0% found this document useful (0 votes)
80 views4 pages

Probe Lu Siemens

This document provides information about the SITRANS Probe LU, a 2-wire loop powered ultrasonic transmitter for level, volume, and flow monitoring of liquids in storage vessels and simple process vessels. Key details include an overview of features, benefits, applications, technical specifications including measuring range and accuracy, and dimensional drawings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
80 views4 pages

Probe Lu Siemens

This document provides information about the SITRANS Probe LU, a 2-wire loop powered ultrasonic transmitter for level, volume, and flow monitoring of liquids in storage vessels and simple process vessels. Key details include an overview of features, benefits, applications, technical specifications including measuring range and accuracy, and dimensional drawings.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

SITRANS L Level instruments

Continuous measurement - Level transmitters


SITRANS Probe LU

■ Overview ■ Technical specifications


Mode of operation
• Measuring principle Ultrasonic level measurement
• Typical application level measurement in storage
vessels and simple process ves-
sels
Inputs
Measuring range
• 6 m (20 ft) model 0.25 to 6 m (10” to 20 ft)
• 12 m (40 ft) model 0.25 to 12 m (10” to 40 ft)
Frequency 54 KHz
Outputs
mA
• range 4 to 20 mA
• accuracy ± 0.02 mA
• span proportional or inversely propor-
SITRANS Probe LU is a 2-wire loop powered ultrasonic transmit- tional
ter for level, volume, and flow monitoring of liquids in storage
vessels and simple process vessels, as well as in open chan-
nels.
Beam angle
Performance
10°
5
• Resolution ≤ 3 mm (0.12”)
■ Benefits • Accuracy ± the greater of 0.15% of range or
• Continuous level measurement up to 12 m (40 ft) range 6 mm (0.25”)

• Easy installation and simple startup • Repeatability ≤ 3 mm (0.12”)

• Programming using infrared intrinsically safe handheld pro- • Blanking distance 0.25 m (10")
grammer, SIMATIC PDM, or HART® Communicator • Update time at 4 mA ≤ 5 seconds
• Communication using HART • Temperature compensation built-in to compensate over tem-
• ETFE or PVDF transducers for chemical compatibility perature range
• Patented Sonic Intelligence signal processing Rated operating conditions
• Extremely high signal-to-noise ratio • Ambient conditions
• Auto False-Echo Suppression for fixed obstruction avoidance - Location Indoor/outdoor
- Ambient temperature -40 to 80 °C (-5 to 176 °F)
■ Application - Relative humidity/ingress protec- Suitable for outdoor
The SITRANS Probe LU is ideal for level monitoring in the water tion
and wastewater industry and chemical storage vessels. - Installation category I
The range of SITRANS Probe LU is 6 or 12 meters (20 or 40 feet). - Pollution degree 4
Using Auto False-Echo suppression for fixed obstruction avoid- • Medium conditions
ance, as well as an improved signal-to-noise ratio, and improved
accuracy of 0.15% of range or 6 mm (0.25”), the Probe LU pro- - Temperature at flange or threads -40 to 85 °C (-5 to 185 °F)
vides unmatched reliability. - Pressure (vessel) ambient, vented to atmosphere
SITRANS Probe LU includes Sonic Intelligence® signal process- Design
ing from the field-proven Milltronics Probe, and incorporates new • Material (enclosure) PBT (Polybutylene Terephthalate)
echo processing features and the latest micro-processor and
communications technology. • Degree of Protection Type 4X/NEMA 4X, Type 6/NEMA
6/IP67/IP68 enclosure
The transducer on the Probe LU is available as ETFE or PVDF to
• Weight 2.1 kg (4.6 lbs)
suit the chemical conditions of your application. As well, for ap-
plications with varying material and process temperatures, the • Cable inlet 2 x M20x1.5 conduit gland or 2 x
Probe LU incorporates an internal temperature sensor to com- ½” NPT thread
pensate for temperature changes. • Transducer (2 options) ETFE (Ethylene Tetrafluoroethyl-
ene) or PVDF (Polyvinylidene Flu-
oride)
• Process connection
- threaded connection 2” NPT, BSP, or G/PF
- flange connection 3” (80 mm) universal flange
- other connection FMS 200 mounting bracket or
customer supplied mount

