ABSTRACT
In this paper, a comparative study of different edge detection methods had been done to determine the vessel wall elasticity for
early diagnosis of the Deep Vein Thrombosis condition. Currently, in most research found that the measurement of the vessel
detection conducted solely on the raw image obtains from the ultrasound. Thus, the precision of the measurement could be
controvertible from time to time. As a matter of fact, the image consists of its individual characteristics or properties that
cannot be verified distinctly. Therefore, various methods of edge detection techniques had been applied to the B-mode
ultrasound image. There are several edge detection techniques available for pre-processing in computer vision. Though,
Canny, Sobel and Roberts are some of the most applied methods. This paper compares each of the methods by the evaluation
of the Mean Squared Error (MSE) and Peak Signal to Noise Ratio (PSNR) of the output image. The evaluations are using
Matlab software, the algorithms applied on the B-mode image of 10 subjects that had been volunteering for the purpose of
the study. Both the MSE and the PSNR were in numeric values, that includes the vessel wall elasticity measurement of the
popliteal vein, therefore, the performance of the algorithm is determined.
%
PSNR 10 log# $ (
&'
(2)
VOL. X, NO. X, XXXXXXXX ISSN 1819-6608
ARPN Journal of Engineering and Applied Sciences
©2006-2013 Asian Research Publishing Network (ARPN). All rights reserved.
E. Image Processing
Edges are detected by using the Sobel, Canny, Roberts,
combination of Canny and Sobel, and also combination
method of Canny and Roberts. For Sobel and Roberts
operators, the method applied by thresholding the gradient
function (Mamta Juneja,2009). While for the Canny
method, a threshold is applied to the gradient using the
derivatives of Gaussian filter. Figure 2 shows an example of
the original image used for the edge detection method.
0 ,1
Figure 2: The original image of the popliteal vein (B-mode
ultrasound image). 0* 1 2
1 0
(5)
,1 0
i. Detection using the Sobel operator.
0/ 1 2
0 1
Sobel filter has a simple approximation concept of (6)
gradient with smoothing. The 3x3 convolution mask is
usually realized to detect gradients in X and Y directions.
Equation (3) and (4) show the operator that consists of a
pair of 3x3 convolution masks (kernels) (O. Marques, 2011).
Figure 3 shows the example of an image using the Sobel
operator.
,1 0 1
)* +,2 0 2. (3)
,1 0 1
,1 ,2 ,1
)/ + 0 0 0. (4)
1 2 1
Figure 4: Roberts edge map of Figure 2.
iv. Combination of Canny with Sobel and Canny with PERFORMANCE EVALUATION
A. MSE and PSNR measurement
Roberts.
As mentioned earlier, the edge detection methods were
In the combination of Canny with Sobel and Canny
assessed by objective quality measure to find reliable
with Roberts operators, the performance evaluation
indicated by the measurement of MSE and PSNR. Thus, statistical evidence in distinguishing among the output
images obtained. The numerical results shown in Table 1 and
the quality of the output image from both combinations can
Table 2 were the outcomes from the Matlab program which is
be compared and determine the acceptable method for edge
developed to process the B-mode images and to compute
detection in vessel wall elasticity measurement. Figure 6
and Figure 7 shows the example of image process by the MSE and PSNR value of the output images. The MSE
both combinations Canny with Sobel and Canny with values measured in the range of 1.60 to 16.73. The PSNR
value measured at range start from 5.9 dB to 15.98 dB. Based
Roberts operators.
on the measurement obtained, there were an absolute
difference in the measurement of MSE and PSNR for
different edge detection operators.Figure 8 and Figure 9 show
the comparison of average value between the applied
algorithm in the measurement of MSE and PSNR calculation.
It shows the pattern of the value for the mean MSE and the
mean PSNR for each edge detection method applied.
