again...
2.10.2 O p e n a s e pa r a t e r e p o r t w i n d o w
To d i s p l a y t h e v i e w p r o p e r t i e s s e t . . .
...right click into the report window and select the Properties option
from the pop-up menu. The View Properties dialog box opens.
2 . 11 Too l b a r
AB Absorption spectrum
ANNO Annotation
CLASS Classification
3D 3D spectrum
HISTORY Protocol
IFG Interferogram
INTERPR Interpretation
KM Kubelka-Munk spectrum
ME Multi Evaluation
METHOD Method
QC Quick Compare
SIGNAT. Signature
TR Transmittance spectrum
TRC Trace
2.13 Decimals
The OPUS status light can have the following colors: green, gray
yellow and red. The following table explains the meaning of these
colors.
If you click the instrument status light, the Diagnosis dialog pops
up. This dialog shows the status of the single hardware compo-
nents in the upper icon row.
The second and third row refer to the possible active test channels.
For further information on this subject refer to chapter 7.8 and 13.
The following table includes the file name extensions used in the
OPUS software.
Extension Definition
Extension Definition
Extension Definition
H o w t o s e t u p c o m m a n d l i n e pa r a m e t e r s ?
1 Click the Windows Start button and right click the OPUS
icon.
2 A menu appears. Select Properties. The following dialog
opens:
1 Click the Windows Start button and right click the OPUS
icon.
2 A menu appears. Select Properties. The following dialog
opens:
Language Parameter
Chinese /LANGUAGE=CHINESE
English /LANGUAGE=ENGLISH
French /LANGUAGE=FRENCH
German /LANGUAGE=GERMAN
Japanese /LANGUAGE=JAPANESE
Korean /LANGUAGE=KOREAN
Polish /LANGUAGE=POLISH
Russian /LANGUAGE=RUSSIAN
Spanish /LANGUAGE=SPANISH
If OPUS hangs for quite some time and cannot be closed by the
icon, you can abort the program by using the Windows Task Man-
ager. Data not stored before aborting the program will be lost.
This also applies to the workspace settings made during the pro-
gram run (e.g. toolbar configuration, directory name etc.)
Note that in this case possible changes made, e.g. in the work-
space, are only stored temporarily. They will not be available when
restarting OPUS from CD.
A B C D E F G
Principle • Load the spectrum file and select the Unload File
command.
• If required, drag & drop the spectrum file into the
File(s) to unload selection field.
• Click Unload.
Definition The Unload All Files command removes all files from
the OPUS browser and spectrum window.
If you select the Save File command, the following dialog opens:
C
D
A The name of the spectrum file lately used is shown in the File
to save selection field.
B In this entry field you enter the appropriate file name.
C You change the path shown either manually or by using the
Change Path button.
D If you activate the Overwrite check box, existing files with the
same file names will be replaced by new files. Note: This
option cannot be used when working in GLP1 mode.
E To save the spectrum file click Save.
The Mode tab allows to set up the output format for the spectrum
data to be saved.
A Apart from the OPUS and Data Point Table format there are
also other formats available, e.g. GRAMS1 (Galactic), InStep2
(Pirouette) or JCAMP.DX as an additional ASCII format. The
spectrum can also be saved in the XML format.
B Additional saving options are available, provided you have
checked the OPUS format option button first. If you check the
Save all option button, all data blocks will be saved. You can
also move the original spectrum file, or remove all copies of
the spectrum file.
C Click Save to save the spectrum file.
The Data Point Table tab allows to define the structure of the out-
put file.
A You enter the number format for x and y values, the number
of decimals as well as the separator format into these entry
fields. If you activate the Separator is TAB check box, the col-
umns will be separated by tabs.
B You can limit the output file size to 64KB by checking the
Max. 64KB option button.
C This group field is enabled if you have activated the XML
option button on the Mode tab. Two types of data format are
available for converting the binary area of a spectrum:
The Base 64 data format converts the binary spectrum area
into coded text. Advantage: as the data are more compressed
than in case of the X/Y Table format, less storage capacity is
required.
The X/Y Table data format converts the binary spectrum area
into a data table. This format is more readable than the
Base 64 format.
D To save the spectrum file click Save.
70 OPUS Reference Manual Bruker Optik GmbH
Undo Changes Chapt. 3
A The red document symbol indicates that the spectrum file has
been edited.
B As soon as you click Undo, the color of the document symbol
changes from red to bright blue in the browser.
Principle • Load the spectrum file and select the Send File
command.
• If required, drag & drop the spectrum file into the
File(s) to send selection field.
• Click Send.
Note: If you use the mail program for the first time, you will
be prompted to configure it. The data will be sent as
attachment. If the addressee uses an appropriate mail
program, it is possible to drag and drop the attach-
ment from this program directly to the OPUS spec-
trum window.
Principle • Load the spectrum file and select the Send File to
GRAMS command.
• If required, drag & drop the spectrum file into the
File(s) to send selection field.
• Click SendToGrams.
Analog to Send File to GRAMS (chapter 3.7) you can also send a
file to the InStep program.
Principle • Load the spectrum file and select the Delete Data
Blocks command.
• If required, drag & drop the spectrum file into the
File(s) to delete data blocks selection field.
• Click Delete.
Basic Features
You can only start external programs which run on Windows as the
appropriate system functions will directly be activated when start-
ing the program. You can also use batch files which can be identi-
fied by the following extensions:
•*.exe
•*.com
•*.bat
Selecting Program
A Usually, each program runs in its own window. The size and
type of this window can be defined, according to the following
options:
• Normal: the program runs in a window of standard size.
• Maximized: the program runs in a window of maximum size.
• Minimized: the program runs in a minimized window, i.e. the
window is reduced in size and can be enlarged by selecting
the respective icon on the task bar.
• Hidden: the program starts without any visible window and
must be made visible, if desired, by using special com-
mands.
• Run as private VDM (for 16 bit applications only): if you acti-
vate this check box, you allocate a separate VDM Virtual
DOS Machine) to the external program. This may cause
increased administrative costs for the systems, but will also
DDE Command
On this tab you can make settings which are called via the DDE
interface. This can apply to a server program which has to be
started or which has already been started.
A
C
B
script from the dialog that opens, the script name will be dis-
played in the Script entry field, but not started until you click
Execute.
Note: Directly after the OPUS installation the drop-down
list will be empty. Entries will be added to the list at the
beginning of a new program selection.
B If you activate this check box, OPUS waits for the script to ter-
minate after starting the script. This is useful if you activate
this function from within macros and scripts and you have to
wait for the end of the script loaded, or have to retrieve return
parameters.
C If you activate this check box, the script will run in the back-
ground instead of being displayed when starting the script.
D Additional parameters to be used in combination with the
script can be specified in this entry field. They will be
exchanged in string format.
E Drag & drop the spectrum data blocks from the browser to the
entry field. Make sure that the data blocks can be processed
by the script.
Make sure that you know the type of data point table
(type of spectra, configuration of x and y values). You
can select the data range directly from the table by
either numeric entry or interactive mode. The preview
window displays the data as spectrum.
Principle • Click the Load Data Point Table command and click
the respective file in the dialog that opens.
• Click Open after you have optionally made the set-
tings described in Definition.
Principle • Before you can use this search command, you have
to create a database.
• To do so click the Scan OPUS Files command. The
following dialog opens.
To c r e a t e a d a ta b a s e . . .
3 Click the Scan Files button after the settings have been
made. Note: Depending on the number of spectrum files
being stored, the scanning can take a while. During scan-
ning, the status bar indicates this background task by Scan
OPUS Files.
A If you want to find text in files, activate the check box. Enter
the text to be searched for into the entry field. You can specify
whether to find Exact wording only, any word, or All words
you have entered.
B You can even further narrow down the search by activating
the Find in period check box. Specify the date and select
between after, before or between this date to narrow down
your search.
Principle • In either case load the spectrum file and select the
respective command.
• If required, drag & drop the spectrum file into the
File(s) to clone selection field.
• Click Clone.
Principle • Load the spectrum file and select the Add Comment
command.
• If required, drag & drop the spectrum file into the
File(s) to comment selection field.
• Enter the respective comment into the Comment
text entry field.
• Click Add.
...in the data block of the spectrum file. This data block is
added to the spectrum file in the OPUS browser.
To h a v e t h e c o m m e n t d i s p l a y e d
Right click on the data block and select the Show Report
option. The report view of the data block is displayed, including the
comment(s) added.
Figure 60: Comments added and displayed in the history data block
A
G
B H
C
I
D
E
J
F
K
L M
The RS 232 Settings tab allows to set the different options for the
serial interface.
A
G
B H
I
C
Clicking the Change Settings button (H in figure 62) opens the fol-
lowing dialog:
• Parity
• Stop bits
• Flow control
The same dialog opens if you set up the parameters of the serial
interface by using the System Settings command of the operating
system.
Principle • Load the spectrum file and select the Copy Data
Block command.
• If required, drag & drop the data block(s) to be cop-
ied into the Block(s) to copy selection field.
• Define the file where to copy the data blocks in the
Destination file field.
• Click Copy Block.
Principle • Load the spectrum file and select the Change Data
Block command.
• If required, drag & drop the data block(s) to be
changed in to the Block(s) to change selection field.
• On the Choose block type drop-down list, select the
block type. It may be possible that not all data block
types are included in the drop-down list.
• Click Change Block Type.
A ft e r c o m pa r i n g . . .
Besides, the result view shows those parameters which are either
available in one spectrum file only, or in both files with different val-
ues (see figure 67).
3.21 LabJournal
To c r e a t e a l a b j o u r n a l . . .
D e p e n d i n g o n t h e u s e r r i g h ts . . .
...defined (see chapter 2.5) you either have limited or full access to
the lab journals in OPUS.
• Operator: you can only access the lab journal assigned
to your account.
• Administrator or Labmanager: you can access the lab
journals of other users as well.
A B
When renaming, the date of the first and last entry are added to the
previous file name. Thus, the new file name indicates the time
period of the different entries. The following format applies to the
new file name:
Example: LabJournal_Default_2006-03-15_2006-04-10
The lab journal can also be printed with each lab journal entry
being on a separate page. On the File menu, click Print.
Clicking again the particular column title restores the original sort-
ing order.
Confirm the settings made by clicking OK. The list will be sorted
again, considering only those evaluations or measurements
selected within the defined period of time.
To undo this setting double click again the Date column. Deactivate
the Restrict time span check box and click OK.
3.22 New/Open
• OPUS
• PLE
• VBScript
• GMacro
3.23 Print
Principle You can define the paper size and print-out format.
These settings will be standard for all OPUS print-
outs.
3.26 Exit
If you click the Exit command, the OPUS program will close.
4.1 E d i t P arameter
Principle • Load the spectrum file and select the Edit Parame-
ter command.
• If required, drag & drop the spectrum file into the
Select file selection field.
• Make the changes required.
• Click Change.
a. GLP = Good Laboratory Practice; regulations for laboratories and
producers on how to plan, perform and control tests and measure-
ments.
The Axes Labels tab allows to edit the axes labeling of the spectra,
and set a scaling factor for the axes. The settings made are consid-
ered when printing the spectrum file.
4.2 Replay
Principle • Load the spectrum file and select the Replay com-
mand.
• If required, drag & drop the spectrum file into the
File to replay selection field.
• Select the appropriate editing command.
• Click Replay.
a. GLP = Good Laboratory Practice; regulations for laboratories and
producers on how to plan, perform and control tests and measure-
ments.
Principle • Load the spectrum file and select the Attach Video
Image command.
• If required, drag & drop the spectrum file into the
Select OPUS spectrum selection field.
• Click the Browse button to select the respective
video image. The image is displayed as preview.
• Optionally, make further settings on the Options tab.
• Click Attach Video Image.
1. If you attach the video image to a 3D-MAP file, you can define the position of the
video image in relation to the chemical image. In this case, the x and y position
coordinates defined refer to the lower left end of the chemical image.
114 OPUS Reference Manual Bruker Optik GmbH
Attach Video Image Chapt. 4
After the video image has been attached, the detector area is
stored in the datablock as annotation. This data block is
attached to the spectrum file in the explorer. To show the data
block parameters right click the data block and select the
Show Report option.
Loading Configuration
Saving Configuration
4.3.2 Sp e c t r u m f i l e i n t h e b r o w s e r
4.3.3 R e p o r t v i e w o f g r a p h d a ta b l o c k
If you right click the graph data block ( ) attached to the spec-
trum file and displayed in the browser, the Show Report option
pops up. Select this option to open the report view.
A F
G
B
E d i t i n g i n f o r m a t i o n d a ta b l o c k
Edit the existing data or add new data. Alternatively, you can select
a different kind of information mask. In this case all inputs made so
far will be deleted in the dialog.
D i s p l a y i n g a n d p r i n t i n g i n f o r m a t i o n d a ta
block
You can also include information mask texts into the plot, by using
Tnn as parameter name (e.g. T01 = text in first line, T12 = text in
twelfth line).
4 You have to define all entry fields on the first page before
being able to open a new page by the Next Page button.
5 Save the information mask after you have filled in all lines.
2 Activate the New structure file check box. The path and file
name definition are displayed by default. If required, change
these data accordingly.
Sta r t i n g s t r u c t u r e e d i t o r
• Drawing/Editing bonds:
• Scaling bonds:
• Arranging bonds:
• Refining/Rotating bonds:
On the left side you find the abbreviations of the most important
atoms. If you click PSE, the periodic table of elements opens.
Basic features
##TITLE=Butanoic acid
...
...
##CAS REGISTRY NO=107-92-6
...
450.0 637 638 621 624 638 665 702 723 741 798
490.0 839 862 888 935 928 967 1021 1000 996 1098
530.0 1133 1222 1328 1382 1488 1531 1571 1558 1559 1570
570.0 1572 1637 1678 1756 1802 1942 1924 1905 1927 1920
...
Sta r t i n g c o n v e r s i o n
2 Enter the name of the 3-D JCAMP file into the 3-D JCAMP
file entry field. Either type in the file name manually with the
complete path, or click the Browse button to select the file in
the respective directory.
3 Optionally, you can enter the file name of the molecule struc-
ture file. Observe the additional information below. Note: If
you skip the Molecule structure file entry field, the OPUS
spectra created will not include a structure data block.
4 Define the OPUS format which the JCAMP spectra have to
be converted to. Observe the additional information below.
5 Activate the X Units check box. JCAMP files with x units
being indicated in wavelengths (i.e. nm or µm) cannot be
used in OPUS without converting the x units into wavenum-
bers (cm-1).
6 Click the Convert button to start the conversion. The conver-
sion progress will be indicated by the status bar on the bot-
tom right. Note: If one of the information lines includes
the sum formula, the molecular mass will be calculated
c:75014
C2H3Cl
2:(303,155,320,165;2)(337,156,320,165;1)
1:(3,Cl)
0:
The first line includes the CAS number followed by the sum for-
mula, bonding data and the names of foreign atoms.