Siemens FI 01 · 2005 5/45


SITRANS L Level instruments
Continuous measurement - Level transmitters
SITRANS Probe LU
Display and Controls
• Interface HART: standard, integral to ana-
log output
• Configuration Using Siemens SIMATIC PDM
(PC) or HART handheld commu-
nicator, or Siemens Milltronics
infrared handheld programmer
• Memory non-volatile EEPROM, no battery
required
Programmer (optional infrared key-
pad)
• approval ATEX II 1 G, EEx ia IIC T4
• ambient temperature -20 to 40 °C (-5 to 104 °F)
• interface proprietary infrared pulse signal
• power 3 V lithium battery (non-replace-
able)
Power supply
Standard nominal 24 V DC with 550 ohm
maximum; maximum 30 V DC

5
4 to 20 mA
Approvals
General CSAUS/C, FM, CE
Hazardous
• Europe ATEX II 1G EEx ia IIC T4
• USA FM Class 1, Div. 1, Groups A, B,
C, D (barrier required); Class II,
Div. 1, Groups E, F, G; Class III
• Canada CSA Class 1, Div. 1, Groups A, B,
C, D (barrier required); Class II,
Div. 1, Group G; Class III

5/46 Siemens FI 01 · 2005


SITRANS L Level instruments
Continuous measurement - Level transmitters
SITRANS Probe LU

■ Dimensional drawings

139.7 mm
(5.50”)

hinged lid
130.1 mm
(5.12”)

198.9 mm
(7.83”)
5
electronics

mounting thread

transducer
51.1 mm
(2.01”)

54 mm
(2.13”)

Note: Above model is shown without M20 conduit glands or ½” NPT conduit connectors.

SITRANS Probe LU dimensions

■ Schematics
Connect the
wires to the
terminals as
shown: the
lid screws polarity is
identified on
terminals for the terminal block.
loop current

cable

watertight gland

Notes:
- Model above is shown with M20 conduit glands. ½” NPT threaded connection is also available.
- DC terminal shall be supplied from an SELV source in accordance with IEC-1010-1 Annex H.
- All field wiring must have insulation suitable for rated input voltages.
- Use shielded twisted pair cables (wire gauge 14-22)
- Separate cables and conduit may be required to conform to standard instrumentation wiring practices or electrical codes.

SITRANS Probe LU connections

Siemens FI 01 · 2005 5/47


SITRANS L Level instruments
Continuous measurement - Level transmitters
SITRANS Probe LU

Ordering data Order No.


SITRANS Probe LU C) 7 M L 5 2 2 1 -
2-wire loop powered ultrasonic transmitter for level, 7 7 7 77
volume, and flow monitoring of liquids
Enclosure
Plastic (PBT), Qty 2 x M20x1.5 1
Plastic (PBT), Qty 2 x 1/2” NPT 2
Range / Sensor material
6 meter (20 ft), ETFE A
6 meter (20 ft), PVDF Copolymer B
12 meter (40 ft), ETFE C
12 meter (40 ft), PVDF Copolymer D
Process Connection
2” NPT A
2” BSP B
2” G / PF2 C
Communication / Output
4 to 20 mA, HART 1
Approvals

5 General purpose, FM, CSA, CE


Intrinsically Safe, FM Class 1, Div. 1, Groups A, B,
1
2
C, D (barrier required); Class II, Div. 1, Groups E, F,
G; Class III, ATEX II 1G EEx ia IIC T4
Intrinsically Safe, CSA Class 1, Div. 1, Groups A, B, 3
C, D (barrier required); Class II, Div. 1, Group G;
Class III
Instruction Manual
English C) 7ML1998-5HT01
German C) 7ML1998-5HT31
Note: The instruction manual should be ordered as
a separate item on the order.
Additional Quick start manual
Multi-language Quick start manual C) 7ML1998-5QR81
Note: Due to ATEX regulations, one Quick start
manual is included with every product.
Optional Equipment
Hand programmer, Intrinsically Safe, EEx ia 7ML5830-2AH
HART modem (for use with a PC and SIMATIC A) 7MF4997-1DA
PDM)
Siemens Intrinsically Safe Barrier (DC powered), 7NG4122-1AA10
ATEX II 1 G, EEx ia
2" NPT locknut, plastic 7ML1830-1DT
2" BSP locknut, plastic 7ML1830-1DQ
Universal Mounting Adapter, 2" NPT 7ML1830-1BT
Universal Mounting Adapter, 2" BSP 7ML1830-1BU
Universal Mounting Adapter, 2" G / PF2 7ML1830-1BV
A) Subject to export regulations AL: N, ECCN: EAR99H
C) Subject to export regulations AL: N, ECCN: EAR99