Telebrineller System Brochure
Telebrineller System Brochure
TELEBRINELLER
SYSTEM
NowNow
You You Can Measure
Can Measure BrinellBrinell
Hardness
Hardness Anywhere
Anywhere
In the
In the FieldField • In
• In the the• Plant
Plant •
In the Lab
Simple, Certain Operation
In the Lab Under
All Conditions
Teleweld, Inc.
416 N. Park St.
Streator, IL 61364
Phone: (815)672-4561
1
Visit our website at www.teleweld.net
Table of Contents
Telebrineller Accuracy 9
Halteman Filletester 14
2
The Telebrineller System
3
System--Its Principle of Operation
The Telebrineller System
MEASURING DIAMETERS
The bar is removed from the instrument and the
Telebrineller Microscope positioned over the appropriate im-
pression. The microscope provides a scale of 1/10 millimeter
divisions and is designed so that the image and scale are bright
and clear to the observer.
4
2
1
7
5
5
5
TELEWELD, INC.
239--21 BRINELL MICROSCOPE
Model 239
QUICK and EASY MEASUREMENTS OF BRINELL IMPRESSIONS
Teleweld, Inc. offers a rugged, stable, free-standing portable
TELEWELD, INC instrument for reading brinell impression. The high quality
optics in the 10x focusable eyepiece and 2x objective provides
20x total magnification.
EYEPIECE
10x wide field (22mm F.O.V.), high eye point with 239--21
Model 239
foldable rubber eye guard.
OBJECTIVE
2x magnification-doublet housed in an aluminum
retainer aligned and calibrated.
RETICLE
Easy to read glass reticle with 0-5mm linear scale
in 0.1mm increments
HOUSING
All aluminum lightweight housing cushioned with
a shock absorbing textured grip
NOSEPIECE
Stainless steel, reversible nosepiece allows the unit
to be free-standing or flip and gain access to those
hard to reach areas.
LIGHTING
The Brinell Microscope may be used under normal
room or outdoor lighting conditions. Attached is a
battery-powered (2-AA) included, adjustable beam
light source. The L.E.D. light source pivots for
compact storage.
CALIBRATION
Factory calibrated for magnification and focus. To
focus to the individual users needs, simply rotate
the focusing eye piece. An optional calibration
standard is available.
ORDERING INFORMATION
Catalog Number Description
239-21 20x Brinell Microscope with focusable eye piece, reticle, and illuminator
Due to a policy of continued improvement, we reserve the right to change product specification without notice
6
OPERATING INSTRUCTIONS
7
Non--Steel Materials with the Telebrineller
Testing Non
8
Telebrineller Accuracy
9
TELEBRINELLER TEST BARS
10
Test Bar Availability
11
Test Bar Availability (Cont’d)
1
6
8
7 3
9 2
4 5
10
3 Test Bar
4 Impression Ball—Standard 239-14 $6.20
5 Impression Ball—Tungsten Carbide 239-29 $36.00
13
The Halteman Filletester
is designed for the
Brinnell Hardness testing
of fillet welds and other
hard to reach areas.
14
Telebrineller Maintenance Tips
1.Reparing the Bar Holder Assembly: All parts of the Bar Holder are replaceable and avail-
able from Teleweld, Inc. (See separate parts price sheet).
Disassembly and reassembly procedures are as follows:
A) For Front End Parts
1) Insert a small screw driver between the bottom of the Rubber Front Piece (3)
and the Bar Tube (8). Pry the rubber down to clear the retaining pins and pull the
rubber assembly from the tube.
2) Push the Impression Ball (5) out of its socket toward the inside of the Rubber
Front Piece.
3) Insert a small screw driver under the head of the Anvil (2), work it past the
rubber collar and under the base of the Anvil. Pry the Anvil up and out of the
Rubber Front Piece.
4) Replace the Anvil by inserting one end of its base into the Rubber Front Piece.
Insert a small screwdriver up through the hole in the Rubber Piece at the opposite
end of the Anvil so that the Anvil can be pried down into position.
5) Replace the Impression Ball by pushing it on the end of the Bar Tube until the
retaining pins seat in their sockets in the rubber Front Piece.
6) Reinstall the entire assembly by pushing it on the end of the Bar Tube until the
retaining pins seat in their sockets in the Rubber Front Piece.
2. Maintaining Accuracy: There are only two elements in the Telebrineller system whose
deterioration can cause inaccurate tests.
A) The Impression Ball: After many tests the Impression Ball (5) may become de
formed and cause out of round impressions. Check the ball by measuring it across
multiple axes with a micrometer. If variations of greater then .0005 inches (.01
mm) are detected, replace the ball.