You can also add the melting and boiling points of the different sub-
stances to the information data block of the OPUS file. Make sure
that you add the data at the end of the JCAMP structure file, line by
line as follows:
,CAS_NO,MELTINGPOINT,BOILINGPOINT,
Example:
,1571864,75.0,263.0,
In the Target group field (figure 91) you specify the OPUS format
which the JCAMP spectra have to be converted to. You can select
between three option buttons:
• 1-D OPUS files:
Each JCAMP spectrum is stored separately as OPUS
spectrum. The file name is a combination of the TITLE
and CAS REGISTRY NO information lines used in the
JCAMP file. If you activate the 1-D OPUS files option
button, you also have to specify the target path. Either
type in the target path name manually or click the
Browse button to select the path from the directory
structure.
• 3-D OPUS files:
The 3D JCAMP file is converted into a 3D OPUS file.
This is useful if the spectra have been acquired by chro-
matography, or if all spectra are a compound of always
the same substance which comes in different concentra-
tions. You can also define this substance concentration
in the information line for each spectrum. The format in
the JCAMP file must be as follows:
##CONCENTRATIONS= (NCU)
(<Glucose>, 2.100, %)
(<Lnvisco104>, 1.131, %)
In this case the substance name, quantity and quantity
unit will be specified in a text file (in table format) which
will be created apart from the OPUS file, and with the
same file name. This is helpful if the data are used to
perform a QUANT analysis. In this case the concentra-
tion data can be used from the text file using the clip-
board.
Either type in the target path name manually or click the
Browse button to select the path from the directory.
Specify the OPUS target file name without any exten-
sion. Note: If a 3D OPUS file already exists with this
name, it will not be overwritten, but the name will be
incremented.
• OPUS library
All spectra are stored in an OPUS library, including
spectra information and possible structure data blocks.
Either type in the target path name manually or click the
Browse button to select the path from the directory.
Specify the OPUS target file name without any exten-
sion. Note: If the OPUS library already exists with
this name, it will be overwritten.
I f t h e i n t e r p r e ta t i o n d a ta a r e n o t d i s p l a y e d
a u t o m a t i c a l l y, . . .
A B C D
To e v a l u a t e s i n g l e s p e c t r a . . .
...you can also add specific parameter values to the table in the
form of a column. Make sure that you always enter the right param-
eter code for the specific value into the entry field. The parameter
code can be obtained from the Macro Editor, for more information
see the OPUS Programming manual.
Example:
Select the Add Information command. Enter the following data into
the Compound name line (see figure 95):
• name of the OPUS function (A)
• result data block (B) as well as the
• corresponding element from the QUANT report (C), i.e.
main report number, sub-report number, line, column.
A B C
E v a l u a t i o n r e s u l ts f o r f i l e s w i t h i n t h e
spectrum file list...
When loading a spectrum file which contains a trace data block the
results are displayed within separate columns. Figure 96 exempli-
fies the results of an integration recently performed.
4.12 Cut
• *.bmp (Bitmap)
• *.pdf (Portable Data Format)
• *.emf (Enhanced Metafile)
5 .1 To o l b ars
5.2 Stat u s B a r
5.3 B rowser
Definition The browser contains the spectrum files with the cor-
responding data blocks.
5.5 I nformation Vi ew
6.2 N ew 3D Window
6.3 N e w R e p o r t Win d o w
Window To display...
Library libraries
Window To display...
6.4.1 Te x t Vi e w
To enter text select the TextView window type from the drop-down
list, which is indicated by the tab. There are
three possibilities to enter text:
1 Keyboard
Position the cursor into the window and enter text using your
keyboard.
2 Copy and Paste
It is possible to copy and paste text by using the shortcuts
CTRL+C and CTRL+V.
3 Drag & Drop a text file
Drag and drop a text file into the TextView window. You can
only drag one single file into the TextView window. If there
has already been a text file in the TextView window, this file
will be replaced and the text overwritten unless you press
the CTRL key during drag & drop. In this case the text of the
dragged file is appended to the content of the file which has
already been displayed in the window.
To s e l e c t t e x t
1 Move the cursor to the relevant text passage and press the
left mouse button.
2 You can also use the Shift and an arrow key to select a text
passage.
To d e l e t e t e x t
To s a v e t e x t
6.5 C ascade
6.6 Ti l e
Measuring Modes
Measurement Parameters
Predefined Parameters
W h e n d o y o u p e r f o r m w h i c h k i n d o f m e a-
surement?
2. Sample measurement
Right clicking the status bar (figure 106) opens a pop-up menu
which you can use to abort the measurement.
Select the corresponding task from the list and click the Stop Task
or Abort Task button. If the measurement has been completed, the
No Active Task message is displayed in the status bar.
7.1.1 Basic
To w o r k w i t h a u t o m a t e d pa r a m e t e r s . . .
1 ...click the Auto button (see figure 109) to open the following
dialog:
4 Click the Add at cursor position button. The entry field con-
tains the parameter in pointed brackets. Below the expanded
value is shown in full.
7.1.2 Advanced
The settings which can be made on this tab are not essential for
the actual measuring procedure. This tab is identical to the one of
the Setup Measurements Parameters command (see
chapter 7.2.2).
A You can specify the directory path, and the data file name of
the spectra file to be saved. Starting with the second mea-
surement, the data file name will be incremented, indicating
the current number of measurements. Enter the path either
7.1.3 Optic
The settings which can be made on this tab are not essential for
the actual measuring procedure. Which kind of settings you can
make depends on whether you work with predefined parameters
(see introduction to chapter 7).
7.1.4 Acquisition
7.1.5 FT
7.1.6 Display
On the Display tab you define the axes settings for the spectrum
display during measurement.
7.1.7 Background
To z o o m i n t o a s p e c i f i c s i g n a l r a n g e . . .
To z o o m o u t a s p e c i f i c s i g n a l r a n g e . . .
C
D
E
F
G
St e p p e r M o t o r A d j u s t m e n t
C h a n g i n g o p t i c pa r a m e t e r s
In this case the white area (figure 119) representing the currently
selected spectral range should not be larger than the filled area of
the colored bars.
1.The spectral range in which the spectrometer can select an IR signal depends on
the optical components (source, beamsplitter, detector etc.) used. To achieve
good measuring results make sure that the spectral range interactively defined
for measurement is not larger than the spectral range covered by the optical
components.
1 Position the cursor on the red line between the white and
gray area. The cursor changes to .
2 Press the left mouse button and move the line to the position
desired. If required, perform step 1 and 2 for both red lines.
O p t i c a l c o m p o n e n ts
The optical components best suited for the currently selected spec-
tral range of the spectrometer connected are indicated below the
graphics (A in figure 120).
A B
Figure 120: Example of best (A) and additional (B) optical components
7 . 1 . 11 R a m a n M e a s u r e m e n ts
• Laser power
In case of Raman measurements you can also specify
the laser power on the Basic tab.
In the standard configuration the laser will automatically
be switched off after each measurement. For some
applications, e.g. series measurements, you can deacti-
vate this laser option. Uncheck the Auto Laser OFF
check box.
The laser power can be set using the slider on the
Check Signal tab, i.e. the actual performance will be dis-
played.
• Frequency Range
On the Advanced tab the frequency range for data stor-
age will be displayed in relation to the laser excitation
frequency.
• Acquisition
On the Acquisition tab all frequencies are displayed in
relation to the laser excitation energy.
Definition This command allows to set up, edit and load param-
eters for a measurement experiment.
The warning symbols and their meaning are described in the fol-
lowing table:
7.2.1 Basic
7.2.2 A d v a n c ed
A You can specify the directory path, and the data file name of
the experiment file to be saved. Enter the path either manu-
7.2.3 O p t ic
7.2.4 A c q u i s i t io n
A When defining the upper and lower frequency limit the result-
ing actual frequency limits calculated may deviate from the
values defined and are therefore displayed next to the
Wanted high or low frequency limit entry field on the right.
Note: Restricting the sampling bandwidth reduces the
amount of data and may be useful when recording high
resolution spectra. However, electronic and/or optical filters
have to be used to ensure that signal intensity is 0 beyond the
selected frequency limits. Otherwise, the signals may be con-
voluted or folded to the selected range, and may adulterate
the intensity.
B The High pass filter limits the bandwidth within the high fre-
quency ranges. If you select On from the drop-down list, you
activate the filter. If you select Open, you deactivate the filter.
Note: This electronic filter is not available for all spec-
trometers.
The Low pass filter limits the bandwidth in the low frequency
ranges. As the mapping of IR wavenumbers to frequencies in
the low frequency range is proportional to the scanner veloc-
ity v, the fixed cut-off frequency f of the filter (in Hz) corre-
sponds to a wavenumber which depends on v as follows:
cm – 1 = f Hz v cms –1
Acquisition Mode
Mode Definition
Mode Definition
Fast Return and Forward/Backward modes are not available for all
types of spectrometers. On condition that the measurement times
are identical, the signal-to-noise ratio improves in the following
order:
S/N (no Fast Return) < S/N (Fast Return) < S/N (Forward/Back-
ward)
Correlation Mode
AQP instruments Full Igram Length All interferogram data points are compared to
with Raman each other. An interferogram is rejected if the
points contain more than 10 defective areas
or if the number of defective points exceeds
one eighth of the total number of interfero-
gram points.
Gate Scans CTR This mode rejects all scans during the time
when the DI1/Pin11 inlet is not connected to
mass Pin1. Scans during which the signal
changes are also rejected.
The measurement continues until the desired
number of scans is reached. No time-out nor
any maximum number of rejected scans are
part of this mode. You can interrupt measure-
ment as often as you like by gate periods.
Note: To feed the gate signal a trigger box
with BNC connectors has to be connected
to the spectrometer (Bruker part no:
E525/Z).
If this mode is not automatically selectable
from the drop-down list, you first have to
define it.
• On the Measure menu, click the Optic Setup
and Service command.
• Click the Devices/Options tab.
• Click the Setup button next to Correlation
mode.
• Click the Add New Item button in the dialog
that opens. Enter the following parameter:
512=Gate Scans CTR.
Gate Scans SCT This mode rejects all scans during the time
when the probe trigger is not activated, or if at
least 5V are not applied at the trigger cable
(E516/B). Scans during which the signal
changes are also rejected.
The measurement continues until the desired
number of scans is reached. No time-out nor
any maximum number of rejected scans are
part of this mode. You can interrupt measure-
ment as often as you like by gate periods.
Note: To feed the gate signal a trigger
cable with BNC connectors has to be con-
nected to the spectrometer (Bruker part
no: E516/B).
If this mode is not automatically selectable
from the drop-down list, you first have to
define it.
• On the Measure menu, click the Optic Setup
and Service command.
• Click the Devices/Options tab.
• Click the Setup button next to Correlation
mode.
• Click the Add New Item button in the dialog
that opens. Enter the following parameter:
1024=Gate Scans SCT.
7.2.5 FT
7.2.6 Display
The settings which can be made on this tab are identical to the one
of the Advanced Measurement command (see chapter 7.1.6).
7.2.7 Background
The settings which can be made on this tab are identical to the one
of the Advanced Measurement command (see chapter 7.1.7).
The settings which can be made on this tab are not essential for
the actual procedure of setting up measurement experiments. This
tab is identical to the one of the Advanced Measurement command
(see chapter 7.1.8).
The Options included on the other tabs are identical to those of the
Advanced Measurement command (see chapter 7.1).
7.5 R a p id S ca n Ti me R e so l ved
Measurement
This tab is used to realize the connection between OPUS and the
spectrometer.
7.7.2 Devices/Options
Configuring options
Step 1:
Next to Correlation mode, click the
Setup button.
Step 2:
• Select the 64=Signal Amplitude Lim-
its.
Step 3:
• Enter a lower and upper limit for the
signal amplitude.
• Confirm the entry by OK.
Step 4:
Click the Add New Item button.
Step 5:
• Enter a figure.
• Then, click the Setup button.
Step 6:
• Select Start... and Stop on Signal
Amplitude Limit.
• If necessary, modify the limits.
• Confirm the entry by OK.
Step 7:
The new entry is numbered and dis-
played.
Step 8:
The new correlation mode needs to be
selectable on the Measure menu, click-
ing the Acquisition tab.
7.7.4 Interferometer/AQP
7.7.6 Service
B This button resets both the counter of the total source operat-
ing time and the counter of the on/off source cycles after
replacing the source.
C To update the parameters of all motors click this button.
7.7.7 C o n tr o l P a n e l
The options available on this tab apply to the IFS 125HR spectrom-
eter only and are described in the IFS 125HR user manual.
OK (green):
Component is okay
WARNING (yellow):
The exact meaning of a warning depends
on the specific component. In case of the
source a warning means:
ERROR (red):
Component is defective. In this case mea-
suring is not possible.
INACTIVE (yellow):
The single tests of the particular test cat-
egory are disabled.
PASSED (green):
Instrument test configured or passed.
Test still valid.
FAILED (red):
Last instrument test failed.
To p e r f o r m f a u l t d i a g n o s t i c s o f a pa r t i c u -
lar spectrometer component
I n s t r u m e n t Te s t R e s u l t
The result of the instrument test (see figure 135) refers to the active
test channel. Each time you load an experiment file, the corre-
sponding channel is set as active channel. The status of this new
7.8.1 Sta t u s d i s p l a y f o r n o n - E t h e r n e t
based spectrometers
1 Right click the OPUS icon from the Windows Start menu.
2 Select the Properties command and click the Shortcut tab.
3 Enter the command line into the Target entry field. In
figure 138 the /DIAGTIMER=5 command polls the status
once every 5 minutes.
C
A
B
1 s t Ste p : C r e a t i n g S e t u p F i l e
On the so-called setup file you define the single sample wheel posi-
tions. For each sample wheel position to be measured you can
define a separate experiment method file, a user-defined file name
and path to save the spectra files measured.
A
B
2 n d St e p : M e a s u r e m e n t / R e f e r e n c e
If you deactivate the check box, only one single background mea-
surement will be performed before each sample measurement.
During measurement...
After measurement...
...all spectra measured will be saved, using the file name and path
indicated in the setup file. If the setup file does not contain any file
name and path, OPUS reads the two parameters from the experi-
ment.
3 r d St e p : E d i t i n g S e t u p F i l e
1 Load the particular setup file. Click the Load button on the
Setup tab.
2 To edit single sample wheel positions left click the corre-
sponding position (A in figure 142). The settings defined for
the position selected are displayed on the left. Edit these set-
tings, as desired.
3 To delete the settings made for all positions and the global
experiment file entry click the Clear All button.
If you have edited a loaded setup file and you click the Measure
tab, you will be asked whether these changes made are to be
stored. If you click Yes, the changes will be stored and become
effective during measurement.