Significantly, from figure 8 and figure 9, the
combination of Canny and Sobel operator had the lowest
value of PSNR which is 9.62 dB and with the highest average
value of MSE that is 7.68. While for the Roberts operator, it
has the highest average value of PSNR that is 12.82 dB and
lowest average range of MSE which is 4.49. Moreover,
based on the output image observed for the Sobel and
Roberts operators, the operators were not acceptable for the
Figure 6: Combination of Canny and Sobel edge map of analysis of edge detection for the vessel wall measurement,
Figure 2. eventhough that both operators having the higher value of
PSNR indicate that the image is good in image quality. The
resultant images of the Sobel and Roberts did not allow for
the wall measurement because the vessel wall of the vein
cannot be seen clearly and the measurement cannot be done
based on that images. Therefore, for the image analysis of the
B-mode ultrasound image edge detection using the
combination of the canny and Robert is the best option since
the PSNR value is high and the MSE is low indicates the good
quality of image produced.
VOL. X, NO. X, XXXXXXXX ISSN 1819-6608
ARPN Journal of Engineering and Applied Sciences
©2006-2013 Asian Research Publishing Network (ARPN). All rights reserved.
Table 1: Mean MSE of different edge detection in the B-mode image popliteal vein (for 10 frames) for each subjectsa.
Table 2: Mean PSNR (dB) of different edge detection in the B-mode image popliteal vein (for 10 frames) for each subjectsa .
9 14 12.82
7.68 12.04
8 10.97
12 10.77
7 6.25 5.98 9.62
6 10
5.01
PSNR (dB)
5 4.49 8
MSE
4 6
3
4
2
1 2
0 0
Figure 8 : Average value of MSE of edge detector. Figure 9 : Average value of PSNR of edge detector.
VOL. X, NO. X, XXXXXXXX ISSN 1819-6608
ARPN Journal of Engineering and Applied Sciences
©2006-2013 Asian Research Publishing Network (ARPN). All rights reserved.
B. Vessel wall elasticity Based on this equation, the strain, F measurement is the
From the B-mode ultrasound image, the edge detector changes of the vessel wall displacement. Therefore, the
algorithm chosen was applied. The edge detection edge detection of the vessel wall which is for the wall
algorithms chosen was the combination of Canny and displacement measurement is crucial for this particular
Robert’s algorithm since it has the smallest MSE and purpose of the elasticity measurement. The elasticity of the
largest value of PSNR. This edge detection implementation blood vessel is an important parameter for the early
to the B-mode ultrasound image is important for the detection of the Deep Vein Thrombosis condition. A
measurement of the wall displacement of the popliteal vein number of subjects will be examined using the ultrasound
because of the requirement to measure the elasticity of the and the assessement of the vessel wall elasticity for the
blood vessel. Based on the Hooke’s Law anology, which subjects will be done. Thus, the measurement of the vessel
describes the relationship between stress and strain (Walter F. wall elasticity of the subjects will be compared to the
Boron, 2008): vessel wall elasticity of the subject with Deep Vein
Thrombosis condition.
=
&45677 ; BC?
E
Pa &4589: < @A@?? D
Based on Figure 10, shows an example of the
,
> vessel wall elasticity measurement for a subject with no
? ∆C
(7)
@?
history of Deep Vein Thrombosis with the implementation
where of the combination Canny and Robert’s algorithm. Based
on the result obtained, the average range of the strain and
E : Young's modulus (modulus of elasticity) elasticity measurement were 1.05% and 0.420MPa
respectively. In addition, for the future development of this
F : Force exerted on an object under tension
study, more subjects will be examined to differentiate the
A0: Original cross-sectional area through which the force difference of the vessel wall elasticity measurement for the
is applied subject with no history of Deep Vein Thrombosis and the
∆L: Amount by which the length of the object changes subject that has Deep Vein Thrombosis condition.
L0 : Original length of the object
Strain (%)
3.5 Vessel Wall Elasticity Elasticity (MPa)
3.5
Strain (%)
3 3
Elasticity
2.5 (MPa) 2.5
2 2
1.5 1.5
1 1
0.5 0.5
0 0
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
Time (s)
Figure 10 : Edge detection Canny with Roberts for vessel wall elasticity measurement.