Ftir Lab Report
Ftir Lab Report
Introduction
c
(1)
Where c is the speed of light that has quatity of 3 x 1010 cm/second. The energy of
particle often referred to as a photon (E in Joules) which related to wavelength and frequency
as follows (Equation 2)
hc
E h (2)
Based on provided spectrum region (Figure 1), -ray and X-ray become the higest
energy radiation which shows a great energy needed to break bonds in molecule. Whereas,
radio wave have very low energies, only enough to make spin transition in nuclear region and
Infrared region gives significant information about the molecular vibrations [2]. Infrared has
three regions includess near (12820-400 cm-1), mid (4000-400 cm-1) and far (400-33 cm-1)
infrared.
Molecules that can be detect in infrared spectroscopy must have specific frequencies
at which they rotate or vibrate corresponding to dicrete energy levels known as resonant
frequencies. Those are detemined by the mass of the atoms, the shape of the molecular
potential energy surfaces and by the associated vibronic coupling. Infrared spectroscopy can
detect interaction of polarity of the chemical bonds of the molecules called dipole moments
[4]. Dipole moment is a quantitative measure of the polarity of a molecule bond which owned
by diatomic molecules containing atoms with different elements named polar molecules.
Dipole moment can be used as distinguisher between molecules that have the same formula
but with different structures [5]. Infrared active molecule must undergo a change of its dipole
moment when vibrating. The simplest modes of vibration that are infrared active are
streching and bending.
Analog-
Source Interferometer Sample Detector Amplifier computer
to-digital
FTIR has several basic components as shown in Figure 2 . The radiation produced by
source is passes trough an interferometer to the sample before reaching detector. By signal
amplification, in which high-frequency contributions have been eliminated by a filter, the
data are converted to digital form using an analog-to-digital converter and transferred to the
computer for Fourier-transformation [6]. The main benefit of FT-IR technique over the
conventional IR is the higher signal-to-noise ratios for spectra recorded for the same
measurement time [7]. It also has high precision of wavenumber kalibration (0.01 cm-1) that
known as Connes advantage. Besides its high wavenumber accuracy, FT-IR utilizes circular
apetures which has a larger area than the linear slits used in grating spectrometers that can
provide higher troughput of radiation[8].
Measurements for four different samples (Figure 1 Appendix A) were carried out by
using FTIR Instrument. Polymer was taken from hand glove, small piece of plastic hose,
pippete tip and syringe pump. Each sample was cut it into a small piace to make easier in
term of positioning them to the sample place. Then twice running for each sample was
collected. As with all FT-IR measurements, an infrared background was collected.
The raw data spectra was plotted by using Origin Pro software to be analyzed. The
plotting result is a graph that shows transmittance (%) vs wavenumber (cm-1) in y and x axis
respectively. Then, we bravely analyzed the graph by determining the peak and comparing
with several references to know various materials.
FTIR spectra were recorded in the transmittance mode. The FTIR spectra of A-D
sample are shown in Figure 3A-3D, respectively. These bands can be observed to determine
functional group for make certain the material. The reference and list of the functional gorup
in each peaks also given in Table 1A-1D.
- Sample A
A
110
-1 Fingerprint region
1683-1614 cm
100
90
Transmittance(%)
80
2237 cm-1
2851 cm-1
2359 cm-1
70 2928 cm-1
1435 cm-1
60 3419 cm-1
1176 cm-1
50
40
4000 3500 3000 2500 2000 1500 1000 500
-1
Wavenumbers(cm )
For Figure 3A, a broad band at 3419 cm-1 is assigned to O-H streching. Another peaks
are showed at 2928, 2851, 1614-1683, 1435 and 1176 cm-1 which are correspond to C-H
vibration, CH2, C = C Streching, CH2 Scissoring and C-C streching respectively. The
fascinating peak to be observe are at wavenumber of 2539 and 2237 cm-1 that are concerned
with C N streching. Hence, Based on C N functional group, it can be guessed that A
sample contain of nitrile material. Furthmore, The FTIR spectra of A sampla also has similar
shape with nitrile material that has been worked by Sadeghalvaad, et al. [9].
B
120
110
-1 -1
2838-2990 cm 1409-1473 cm
100
-1
Transmitance (%)
1038 cm
90
80
70
-1
742 cm
60 -1
958 cm
-1
1124 cm
50
-1 -1
1721 cm -1 1069 cm
1255 cm
40
3500 3000 2500 2000 1500 1000 500
-1
Wavenumbers(cm )
According to the Figure 3B, peaks at 2838-2990 are assigned to the C-H streching.
Other characteristic of sample B at 1721, 1409-1473,831, 958 and 742 cm-1 are corresponded
to C =O streching, CH2, CH rocking, Trans CH wagging, C-Cl and –CH=CH- (cis)
respectively. The appearance of C-Cl bending can be use to determine this material to be
specified. Those, corresponds with polyvinyl chloride (PVC). Because the FTIR spectra of
sample B is also match with characteristic of PVC spectra which has been done by Ramesh et
al. in previous work.
C
105
-1
898 cm
-1
2997-2792 cm
100 1045 cm
-1
95
Transmitance (%)
90
2722 cm-1
-1
1102 cm
85
-1
1166 cm
80 -1
1358 cm
75
-1
1455 cm
70
65
-1
1375 cm
60
3500 3000 2500 2000 1500 1000 500
-1
Wavwnumbers (cm )
The characteristic spectra of sample C is shown at 2792-2997 and 2771 cm-1 and
expected to be due to C-CH3. The absorption band at 1455, 1375 and 898 cm-1 are bending
vibration of CH2, C-H and C=CH2 respectively. These peaks are close to the characteristic of
polyprophylene (PP). Wang et al. already reported FTIR spectra for pure polyprophylene
polymer that showspeaks at 3000 for C-H streching and also at 1460 and 1375 cm -1 for C-H
bending. Another report was given by Sarker el al. which exhibited PP peaks at 2841.66-
2955.87, 1465.95 and 1377 cm-1 for C-CH3, CH2 and CH3 bending respectively. Therefore,
with all of this study, it can be assume that sample C is PP polymer.
- Sample D
The FTIR spectra of sample D is shown in Figure 3D. The chacacteristic of peak of
sample also listed in Table 1D. The peak at 2915-2948 cm-1 are assigned to CH2 asymetric
streching. Where, CH2 symmetric streching group is indicated at wavenumber of 2836-2866
cm-1. In addition, CH3 symmetric group is corresponded to the peak at 1375 and 1357 cm-1.
[19] [20]. This study was well matched with the work that has been done by Muhonjaet al.
They investigated biodegradable polyethylene and reported that the main bands of LDPE
consist of a situated band at 2900 which assignable to CH2 as an asymmetric streching, a
band around 1461-1466 cm-1 revealing a bending deformation [21] . Given the functional
group, it can be concluded that the structure of this polymer has to be polyethylene.
D
110
100
90
Transmittance(%)
-1
2721cm
80 1263 cm
-1
-1
70 1357 cm
2836 cm-1 -1
60 1456 cm
2866 cm-1
50
2948 cm-1 -1
1375 cm
40 -1
2915 cm
Conclusion
References
A B
C D
Appendix B Product data sheet of materials
PVC
NITRILE
PE