General Features
Before the initial use the Twister stacker has to be calibrated. For
calibration you need a microtiter plate which you have to place into
the respective position to be calibrated.
7.12.1 C a l i b r a t i n g Tw i s t e r
7.12.2 C o n t r o l l i n g Tw i s t e r
Control by macro
Parameter Definition
TPT = Twister Plate Thick- Internal parameter which stores the microtiter
ness plate thickness measured
TPL = Twister Last Vertical Internal parameter which identifies the posi-
Stack tion from which the plate has been picked up
General Features
Before using the pump for the very first time, the pump has to be
calibrated which cannot be done by the OPUS software.
W h a t k i n d o f p u m ps a r e s u p p o r t e d b y t h i s
OPUS command?
• AIM3000
• ISMATEC REGLO Digital
• Metrohm 838
• Sapphire VFP17
P u m p Vo l u m e
The deviation is due to the fact that the pump control rounds off the
volume to complete cam steps. For further details refer to the docu-
mentation provided by the pump manufacturer.
Operation Mode
7.17.1 R e f e r e n c e m e a s u r e m e n t o f ATR
c r y s ta l
1 Select the Cleanness Test Setup command. The following
dialog opens:
7.18.2 Te s t r e s u l t
Manipulating spectra...
T h e m a n i p u l a t i o n r e s u l ts . . .
If you exit OPUS without saving these work files, the data gener-
ated during manipulation get lost. Normally, a warning pops up
when you want to exit OPUS indicating the existence of data which
have not yet been saved.
8.1.1 Sta n d a r d B a s e l i n e C o r r e c t i o n
1 On the File menu, click Load File. Load the spectrum file.
2 On the Manipulate menu, click Baseline Correction. The fol-
lowing dialog is displayed:
8.1.2 In t e r a c t i v e B a s e l i n e C o r r e c t i o n
S e t t i n g B a s e l i n e P o i n ts
D e l e t i n g B a s e l i n e P o i n ts
M o v i n g B a s e l i n e P o i n ts
Z o o m i n g I n / O u t Sp e c t r a
You can also zoom in or out spectra. Right click the original spec-
trum and select the Zoom command from the pop-up menu.
Scattering Correction
In case of scattering correction the points are set such that the
baseline shows monotonic decreasing and as small as possible
negative increasing. This, of course, requires the baseline to touch
the corresponding curve minima.
Rubberband Correction
trum from the bottom up (in case of absorption spectra) with vary-
ing intensity.
In figure 166 the concave baseline has been calculated, using 40,
15 and 8 iterations (top down). The result spectra are not displayed
to avoid confusion.
To s e t u p a c o r r e c t i o n m e t h o d . . .
Figure 169 shows the original spectrum together with the result
spectrum after baseline correction has been performed. In this
case the rubberband correction method has been used.
I n t e r a c t i v e Spe c t r u m S u b t r a c t i o n
Click the Start Interactive Mode button (see figure 170). The follow-
ing view opens:
H
C
I
F
T h e s u b t r a c t i o n d a ta b l o c k c o n ta i n s . . .
The first column on the right shows the factors used, and the mark
in the Modified column indicates the spectrum which the subtrac-
tion data block has been attached to.
AB = – log TR
TR = 10 – AB
In figure 177 the range between 2,550 and 2,360cm-1 has been
replaced by a straight line. The dotted line shown will not be avail-
able any more in the manipulated spectrum.
General
To c a l c u l a t e s p e c t r a . . .
I
F
J
Sine
Hyperbolic sine
Cosine
Hyperbolic cosine
Tangent
Hyperbolic tangent
Trigonometric number
Square root
Exponent (general)
Exponential function
Wavenumber value
D C clears the display in (H) and (I). You can use the brackets
to form mathematical groups that will be calculated first.
E The four basic arithmetic operations can be used.
F Numerical buttons consist of the figures 0 - 9, the decimal
point and exponential function E (e.g. 4.321E3 = 4321).
G This button starts the calculation. The result will be displayed
in the entry field (H).
H The loaded spectrum file to be calculated is available in this
entry field. You can also enter algebraic terms and numerical
values into this entry field.
I Select the result data block from the drop-down list.
J The display field shows the calculation results, or indicates
ERROR in case of an incorrect entry.
Miscellaneous
dent terms. This can be done using the function, which rep-
resents the frequency values of the spectral data points. The term
8.6 C ut
You can also use the Cut command to invert the frequency limits of
spectra.
1 Select upper and lower limits which are beyond the spectral
range and define them as new frequency limits in reverse
order.
2 Right click the spectrum and select the Properties command
from the pop-up menu.
3 Invert the axes settings using the Axes tab. This change may
not immediately appear in the spectrum window.
8.7 N or malization
Normalization methods
• Min/Max normalization
This method is used with spectra which optical layer
thickness substantially varies, to be able to compare
these spectra with each other. For example, the spectra
of a powder sample which have been compacted
unequally during separate measurements.
• Offset correction
This method shifts spectrum intensities to the effect that
the minimum absorbance unit will be 0.
Example:
8.9 C onvert Sp ec t r a
Conversion Method
E
A
F
B
G
C
H
D
I
8.10 S m ooth
Sm oothi n g.. .
To s m o o t h a s p e c t r u m . . .
Interactive smoothing
A C
B
D
8.11 Deriv a t i v e
To p e r f o r m a d e r i v a t i v e . . .
...a data block indicating the actual derivative order of the original
spectrum. If a spectrum of 1st order is subject to a 1st order deriva-
tion again, the result will be a 2nd order spectrum. The kind of deriv-
ative order is displayed in the data block of the resulting spectrum.
A ft e r c a l i b r a t i o n . . .
...you can have the parameter list displayed. Right click the name
of the spectrum file in the OPUS browser. Select the Show
Para-meters command from the pop-up menu.
A The frequencies of the first and last data point have been-
changed.
B The two new frequency values, factor M (Mult. for Freq. Calib)
and addend A (Add for Freq. Calib.) have been included in
the parameter list.
The two current frequencies are stored in the spectrum data block
after calibration. This allows to undo the calibration.
3 Click Calibrate.
4 In the parameter list the frequencies of the first and second
data point need to display the original calibration values. The
two parameters Mult. for Freq. Calib. and Add for Freq.
Calib. need to have the default values 1 and 0.
The two parameters factor M and addend A (see figure 193) have
to be entered as current frequencies.
M A
old ,1 = old M 1 + A 1
and later:
new ,2 = new ,1 M 2 + A 2
For the first run, select the M1 and A1 parameters, and for the sec-
ond the M2 and A2 parameters. The parameter list will then include
the Mtotal and Atotal parameters, which represent the total change
compared to the original calibration.
new ,1 = old M 1 + A 1
new ,2 = new ,1 M 2 + A 2
M total = M 1 M 2
A total = A 1 M 2 + A 2
8.13 R am a n Correction
A ft e r c o r r e c t i o n . . .
3 Click Correct.
4 The correction parameters need to display the original cor-
rection values.
C o r r e c t i o n M e t h o d ta b
A C
4
Rayleigh's law relates the scattering as a function of the wave-
number. This effect increases with the spectral distance of the line
of interest to the wavenumber of the Raman laser. In case of scat-
ter correction the Raman spectrum is multiplied point by point
using: Laser 4
8.14 B l a c k Body
A ft e r c a l c u l a t i n g t h e e m i s s i o n s p e c t r u m . . .
With:
• T = Temperature [K]
• c = 2.99792458 x 108 [ms-1]
• h = 6.626176 x 10-34 [m2kgs-1]
• k = 1.380662 x 10-23 [m2kgs-2K-1]
• = Wavenumber [cm-1]
• Apodization
• Phase computation
• Zerofilling
• Fourier Transformation of the interferogram
• Phase correction
Apodization
On this tab you select the apodization function and zerofilling fac-
tor.
L i m i t D a ta
On this tab you select the resolution and define which direction of
mirror travel has to be considered during conversion.
A If you activate the check box, you can vary the resolution
entering a value which is greater than or at least equal to the
value used during measurement. In this case only a fraction
of the measured interferogram is used for computation.
B In this group field you specify how precise the phase has to
be determined. Generally, you should enter the same value
as being used for the measurement. The value is limited by
the length of the double-sided part of interferogram.
C In case of interferograms recorded in forward/backward mode
you can specify which direction(s) of mirror travel have to be
evaluated during transformation. If both mirror travels should
be evaluated, the forward and backward scans will be trans-
formed separately, phase corrected and averaged.
D If data have been recorded in multiplex mode, the interfero-
gram contains alternating data from two analog-to-digital con-
verters (ADCs). The Even and Odd option allow the data from
Phase Correction
On this tab you define the phase correction mode. The phase cor-
rection can be compared with an interferogram symmetrizing,
which is always necessary due to the asymmetry of any measured
interferogram.
Non-Linearity
Peak Search
This tab allows to select the position of the zero path difference
(ZPD).
A This drop-down list is used to select the peak search mode for
ZPD. The single modes are described in the following.
• Absolute largest value searches the peak with the
highest absolute intensity.
• Maximum searches the highest peak with the largest
positive value.
• Minimum searches the lowest peak, if known that the
peak has a negative intensity.
• Mid between Min./Max. calculates a value between
the minimum and maximum limit.
• No peak search uses the position saved in the inter-
ferogram. If this value is known, it can be entered
manually.
• Mid between largest two searches the peak between
the two largest values.
• Take from stored phase uses the value calculated for
the phase stored.
St o r e
This tab allows to select the frequency range for the spectrum file
and define which kind of spectrum has to be additionally stored.
8.16 Inver se FT
Example 1:
Example 2:
In both cases the digital resolution has been 8 cm-1. To avoid arti-
facts allow at least 50 more data points on each side of the desired
interval. This frequency range must also contain meaningful spec-
tral information.
If you click the Zerofill button (figure 207), the spectra file(s)
selected will be interpolated using the additional zerofilling factor
specified.
before zerofilling
Interactive Mode
Adjust Parameter
If you do not work in the interactive mode, you can also make the
settings required by using this tab.
8.18.1 F o u r i e r S e l f - D e c o n v o l u t i o n T h e-
ory
If the shape and width of the LBF are known, its effect can be arith-
metically excluded from the spectrum. It is recommended to per-
form this arithmetic exclusion on the basis of the interferogram as
in this case the deconvolution is equivalent to a simple interfero-
gram division caused by the LBF Fourier transformation.
Deconvolution
Signal-to-Noise Ratio
Noise Reduction
The noise reduction factor should range from 0.0 to 1.0. It is the
fraction of the interferogram, which apodization function has been
decayed to 0 and thus used for evaluation.
Example
Deconvoluted spectrum
Original spectrum
8.20 Kramers-Kronig-Transforma-
tion
C o n d i t i o n s r e q u i re d f o r a s u c c e s s f u l
K r a m e r s - K r o n i g - Tr a n s f o r m a t i o n
A B
A If you activate one of these two option buttons, the result file
includes two data blocks: the real and imaginary part of the
functions indicated by R or I on the data block ( ).
B If an absorbance spectrum is calculated, it only includes pure
absorption contributions, but no reflection losses. To be able
to compare the absorbance spectrum directly with a transmis-
sion spectrum, you have to consider the reflection losses at
both boundaries of the sample layer by means of the correc-
tion -2 lg (1 - [Input - Reflectance spectrum]), using the OPUS
spectrum calculator.
8.21 S p l i t I n t erferograms
Basic Features
If you have selected the Double Sided acquisition mode, two inter-
ferograms will be generated during splitting. In case of a dou-
ble-sided Forward-Backward acquisition mode, four interferograms
will be generated during splitting. This yields to the maximum pos-
sible time resolution of single spectra at a given mirror velocity.
8.23 E x t rapolation
8.23.1 E x t r a p o l a t i o n t o 0 c m -1
a2
R s – 1 + a 4 s – ----4- s
2-
a 5 = --------------------------------------------------
s 1 5
1 = R 0 + a1 2 + a2 4
R s = R 0 + a1 s 2 + a2 s 4
R s – R 0 – a1 s2
a 2 = ---------------------------------------------
s4
The remaining function is:
4
1 = R 0 + a 1 2 + R s – R 0 – a 1 s 2 ---
s
Now, specify a1 in a way that the sum of all error squares within the
overlapping area can be minimized.
P h y s i c a l E f f e c ts
1. All correction methods are based on the Advanced ATR Transformation by Koi-
chi Nishikida, N & K Spectroscopy, LLC, an Honorary Associate/Fellow, Materi-
als Science and Engineering, University of Wisconsin
Select Files
A
B
Scaling
400 25.00
1.19
420 23.81
1.08
440 22.73
0.99
460 21.74
0.91
480 20.83
0.83
500 20.00
Interpolation
The minimum number of data points for the micron spectrum cre-
ated is equivalent to the number of data points found in the cm-1
spectrum. The MCF defines the quality of the conversion properties
within the most inappropriate range, i. e. in the high frequency end
of the spectrum (wavenumbers) which is converted to the low
wavelength end (microns).
Example:
The MCF can vary between 0.5 and 50. However, the new spec-
trum has never fewer data points than the original spectrum.
Therefore, it may occur that the number of data points, e.g. starting
from a factor of 20, remain unchanged.
In case of a small MCF the number of data points of the new spec-
trum may become quite large. This depends on the upper and
lower frequency limits as well as on the distance of data points in
the original spectrum. By reconverting the new spectrum and com-
paring it with the original one it can be verified whether the factor
selected has been reasonable.
Example:
The upper left spectrum in figure 225 shows the original spec-
trum scaled in wavenumbers. On the right the spectrum has
been converted to wavelengths, with the intensities remaining
unchanged. In the lower left spectrum the intensities have
been converted without changing the integrals.
D If you activate this check box, spectra files are weighted pro-
portionally to their number of scans during measurement.
E Activating this checkbox calculates the standard deviation of
single-spectra intensities from the average spectrum intensi-
ties. An additional entry field will be displayed (only if C has
been activated before) next to this check box function. If you
leave this entry field empty, a new file will be generated which
contains the standard deviation spectrum. If you drag & drop
a spectrum file from the browser into this entry field, this spec-
trum file will be updated.
F If you activate this checkbox, an identity comparison will be
performed between the original set of spectra and the aver-
age spectrum. This will result in an average report which
includes information on the single spectra used to create the
average spectrum.
The report also outlines how much each single spectrum
deviates from the average spectrum, and which frequency
ranges and method have been selected for averaging.
The single-spectra deviation is indicated in multiples of the
standard deviation to make the detection of outliers easier.
The Report Method button is enabled as soon as you have
activated the Compute Av. report check box. Select the
respective kind of method used to calculate the spectral dis-
tance by clicking this button.
8.26.1 Av e r a g i n g T h e o r y
n
2
yi – y
i=1
= ------------------------------------
n–1
Principle • Load the spectrum files. Note: You can only use
one file type, i. e. either only absorption (AB) or
transmission (TR) spectrum files. Each spec-
trum file must contain a reference data block for
the single-channel spectrum ( ).
• Mark the result data block (e.g. or )
of the spectrum files in the browser.
• Select the Merge Spectral Ranges command.
• Click Merge to start this manipulation command.
The usable regions on the edges of spectra files, i.e. the overlap-
ping ranges of spectra files can be set interactively. By default, the
signal intensity is predefined by at least 10 % from the maximum of
the reference spectrum. This value should only be changed if nec-
essary.
Note: Once spectra have been merged, the latest range set-
tings made for the overlapping range, either interac-
tively or manually, remain valid. Clicking the Clear list
button (figure 229) allows to delete the range settings.
A
B
A ft e r c o m p e n s a t i o n . . .
General
• Wavenumber ranges:
- for H2O compensation (Standard, MIR): from 3600 to
4000 cm-1
- or H2O compensation (Aqueous solution, MIR): from
1900 to 1800 cm-1
- for H2O compensation (NIR): from 5000 to 5600 cm-1
- for CO2 compensation: from 2300 to 2400 cm-1
Sample absorption or a low signal to noise level within
these wavenumber ranges can affect the compensation
quality.
• The sample and reference spectrum have to be
acquired using the same spectrometer resolution. Differ-
ent zero-filling factors are possible.
• The spectrum should not end at a wavenumber which
shows strong absorption of atmospheric gases. In case
of a H2O compensation (see figure 233), the lower wav-
enumber limit must not be around 1700 cm-1.
• For compensation in aqueous solutions the spectrum
has to include wavenumber ranges of 2300 - 2400 cm-1
and 1800 - 1900 cm-1 for both CO2 and H2O. These
wavenumber ranges must be free of sample absorption
bands.
A t m o s p h e r i c c o m p e n s a t i o n u s i n g s i n-
gle-channel spectra from one file
In this case the two single-channel spectra data blocks are stored
in one spectrum file.
Example:
Select the spectrum file and drag & drop it into both selection fields
on the Select Files tab.
If you have selected several spectrum files, a data block will be dis-
played in both selection fields for each spectrum file selected.
A t m o s p h e r i c c o m p e n s a t i o n u s i n g s i n-
gle-channel spectra from different files
Example:
tion field.
6 Click Calculate.
Principle • Load the spectrum file (only data blocks with ratio
spectra, e.g. absorbance, transmittance,
Kubelka-Munk, reflectance or log-reflectance
files).
• Select the Straylight Correction command.
• If required, drag & drop the spectrum file into the
Spectra to apply straylight correction selection field.
• Define the type of window material.
• Click the Record new spectra tab to create new cor-
rection spectra (chapter 8.29.1).
• Click Calculate to start this manipulation command.
Sta n d a r d m a t e r i a l r e q u i r e d
To c r e a t e c o r r e c t i o n s p e c t r a . . .
The following ratio spectra are calculated from these three different
single-channel spectra:
• ExtRef: SCextBG/SCintBG
• Stray: SCStray/SCintBG
The ExtRef and Stray ratio spectra are used to perform straylight
correction on the spectra measured.
TM – Stray
TM corrected = ------------------------------------
Extref – Stray
TM – Stray Extref
TM corrected = -----------------------------------------------
1 – Stray Extref
1+c - 2
------------------------ where c = R 1 + 1 – R
1 + TM c
8.29.3 St r a y l i g h t c o r r e c t i o n a s pa r t o f
the measurement process
7 Click Calculate.
C A
B
8.33.1 In t r o d u c t i o n
Absolute emission
Measurement conditions
B e a m pa t h
The beam path for the sample and the reference source must be
identical. The emitting surfaces (of both sources) have to be larger
than the spot size received by the spectrometer at the emission
input, and have to be positioned in or imaged to the focal plane of
the spectrometer input.
Reference source
8.33.2 T h e o r e t i c a l b a c k g r o u n d i n f o r m a-
tion
[1] S T = m L T + b
S c T c
[5] b = ---------------------
- – B T c + 1 – B T amb
m
with
B T h – B T c
[8] Slope = ---------------------------------------------------------------
-
S h T h – S c T c
Offset T c = B T c + 1 – B T amb
[9] = 1 – r –t
Then:
[10] L T u = u B T u + 1 – u B T amb
8.33.3 Sta n d a r d p r o c e d u r e o f r a d i o m e t-
ric calibration
1 Measuring the cold reference source.
➣The temperature of the cold reference source should be
lower than the range of expected sample temperatures, e.g.
at room temperature.
2 Measuring the hot reference source.
➣The temperature of the hot reference source should be
close to the highest expected sample temperature, but not
higher. Thus, excessive detector non-linearity is avoided.
3 Measuring the emission sample.
➣Emission sample measurements may be repeated under all
desired sample conditions (temperatures).
4 Generating a radiometric calibration file (chapter 8.33.5),
using the files from (1) and (2).
5 Performing radiometric calibration (chapter 8.33.6) of the
files from (3), using the calibration file generated in (4).
➣The calibration file remains valid until the temperature of the
spectrometer or ambient change substantially.
8.33.4 C u s t o m i z i n g M a n i p u l a t e menu
Note: The two commands available for radiometric calibra-
tion in OPUS are integrated in the Manipulate menu.
As the commands may not be seen in the default
menu configuration, you have to customize the
Manipulate menu.
1 On the File menu, select the Load command and load the
spectrum file of the cold and hot black body into the browser.
2 On the Manipulate menu, select the Build Radiometric Cali-
bration File command.
3 On the dialog that opens, make sure that the interferogram
(or spectrum) file for the cold and hot black body is loaded
into the particular list box.
➣If the file contains a ’SIFG’ data block, this data block is
preferably loaded by default. Details on the further setting
options are described in chapter 8.33.5.1.
4 Enter the temperature for the cold and hot black body.
8.33.5.1 D e ta i l s o n t h e s e t t i n g o p t i o n s
1
3
2 4
10
11
Komponente Definition
1 Interfero- The entry field must contain the interferogram (or spec-
gram/spectrum trum) file of the cold black body.
file - cold
2 Interfero- The entry field must contain the interferogram (or spec-
gram/spectrum trum) file of the hot black body.
file - hot
3 Temperature (°C) The temperature of the cold black body has to be entered.
The temperature is defined in C°.
4 Temperature (°C) The temperature of the hot black body has to be entered.
The temperature is defined in C°.
7 Use the following If the option button is activated, the non-linearity settings
non-linearity set- have to be defined as follows:
tings • Detector cutoff: the detector cutoff describes the spectral
limit (in cm-1), below which the detector is not sensitive
anymore. If you use a preamplifier which reverses the sig-
nal polarity, the value of the detection limit has to be multi-
plied by -1.
• Modulation efficiency: the modulation efficiency causes
an additional multiplicative correction of the entire spec-
trum. It can be set either to the correct value (typically
between 0.7 and 1.0), or to 1.0 if the correct value is
unknown.
More details on non-linearity are described in chapter 8.15
et seq.
8 NL correction dis- Activating the option button makes sense if you use a linear
abled detector type, e.g. DTGS.
Komponente Definition
9 Use constant The option button should only be activated if the reference
emissivity source has a constant spectral emissivity of <1. In case of a
cavity black body source the typical emissivity is 0.99, in
case of an extended-area black body source the emissivity
is mostly <0.99.
• If activated, enter the value into the entry field next to the
option button.
11 File name Enter a unique file name, perhaps referring to the measur-
ing setup.
8.33.6 P e r f o r m i n g r a d i o m e t r i c c a l i b r a-
tion
Note: It is highly recommended to use emission sample
spectrum files which contain both the interferogram
and the single-channel spectrum. The interferogram
allows an optimal phase correction, whereas the FFT
parameters, e.g. apodization and non-linearity correc-
tion can be pre-defined by the spectrum used. The
measurement parameters used to generate the emis-
sion sample spectrum file, must not differ from the
measurement parameters used to generate the black
body spectrum file, except for the number of scans.
1 On the File menu, select the Load command and load the
spectrum file of the emission sample.
2 On the Manipulate menu, select the Radiometric Calibration
command.
3 On the dialog that opens, make sure that the data block for
the interferogram (SIFG) is loaded (figure 245).
➣If the spectrum file does not contain a SIFG data block, use
the data block for the single-channel spectrum (SSC).
➣If the spectrum file does contain a SIFG and SSC block, the
operator determines whether interferograms or spectra are
used to perform radiometric calibration by selecting either
one of them. Performing radiometric calibration based on
interferograms has some advantages but also some limita-
tions, as outlined at the end of chapter 8.33.2. Radiometric
calibration based on spectra is always possible.
9.1 C ur ve F i t
B e f o r e y o u s ta r t a c u r v e f i t . . .
changes to .
This kind of cursor allows to move the curve. Click the left
mouse button and drag the curve to the position desired
while pressing the mouse button.
5 Position the cursor slightly below the top of the curve. The
cursor changes to .
The peak parameters can be defined in the lower end of the curve
fit view. Position, intensity and width can be specified more pre-
cisely than by using the cursor.
1 ...click into one of the entry fields and enter the value
manually, or use the arrow buttons above the respective
entry field.
3 Each peak is shown in one line of the peak list including all
peak parameters. The lines with non-selected peaks are
marked green and the currently selected peak is marked vio-
let. The peaks in the list are always sorted by peak position.
Defining Method
D e f i n i n g I t e r a t i o n Tim e
A ft e r c a l c u l a t i o n . . .
3 peak
spectra
Note: Once a suitable model has been found (see next para-
graphs), it can be applied to other spectra of the same
type, there is no need to set up a new model for each
spectrum.
9.1.3.2 A s d a ta b l o c k a d d e d t o t h e
original spectrum
1 Click the Save Peaks and Reps button.
2 The applied model is stored as report data block ( ),
added to the original spectrum displayed in the OPUS
browser.
To d i s p l a y t h e pa r a m e t e r s o f p e a k c a l c u l a -
tion...
L e v e n b e r g - M a r q u a r dt
B e f o r e y o u c a n s ta r t i n t e g r a t i o n . . .
A
D
G If you click this button, all areas including the relevant infor-
mation will be deleted. Note: Before you set up a new
method click this button to delete possible default set-
tings caused by the previously loaded integration
method.
H Click this button to load or edit an existing method.
I To save all the settings made click this button and define an
appropriate name for the currently created method.
Integration Modes
Looks for the highest peak between the left and right
edge, and for the minima on the left and right of this
highest peak. The baseline connects these minima.
changes to .
2 Press the left mouse button and reduce or enlarge the area
to the size desired.
9.2.3 Sta r t i n g I n t e g r a t i o n
1 Load the spectrum file.
2 Select the Integration command. A dialog opens which con-
tains the name of the spectrum file in the File(s) to integrate
selection field.
To d i s p l a y t h e i n t e g r a t i o n r e s u l t i n a
report view
St o r i n g R e p o r t
Principle A wizard will guide you step by step through the setup
of a QUANT method. Follow the on-screen instruc-
tions.
A Q U A N T m e t h o d c o n s i s ts o f :
is fitted to the data obtained from the calibration samples, and this
function is then used to determine the concentrations of the
unknown sample.
To s e t u p a Q U A N T m e t h o d . . .
The first dialog gives a short overview of the steps required to set
up a method. You can load and edit an existing method. Click Next
to continue with the next dialog.
Specify the component names and the appropriate unit. Click the
Add Component button.
Click again the Add Component button until the desired number of
components has been reached.
The terms for the component and unit can be changed. Select the
respective component and enter the definition into the appropriate
entry field. Click the Set button to update the list.
To delete a component from the list, select it and press the Del key
on your keyboard.
Click Next.
Based on the type of unit specified enter the value of each compo-
nent for each sample of the calibration set.
To change the path of the spectra directory click the Change Path
button. If you want to remove a spectrum from the calibration set,
select the spectrum and click the Remove Selected Spectra button.
Now, you have to define one isolated peak for each single compo-
nent. Either the peak height or integral is used for calibration. Click
the Set Integration Area button. The following dialog is displayed:
Define the left and right edge of the integration area by either enter-
ing the values in the respective entry field or clicking the Interactive
button. In case of some integration modes you have to make fur-
ther settings. For information on integration modes refer to
chapter 9.2.1.
If you activate the Use reference peak button, make sure that you
also select the Reference Peak option from the drop-down list.
If all settings have been made, click OK to get back to the wizard.
In the next dialog you have to select the type of calibration curve.
If you click Calibrate, you will be asked to store the method. After-
wards the calibration curve is calculated for each component and
displayed accordingly.
Below the graphics, the method file, correlation coefficient and cali-
bration equation are indicated.
Apart from the Fit vs True (fit vs the known values) depiction type
you can choose between Difference vs. True (difference vs known
values) and True vs. X (known values vs peak intensity).
You can print the results, figures as well as data tables by clicking
Print. The calibration results, graphics and tables as well as the cal-
ibration equation are stored together with the method file.
Click Next to have the calibration report displayed for each compo-
nent separately.
A table shows the known value (true) of the sample and the value
calculated by the fit. It also shows the difference between these
values, the error of the calculated value and the peak size X calcu-
lated from the spectra.
You can have the results displayed in a report view or table. Acti-
vate or deactivate the Calibration report check box.
1 2
St o r i n g Q U A N T m e t h o d
3 Enter the desired method file name and define the path.
QUANT methods have the extension *.q1.
4 If the settings have been made, click Save.
B
Figure 287: QUANT Builder - Change method
T h e r e s u l t o f a q u a n t i ta t i v e a n a l y s i s . . .
To d i s p l a y t h e q u a n t i ta t i v e a n a l y s i s r e s u l t
i n a r e p o r t w i n d o w. . .
1. Sigma is the standard deviation of the concentration value calculated. The sigma
value is determined by the quality of the QUANT method set up. Ideally, the sig-
ma value should be as small as possible.
You can apply more than one QUANT method to analyze a spec-
trum. In this case the results are stored in one single QUANT
report.
On the left side of the report view there will be several Quantitative
Analysis 1 Report folders. Clicking the respective folder shows the
corresponding analysis parameters on the right side.
The difference between the true component values (True) and the
predicted values can be indicated by . Note that the true values
are also prone to statistical errors, depending on how carefully they
have been determined. The correlation coefficient r is an indicator
for the quality of calibration:
Zi – Zm Yi – Ym
r = --------------------------------------------------------------------
Zi – Zm 2 Yi – Ym 2
9.6 S i gnal-to-Noise-Ratio
On the Frequency Range tab you can select between different fre-
quency ranges which have been pre-defined.
You can also interactively select the frequency range or use the
current file limits (see also chapter 2.9). If you click Calculate, the
signal-to-noise ratio will be determined.
To s t o r e t h e r e s u l t . . .
To s h o w t h e r e s u l t i n a pa r a m e t e r l i s t
Signal
• SN (RMS) = -----------------
RMS
Signal
• SN (PP) = ---------------------------
PP Noise
9.7 P e a k Picking
Select Files
Frequency Range
Y- L im its
Mode
This tab allows to define the direction of peaks and the peak pick-
ing method. In case of absorbance spectra usually the maxima are
of interest, whereas for transmittance or reflectance spectra the
minima are the important features.
To d i s p l a y t h e p e a k p i c k i n g r e s u l t i n a
peak list
9.7.1 In t e r a c t i v e P e a k P i c k i n g
1 Load the spectrum file.
2 Select the Peak Picking command. The Peak Picking dialog
(see figure 296) is displayed.
3 Click the Start Interactive Mode button. The following view is
displayed:
Wor k i n g i n i n t e r a c t i v e m o d e
Moving Line
Left click the line point and move the line upwards or down-
wards.
To remove the line from the view, just double click the line point.
Drag the line point to the position desired while pressing the left
mouse button.
St o r i n g I n t e r a c t i v e P e a k P i c k
Click Store (see figure 300). The interactive mode view closes and
the spectrum window is displayed again.
Sta n d a r d M e t h o d
Half width
Relative
intensity
Local baseline
Figure 301: Spectrum with line width, relative intensity and baseline indicated
Peaks will be included into the peak pick table if their relative inten-
sity is within the interval defined on the Y-Limits tab. This interval
has been normalized such that the difference between maxima and
minima corresponds to 100%.
The x-positions of the bands are the x-values of the minima (or
maxima) second derivative spectrum which has been smoothed.
This is particularly useful for seriously overlapping bands because
in this case an extremity of the spectrum is frequently not shifted
with respect to the true band position. Weak shoulders can also be
recognized using the second derivative method.
In order to calculate the relative intensities and the half width, the
two zero-points and the minima of the second derivative are
searched.
• [ –1 ,0 ] (zero-point)
• [ 0 ,I 0 ] (minimum)
• [ –1 ,0 ] (zero-point)
These points are typical for the original peaks, i.e. before being
overlapped by other lines as the following examples show
(figure 304) where two Lorentzian curves are superimposed. The
zero-points and the minima of both signals are almost identical
(both for the derivatives of the isolated signals as well as for their
superposition). The maxima of the overlapped signals, however,
are shifted and therefore misleading.
454 OPUS Reference Manual Bruker Optik GmbH
Peak Picking Chapt. 9
If the fraction of peak height is too small, the band will not be listed
on the band list. Too small a fraction of peak height could decrease
the number of measuring points, used to calculate the band posi-
tion, to less than 3.
Based on the band maximum Y max X max , which is usually not the
maximum of the measuring values, the measuring points are con-
secutively numbered to the left and right.
The numbering starts from 0 for both sides, while the y- value of
each measuring point must be greater than
Y i Y max – Y max – Y min Fraction , with Y i being the measurement values,
and the Fraction term being the percentage value defined for the
peak height. This yields to the following sums:
with x max being the next smaller integer of the x-value for the
band maximum.
PeakPos COG v = PeakPos COG v + v min and v being the wavenumber dis-
tance between two measuring points and vmin being the wavenum-
ber of the first data point.
If you activate the Use file limits check box the complete spectra
are compared with each other. For further details on defining fre-
quency ranges refer to chapter 2.9.
To d i s p l a y t h e q u i c k i d e n t i t y r e s u l t i n a
report view
The closer the hit quality is to 0, the better the spectra matches
within the frequency ranges used. In case of the quick identity test
the following report lines are important:
• Method file: shows the name of the method used for
the test
• Expected Reference: shows the name of the principal
file used
• Vector normalizes spectra: indicates whether a vector
normalization has been used
• X-Ranges: number of frequency ranges
• Order of Internal Derivation: defines the order of the
derivative used
• Smoothing Points for Internal Derivation: defines the
number of smoothing points
Q I D N T. 0 F i l e
9.9 Qu al i t y Te st
B e f o r e y o u c a n s ta r t a q u a l i t y t e s t . . .
To d i s p l a y t h e q u a l i t y t e s t r e s u l t i n a
report view
The first line of the report indicates the test result. A value of 1
means test passed, whereas 0 means test failed. For further analy-
sis do not use spectra which have not passed the test. The second
line includes the name and path of the test method used.
The lower end of the report shows all parameters and their limit val-
ues (minimum and maximum) as well as their measured values.
These parameters set can be edited on the Parameters tab of the
Quality Test dialog.
To c h a n g e t h e pa r a m e t e r s e t t i n g s . . .
...click into the respective entry field and enter the desired value
manually.
To r e s e t t h e pa r a m e t e r s . . .
Range Definition
Range Definition
Range Definition
General Features
Based on the results obtained by this IDENT analysis, you can con-
tinue with a quantitative (QUANT)1 analysis using the same sam-
ple.
1. You can also perform a QUANT evaluation by using PLS models which have
been created by Unscrambler 9.8. The following preconditions must be met:
- Install the OLUP 9.8 version (this is the only version possible) produced by
Camo into the respective program directory (e.g. C:\Program
Files\Bruker\OPUS_<version>).
- Copy the OLUP.SYS file from the OLUP 9.8 installation directory into the OPUS
root directory.
- To be able to use the OLUP prediction engine in OPUS, you MUST have an
OPUS registry which contains the QUANT or QUANTRUNTIME packages.
- Copy ALL files (e.g. *.42m etc.), which belong to an Unscrambler model, into
the *\OPUS_<version>\ME_Base\Quant directory.
You can use this PLS model for the OPUS Multi Evaluation Setup by adding a
QUANT 2 method.
M E M e t ho d
1. Only possible with QUANT 2 methods (chapter Adding next step QUANT 2)
If only one substance group applies to the product, you can con-
tinue multi evaluation by performing a conformity test and/or
QUANT evaluation.
9.10.1 F i r s t s t e ps
1 Generate evaluation methods for qualitative and quantitative
analysis (IDENT and QUANT methods). Optionally,
generate methods for integration and conformity test as well.
2 Store the evaluation methods generated in the correct
OPUS directory available for multi evaluation:
- *\OPUS\ME_BASE\Ident or
- *\OPUS\ME_BASE\Quant
9.10.2 S e t t i n g u p a n M E me t h o d
You can add methods to each group and define the method param-
eters separately.
Step 1:
Click the appropriate table cell.
Step 2:
Select the existing group name.
Step 3:
Overwrite the existing group name.
Confirm the setting made by the Return
key.
Step 1:
• Right click the appropriate group
name.
• Select Set Expected Reference option
from the pop-up menu.
Step 1:
• Right click the appropriate group
name.
• Click the Set Expected Reference
option from the pop-up menu.
A d d i n g a n o t h e r e v a l ua t i o n m e t h o d t o a
group
You can add one or several methods to the single groups belong-
ing to a method. The procedure on how to add evaluation methods
is the same for all types of methods
Step 1:
• Right click the appropriate group
name.
• Select the Add Quant 2 Analysis
option form the pop-up menu.
Step 2:
Click the Add Method button.
Step 3:
• Select the method from the dialog that
opens.
Step 4:
Click the Apply button.
Setting formulae
Step 1:
On the Calculation table cell, click the
Set button.
Step 2:
Select the type of formula.
Step 3:
Combine e.g. the method component
with a numerical value. Double click the
component.
Step 4:
Use the numerical pad to enter a
numerical value.
Step 5:
• Optionally, enter a name for the for-
mula. By default, an asterisk is set as
formula name.
• Confirm the settings by clicking the
OK button.
S e t t i n g v a r i a b l es
Step 1:
On the Calculation table cell, click the
Set button.
Step 2:
Select the type of formula, e.g. General
Formula.
Step 3:
Combine e.g. a variable with the predic-
tive value. Double click the variable.
Step 4:
• Optionally, enter a name for the vari-
able.
• Confirm the settings by clicking the
OK button. The name of the variable
defined in the calculator remains the
same.
Step 1:
Click the name of the first method com-
ponent.
Step 2:
Click the Remove Method button.
Step 3:
On the dialog that opens, click the
method.
Step 4:
Click the Remove Method button.
When replacing one method by another make sure that both meth-
ods contain the same components. Otherwise, replacing the
method is not possible.
Step 1:
Click the name of the first method com-
ponent within a group.
Step 2:
Click the Replace Method button.
Step 3:
On the Available Methods list box, click
the method you want to use to replace
the previous one.
Step 4:
On the Selected Methods list box, click
the method you want to replace.
Step 1:
• Right click the real component of the
appropriate group.
• Click the Add Calculated Component
option from the pop-up menu.
Step 2:
• Enter an appropriate component
name.
• Enter reasonable values into the entry
fields.
• Click the Calculation table cell and
enter an appropriate formula.
Step 1:
On the browser view, click the name of
the component added.
Step 2:
Click the Remove Component button.
Step 3:
• On the dialog that opens, select the
component.
Example 1:
When using the Add Next Step Quant 2 option the table of contents
of the data block should contain only the result of the last
QUANT step. This requires all component names in all QUANT
methods (*.q2) used to be identical, upper and lower case must be
observed.
• The QUANT method (A) used in the first step contains
the ethanol component (apart from methanol and propa-
nol) within the concentration range of 0 to 100%.
• The QUANT method (B) used in the next step contains
the ethanol component within the concentration range of
5.9 to 13.5%, which can improve the accuracy of the
prediction value.
Example 2:
When using the Add Next Step Quant 2 option the table of contents
of the data block should contain the results of all components
included in the QUANT methods used.
• The QUANT method used (soybean protein) contain the
protein component within a large concentration range.
Calibration samples of many soybean types with differ-
ent composition form the basis for this concentration
range. The fat and ash components, however, cannot be
determined with sufficient reliability by this QUANT
method.
• Depending on the protein content found in the first step,
individual QUANT methods for fat and ash are now com-
bined in the next step.
Based on the Add Next Step Quant 2 option, the result achieved in
example 2 contains the prediction value for protein (from the first
step) and for fat and ash (from the last step).
Procedure:
Step 1:
• On the browser view, right click the
name of the component which is part
of the first QUANT 2 method.
Step 2:
• Depending on the calibration range
specified, enter a concentration range
which the second QUANT 2 analysis
is to apply for.
Step 3:
• Click the Add Method button.
Step 4:
• Check whether the values specified
for the concentration range corre-
spond to the method selected, and do
not overlap. Click the Check button.
Step 5:
• As soon as the values are correct you
can make further settings on the sin-
gle analysis parameters.
Step 1:
Click the name of the component dedi-
cated for a QUANT 2 analysis.
Step 2:
Click the Remove Method button.
Setting Output
The Set Output option allows to define which evaluation data are
displayed in the ME report (see also chapter 9.11).
Step 1:
• Right click the name of a method
component. By default, the Set Output
option is checked.
• Click the Set Output option if the eval-
uation data of this component are not
to be displayed in the ME report.
Step 1:
• Right click the name of the method
component whose evaluation data
had been hidden in the ME report.
• Click the Set Output option.
9.10.3 St o r i n g M E m e t h o d
1 Enter a name for the ME method into the lower entry field.
2 Click the Save button. Now, the name of the ME method has
to be shown on the browser view.
9 . 11 . 1 I n p u t Va r i a b l e s
To d i s p l a y t h e m u l t i e v a l u a t i o n r e s u l t i n a
r e p o r t v i e w. . .
The ME test report contains the method name and path as well as
the results of the different method tests, e.g. QUANT, IDENT, con-
formity test or integration.
D
Figure 324: Report overview
Component Definition
1
2
3
9.12.1 T h e o r e t i c a l b a c k g r o u n d i n f o r m a-
tion
Procedure
1 2
Component Definition
Component Definition
9.12.2.2 S e t t i n g o p t i o n s o f pa r a m e t e r s
1
2
Component Definition
Component Definition
9.12.2.3 D e ta i l s o n t h e v a l i d a t i o n o f r e f-
erence spectra
Note: Validating reference spectra checks the similarity
between the spectra to each other. Each reference
spectrum is compared with all other reference spec-
tra. This kind of spectrum comparison is based on
the correlation coefficient, which defines the correla-
tion1 between at least two spectra.
2
1 3
Component Definition
Component Definition
Component Definition
9.12.3 P e r f o r m i n g q u i c k c o m pa r e
1 On the File menu, select the Load File command and load
the spectrum file.
➣The data blocks of the query and reference files must have
the same spectrum type, i.e. AB (absorption) or TR (trans-
mission). The wavenumber range of the sample spectrum
has to be at least as large as the range defined for the ref-
erence files.
2 On the Evaluate menu, select the Quick Compare com-
9.12.3.1 S e t t i n g o p t i o n s o f q u i c k c o m-
pa r e
Component Definition
Component Definition
9.12.4 Q u i c k c o m pa r e r e s u l t
The following table contains the possible result display for the
Material verification and Material identification modes.
The structure of the result view is the same for both evaluation
modes. Figure 301 exemplifies the result view for the Material veri-
fication mode.
Component Definition
9.12.4.2 Vi e w i n g r e s u l t i n t h e r e p o r t
view
1 Right click the QC data block added to the spectrum file and
displayed in the browser.
2 From the pop-up menu that opens, select the Show Report
option.
The report contains the parameter values used and the result cal-
culated as well as the number of hits found.
Note: Templates for the analysis report have the file exten-
sion *.art. By default, the MIR.art template is part of
OPUS and can be used to create analysis reports for
quick compare.
1. Detailed information on the Setup Analysis Report Template command are de-
scribed in chapter 11.2.
2. Single sample report: one report for each single spectrum file; Multiple sample
report: one report for all spectrum files
6 On the dialog that opens, enter a file name for the template.
➣By default, the templates for the analysis report are stored
in the *\User\Public\Public Documents\Bruker\OPUS direc-
tory.
7 Click the Save button.
Page 2
Figure 334: Example of analysis report - Page 1
To h a v e t h e r e s u l t d i s p l a y e d
If the results are to be saved with the spectrum, click Yes. The
results are displayed in the parameter list.
To open this list right click the file name in the browser window and
select Show Parameters from the pop up menu.
9.14 H 18
Setting options
To d i s p l a y t h e H 1 8 e v a l u a t i o n r e s u l t i n a
report view
9.14.1 Theory
A:
1 4 ExtractVol a a
conc ------- = ----------------------------------------------------------- ----1- + ----2- DilFac
mg
l WaterVol CellThickn c 1 c 2
B:
1 3 ExtractVol a a a
conc ------- = ----------------------------------------------------------- -----1- + ----2- + ----3- DilFac
mg
l WaterVol CellThickn c1 c 2 c 3
Symbol Description
Symbol Description
In the H18 report the peak heights used and the associated wave-
numbers are listed according to the spectral data.
To p e r f o r m s p e c t r u m s e a r c h . . .
Sp e c t r u m S e a r c h
Component Definition
Show search report imme- The search report is shown immediately after the spec-
diately trum search has been completed.
Use search report for When activated, a list box is displayed below the check
search box. You can select an already existing search report
from this list box. In this case, searching is not performed
by means of the settings made on the Select Libraries
tab, but only by those spectra which are included in the
search report. This enables to refine a search query pre-
viously performed.
1 On the query spectrum, position the cursor onto the red line
(figure 342).
➣The cursor changes to .
2 While pressing the left mouse button, move the cursor to the
spectral range limit desired on the x-axis.
➣Alternatively, you can manually enter the limits into the
table.
Alternatively, you can also add a new range by entering the limits
into the table. Figure 343 exemplifies two spectral ranges added to
the query spectrum.
Search Parameters
Component Definition
Search algorithm:
Component Definition
Using existing peak table The Use existing peak table search algorithm uses an
already existing peak table.
A peak table is generated by the Peak Picking or Single
Peak Pick command. This command allows to limit spec-
trum search to selected bands only, and to ignore other
types of bands being part of the spectrum as well.
Normalization method
Types of derivative
Select Libraries
A search can involve either one or more libraries, but creates only
one search report in either case.
8 From the dialog that opens, select the name of the library
list.
➣Library lists have the file extension *.LLS.
9 Click the Save button.
➣The library list is overwritten by the new settings made.
On the Select Libraries tab, check the Status column. Libraries can
have the following status:
• Green checkmark: library is OK
• Red cross: library could not be found (e.g. library
deleted or library directory changed
• Red exclamation mark: library is not accessible (e.g.
registration not valid)
...is added to the query spectrum in the form of the data block
and displayed in a separate result view.
T h e s p e c t r u m s e a r c h r e s u l t v i e w. . .
If you have not activated this check box, double click the data
block in the browser to open the spectrum search result view.
Figure 346: Example of a search result view with spectrum and structural formula
Definition
E Query spectrum
10.2 Stac k e d
Definition If more than one spectrum has been loaded into the
same spectrum window, you can have them dis-
played in stacked mode. In this case the spectra will
not overlap.
To d i s p l a y s e pa r a t e a x e s f o r e a c h s p e c-
trum...
Figure 349 shows two spectra with separate axes for each spec-
trum.
changes to .
• Left click on the desired peak. The wavenumber of
the generated peak is displayed. In case of equally
large peaks only the peak close to the cursor will
be displayed.
• After a single peak pick the PEAK data block is gen-
erated and displayed in the browser.
• To delete single peak labels right click the wave-
number displayed for the particular peak and select
the Delete peak label command from the pop-up
menu.
cursor changes to .
• Stretch the curve up or down, by pressing the left
mouse button.
• To undo the action select the Reset Shift Curve
command. The spectrum is displayed in its original
size ratio again.
11 .1 Print Spectra
Principle • Load the spectrum file and select the Print Spectra
command.
• If required, drag & drop the spectrum file into the
File(s) to print for selected frame selection field.
• Select a plot layout.
• If necessary, click the Frequency Range tab and
define the range desired (see also chapter 2.9).
• Define the output format on the Options tab.
• Click Print.
The different tabs of the Print Spectra dialog are described in the
following.
Select Files
A B
Show Preview
Frequency Range
A If you activate this option button, the axis of all spectra in all
frames will be set to their maximum size.
B If you activate this option button, the frequency limits of the
plot layout selected before will be used.
C If you activate this option button, the Select frequencies for...
group field will be enabled. Either type the limit values manu-
ally or define them interactively (see chapter 2.9).
If you click the Get Display Limits button, the current limits
displayed in the spectrum window are set in the appropriate
entry field.
Options
11 .2 .1 D ia lo g
11 . 2 . 2 R e p o r t Ty p e s
11 . 2 . 3 S e t t i n g o p t i o n s f o r r e p o r t d a ta
The setting options (C in figure 358) for report data depend on the
report type selected. Not every setting option is available for each
report type.
11 . 2 . 4 R e p o r t St r u c t u r e
Sp e c t r u m F r a m e
The spectrum frame is displayed below the general report data and
product-specific data. This kind of report data is optional and dis-
played either always, never or only if the evaluation result is not ok.
Step 1:
Click the Setup Report Template com-
mand and select a report type from the
drop-down list.
Step 2:
Click on one of the setting options from
the list box. A second group field may
be displayed below the setting options.
Step 3:
Activate the respective checkboxes. Fill
in the entry fields or select an option,
where required.
Step 4:
Click the Save button to save the tem-
plate.
A Report header
B Analysis data in table format
C Spectrum view
11 . 2 . 7 Sp e c i a l f e a t u r e s i n c a s e o f a m u l t i
sample report
Sample-specific data are written in tabular form with the multi sam-
ple report. The setting options (chapter 11.2.3) to be selected are
the headers of the table columns, the samples form the name of
the respective table row. There is one table row for each sample.
Page 1 C Page 2
A Report header
B Analysis data in table format
C Spectrum view
11 . 2 . 7 . 2 S i n g l e e v a l u a t i o n r e s u l ts
11 . 2 . 8 Multi Evaluation
11 . 3 . 1 Report Layout
A multi sample report lists all spectrum files with the respective
data in one report.
11 . 3 . 2 Sta n d a r d Te m p l a t e
When generating an analysis report for the very first time OPUS
uses the Default.art standard template. The data provided in this
template considers the typical setting options available for the three
different report layouts (chapter 11.3.1), with regard to e.g. sample
parameters and evaluation results.
11 . 3 . 3 Report Output
If you activate the PDF option button, the File name and path entry
fields are enabled. To save the report define the file name and path
or edit the path, using the Select Path button.
11 .4 Quick Print
11 .5 New Layout
To c r e a t e a n e w l a y o u t t e m p l a t e . . .
3 Store the template. On the PLE view, select the File menu
and click Save as. Template layouts have the file exten-
sion *.PLE.
11 . 5 . 1 P L E Vie w
C
D
B
A The PLE drawing area consists of a black grid and a red bro-
ken-line frame. This frame represents the available print area
of the printer.
The PLE drawing area visually represents the plot view of the
current template, containing all the frames. Each frame can
be scaled and positioned individually.
To resize the drawing area right click somewhere on the
drawing area outside any frame. The Ple Page dialog opens.
PLE Icons
Icon Definition
Drawing a frame
Scaling spectrum
Icon Definition
Deleting items
11 . 5 . 2 Pop-up menu
If you right click a frame, the PLE pop-up menu opens. This menu
allows to individually configure the frame.
The Remove command deletes the frame from the drawing area,
while the Snap to Grid command aligns this frame exactly on the
closest grid lines.
11 .5 .3 Resizing Dr a w i ng Ar e a
1 ...right click somewhere on the drawing area.
2 The Ple Page dialog opens.
3 Use the different tabs of this dialog to define the general set-
tings for the drawing area.
4 To confirm the settings click OK.
The different tabs of the Ple Page dialog are described in the fol-
lowing.
Page
A
E
F
B
G
D H
A To specify the unit that you want to work with (inches or centi-
meters) activate one of the option buttons.
B You can print either in landscape or portrait.
C The maximum print area of the printer can be defined. Each
printer has a particular margin used for paper feed, and which
will not be printed on. These default settings are suitable for
all common printers and have to be rarely adjusted. The max-
imum print area is represented by a dotted red line on the
PLE document.
D The option buttons allow to define the paper size. The equiva-
lent unit has been defined by (A).
E Check this option button if the drawing area is to be enlarged
up to the size of the whole page.
F Check this option button if the drawing area is to be enlarged
up to the page width.
G If you check this option button, the entry field will be enabled.
Enter the scale factor manually. The scale factor required to
view the PLE pages is indicated in per cent. Most monitors
are not large enough to allow viewing the entire page at 100%
580 OPUS Reference Manual Bruker Optik GmbH
New Layout Chapt. 11
Grid
C
E
D
A If you activate the check box, the grid will be visible. The grid
can be useful to align objects on the drawing area.
B If you activate the check box, new objects created are posi-
tioned and sized to line up with the closest major grid lines.
C Usually, the grid is only displayed on the screen to simplify
object alignment. Activate the check box if you want to have
the grid printed on paper.
D The Major and Minor settings define the grid size in the unit
specified on the Page tab. Figure 378 exemplifies a grid size
of 0.25 x 0.25 inches. The grid can be further divided by filling
in the Minor entry fields.
E If you click this button, a color palette opens. Select or define
the color for the grid.
Header
The header and footer are printed on every page. The Header and
Footer tabs are identical, therefore only the Header tab will be
described in the following.
Te m p l a t e
The Save As command from the File menu allows to save tem-
plates, using any file name and directory. To load these templates
use the File menu and click Load File.
11 .5 .4 A d d i n g e l e m e n ts , m o d i f y i n g s p e c -
trum frame
Adding spectrum
1 Drag & drop the spectrum from the browser directly into the
spectrum frame on the drawing area. Load the spectrum file
from within the PLE view. The spectrum window is displayed
again. Drag the spectrum data block (AB or TR) from the
browser to the PLE tab ( ). Wait until the PLE view
opens by pressing the left mouse button. Drop the spectrum
block into the spectrum frame.
A d d i n g g r a p h i c a l e l e m e n ts a n d p e a k l a b e l
Graphical elements
9 Click the icon and draw a line to the circle. If you want
the line to have no arrow head, right click the line. From the
pop-up menu, select the Properties command and deacti-
vated the Arrow Head checkbox.
Text
To h a v e t h e s ta n d a r d s p e c t r u m pa r a m e t e r s
displayed...
To m o d i f y t h e s p e c t r u m f r a m e . . .
I t em
C
B
D
E
L i m i ts
A
D
B
F
C
G
A The entry fields allow to define the frame limits. The unit of the
values entered has been specified in the PLE Page dialog on
the Page tab (see chapter 11.5.3).
B The style type selected determines the data depiction within
the frame. You can have the spectra displayed in overlaid or
stacked mode.
C Activate the check box if you want to have the logo displayed
within the spectrum frame.
D If you activate the check box, the spectra x axis will be auto-
scaled within the frame.
E If you activate the check box, the y axis of the spectra will be
autoscaled within the frame.
Axes
The X Axis and Y Axis tabs are identical except for the Rotate
Numbers check box on the Y Axis tab. This option allows to rotate
the y axis annotations by 90°.
A
I
B
J
C
K
D
E
L
F
A Activate the check box if the axes, tick marks and labels have
to be displayed.
B The drop-down list allows to determine the thickness of the
axis lines. You can also enter the thickness manually.
C This check box is activated by default, i.e. each spectrum is
displayed with a separate x axis. Otherwise, one common
x axis is used for all spectra loaded.
D If you activate the check box, wavenumbers above 2000cm-1
are displayed using a smaller linear scale.
E If you activate the check box, the tick marks of the x axis are
displayed on top of the spectral frame.
F The check box is activated by default, i.e. the tick marks of the
x axis are displayed at the bottom of the spectral frame.
G Activate this check box if you want to have the grid lines dis-
played on the spectral frame.
H The axes are labelled according to the spectrum file loaded
by the editor. To change the axis label enter text into the entry
field.
I Using this button allows to define the axes color. If you click
the button, a color palette opens. Select or define the color
desired.
J If you click the button, a list box is displayed. Use this list box
to determine the font and size of the axis label.
K If you click the button, a list box is displayed. Use this list box
to determine the font and size of the tick marks.
L The check box is activated by default. In this case, the editor
calculates the optimum settings for the tick marks.
If you deactivate the checkbox, you determine the position
and interval of the tick marks manually. In this case, interval
means the wavenumber range between two tick marks. You
can manually enter the number of digits (maximum: 20) used
for the numeric tick mark labels into the Digits check box.
D a ta
A C
B
E
A Spectra added to a frame, are put into a slot. This slot deter-
mines the position of spectra. To change a spectrum view,
you first have to select the corresponding slot of the spec-
trum. Then, you have to define the properties for that slot
using the control elements for color, line style and line width.
B If you click the button, the spectra data selected in (A) is
deleted from the slot.
C The color palette allows to determine the color of the spec-
trum displayed.
D Use the list box to determine the line style of the spectrum
displayed.
E Define the line width of the spectrum displayed. Separate line
widths can be specified for the screen view and printout.
Peaks
A E
F
B
G
C
H
D
A Spectra added to a frame, are put into a slot. This slot deter-
mines the position of spectra. To change a spectrum view,
you first have to select the corresponding slot of the spec-
trum.
B Define the stroke width of the spectrum displayed. Separate
line widths can be specified for the screen view and printout.
Click Color to determine the color for the strokes.
C Activate this check box if the stroke color has to be the same
as the one for the spectrum labels. This helps to get a clear
display, which is especially useful if the PLE view consists of
more than one spectrum. By means of the Color button (B)
you can define an individual stroke color for each spectrum,
which will be the same for the spectrum label if you activate
the check box (C).
D The check box is activated by default. Therefore, extra space
is provided between the highest peak and the spectrum frame
for the peak labels.
E Use the list box to select the type and position of the labels
you want to assign to the peaks.
F This option is only enabled if the spectrum file contains a
data block. Starting with the highest peak you specify
the number of peaks to be labelled, using the All and None
buttons.
G This option is only enabled if the spectrum file contains a
data block. Enter the number of digits and decimals for
the peak labels.
H Click the button to determine the font type used for the labels.
Text can be included using text frames, i.e. either enter text manu-
ally into or import an existing text file to a frame.
Note: In case of manual text input you can also use meta
strings. The meta strings used for text parameters
can be directly typed into a text frame, and will be
To r e s i z e t h e t e x t f r a m e . . .
1 ...left click into the frame and scale it up or down using the
marker positions.
2 To edit text use the Edit command from the pop-up menu.
Right click into the text frame to select this command.
To i m p o r t a t e x t f i l e . . .
To m o d i f y t h e t e x t f r a m e . . .
The Text and Parameters tabs of the Text Properties dialog are
described in the following. For details on the Item tab refer to
chapter 11.5.4.
Te x t
B
A
A Defines the position of tab stops for the text in the frame. The
value entered represents the distance to the next tab. Multiple
positions can be used, separated by spaces, e.g. 8 12 20 40.
The last tab spacing is used for subsequent tab stops, if more
tabs are required.
B This button allows to determine the font and font size of the
text.
C To better align text within a frame activate the check box, as
the text lines will be horizontally centered.
Parameter
11 .5 .6 A d d i n g ta b l e a n d m o d i f y i n g ta b l e
frame
To a d d a ta b l e . . .
To a d d t e x t t o a ta b l e . . .
To m o d i f y t h e ta b l e f r a m e . . .
rate table cell is available for each spectrum within the table
frame.
D The more spectra you add to a frame, the more additional
table cells can be created to be able to display the parame-
ters for each spectrum. If you activate the If Multiple, automat-
ically add rows for new spectra check box, the number of
rows set on the Format tab will become meaningless.
Format
A G
B H
I
C
J
D
11 .6 Open Layout
Principle When using this command for the first time after the
OPUS installation you have to navigate to the particu-
lar directory. By default, templates are stored in
C:\Users\Public\Public Documents\Bruker\
OPUS_<version> and have the file extension *.PLE.
Plot layouts have the extension .*PLE and can consist of different
frames, e.g. spectrum or text frames, as well as of tables. For
detailed information on the plot layout elements refer to
chapter 11.5.
3D.PLE 3D depictions
LIBRARY.PLE library
REPORT.PLE reports
Definition You can view the result of a printout using the Print
Preview command. The Print command starts printing
immediately.
11 .8 Print Setup
Definition
This command allows to store data from the
data block in a script.
B A
12.3 M a cr o Converter
12.4 R un Macro
A B C D E
12.6.2 C o r r e c t i n g m a c r o s y n ta x e r r o r s
1 Click the Open Macro button to load an existing macro.
2 If the macro includes syntax errors, the following error mes-
sage pops up: Suppress Error Messages and load with Auto
Correct Option? Click Yes. The macro is loaded and all syn-
tax errors detected are automatically corrected.
3 A respective message appears. Confirm the message. The
Save File As dialog opens.
4 Store the corrected macro.
Measure 1
Manipulate 2
Evaluate 3
Display 4
Print 5
Macro 6
Edit 7
Validation 8
Settings 9
File 10
Definition Macros are written and stored as text files. This com-
mand allows to convert these text files to executable
binary codes. Compiled macros are started faster
than macros in text format.
12.8 New P ro ce d u r e
Wor k i n g w i t h t h e m a c r o w i z a r d
Step 1:
Step 2:
Step 3:
Step 4:
For the last procedure step define whether the file is to be saved or
printed. If you activate the Loop check box the procedure
previously defined will be repeated n times.
Double click the procedure block to define the parameters for the
respective block. The OPUS command dialog opens which allows
to set the parameters for the block selected.
Left click the block and press the Del key on the keyboard to
delete a procedure block.
Macro
F
Figure 410: Pop-up menu - Macro
E Saves the procedure. Define the file name and directory path
in the dialog displayed. The macro procedure is stored as
OPUS basic script and has the file extension *.obs.
F This is a global setting which specifies how the script should
react in case of an error. If you click Error Management, the
following dialog opens:
To o l s
Blocks
F
G
Figure 412: Pop-up menu - Blocks
Enter the name of the variable into the Variable entry field.
The pre-defined names of each variable as well as the
description of the corresponding variable value are included
in the list below. Enter the mathematical term for the variable
defined into the Expression entry field. Example:
Edit
1 2.9 E d i t P ro ce d u r e
1. In some cases OPUS also uses OVP (OPUS Validation Program) as reference
with regard to the command used to set and perform the instrument test. The in-
strument test itself is sometimes called PQ test (Performance Qualification).
Te s t c h a n n e l c o n f i g u r a t i o n
9
1 10
4 11
5
12
6
7 13
Component Definition
1 Current test channel The drop-down list shows the test channel
used in the current configuration
2 Test channel config- For each spectrometer test channel you have
uration to assign a measurement configuration to be
used during actual validation measurement.
This is a valid combination of source, beam-
splitter, measurement channel and detector
type.
The respective entry is pre-defined in the
appropriate selection field by default. As soon
as you have selected the correct entry the
appropriate instrument configuration (4) and
the bitmap (9) specified for the measurement
channel selected will be proposed.
Component Definition
Component Definition
11 Clear current config- Clicking the button deletes the current configu-
uration ration.
Note: Once deleted the settings cannot be
reloaded. If you have deleted a channel,
you have to repeat the complete setup
again.
13.1.1 M e a s u r e L a s e r Wa v e n u m b e r
Before starting the instrument test for the current configuration you
may first have to calibrate the laser wavenumber (LWN). Measur-
ing and assigning the laser wavenumber for the current test chan-
nel ensures the wavenumber accuracy of spectral data.
General features
To c a l i b r a t e t h e l a s e r w a v e n u m b e r.. .
1 ...make sure that the correct test channel has been selected.
When calibrating...
A ft e r c a l i b r a t i o n . . .
Figure 417a: LWN calibration okay 378a: New LWN setting required
If the LWN has to be set (see figure 417b), click the YES but-
ton.
2 ...a green check mark is added to the button in
the Setup OVP dialog.
• The LWN test result shows both the water bands speci-
fied and measured.
• You can also see the water band deviation as well as
the laser wavenumber, i.e. both the specified and newly
calculated one.
If the current LWN for a particular channel is within a
certain tolerance, the test has passed and no further
changes are required.
If the deviation is beyond a certain tolerance, the LWN
value for the current measurement channel is set to the
currently computed optimal value when clicking OK.
... the water vapor concentration of your instrument may be too low
or too high, and a warning pops up in OPUS.
In this case you have to test the LWN_Water.x water vapor spec-
trum (with x being any number) which is located in the
*\OPUS_<version>\Validation\Data\LWNCalibra-
tion\<Date>\<Time>1 directory.
A d d i t i o n a l i m p o r ta n t i n f o r m a t i o n
13.1.2 S e t t i n g u p O V P t e s ts
The OVP Test Setup tab contains the OVP test parameters.
H o w t o s e l e c t t h e t e s ts ?
2 Type the interval period into the entry field and select the
units from the drop-down list. Available units are: hours,
days, or months.
3 Click Setup to have the available PQ tests displayed:
If the reference spectra have not yet been measured, the following
warning symbol is displayed in the OVP - PQ Test Protocol dialog.
Reference spectra used for the tests are stored in the user data-
base. Whenever used the reference data are temporarily extracted
and afterwards removed again. This avoids accidental deleting or
overwriting. A copy of the reference spectra is stored in the follow-
ing directory:
*\OPUS_<version>\Validation\Data\PQ_Reference\
<Date>\<Time>
A ft e r m e a s u r i n g r e f e r e n c e s p e c t r a . . .
13.1.2.1 P Q Te s t D e s c r i p t i o n
Energy Test Signal to Noise Test This test verifies that the
signal-to-noise ratio is better
than a specified limit.
The S/N ratio is calculated separately for each spectrum, using the
mean value of all 10 results.
E n e r g y D i s t r i b u t i o n ( S i n g l e - C h a n n e l Sp e c-
trum)
Figure 427 and 428 show the single-channel spectra of the refer-
ence file in MIR and NIR region. The integral over the total refer-
ence single-channel spectrum is set to 100%.
I c e B a n d Te s t
The ice band test is part of the energy test and compares the inte-
gral in the ice band region (3485 - 3050cm-1) with the integral of the
reference spectrum measured for the PQ test. The integral should
not exceed a preset limit.
S e a l i n g Te s t
X - A x i s F r e q u e n c y C a l i b r a t i o n Te s t ( Wa v e -
number Accuracy)
Figure 431 and 432 show typical water vapor spectra in the MIR
and NIR region.
Y- A x i s R e p r o d u c i b i l i t y Te s t ( P h o t o m e t r i c
Accuracy)
Figure 433 and 434 show the spectra of two glass filters in NIR and
MIR.
P o s i t i o n Te s t
This test is only used with the HTS-XT module. For detailed infor-
mation on the position test refer to the HTS-XT user manual.
13.1.2.2 O Q Te s t D e s c r i p t i o n
Figure 435:
2 Type the interval period into the entry field and select the
units from the drop-down list. Available units are: hours,
days, or months.
3 Click Setup to have the available PQ tests displayed.
Water Vapor Test This test verifies that the water vapor con-
centration does not exceed a specified value
(only with older spectrometer types).
Tests which are not available for the selected instrument type are
disabled, all the other tests are activated. If you do not want to per-
form a specific test, deactivate the respective check box.
P h o t o m e t r i c A c c u r a c y Te s t
Figure 438: Water spectrum for NIR with different layer thickness
S e n s i t i v i t y Te s t
S/N ratio is different for systems used in the MIR and NIR region,
and is displayed in the corresponding report.
E n e r g y D i s t r i b u t i o n Te s t
Wa v e n u m b e r A c c u r a c y Te s t
To ensure that the water vapor band is completely resolved the test
is performed by a high-resolution setting. The NIR region uses the
band at 7306.74cm-1 for calibration, and the MIR region at
1554.353cm-1.
Figure 442 and 443 show a typical water vapor spectrum in the
MIR and NIR region. Depending on the source the BRM 2065 is
alternatively used in NIR for testing if the water vapor band speci-
fied is beyond the limits defined.
R e s o l u t i o n Te s t
This means for most spectrometer types: 1cm-1 for MIR and
2cm-1 for NIR, see figure 444. The spectrum is first logarithmized.
Then, the full width at half maximum (FWHM) of a water band
(7306,74cm-1 for NIR and 1554,353cm-1 for MIR) is determined.
The result is compared to the specified resolution of the spectrom-
eter.
S c a n Tim e Te s t
Wa t e r Va p o r Te s t
A l i g n m e n t Te s t
The alignment test verifies that the absolute peak position of the
interferogram is within the range specified.
L i n e a r i t y Te s t
R e p r o d u c i b i l i t y Te s t
The reproducibility test confirms that all moving parts (mirrors etc.)
of a multi-channel instrument are positioned correctly and in a
reproducible way.
P o s i t i o n Te s t
This test is only used with the HTS-XT module. For detailed infor-
mation on the position test refer to the HTS-XT user manual.
R a t i o Te s t
This test is only used with the HTS-XT module. For detailed infor-
mation on the position test refer to the HTS-XT user manual.
S e a l i n g Te s t
13.1.2.3 A d d i t i o n a l Te s ts ( o p t i o n a l )
H o w t o s e l e c t t h e t e s ts ?
1 Activate the Run check box and select the respective type of
test.
1
2
Component Definition
2 Test category The tests set up for the active test channel
selected, as well as the their status are dis-
played. The following status types are possi-
ble:
• test passed and expires in
• test not yet run
• test failed
• test expired since
Note: Tests which have expired, or which
have not yet been performed are automati-
cally activated for measurement. Tests
which have not yet expired are not acti-
vated, and their remaining time is dis-
played accordingly.
If the validity period of a test has not expired,
but the remaining time is e.g. 2 hours, you can
activate the check box for that test and run the
test earlier.
If the validity periof of a test has already
expired, but you are not prepared to perform
that test (external standards required), you
can deactivate the check box for that test.
Note: If you decide not to perform a test
which has already been expired, you may
not be allowed to use this test channel in
regulated environment.
Component Definition
W h e n s ta r t i n g t h e t e s ts . . .
...they run in the background, invisibly for the user. The names of
the single tests are shown on the OPUS status bar.
The spectra produced when running the tests are stored in a direc-
tory structure which will be automatically generated. In the
*\OPUS_<version>\Validation\Data directory a sub-directory is cre-
ated for the current date, supporting the following format: YYYYM-
MDD1 (e.g. 20090212).
The spectrum file names always start with either PQ or OQ, the
spectrometer configuration and a test definition:
e.g. PQ_MPA_Sample_Compartment_Energy.0.
The date and time is also used for the report name to allow a clear
correlation between spectra and reports. The spectra contain the
test result together with other important data in the information data
block ( ).
A ft e r t h e t e s t r u n . . .
Te s t r e p o r t
The test report contains one summarized test result page and sev-
eral different pages for each test with all the test spectra and mea-
suring parameters included. Reports are stored in the separate
VALIDATION\REPORTS directory. Figure shows the test report of
a PQ test.
To o p e n t h e d i a l o g . . .
...click the OPUS status light at the lower right end of the OPUS
interface.
W h a t d o e s t h e s ta t u s l i g h t i n d i c a t e ?
The status light always indicates both the hardware status as well
as the current test status of the active test channel. The test status
is displayed in different colors and can be as follows:
• Green: all tests have been configured, or have passed
and are still valid
• Yellow: at least one test has expired
• Red: last test has failed
I n s t r u m e n t St a t u s d i a l o g
A The first row of icons represents the test status of the instru-
ment hardware components (see also chapter 7.8).
B The second and third row of icons are reserved for OVP tests.
If more than one test channel has been set up in OVP, the
corresponding icons are displayed for all channels.
The test status of the OVP tests is indicated as follows:
- INACTIVE (yellow): the single tests of the particular test
category are disabled
- PASSED (green): test configured or passed and still valid
- EXPIRED (light-blue): test expired
- FAILED (red): last test failed
To s ta r t t h e O V P t e s t . . .
2 Activate the check box for the respective test category. The
single tests of the test category are displayed in a
browser-like view.
3 Define whether the test report is to be printed and/or dis-
played automatically, with or without the spectrum pages
created for each test. Note: To have the reports displayed
make sure that the Adobe Acrobat Reader is installed on
your computer. The Adobe Acrobat Reader is available
on the OPUS installation stick.
4 If desired, add a comment into the respective entry field.
5 To start the test(s) click Run Tests.
W h e n s ta r t i n g t h e t e s ts . . .
...they run in the background, invisibly for the user. The names of
the single tests are shown on the OPUS status bar (see also
chapter 13.2).
A ft e r t h e t e s t r u n . . .
Te s t r e p o r t
13.2.2 Sta r t i n g O V P t e s ts f o r m u l t i p l e
test channels sequentially
Note: The sequential starting of OVP tests for multiple test
channels makes sense in case of OVP tests which do
not require any user interaction.
M e a s u r i n g St e ps :
3 Position the cursor above a yellow drop down list, a tool tip is
displayed.
The tool tip indicates what kind of problem has occurred and
how it can be solved.
The Open Planner command from the Setup menu allows to auto-
matically run the OVP PQ test. For detailed information on the
Open Planner command refer to chapter 14.11.
6 Define the start of the test either manually or use the arrow
buttons.
7 Activate the No end date option button for the range of recur-
rence.
8 Confirm the settings made by clicking OK.
2 Click the OVP Trend Chart button. The test result is dis-
played.
E
Figure 460: Trend Chart of a sensitivity test
A Name of test
B Limit: the test-specific limit is displayed without data points as
this line is a constant
C Measured value: the data points represent the values mea-
sured
D Y-axis: is automatically scaled
E X-axis: shows date and time of measurement
Depending on which test you have selected, there may be two lim-
its shown in the trend chart. In this case, the measured value has to
be within these two limits (figure 462).
Right click into the trend chart. The following menu pops up:
D This option allows to show and hide the lines displayed in the
trend chart. Click this option to open the following dialog:
Depending on whether you want to hide the line displayed for the
limit or the measured value, click the respective Visible column to
remove the check mark.
When installing an accessory for the very first time, the accessory
has not been registered yet in OPUS. If an accessory is positioned
into the sample compartment, OPUS reads the data of the acces-
sory and checks whether the accessory has already been used in
one of the OVP test channels defined.
If this is not the case, OPUS prompts you to prepare a new test
channel for the current accessory used.
9 Set up the OVP tests on the OVP Test Setup tab. More
details are described in chapter 13.1.2.
10 Click Save and Exit to store all settings made.
4 On the Advanced tab, click the Save button to save the mea-
surement experiment for the accessory.
13.6.3 P e r f o r m i n g O V P t e s ts
After setting up the test channel for the accessory perform the OVP
tests.
13.6.4 A u t o m a t i c P e f o r m a n c e Te s t
In case of the performance test only one reference and one sample
measurement starts. Wait for the test result before you continue
measuring.
13.6.5 Wo r k i n g w i t h s p e c i a l t y p e s o f
accessories
AT R a c c e s s o r i e s w i t h e x c h a n g e a b l e c r y s -
ta l s ( T E N S O R a n d V E R T E X s e r i e s o n l y )
The ATR accessory is now registered and can be selected from the
Accessory drop-down list of the Measurement dialog. To activate
this drop-down list proceed as follows:
F i b e r p r o b e s ( F T- N I R s y s t e m s o n l y )
If the exact length is not listed, always select the next higher
length, e.g. select 100m if the length of the fiber used is 95m.
The fiber probe is now registered and can be selected from the
Accessory drop-down list of the Measurement dialog. To activate
this drop-down list proceed as follows:
5 If required, on the OVP Test Setup tab select the OVP test
category.
➣ Details on the OVP test categories are described in
chapter 13.1.2.
6 Click the Save and Exit button.
1 4.1 Lo g o u t
Definition This command allows to log out the current user and
login a new one.
If you click Exit from OPUS, you close the login dialog
and terminate OPUS.
Note: The Logout command does not unload files from the
browser, nor stops processes still running (e.g. mea-
suring).
Basic Features
A conversion table is a text file with the extension *.PCT, and can
be generated by any kind of text editor. Additionally, you have to
specify the file containing the info text definition.
The general syntax of the file with the *.PCT extension is as fol-
lows:
#Ixx:yyy:LABEL with:
Ixx defining the number of the line in the info block, xx being a
two-digit number starting with 01.
#I01::SNM:TITLE
#I02::MOLFORM
#I03::MW
#I04::CAS REGISTRY NO
#I05::MP
#I06::BP
#I07:SFM:SAMPLING PROCEDURE
#I08::
#I09:CNM:ORIGIN
#I10::CROSS REFERENCE
#I11::CHARGE NO
#I112:HIS:DATA PROCESSING
U s e r D a ta b a s e
Color Definition
Setup Record
F
G
H
E Enter the user name. The user name is stored in newly mea-
sured spectra and in the audit trail of spectra and methods,
and shown accordingly.
F The user group specifies the access rights of a user. The
drop-down list includes the Administrator, Labmanager and
Operator user group. The first two user groups have unlimited
access to all OPUS workspaces. The Operator user group
has only access to the workspace assigned. Note: For users
with Operator rights at least one workspace must be
assigned.
If you select the Operator user group, the list of assigned
workspaces is displayed on the right.
Global Options
A u d i t Tr a i l
The Audit Trail tab indicates the audit trail of the following actions
for each user.
• Successful login
• Login attempt of a locked user
• Login with password expired
• Changing password
C
E
D
Note: The Save and Clear Audit Trail actions are recorded in
the audit trail of the user currently logged in.
Definition Each user can change his own password. There are
different reasons for changing a password:
I f y o u r pa s s w o r d h a s a l r e a d y e x p i r e d . . .
Click Change Password. The OPUS Login dialog opens. You can
now log in OPUS by using the new password.
I
G
J
Figure 489: Setup User Macro List - User Macro Entry
A Select the correct language for the entry from the drop-down
list. The default language is the language in which OPUS has
been started. You can only edit the entries for the selected
language. If the language has been changed, you can access
the entries for the language newly selected.
724 OPUS Reference Manual Bruker Optik GmbH
Setup User Macro List Chapt. 14
[ENGLISH]
.\macro\it.mtx@8@Instrument Test@Instrument Test@Run
the Instrument Test
The data required (key, serial number etc.) to register OPUS in the
above mentioned cases can be obtained from the OPUS stick.
The registration has to be performed only once and is valid for all
users.
1. If you work without any spectrometer connected, you can use OPUS for 30 days
without registering, after the program has been installed.
To r e g i s t e r O P U S . . .
D
Figure 490: Registration Dialog
4 Click the icon to delete the toolbar from the OPUS inter-
face.
B
A
To a d d a c o m m a n d t o t h e t o o l b a r.. .
To d e l e t e a c o m m a n d f r o m t h e t o o l b a r
1 Click the appropriate icon in the toolbar and drag it from the
toolbar while pressing the left mouse button. If you release
the mouse button, the icon is deleted from the toolbar. Note:
Make sure that the Customize dialog is open.
2 You can also right click the icon and select the Delete com-
mand from the pop-up menu. Note: Make sure that the
Customize dialog is open, otherwise no pop-up menu
will be displayed.
To a d d s i n g l e m e n u s o r s u b - m e n u s t o t h e
t o o l b a r.. .
14.7.5 C r e a t i n g Too l b a r
B
A
To c r e a t e n e w t o o l b a r s . . .
2 Enter the name of the new toolbar into the Toolbar Name
entry field.
3 Confirm the name by clicking OK. The new entry will be dis-
played on the lower end of the Toolbars list box (A in
figure 492).
To rename or delete this new entry, select it and click the
Rename or Delete button.
Note: Toolbars which have not been created by the user
cannot be deleted. If you select such a toolbar the
Reset button will be enabled. This button allows to
reset the standard icons of the toolbar.
O p t i o n s ta b
For details on the menu settings in this dialog refer to chapter 14.8.
2 Position the cursor on the edit field arrow head. Another edit
field is displayed.
• Image:
• Text:
Right click any icon in the toolbar and select the appropriate
depiction type from the pop-up menu. A check mark will be set in
front of the depiction type selected.
To a d d a d i v i d i n g l i n e . . .
To d e l e t e a d i v i d i n g l i n e b e t w e e n . . .
...right click one of the icon and remove the check mark at the Start
Group command in the pop-up menu.
If you want to create a name for the icon, click the icon and
activate the Image and text option button. The Button text
entry field will be enabled. Now, enter an appropriate name
for the icon.
- Unfold
- Slide
- Fade
D The check box is activated by default and displays the menus
with a light grey shadow on the right side.
E This button allows to open the online help of the respective
tab.
A If you activate the check box, only those commands are dis-
played in the menus which have been used recently. To open
the complete menu view click the double arrow ( ) below
the last command shown.
B You can have the complete menu view be displayed after a
short period of time. Activate the check box.
C If you click this button, all the default settings are restored.
14.8.1 A d d i n g o r d e l e t i n g m e n u c o m-
mands
14.8.2 M o v i n g o r c o p y i n g m e n u c o m-
mands
When moving click the menu or menu command and drag and
drop it to the desired position in the toolbar.
When copying press the Ctrl key and drag and drop the menu
command or menu to the desired position in the toolbar.
3 Open the menu which you want to edit and drag and drop
the New Menu entry into this menu. The new entry is now
displayed in the menu:
To c h a n g e t h e n a m e o f t h e m e n u e n t r y.. .
The new menu entry has a small arrow head on the right side. If
you click this arrow head, a small field opens in which you can drag
and drop the sub-menu commands.
To a d d a d i v i d i n g l i n e . . .
To d e l e t e a d i v i d i n g l i n e . . .
14.9 Shortcuts
14.9.1 A s s i g n i n g s h o r t c u ts
1 On the Customize dialog, click the Keyboard tab .
2 Select the OPUS option from the Set Accelerator for drop-
down list.
5 Use the Press New Shortcut Key entry field to enter the
shortcut to be assigned to the command selected. Press e.g.
STR+ and the key desired. Note: If a shortcut has already
been available for the command selected, this shortcut
will be displayed in the Current Key list box. An existing
shortcut can also be deleted (see chapter 14.9.2).
6 Whether the shortcut has already been assigned to a differ-
ent command or not is displayed accordingly.
14.9.2 D e l e t i n g s h o r t c u ts
1 From the Category drop-down list, select the OPUS menu
which contains the command with the respective shortcut.
2 In the Command list box, click the command definition. The
shortcut is displayed in the Current Keys list box.
3 First, click the shortcut and then Remove.
14.9.3 R e s e t t i n g s h o r t c u ts
1 Select the appropriate option from the Set Accelerator for
drop-down list, whose shortcuts you want to delete.
2 Click Reset All.
The different tabs of the User Settings dialog are described in the
following.
General
2 1 C F R 11 R i g h ts
1. GLP means Good Laboratory Practice and is some kind of quality assurance di-
rective.
Preferences
Display
If you activate the Whole file name option button, all spectra
with the same file name (without extension) will be displayed
in the same color.
Activate the File name check box if only a specific number of
file name characters has to be identical. Example: if a file
name has 12 characters and you enter 1 into the First charac-
ter entry field and 8 into the Length entry field, the first 8 char-
acters of any file name will be checked to see whether they
are conform with the file name series.
Click the Spectra Colors... button to assign a color sequence
to the spectra. The following dialog opens:
New File
Evaluations
On the Evaluations tab you define whether the result of the evalua-
tion performed should be directly displayed in OPUS, or printed.
Diagnostics
C o m pa n y S e t t i n g s
1 4 . 1 0 . 1 In s t r u m e n t Te s t ( o n l y i n c a s e o f
n o n - O V P s u p p o r t e d i n s t r u m e n ts )
The test channels of the Instrument Test 2 tab correlate with the
third row of the Instrument Status dialog. Thus, a total of 12 differ-
ent test channels can be assigned, if required.
First, activate the IT check box which represents the active test
channel, and either enter the path manually or click the Path but-
ton.
N o t e s w h e n s e l e c t i n g t h e pa t h :
If you have selected a path, OPUS will be informed that you want to
assign the Instrument Test macro (which can be installed during
OPUS installation) to a particular channel.
OPUS assumes that the IT.mtx macro is located in this path. If the
macro cannot be found, a Runtime Error in Macro message pops
up when you attempt to run the test.
Example:
If you have entered a path into the appropriate IT entry field on the
Instrument Test tab, an icon is assigned for that particular channel
on the Instrument Status dialog (see chapter 7.8).
M e a s u r e m e n t e x p e r i m e n ts :
You can add as many XPM files as you like. It is important that you
only select experiment files which have the same basic spectrome-
ter configuration (source, beam splitter, measurement channel
etc.).
B C
A D
14.11 O p en P l an n e r
Basic Features
If you double click anywhere on the planner, you can start defining
the entry. It is also possible to colorize the entry to see at once
what kind of appointment or event it is about.
OPUS has to run, before being able to start the scheduled item at
the pre-defined time. If you select the Open Planner command, the
following dialog opens:
1 4 . 11 . 1 Vie w O p t i o n s
To c h a n g e a p l a n n e r v i e w. . .
The following planner view options are set by default in the pop-up
menu:
1 4 . 11 . 2 C r e a t i n g p l a n n e r i t e m s
1 Right click the respective day and select the New
Appointment command from the pop-up menu.
2 Define the settings in the dialog displayed.
3 Save the settings made.
D
B
Starting from today’s date you can also use the arrow keys
to define the date desired. The time can be set either man-
ually or by using the arrow keys.
D If you activate the check box, the entry becomes an event that
lasts at least one day. Note: An all day event does not
occupy blocks of time in your planner. Instead, it is dis-
played at the top of the date which you have specified in
the planner. This kind of event can also be created by using
the New All Day Event from the pop-up menu (see
figure 519).
E The Show time as drop-down list allows to define the time sta-
tus of the entry. This is recommended if others have access
to your planner.
F To save your settings click Save and Close.
To c r e a t e r e c u r r i n g i t e m s . . .
1 ...right click the planner view and select the New Recurring
Appointment or New Recurring Event command. The dialog
that opens is the same for both commands.
2 Define the settings.
3 Confirm the settings made by clicking OK.
4 At last, define the name of the recurring appointment or
event.
D
Figure 522: Appointment Recurrence
You can also right click the appointment or event and select the
Open command from the pop-up menu to have the Edit
Appointment dialog displayed.
To r e n a m e s i n g l e a p p o i n t m e n ts o r e v e n ts
i n t h e p l a n n e r. . .
14.12 N e w Wor k s pa c e
1 4.14 Cl o se Wor k s pa c e
1 4.15 S en d Wo rk s pa c e
If you click this command, the OPUS online help starts. The online
help can also be started by the F1 key on the PC keyboard.
If you position the cursor on this command, the names of the online
manuals available in OPUS are displayed. The online manuals are
provided as PDF files.
If you select this command, a dialog opens and displays the OPUS
version used, the user name registered and the OPUS software
package available (see also chapter 1).
ln re i = ln r + i (B-3)
depend on each other due to the causality principle. The two parts
can be transformed into each other by the so-called Kramers-Kro-
nig Transformation (KKT). OPUS uses KKT to first calculate the
phase rotation angle as function of the wavenumber v from
the measured reflectance spectrum R = r 2 of an optically
thick sample. Then, the reflectivity of the air/sample edge (vertical
incidence) is calculated from r and using the Fresnel
equation:
i
e = n – 1 n + 1 (B-4)
The real and imaginary part of the refractive index are calculated
as follows:
12
n = 1 – R 1 + R – 2R cos (B-5)
= 2R 1 2 sin 1 + R – 2R 1 2 cos (B-6)
= n 2 – 2 (B-7)
= 2n (B-8)
Performing KKT
When selecting the frequency range make sure that the reflectance
spectrum is flat at the edges. Avoid any significant structures at the
edges of the spectrum.
Measuring Accuracy
The reflectance spectrum should not contain any zero lines as oth-
erwise the ln R KKT integrand becomes singular. In case of
highly absorbing material, e.g. NaCl, SiO2, the reflection decreases
–3 –4
in some positions from 10 to 10 which may lead to the fact that
the measuring errors are greater than the wanted signal.
Incorrect Model
If the sample you use is not thick enough, roughen the sample to
avoid back reflection. Strictly speaking, the calculation of optical
Examples
(2.3)
j =1 j =2 j=3
pj 284 60 40
rj 5.5 28 16
(2.4)
j =1 j =2 j=3 j=4
rj 30 70 20 70
Polystyrene (5 oscillators):
2 ,4
pj 57 100 65 40 50
rj 60 7 12.5 5 10
V
Valve Control 241
VisualBasic Script 82
W
Window
Cascade 150
Window Display 540
Wizard 8, 142
Workspace 8, 11
Close 775
New 774
Send 776
Z
Zerofilling 300